a National Science Foundation Engineering Research Center in the MSU College of Engineering Center for Biofilm Engineering INTRODUCTION Evaluation of Fluid Path Colonization in Needle-Free Connectors and Biofilm Formation in Central Venous Catheters E. deLancey Pulcini 1 , A.E. Parker 1 , M. Ryder 2 , G. James 1 (1) Center for Biofilm Engineering Montana State University-Bozeman (2) Ryder Science Inc. Escondido, CA Poster#275 ASM Biofilms 2012 In the United States, approximately 80,000 catheter-related bloodstream infections (CRBSI) occur in ICUs annually 1 and approximately 250,000 CRBSIs occur hospital wide annually 2 . Needle-free connectors collected from patients have been shown to contain bacterial biofilms 3 . Previous studies have shown bacterial contamination and bacterial transfer through needle-free connectors 4 . Needle-free connectors were developed for use with vascular catheters to prevent needle stick injuries to health care workers. The use of needle-free connectors has been implicated in catheter related bloodstream infections. The objective of this study was to determine if there was a difference, by connector type, in the colonization of the fluid pathway in inoculated needle-free connectors and in attached vascular catheters after repeated flushing with no disinfection. Figure 1. The connector-catheter sets were placed in conical vials between flushes. The technician is flushing one of the connector- catheter sets. Figure 3. Image of a triple lumen catheter. For this project, single lumen catheters were used. 1. Mermel LA. Prevention of intravascular catheter-related infections. (Erratum: Ann Intern Med 133:395, 2000). Ann Intern Med 2000. 132:391–402. 2. Maki DG, Kluger DM, Crnich CJ. The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies. Mayo Clin Proc 2006. 81:1159–1171 3. Donlan RM, Murga R, Bell M, Toscano CM, Carr JH, Novicki TJ, Zuckerman C, Corey LC, Miller JM. Protocol for detection of biofilms on needleless connectors attached to central venous catheters. J Clin Microbiol. 2001. 39:750-3. 4. Ryder M, Fisher S, Hamilton G, Hamilton M, James G. Bacterial transfer through needlefree connectors: comparison of nine different devices. Presented at the Society for Healthcare Epidemiology of America Annual Scientific Meeting, April 2007 The connector septum was inoculated twice a day with 10 6 CFU Staphylococcus aureus ATCC # 6538. The inoculated connector was allowed to dry for 30 minutes and then was attached to a catheter. Each connector-catheter set was flushed with 3.0 ml sterile saline which was collected and plated (First Flush). The catheter-connector sets were sterile saline (SS) flushed twice more, locked with sterile Brain Heart Infusion Broth (BHI) for 1 hour and SS flushed three more times. The last flush was also collected and plated (Last Flush). The connector-catheter sets were inoculated a second time each day after the 6th sterile saline flush followed by a second round of flushing, plating and locks for a total of 18 connector accesses daily, considered to be a routine number of accesses in an intensive care unit. The entire procedure, inoculation and flushing, was repeated each day for 5 days. On Days 4 and 5, two connector-catheter sets for each connector type were destructively sampled for bacterial counts and microscopy. Statistical analysis was performed using mixed effect ANOVA analysis and Tukey’s tests to determine significant mean differences of bacteria in the flush, hub, catheter segment or connector amongst the different needle-free connectors. A multiple linear regression was used to determine if any combination of the variables measured in the study could significantly predict the log density of bacteria either in the flush. METHODS RESULTS REFERENCES RESULTS A total of 8 needle-free connectors were evaluated in this study. Three of each connector type were evaluated in three replicate runs (n=9) with the MicroClave® (ICU Medical Inc.) serving as the matched control for every run (n =27). METHODS ACKNOWLEDGEMENTS This research was funded by ICU Medical. RESULTS Surface Inoculation The mean log densities (LD) of the surface inoculations averaged across all days were statistically equivalent as long as mean differences as large as 0.37 were assumed to be negligible. For each day individually, the mean LD of the surface inoculum (averaged across the two inoculations for each day) were shown to be statistically equivalent as long as mean differences as large as 0.53 were assumed to be negligible. 4 3 2 1 0 5.5 5.0 4.5 4.0 3.5 3.0 D ay m ean log(C FU /flush) M icroclave Sm artSite C learLink Invision M axim us Q -Syte O ne-Link B ionector C onnector D aily leastsquare m ean bacterialdensities in the flush The mean log density (LD) in the daily flush for the MicroClave connector was significantly smaller compared to any of the other connector types tested (p<0.0005). The Q-Syte had the significantly largest mean LD of bacteria in the flush compared to any of the other connector types. (p <0.0005). The MicroClave connector had the significantly smallest mean log density (LD) of bacteria in the flush compared to any of the other connector types (p<0.0005). The mean LD of bacteria in the hub for the MicroClave connector was significantly smaller than the SmartSite, Invision, Q-Syte and Bionector connectors. The mean LD of bacteria in the catheter segment for the MicroClave was significantly smaller that the SmartSite, Invision, Q-Syte and Bionector connectors. The mean LD of bacteria in the MicroClave connector was significantly smaller than the SmartSite, Invision, and Bionector connectors. The daily mean LD of bacteria attached in the hub and in the connector were better predictors of the daily mean LD of bacteria recovered from the flush than the mean LD of bacteria in the catheter segment. The least square mean for all flushes for all days was calculated. The color scheme indicates the significant groups (p< 0.05). LeastSquare M ean (Log C FU /flush) Microclave 3.128 A B ionector 3.637 B O ne-Link 3.907 B C C learLink 4.176 C D Sm artSite 4.176 C D Invision 4.368 D E Maxim us 4.573 E Q -Syte 5.276 F SignificantG roups M ean Flush forallD ays and allFlushes LeastSquare M ean (Log C FU /hub) M icroClave 1.594 A O ne-Link 1.694 A B C learLink 1.853 A B C Maxim us 2.101 A B C D Sm artSite 2.095 B C D E Q -syte 2.925 C E F B ionector 2.905 D E F Invision 3.14 D F SignificantG roups H ub D ays 3 and 4 The least square mean for the destructive sampling of the catheter hubs was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05) The least square mean for the destructive sampling of the connectors was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05). LeastSquare M ean (C FU /connector) M icroClave 2.544 A O ne-Link 2.592 A B C learLink 3.005 A B C Sm artSite 3.274 B C Maxim us 3.902 A C Q -Syte 3.936 A C B ionector 3.481 C Invision 3.79 C SignificantG roups C onnectorD ays 3 and 4 4 3 2 1 0 6.6 6.5 6.4 6.3 6.2 6.1 6.0 5.9 D ay M ean log(C FU /connector) M icroclave Sm artSite C learLink Invision M axim us Q -syte O ne-Link B ionector Catheter Hub Catheter Tubing CONCLUSIONS LeastSquare M ean (Log C FU /segm ent) SignificantG roups M icroClave 0.845 A O ne-Link 1.14 A B C learLink 1.194 A B Maxim us 1.426 A B Invision 1.541 B Sm artSite 1.58 B B ionector 1.646 B Q -Syte 1.797 B C atheterSegm entD ays 3 and 4 Figure 2. For surface inoculation controls, the connector was swabbed in order to determine the concentration of bacteria on the connector septum. The least square mean for the destructive sampling of the catheter segments was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05). MicroClave® (ICU Medical Inc.) SmartSite® (CareFusion Corp.) ClearLink® (Baxter Inc.) Invision® (RyMed Technologies Inc.) Q-Syte (BD and Co.) Maximus® (MaxPlus) (CareFusion Inc.) One-Link® (Baxter Inc.) Bionector® (Vygon Inc.)

A National Science Foundation Engineering Research Center in the MSU College of Engineering Center for Biofilm Engineering INTRODUCTION Evaluation of Fluid

Embed Size (px)

Citation preview

Page 1: A National Science Foundation Engineering Research Center in the MSU College of Engineering Center for Biofilm Engineering INTRODUCTION Evaluation of Fluid

a National Science Foundation Engineering Research Center in the MSU College of Engineering

Center for Biofilm Engineering

INTRODUCTION

Evaluation of Fluid Path Colonization in Needle-Free Connectors and Biofilm Formation in Central Venous Catheters

E. deLancey Pulcini1, A.E. Parker1, M. Ryder2, G. James1

(1) Center for Biofilm Engineering Montana State University-Bozeman (2) Ryder Science Inc. Escondido, CA

Poster#275ASM Biofilms

2012

In the United States, approximately 80,000 catheter-related bloodstream infections (CRBSI) occur in ICUs annually1 and approximately 250,000 CRBSIs occur hospital wide annually2.

Needle-free connectors collected from patients have been shown to contain bacterial biofilms3. Previous studies have shown bacterial contamination and bacterial transfer through needle-free connectors4.

Needle-free connectors were developed for use with vascular catheters to prevent needle stick injuries to health care workers. The use of needle-free connectors has been implicated in catheter related bloodstream infections.

The objective of this study was to determine if there was a difference, by connector type, in the colonization of the fluid pathway in inoculated needle-free connectors and in attached vascular catheters after repeated flushing with no disinfection.

Figure 1. The connector-catheter sets were placed in conical vials between flushes. The technician is flushing one of the connector-catheter sets.

Figure 3. Image of a triple lumen catheter. For this project, single lumen catheters were used.

1. Mermel LA. Prevention of intravascular catheter-related infections. (Erratum: Ann Intern Med 133:395, 2000). Ann Intern Med 2000. 132:391–402.

2. Maki DG, Kluger DM, Crnich CJ. The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies. Mayo Clin Proc 2006. 81:1159–1171

3. Donlan RM, Murga R, Bell M, Toscano CM, Carr JH, Novicki TJ, Zuckerman C, Corey LC, Miller JM. Protocol for detection of biofilms on needleless connectors attached to central venous catheters. J Clin Microbiol. 2001. 39:750-3.

4. Ryder M, Fisher S, Hamilton G, Hamilton M, James G. Bacterial transfer through needlefree connectors:comparison of nine different devices. Presented at the Society for Healthcare Epidemiology of America Annual Scientific Meeting, April 2007

The connector septum was inoculated twice a day with 106 CFU Staphylococcus aureus ATCC # 6538. The inoculated connector was allowed to dry for 30 minutes and then was attached to a catheter.

Each connector-catheter set was flushed with 3.0 ml sterile saline which was collected and plated (First Flush). The catheter-connector sets were sterile saline (SS) flushed twice more, locked with sterile Brain Heart Infusion Broth (BHI) for 1 hour and SS flushed three more times. The last flush was also collected and plated (Last Flush).

The connector-catheter sets were inoculated a second time each day after the 6th sterile saline flush followed by a second round of flushing, plating and locks for a total of 18 connector accesses daily, considered to be a routine number of accesses in an intensive care unit.

The entire procedure, inoculation and flushing, was repeated each day for 5 days. On Days 4 and 5, two connector-catheter sets for each connector type were destructively sampled for bacterial counts and microscopy.

Statistical analysis was performed using mixed effect ANOVA analysis and Tukey’s tests to determine significant mean differences of bacteria in the flush, hub, catheter segment or connector amongst the different needle-free connectors. A multiple linear regression was used to determine if any combination of the variables measured in the study could significantly predict the log density of bacteria either in the flush.

METHODS RESULTS

REFERENCES

RESULTS

A total of 8 needle-free connectors were evaluated in this study. Three of each connector type were evaluated in three replicate runs (n=9) with the MicroClave® (ICU Medical Inc.) serving as the matched control for every run (n =27).

METHODS

ACKNOWLEDGEMENTS

This research was funded by ICU Medical.

RESULTS

Surface Inoculation

The mean log densities (LD) of the surface inoculations averaged across all days were statistically equivalent as long as mean differences as large as 0.37 were assumed to be negligible. For each day individually, the mean LD of the surface inoculum (averaged across the two inoculations for each day) were shown to be statistically equivalent as long as mean differences as large as 0.53 were assumed to be negligible.

43210

5.5

5.0

4.5

4.0

3.5

3.0

Day

me

an lo

g(C

FU

/flu

sh)

MicroclaveSmartSiteClearLinkInvisionMaximusQ-SyteOne-LinkBionector

Connector

Daily least square mean bacterial densities in the flush

The mean log density (LD) in the daily flush for the MicroClave connector was significantly smaller compared to any of the other connector types tested (p<0.0005).

The Q-Syte had the significantly largest mean LD of bacteria in the flush compared to any of the other connector types. (p <0.0005).

The MicroClave connector had the significantly smallest mean log density (LD) of bacteria in the flush compared to any of the other connector types (p<0.0005).

The mean LD of bacteria in the hub for the MicroClave connector was significantly smaller than the SmartSite, Invision, Q-Syte and Bionector connectors.

The mean LD of bacteria in the catheter segment for the MicroClave was significantly smaller that the SmartSite, Invision, Q-Syte and Bionector connectors.

The mean LD of bacteria in the MicroClave connector was significantly smaller than the SmartSite, Invision, and Bionector connectors.

The daily mean LD of bacteria attached in the hub and in the connector were better predictors of the daily mean LD of bacteria recovered from the flush than the mean LD of bacteria in the catheter segment.

The least square mean for all flushes for all days was calculated. The color scheme indicates the significant groups (p< 0.05).

Least Square Mean (Log CFU/flush)

Microclave 3.128 ABionector 3.637 BOne-Link 3.907 B CClearLink 4.176 C DSmartSite 4.176 C DInvision 4.368 D EMaximus 4.573 EQ-Syte 5.276 F

Significant Groups

Mean Flush for all Days and all Flushes

Least Square Mean (Log CFU/hub)

MicroClave 1.594 AOne-Link 1.694 A BClearLink 1.853 A B CMaximus 2.101 A B C DSmartSite 2.095 B C D EQ-syte 2.925 C E FBionector 2.905 D E FInvision 3.14 D F

Significant Groups

Hub Days 3 and 4

The least square mean for the destructive sampling of the catheter hubs was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05)

The least square mean for the destructive sampling of the connectors was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05).

Least Square Mean (CFU/connector)

MicroClave 2.544 AOne-Link 2.592 A BClearLink 3.005 A B CSmartSite 3.274 B CMaximus 3.902 A CQ-Syte 3.936 A CBionector 3.481 CInvision 3.79 C

Significant Groups

Connector Days 3 and 4

43210

6.6

6.5

6.4

6.3

6.2

6.1

6.0

5.9

Day

Me

an lo

g(C

FU

/co

nn

ect

or)

MicroclaveSmartSiteClearLinkInvisionMaximusQ-syteOne-LinkBionector

Catheter Hub

Catheter Tubing

CONCLUSIONS

Least Square Mean (Log CFU/segment) Significant Groups

MicroClave 0.845 AOne-Link 1.14 A BClearLink 1.194 A BMaximus 1.426 A BInvision 1.541 BSmartSite 1.58 BBionector 1.646 BQ-Syte 1.797 B

Catheter Segment Days 3 and 4

Figure 2. For surface inoculation controls, the connector was swabbed in order to determine the concentration of bacteria on the connector septum.

The least square mean for the destructive sampling of the catheter segments was calculated for Days 3 and 4 combined. The color scheme indicates the significant groups (p< 0.05).

MicroClave® (ICU Medical Inc.)

SmartSite® (CareFusion Corp.)

ClearLink® (Baxter Inc.)

Invision® (RyMed Technologies Inc.)

Q-Syte (BD and Co.)

Maximus® (MaxPlus) (CareFusion Inc.)

One-Link® (Baxter Inc.)

Bionector® (Vygon Inc.)