8/14/2019 AccuPoint
1/2
78 Sales & Service: 800/234-5333 (USA/Canada) or 517/372-9200 Fax: 517/372-2006 www.neogen.com
Food Safety Online Orderingwww.neogen.com/foodsafetyorder.htmSanitation
AccuPoint
Sanitation Monitoring System
Intended Use
Neogens AccuPoint ATP Sanitation Monitoring System is a lightweight,handheld system that validates the effectiveness of a sanitation programby detecting food residues and microorganisms present on surfaces and inliquids. The system utilizes easy-to-use AccuPoint samplers, which auto-matically mix prepared reagents to quickly generate results.
The Test
The AccuPoint system utilizes adenosine triphosphate (ATP) biolumi-nescence to determine the cleanliness of test samples. ATP is a chemicalcompound found in all living cells, including bacteria, food debris, yeastand mold. Bioluminescence is a chemical reaction that produces light.
ATP bioluminescence occurs when ATP from a sample comes intocontact with luciferase, an enzyme found in fireflies, and luciferin, asubstrate. The amount of light emitted in this reaction is proportional
to the amount of ATP detected in a sample.
After a sample is taken the sampler is pressed into its cartridge in themonitor, breaking its seals and initiating the mixing of reagents. Thecartridge contains a buffering solution, which flushes the ATP ontoa pad containing the reagents. The reaction takes place within thecartridge and a detector measures the amount of light produced. Thereading is displayed on the LCD display in relative light units (RLU).
According to preset limits, an icon is displayed indicating a pass, mar-ginal or fail result. These limits are defined by the operator or by usingthe system presets.
Consistently Accurate
The samplers vertical sampling technique increases downwardpressure on the surface. This helps penetrate surface crevasses, andbreak-up existing biofilms to uncover previously protected organisms.
The sponge will reabsorb the liquid, with any present organic matteror bacteria, for consistently accurate results.
Sampler extender (optional)
Sampler
Sample Pad
Cartridge
Buffer
Luciferaseand
Luciferin
Three samplers engineered for precise sampling in each
unique applicationflat surfaces, water samples and
tight spaces such as small orifices and channels
http://www.neogen.com/foodsafetyorder.htmhttp://www.neogen.com/foodsafetyorder.htmhttp://www.neogen.com/foodsafetyorder.htm8/14/2019 AccuPoint
2/2
Sales & Service: 800/234-5333 (USA/Canada) or 517/372-9200 Fax: 517/372-2006 www.neogen.com
Food Safety Online Orderiwww.neogen.com/foodsafetyorderSanitation
Results
AccuPoint
Sanitation Monitoring System
Ordering Information
Prod.# Product description
9600 AccuPoint Sanitation Monitoring System - Includes monitor
NiMH battery pack, Docking station with power supply data
cable, DataManager Software and Users Guide
9605 AccuPoint ATP Surface Samplers (100)
9606 AccuPoint Water Samplers (100)
9607 AccuPoint Access Samplers (100)
9610 NiMH Battery Replacement pack
9611 Sampler extender
9612 AccuPoint Cleaning Kit (5 swabs)
9613 AccuPoint Stand-up Holster
9615 AccuPoint Docking Station
9616 AccuPoint Communications Cable
9617 AccuPoint DataManager Software
9618 Hawking Technologies USB Serial Converter
9619 ATP Standards Kit
2. Insert sampler into cartridge to first stop
3. Insert cartridge into device and
fully depress sampler. Close unit
door and read results.
1. Sample surface or water
The Procedure
AccuPoint comes with Data Managersoftware that allows customers to:upload results to a PC, set pass/caution/fail levels, create and print
trend analysis graphs, export data tospreadsheets, and keep a record of sanitation effectiveness to veSSOPs. Results can be presented in a clear chart format showingactual measurement in color coded green (pass), yellow (cautand red (fail) for extra clarity. Accumulated results can be grapusing bar or pie charts, over a variety of criteria, for clear pictof sanitation testing results and trends.
After the countdown, the data reading is displayed. Easy to intericons indicate a pass, marginal or fail result. Results can be recoror using AccuPoints docking station, simply uploaded to a PC
easy data analysis.
Use the Access sampler
for tight spaces