A Study and Optimization of Lumbar Spine X-Ray Imaging

Embed Size (px)

Citation preview

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    1/12

    A study and optimization of lumbar spine X-ray imaging

    systems

    1G MCVEY, D.Phil,

    2M SANDBORG, PhD,

    1D R DANCE, PhD, FIPEM and

    2G ALM CARLSSON, PhD, FInstP

    1Joint Department of Physics, The Royal Marsden NHS Trust, Fulham Road, London SW3 6JJ, UK and 2Department of

    Radiation Physics, Faculty of Health Sciences, Linkoping University, SE581 85 Linkoping, Sweden

    Abstract. A Monte Carlo program has been developed that incorporates a voxel phantom of an adult patient in

    a model of the complete X-ray imaging system, including the anti-scatter grid and screenfilm receptor. This

    allows the realistic estimation of patient dose and the corresponding image (optical density map) for a wide

    range of equipment configurations. This paper focuses on the application of the program to lumbar spine

    anteroposterior and lateral screenfilm examinations. The program has been applied to study the variation of

    physical image quality measures and effective dose for changing system parameters such as tube voltage, grid

    design and screenfilm system speed. These variations form the basis for optimization of these system

    parameters. In our approach to optimization, the best systems are those that can match (or come close to) the

    calculated image quality measure of systems preferred in a recent European clinical trial, but with lower patient

    dose. The largest dose savings found were 21% for a 400 speed class system with a grid having a strip density of40 cm21 and a grid ratio of 16. A further dose saving of 13% was possible when a 600 speed class system was

    employed. The best systems found from the optimization correspond to those recommended by the European

    Commission guidelines on image quality criteria for diagnostic radiographic images.

    Lumbar spine radiographs allow clinicians to judge the

    configuration and alignment of bones with a high degree

    of accuracy. Malalignment or other changes in the shape

    of the vertebrae can then be identified and may imply the

    presence of a tumour, fracture or infection. Lumbar spine

    radiography is a routine examination for lower back pain,

    which is very common; 27 patients per 1000 inhabitants in

    the UK undergo plain radiography of the lumbar spineeach year [1]. These examinations contribute 4.3% of the

    annual collective effective dose for all medical and dental

    exposures compared with 0.9% for chest examinations in

    the UK [2].Optimization is necessary to balance the requirement for

    good image quality with low patient dose. The Commis-

    sion of the European Communities (CEC) image quality

    criteria [3] describe the presentation of the normal

    anatomy in a lumbar spine radiograph. Almen et al [4]

    have evaluated the image quality of lumbar spine radio-

    graphs using the CEC criteria [3]. These studies showed

    that systems using a low tube voltage (70 kV) and a

    medium speed class (400) for the screenfilm receptorfulfilled more of the image criteria for the anteroposterior

    (AP) projection than those using high tube voltage and

    high speed class. The systems using high speed class (600)

    and low tube voltage (77 kV) fulfilled more criteria for the

    lateral (LAT) projection than those using low speed classand high tube voltage. Vano et al [5] optimized lumbarspine imaging by varying different technical parametersand found the largest dose saving by decreasing the optical

    density by changing the settings of the automatic exposurecontrol (AEC). Almen et al did not study the effect ofoptical density as they did not use AEC.

    The assessment by Almen et al [4] of clinical imagequality has been complemented by theoretical modelling aspart of the same project. A realistic Monte Carlo model ofthe patient (voxel phantom) and the complete imagingsystem has been developed [6] for this purpose. The modelcan be used to calculate physical measures of imagequality and patient dose. In Sandborg et al [7], thecorrelations between our calculated physical measures ofimage quality and the clinical assessments of image qualityare presented for chest and lumbar spine radiographs.

    For the latter, the signal-to-noise ratio (SNR) of trabe-cular structures was found to be a good predictor ofclinical image quality. This paper presents the application

    of the Monte Carlo program to the study and optimization

    of lumbar spine imaging. The optimization approach issimilar to that used, with the same model, for chestradiography [8] and involves the use of a reference systemknown to be of good image quality. Preliminary results forthis study are outlined in Dance et al [9]. In this paper, ourpreliminary study is considerably extended so that theinfluence of tube voltage, grid design, screenfilm speedand operating optical density are all considered for bothAP and LAT projections.

    Methods and materials

    Monte Carlo model and voxel phantom

    A Monte Carlo computer program has been developedto simulate diagnostic X-ray examinations. It is based on

    Received 8 April 2002 and in revised form 30 August 2002, accepted 21October 2002.

    Current address for G McVey: North Wales Medical Physics, Glan

    Clwyd Hospital, Bodelwyddan, Denbighshire LL18 5UJ, UK.

    This work was supported by grants from the Commission ofEuropean Communities (Nos. FI4P CT950005 and FIGM-CT2000-00036). The Swedish authors were supported by grants from the

    Swedish Radiation Protection Institute, SSI (Nos. SSI P1018.97 andSSI P1083.98), and the Swedish Foundation for Strategic Research(No. R98:006).

    The British Journal of Radiology, 76 (2003), 177188 E 2003 The British Institute of Radiology

    DOI: 10.1259/bjr/52734084

    177The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    2/12

    our earlier work, which employed a homogeneousphantom in a model of the complete imaging system[10, 11]. For the present model, the program has beenextended by the inclusion of a voxel phantom to providea more accurate model of the patient. The programtransports photons through the patient and the anti-scattergrid to the imaging device. The energy imparted to the

    voxel phantom allows the patient dose to be calculatedand the energy imparted to the screen allows the imagequality measures to be calculated. These parameters are

    discussed in more detail below.The voxel phantom used is that developed by Zubal et al

    [12, 13] and was obtained by segmentation of a series ofCT slices of an adult male. Female organs (breasts, uterusand ovaries) were added by us to facilitate the calcula-tion of effective dose [14]. An extra layer of voxels wasincluded in the phantom to model the couch top. Forthe AP view, the dimensions of the voxel phantomwere 899 mm from the top of the head to the bottomof the pelvis (236 voxels), 356 mm wide (128 voxels)and 214 mm thick (77 voxels). The phantoms length was

    adjusted to correspond to the sitting height of theaverage European male [15]. The phantoms width and

    thickness were found by the comparison of calculations ofentrance air kerma for the voxel phantom with measure-ments from a patient study [6, 16]. Each voxel in the

    phantom belongs to 1 of 55 organs and each organ isassociated with one of four tissue types: average softtissue (1030 kg m23); lung (260 kg m23); average bone(1490 kg m23); or bone spongiosa (1180 kg m23). Tissuecompositions were obtained from the InternationalCommission on Radiation Units and Measurements(ICRU) [17], except for average bone, which was takenfrom Kramer [18].

    Figure 1 shows the model of the voxel phantom and thecomponents of the imaging system. The photon spectrumwas obtained from Birch et al [19]. A grid was used as theanti-scatter technique and was specified in terms of stripdensity N, lead strip width d, grid ratio r and the materials

    in the interspaces and covers. The model of the imagereceptor included the cassette front, fluorescent screen andfilm characteristic (H and D) curve, measured by Dr FVerdun, Lausanne (personal communication). The Monte

    Carlo code calculates the contrast and SNR of anatomicaldetails at different positions in the image to provide aphysical measure of image quality. These parameterswere calculated with a large number of photon historiesso that the uncertainty of their values is less than 3%

    (1 standard deviation).

    Important contrast details

    The important contrast details used for the calculationof image quality were carefully selected to correspond tothe diagnostic requirements described in the CEC imagequality criteria [3] and following discussions with localradiologists in London and Linkoping. Lumbar spineX-ray images help the clinician appraise the presentationof the lumbar spine vertebrae and thus, all the detailschosen represent bony anatomy. For modelling the APprojection, the L1, L3 and L5 transverse processes were

    selected as low contrast details with thicknesses of2.0 mm (L1T), 3.5 mm (L3T) and 5.0 mm (L5T), respec-tively. For modelling the LAT projection, the L1, L3 and

    L5 spinous processes were selected as low contrast

    details with thicknesses of 5.0 mm (L1S), 5.5 mm (L3S)

    and 6.0 mm (L5S), respectively. The thicknesses were

    obtained from measurements on a skeleton. All bony

    processes were simulated as cortical bone (1920 kg m23)

    and their contrast was calculated against a background of

    soft tissue.

    Small high contrast details were also chosen. These werethe trabecular structures on the L1, L3 and L5 vertebrae

    in the AP projection, referred to subsequently as L1D,

    L3D and L5D, respectively. For the LAT projection, the

    trabecular structures were selected to be at an anterior

    position on the L1 and L5 vertebrae and at a posterior

    position on the L5 vertebra, referred to subsequently as

    L1F, L5F and L5B, respectively). All of the trabecular

    structures were 1 mm thick. This is quite similar to the

    important detail size of 0.3 mm to 0.5 mm given in the

    CEC image quality criteria document [3]. Trabecular

    structures were simulated as bone marrow cavities

    (1030 kg m23) and their contrast was calculated against

    a background of cortical bone. The compositions of theanatomical details and tissue backgrounds were taken

    from the ICRU [17].

    Figure 1. The imaging system included in the Monte Carlomodel of the lumbar spine anterior-posterior projection. The

    bony structures in the voxel phantom have been highlighted.

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    178 The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    3/12

    Image quality and patient dose parameters

    ContrastContrast was calculated in the Monte Carlo program as

    the difference in optical density (DOD) beside and behind theimportant details superimposed on the voxel phantom. The

    effects of film gradient and imaging system unsharpness weretaken into account in the calculation ofDOD using:

    DOD~ log10 (e)|c(ODdet)|Ce|cdfMTF 1

    The H and D curve was measured in accordance with theISO-standard [20] by Dr F Verdun, Lausanne (PrivateCommunication, 1998). The film gradient (c) was derivedfrom the H and D curve for the OD beside the detail (ODdet).The quantitycdfMTF is thereduction in contrast causedby thetotal system unsharpness (total modulation transfer func-tion, MTFtot). Image receptor (screenfilm) and geometricunsharpness (focal spot size and magnification) are all takeninto account in the calculation of the MTFtot. The MTFs

    of the screenfilm combination were also measured by Dr FVerdun, Lausanne (personal communication). Sandborg

    et al [8] describes the calculation of cdfMTF.The object contrast C

    ewas found from Monte Carlo

    calculations of energy imparted to the fluorescent screenper unit area:

    C~E(p1){E(p2)

    E(p1)|

    1

    1z E(s)=E(p1) 2

    Here, ep1 and ep2 are the energy imparted to the receptorper unit area by primary photons beside and behind thedetail, respectively, and es is the energy imparted to thereceptor per unit area by scattered photons. The notationE denotes the expectation value. It was assumed that thedetail does not alter the distribution of scattered photons

    in the imaging plane.

    Signal-to-noise ratioThe SNR of the ideal observer, SNRI [21], of a small

    detail at an optical density ODdet was obtained using:

    SNRI(ODdet)~SNRMC(edet)|

    ffiffiffiffiffiffiffiffiffiffiA

    AMC

    s|

    SNRDF(ODdet)

    3

    The SNRMC (edet) was calculated by the Monte Carloprogram. It was obtained from the energy imparted to anarea of the detector AMC with and without the detail being

    present assuming that the only noise source is quantummottle and neglecting image unsharpness. The SNR2I wasscaled from the area of the detector element AMC50.25 mm2 to the area of the detail A. The SNRMC due to

    quantum noise has been shown to give good agreementwith experiments [22, 23] for details with diameters largerthan or equal to 3 mm. Hence, as many of the details usedin our Monte Carlo model were similar to or smaller than3 mm, the model needs to take into account the effectof image unsharpness. This was implemented using theSNR degradation factor SNRDF, which also accounts for:(i) the different efficiencies with which the signal andquantum noise are transferred through the screen caused

    by light emitted from different depths in the screen [24]; (ii)the statistical variations in the transport of light from thescreen to the film [25]; and (iii) the total system noise

    including that from the film. These corrections are derivedfollowing the methods of Nishikawa and Yaffe [26]. Amore detailed description of the implementation is givenby Sandborg et al [8].

    Calculation of entrance air kermaThe Monte Carlo program calculates air kerma, without

    backscatter, at the entrance surface of the phantom, airkerma at the surface of the cassette front and energyimparted to the screen per unit area. The entrance airkerma for a fixed OD can be calculated using thesequantities combined with the H and D curve measured interms of the cassette entrance air kerma. The calculationwas implemented in two parts.

    In the first part, the experimental set-up used to measurethe H and D curve was simulated and the air kerma at thesurface of the cassette front and the energy imparted to thescreen per unit area calculated. In this way, the H and Dcurve was expressed as the OD for a given value of theenergy imparted to the screen per unit area.

    In the second part, the voxel phantom in the lumbarspine imaging system under investigation was simulated.Ratios of energy imparted to the screen per unit area tothe incident air kerma at the phantom were calculated forapproximately 200 evenly spaced points of interest acrossthe whole image and the median ratio found. The cali-brated H and D curve was used to convert an OD to be

    used as a normalization point, for example, the medianOD of a radiograph or set of radiographs, to an energyimparted per unit area. The entrance air kerma was then

    calculated by this value of the energy imparted divided bythe median value of the ratio.

    Effective doseEffective dose has been used in this work to quantify the

    radiation risk. The voxel phantom was segmented intoorgans each with known mass. The Monte Carlo codecalculated the energy imparted to each voxel associatedwith an organ. The organ dose was obtained by dividingthe sum of the energy imparted to all voxels of an organwith the mass of that organ. The effective dose was thenfound by combining the organ doses with the tissueweighting factors according to the InternationalCommission on Radiological Protection [14]. The MonteCarlo code calculates the ratio of the effective dose tothe incident air kerma at the voxel phantom surface. Theeffective dose for a given situation was found from the

    product of this ratio and the incident air kerma (seeprevious section).

    Validation of the modelThe Monte Carlo program has been validated in two

    parts. Firstly, Monte Carlo calculations of OD behindpolymethyl methacrylate (PMMA) phantoms were com-pared with measurements carried out under carefullycontrolled conditions. Good agreement, within 13% wasfound providing that there was detailed knowledge of theimaging system [6, 16]. Secondly, patient images werecollected and the entrance air kerma measured for chestand lumbar spine examinations in both frontal and lateral

    projections. The images were digitized and analysed. Mea-surements of contrast were extracted from the digitizedradiographs. For the lumbar spine AP projection, it was

    Optimization of lumbar spine X-ray imaging systems

    179The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    4/12

    found that the calculated entrance air kerma was slightlylower than the minimum value in the range of measuredentrance air kermas. This was due to the voxel phantombeing slightly thinner than required. However, as thecalculated entrance air kermas were within the range ofmeasured values for the other projections, it was decidednot to increase the thickness of the voxel phantom for the

    lumbar spine AP projection as the calculated value wasstill reasonably representative of the range of calculatedvalues. The program was also successfully validated by

    comparing the calculated contrast of important ana-tomical details and the calculated dynamic range of theimage with the range of measured values [6, 16]. The voxelphantom was thus found to be sufficiently representativeof a patient undergoing both chest and lumbar spine X-rayexaminations.

    Reference system and optimizationIn order to optimize the parameters used in X-ray

    imaging systems, one system had to be identified that

    provided good image quality, and this was designated asthe reference system. Thus, we determined a suitablereference system to be the imaging system that producedimages with the highest image quality as judged by anexpert panel of European radiologists in a recent clinicaltrial [4]. These preferred images were thus the referenceimages. Table 1 shows the characteristics of the reference

    imaging systems for the AP and LAT views. The refer-ence systems used 72 kV with a 400 speed class screenfilmsystem for the AP view and 77 kV with a 600 speed class

    screenfilm system for the LAT view. The preferred systemfrom the clinical trial corresponds to the good radio-graphic practice outlined in the CEC image criteria docu-

    ment [3], except that a lower tube voltage was used thansuggested by the guidelines.In our theoretical study, we have investigated what

    happens to the image quality and the patient dose if theimaging parameters are varied from their reference values.The range of the parameters studied is also given inTable 1. This study allows a greater understanding of theoptimization results.

    A good quality image may be one that fulfils itsdiagnostic purpose but may not always be an image withthe highest possible contrast or SNR [27]. In ouroptimization scheme, it was decided to use the bestsystems from the clinical trials as the reference systems andthe images they produce as the reference images. It was

    assumed that an image for which the contrast or SNR

    were 10% lower than those in the reference image wouldstill fulfil its diagnostic purpose. Values of SNRI and DODwere calculated for each detail for a specified scatter-

    rejection technique, speed class, OD and film type for tubevoltages between 60 kV and 110 kV in the AP view andbetween 70 kV and 110 kV in the LAT view. The tubevoltages required to give 0.9 of the appropriate SNRI andDOD value for each detail were then deduced. The detailrequiring the lowest tube voltage is referred to as thelimiting detail. This tube voltage is the highest employablethat ensures all details fulfil the criterion of the associatedimage quality measure being greater than or equal to 0.9of that for the reference system. The effective dose iscalculated for this limiting tube voltage and compared

    with the values for the reference system. The procedure isthen repeated for different imaging systems and the system

    resulting in the lowest effective dose is the optimum.

    Results

    Effect of varying image system parameters on patient

    dose and image quality

    Tube voltageFigure 2 shows the results for the AP projection of

    varying the tube voltage between 60 kV and 110 kV on (a)

    the effective dose, (b) the contrast of the L5 transverseprocess and (c) the SNR of a trabecular structure on theL1 vertebra. The reference system gives an incident airkerma without backscatter of 0.88 mGy and an effectivedose of 0.12 mSv. The calculated incident air kerma iswithin the range of entrance surface doses given in Hartet al [28]. The calculated effective dose is lower than wouldbe expected for example, from the effective doses given in

    Table 1. The parameters for the anteroposterior (AP) and lateral reference imaging systems. The range of imaging system parametersis also given

    AP imaging systems Lateral imaging systems

    Parameters Range Reference system Range Reference system

    Tube voltage 60110 kV 72 kV 70110 kV 77 kV

    Filtration 4.7 mmAl 4.7 mmAl 4.7 mmAl 4.7 mmAl

    Focal spot size 0.9 mm 0.9 mm 0.9 mm 0.9 mm

    Focusfilm distance 1.46 m 1.46 m 1.46 m 1.46 m

    Grid ratio 816 10 816 10

    Strip density 4070 cm21 52 cm21 4070 cm21 52 cm21

    Strip width 2040 mm 36 mm 2040 mm 36 mm

    Cover material Aluminium and carbon fibre Aluminium Aluminium and carbon fibre Aluminium

    Interspace material Aluminium and carbon fibre Carbon fibre Aluminium and carbon fibre Carbon fibre

    Speed class 320, 400 and 600 400 320, 400 and 600 600

    Screen material Gd2O2S Gd2O2S Gd2O2S Gd2O2S

    Screen types Lanex Medium/Regular, Lanex

    Regular Plus and Lanex Fast

    Lanex Regular Plus Lanex Medium/Regular, Lanex

    Regular Plus and Lanex Fast

    Lanex Fast

    Film type aKodak TML aKodak TML aKodak TML aKodak TML

    Median OD 0.23.0 1.36 0.23.0 1.36

    OD, optical density. aEastman Kodak Campany, Rochester, NY.

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    180 The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    5/12

    Hughes [29], owing to the voxel phantom thickness being

    slightly thinner than is required. However, this will notaffect the results as they are quoted relative to thereference system values in this paper. The effective dosedecreases by 73% between 60 kV and 110 kV. The threetransverse processes show approximately the same varia-tion of contrast with tube voltage. The same applies to theSNR for the three trabecular structures. For example, thecontrast of the L5 transverse process decreases by 54%between 60 kV and 110 kV, with a similar decrease inSNR of 58% for the L5 trabecular structure.

    Figure 2 also shows the results for the LAT projectionof varying the tube voltage between 70 kV and 110 kV on

    (d) the effective dose, (e) the contrast of the L3 spinousprocess and (f) the SNR of a trabecular structure on thefront of the L1 vertebra. The reference system gives anincident air kerma of 2.57 mGy and an effective dose of0.14 mSv. Again, the calculated incident air kermacompares well with Hart et al [28] and the calculated

    effective dose is lower than expected [29] due to thethickness of the voxel phantom. The effective dosedecreases by 59% between 70 kV and 110 kV, which is a

    smaller decrease than for the AP view owing to the smallervoltage range. The SNR and contrast show a similarvariation with tube voltage. There is a 47% decrease inthe SNR of a trabecular structure on the L5 vertebra

    and a 43% decrease in the contrast of a L5 spinous pro-cess between 70 kV and 110 kV. The variation of thecontrast and SNR is less for the LAT projection than

    the AP projection as a smaller range of tube voltages

    was studied.

    Grid designFigure 3 shows the results for the AP projection of

    increasing the grid ratio (r5816) for three grids: (1) stripdensity N540 cm21, strip width d540 mm, aluminiumcovers and interspaces; (2) the same parameters exceptwith carbon fibre covers and interspaces; and (3)N570 cm21, d520 mm, carbon fibre covers and inter-spaces. The figure shows the variation of (a) effective dose,(b) contrast of the L3 transverse process and (c) the SNR

    of the trabecular structure on the L3 vertebra. The resultsare shown relative to the reference system, which has agrid constructed with N552 cm21, r510, d536 mm withaluminium covers and carbon fibre interspaces. The effec-

    tive dose increases for increasing grid ratio for all grids,for example, increasing by 34% for the aluminium gridbetween r58 and r516. The carbon fibre grids give thelowest effective dose. For the carbon fibre grid withN540 cm21, the effective dose is lower by 11% (r58)compared with the mixed material grid, and lower by 13%(r58) compared with the aluminium grid. There is afurther dose reduction by increasing the strip density anddecreasing the strip width. The effective dose for the

    N570 cm21

    , d520 mm grid is lower by 19% (r58) than forthe N540 cm21 carbon fibre grid.

    There is a contrast and SNR advantage to using the

    Figure 2. The effect of tube voltage on (a) effective dose, (b) optical density (OD) difference of the L5 transverse process and (c)signal-to-noise ratio (SNR) of a trabecular structure on the L1 vertebra for the anteroposterior projection. The effect of tube voltage

    on (d) effective dose, (e) OD difference of the L3 spinous process and (f) SNR of a trabecular structure on the front of the L1

    vertebra for the lateral projection.

    Optimization of lumbar spine X-ray imaging systems

    181The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    6/12

    carbon fibre grids (r.8) rather than the mixed material oraluminium grids. There is a 13% increase in the L3

    transverse process contrast and a 14% increase in the SNR

    of the L3 trabecular structure for a grid with N540 cm21,

    r516. The contrast and SNR advantage is less for

    increasing strip density and decreasing strip width. There

    is only a 5% increase in the L3 transverse process contrast

    and a 4% increase in the L3 trabecular structure SNR for

    a grid with N570 cm21, d520 mm, r516. The loss of

    contrast and SNR for reducing the lead strip width is only

    slightly compensated for by increasing the strip density.Figure 3 also shows the results for the LAT projection

    for increasing grid ratio for the three grids mentioned

    above. The results show the variation of (d) effective dose,(e) the contrast of the L3 spinous process and (f) the SNR

    of the trabecular structure at an anterior position on the

    L5 vertebra. The dose reductions obtained with a carbon

    fibre grid are less for the LAT view than the AP view

    owing to the higher tube voltage. The effective dose for

    the N540 cm21 grid (r58) is 6% less than the reference

    system. By increasing the strip density and decreasing the

    strip width the dose is decreased by a further 14%.The contrast and SNR advantages from using carbon

    fibre grids in the LAT view are generally the same or

    smaller than for the AP view. The contrasts obtained using

    the N540 cm21, r516 and N570 cm21, d520 mm, r516

    grids are 8% and 5% greater than for the reference system.The corresponding increases in SNR for these grids

    compared with the reference system are 14% and 3%,

    respectively. For grids with low grid ratios where there isless contrast or SNR than the reference system, the tubevoltage does not need to be decreased significantly,especially if carbon fibre grids are used since the loss ofcontrast and SNR is small. For carbon fibre grids withhigh grid ratios, the tube voltage may be increased without

    losing contrast or SNR and therefore, such a system mayhave a significantly reduced dose.

    Screenfilm speedFigure 4 shows the results for the AP projection of

    varying the speed class between 320 and 600 (all usingTML film) on (a) the effective dose, (b) the contrast of the

    L3 transverse process and (c) the SNR of the trabecularstructure on the L3 vertebra. The results are shown at both72 kV and 90 kV. The effective dose decreases by 42% asthe speed class increases from 320 to 600 for both 72 kV

    and 90 kV X-rays. At 72 kV, the contrast of the L3transverse process is near its maximum value for the 400speed class system. The contrast decreases by 10% and 3%when the 400 speed class system is replaced by a 320 and600 speed class systems, respectively. This is due todifferences in the shape of the H and D curves for thedifferent screenfilm combinations. At 72 kV, the SNR ofthe trabecular structure varies by a greater amount thanthe contrast. The SNR decreases by 19% for increasing the

    speed class from 320 to 600. Similar variations of contrastand SNR are observed at 90 kV. If a 600 speed classsystem is used instead of a 400 speed class system, the tube

    Figure 3. The effect of two grids with strip density N540 cm21 with aluminium and carbon fibre covers and interspaces and a thirdgrid with a strip density N570 cm21, strip width d520 mm with carbon fibre covers and interspaces on (a) effective dose, (b) optical

    density (OD) difference of the L3 transverse process and (c) signal-to-noise ratio (SNR) of the trabecular structure on the L3 verte-

    bra for the anteroposterior projection. The effect of the same grids on (d) effective dose, (e) OD difference of the L3 spinous process

    and (f) SNR of the trabecular structure at an anterior position on the L5 vertebra for the lateral projection.

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    182 The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    7/12

    voltage has to be decreased slightly to regain the loss of

    contrast but significantly more to regain the loss of SNR.However, large dose reductions are still possible due tothe greater sensitivity of the system, despite significantlylower tube voltages being required to maintain contrastand SNR.

    Figure 4 also shows the results for the LAT projectionof varying the speed class between 320 and 600 on (d)effective dose, (e) contrast of the L3 spinous process and(f) the SNR of the trabecular structure at an anteriorposition on the L5 vertebra. The effective dose decreases

    by 43% with increasing speed class from 320 to 600 forboth 77 kV and 95 kV. At 77 kV, the contrast of the L3spinous process is lower by 10% and the SNR of the

    trabecular structure is higher by 23% for the 320 speed

    class system. There are similar results at 95 kV.

    Optical densityFigure 5 shows the results for the AP projection of

    varying the value of the median OD between 0.2 and 3.0using the Lanex Regular (Eastman Kodak Campany,Rochester, NY) screen with TML film on (a) effectivedose, (b) the contrast of the L1, L3 and L5 transverseprocesses and (c) the SNR of the trabecular structures onthe L1, L3 and L5 vertebrae. The effective dose increaseslinearly with OD between 0.4 and 1.6. There is a rapidincrease in effective dose above an OD of 1.6 due tothe shape of the TML H and D curve. The effec-

    tive dose is 22% greater at a median OD of 1.6 than atthe median OD of 1.36 used in the reference system. Thetransverse processes have a maximum contrast at different

    median ODs due to the differing OD beside each

    anatomical detail and, therefore, their position onthe H and D curve. The L1, L3 and L5 transverse

    processes have maximum contrasts at ODs of 1.6, 1.4

    and 1.2, respectively. The contrast of the L3 process

    at an OD of 1.6 is very similar to that at 1.36. The

    trabecular structures also have a maximum SNR atdifferent median ODs. The details on the L1, L3 and

    L5 vertebrae have maximum SNRs at ODs of 2.6, 2.4

    and 2.2, respectively. The maximum SNR values are

    47%, 33% and 20% greater than the SNR values for the

    L1, L3 and L5 trabecular structures using the reference

    system.Figure 5 also shows the results for the LAT projection

    of varying the median OD between 0.4 and 3.0 on (d)

    effective dose, (e) the contrast of the L1, L3 and L5spinous processes and (f) the SNR of the trabecular

    structures on the L1 and L5 vertebrae. The effective dose

    shows the same variation as for the AP projection with a

    23% increase at a median OD of 1.6 compared with the

    effective dose at a median OD of 1.36. The maximum

    contrast values occur at an OD of 1.0 for the L1 and L3processes and at an OD of 1.8 for the L5 process. These

    maximum contrasts are at most 8% greater than the

    contrast of the details obtained with the reference system.

    The maximum SNR values occur at an OD of 2.0 for the

    details on the anterior position of the L1 and L5 vertebra

    and at an OD of 2.6 for the detail on the posterior posi-

    tion on the L5 vertebra. The maximum SNR values are38%, 12% and 16% greater, respectively, than the SNR

    values for the posterior positioned detail on the L5

    Figure 4. The effect of screenfilm speed class on (a) effective dose, (b) optical density (OD) difference of the L3 tranverse processand (c) signal-to-noise ratio (SNR) of the trabecular structure on the L3 vertebra for the anteroposterior projection. The effect of

    screenfilm sensitivity class on (d) effective dose, (e) OD difference of the L3 spinous process and (f) SNR of the trabecular structure

    at an anterior position on the L5 vertebra for the lateral projection.

    Optimization of lumbar spine X-ray imaging systems

    183The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    8/12

    vertebra and the anterior positioned details on the L5 andL1 vertebrae using the reference imaging system.

    Results of optimization

    Scatter rejection techniqueTable 2 shows the tube voltages for the six important

    details which give 0.9 of the contrast and SNR values for

    the lumbar spine AP reference system. These results are for

    an imaging system using a grid with N540 cm21 and r58

    and a Lanex Regular/TML screenfilm system (400 speed

    class). The table shows that there are differences in the

    voltage required for each detail. The lowest tube voltage is

    found for the L1 trabecular structure. The imaging

    requirement for this detail then limits the tube voltage

    to be less than or equal to this value so that the image

    quality criterion is met for all six details. The L1 trabecular

    structure was found to be the limiting detail for this grid,

    and for some of the other grids investigated (Table 3), as in

    these cases the SNR for this detail had the largest response

    to tube voltage. Thus for the grid under investigation, the

    largest dose saving that can be achieved is 18%.Table 3 shows the optimization of different grid designs

    using the Lanex Regular/TML screenfilm system (400speed class) for the AP projection. The highest tube

    voltages that satisfy the image quality criterion and the

    corresponding effective doses calculated with these systems

    are compared with the reference system, which is also 400

    speed class. All scatter-rejection techniques produce a dose

    saving compared with the reference system except for the

    grid with N570 cm21, r58 and d536 mm. The largest

    dose saving is for a grid with N570 cm21, r516 and

    d520 mm, which gives 22% lower effective dose than the

    reference system. These dose reductions are partly owing

    to the lower attenuation of the carbon fibre covers and

    interspace of the grids studied compared with the mixed

    material grid used in the reference system. The large dosereductions obtained for a large grid ratio are also owing

    to the fact that the tube voltage has to be increased

    Figure 5. The effect of median optical density (OD) on (a) effective dose, (b) OD difference and (c) signal-to-noise ratio (SNR)for the anteroposterior projection. The effect of median OD on (d) effective dose, (e) OD difference and (f) SNR for the lateral

    projection.

    Table 2. The tube voltages which produce 10% lower contrast(DOD, difference in optical density) or signal-to-noise ratio

    (SNR) for the six anatomical details (see Important Contrast

    Details section) than obtained with the lumbar spine anterior-

    posterior reference system using an imaging system with a

    N540 cm21, r58, d540 mm grid and a Lanex Regular/TML

    screenfilm system (400 speed class). The corresponding effec-

    tive dose relative to the value for the reference imaging system

    (Eastman Kodak Campany, Rochester, NY) is also given. The

    detail which limits the optimization, i.e. the one which requires

    the lowest tube voltage, is written in bold italics

    Detail Image quality

    measure

    Tube

    voltage (kV)

    Relative

    effective dose

    L5T DOD 78.3 0.76

    L3T DOD 77.1 0.78

    L1T DOD 77.0 0.78

    L5D SNR 76.4 0.80

    L3D SNR 75.8 0.81L1D SNR 75.5 0.82

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    184 The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    9/12

    substantially in order to reduce the contrast and SNR toexactly match the image quality criterion. The oppositewas found for the chest AP projection [8] where gridswith a low grid ratio were found to be optimal. This wasdue to the increase in effective dose with increasing tube

    voltage above 110 kV.Table 4 shows the optimization of different grid designs

    using the Lanex Regular/TML screenfilm system for theLAT projection. It was found that there are no dosesavings for these grids compared with the reference system.This is due to the reference system using the more sensi-

    tive Lanex Fast screenfilm system (600 speed class).Therefore, a compromise for a 400 speed class imagingsystem would be to use a N540 cm21 and r58 grid in boththe AP and LAT projections. This provides a smalloverall dose saving of 5% compared with the respectiveAP and LAT reference imaging systems.

    Screenfilm speedTable 5 shows the results of the optimization of scatter-

    rejection technique using the 600 speed class system for the

    AP projection. The highest tube voltages for the 600 speedclass system are on average 5 kV less than the highest tubevoltages for the 400 speed class system (Table 3). The tubevoltage is lower than for the 400 speed class system inorder to recover the reduction in SNR for the faster

    system (Figure 4). Overall, the use of the faster screenfilmsystem results in greater dose savings than the 400 speedclass systems. For example, for the N540 cm21 and r516grid, the effective dose is 34% lower than for the referencesystem.

    There is a similar effect on the LAT projection using a

    faster screenfilm system. The highest tube voltage whichmeets the imaging requirements for the 600 speed class isabout 4 kV lower than the highest tube voltage for the 400speed class. The largest dose reductions are for the gridswith N540 cm21 with the effective dose values between12% and 15% smaller than the effective dose produced bythe reference imaging system.

    Optical densityFigure 6 shows the optimization of the median OD,

    ODmed in the AP projection. The median OD was variedbetween 80% and 150% of the reference system value of1.36. The system studied used a grid with N540 cm21 and

    r512 and a Lanex Regular/TML screenfilm system.Figure 6a shows the variation of the highest tube voltagethat fulfills the image quality requirement as a functionof ODmed. The corresponding limiting detail and image

    quality parameter type are shown for each data point. Thehighest tube voltage increases with increasing ODmed untila maximum value of 81 kV is reached at an ODmed of 1.36and then decreases. Below an ODmed of 1.36, the contrastand SNR of each detail all increase with increasing ODmed.The limiting detail is the trabecular structure on the L1vertebra as its SNR has the largest response with ODmed.Above an ODmed of 1.36, the contrasts of the L5 and L3transverse processes decrease with increasing ODmed.

    Therefore, the tube voltage has to be decreased in orderto recover the reduced contrast to meet the required imagequality criterion. The L5 transverse process is the limiting

    Table 4. The best tube voltages and the corresponding valuesof the relative effective dose for each grid studied with a Lanex

    Regular/TML screenfilm system (Eastman Kodak Campany,

    Rochester, NY) (400 speed class) for the lateral projection. Thelimiting detail (see Important Contrast Details section) and the

    image quality measure (difference in optical density (DOD) or

    signal-to-noise ratio (SNR)) are also given. The system which

    gives the lowest patient dose is written in bold italics

    Scatter rejection

    technique

    N (cm21), r, d (mm)

    Detail Image

    quality

    measure

    Best tube

    voltage

    (kV)

    Relative

    effective

    dose

    40,08,40 L1S DOD 84 1.09

    40,12,40 L3S DOD 90 1.12

    40,16,40 L1S DOD 90 1.22

    70,08,36 L5B SNR 82 1.18

    70,12,36 L3S DOD 87 1.24

    70,16,36 L1S DOD 90 1.26

    N, strip density; r, grid ratio; d, lead strip width.

    Table 5. The best tube voltages and the corresponding values ofthe relative effective dose for each grid studied with a Lanex

    Fast/TML screenfilm system (Eastman Kodak Campany,

    Rochester, NY) (600 speed class) for the anterior-posterior projec-

    tion. The limiting detail (see Important Contrast Details section)

    and the image quality measure (difference in optical density

    (DOD) or signal-to-noise ratio (SNR)) are also given. The system

    which gives the lowest patient dose is written in bold italics

    Scatter rejection

    technique

    N (cm21), r, d (mm)

    Detail Image

    quality

    measure

    Best tube

    voltage

    (kV)

    Relative

    effective

    dose

    40,08,40 L1D SNR 71 0.74

    40,12,40 L3D SNR 76 0.67

    40,16,40 L5D SNR 79 0.66

    70,08,36 L3D SNR 68 0.89

    70,12,36 L3D SNR 74 0.78

    70,16,36 L5D SNR 77 0.74

    70,08,20 L1D SNR 65 0.83

    70,12,20 L1D SNR 70 0.73

    70,16,20 L3D SNR 74 0.67

    N, strip density; r, grid ratio; d, lead strip width.

    Table 3. The best tube voltages and the corresponding values ofthe relative effective dose for each grid studied with a Lanex

    Regular/TML screenfilm system (Eastman Kodak Campany,

    Rochester, NY) (400 speed class) for the anterior-posterior projec-

    tion. The limiting detail (see Important Contrast Details section)

    and the image quality measure (difference in optical density

    (DOD) or signal-to-noise ratio (SNR)) are also given. The systems

    which give the lowest patient dose are written in bold italics

    Scatter rejection

    technique

    N (cm21), r, d (mm)

    Detail Image

    quality

    measure

    Best tube

    voltage

    (kV)

    Relative

    effective

    dose

    40,08,40 L1D SNR 76 0.82

    40,12,40 L1D SNR 81 0.80

    40,16,40 L5T DOD 85 0.79

    70,08,36 L3D SNR 72 1.02

    70,12,36 L1D SNR 79 0.89

    70,16,36 L5D SNR 82 0.86

    70,08,20 L1D SNR 69 0.92

    70,12,20 L1D SNR 57 0.82

    70,16,20 L1D SNR 79 0.78

    N, strip density; r, grid ratio; d, lead strip width.

    Optimization of lumbar spine X-ray imaging systems

    185The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    10/12

    detail as its contrast has the largest decrease withincreasing ODmed. Figure 6b shows that there are dosesavings below an ODmed of 1.50 with the effective dosebeing 25% lower than the value for the reference system at

    an ODmed of 1.09.There is a similar variation with ODmed for the lateral

    projection. The system used is a 40 cm21, r512 grid with aLanex Regular/TML screenfilm system. Figure 6c showsthat the highest tube voltage reaches a maximum value of90 kV at an ODmed of 1.22. Figure 6d shows that there is aminimum dose at an ODmed of 1.09 with the effective dosebeing 7% lower than reference system value. The highesttube voltage is lower at an ODmed of 1.09 than at 1.36 in

    order to recover the lower SNR as ODmed decreases(Figure 5). There is a small dose saving due to using thecarbon fibre grid rather than the mixed material grid.

    Discussion

    Our work on the optimization of the scatter rejection

    technique has shown that the tube voltage could be

    decreased or increased in order to produce a dose reduc-

    tion depending on the grid design. Vano et al [5] increased

    the tube voltage from 60 kV to 90 kV to produce a dose

    reduction of 35% whilst maintaining image quality for the

    lumbar spine AP examination. However, Almen et al [4]

    have shown that increasing the tube voltage from 70 kV

    to 90 kV significantly alters the image quality of AP

    films, whereas increasing from 77 kV to 95 kV does not

    significantly alter the image quality of LAT films. In our

    optimization studies, the tube voltages that fulfilled theimage quality criterion were less than 85 kV for the AP

    films and 90 kV for the LAT films.

    Figure 6. The optimization of median optical density with (a) the highest tube voltage consistent with the requirement to obtain atleast 90% of the image quality of the reference system for all details considered and (b) the corresponding values of the effective

    dose relative to the reference system values for the anterior-posterior view. The optimization of median optical density with (c) the

    highest tube voltage and (d) the corresponding effective dose for the lateral view.

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    186 The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    11/12

    Further evidence that our work closely corresponds tothat of Almen et al is given by Sandborg et al [30].Sandborg showed that the physical parameters such ascontrast and SNR could be used to predict the order thatthe imaging systems were ranked by the European radiol-

    ogists [4]. For example, in the AP projection Almen et alfound significant differences in image quality for changing

    tube voltage, but not for changing speed class. This canalso be demonstrated from our study of changing theimage parameters and observing their effect on calculatedimage quality. By increasing the tube voltage from 70 kV

    to 90 kV, a large decrease of 28% was observed in thecalculated contrast and SNR whereas only a smalldecrease of 10% was seen in the SNR for increasing thespeed class from 400 to 600. It is therefore reassuringthat the work in this paper is consistent with changes inimage quality observed clinically.

    Conclusions

    The results of varying the different imaging parametersshows how straightforward it is to have high image qualityand high patient dose, e.g. low tube voltage and to have

    low image quality and low patient dose, e.g. high tubevoltage. Conversely, it is difficult to balance high imagequality and low patient dose. The optimization of radio-graphic imaging involves several different parameters.Therefore, it is very useful that a Monte Carlo model canbe used to point out imaging systems that give low patientdose whilst still maintaining the same image quality asreference systems. These systems are worth investigation infuture, more time-consuming, clinical trials.

    For 400 speed class systems using grids in the APprojection, a dose reduction of between 8% and 22% canbe achieved. A further dose reduction of 13% is possiblewith a 600 speed class system using a grid. Table 6 shows

    the imaging system configuration that produced the largestdose reduction in our work. Dose reductions of a similarsize can be obtained for a grid with a high grid ratio(r516), a high strip density (N570 cm

    21) and a small leadstrip width (d520 mm).

    For 400 speed class systems using grids in the LATprojection, only a small dose reduction of 7% could beachieved by reducing the operating OD from 1.36 to 1.09.The largest dose reduction of 15% was obtained using the

    600 speed class screenfilm system shown in Table 6. Thescope for large dose reductions in the LAT projection wasrestricted as a 600 speed class screenfilm system was used

    as the reference system. For both AP and LAT projec-tions, the dose advantage of using carbon fibre compo-nents has been shown throughout this work as thereference system used a grid constructed from aluminiumand carbon fibre.

    Our work clearly shows that the largest dose reductionsare for 600 speed class systems. However, in a recent

    review [28] of patient doses from screenfilm imaging inthe UK for the year 2000, the National RadiologicalProtection Board (NRPB) shows that there are signifi-cantly fewer 600 speed class systems in use compared with

    400 speed class systems. The review [28] also shows thecontinuing trend for lower dose per lumbar spine radio-graph of 37% in the period from 1984 to 1995 and 18% inthe period from 1995 to 2000. The NRPB state that this isdue to the increasing use of faster screenfilm combina-tions. Therefore, our work highlights that there are stillpotential optimizations to be made in lumbar spineradiography. It is also reassuring to know that the systems

    found by the optimizations are similar to those recom-mended by the CEC guidelines [3] as given in Table 6.

    Acknowledgment

    Dr F R Verdun (Lausanne, Switzerland) is thanked forsupplying measured H and D curves, modulation transferfunction and noise power spectra of the screenfilmcombinations used in this work. Alexandr Malusek isacknowledged for the image of the voxel phantom inFigure 1.

    References

    1. Espeland A, Korsbrekke K, Albreksten G, Larsen JL.

    Observer variation in plain radiography of the lumbarsacral

    spine. Br J Radiol 1998;71:36675.

    2. Hart D, Wall BF. Radiation exposure of the UK population

    from medical and dental X-ray examinations. NRPB-W4,

    Didcot, Oxon: NRPB, 2002.

    3. Carmichael JH, Maccia C, Moores BM, Oestmann JW,

    Schibilla H, Van Tiggelen R, Wall B, editors. European

    guidelines on quality criteria for diagnostic radiographic

    images. Report EUR 16260 EN. Luxembourg: European

    Commission, 1996.

    4. Almen A, Tingberg A, Mattsson S, Besjakov J, Kheddache S,

    Lanhede B, et al. The influence of different technique factorson image quality of lumbar spine radiographs as evaluated by

    established CEC image criteria. Br J Radiol 2000;73:11929.

    Table 6. The optimum imaging system configuration found in this work for the lumbar spine anteroposterior and lateralprojections compared with the examples of good radiographic technique given by the Commission of the European Communities

    (CEC) guidelines [3]

    Parameters AP imaging systems Lateral imaging systems

    CEC guidelines Results of optimization CEC guidelines Result of optimization

    Tube voltage 7590 kV 79 kV 8095 kV 82 kV

    Filtration 3.0 mmAl 4.7 mmAl 3.0 mmAl 4.7 mmAlFocal spot size ,1.3 mm 0.9 mm ,1.3 mm 0.9 mm

    Focusfilm distance 1.15 m (1.0 to 1.5 m) 1.46 m 1.15 m (1.0 to 1.5 m) 1.46 m

    Grid ratio 10 16 10 8

    Strip density 40 cm21 40 cm21 40 cm21 40 cm21

    Strip width 40 mm 40 mm

    Speed class 400 600 400 600

    Optimization of lumbar spine X-ray imaging systems

    187The British Journal of Radiology, March 2003

  • 8/2/2019 A Study and Optimization of Lumbar Spine X-Ray Imaging

    12/12

    5. Vano E, Oliete S, Gonzalez L, Guibelaude E, Velasco A,

    Fernandez JM. Image quality and dose in lumbar spine

    examinations: results of a 5 year quality control programme

    following the European quality criteria trial. Br J Radiol

    1995;68:13325.

    6. Sandborg M, McVey G, Dance DR, Persliden J, Alm

    Carlsson G. A voxel phantom based Monte Carlo computer

    program for optimization of chest and lumbar spine X-ray

    imaging. Rad Prot Dosim 2000;90:1058.7. Sandborg M, Tingberg A, Dance DR, Lanhede B, Almen A,

    McVey G, et al. Demonstration of correlations between

    clinical and physical image quality measures in chest and

    lumbar spine screenfilm radiography. Br J Radiol 2001;

    74:5208.

    8. Sandborg M, McVey G, Dance DR, Alm Carlsson G.

    Schemes for the optimization of chest radiography using a

    computer model of the patient and X-ray imaging system.

    Med Phys 2001;28:200719.

    9. Dance DR, McVey GH, Sandborg M, Alm Carlsson G,

    Verdun FR. The optimization of lumbar spine AP radio-

    graphy using a realistic computer model. Rad Prot Dosim

    2000;90:20710.

    10. Sandborg M, Dance DR, Persliden J, Alm Carlsson G. A

    Monte Carlo program for the calculation of contrast, noiseand absorbed dose in diagnostic radiology. Comput Methods

    Programs Biomed 1994;42:16780.

    11. Dance DR, Lester S, Alm Carlsson G, Sandborg M, Persliden

    J. The use of carbon fibre material in radiographic cassettes:

    estimation of the dose and contrast advantages. Br J Radiol

    1997;70:38390.

    12. Zubal G, Harrell CR. Voxel based Monte Carlo calculations

    of Nuclear Medicine images and applied variance reduction

    techniques. Image and Vision Computing 1992;10:3428.

    13. Zubal G, Harrell CR, Smith EO, Rattner Z, Gindi G, Hoffer

    PB. Computerised three dimensional segmented human

    anatomy. Med Phys 1994;21:299302.

    14. International Commission on Radiological Protection. 1990

    Recommendations of the International Commission on

    Radiological Protection, ICRP Publication 60. Ann ICRP1991:21.

    15. International Commission on Radiation Units and Measure-

    ments. Phantoms and computational models in therapy,

    diagnosis and protection. ICRU Report No. 48. Bethesda,

    MD: ICRU, 1992.

    16. Dance DR, McVey G, Sandborg M, Persliden J, Alm

    Carlsson G. Calibration and validation of a voxel phantom

    for use in the Monte Carlo modeling and optimization of

    X-ray imaging systems. Proc SPIE 1999;3659:54859.

    17. International Commission on Radiation Units and Measure-

    ments. Photon, electron, proton and neutron interaction data

    for body tissues. ICRU Report No. 46. Bethesda, MD:

    ICRU, 1992.

    18. Kramer R. Determination of conversion factors between

    body dose and relevant radiation quantities for external

    x- and c-radiation. Neuherberg, Germany: GSF, 1979.

    19. Birch R, Marshall B, Ardan GM. Catalogue of spectral data

    for diagnostic X-rays. Scientific Report Series 30. London:The Hospital Physicists Association, 1979.

    20. International Standards Organization. Photographysensito-

    metry of screenfilm systems for medical radiographyISO

    method for determination of sensitometic curve shape, speed

    and average gradient. ISO 9236. Geneva, Switzerland: ISO,

    1993.

    21. International Commission on Radiation Units and Measure-

    ments. Medical imagingthe assessment of image quality.

    ICRU Report 54. Bethesda, MD: ICRU, 1996.

    22. Tapiovaara MJ, Sandborg M. Evaluation of image quality in

    fluoroscopy by measurements and Monte Carlo calculations.

    Phys Med Biol 1995;40:589607.

    23. Tapiovaara MJ, Sandborg M, Dance DR. A search for

    improved technique factors in paediatric fluoroscopy. Phys

    Med Biol 1999;44:53759.

    24. Lubberts G. Random Noise produced by X-ray fluorescent

    screens. J Opt Soc Am 1968;58:147583.

    25. Swank RK. Absorption and noise in X-ray phosphors. J Appl

    Phys 1973;44:4199203.

    26. Nishikawa RM, Yaffe MJ. Model of the spatial-frequency

    dependent detective quantum efficiency of phosphor screens.

    Med Phys 1990;17:894904.

    27. Tingberg A. PhD thesis: quantifying the quality of medical

    X-ray images: an evaluation based on normal anatomy for

    lumbar spine and chest radiography. Lund, Sweden: Lund

    University, 2000.

    28. Hart D, Hillier MC, Wall BF. Doses to patients from medical

    X-ray examinations in the UK2000 Review. NRPB-W14.

    Didcot, UK: NRPB, 2002.29. Hughes JS. Ionising radiation exposure of the UK popula-

    tion: 1999 review. NRPB-R311. Didcot, UK: NRPB, 1999.

    30. Sandborg M, McVey G, Dance DR, Alm Carlsson G.

    Comparison of model predictions of image quality with

    results of clinical trials in chest and lumbar spine screenfilm

    imaging. Rad Prot Dosim 2000;90:1736.

    G McVey, M Sandborg, D R Dance and G Alm Carlsson

    188 The British Journal of Radiology, March 2003