Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems Matthew L. Paden, MD Division of Pediatric Critical Care Emory University Children’s Healthcare of Atlanta at Egleston

Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

  • Upload
    ashanti

  • View
    183

  • Download
    0

Embed Size (px)

DESCRIPTION

Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems. Matthew L. Paden, MD Division of Pediatric Critical Care Emory University Children’s Healthcare of Atlanta at Egleston. The Clinical Problem. - PowerPoint PPT Presentation

Citation preview

Page 1: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

Matthew L. Paden, MDDivision of Pediatric Critical Care

Emory UniversityChildren’s Healthcare of Atlanta at Egleston

Page 2: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

2

The Clinical Problem

Multiple ECMO/CVVH patients who are 10+ liters positive on paper, yet are clinically dehydrated

Hypothesis : Both IV pump and Braun Diapact driven CVVH systems when used with ECMO have error rates greater than 5% / hour.

Page 3: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

3

Methods for CRRT on ECMO

Commercially available CRRT machines• B Braun Diapact• Edwards Aquarius• Fresenius• Gambro

Prisma/Prismaflex

Advantage – Standardized equipment, “built for the purpose”

Page 4: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

4

Methods for CRRT on ECMO

IV Pump Driven• Described by Weber, et al 1998• IV pumps used to create ultrafiltrate and

deliver replacement fluid• Ultrafiltrate is measured using a

urometer Advantage – low cost, no additional training

for ECMO specialists, less extracorporeal blood volume

Page 5: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

5

Page 6: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

6

Page 7: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

7

The Problem

Inaccuracy has been described in CRRT in patients not on ECMO

The pumps are the problem• PCRRT 2000 –

7% extra ultrafiltrate removed 2% less replacement fluid delivered

• ASAIO 1992 – Up to 12.5% error rate Error correlates with pressure drop

across membrane and pump type

Page 8: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

8

Alaris Pump Accuracy

Sucosky et al. Awaiting publication. 2008

Page 9: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

9

Methods

Two identical saline primed ECMO circuits• Stockert S3, ½ inch drain, 3/8 inch return• 4.5 m2 Medtronic oxygenator• PAN 6 hemofilter

CVVH prescribed : IV pump vs. Diapact• Net even fluid balance• Varying ultrafiltration rates (0.5 – 2 L /

hour) Hourly weights of UF/Replacement bags,

circuit pressures

Page 10: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

10

Results

48 hourly measurements• 26 hours Alaris• 22 hours Diapact

No correlation between error rate and• Prescribed ultrafiltration rate• ECMO flow rate• CVVH blood flow rate

Page 11: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

11

Results - Alaris

Created a median 0.8% less ultrafiltrate per hour than prescribed (+7% to -12%)

Delivered a median of 4.3% less replacement fluid per hour than prescribed (+3% to -25%)

NET 3.5% of prescribed UF per hour dehydration to patient

Page 12: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

12

Examples with Alaris

4 kg neonate on ECMO with 100 ml/hour prescribed UF rate and even fluid balance• 84 ml (21 ml/kg) fluid negative per day

10 kg child on ECMO with 300 ml/hour prescribed UF rate and even fluid balance• 252 ml (25 ml/kg) fluid negative per day

45 kg child on ECMO with 2000 ml/hour prescribed UF rate and even fluid balance• 1.68 L (37 ml/kg) fluid negative per day

Page 13: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

13

Results - Diapact

Created a median 1% more ultrafiltrate per hour than prescribed (+6% to -8%)

Delivered a median of 1% more replacement fluid per hour than prescribed (+10% to -7%)

Page 14: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

14

Examples Braun

Accurate when looking at medians over time

Hourly variation could be important in a hemodynamically unstable patient

Hourly Fluid Balance

-5

35

10

10

35

-40

-5

0

-10

-30

Page 15: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems

15

Conclusions

In this in vitro CVVH/ECMO model• Both the Alaris and Diapact methods had

error rates that could be clinically significant

• Careful physical assessment of the patient’s volume status is necessary

Further work is needed to develop more a accurate fluid management system for CRRT on ECMO