1 Motion Correction on 3D PET/CT Data with Optical Flow Alg A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data M Dawood 1,2 , C Brune 2 , F Büther 1 , KP Schäfers 1 European Institute for Molecular Imaging Department of Computer Science, University of Münster, Germany

A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

  • Upload
    matteo

  • View
    30

  • Download
    0

Embed Size (px)

DESCRIPTION

A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data. M Dawood 1,2 , C Brune 2 , F Büther 1 , KP Schäfers 1 European Institute for Molecular Imaging Department of Computer Science, University of Münster , Germany. Cardiac Motion and Partial Volume. - PowerPoint PPT Presentation

Citation preview

Page 1: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

M Dawood1,2, C Brune2, F Büther1, KP Schäfers1

European Institute for Molecular Imaging Department of Computer Science,

University of Münster, Germany

Page 2: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

2

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Cardiac Motion and Partial Volume

Coronal slice through non-attenuated PET1 h. p. i., 18FDG, CHD patient

Cardiac motion

Ungated dataLarge blur, low noise(Problem in plaque imaging)

One phaseSmall blur, high noise

Page 3: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

3

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Step 1: Gating

ECG signal

Series of images reconstructed from cardiac gated PET acquisition

Page 4: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

4

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Mass conservation in cardiac data

Systole Diastole

Mass Conservation

Page 5: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

5

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Step 2: Motion Estimation

Page 6: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

6

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

All gates deformed to Diastole

Visual result

Cardiac phases

Page 7: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

7

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Ungated dataLarge blur, low noise Noise 25

One phaseSmall blur, high noise Noise 36

All phases motion correctedSmall blur, low noise Noise 22

Visual result

Page 8: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

8

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Quantitative results on patient data

Data:

14 patients with known CHDca. 4 MBq/Kg body weight 18F-FDGScan time ca. 15 minutes, 1:15 hours post injectionListmode acquisition on Siemens Biograph 16 scanner

Quatification methods:

Correlation of ROI (40x40x40) with target phaseMyocardial thickness. FWHM of Gaussian fit to line profileMean activity in blood pool in LV

Page 9: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

9

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Quantitative results 1: Correlation of end-systolic gate with target phase

1 2 3 4 5 6 7 8 9 10 11 12 13 140.700000000000001

0.750000000000001

0.800000000000001

0.850000000000001

0.900000000000001

0.950000000000001

1

BeforeAfter

Patient No

Corr

elati

on c

oeffi

cien

t

Page 10: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

10

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Quantitative results 2: Myocardial thickness

1 2 3 4 5 6 7 8 9 10 11 12 13 14 Avg3

3.5

4

4.5

5

5.5

6

Patient Number

Myo

card

ial t

hick

ness

[mm

]

End-systoleEnd-diastoleEnd-systole after MC

Page 11: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

11

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Quantitative results 3: Mean activity in blood pool

Page 12: A Mass Conservation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data

12

Motion Correction on 3D PET/CT Data with Optical Flow Algorithms

Thank you

To conclude:

A method for cardiac motion and partial volume correction was presented.

The results on patient data show that the motion was corrected precisely.