A method for 237Np determinationwith LSC in the experiment ofneptunium sorption onto bentonite

Wangsuo Wu, Ping Li, Keliang Shi

Radiochemistry LabSchool of Nuclear Science and Technology, Lanzhou University

[email protected] 2017 Denmark

School of Nuclear Science and Technology

Page 2: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

2 Radiochemistry Laboratory Lanzhou University

OUTLINE OF THE TALK

Summary

Results and Discussion

Introduction

Page 3: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

3 Radiochemistry Laboratory Lanzhou University

Introduction − Background237Np will be a major contributor to environmental

radioactivity from the disposal of high- or intermediate-levelradioactive waste because of its long half-life (t1/2 = 2.14×106

yr), suspected high radiotoxicity and low affinity for manygeological minerals. (usually present as the pentavalent NpO2

+

cation under several environmental conditions)

237Np

Mean Dose

Time (y)1E61E51E41E3

The so-called “key nuclide” in the HLW disposal repository

Page 4: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

4 Radiochemistry Laboratory Lanzhou University

Introduction − Background

However, determination of 237Np is rather a complicated

task because of its low specific activity and lack of a suitable

yield tracer such as high purity 236Np.

Natural barriers

Buffer and backfill

Metallic containers Waste

System of Multiple Barriers

Page 5: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

5 Radiochemistry Laboratory Lanzhou University

Introduction − Methods for determination of 237Np

At present, analytical techniques for determination of 237Npincludes:

(1) alpha spectroscopy,

(2) Neutron Activation Analysis (NAA),

(3) gama spectroscopy,

(4) 239Np tracer,

(7) Photon Electron Rejecting Liquid Alpha Spectroscopy (PERALS),

(5) Liquid Scintillation Counting (LSC) (6) ICP-MS,

and methods unusually used such as spectrophotometric method, luminescence methods, X-ray fluorescence spectroscopy (XRF), electrochemical methods, Mössbauer spectroscopy, etc.

However,

• Both alpha and NAA methods require tedious chemical separations of analytes and long counting times. Both methods suffer from potential interferences from the presence of high levels of uranium, thorium, polonium, etc., in environmental samples, as well as the influence of self-adsorption.

• 239Np tracer has a too short half-life (T1/2 = 2.36 d).

• In PERALS system, a process of extraction is needed.• Peak tailing of 238U and molecular interferences (e.g. 235U1H2+,

236U1H+) in the method of ICP-MS. • ……

Page 6: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

6 Radiochemistry Laboratory Lanzhou University

Introduction − Determination by LSC

LSC is widely used in the field of radionuclide metrology,because of its numerous advantages over conventional techniques:− Virtually 100% counting efficiency;

− simple sample preparation;

− Large sample capacity;

− high throughput-automatic, multiple sample counting;

− no disturbance from the self-adsorption;

− Low backgrounds and great sensitivity for alpha counting, etc.

Very proper for samples determination in batch sorption experiment.

Page 7: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

7 Radiochemistry Laboratory Lanzhou University

Introduction − Determination by LSC

However,

The β active daughter 233Pa (t1/2 = 27.00 d) has a radioactivity

2.89×107 times higher than 237Np.

To solve this problem:

The use of a pulse shape analysis (PSA) technique.

Page 8: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

8 Radiochemistry Laboratory Lanzhou University

Introduction − Determination by LSC

Theory:

-50 0 50 100 150 200 250 300

Inten

sity (

norm

alized

)

t (ns)

beta pulse alpha pulse

Fig. 1 Principal illustration of the shape of α- and β-pulses obtained in liquid scintillation measurements.

De-excite of alpha particles to the ground state slower than singlet states produced by beta particles.

Measurement of the pulse decay timeor length allows identification of theparticle which caused it and enablessimultaneous recording of pure alphaand beta spectra, respectively.

By optimizing the PSA setting, a high efficiency 98% for alpha counting and a discrimination of beta emitters of 99.9% could be achieved.

Page 9: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

9 Radiochemistry Laboratory Lanzhou University

Results and Discussion − PSA-level Optimization

Usually, a pure α emitter and a pure β emitter are needed. The bestpure α and β emitters should be the same nuclides as are present inthe sample. Otherwise, some alternative nuclides with the same energydistribution could be used practically. Nevertheless, in some cases suchas the determination of 237Np/233Pa, it is not convenient to obtain apure α emitter or a pure β emitter with the same energy distribution as233Pa.

Therefore, we set the optimum PSA by measuring the sample

itself of mixed α/β emitters as well as a background sample.

Page 10: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

10 Radiochemistry Laboratory Lanzhou University

Results and Discussion − PSA-level Optimization

The bulk of beta events are regarded as long pulses

0 40 80 120 160 200 240

0

5

10

15

20

25

Background Sample Mixed Sample

PSA-level

back

grou

nd co

untin

g ra

te (c

pm)

0

1000

2000

3000

4000

5000

6000

7000sam

ple counting rate (cpm)

The true alpha events are regarded as alpha pulses

Beta events start to spill into the alpha spectrum

38

Page 11: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

11 Radiochemistry Laboratory Lanzhou University

Results and Discussion − PSA-level Optimization

Byegẩrd J, Ramebäck H, Widestrand H. International Progress Report IPR-02-68. 2002

222Rn

This is similar to the measurement of 222Rn:

Page 12: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

12 Radiochemistry Laboratory Lanzhou University

Results and Discussion − PSA-level Optimization

Reasonable?

0 10 20 30 40 50 60 70 800

100

200

300

400

Coun

ts

Channel0 20 40 60 80

0

200

400

600

800

Coun

tsChannel

deteced alpha counts deteced beta counts

Without PSA-level Use PSA-level=38

Page 13: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

13 Radiochemistry Laboratory Lanzhou University

Results and Discussion − Validation of the selected PSA-level

If the PSA-level is set higher than the optimum, many of the pulses originatingfrom β-activity will be falsely accepted as α-decay.

The misclassified β events of 233Pa into α-MCA will greatly influence theresults of neptunium sorption experiment since Pa5+ exhibits very different sorptionability to NpO2

+.

The main Gaussianpeak is compounded ofpeak from Eα = 4766keV to 4817 keV andrepresents 85.5% of thetotal α emission.

Use the mathematic method of Aupiais et al. to validate:

Pure αspectrum

We analyzed the corresponding peaks.

Page 14: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

14 Radiochemistry Laboratory Lanzhou University

0.0 0.5 1.0 1.5 2.0 2.5 3.0

0

1000

2000

3000

4000

mathematical treatment PSA level=38

Coun

ting

rate

(CPM

)

m (tracer) /g

y=1445.49xR2=0.9999

y=1438.15xR2=0.9994

Results and Discussion − Validation of the selected PSA-level

APSA-level=38, the detected α emissionaccounts for ~86% of the total 237Np αemission and ~14% of α emission ismisclassified into β-MCA.

The selected PSA-level is slightlylower than the optimum.We can then obtain the optimumPSA-level by increasing PSA-leveluntil the exact value of total 237Npα emission is reached.

A series of samples are determined at PSA-level=38 as well as analyzed by themathematic method. The results are compared:

Page 15: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

15 Radiochemistry Laboratory Lanzhou University

Results and Discussion − Validation of the selected PSA-level

However, the PSA-level is not further optimized on account of the fact thatin the experiment of neptunium sorption onto bentonite, sorptionpercentages of neptunium can be obtained by:

where CPMt is the total α emission of the sample in batch sorptionexperiment; CPMs is the α emission in supernatant.

Inferring from the equation, the calculated sorption percentage doesnot change while ~86% of true CPMt and CPMs is detected, i.e.the misclassified ~14% of α emission has no influence on the sorptionresults.

% t s

t

CPM CPMSorptionCPM−

=

PSA-level=38 is suitable for the determination of 237Np samples.

Page 16: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

16 Radiochemistry Laboratory Lanzhou University

Results and Discussion − Quenching of bentonite

Influence of bentonite suspension on LSC measurement at PSA-level=38.

Almost ~50% of 237Np(V) isadsorbed on betonite underthis condition, but no obviousinfluence was observed, themeasurement was notinfluenced.

0.0 0.5 1.0 1.5 2.0 2.5 3.0

0

1000

2000

3000

4000

5000

m/V=0.2g/L pH=2.0 m/V=1.0g/L pH=2.0 m/V=1.0g/L pH=8.8 m/V=10.0g/L pH=8.5

Coun

ting r

ate (C

PM)

m (tracer) /g

Solid to Volume ratio

(g/L)

Solution pH

Linear relationship between the detected α emission and 237Np tracer content at PSA-level=38

R2

0.2 2.0 y=1391.96x 0.99971.0 2.0 y=1377.03x 0.99961.0 8.8 y=1389.98x 0.9999

10.0 8.5 y=1342.53x 0.9995

Therefore, suspension samples can be directly determined. In conclusion, the second mode is more suitable for the determination of batch experiment samples.

Page 17: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

17 Radiochemistry Laboratory Lanzhou University

Summary

• 237Np samples were well measured on Wallac Guardian 1414 at PSA-level = 38. At this PSA-level, practically all pulses from β-activity of 233Paare discarded and ~86% of total α emission was detected.

PSA-level Optimization

• Bentonite solid particles in sample have no influence on 237Np samplemeasurement by LSC. As a result, sorption percentage can be calculatedby counting suspension and equivalent supernatant from the samesample.

Determination Modes

• The numerical setting of the PSA-levels works on the Wallac Guardian1414 is defined in a different way than the corresponding settings onsome other LSC such as Wallac Rackbeta 1219 (Chalmers). Therefore,the selected PSA-level in this study cannot be directly applied to 237Npmeasurement on some other LSCs.

Notice

Page 18: A method for Np determination with LSC in the experiment ...lsc2017.nutech.dtu.dk/wp-content/uploads/9-Wangsuo-Wu-LSC2017-139.pdf · A method for 237Np determination with LSC in the

Thanks for your attention!