Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
For:• biomaterials• semiconductors• ceramics• polymers• forestry / pulp & paper
• catalysis• environmentalstudies
• minerals• metallurgy • and more
A core resource facility allowing researchers ready access to:
• state-of-the-art equipment • the related expertise
Providing:• surface analytical services• a means for collaborative research
Surface Interface Ontario
Why the surface?• Boundary layer between the solid at its interface with its environment
• Understanding this interface is of fundamental and practical importance to many disciplines.
Modern methods of surface analysis allow for very detailed characterisation of the surface.
•Elemental Information- type and quantification
•Chemical Information- surface chemistry and elemental state
•Distribution- both lateral and into the bulk
Spectroscopic Methods
x, y, z
rotn.
+ tilt
SOURCE ANALYSER
Penetration depthof 1 keV particles
Photons - 1000 nmElectrons - 2 nmIons - 1 nm
Thus either the source or analysed particle will usually include an electron or ion for surface specificity
Beam OutBeam In
Photons Electrons Ions
Photons FTIR
Laser Raman
X-RayFluorescence
EXAFS
XPSUV PhotoectronSpectroscopy
SEXAFS
Laser Mass
Spectroscopy
Electrons ElectronMicroprobe
EDAX
AugerTEM / SEM
LEED
EELS
Electron-induced IonDesorption
Ions IonMicroprobe
SIMSIon ScatteringSpectroscopy
RBS
Surface Science UnitCentre for Biomaterials• Established in 1989• Funded by OCMR• Precursor of SI-Ontario
XPS – Leybold MAX200• Dual anode (Mg/Al) Kα• Monochromatic Al Kα• Flood gun (ineffective) • Analysis area (defined by aperture)
– 200 µm, 500 µm, 1 mm2, 2 x 4 mm2, 4 x 7 mm2
• Ion gun – depth profiling/cleaning
X-ray Photoelectron Spectroscopy (XPS)[Also known as Electron Spectroscopy for Chemical Analysis]
Principles• Incident X-rays cause photo-emission of electrons from the surface - energy analysed
• Binding energies are characteristic of each element -can be used for identification
• Binding energy of a particular electron affected by the atom’s environment - chemical information via the “chemical shift”
Information• Elements Li and up, on both insulating and conducting samples
• 0.1 - 1 at.% detection limit, straight-forward quantification (to ±5%)
• “Chemical-shifts” – bonding information
• Surface specific (2 - 10 nm)
• Angle-resolved XPS - non-destructive depth information (to 10 nm)
• Ion sputtering - deeper depth profiles
• High spectral resolution with monochromatic source
• Mapping capabilities – when utilising small analysis spotEB = hν - EK + Φ
The binding energies are characteristic of each element
- use for elemental identification
- peak area allows quantification
Survey spectrum of dentin from a cow’s tooth C 1s spectrum of PMMA obtained using monochromatic Al KαX-rays
note 3:1:1 C ratio
Chemical information via the chemical shift – thus information on
- chemical bonding of the element
- its particular oxidation state- this is also quantifiable
Angle-Resolved XPS (ARXPS)Non-destructive depth information (to ~10 nm)
Surface selective modification of fluoropolymers by Al deposition
McKeown et al. Langmuir 7 (1991) 2146
IB = IB 0 (1 - exp ( - dB / λBB sinθ ))
IA = IA 0 exp ( - dB / λAB sinθ ))
Making use of: I = I 0 exp ( - d / λλλλ sinθθθθ )
Models can be built to estimate amount and type of coverage. Thus:
B
A
θ
A
B
dB
Or, for partial coverage:
IA = IA0 . [ (1-f) + f . exp ( -dB / λB sinθ ) ]
IB = f . IB0 . [ (1 - exp ( -dB / λB sinθ ) ]
Applying this to the previous example:Depth of modification – 3 nm, with 89% degree of modification (“coverage”)
More details on this in John Wolstenholme’s talk
Estimate of overlayer thickness0.2% 0.8 nm
1.0% 1.8 nm
HA
CHX
0.2%
1%
Can apply to films/coatings
Retention of Antibiotics on Surfaces
Chlorhexidene (CHX)0.2 and 1% solutions on hydroxyapatite (HA)
P – unique to substrate
N – unique to adsorbate
Sodhi et al. J. Dental Res. 71 (1992) 1493
• Assume uniform overlayer• ARXPS not performed since “rough”– meaningless to tilt
Sputter Depth Profiling – for thicker layers
Evaluating Sol-Gel Ceramic Thin Films for Metal Implant Applications : III. In Vitro Ageing of Sol-Gel Derived Zirconia Films on Ti6Al4V
P. B. Kirk et al, Applied Biomaterials (1999)
Sol-Gel zirconia films deposited on Ti6Al4V and aged in Hanks’balanced salt solution. Precipitate, predominately Calcium Phosphate, formed upon aging.
Ar+ ions used to sputter profile through coating.
Over 12 h for profile
Profile
O 1s in Zr region – note: energy scale not corrected for charging
Surface Interface OntarioDepartment of Chemical Engineering & Applied Chemistry
• 2002• Following successful a CFI application
ToF-SIMS – ION-TOF IV• Liquid Metal Ion Gun (Ga)• Dual Column Source (Cs, EI – (Ar, O2, SF6))
• Dual beam depth profiling • Effective charge compensation • Variable temperature (-130°C to +600°C)• Preparation / sample treatment chamber
XPS – Data Acquisition Unit + Software (Specs)
Principles• Pulsed primary ion beam
• Sputters off ∼ 0.1% of surface (static-SIMS)
• Secondary ions (& neutrals) emitted
• Secondary ions mass analysed by
time-of-flight
• Obtain composition, distribution &
molecular information of the surface
Time-of-Flight Secondary Ion Mass Spectrometry
Collision Cascade
Energy transferred back to surface –atoms & molecular fragments ejected
Whilst most secondary particles come off neutrally charged, a small proportion come off as either positive or negative ions
Why Time of Flight?• Parallel detection of ions - sensitivity
• 1 - 2 monolayer analysis• High mass range (up to 10 k Daltons)
• Excellent mass resolution(> 9,000)
• Effective charge compensation
Other Advantages• Elemental and chemical analysis • PPM sensitivity• High spatial resolution (100 nm) • SEM images possible (60 nm) without coating of insulators.• Depth information
- by profiling (dual beam – provides “dynamic SIMS” information)
- imaging of cross-section
Ga - main work horse, various modes• high mass resolution mode• high spatial resolution mode (200 nm), collimated mode (100 nm)• combination mode (250 nm)
Cs - enhances yield of electronegative ions• implantation reduces work function allowing 2º electrons to overcome this• use as sputter source for cleaning or profiling for negative species
O - enhances yield of electropositive ions• high electron affinity - metal more likely to give up electron• explains great increase (as much as 104) in M-O signal over M• equivalent can be accomplished by O2 bleed
Ar - neutral• use as sputter source for cleaning or profiling for positive species
SF6 - heavier fragment yields + monolayer depth resolution• higher mass + polyatomic species
Imaging from macro to micro – insulators & conductors
Ultimate Spatial Resolution
100 nm (SIMS) 60 nm (SEM)
Co-poly ferrosilanemicrobars on Si -nominally 0.5 x 3 µm2
Manner’s group - Chemistry
Cross-Section of Plant Membrane StructureCooper’s group - Forestry
Line scan showing Br penetration
Depth Information
AlH
Si 27.977
C2H428.032
CH2N
InGaAs\GaAs- Ga acquisition;
SF6 sputter
Dual Beam - interlaced mode - use of SF6 allows shallow profiling with improved
resolution
28
Mass Resolution (28 amu) >8000
Other Advantages
Molecular Information
Excellent mass resolution & sensitivity
InCs+
Compare with Cs sputter
mass / u180 200 220 240 260
mass / u300 320 340 360 380 400 420
mass / u440 460 480 500 520 540 560 580
90° 20°C O C O
Sample 1 66.6 33.4 70 30Sample 2 66.6 33.4 70 30
Sample 1 – 90° Sample 1 – 20°
Sample 2 – 90° Sample 2 – 20°
No difference in composition from survey spectra and little difference in C 1s envelope
However, ToF-SiMS shows a difference – presence of fatty acids in Sample 1
Need for Complementary Techniques
Two different paper samplesSample 1 – bad ink transfer
-1000 -900 -800 -700 -600 -500 -400 -300 -200 -100 00
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
Binding Energy (eV)
Inte
nsity
(arb
. uni
ts)
315
1A2-4GG/GEL2/72HOURS
ToF-SIMS confirms presence
Cl?
Adsorption of Antibiotic on TeethAdsorption of Antibiotic on Teeth
• Recall CHX on HA – presence confirmed by XPS
• More difficult on root canal• N & C both present in substrate• Cl signal too weak
CCa
O
N
P
Analysis Parameters:
Energy:Current:Area:
Sample Parameters:
Sample:Origin:
Polarity:
BCG1 su c6934
positiveFile: G1PR_1P.tfd PIDD: Unknow
99.6x99.6 µm²Unknow25 keV
Sputter Parameters:
PI:Energy:Current:Area:PIDD: 1.57E+017 Ions/cm²
300.0x300.0 µm²6.40 nA1 keV
Comments:
Ga+ Ar+PI:
Note: left side close to crater edge - thus ROI - rhs
TOF-SIMS IV
Depth / nm20 40 60 80 100 120 140 160 180 200 220 240
110
210
310
410
510
Inte
nsity
Substance Mass Color
Na 22.99
Al 26.98
C2H3 27.02
Si 27.98
K 38.96
Ca 39.9658Ni 57.9358NiO 73.93
Au 196.97
1 2 3 spectra extracted from these points
Can obtain depth profiles and complete mass spectra for each point
Depth scale calculated assuming literature sputter parameters
Na K 58Ni
mass / u10 20 30 40 50 60 70 80
5x10
0.2
0.4
0.6
0.8
Inte
nsi
ty
mass / u90 100 110 120 130 140 150 160
3x10
1.02.03.04.05.0
Inte
nsi
ty
Au
mass / u170 180 190 200 210 220 230 240
3x10
0.5
1.0
1.5
Inte
nsi
ty
mass / u250 260 270 280 290 300 310 320
3x10
0.20.40.60.81.01.2
Inte
nsi
ty
mass / u330 340 350 360 370 380 390 400
2x10
0.5
1.0
1.5
2.0
Inte
nsi
ty
Na
mass / u10 20 30 40 50 60 70 80
4x10
0.51.01.52.02.5
Inte
nsi
ty
mass / u90 100 110 120 130 140 150 160
3x10
0.2
0.4
0.6
0.8
Inte
nsi
ty
mass / u170 180 190 200 210 220 230 240
2x10
1.0
2.0
3.0
4.0
Inte
nsi
ty
mass / u250 260 270 280 290 300 310 320
2x10
0.51.01.52.02.5
Inte
nsi
ty
mass / u330 340 350 360 370 380 390 400
2x10
0.51.01.52.02.5
Inte
nsi
ty
mass / u10 20 30 40 50 60 70 80
4x10
0.5
1.0
1.5
Inte
nsi
ty
mass / u90 100 110 120 130 140 150 160
3x10
0.20.40.60.81.0
Inte
nsi
ty
Au
mass / u170 180 190 200 210 220 230 240
2x10
0.5
1.0
1.5
2.0
Inte
nsi
ty
mass / u250 260 270 280 290 300 310 320
1x10
2.0
4.0
6.0
Inte
nsi
ty
mass / u330 340 350 360 370 380 390 400
2x10
1.0
2.0
3.0
4.0
Inte
nsi
ty
Point 1
Point 2
Point 3
Surface Interface OntarioExpansion of Lab Facilities
• 2007• Following successful a CFI-LEF application
ToF-SIMS – ION-TOF IV• Upgraded – 2 new ion sources (February 2008)
XPS – 2 new instruments• Thermo Scientific K-Alpha (August 2007)• Thermo Scientific Theta (March 2008)
Other Equipment• Surface Profilometer - KLA Tencor P16+ (August 2007)• Target Sectioning System – Leica EM TXP (December 2007)• Cryo-Ultramicrotome – Leica EM UC6/FC6 (~ May 2008)• Vacuum/Cryo “Suitcase” – Thermionics (~ June 2008)• Preparation/Reaction Chambers – Old MAX (ongoing)
ION-TOF ToF-SIMS IV
• Multiple Ion Sources• Liquid metal ion gun (Bin) - rapid submicron imaging, cluster source (to Bi7) - enhances fragmentation and yield
• Third ion gun column - C60 source - enhances fragmentation and yield, shallow & molecular
depth profiling
• Dual Beam Depth Profiling • Cs / EI / C60 for sputtering, Bi, Cs or EI for spectral acquisition
• Variable Temperature• Temperature stage (-150°C to +600°C)
• Pulsed secondary electron detector• (SEM on insulators)
• Updated software control
Use of heavier & polyatomic fragments - Aum
New development (2002) (superseded by Bim (Sept. 2004))
Different clusters (Au, Au2, Au3) selectable.
Unlike SF6, keeps the imaging capabilities of a LMIG
No features discernable with Ga
More on this from Albert Schnieders
~ 19x enhancement with Au3+ over Au1
+
Note this type of work is totally impossible with Ga gun
Cholesterol
PhospholipidsDiglycerides
Au3+ - mouse brain slice - no treatment
195 360 390
450 550 680 900
Tissue Molecular Ion Imaging with Au Clusters
Touboul et al. Analytical Chemistry 76 (2004) 1550
Peter Sjövall will probably have more examples
Use of heavier & polyatomic fragments – C60
• Enhances fragmentation and yield
• Shallow depth profiling
Tt=29 ps
•Molecular depth profiling possibilities on some systems
Larger volume is altered for Ga15x more material removed with C60
t = 29 ps15 keV Ga 15 keV C60
Barbara Garrison’s Group – Penn State: Anal. Chem., 75, 4402-4407 (2003)
• Automated, compact XPS• Monochromated Al Kα X-ray source• Variable spot - 30 - 400 µm• Effective charge compensation• Ar+ ion gun for depth profiling, Zalar rotation possible
• 128 channel detector - rapid spectral acquisition
• Large area sample platen for mounting sample(s)
• Ease of selection of analysis area• Wedges (20° & 30°) allow ARXPS to be performed
• All aspects digitally controllable and capable of automation, including data work-up and report generation
• High volume, rapid throughputS-15S-14S-13
S-12S-11S-10
S-9S-8S-7
S-6S-5S-4
S-3S-2S-1
Thermo Scientific K-Alpha
0.00E+00
1.00E+05
2.00E+05
3.00E+05
4.00E+05
020040060080010001200
Cou
nts
/ s
Binding Energy (eV)
Survey1 Scan, 1 m 8.1 s, 400µm, CAE 200.0, 1.00 eV
C1s
F1s
N1s
O1s
Si2p
3.0940255.502.53102.220.817Si2p
17.94669633.373.02532.352.930O1s
1.9448203.722.95399.941.800N1s
1.3266877.692.21689.334.430F1s
75.701107556.572.66285.031.000C1s
At. % Area (P) CPS.eVFWHM eVPeak BESF Name
PHEMA : 4.7kDA HA(5g/L)
Experiment\X-Ray000 400um - FG ON\S-6
0.00E+00
2.00E+03
4.00E+03
6.00E+03
8.00E+03
1.00E+04
1.20E+04
1.40E+04
1.60E+04
1.80E+04
2.00E+04
280282284286288290292294296298
Cou
nts
/ s
Binding Energy (eV)
C1s Scan #210 Scans, 400µm, CAE 20.0, 0.10 eV
C1s
C1s A
C1s BC1s C
C1s D
C1s F
2.921008.931.02289.531.000C1s F
18.456383.931.12285.671.000C1s D
4.321494.351.14287.711.000C1s C
4.001381.321.08288.891.000C1s B
12.474314.781.10286.701.000C1s A
57.8420021.521.19285.031.000C1s
At. % Area (P) CPS.eVFWHM eVPeak BESF Name
Survey and high resolution XPS on polymer sample
• High-end, small-spot XPS• Monochromated Al Kα X-ray source• Variable spot - 15 - 400 µm• Dual anode (Mg/Ag) nonmonochromated X-ray source
• Effective charge compensation• Ar+ ion gun for depth profiling, Zalar rotation possible
• Parallel ARXPS (60° range)• 2-d multi-channel detector
110 channels for energy, 96 channels for angle
• Large area sample platen• Special mount for heating/cooling
(-150°C to +600°C)
• Field emission gun for Auger/SEM (95 nm resolution)
• Multi-port preparation chamber with heating/cooling, sample parking and port for vacuum/cryo “suitcase”
Thermo Scientific Theta Probe
(See poster by Joyce Koo)
Examples
ARXPS – ATRP on Si (Krull’s group – UTM)
Contamination on Au wire – coating thickness
Snapshot Imaging – Spectrum Mapping
NiCr Alloy
(Coyle’s Group
U. Of T.)
Depth Profiling GaAs multi-stack layer
3 keV Ar+, 3 µA, 3 x 3 mm2 - rotated
Ruda’s Group – U. of T.
ToF-SIMS Dynamic SIMS Auger XPS
ElementalSensitivity XXXX XXXXX XX XX
BondingInformation XXX X XX XXXXX
MolecularInformation XXXXX - - -
SpatialResolution XXXX XXXX XXXXX XX
Depth ProfilingSpeed XXXX XXXXX XXX XX
Primary Organics Inorganics Inorganics OrganicsApplication & Inorganics & Inorganics
The following table summarises some of this information for the most common techniques
Other EquipmentKLA-Tencor P16+ Surface Profilometer• 2-d line profiles• 3-d mapping• Low force option• Extended (1 mm) range (z)• Long scan (200 mm) length (x)• 20 site sequencing (line/map) capability
Leica EM UC6/FC6 Cryo-Ultramicrotome• Sectioning of biological and other samples at temperatures from -15° to -185°C
• Virtual connection to ToF-SIMS, XPS and preparation chambers via side port for vacuum/cryo “suitcase”
Leica EM TXP Target Sectioning System• Stereo microscope to easily view target area• Ease of use. Sawing, milling, grinding and polishing exactly to target - no need to remove sample
Vacuum/Cryo “Suitcase”• Virtual connection of preparation chambers, ToF-SIMS, Thetaprobe XPS and cryo-ultra microtome
Preparation/Reaction Chambers
• SI-Ontario allows scientists and engineers ready access to modern surface analytical equipment andthe related expertise
• Leading edge equipment in major areas of surface analysis
• Over the years this core-resource has served both academia and industry – truly
“creating partnerships for innovative research”
Acknowledgements
Surface Interface Ontario was made possible through funding provided by: Canadian Foundation for Innovation, Ontario Research Fund (formerly Ontario Innovations Trust), University of Toronto, McMaster University, Environment Canada, Celestica, VALE INCO, Novelis, Amtel and ITL.
Further support was received from several departments, both at Toronto and McMaster, via faculty participation.
This Workshop was made possible with donations from OCE, SFR, Datacomp, Ion-TOF USA, Thermo Scientific, Leica & KLA-Tencor