89
9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 1 of 4 [Customer Version] 1. Objective: This report details the correlation result of the 9RS08KA8(TSSOP 16, PDIP 16) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN- FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN- FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility. Note: 9RS08KA8 already qualified package SOIC 20W in KES-CHINA. Please refer to enclosed file for detail. G9DTKA82_Correlat ion_Report_to_Cus 2. General Information * Correlation Lot Information: Lot # Qty. Engineer lot 30 3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G. Correlation Plan Name Detail Description Qual per Correlatio n Plan, (Y / N ?) Acceptance Criteria c Parametric Comparison Goal: Parametric comparison for Identify Parametric Tests Lot quantity: 1 Units per lot: 30 Temperature: 25C Type of units: good Site: FSL-TJN-FM/ KES-CHINA Y t-test and f-test should not demonstrate significant difference. P-Unit of Measure should be greater than 0.05. 3.2 General Method: Using the production test program, perform 100% screening of 30 units at 25C temperature, in the current FSL-TJN-FM. Test the same 30 units under the same conditions in KES-CHINA.

9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9RS08KA8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 4

[Customer Version]

1. Objective:

This report details the correlation result of the 9RS08KA8(TSSOP 16, PDIP 16) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN- FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility.

Note: 9RS08KA8 already qualified package SOIC 20W in KES-CHINA. Please refer to enclosed file for detail.

G9DTKA82_Correlation_Report_to_Customer.doc

2. General Information

* Correlation Lot Information:

Lot # Qty.

Engineer lot 30

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

3.2 General Method: Using the production test program, perform 100% screening of 30 units at 25C temperature, in the current FSL-TJN-FM. Test the same 30 units under the same conditions in KES-CHINA.

Page 2: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9RS08KA8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 4

Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results: 4.1 Parametric Comparison:

Tables below summarize the parametric data on 30 good units The following are significant parameters chosen for this comparison:

• PORTA VOL LV

• IDD stop off

• PMC vregon bg

PORTA VOL LV Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(mV) (mV)

Pat#

(mV) (mV)

1 206.6981 203.7977 16 201.668 206.8748

2 195.9585 196.6969 17 199.6893 206.7129

3 196.5364 199.7323 18 196.7448 199.8295

4 206.5939 208.1154 19 202.9653 206.9792

5 203.8029 208.0231 20 196.4088 197.2317

6 196.9012 199.8364 21 203.6426 207.5011

7 206.5417 202.4905 22 199.6109 204.0582

8 198.8293 203.954 23 196.4273 197.4798

9 196.0106 196.7491 24 207.3437 213.0163

10 196.9295 199.8885 25 206.8023 208.6886

11 203.6426 207.9069 26 196.5129 197.5842

12 204.0925 208.0231 27 207.9165 216.3503

13 197.1378 199.8885 28 196.1716 197.1145

14 197.3981 202.0729 29 216.8206 218.3575

15 207.2898 210.0497 30 223.9544 222.1857

Page 3: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9RS08KA8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 4

IDD stop off Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 2.2929 2.4668 16 2.3797 2.4965

2 2.546 2.4555 17 2.8196 2.7971

3 2.5255 2.7703 18 2.6221 2.8113

4 2.6082 2.7697 19 2.4292 2.5076

5 2.3572 2.48 20 2.4489 2.3469

6 2.2074 2.3693 21 2.2053 2.3679

7 2.4124 2.3947 22 2.4483 2.5666

8 2.6439 2.4405 23 2.5668 2.3644

9 2.4157 2.5054 24 2.5048 2.715

10 2.615 2.7854 25 2.6332 2.7835

11 2.6716 2.8562 26 2.6598 2.4838

12 2.646 2.455 27 2.3833 2.5024

13 2.2279 2.4027 28 2.5514 2.459

14 2.5457 2.4483 29 2.4801 2.5875

15 2.4378 2.4115 30 2.4343 2.5317

PMC vregon bg Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 1.1913 1.197 16 1.2022 1.197

2 1.1691 1.1692 17 1.1911 1.1914

3 1.1858 1.1914 18 1.1802 1.1859

4 1.1858 1.1914 19 1.1746 1.1803

5 1.1966 1.197 20 1.1911 1.1914

Page 4: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9RS08KA8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 4

6 1.1969 1.197 21 1.1802 1.1859

7 1.1858 1.1914 22 1.1913 1.1914

8 1.1969 1.197 23 1.1799 1.1803

9 1.1746 1.1692 24 1.2025 1.197

10 1.1858 1.1914 25 1.1799 1.1803

11 1.1858 1.1914 26 1.1802 1.1803

12 1.1969 1.197 27 1.1802 1.1803

13 1.1969 1.197 28 1.1802 1.1803

14 1.1858 1.1914 29 1.2025 1.2026

15 1.1969 1.197 30 1.1913 1.1914

25C PORTA VOL LV IDD stop off PMC vregon bg

t- test (P-value) 0.124 0.182 0.523 f- test (P-value) 0.975 0.696 0.758

Results: Pass

5.0 Analysis Conclusions and Verification Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 26 Jun

2010

Ji Chang (KES-CHINA, TPE)

Wu Xiangyuan (TJN MSG PE)

Page 5: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC/AW16/32/60 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 7

[Customer Version]

1. Objective: This report details the correlation result of the 9S08AC/AW/16/32/60 (QFN 48) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

Correlation Lot Information:

Lot # Qty.

LJMEA0N4PU01 1208

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C,-40C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 25C, 135C,-40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C,-40C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 6: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC/AW16/32/60 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 7

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C and -40C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: � Tables below summarize the correlation results. All numbers filled in green boxes means

correlation was achieved. � All numbers filled in yellow boxes means results did not correlate and requires further

explanation.

Bin-to-Bin Correlation Matrix for Room & Cold Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FCN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

Page 7: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC/AW16/32/60 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 7

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Room 25C Hot 135C Cold -40C

Lot # Pass # Fail # Pass # Fail # Pass # Fail

TJMEA0N4PU01 100% 0 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison: For 9S08AC:

• sidd2_fbe_AllOff(FRN)

• stop3_rti_hv

• vreg_on_vddtest_hv(FCN)

sidd2_fbe_AllOff Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

1 857.654 844.36 16 922.548 919.307 2 877.160 877.421 17 831.390 812.320 3 856.455 834.863 18 821.701 802.914 4 907.657 902.154 19 883.637 885.590 5 863.544 855.031 20 901.996 894.803 6 982.000 939.829 21 855.792 825.161 7 914.669 906.447 22 920.916 909.965 8 1007.100 964.537 23 838.173 812.780 9 832.640 812.55 24 924.614 920.583

10 869.485 866.522 25 866.425 862.613 11 874.763 869.608 26 821.012 798.625 12 930.912 921.477 27 986.600 944.027 13 772.488 767.643 28 908.090 904.069 14 812.827 778.256 29 899.115 894.065 15 839.830 822.379 30 948.090 939.115

Page 8: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC /AW Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 7

vreg_on_vddtest_hv Parametric Comparison FCN

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

1 2.5133 2.5648 16 2.5472 2.5729

2 2.5418 2.5160 17 2.5459 2.5702

3 2.5215 2.5675 18 2.5770 2.5472

4 2.5486 2.5580 19 2.5025 2.5173

5 2.5269 2.5323 20 2.5404 2.5512

6 2.5838 2.5431 21 2.5486 2.5458

7 2.5270 2.6122 22 2.5418 2.5160

8 2.5364 2.5607 23 2.4903 2.5675

9 2.5201 2.5580 24 2.5513 2.5539

10 2.5242 2.5783 25 2.5255 2.5282

11 2.5787 2.5905 26 2.5876 2.5648

12 2.5581 2.5621 27 2.5418 2.4835

13 2.5976 2.6014 28 2.5255 2.5512

14 2.5581 2.5146 29 2.5201 2.5458

15 2.5662 2.5743 30 2.5981 2.5946

stop3_rti_hv Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(ms)

FSL-TJN

Unit of

Measure

(ms)

Pat# KES-

CHINA

Unit of

Measure

(ms)

FSL-TJN

Unit of

Measure

(ms)

1 7.3466 7.4618 16 8.2735 8.2128 2 7.9716 7.7959 17 7.0466 7.0963 3 7.4423 7.3654 18 7.3466 7.0763 4 7.1415 7.2086 19 8.3051 7.688 5 6.8732 6.9004 20 7.4399 7.2147 6 7.0417 6.9732 21 6.3487 6.914 7 7.3451 7.2084 22 7.5466 7.4029 8 7.9761 8.0217 23 7.6477 7.3321 9 7.3225 7.2352 24 8.9716 8.7302

Page 9: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC /AW Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 7

10 7.3466 7.3916 25 7.3533 7.0737 11 7.1433 7.141 26 7.4466 7.6357 12 7.7456 7.7636 27 7.9716 7.7997 13 7.3866 7.6543 28 8.9716 8.2184 14 8.1719 8.2315 29 7.2133 6.9877 15 7.9725 7.9612 30 7.9716 8.1445

FCN

Pat# KES-

CHINA

Unit of

Measure

(ms)

FSL-TJN

Unit of

Measure

(ms)

Pat# KES-CHINA

Unit of

Measure

(ms)

FSL-TJN

Unit of

Measure

(ms)

1 6.9016 7.5725 16 8.2799 7.9587

2 7.5035 8.0412 17 6.9727 7.213

3 7.2341 7.2471 18 7.2287 7.4116

4 7.6692 7.858 19 7.9215 8.0224

5 7.3567 7.5851 20 8.2543 8.0878

6 8.3567 8.3689 21 7.8959 7.4634

7 8.2543 8.3705 22 8.2031 7.9568

8 8.3055 7.7978 23 7.2287 7.3791

9 8.3311 8.5351 24 8.1519 7.8769

10 7.2799 7.6413 25 7.1263 7.3263

11 7.0495 6.9368 26 7.1007 7.3626

12 7.1263 7.5836 27 7.1689 6.9016

13 7.5519 7.2332 28 7.1799 7.5035

14 7.2031 6.8532 29 7.1909 7.2341

15 8.1263 7.6702 30 8.202 7.6692

25C sidd2_fbe_AllOff stop3_rti_hv

t- test (P-value) 0.298 0.585 f- test (P-value) 0.908 0.317

-40C vreg_on_vddtest_hv stop3_rti_hv

t- test (P-value) 0.171 0.934 f- test (P-value) 0.842 0.362

For 9S08AW:

• sidd1_ram_3v

• stop3_bgref_at

Page 10: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC/AW Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 7

sidd1_ram_3v Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(uA) (uA)

KES-

CHINA

Unit of

Measure

(uA) (uA)

1 120.2303 119.5599 16 114.4835 117.4364

2 119.2779 119.359 17 122.2241 122.7704

3 119.4171 120.8338 18 117.7785 120.2526

4 116.4995 119.4013 19 120.6283 122.1935

5 120.7376 118.1765 20 119.4044 123.4702

6 126.1586 124.0155 21 119.45 115.8499

7 118.8537 120.8868 22 116.8949 118.3264

8 114.8838 120.5073 23 121.2623 117.8316

9 124.4943 122.9668 24 116.4725 116.0856

10 117.02 121.7919 25 118.865 113.4958

11 115.6039 119.8265 26 115.0469 119.3018

12 113.7032 117.8326 27 112.8127 119.0138

13 120.8615 122.8395 28 117.2862 122.1245

14 118.4617 120.8506 29 115.7077 116.7068

15 117.7825 119.3497 30 114.3339 115.672

stop3_bgref_at Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(V) (V)

KES-

CHINA

Unit of

Measure

(V) (V)

1 1.204 1.2014 16 1.2003 1.2021

2 1.2018 1.2006 17 1.201 1.2006

3 1.201 1.2014 18 1.201 1.2021

4 1.2018 1.2021 19 1.201 1.2006

Page 11: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08AC/AW Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 7

5 1.1977 1.1999 20 1.2032 1.2006

6 1.2032 1.2021 21 1.204 1.2006

7 1.2032 1.2036 22 1.204 1.2021

8 1.2025 1.2028 23 1.2018 1.2014

9 1.2069 1.2043 24 1.201 1.1992

10 1.2003 1.2006 25 1.2025 1.2006

11 1.1977 1.1999 26 1.2025 1.2028

12 1.2018 1.2021 27 1.1977 1.197

13 1.2047 1.2014 28 1.204 1.2043

14 1.2047 1.2036 29 1.2032 1.1999

15 1.2003 1.2006 30 1.204 1.2006

95C sidd1_ram_3v stop3_bgref_at t- test (P-value) 0.066 0.137 f- test (P-value) 0.248 0.085

Results: Pass

5.0 Analysis Conclusions and Verification Bin-to-bin activity shows 100% correlation of 1208 units at room&cold temperature between the new KES-

CHINA and the current FSL-TJN-FM.

The QA Gate correlation passes at KES-CHINA.

Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 28 Jun

2010

Jennifer Li (KES-CHINA, TPE)

Zhang Guanyu (TJN MSG PE)

Page 12: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08GT16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 5

[Customer Version]

1. Objective:

This report details the correlation result of the 9S08GT16 (QFN 32) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

Note: 9S08GT already qualified package QFP 44 in KES-CHINA. Please refer to enclosed file for detail.

G9GSGT1A Correlation Report to Customer.doc

2. General Information

* Correlation Lot Information:

Lot # Qty.

Engineer lot 30

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 135C,-40C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

3.2 General Method: Using the production test program, perform 100% screening of 30 units at 135C and -40C temperature, in the current FSL-TJN-FM. Test the same 30 units under the same conditions in

Page 13: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08GT16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 5

KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results: 4.1 Parametric Comparison:

Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• lvd_trim

• rti_stop3_128ms

• sidd1_ram_3v

lvd_trim Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(V) (V)

KES-

CHINA

Unit of

Measure

(V) (V)

1 2.14 2.135 16 2.115 2.115

2 2.1275 2.125 17 2.14 2.1425

3 2.1025 2.105 18 2.1025 2.1025

4 2.1025 2.1025 19 2.1275 2.1225

5 2.13 2.1275 20 2.105 2.1075

6 2.11 2.11 21 2.13 2.1325

7 2.13 2.1325 22 2.1125 2.11

8 2.1225 2.12 23 2.13 2.13

9 2.125 2.12 24 2.13 2.1275

10 2.115 2.1125 25 2.135 2.1325

11 2.13 2.13 26 2.115 2.1125

12 2.115 2.1175 27 2.13 2.13

13 2.1275 2.125 28 2.115 2.11

14 2.1125 2.11 29 2.115 2.1125

15 2.13 2.1325 30 2.13 2.125

rti_stop3_128ms Parametric Comparison

Page 14: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08GT Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 5

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(ms) (ms)

KES-

CHINA

Unit of

Measure

(ms) (ms)

1 136.1454 136.9818 16 133.5999 134.0934

2 129.082 131.0527 17 135.9014 136.728

3 146.0986 146.4033 18 127.1181 128.9992

4 140.7813 142.4731 19 120.5136 120.7731

5 129.1337 131.8784 20 134.5617 135.5977

6 137.1361 137.6665 21 137.3252 138.2802

7 131.814 133.8802 22 134.3109 134.2942

8 140.3735 141.4099 23 139.1231 139.7488

9 124.0753 124.1262 24 134.1981 134.26

10 137.3049 137.9415 25 135.7242 136.3131

11 125.8375 125.8232 26 139.8364 140.4863

12 127.6026 130.9428 27 134.4563 135.1947

13 126.5038 127.1243 28 126.2626 127.0664

14 124.2268 124.4134 29 138.0907 138.4502

15 139.8885 140.6186 30 131.8327 134.0636

FCN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(ms) (ms)

KES-

CHINA

Unit of

Measure

(ms) (ms)

1 138.834 137.1239 16 131.0768 129.2661

2 148.8691 149.3744 17 122.0742 120.6099

3 143.1933 142.8352 18 138.3694 135.8266

4 140.6485 138.8547 19 138.8492 138.3655

5 136.6558 133.4688 20 137.4214 135.288

6 143.1888 141.395 21 141.3884 139.157

7 125.9196 124.2958 22 137.0815 134.7328

8 140.4371 138.7921 23 138.5947 136.2225

9 133.5121 131.9616 24 143.0831 141.0668

10 128.7779 126.6273 25 138.273 135.4661

Page 15: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08GT Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 5

11 128.9395 127.2191 26 129.7228 128.4633

12 127.7239 124.815 27 139.8419 138.5689

13 141.5078 140.2307 28 137.4435 135.4384

14 136.8506 134.4518 29 130.8491 129.0802

15 138.6806 136.3425 30 136.7868 133.8407

sidd1_ram_3v Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 41.65 45.4869 16 146.6824 139.8682

2 104.7507 98.1181 17 105.6124 112.7169

3 100.7299 93.8798 18 175.4413 160.2224

4 57.2236 54.2747 19 120.8965 132.7125

5 73.4487 68.7244 20 151.8332 161.4212

6 96.0146 104.7966 21 97.9738 89.4057

7 76.1506 71.119 22 158.5746 149.1016

8 105.2135 112.3211 23 90.2338 85.6566

9 140.6655 130.06 24 169.6938 159.7888

10 236.6312 219.3977 25 102.1281 108.7892

11 106.3905 117.6903 26 124.3604 133.262

12 224.8965 215.4697 27 99.7669 93.0529

13 64.5771 60.203 28 152.4647 138.8479

14 102.9034 111.3773 29 50.5021 55.7652

15 100.8201 94.8637 30 109.6335 120.1516

FCN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 40.345 44.2627 16 130.3447 120.3301

2 39.2728 42.2545 17 122.8046 132.4319

Page 16: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08GT Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 5

3 35.3232 37.9973 18 116.2443 109.4643

4 77.9003 74.4948 19 97.6541 94.5322

5 85.6027 94.6342 20 90.632 91.8393

6 117.4785 109.6684 21 94.9089 96.7001

7 110.7666 120.3938 22 120.1667 114.0555

8 149.4016 138.8479 23 77.9152 74.8426

9 132.2821 121.1208 24 104.7485 96.507

10 80.4329 87.0186 25 61.5872 62.9065

11 60.5154 55.8673 26 72.2302 66.1287

12 75.9653 73.8827 27 85.6508 79.3569

13 105.0365 97.0571 28 65.0689 61.9966

14 120.5972 113.979 29 37.3502 40.9981

15 105.8958 103.4077 30 75.0721 68.8834

135C lvd_trim rti_stop3_128ms sidd1_ram_3v

t- test (P-value) 0.682 0.543 0.882 f- test (P-value) 0.929 0.972 0.706 -40C N/A rti_stop3_128ms sidd1_ram_3v

t- test (P-value) N/A 0.247 0.781 f- test (P-value) N/A 0.823 0.696

Results: Pass

5.0 Analysis Conclusions and Verification Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 26 Jun

2010

Ji Chang (KES-CHINA, TPE)

Wang Lei (TJN MSG PE)

Page 17: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 7

[Customer Version]

1. Objective: This report details the correlation result of the 9S08JS16(QFN 24) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINAFinal Test facility

2. General Information

* Correlation Lot Information:

Lot # Qty.

LJMEA0N4RA01 1151

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C,95C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 95C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C,95C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 18: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 7

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C and 95C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: � Tables below summarize the correlation results. All numbers filled in green boxes means

correlation was achieved. � All numbers filled in yellow boxes means results did not correlate and requires further

explanation.

Bin-to-Bin Correlation Matrix for Room & Hot Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

Page 19: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 7

4.2 QA Gate results (tested at spec limits):

Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 95C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJMEA0N4RA01 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• SIDD3_ALL_SO

• RIDD_PEE_48M

• Stop2_RTC_ON

SIDD3_ALL_SO Parametric Comparison

FHN

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 7.6856 7.3761 16 7.7449 7.3804

2 7.8951 7.768 17 7.8805 7.6248

3 7.5888 7.3627 18 7.9486 7.928

4 7.3249 7.2195 19 8.2679 8.0021

5 6.3611 6.6778 20 8.1645 8.2561

6 6.9324 7.0218 21 7.423 7.4311

7 7.6749 7.3711 22 7.8314 7.8212

8 7.5808 7.2855 23 6.3994 6.4048

9 8.24 8.0675 24 7.7907 7.9925

10 7.8809 7.6728 25 8.0877 7.4207

11 7.9388 7.9209 26 7.8045 7.7513

12 7.1468 7.1048 27 7.8885 8.012

13 7.3496 7.2337 28 6.5575 7.1996

14 7.2337 7.2192 29 7.2016 7.1704

15 7.5072 7.2663 30 8.2271 8.2308

Page 20: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 7

FRN

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 3.0391 3.0502 16 3.2281 3.1773

2 3.0821 3.0682 17 2.7419 2.6077

3 2.9851 3.0383 18 3.2484 3.2189

4 2.9794 3.0066 19 3.4715 3.3194

5 3.0819 3.0574 20 2.857 2.9312

6 4.6636 4.286 21 3.0865 3.1302

7 3.2823 3.2229 22 2.7894 2.6221

8 2.8443 2.8441 23 2.8719 2.9644

9 2.7088 2.5602 24 2.8157 2.8201

10 3.5222 3.3453 25 3.2338 3.1952

11 3.0151 3.0445 26 2.9057 2.9714

12 2.8142 2.8006 27 3.1911 3.1373

13 2.6486 2.4582 28 2.8509 2.8572

14 3.9333 3.9105 29 3.3078 3.2409

15 3.4567 3.286 30 3.3252 3.2438

RIDD_PEE_48M Parametric Comparison

FHN

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

1 20.0168 19.9288 16 20.002 20.1498

2 19.9153 19.8786 17 19.9811 19.8377

3 19.9723 19.8377 18 20.0792 19.9165

4 19.9102 19.9117 19 20.1588 19.9738

5 20.0604 19.9732 20 19.9984 19.9737

6 19.957 19.8574 21 19.9988 19.8605

7 19.6598 19.5614 22 20.0538 20.0897

8 20.0101 20.0578 23 20.0973 19.938

Page 21: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 7

9 19.9515 19.892 24 19.8877 19.8371

10 19.9448 19.9775 25 19.9764 19.9265

11 20.062 19.9672 26 19.8724 19.9028

12 20.0492 20.1395 27 20.1107 19.9943

13 20.0127 19.8991 28 19.8745 19.8886

14 19.7859 19.9765 29 19.8377 19.8431

15 19.967 19.856 30 20.0543 19.9665

FRN

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

1 19.5185 19.5445 16 20.1499 20.0068

2 19.5471 19.5777 17 20.0077 20.0007

3 19.3935 19.4061 18 19.5746 19.5978

4 19.6611 19.7809 19 19.6662 19.8214

5 20.019 20.0007 20 19.6744 19.878

6 19.7224 19.915 21 19.8347 19.9245

7 19.8524 19.9556 22 19.5165 19.4679

8 19.7622 19.9211 23 19.6459 19.7718

9 19.542 19.5614 24 19.6254 19.7074

10 19.7102 19.8931 25 19.9279 19.9997

11 19.6163 19.7043 26 19.2586 19.3868

12 19.9407 20.0004 27 20.2279 20.016

13 19.8521 19.9469 28 19.6326 19.7418

14 19.7611 19.9201 29 19.2336 19.3827

15 19.7222 19.8982 30 19.9168 19.9877

Stop2_RTC_ON Parametric Comparison

FHN

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 5.8863 5.9501 16 5.7624 5.8484

Page 22: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 7

2 5.9964 5.6934 17 6.0138 5.9874

3 5.8626 5.5881 18 5.9898 5.7632

4 5.9152 5.7183 19 6.871 6.7867

5 5.6438 5.7444 20 5.9796 5.6064

6 5.486 5.3583 21 5.8294 6.0794

7 5.0534 5.3335 22 5.5157 6.0281

8 5.4014 5.4318 23 5.5462 5.8087

9 5.7793 5.7369 24 6.5275 6.7841

10 5.6248 5.6873 25 5.9235 5.7944

11 6.042 5.681 26 5.9646 5.505

12 5.8305 5.8857 27 5.4574 5.6039

13 6.477 6.689 28 5.313 5.5644

14 5.4523 5.6511 29 5.6492 5.7566

15 5.1769 5.6529 30 6.2044 5.7258

FRN

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 2.665 2.5636 16 3.1282 3.0529

2 3.0429 2.9907 17 2.9127 2.8757

3 2.957 2.9191 18 3.1821 3.1763

4 3.1318 3.1146 19 2.9805 2.9663

5 3.0049 2.9699 20 2.9727 2.9519

6 3.7405 3.7745 21 2.7327 2.604

7 3.2538 3.2195 22 2.9817 2.9679

8 2.7872 2.6229 23 2.877 2.7459

9 2.7139 2.5947 24 2.8366 2.6677

10 2.9995 2.9697 25 2.8449 2.6974

11 2.9993 2.9679 26 2.897 2.7585

12 2.9085 2.8571 27 3.1018 3.0397

13 2.827 2.6396 28 2.9054 2.7661

14 3.4203 3.2542 29 3.1752 3.1409

15 3.3609 3.2301 30 3.116 3.0516

Page 23: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08JS16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 7

95C SIDD3_ALL_SO RIDD_PEE_48M Stop2_RTC_ON t- test (P-value) 0.528 0.084 0.927 f- test (P-value) 0.418 0.777 0.680

25C SIDD3_ALL_SO RIDD_PEE_48M Stop2_RTC_ON t- test (P-value) 0.602 0.202 0.225 f- test (P-value) 0.548 0.564 0.476

Results: Pass

5.0 Analysis Conclusions and Verification Bin-to-bin activity shows 100% correlation of 1151 units at hot&room temperature between the new KES-

CHINA and the current FSL-TJN-FM.

The QA Gate correlation passes at KES-CHINA.

Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 17 Jun

2010

Jennifer Li (KES-CHINA, TPE)

You Zhou (TJN MSG PE)

Page 24: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 7

[Customer Version]

1. Objective: This report details the correlation result of the 9S08QB8 (TSSOP16, SOIC28W) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN- FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

TJMEA0LEPN00 1256

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were base on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C -40C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 95C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C,-40C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 25: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 7

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C and -40C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Room& Cold Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FCN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Page 26: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 7

Results: Pass

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 95 C Cold -40C

Lot # Pass # Fail # Pass # Fail

TJMEA0LEPN00 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• ics_meas_20MHz_int_max_ti

• stopidd2_max_ti

• stopidd3_max_ti

ics_meas_20MHz_int_max_ti Parametric Comparison

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(Mhz) (Mhz)

KES-

CHINA

Unit of

Measure

(Mhz) (Mhz)

1 16.75 16.75 16 16.74 16.74

2 16.75 16.75 17 16.76 16.75

3 16.74 16.74 18 16.74 16.74

4 16.78 16.77 19 16.75 16.74

5 16.78 16.81 20 16.75 16.74

6 16.77 16.76 21 16.72 16.73

7 16.79 16.81 22 16.8 16.82

8 16.7 16.73 23 16.73 16.73

9 16.74 16.74 24 16.77 16.76

10 16.68 16.68 25 16.8 16.84

11 16.74 16.74 26 16.77 16.75

Page 27: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 7

12 16.75 16.75 27 16.75 16.76

13 16.75 16.75 28 16.78 16.75

14 16.76 16.75 29 16.78 16.73

15 16.79 16.81 30 16.73 16.8

FCN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(Mhz) (Mhz)

KES-

CHINA

Unit of

Measure

(Mhz) (Mhz)

1 16.72 16.74 16 16.81 16.77

2 16.72 16.72 17 16.74 16.78

3 16.8 16.92 18 16.76 16.77

4 16.77 16.91 19 16.72 16.82

5 16.8 16.76 20 16.76 16.77

6 16.76 16.76 21 16.78 16.74

7 16.75 16.76 22 16.88 16.75

8 16.78 16.79 23 16.79 16.77

9 16.71 16.79 24 16.74 16.76

10 16.83 16.77 25 16.72 16.73

11 16.76 16.83 26 16.76 16.81

12 16.84 16.72 27 16.72 16.76

13 16.76 16.74 28 16.91 16.78

14 16.79 16.7 29 16.76 16.87

15 16.78 16.78 30 16.79 16.77

stopidd2_max_ti Parametric Comparison

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 192.764 187.852 16 185.1891 199.6735

2 296.7598 292.8617 17 178.5519 187.7246

Page 28: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 7

3 172.2977 184.9315 18 285.3558 268.258

4 170.8937 183.9861 19 164.0013 174.8647

5 184.5509 195.4982 20 180.3171 164.9825

6 217.5054 200.0562 21 243.1859 251.4431

7 183.9127 188.4994 22 265.8663 268.7679

8 209.3579 206.6882 23 167.9138 186.4606

9 168.0857 177.2747 24 168.5963 180.4149

10 189.5288 204.6455 25 263.4412 254.1193

11 214.8062 198.0565 26 311.0995 295.2838

12 161.1933 174.293 27 370.9561 364.1628

13 174.1478 164.531 28 345.3949 319.3621

14 203.1859 216.3737 29 331.9929 311.8315

15 160.938 170.2773 30 336.4602 317.3199

FCN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 289.2188 274.127 16 257.8252 245.8431

2 234.9338 231.5438 17 221.0214 228.2496

3 241.2323 231.4424 18 211.1933 223.2576

4 219.8065 208.8564 19 120.3408 115.3563

5 145.3845 153.9537 20 177.95 168.9363

6 296.8379 273.7397 21 146.2053 152.9342

7 170.6273 176.2664 22 250.167 240.4674

8 173.8145 182.5831 23 214.3842 227.8744

9 131.1057 119.928 24 162.0855 164.7274

10 167.958 162.9596 25 177.7666 192.3123

11 199.0415 205.7021 26 271.3105 246.8414

12 108.556 103.6356 27 172.1363 178.4327

13 175.4002 164.7859 28 220.914 201.0765

14 237.4866 236.5155 29 193.0688 210.9181

15 175.1996 188.2446 30 208.13 213.6966

Page 29: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 7

stopidd3_max_ti Parametric Comparison

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 220.5251 212.9571 16 253.0726 246.2895

2 275.4091 265.3578 17 287.8538 270.3039

3 230.5681 216.6927 18 361.4153 367.8199

4 209.3454 210.2594 19 219.7274 210.4806

5 238.2348 219.1697 20 166.6445 175.1858

6 237.4885 220.4212 21 258.0186 248.1617

7 249.084 235.4882 22 292.002 285.2766

8 207.9209 194.1574 23 146.3254 141.9885

9 264.56 249.5953 24 257.6995 247.2461

10 296.7884 324.9547 25 142.0603 137.9285

11 289.9279 272.1234 26 253.7108 247.0326

12 200.6018 188.0992 27 411.4667 436.7884

13 208.24 207.7003 28 324.39 359.4083

14 295.1929 302.1677 29 425.5068 435.8311

15 267.2723 261.6107 30 371.2505 385.2548

FCN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 321.9419 337.8456 16 236.2439 225.086

2 334.6908 320.6523 17 192.8102 182.0154

3 149.5114 156.8019 18 243.8189 268.3855

4 181.9018 195.9111 19 180.3603 192.4815

5 222.3799 206.1074 20 152.9131 161.1563

6 255.5671 278.7433 21 164.5085 169.4692

7 161.1447 165.8811 22 250.2522 239.7025

8 162.7383 154.6199 23 186.4745 174.1639

Page 30: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QB8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 7

9 164.0132 180.1421 24 159.5059 166.8956

10 267.2004 260.418 25 255.2496 236.2144

11 268.9931 247.2752 26 246.3664 230.6196

12 191.7023 181.8816 27 166.4179 163.9697

13 185.3676 177.1154 28 286.6803 261.2755

14 254.9924 241.9334 29 196.0936 209.2933

15 265.6064 247.4344 30 300.2418 282.2489

25C ics_meas_20MHz_int_max_ti stopidd2_max_ti stopidd3_max_ti t- test (P-value) 0.738 0.987 0.878 f- test (P-value) 0.304 0.447 0.557 -40C ics_meas_20MHz_int_max_ti stopidd2_max_ti stopidd3_max_ti t- test (P-value) 0.732 0.898 0.824 f- test (P-value) 0.687 0.646 0.709

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1256 units at Room&Cold temperature and between the

new KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Rev Date Originator

O 17 Jun

2010

Ji Chang (KES-CHINA, TPE)

Dai Xiaojing (TJN MSG PE)

Page 31: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 10

[Customer Version]

1. Objective: This report details the correlation result of the 9S08QE128(QFN 48) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

LJMEA0N4RX00 1040

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin Correlation

Goal: Correlate bin-to-bin per production test flow Lot quantity: 1 Units per lot: 1000

Temperature:25,95C,-40C Type of units: good, reject Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures Lot quantity: 1 Units per lot: 1000 Temperature: 95C, -40C Type of units: good Site: KES-CHINA

Y 0 failures.

c Parametric Comparison

Goal: Parametric comparison for Identify Parametric Tests Lot quantity: 1 Units per lot: 30

Temperature: 25,95C,-40C Type of units: good Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 32: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 10

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C,95C, and -40C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Room, Hot, Cold Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Page 33: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 10

FCN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 95 C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJMEA0N4RX00 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• bandgap_meas_ti • osc_bias_hgo0_rng0_gnd

• stopidd3_max_ti

bandgap_meas_ti Parametric Comparison

FRN

Pat# KES-CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

1 1.1731 1.1753 16 1.1721 1.1753 2 1.1749 1.1761 17 1.1753 1.1745

Page 34: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 10

3 1.1749 1.1737 18 1.1773 1.1793 4 1.1777 1.1769 19 1.1745 1.1769 5 1.1733 1.1745 20 1.1753 1.1777 6 1.1777 1.1769 21 1.1721 1.1753 7 1.1741 1.1753 22 1.1721 1.1753 8 1.1777 1.1769 23 1.1745 1.1769 9 1.1741 1.1753 24 1.1729 1.1761

10 1.1721 1.1753 25 1.1721 1.1745 11 1.1769 1.1761 26 1.1739 1.1729 12 1.1755 1.1777 27 1.1753 1.1777 13 1.1775 1.1769 28 1.1749 1.1761 14 1.1745 1.1769 29 1.1753 1.1777 15 1.1749 1.1721 30 1.1749 1.1761

FHN

Pat# KES-CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

1 1.1751 1.1761 16 1.1755 1.1745 2 1.1751 1.1769 17 1.1717 1.1737 3 1.1709 1.1721 18 1.1781 1.1761 4 1.1717 1.1737 19 1.1739 1.1729 5 1.1719 1.1729 20 1.1771 1.1761 6 1.1721 1.1761 21 1.1707 1.1729 7 1.1627 1.1645 22 1.1753 1.1753 8 1.1709 1.1729 23 1.1733 1.1753 9 1.1709 1.1713 24 1.1751 1.1769

10 1.1755 1.1745 25 1.1717 1.1737 11 1.1771 1.1761 26 1.1743 1.1753 12 1.1791 1.1793 27 1.1717 1.1729 13 1.1721 1.1761 28 1.1751 1.1777 14 1.1763 1.1753 29 1.1733 1.1753 15 1.1719 1.1729 30 1.1741 1.1753

Page 35: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 10

FCN

Pat# KES-CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

1 1.1763 1.1761 16 1.1737 1.1753 2 1.1773 1.1777 17 1.1737 1.1753 3 1.1681 1.1697 18 1.1737 1.1753 4 1.1721 1.1737 19 1.1787 1.1785 5 1.1721 1.1754 20 1.1767 1.1793 6 1.1721 1.1769 21 1.1745 1.1769 7 1.1721 1.1761 22 1.1745 1.1753 8 1.1721 1.1769 23 1.1745 1.1753 9 1.1721 1.1729 24 1.1745 1.1753

10 1.1721 1.1753 25 1.1745 1.1745 11 1.1721 1.1753 26 1.1745 1.1729 12 1.1769 1.1793 27 1.1775 1.1785 13 1.1759 1.1769 28 1.1753 1.1761 14 1.1759 1.1777 29 1.1761 1.1777 15 1.1737 1.1729 30 1.1761 1.1777

osc_bias_hgo0_rng0_gnd Parametric Comparison

FRN

Pat# KES-CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 -1.4694 -1.5633 16 -1.532 -1.6259 2 -1.4068 -1.4904 17 -1.3441 -1.3651 3 -1.3441 -1.331 18 -1.4068 -1.5189 4 -1.4694 -1.6127 19 -1.6572 -1.8006 5 -1.3441 -1.3023 20 -1.4694 -1.6154 6 -1.5068 -1.5633 21 -1.7198 -1.8765 7 -1.532 -1.6157 22 -1.9703 -1.9289

Page 36: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 10

8 -1.3441 -1.331 23 -1.4068 -1.3936 9 -1.5946 -1.7379 24 -1.4068 -1.4875

10 -1.532 -1.6781 25 -2.0329 -2.0538 11 -1.532 -1.6886 26 -1.7572 -1.8765 12 -1.4068 -1.4904 27 -1.9198 -1.9289 13 -1.4068 -1.4563 28 -1.4694 -1.5815 14 -1.4068 -1.4249 29 -1.5946 -1.7379 15 -1.3441 -1.2397 30 -1.5946 -1.7407

FHN

Pat# KES-CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 -1.4694 -1.553 16 -1.5946 -1.7512 2 -1.4068 -1.5189 17 -1.7694 -1.6783 3 -1.4068 -1.5501 18 -1.2189 -1.2684 4 -1.5694 -1.6781 19 -1.4068 -1.5501 5 -1.2937 -1.2501 20 -1.2815 -1.3023 6 -1.6572 -1.8139 21 -1.532 -1.6886 7 -1.6694 -1.6783 22 -1.532 -1.6886 8 -1.2815 -1.3936 23 -1.5946 -1.741 9 -1.4068 -1.4875 24 -1.3441 -1.4563

10 -1.4694 -1.6154 25 -1.732 -1.7379 11 -1.7946 -1.8139 26 -1.5946 -1.7407 12 -1.4694 -1.553 27 -2.1198 -2.0644 13 -1.632 -1.7068 28 -1.4068 -1.553 14 -1.732 -1.7379 29 -1.9198 -1.9573 15 -1.4815 -1.4275 30 -1.2815 -1.3623

FCN

Pat# KES-CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 -1.3441 -1.4904 16 -1.5946 -1.7512

Page 37: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 10

2 -1.2189 -1.331 17 -1.0937 -1.1431 3 -1.4694 -1.6127 18 -1.5694 -1.6753 4 -1.0937 -1.1144 19 -1.3815 -1.4275 5 -1.4694 -1.5633 20 -1.4068 -1.553 6 -1.6572 -1.7512 21 -1.4694 -1.6157 7 -1.2189 -1.3025 22 -1.3441 -1.438 8 -1.4068 -1.5189 23 -1.632 -1.7068 9 -1.9058 -1.9258 24 -1.4068 -1.5501

10 -1.3441 -1.4275 25 -1.4694 -1.6154 11 -2.0329 -1.8765 26 -1.7824 -1.8765 12 -1.2815 -1.3651 27 -1.4694 -1.6157 13 -1.2815 -1.3936 28 -1.4441 -1.5189 14 -1.3441 -1.4875 29 -1.3441 -1.4875 15 -1.3441 -1.4275 30 -1.732 -1.7407

stopidd3_max_ti Parametric Comparison

FRN

Pat# KES-CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

1 186.3388 175.1294 16 328.3542 305.0848 2 186.3388 193.2407 17 336.0035 309.0342 3 218.3786 201.6586 18 308.2983 312.12 4 199.0558 213.2651 19 324.1628 313.9794 5 216.8645 223.807 20 337.6766 326.0172 6 210.9441 230.9938 21 356.3459 326.4881 7 218.9759 234.9189 22 350.298 328.2844 8 218.9759 241.6536 23 325.78 340.1542 9 229.7408 249.9578 24 351.772 342.7194

10 269.7408 250.6288 25 361.4612 344.8869 11 262.4181 256.12 26 376.8314 351.1413 12 254.0754 261.7686 27 345.6998 356.7441 13 310.9598 295.8003 28 360.2335 364.4535 14 316.0594 298.014 29 378.4644 368.6577

Page 38: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 8 of 10

15 327.7168 301.4438 30 434.0739 415.6585

FHN

Pat# KES-CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

1 482.4 479.8325 16 589.7 582.7606 2 748 756.0666 17 927.8 868.7745 3 2319.5 2473.8 18 1460.5 1586.2 4 945.4 891.6375 19 1058.6 1018.9 5 4178.3 4067.8 20 3161.5 3163.2 6 919.4 848.9453 21 337.6 359.0481 7 1128.9 1259.4 22 459.1 476.9744 8 3047.8 2852.2 23 775 783.7262 9 4166.8 4013.1 24 1531.7 1417.9

10 945.4 973.6835 25 510.3 474.4003 11 470.8 449.477 26 326.8 307.6478 12 616.1 610.7194 27 428.6 427.3778 13 980.3 1033.4 28 890 844.6663 14 2310.5 2453 29 1698.2 1717 15 617.8856 619.7645 30 470.5 445.6077

FCN

Pat# KES-CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

1 424.9647 403.306 16 244.2486 216.3262 2 532.4221 555.3854 17 323.2916 305.8726 3 231.3636 240.5038 18 143.9865 137.2109 4 508.5816 537.7184 19 300.4153 279.6047 5 112.5212 120.6069 20 319.467 292.6286 6 467.2753 441.3141 21 221.4194 237.4376 7 480.7891 471.8745 22 194.5996 185.7156

Page 39: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE128 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 9 of 10

8 185.6754 177.7751 23 219.0688 239.8288 9 169.2852 170.9297 24 120.457 127.6316

10 476.582 453.8888 25 243.7701 225.7142 11 327.3713 318.7442 26 207.834 218.3669 12 583.4663 534.3483 27 337.3154 350.4968 13 326.7338 311.0256 28 215.5636 207.9258 14 210.4641 195.4525 29 159.2136 149.6 15 474.7971 447.3913 30 329.921 335.4062

Parametric Comparison

25C bandgap_meas_ti osc_bias_hgo0_rng0_gnd stopidd3_max_ti

t- test (P-value) 0.184 0.168 0.736

f- test (P-value) 0.433 0.573 0.513

95C bandgap_meas_ti osc_bias_hgo0_rng0_gnd stopidd3_max_ti

t- test (P-value) 0.141 0.151 0.977

f- test (P-value) 0.404 0.728 0.923

-40C bandgap_meas_ti osc_bias_hgo0_rng0_gnd stopidd3_max_ti

t- test (P-value) 0.145 0.090 0.844

f- test (P-value) 0.294 0.627 0.881

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1040 units at Room Hot Cold temperature and between

the new KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Page 40: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

G9ZZQE28 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Transfer

Page 10 of 10

Rev Date Originator

O 28 Jun

2010

Ji Chang (KES-CHINA, TPE)

Liu Qingfeng (TJN MSG PE)

Page 41: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 5

[Customer Version]

1. Objective: This report details the correlation result of the 9S08QE32(LQFP 32, LQFP 44) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

LJMEA0NDNK00 1229

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C

Type of units: good reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 95C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 42: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 5

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Room Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 95 C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJMEA0NDNK00 100% 0 100% 0

Results: Pass

Page 43: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 5

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• stopidd3_max_ti

• bandgap_meas_ti

• ics_meas_max_ti

stopidd3_max_ti Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

Pat# KES-

CHINA

Unit of

Measure

(nA)

FSL-TJN

Unit of

Measure

(nA)

1 284.3824 301.6279 16 270.2178 257.3967 2 258.5267 244.5977 17 287.058 302.9023 3 344.691 350.5128 18 343.2678 341.7451 4 281.647 282.5868 19 276.055 265.8614 5 294.3334 324.6176 20 381.4436 362.6448 6 233.5618 221.3454 21 250.1029 239.3456 7 338.6594 357.7513 22 328.9519 332.5187 8 288.2016 310.9893 23 263.632 247.2493 9 283.3566 294.1856 24 343.7778 341.7961

10 273.7752 257.9576 25 265.572 255.459 11 332.1754 335.883 26 278.3649 266.6591 12 264.8526 249.6969 27 279.1695 268.258 13 288.4619 317.4301 28 330.3782 351.9401 14 271.2763 257.8047 29 279.7822 270.0937 15 298.8145 329.6131 30 298.6565 327.88

bandgap_meas_ti Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Page 44: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 5

1 1.1677 1.1678 16 1.1661 1.1686 2 1.1726 1.1702 17 1.1661 1.1653 3 1.1677 1.1678 18 1.1694 1.1686 4 1.1669 1.1662 19 1.1694 1.1686 5 1.1677 1.1678 20 1.1685 1.1678 6 1.171 1.1694 21 1.1661 1.1645 7 1.1661 1.1653 22 1.1669 1.167 8 1.171 1.1694 23 1.1677 1.167 9 1.1661 1.1653 24 1.1726 1.1702

10 1.1694 1.1686 25 1.1685 1.1678 11 1.1694 1.1702 26 1.171 1.1694 12 1.1685 1.1678 27 1.1726 1.1694 13 1.1645 1.1645 28 1.1734 1.1702 14 1.171 1.1694 29 1.1677 1.167 15 1.1661 1.1653 30 1.171 1.1686

ics_meas_max_ti Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

1 50.35 50.39 16 50.62 50.53 2 50.58 50.53 17 50.63 50.55 3 50.47 50.49 18 50.54 50.58 4 50.54 50.52 19 50.57 50.46 5 50.45 50.39 20 50.46 50.41 6 50.55 50.52 21 50.52 50.48 7 50.47 50.42 22 50.53 50.57 8 50.55 50.47 23 50.38 50.5 9 50.46 50.48 24 50.57 50.51

10 50.46 50.39 25 50.39 50.42 11 50.49 50.5 26 50.42 50.52 12 50.61 50.53 27 50.46 50.42 13 50.37 50.4 28 50.41 50.45 14 50.52 50.44 29 50.53 50.39 15 50.46 50.48 30 50.46 50.43

Page 45: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08QE32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 5

25C stopidd3_max_ti bandgap_meas_ti ics_meas_max_ti t- test (P-value) 0.852 0.094 0.214 f- test (P-value) 0.274 0.107 0.151

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1229 units at room temperature and between the new

KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Rev Date Originator

O 25 Jun

2010

Ji Chang (KES-CHINA, TPE)

Liu David (TJN MSG PE)

Page 46: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 6

[Customer Version]

1. Objective: This report details the correlation result of the 9S08SE8(TSSOP16, SOIC28W) device to

demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN- FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

LJMEA0N4Y301 1104

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Package: SOIC28

Lot quantity: 1

Units per lot: 1000

Temperature: 95C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Package: SOIC28

Lot quantity: 1

Units per lot: 1000

Temperature: 95C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

Page 47: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 6

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Package: SOIC28 and TSSOP16

Lot quantity: 1lot per package

Units per lot: 30

Temperature: 95C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 95C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Hot Test

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

Page 48: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 6

QA Gate Results

QA Gate Hot 95 C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJMEA0N4Y301 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison: For SOIC 28:

• se8_widd_hv

• jvt_vdd_par_28pin

se8_widd_hv Parametric Comparison

FHN

KES-

CHINA

FSL-TJN KES-

CHINA

FSL-TJN

Unit of

Measure

Unit of

Measure

Unit of

Measure

Unit of

Measure

Pat#

(mA) (mA)

Pat#

(mA) (mA)

1 1.4615 1.4609 16 1.4462 1.4368

2 1.4432 1.4386 17 1.4366 1.4595

3 1.4505 1.4286 18 1.3941 1.4015

4 1.4035 1.4246 19 1.4253 1.4164

5 1.4207 1.4393 20 1.4281 1.4495

6 1.4579 1.4743 21 1.4535 1.4671

7 1.44 1.461 22 1.4667 1.4597

8 1.4597 1.4498 23 1.4454 1.4644

9 1.431 1.451 24 1.432 1.4532

10 1.4018 1.4185 25 1.4631 1.4539

11 1.4096 1.4325 26 1.4571 1.4717

12 1.4458 1.4649 27 1.4416 1.4642

13 1.4599 1.4498 28 1.4792 1.4654

14 1.4009 1.4181 29 1.4509 1.4364

15 1.4096 1.4288 30 1.4636 1.4548

Page 49: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 6

jvt_vdd_par_28pin Parametric Comparison

FHN

KES-

CHINA

FSL-TJN KES-

CHINA

FSL-TJN

Unit of

Measure

Unit of

Measure

Unit of

Measure

Unit of

Measure

Pat#

(mA) (mA)

Pat#

(mA) (mA)

1 290.0629 289.5444 16 292.5495 293.3007

2 286.9331 287.3533 17 292.2367 292.0486

3 286.6066 287.0403 18 291.6111 291.1096

4 287.8721 287.9793 19 292.8797 294.8658

5 284.4293 286.7273 20 292.2538 292.3616

6 289.1089 288.9184 21 291.6111 291.4226

7 291.6111 291.4226 22 292.2538 292.6747

8 290.6728 290.1705 23 292.2538 292.6747

9 291.2984 291.1096 24 288.498 288.6054

10 289.4217 289.2314 25 292.8623 293.3007

11 289.124 288.9184 26 290.9856 290.7965

12 292.2367 292.3616 27 291.9239 291.7356

13 290.6889 290.7965 28 290.6889 290.4835

14 291.9408 292.0486 29 292.2538 292.3616

15 289.7499 289.2314 30 291.9239 291.7356

95C se8_widd_hv jvt_vdd_par_28pin

t- test (P-value) 0.186 0.812 f- test (P-value) 0.326 0.879

For TSSOP16:

• ics_fei_factory_trim_HT_max

• se8_widd_hv

ics_fei_factory_trim_HT_max Parametric Comparison

Page 50: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 6

FHN

Pat#

FSL-

TJN

Unit of

Measure

Pat#

FSL-

TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(mHz) (mHz)

KES-

CHINA

Unit of

Measure

(mHz) (mHz)

1 5.0174 5.0208 16 5.0223 5.0208

2 5.0125 5.0061 17 5.0125 5.011

3 5.0076 5.0012 18 5.0125 5.011

4 5.0174 5.0208 19 5.0076 5.0012

5 5.0174 5.0208 20 5.0223 5.0208

6 5.0272 5.0256 21 5.0223 5.0208

7 5.0174 5.0208 22 5.0125 5.0012

8 5.0321 5.0305 23 5.0125 5.0012

9 5.0174 5.0159 24 5.0223 5.0208

10 5.0028 4.9963 25 5.0223 5.0208

11 5.0125 5.011 26 5.0174 5.0159

12 5.0272 5.0256 27 5.0223 5.0208

13 5.0125 5.011 28 5.0125 5.0012

14 5.0174 5.0208 29 5.0369 5.0305

15 5.0125 5.0061 30 5.0272 5.0256

se8_widd_hv Parametric Comparison

FHN

Pat

#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(mA) (mA)

KES-

CHINA

Unit of

Measure

(mA) (mA)

1 1.38 1.3983 16 1.4338 1.4538

2 1.4007 1.401 17 1.4144 1.4277

3 1.401 1.443 18 1.4144 1.4294

4 1.441 1.4576 19 1.4446 1.4329

5 1.4014 1.4474 20 1.4148 1.4354

6 1.4527 1.4521 21 1.4148 1.4202

7 1.4321 1.4458 22 1.4449 1.4723

8 1.4322 1.4472 23 1.4149 1.4115

Page 51: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SE8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 6

9 1.3824 1.3832 24 1.4094 1.4145

10 1.4326 1.416 25 1.4151 1.4373

11 1.4334 1.4463 26 1.4054 1.4063

12 1.4136 1.4152 27 1.4156 1.418

13 1.4138 1.4195 28 1.4177 1.377

14 1.3838 1.3941 29 1.3956 1.3846

15 1.4132 1.4187 30 1.4323 1.4468

95C ics_fei_factory_trim_HT_max se8_widd_hv

t- test (P-value) 0.239 0.136

f- test (P-value) 0.183 0.172

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1104 units at Hot temperature and between the new KES-

CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Rev Date Originator

O 28 Jun

2010

Ji Chang (KES-CHINA, TPE)

Qing Jia (TJN MSG PE)

Page 52: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 6

[Customer Version]

1. Objective: This report details the correlation result of the 9S08SF4(TSSOP 20) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

LJXE94227400 1035

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C ,135C Type of units: good, reject Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 135C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C ,135C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 53: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 6

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C and 135C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Room& Hot Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

Page 54: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 6

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 135 C Cold -40C

Lot # Pass # Fail # Pass # Fail

TJMEA0LEPN00 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• ridd_1mhz_bus_max_ti

• ics _fei_max_trim_hex_FF_fine_max_ti

• pmc_temp_vregon_adc_par_max_ti

ridd_1mhz_bus_max_ti Parametric Comparison

FHN

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

1 1.3354 1.3101 16 1.2925 1.281

2 1.2797 1.2881 17 1.3102 1.3216

3 1.2654 1.277 18 1.3158 1.3417

4 1.2654 1.265 19 1.2903 1.2978

5 1.2807 1.2933 20 1.3077 1.3208

6 1.2898 1.2977 21 1.2904 1.3044

7 1.2912 1.2973 22 1.2965 1.3138

8 1.2904 1.3039 23 1.2999 1.3099

9 1.2766 1.2877 24 1.2919 1.3091

10 1.2713 1.2626 25 1.2959 1.3106

11 1.3077 1.3191 26 1.2725 1.279

12 1.2715 1.2779 27 1.3036 1.3152

13 1.2919 1.308 28 1.3165 1.3131

14 1.3109 1.3045 29 1.3207 1.3342

Page 55: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 6

15 1.2766 1.287 30 1.327 1.314

FRN

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

Pat# KES-

CHINA

Unit of

Measure

(mA)

FSL-TJN

Unit of

Measure

(mA)

1 1.298 1.3142 16 1.2781 1.2885 2 1.271 1.2728 17 1.2685 1.2621 3 1.3005 1.3158 18 1.2542 1.2564 4 1.2934 1.3044 19 1.2751 1.2779 5 1.3117 1.3209 20 1.2639 1.2611 6 1.2934 1.3038 21 1.2766 1.2811 7 1.2685 1.2713 22 1.2868 1.2901 8 1.2893 1.2978 23 1.243 1.2508 9 1.2878 1.2969 24 1.2873 1.2956

10 1.2914 1.2982 25 1.2736 1.2759 11 1.3031 1.3174 26 1.2771 1.2825 12 1.297 1.3079 27 1.2873 1.2954 13 1.3041 1.3198 28 1.241 1.2481 14 1.2965 1.3054 29 1.2781 1.2846 15 1.2924 1.302 30 1.2532 1.2529

ics _fei_max_trim_hex_FF_fine_max_ti Parametric Comparison

FHN

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

1 3.9795 3.9893 16 3.999 4.0049 2 3.9482 3.9404 17 3.9834 3.9912 3 4.0303 4.0381 18 4.0225 4.0244 4 3.9795 3.9893 19 4.0186 4.0166 5 4.0049 4.0147 20 3.96 3.96 6 4.1221 4.126 21 3.9756 3.9834 7 4.0498 4.0654 22 3.9756 3.9873

Page 56: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 6

8 4.0244 4.0361 23 4.0694 4.0752 9 3.9248 3.9209 24 3.9873 3.999

10 3.9248 3.9189 25 4.0889 4.0772 11 4.0459 4.0596 26 3.9385 3.9248 12 3.9951 4.0049 27 3.9893 4.001 13 4.0459 4.0537 28 3.8935 3.9072 14 4.0225 4.0225 29 4.0264 4.0361 15 3.9521 3.96 30 3.9111 3.9131

FRN

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

Pat# KES-

CHINA

Unit of

Measure

(Mhz)

FSL-TJN

Unit of

Measure

(Mhz)

1 4.0283 4.0908 16 3.9756 4.0147 2 3.8916 3.917 17 4.1006 4.0029 3 3.956 4.0029 18 3.9951 4.044 4 3.9013 3.9306 19 3.9306 3.9756 5 4.081 4.0576 20 3.8838 3.8955 6 3.9267 3.958 21 4.0127 4.0615 7 4.0049 4.0576 22 4.0576 4.1123 8 3.9287 3.9639 23 3.9889 3.9502 9 4.044 4.0947 24 4.0244 4.0713

10 4.0814 4.0186 25 4.0537 4.0967 11 3.9385 3.9795 26 4.0967 3.9287 12 3.9873 4.0205 27 3.9541 3.9814 13 3.9912 4.0361 28 3.9932 4.04 14 3.9971 4.0459 29 3.9912 4.0244 15 4.099 4.0518 30 3.9697 4.0049

25C ridd_1mhz_bus_max_ti ics _fei_max_trim_hex_FF_fine_max_ti

t- test (P-value) 0.179 0.243 f- test (P-value) 0.303 0.651 135C ridd_1mhz_bus_max_ti ics _fei_max_trim_hex_FF_fine_max_ti

t- test (P-value) 0.152 0.715 f- test (P-value) 0.865 0.874

Page 57: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SF4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 6

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1035 units at Room &Hot temperature and between the

new KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History: Rev Date Originator

O 17 Jun

2010

Ji Chang (KES-CHINA, TPE)

Cui Yujie (TJN MSG PE)

Page 58: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 6

[Customer Version]

1. Objective:

This report details the correlation result of the 9S08SH32(TSSOP 16, TSSOP 20, TSSOP 28, SOIC 28W) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

LJME00480T00 1104

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on of the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Package: TSSOP28

Lot quantity: 1

Units per lot: 1000

Temperature: 25C

Type of units: good reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Package: TSSOP28

Lot quantity: 1

Units per lot: 1000

Temperature: 135C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

Page 59: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 6

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Package: TSSOP28 and SOIC28

Lot quantity: 1lot per package

Units per lot: 30

Temperature: 25C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Room Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

Page 60: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 6

QA Gate Results

QA Gate Hot 135 C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJME00480T00 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison: For TSSOP 28

• stop3_bgref

• iddq_measure_10vect_max

stop3_bgref Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

Pat# KES-

CHINA

Unit of

Measure

(V)

FSL-TJN

Unit of

Measure

(V)

1 1.202 1.1996 16 1.2021 1.2023

2 1.2048 1.2037 17 1.2034 1.205

3 1.2021 1.2023 18 1.2021 1.201

4 1.202 1.201 19 1.2034 1.2023

5 1.2034 1.201 20 1.2047 1.2037

6 1.2021 1.2023 21 1.2048 1.2037

7 1.2034 1.2037 22 1.2034 1.2064

8 1.2021 1.2023 23 1.2048 1.2037

9 1.2048 1.2037 24 1.2021 1.201

10 1.2021 1.2023 25 1.2034 1.2023

11 1.2021 1.2023 26 1.2021 1.201

12 1.2021 1.2023 27 1.2047 1.2037

13 1.2061 1.205 28 1.2034 1.2023

14 1.2021 1.2023 29 1.2034 1.2023

15 1.2048 1.2037 30 1.2061 1.2037

Page 61: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 6

iddq_measure_10vect_max Parametric Comparison

FRN

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

Pat# KES-

CHINA

Unit of

Measure

(uA)

FSL-TJN

Unit of

Measure

(uA)

1 324.0502 329.0214 16 450.0906 422.0099

2 342.4476 351.7394 17 490.1889 484.2359

3 475.8062 459.7713 18 458.7841 455.577

4 496.1307 486.8239 19 371.9904 362.5503

5 358.9708 362.3718 20 420.4684 404.0853

6 450.5623 426.5739 21 399.6083 375.5157

7 488.7098 472.061 22 539.8399 573.2979

8 453.2907 429.4933 23 384.801 372.8512

9 354.585 354.7099 24 418.5558 401.5483

10 476.6221 461.2502 25 379.5516 370.225

11 393.0881 373.8711 26 404.7979 384.6436

12 379.4463 364.5263 27 423.9366 409.3123

13 530.3735 520.85 28 414.3353 393.2107

14 315.5719 328.9322 29 409.2861 389.9981

15 354.0878 352.0327 30 325.8096 339.4243

25C stop3_bgref iddq_measure_10vect_max

t- test (P-value) 0.096 0.512 f- test (P-value) 0.657 0.885

For SOIC28:

• iddq_measure_10vect_max_long

• stop3_bgref

iddq_measure_10vect_max_long Parametric Comparison

FRN

FSL-TJN FSL-TJN Pat# KES-

CHINA

Unit of

Measure

Unit of

Measure

Pat# KES-

CHINA

Unit of

Measure

Unit of

Measure

Page 62: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 6

(uA) (uA) (uA) (uA)

1 1.1872 1.1196 16 1.1383 1.0953

2 1.1778 1.1153 17 1.0978 1.0693

3 0.6001 0.5911 18 0.59805 0.5843

4 1.1901 1.1234 19 1.2079 1.1817

5 1.1283 1.0909 20 1.1464 1.1070

6 0.6523 0.7025 21 0.7077 0.6493

7 1.0957 1.0563 22 1.2078 1.1466

8 1.1347 1.0936 23 1.1229 1.0907

9 0.5497 0.5977 24 0.5996 0.6330

10 1.1471 1.1108 25 1.1096 1.0861

11 1.0832 1.0532 26 1.1897 1.1221

12 0.5882 0.6007 27 0.6634 0.6397

13 1.1672 1.1129 28 1.1882 1.1206

14 1.1187 1.0870 29 1.171 1.1137

15 0.6338 0.6232 30 0.5982 0.6136

stop3_bgref Parametric Comparison

FRN

FSL-TJN FSL-TJN KES-

CHINA

Unit of

Measure

Unit of

Measure

KES-

CHINA

Unit of

Measure

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 1.2006 1.2019 16 1.202 1.2019

2 1.2006 1.2032 17 1.2033 1.2019

3 1.206 1.2023 18 1.1966 1.1992

4 1.202 1.2019 19 1.2033 1.2019

5 1.206 1.2072 20 1.206 1.2059

6 1.1993 1.2005 21 1.2033 1.2019

7 1.1993 1.2005 22 1.1993 1.2005

8 1.202 1.2019 23 1.2033 1.2019

Page 63: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH32 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 6

9 1.202 1.2019 24 1.2047 1.2072

10 1.206 1.2045 25 1.2047 1.2019

11 1.206 1.2045 26 1.2047 1.2072

12 1.202 1.2019 27 1.2033 1.2032

13 1.202 1.2019 28 1.2033 1.2045

14 1.202 1.2032 29 1.2006 1.2005

15 1.1966 1.1992 30 1.2033 1.2032

135C iddq_measure_10vect_max_long stop3_bgref t- test (P-value) 0.650 0.778

f- test (P-value) 0.585 0.324

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1104 units at Room temperature and between the new

KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Rev Date Originator

O 28 Jun

2010

Ji Chang (KES-CHINA, TPE)

Ren Tao (TJN MSG PE)

Page 64: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 6

[Customer Version]

1. Objective: This report details the correlation result of the 9S08SH8/SH4/SQ4(TSSOP 16, SOIC 20W, SOIC 8) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA HINA Final Test facility

Note: 9S08SH8 already qualified package TSSOP20 in KES-CHINA Please refer to enclosed file for detail.

KESTJ FT QUAL REPORT Jumper(TSSOP20).doc

2. General Information

* Correlation Lot Information:

Lot # Qty.

Engineer lot 30

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Package: TSSOP16 and SOIC20 W

Lot quantity: 1lot per package

Units per lot: 30

Temperature: 25C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 65: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 6

3.2 General Method: Using the production test program, perform 100% screening of 30 units at 25C temperature, in the current FSL-TJN-FM. Test the same 30 units under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results: 4.1 Parametric Comparison:

Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison: For SOIC 20:

• pmc_lpo_8b_5v

• sidd3_dps_5v

• ridd_FBE_2M_dps

pmc_lpo_8b_5v Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(us) (us)

Pat#

(us) (us)

1 862.625 865.125 16 938.5 931.375

2 884.125 886 17 939.5 946

3 946.375 949.75 18 917 912.625

4 924.5 926.625 19 948.25 963.5

5 947.625 960.125 20 977.25 969.125

6 917.25 919.5 21 917.25 922

7 927 926.625 22 938.25 930.125

8 938.5 933.875

23 917 915.375

9 917.875 925.375 24 884.5 902.25

10 884.5 903.875 25 947.25 955.125

11 927.625 929.125 26 947.875 960.375

12 976.625 966.375 27 938.875 935.75

13 863.25 879.75 28 916.625 908.875

14 927.25 926.625 29 947.25 952

Page 66: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 6

15 947.625 956 30 939.125 942.25

sidd3_dps_5v Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 1.1708 1.137 16 1.0595 1.0664

2 1.0524 1.1454 17 0.9972 0.988

3 0.9987 0.9926 18 1.0391 1.0404

4 1.0774 1.0751 19 1.0054 1.0049

5 1.0146 1.034 20 1.0079 1.0267

6 1.0304 1.0399 21 1.0141 1.0284

7 1.0155 1.0345 22 1.0667 1.069

8 1.0483 1.0593 23 1.081 1.0824

9 1.0682 1.0712 24 1.0876 1.0873

10 1.0406 1.044 25 1.0437 1.0544

11 1.0064 1.0069 26 1.0395 1.043

12 1.0222 1.0358 27 1.0616 1.0681

13 1.1611 1.0977 28 1.157 1.0939

14 1.1943 1.1403 29 1.0825 1.0865

15 1.0527 1.063 30 1.0963 1.0929

ridd_FBE_2M_dps Parametric Comparison

FRN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 647.1514 638.0637 16 649.8568 648.7715

2 647.3558 633.7004 17 649.1944 642.3926

3 647.2025 641.9342 18 649.9078 647.0403

4 640.0736 644.5455 19 650.2135 646.7723

Page 67: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 6

5 639.9207 643.5129 20 659.3853 640.2686

6 640.2264 649.2806 21 658.5699 652.7308

7 639.9207 646.976 22 659.7928 641.0323

8 639.2073 649.2806 23 658.5699 634.9572

9 639.1564 639.6425 24 659.0286 624.9938

10 639.3602 647.0913 25 658.4171 633.7349

11 635.2329 659.7589 26 640.8379 642.3561

12 635.2839 639.6575 27 641.0417 646.8741

13 634.5196 639.5406 28 640.2774 653.4048

14 651.2325 648.5169 29 640.0226 650.2863

15 651.3345 644.4806 30 640.2774 659.2093

25C pmc_lpo_8b_5v sidd3_dps_5v ridd_FBE_2M_dps

t- test (P-value) 0.656 0.794 0.229

f- test (P-value) 0.685 0.146 0.407

For TSSOP 16:

• pmc_lpo_8b_5v

• sidd3_dps_5v

pmc _lpo_8b_5v Parametric Comparison

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(us) (us)

KES-

CHINA

Unit of

Measure

(us) (us)

1 946.00 952.00 16 1022.60 1046.60

2 988.50 993.88 17 995.75 1038.90

3 946.63 952.00 18 979.50 984.13

4 988.50 992.88 19 1022.30 1040.40

5 977.88 978.25 20 995.13 1033.50

6 946.00 951.63 21 946.00 928.50

7 993.50 1004.50 22 1023.90 1054.70

8 992.25 1001.60 23 990.13 999.50

9 1074.10 1062.30 24 987.63 985.75

Page 68: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 6

10 994.13 1017.20 25 979.13 979.75

11 977.63 975.13 26 947.00 952.25

12 1034.50 1023.50 27 978.25 978.25

13 977.00 966.38 28 992.88 1002.00

14 992.63 1001.60 29 994.50 1022.90

15 1022.90 1048.20 30 988.50 992.63

sidd3_dps_5v Parametric Comparison

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(uA) (uA)

KES-

CHINA

Unit of

Measure

(uA) (uA)

1 996.00 1003.70 16 1022.60 982.06

2 988.50 976.45 17 975.75 945.48

3 1016.63 1037.90 18 1002.50 999.41

4 978.50 969.30 19 970.23 963.86

5 977.88 999.41 20 995.13 943.27

6 1010.95 998.06 21 1001.20 1021.00

7 923.50 942.93 22 953.90 947.86

8 992.25 972.36 23 970.13 942.93

9 944.10 959.77 24 987.63 978.49

10 1004.13 971.34 25 979.13 943.95

11 1007.62 1017.50 26 987.00 1018.80

12 954.50 936.63 27 998.25 1046.00

13 1100.21 1009.30 28 992.88 980.53

14 962.63 949.39 29 974.50 940.37

15 972.90 955.69 30 998.50 989.72

Results: Pass

25C pmc_lpo_8b_5v sidd3_dps_5v

t- test (P-value) 0.287 0.221

f- test (P-value) 0.309 0.900

Page 69: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

9S08SH8/SH4/SQ4 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 6

5.0 Analysis Conclusions and Verification Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 28 Jun

2010 Jennifer Li (KES-CHINA, TPE)

James Liu (TJN MSG PE)

Page 70: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 9

[Customer Version]

1. Objective:

This report details the correlation result of the HC908JL8(LQFP 32) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

• Correlation Lot Information:

Lot # Qty.

YZME001ZY300 1159

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on of the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C,135C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 135C,25C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C,135C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 71: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 9

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 135C and 25C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: Tables below summarize the correlation results. All numbers filled in green boxes means correlation was achieved. All numbers filled in yellow boxes means results did not correlate and requires further explanation.

Bin-to-Bin Correlation Matrix for Hot& Room Test

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

Page 72: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 9

4.2 QA Gate results (tested at spec limits): Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 135 C Room 25C Cold -40C

Lot # Pass # Fail # Pass # Fail # Pass # Fail

YZME001ZY300 100% 0 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• SIDD_LVI_EN_HV

• SIDD_LVI_DIS_HV

• LVI_Trip_HV

• LVI_Recover_HV

SIDD_LVI_EN_HV Parametric Comparison

FHN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 13.5197 13.8941 16 13.6519 13.5116

2 13.8732 13.4636 17 13.9536 14.0757

3 14.2599 15.1328 18 12.733 13.2199

4 13.9536 14.6592 19 13.877 14.0216

5 13.4603 13.6421 20 13.6774 13.5146

6 14.1925 13.2964 21 13.0393 13.2709

7 13.0138 13.2311 22 13.2134 13.6646

8 14.126 13.9737 23 13.4942 13.8717

9 12.9071 13.628 24 12.805 13.577

10 13.269 13.4606 25 12.754 13.5371

11 13.0603 13.6421 26 13.2062 13.1831

12 13.218 13.3475 27 13.4942 13.73

13 13.1879 13.6646 28 13.754 13.5626

14 14.1879 13.6421 29 13.6983 13.9227

Page 73: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 9

15 13.4686 13.679 30 13.8816 13.577

FRN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 13.6913 13.6901 16 13.3157 13.2566

2 13.5637 13.6421 17 13.385 13.4126

3 13.2137 13.1689 18 13.4177 13.475

4 14.2529 14.0471 19 13.4687 13.5371

5 13.4177 13.4636 20 14.1253 13.9737

6 13.819 13.73 21 14.1508 14.0106

7 14.304 14.9142 22 12.8235 13.1546

8 14.0998 13.8462 23 13.3667 13.3871

9 13.7934 13.7045 24 14.2529 14.0871

10 13.2647 13.1801 25 14.0998 13.9706

11 14.0742 13.7696 26 13.9211 13.7441

12 13.8956 13.73 27 13.4432 13.526

13 14.2784 14.6592 28 13.5452 13.6136

14 13.7424 13.6931 29 14.0742 13.8207

15 13.3412 13.3219 30 13.4871 13.5515

SIDD_LVI_DIS_HV Parametric Comparison

FHN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(nA) (nA)

Pat#

(nA) (nA)

1 431.6724 430.8664 16 337.2245 364.5539

2 535.8841 502.2441 17 489.4897 525.1976

3 334.0401 353.5115 18 452.0936 440.2339

4 324.9081 328.8472 19 395.9354 410.4626

5 472.0679 471.6396 20 492.9359 458.8877

6 472.5148 452.9871 21 334.6719 353.5115

Page 74: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 9

7 358.9348 344.1501 22 264.9338 254.8833

8 429.1198 428.2831 23 450.7788 484.3915

9 431.6724 432.5819 24 512.4635 552.4628

10 314.6975 323.7462 25 295.5655 280.3881

11 490.3833 453.7869 26 453.3315 494.593

12 385.7248 404.5247 27 492.4891 473.3924

13 357.6457 367.1044 28 390.8301 405.3616

14 559.2791 520.0968 29 403.5933 418.0816

15 545.6479 562.6653 30 464.8568 450.4365

FRN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(nA) (nA)

Pat#

(nA) (nA)

1 302.155 318.6453 16 233.4711 219.1765

2 408.936 453.7869 17 327.0259 339.0198

3 395.537 412.1766 18 273.4711 261.6879

4 291.323 295.6909 19 475.7487 438.5179

5 406.822 397.6786 20 488.0739 463.9884

6 380.581 386.67 21 426.1714 442.7845

7 281.122 265.0852 22 142.6674 157.965

8 429.359 453.7869 23 430.643 412.9808

9 468.724 450.4365 24 404.269 394.322

10 427.245 400.2607 25 355.7648 367.1044

11 435.537 412.9808 26 495.7325 476.7403

12 431.585 401.974 27 482.5293 509.1016

13 485.521 456.3712 28 335.3421 351.8015

14 511.05 499.6938 29 370.3798 372.1747

15 551.456 521.8549 30 320.025 328.0049

Page 75: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 9

LVI _ Trip _ HV Parametric Comparison

FHN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(V) (V)

Pat#

(V) (V)

1 4.14 4.11 16 3.99 4.00

2 3.97 3.97 17 4.02 4.03

3 3.99 4.01 18 3.98 3.97

4 4.18 4.17 19 4.03 4.04

5 4.22 4.22 20 3.96 3.96

6 3.86 3.78 21 3.91 3.83

7 3.85 3.7 22 3.98 3.97

8 4.18 4.18 23 4.04 4.04

9 4.32 4.24 24 4.01 4.02

10 4.16 4.12 25 4.21 4.19

11 4.18 4.12 26 3.94 3.84

12 3.89 3.8 27 3.95 3.89

13 3.94 3.85 28 4.04 4.09

14 3.96 3.95 29 4.12 4.1

15 4.02 4.04 30 4.06 4.09

FRN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(V) (V)

Pat#

(V) (V)

1 4.27 4.24 16 3.87 3.79

2 4.05 4.12 17 3.98 4

3 3.91 3.83 18 4.18 4.2

4 4.04 4.11 19 3.97 3.89

5 3.87 3.8 20 4.2 4.21

6 3.98 3.97 21 3.98 4.03

7 3.98 4 22 3.84 3.7

8 3.93 3.86 23 3.98 3.97

Page 76: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 9

9 3.98 4.04 24 4 4.05

10 3.98 3.97 25 4.1 4.18

11 3.97 3.95 26 4.06 4.12

12 3.98 4.02 27 3.98 3.96

13 4.08 4.16 28 4.03 4.09

14 3.98 4.03 29 4.14 4.1

15 3.92 3.84 30 4.14 4.09

LVI _ Recover _ HV Parametric Comparison

FHN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(V) (V)

Pat#

(V) (V)

1 4.28 4.25 16 4.12 4.15

2 4.11 4.11 17 4.16 4.18

3 4.12 4.15 18 4.1 4.1

4 4.32 4.32 19 4.18 4.19

5 4.37 4.36 20 4.09 4.09

6 3.99 3.91 21 4.03 3.95

7 3.99 3.81 22 4.11 4.1

8 4.32 4.31 23 4.17 4.18

9 4.47 4.41 24 4.14 4.15

10 4.32 4.28 25 4.35 4.33

11 4.3 4.25 26 4.07 3.98

12 4.02 3.93 27 4.08 4.02

13 4.07 3.98 28 4.18 4.23

14 4.08 4.09 29 4.19 4.23

15 4.15 4.17 30 4.25 4.24

FRN

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(V) (V)

Pat#

(V) (V)

Page 77: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JL8Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 8 of 9

1 4.4 4.41 16 4 3.91

2 4.2 4.26 17 4.11 4.15

3 4.05 3.95 18 4.34 4.33

4 4.19 4.25 19 4.11 4.02

5 4.01 3.93 20 4.35 4.36

6 4.11 4.1 21 4.12 4.18

7 4.11 4.14 22 3.97 3.81

8 4.05 3.98 23 4.11 4.1

9 4.12 4.19 24 4.14 4.19

10 4.11 4.11 25 4.23 4.31

11 4.11 4.09 26 4.2 4.28

12 4.11 4.15 27 4.11 4.1

13 4.25 4.32 28 4.16 4.23

14 4.11 4.16 29 4.27 4.23

15 4.05 3.98 30 4.27 4.24

135C SIDD_LVI_EN_HV SIDD_LVI_DIS_HV LVI _ Trip _HV LVI_ Recover_ HV t- test (P-value) 0.096 0.910 0.433 0.516 f- test (P-value) 0.596 0.893 0.392 0.364

25C SIDD_LVI_EN_HV SIDD_LVI_DIS_HV LVI_ Trip_ HV LVI_ Recover_ HV t- test (P-value) 0.858 0.879 0.957 0.992 f- test (P-value) 0.963 0.750 0.102 0.076

Results: Pass

5.0 Analysis Conclusions and Verification � Bin-to-bin activity shows 100% correlation of 1159 units at Hot & Room temperature and between the

new KES-CHINA and the current FSL-TJN-FM.

� The QA Gate correlation passes at KES-CHINA.

� Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units,

passes. The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this

product.

6.0 Document History:

Rev Date Originator

Page 78: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

G9F8JL8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Transfer

Page 9 of 9

O 17 Jun

2010

Ji Chang (KES-CHINA, TPE)

Richard Shao (TJN MSG PE)

Page 79: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JW16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 4

[Customer Version]

1. Objective:

This report details the correlation result of the HC908JW16(LQFP 48) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

Note: HC908JW16 already qualified package QFN 48 in KES-CHINA. Please refer to enclosed file for detail.

JW32_QFN48__GEP9JW32__Correlation__Report.doc

2. General Information

* Correlation Lot Information:

Lot # Qty.

Engineer lot 30

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 80C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

3.2 General Method: Using the production test program, perform 100% screening of 30 units at 80C temperature, in the current FSL-TJN-FM. Test the same 30 units under the same conditions in KES-CHINA.

Page 80: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JW16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 4

Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results: 4.1 Parametric Comparison:

Tables below summarize the parametric data on 30 good units The following are significant parameters chosen for this comparison:

• Vreg25_Meas_MinVdd_HT

• sidd_ram_IRCDIS_exe

• ptb_led_ramexe_ptb0_hv_ht

Vreg25_Meas_MinVdd Parametric Comparison

FHN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(V) (V)

Pat#

(V) (V)

1 2.3067 2.3793 16 2.3609 2.4044

2 2.3137 2.3338 17 2.3628 2.3144

3 2.3158 2.3473 18 2.365 2.3721

4 2.317 2.3537 19 2.3679 2.3443

5 2.3201 2.3617 20 2.3689 2.3304

6 2.3297 2.3884 21 2.3697 2.3584

7 2.3303 2.3044 22 2.372 2.3406

8 2.339 2.3831 23 2.3731 2.3253

9 2.3426 2.3543 24 2.3767 2.3377

10 2.3475 2.3617 25 2.3772 2.3814

11 2.3498 2.3687 26 2.3825 2.3198

12 2.3503 2.3659 27 2.3897 2.3564

13 2.3581 2.3113 28 2.39 2.3743

14 2.3589 2.3645 29 2.397 2.3133

15 2.3603 2.3429 30 2.4117 2.3601

Page 81: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JW16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 4

sidd_ram_IRCDIS_exe Parametric Comparison

FHN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat#

(uA) (uA)

Pat#

(uA) (uA)

1 136.487 135.5607 16 142.9292 141.6864

2 122.7153 122.8735 17 124.6391 127.5956

3 133.8202 130.1618 18 141.5643 139.5084

4 134.8437 132.7147 19 131.9134 129.4542

5 137.7244 136.3325 20 137.7894 136.3617

6 142.2348 139.8445 21 139.7189 137.0026

7 133.6352 129.6192 22 136.7154 135.6998

8 136.2499 135.422 23 133.1677 129.4955

9 137.9296 136.4201 24 147.3105 142.1989

10 136.747 136.1949 25 148.3864 143.1362

11 140.0054 137.4953 26 123.1169 127.0462

12 133.8141 130.0498 27 137.1836 136.3164

13 124.7439 128.9145 28 142.8087 140.1161

14 138.1855 136.6808 29 126.871 129.2809

15 134.6213 131.2094 30 135.3674 135.374

ptb_led_ramexe_ptb0_hv_ht Parametric Comparison

FHN

KES-

CHINA

KES-

CHINA

Unit of

Measure

FSL-TJN

Unit of

Measure Unit of

Measure

FSL-TJN

Unit of

Measure

Pat

#

(mA) (mA)

Pat#

(mA) (mA)

1 25.4878 25.6624 16 25.2702 25.5189

2 26.055 26.7859 17 24.9158 25.218

3 25.1894 25.3691 18 24.6609 24.7574

4 24.8599 25.0133 19 24.6174 24.2498

5 24.8864 25.2005 20 25.4085 25.5376

6 25.5515 25.7873 21 25.0837 25.268

7 25.2205 25.3804 22 25.4505 25.6375

Page 82: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908JW16 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 4

8 25.9866 25.9371 23 25.1583 25.3617

9 24.7496 24.9307 24 24.8739 25.0694

10 25.1039 25.3316 25 24.835 24.9557

11 25.7193 25.8115 26 24.4699 24.1831

12 26.0053 26.193 27 25.5204 25.674

13 24.6609 24.6558 28 25.0107 25.2629

14 25.5562 25.7935 29 24.7185 24.8869

15 24.8739 25.1243 30 24.9904 25.2442

80C Vreg25_Meas_MinVdd sidd_ram_IRCDIS_exe ptb_led_ramexe_ptb0_hv_ht t- test (P-value) 0.454 0.338 0.198 f- test (P-value) 0.773 0.180 0.253

Results: Pass

5.0 Analysis Conclusions and Verification Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 25 Jun

2010

Jennifer Li (KES-CHINA, TPE)

Irene Feng(TJN MSG PE)

Page 83: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 1 of 7

[Customer Version]

1. Objective:

This report details the correlation result of the HC908MR8(LQFP 32) device to demonstrate the KES-CHINA, Tianjin, China Final Test facility correlates with the current FSLTJN-FM, Tianjin, China Final Test facility. The correlation units were first tested at the current FSL-TJN-FM Final Test facility and then both rejects and good units were tested at the KES-CHINA Final Test facility

2. General Information

* Correlation Lot Information:

Lot # Qty.

LJMEA0MRZM00 1183

3. Qualification Plan: 3.1 Requirements: The requirements and acceptance criteria were based on the Freescale specification 12MQS10009G.

Correlation Plan

Name Detail Description

Qual per Correlation Plan, (Y / N ?)

Acceptance Criteria

a Bin-to-bin

Correlation

Goal: Correlate bin-to-bin per

production test flow

Lot quantity: 1

Units per lot: 1000

Temperature: 25C,115C

Type of units: good, reject

Site: FSL-TJN-FM / KES-CHINA

Y 100% correlation for all good and reject units.

b QA Gate Qual

Goal: 100% QC Gate, 0 failures

Lot quantity: 1

Units per lot: 1000

Temperature: 115C, -40C

Type of units: good

Site: KES-CHINA

Y 0 failures.

c Parametric

Comparison

Goal: Parametric comparison for

Identify Parametric Tests

Lot quantity: 1

Units per lot: 30

Temperature: 25C,115C

Type of units: good

Site: FSL-TJN-FM/ KES-CHINA

Y

t-test and f-test should not demonstrate significant

difference. P-Unit of Measure

should be greater than 0.05.

Page 84: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 2 of 7

3.2 General Method: Using the production test program, perform 100% screening of 1 lot at 25C and 115C temperature, in the current FSL-TJN-FM. Test the same lots under the same conditions in KES-CHINA. Any miss-correlations at the new Test site name are sent to current Test site name for evaluation.

4.0 Correlation/Qualification Results:

4.1 Bin-to-bin Correlation Results: � Tables below summarize the correlation results. All numbers filled in green boxes means

correlation was achieved. � All numbers filled in yellow boxes means results did not correlate and requires further

explanation.

Bin-to-Bin Correlation Matrix for Room & Hot Test

FRN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

FHN KES-CHINA RESULT

BIN 1 2 3 4 5 6 7 8

1 PASS

2 100%

3 100%

4 100%

5 100%

6 100%

7 100%

FS

L-T

JN

-FM

RE

SU

LT

8 100%

Results: Pass

Page 85: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 3 of 7

4.2 QA Gate results (tested at spec limits):

Table below shows the QA Gate results Qual in KES-CHINA.

QA Gate Results

QA Gate Hot 115C Cold -40C

Lot # Pass # Fail # Pass # Fail

LJMEA0N4RA01 100% 0 100% 0

Results: Pass

4.3 Parametric Comparison: Tables below summarize the parametric data on 30 good units. The following are significant parameters chosen for this comparison:

• widd_exec_ptm0

• sidd_exec_C_k94r_ptm0

• siddLVI_exec_k94r_ptm0

widd_exec_ptm0 Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(mA) (mA)

KES-

CHINA

Unit of

Measure

(mA) (mA)

1 6.2669 6.2234 16 6.0516 6.0644

2 6.4053 6.3335 17 6.0574 6.0654

3 6.0015 6.0144 18 6.1324 6.1551

4 6.1249 6.1082 19 6.2174 6.2275

5 6.3598 6.2408 20 6.2434 6.1806

6 6.2659 6.199 21 6.126 6.1857

7 6.4372 6.3825 22 6.1097 6.1174

8 6.1452 6.251 23 6.1147 6.1643

9 6.0984 6.1225 24 6.0365 6.0929

10 6.1751 6.0715 25 6.3036 6.2377

11 6.2612 6.2418 26 6.2608 6.2469

12 6.1717 6.1888 27 6.2398 6.1877

13 6.1218 6.1643 28 6.1495 6.1847

14 5.9923 6.097 29 6.1874 6.1541

Page 86: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 4 of 7

15 6.0413 6.0042 30 6.1658 6.1582

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(mA) (mA)

KES-

CHINA

Unit of

Measure

(mA) (mA)

1 6.2653 6.2224 16 6.0423 6.0644

2 6.4351 6.3346 17 6.1495 6.1541

3 6.0056 6.0174 18 6.2689 6.2265

4 6.1228 6.1062 19 6.2148 6.1847

5 6.3812 6.2469 20 6.2378 6.1877

6 6.2608 6.2 21 6.1239 6.1123

7 6.3874 6.3845 22 6.1676 6.1633

8 6.3036 6.2561 23 6.0974 6.0868

9 6.125 6.1215 24 6.3334 6.2326

10 6.0547 6.0674 25 6.3354 6.2408

11 6.2618 6.2418 26 6.2414 6.1877

12 6.2577 6.1888 27 6.2195 6.1857

13 6.1658 6.1633 28 6.1605 6.1541

14 6.1106 6.098 29 6.3114 6.3417

15 5.9944 6.0072 30 6.1462 6.152

sidd_exec_C_k94r_ptm0 Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 -645.6233 -664.795 16 -560.748 -539.289

2 -624.7145 -655.102 17 -536.49 -521.943

3 -519.9441 -510.209 18 -581.877 -564.288

4 -477.6103 -461.741 19 -613.495 -612.246

5 -587.9966 -569.9 20 -590.33 -584.696

Page 87: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 5 of 7

6 -559.7277 -536.738 21 -492.912 -465.312

7 -545.9564 -522.453 22 -473.02 -433.681

8 -573.2075 -556.125 23 -590.547 -588.777

9 -552.587 -529.085 24 -597.176 -592.859

10 -508.9512 -502.046 25 -555.869 -532.657

11 -521.9843 -511.229 26 -555.869 -533.677

12 -584.7199 -565.309 27 -573.718 -558.676

13 -578.3072 -561.737 28 -519.434 -505.617

14 -517.9039 -503.576 29 -529.635 -512.249

15 -561.9881 -554.085 30 -523.004 -512.249

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(nA) (nA)

KES-

CHINA

Unit of

Measure

(nA) (nA)

1 -707.5291 -724.487 16 -633.185 -615.307

2 -646.4438 -653.571 17 -616.866 -593.879

3 -590.3907 -581.125 18 -612.276 -588.777

4 -613.6945 -615.307 19 -581.969 -571.941

5 -540.8799 -525.514 20 -526.374 -490.311

6 -614.1016 -589.798 21 -508.012 -546.942

7 -535.5546 -502.046 22 -608.706 -584.696

8 -545.2455 -546.942 23 -549.326 -565.819

9 -577.8884 -565.819 24 -624.515 -596.43

10 -578.3984 -568.88 25 -635.014 -620.409

11 -637.7743 -627.041 26 -595.23 -581.125

12 -628.0849 -613.777 27 -580.949 -569.9

13 -627.0649 -598.981 28 -656.133 -674.489

14 -699.9024 -713.773 29 -586.559 -574.492

15 -695.9109 -695.917 30 -536.065 -516.841

Page 88: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 6 of 7

sidd LVI _exec_k94r_ptm0 Parametric Comparison

FHN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(uA) (uA)

KES-

CHINA

Unit of

Measure

(uA) (uA)

1 270.7845 269.1443 16 269.5298 268.0526

2 272.1209 270.5981 17 267.5879 267.4456

3 263.503 263.3952 18 272.0087 270.2461

4 268.3324 268.0016 19 264.4872 265.7469

5 265.14 266.4151 20 265.3236 266.6957

6 273.3604 272.6743 21 264.1558 264.9511

7 273.8296 272.761 22 271.8251 269.6034

8 266.6189 267.2619 23 266.2976 266.7263

9 271.8251 269.9655 24 271.1773 269.3687

10 273.3349 272.108 25 267.338 267.3231

11 267.5406 267.4048 26 271.82 269.4401

12 269.9328 268.7821 27 265.128 265.9917

13 266.3588 266.9304 28 264.1711 265.3592

14 266.0478 266.7263 29 264.9564 265.8897

15 263.5183 263.4565 30 269.8562 268.323

FRN

Pat#

FSL-TJN

Unit of

Measure

Pat#

FSL-TJN

Unit of

Measure

KES-

CHINA

Unit of

Measure

(uA) (uA)

KES-

CHINA

Unit of

Measure

(uA) (uA)

1 271.8197 269.7615 16 267.9738 267.5476

2 272.4063 270.7868 17 271.9574 269.9604

3 263.3139 263.6044 18 265.0325 266.155

4 269.8152 268.2413 19 266.6542 266.8283

5 266.5165 266.762 20 263.9411 264.9715

6 273.9977 272.9803 21 272.5542 271.3837

7 273.9109 272.7763 22 265.7719 266.6345

8 267.8526 267.4303 23 270.0345 269.0116

Page 89: 9RS08KA8 Correlation Report - NXP Semiconductors · 9RS08KA8 Correlation Report for FSL-TJN-FM to KES-CHINA Final Test Site Expansion Page 4 of 4 6 1.1969 1.197 21 1.1802 1.1859 7

HC908MR8 Correlation Report

for FSL-TJN-FM to KES-CHINA Final Test Site Expansion

Page 7 of 7

9 271.6718 269.7003 24 266.8326 267.216

10 273.054 271.9856 25 272.2227 270.5726

11 268.1382 267.5833 26 264.9407 266.0325

12 271.279 269.0524 27 264.4817 265.2061

13 267.5823 267.3333 28 264.8489 265.4153

14 266.6797 266.9609 29 269.5754 268.0475

15 263.6402 263.686 30 265.2603 266.3131

115C widd_exec_ptm0 sidd _ exec _C_k94r_ptm0 siddLVI_exec_k94r_ptm0

t- test (P-value) 0.798 0.281 0.622 f- test (P-value) 0.140 0.285 0.085

25C widd_exec_ptm0 sidd _ exec _C_k94r_ptm0 siddLVI_exec_k94r_ptm0

t- test (P-value) 0.287 0.512 0.550 f- test (P-value) 0.194 0.576 0.110

Results: Pass

5.0 Analysis Conclusions and Verification Bin-to-bin activity shows 100% correlation of 1183 units at hot&room temperature between the new KES-

CHINA and the current FSL-TJN-FM.

The QA Gate correlation passes at KES-CHINA.

Parametric comparison between new KES-CHINA and the current FSL-TJN-FM for 30 good units, passes.

The results demonstrate that the new KES-CHINA Final Test Site has qualified as a test site for this product.

6.0 Document History: Rev Date Originator

O 17 Jun

2010

Jennifer Li (KES-CHINA, TPE)

Dai Xiaojing (TJN MSG PE)