Citation: Saunders TFC and Osborne MS. A Rare Complication of Nasal Retaining Loop Insertion. Austin J Otolaryngol. 2015;2(4): 1038. Austin J Otolaryngol - Volume 2 Issue 4 - 2015 ISSN : 2473-0645 | www.austinpublishinggroup.com Saunders et al. © All rights are reserved Austin Journal of Otolaryngology Open Access Abstract An 80 year old man was referred by his GP with a 6 week history of foreign body sensation in his left nasal cavity caused by an unidentified metallic wire seen on anterior rhinoscopy. With a combination of radiology and flexible nasendoscopy the cause was found to be a flexible guidewire that had been used to insert a nasal retaining loop (nasal bridle) 12 months earlier. This required removal under general anaesthetic. Fortunately this retained foreign body resulted in no long term sequela to the nasal cavity. To the authors knowledge there are no documented cases reported in the literature, and this case highlights the importance of developing guidelines of correct insertion and removal of these popular devices. Case Report A Rare Complication of Nasal Retaining Loop Insertion Saunders TFC* and Osborne MS Department of Otolaryngology, Royal Stoke University Hospital, UK *Corresponding author: Tom Saunders, Department of Otolaryngology, Royal Stoke University Hospital, UK Received: February 12, 2015; Accepted: March 25, 2015; Published: March 27, 2015 Introduction and Case Presentation An 83 year old man was referred by his GP with a six week history of a foreign body sensation in the leſt nasal cavity, causing irritation and skin rash of the philtrum. ere was no epistaxis, discharge or obstruction of airflow through the affected side of the nose. e patient reported no history of any foreign body insertion into the nose since his coronary artery bypass graſt 12 months previously when he had been admitted to ITU post surgery and required a Bridle fixed nasogastric tube for feeding during. is was as a result of an episode of post-operative confusion. On examination a metallic flexible wire was seen to be coming from the leſt nasal cavity. On gentle manipulation under local anaesthetic it was not freely mobile and seemed to be heading to the post nasal space. A flexible nasendoscopy was performed – the wire was imbedded in nasal floor, covered with granulation tissue, and we were unable to remove at this time due to resistance in the PNS where the orange cap of the wire that was stuck to the floor of the nose. Aſter examining the manufacturer’s brochure for the Bridle NG tube we concluded that the wire corresponded to the flexible guide wire that is used during the insertion, but should be removed prior to fixation of the bridle. is guide wire had incorrectly been pulled through the right nasal cavity, passing around posterior to the Vomer and into the leſt nasal cavity. It was not noticed during removal of the bridle or the NG tube and had been leſt in situ for 12 months. Fortunately for this patient, there was no obvious long-term damage to the nasal cavity mucosa or the septum. Discussion We carried out a literature search on the topic, but no cases have been reported of this particular complication, which is purely iatrogenic. NHS trusts across the country provide clear guidelines [1], adapted from the manufacturer’s instructions, on the correct insertion of bridle nasal loops. e known complications include trauma to the nasal septum, epistaxis and mucosal irritation. Bridle loops (also known as ‘nasal retaining loops’) are used to maintain an Figure 1: A lateral plain film radiograph was taken. Figure 2: e patient was taken the theatre for removal under general anaesthetic. Figure 3: e 14cm wire was removed from the oral cavity.

A Rare Complication of Nasal Retaining Loop Insertiond.researchbib.com/f/9nLKImqTyhpUIvoTymnTyhM2qlo3IjYzAioF9iqT… · Austin Journal of Otolaryngology. Open Access. Abstract. An

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: A Rare Complication of Nasal Retaining Loop Insertiond.researchbib.com/f/9nLKImqTyhpUIvoTymnTyhM2qlo3IjYzAioF9iqT… · Austin Journal of Otolaryngology. Open Access. Abstract. An

Citation: Saunders TFC and Osborne MS. A Rare Complication of Nasal Retaining Loop Insertion. Austin J Otolaryngol. 2015;2(4): 1038.

Austin J Otolaryngol - Volume 2 Issue 4 - 2015ISSN : 2473-0645 | www.austinpublishinggroup.com Saunders et al. © All rights are reserved

Austin Journal of OtolaryngologyOpen Access

Abstract

An 80 year old man was referred by his GP with a 6 week history of foreign body sensation in his left nasal cavity caused by an unidentified metallic wire seen on anterior rhinoscopy. With a combination of radiology and flexible nasendoscopy the cause was found to be a flexible guidewire that had been used to insert a nasal retaining loop (nasal bridle) 12 months earlier. This required removal under general anaesthetic. Fortunately this retained foreign body resulted in no long term sequela to the nasal cavity. To the authors knowledge there are no documented cases reported in the literature, and this case highlights the importance of developing guidelines of correct insertion and removal of these popular devices.

Case Report

A Rare Complication of Nasal Retaining Loop InsertionSaunders TFC* and Osborne MSDepartment of Otolaryngology, Royal Stoke University Hospital, UK

*Corresponding author: Tom Saunders, Department of Otolaryngology, Royal Stoke University Hospital, UK

Received: February 12, 2015; Accepted: March 25, 2015; Published: March 27, 2015

Introduction and Case PresentationAn 83 year old man was referred by his GP with a six week history

of a foreign body sensation in the left nasal cavity, causing irritation and skin rash of the philtrum. There was no epistaxis, discharge or obstruction of airflow through the affected side of the nose. The patient reported no history of any foreign body insertion into the nose since his coronary artery bypass graft 12 months previously when he had been admitted to ITU post surgery and required a Bridle fixed nasogastric tube for feeding during. This was as a result of an episode of post-operative confusion.

On examination a metallic flexible wire was seen to be coming from the left nasal cavity. On gentle manipulation under local anaesthetic it was not freely mobile and seemed to be heading to the post nasal space. A flexible nasendoscopy was performed – the wire was imbedded in nasal floor, covered with granulation tissue, and we were unable to remove at this time due to resistance in the PNS where the orange cap of the wire that was stuck to the floor of the nose.

After examining the manufacturer’s brochure for the Bridle NG tube we concluded that the wire corresponded to the flexible guide wire that is used during the insertion, but should be removed prior to fixation of the bridle. This guide wire had incorrectly been pulled through the right nasal cavity, passing around posterior to the Vomer

and into the left nasal cavity. It was not noticed during removal of the bridle or the NG tube and had been left in situ for 12 months.

Fortunately for this patient, there was no obvious long-term damage to the nasal cavity mucosa or the septum.

DiscussionWe carried out a literature search on the topic, but no cases

have been reported of this particular complication, which is purely iatrogenic. NHS trusts across the country provide clear guidelines [1], adapted from the manufacturer’s instructions, on the correct insertion of bridle nasal loops. The known complications include trauma to the nasal septum, epistaxis and mucosal irritation. Bridle loops (also known as ‘nasal retaining loops’) are used to maintain an Figure 1: A lateral plain film radiograph was taken.

Figure 2: The patient was taken the theatre for removal under general anaesthetic.

Figure 3: The 14cm wire was removed from the oral cavity.

Page 2: A Rare Complication of Nasal Retaining Loop Insertiond.researchbib.com/f/9nLKImqTyhpUIvoTymnTyhM2qlo3IjYzAioF9iqT… · Austin Journal of Otolaryngology. Open Access. Abstract. An

Austin J Otolaryngol 2(4): id1038 (2015) - Page - 02

Saunders TFC Austin Publishing Group

Submit your Manuscript | www.austinpublishinggroup.com

enteral feeding option in a variety of conditions and reduce the risk of decannulation of NG tubes and can reduce the need for percutaneous endogastric (PEG) tube insertions.

This report however highlights that care needs to be taken when inserting the bridle loop and also when removing the NG tube. Medical staff using these devises should be aware of manufacture

insertion guidelines, and these should be used to develop local guidelines to ensure that guidewires are removed correctly and not left in situ in the nasal cavity.