Upload
bm-glover
View
212
Download
0
Embed Size (px)
Citation preview
of patients: (a) 135 patients admitted in intensive care unit, yielding a total
of 195 (PEM, standard) ECG pairs; and (b) 24 ambulatory cardiac patients
recorded during their routine visit at the cardiology hospital of Lyon. Both
the standard and the derived 12-lead ECGs have been reprocessed by the
Lyon Program. A quantitative comparison of the computerized global and
lead-by-lead measurements was performed on data set a (Table 1), and a
qualitative comparison was made on data set b (Table 2) by a cardiologist
who blindly analyzed, in measurements and interpretation, the 2 derived
ECGs with reference to the standard 12-lead ECG. In conclusion, the
patient-specific transform provides better results, similar to the Mason-Likar
lead system.
doi:10.1016/j.jelectrocard.2005.06.039
QT measurement performance in a Holter application
Dirk Q. Feild (Advanced Algorithm Research Center, Philips Medical
Systems, USA)
Faced with moderate sampling rates, limited operator interaction, and high
artifact content, obtaining stable QT measurements in an ambulatory
environment is extremely challenging but clinically important. Automatic
analysis is performed on all available channels of electrocardiographic data.
After cubic spline baseline removal on each electrocardiographic channel,
moving average templates are created. Complexes that are not close to the
current composite are rejected. These templates are then combined into a
vector magnitude signal. A Philips proprietary method is used to detect both
the onset of the Q wave and the offset of the T wave. This algorithm is
insensitive to signal amplitudes and residual isoelectric displacements but
slightly sensitive to large U waves.
The results by Jane et al (Comput cardiol 1997) used the human-
determined best-performing channel per record on the best 73 of the
database and on the full database. The results for the Philips algorithm are
nearly equivalent to the manually chosen results of Jane as shown in
Table 1. QT results in Table 2 compare favorably with expert annotation.
The algorithm has proven robust in a commercially available 3-channel
Holter application.
doi:10.1016/j.jelectrocard.2005.06.040
A novel rectangular biphasic waveform from a radio
frequency defibrillator compared with a conventional
waveform for the transvenous cardioversion of chronic
atrial fibrillation in patients
B.M. Glover a,b,C.J. McCanna,b, S.J. Walsha,M.J. Moorea,b,G.Manoharana,
M.J. Robertsa, C.M. Wilsona, J.D. Allen, J.Mc.C. Andersonc, A.A.J. Adgey a
( aRegional Medical Cardiology Centre, Jordanstown, UK; bRoyal Victoria
Hospital, Belfast, Queens University, Jordanstown, UK; cUniversity of
Ulster, Belfast, Jordanstown, UK)
Purpose: The optimal waveform for the transvenous direct current
cardioversion of atrial fibrillation (AF) is unknown. A novel rectangular
biphasic waveform (6/6-milliseconds duration, phase 2 peak voltage 50% of
phase 1) delivered from a radio frequency (RF)–powered defibrillator was
compared with a conventional capacitor–based, exponential biphasic
waveform of equivalent duration and voltage.
Method: Patients with chronic AF (fully anticoagulated) were randomized
to receive either the RF or a conventional trapezoidal waveform (Ventritex
HVS-02). Defibrillation electrodes were positioned in the right atrial
appendage (cathode) and distal coronary sinus (anode). All shocks were
R-wave synchronized. Phase 1 peak voltage was increased in stepwise
progression from 50 to 300 V. Success was defined as return of sinus
rhythm for 30 seconds. Cardiac troponin and creatine kinase–MB were
checked postprocedure.
Results: Patients (n = 16, 11 male) received 83 shocks (RF, 40;
conventional, 43). Mean age was 63 (F11.6) years, mean body mass
index was 28 (F6), and mean duration of AF was 5.8 (F5.7) months. The
groups were matched in age, sex, body mass index, duration of AF,
etiology, drugs, and echocardiographic features. The RF waveform
performed significantly better than the conventional waveform for the
cardioversion of chronic AF (7 [87%]/8 patients vs 1 [12%]/8 patients
Table 1
The mean F SD correlation coefficients between the PEM and the standard ECG measurements of data set a
Global meas (98% of cases) Lead meas for the same wave profiles
I II V2 V1, V3-V6
Ge Sp
0.85 F 1.07 0.84 F 1.06 0.85 F 1.12 0.91 F 0.04 0.70 F 0.16 0.84 F 0.11
Meas indicates measurements; Ge, generic transform; Sp, patient-specific transform.
Table 2
The physician’s evaluation results on data set b
Opinion Diagnosed Meas (%)
Ge+/Sp+ 22 89.7
Ge+/Sp� 0 0
Sp+/Ge� 1 3.4
Ge�/Sp� 1 6.9
Table 1
Automatic measurement results of T end and Q onset on MIT QT database compared with published data (Comput cardiol 1997)
n Qon (ms) SD (ms) Sensitivity PPV Tend (ms) SD (ms) Sensitivity PPV
Jane-best 73 �0.63 19.27 99.7 99.87
Philips-best 73 �0.91 14.01 100.0 100.0 �2.39 18.6 99.92 99.71
Jane-all 105 �7.82 10.86 99.88 99.97 18.68 29.79 98.92 99.91
Philips-all 105 0.09 14.77 100.0 100.0 �6.13 34.38 98.08 89.41
Table 2
Automatic measurement results of QT interval on the same database
n QT (ms) SD (ms) Sensitivity PPV
Philips-best 73 �1.20 22.54 99.92 99.71
Philips-all 105 �5.52 36.62 98.08 89.41
Poster Session I / Journal of Electrocardiology 38 (2005) 33–3934
success; P = .003). The mean leading edge voltage for the RF was 221 V
(range, 100-300 V) and for the conventional waveform was 240 V. No
significant arrhythmias, sinus pauses, or episodes of hypotension occurred.
There was no elevation of cardiac enzymes.
Conclusions: The novel biphasic waveform has a superior efficacy at a
lower voltage compared with the conventional waveform in the trans-
venous cardioversion of AF. There were no arrhythmic, hemodynamic
complications, or elevation of markers of myocardial injury. Use of this
waveform may improve the efficacy of implantable devices for the
treatment of AF.
doi:10.1016/j.jelectrocard.2005.06.041
Electrocardiographic artifact: a frequently misdiagnosed phenomenon
S. Hanon, J.S. Berger, S. Hurwitz, J. Fine, D.L. Brown, P. Schweitzer
Background: Electrocardiographic (ECG) artifact can imitate wide
complex tachycardia, leading to unnecessary diagnostic or therapeutic
interventions. The current study sought to determine the competency of
physicians in different specialties, with various levels of training, in
diagnosing ECG artifact.
Methods: Two ECGs of artifact simulating wide complex tachycardia were
used for analysis. Physicians recorded their rhythm diagnosis and
diagnostic certainty.
Results: Artifact was not recognized in 307 (82%) of 384 attempts. The
most common recorded diagnoses were ventricular tachycardia (54%) and
supraventricular tachycardia (25%). Failure to identify either artifact was
found in 123 (88%) of 140 residents, 8 (26%) of 31 fellows, and 5 (29%) of
17 attendings. Only 1% of residents, 32% of fellows, and 29% of attendings
correctly identified both artifacts. Among 74 physicians who were 100%
certain of their diagnosis, 34 (50%) were mistaken.
Conclusion: This physician survey suggests that ECG artifact is frequently
misdiagnosed. This finding indicates the need for improved training in the
recognition of artifact and the need for a heightened index of suspicion
among physicians who treat patients on telemetry monitoring.
doi:10.1016/j.jelectrocard.2005.06.042
Philips Medical Systems support for open access and use of
electrocardiographic data
Eric Helfenbein, Richard Gregg, Sophia Zhou (Advanced Algorithm
Research Center, Philips Medical Systems, USA)
The Philips extensible markup language (XML)–based electrocardio-
graphic (ECG) format has been published for more than 2 years. This is
the native format used by Philips electrocardiographs and 12-lead capable
bedside monitors and defibrillators. To support research and clinical trials,
Philips Medical Systems is freely distributing a set of 6 software
tools that provide easy access and conversion of the XML ECG files. The
set of software tools consists of the following. (1) The Food and
Drug Administration converter, which translates the ECG file to the HL7
annotated ECG file format adopted by the US Food and Drug Administra-
tion for digital submission of ECG data for pharmaceutical clinical trials. (2)
The Decompressor software provides unrestricted access to the raw ECG
data by decompressing the ECG into plain text sample values because the
ECG waveform data in the Philips XML file is compressed and base-64
encoded to permit efficient transmission and storage. (3) The Scalable Vector
Graphics (SVG) converter translates the ECG file into an XML SVG image
file that can be viewed using a web browser with the Adobe SVG plug-in
installed. The resulting image allows the researcher to review demogra-
phics, interpretation statements, and calibrated waveform data in a format
similar to the original printed cardiograph report. (4) The PC Receiver
software allows ECG files to be sent directly from a cardiograph to a
personal computer (PC) over a network connection. (5) The NewFilename
tool renames the default Globally Unique Identifier ECG filename to a name
containing the patient’s identification, name, and date/time of the ECG; this
allows the ECG files to be effectively managed on a PC. (6) The Directory
Scanner software automatically runs 1 or more of the conversion tools on
ECGs because they arrive in a specified folder in the PC. This collection of
software provides a valuable ECG tool set for use by clinicians, researchers,
and clinical trial sponsors.
doi:10.1016/j.jelectrocard.2005.06.043
Changes in optical map frequency spectra
characteristics before spontaneous termination
of ventricular fibrillation in isolated rabbit heart model
Suresh E. Joel, Peng-Wie Hsia (Virginia Commonwealth University,
Richmond, VA, USA)
Introduction: Once ventricular fibrillation (VF) sets in, it hardly terminates
spontaneously in humans. However, in small young animals, VF
spontaneously reverts back to normal rhythm. We studied frequency
characteristics of activation during a period before spontaneous termination
(SpT) of VF.
Methods: Optical map recordings (87 episodes) were performed during
SpT VF on isolated rabbit hearts (n = 15) using a high-speed CCD camera
(100 � 100, 256 frames per second). Dominant frequency (DF) maps and
dominant bandwidth maps were computed using continuous Fourier
transform. Regional distribution of epicardial DF and dominant bandwidth
values during VF and changes before termination were studied.
Results: An ordered arrangement of DF domains during VF was observed.
Left ventricular apex and right ventricular (RV) apex had the highest DF
values (DF left ventricular apex, DF RV apex N DF RV base, DF RV apex;
P b .0001), revealing the location of the mother rotor. Left and right sides of
the heart did not show significant differences in DF values. Dominant
frequency values reduced significantly ( P b .01) in all regions 1 second
before termination. Dominant bandwidth values dropped significantly only
in the apex before SpT. All regions that had a significant difference in DF
value distribution during VF also showed a significant decrease in difference
during the last second before SpT, leveling off the DF domain distribution.
Changes observed 1 second before termination were not observed during
other time segments during fibrillation.
Conclusions: The ordered regional organization of DF domains is lost
before termination. The mother rotor resides in the apical region of the
heart and terminates during the last second of fibrillation before SpT. Our
data suggest the disappearance of the highest DF domain as the primary
mechanism of SpT VF. Our results further support that a single, dominant,
high-frequency mother rotor maintains VF in isolated rabbit heart.
doi:10.1016/j.jelectrocard.2005.06.044
n 0 Correct 1 Correct 2 Correct
Residents
Internal
medicine
110 94 (85) 14 (13) 2 (2)
Emergency
medicine
17 17 (100) – –
Family
practice
13 12 (92) 1 (8) –
Fellows
Cardiology 24 4 (16) 10 (42) 10 (42)
Critical care 7 4 (57) 3 (43) –
Attendings
Cardiology 11 1 (9) 5 (45) 5 (45)
Emergency
medicine
6 4 (67) 2 (33) –
Values in parentheses are percentages.
Poster Session I / Journal of Electrocardiology 38 (2005) 33–39 35