9279

UNIVERSALCLAMP ON CT

INSTRUCTION MANUAL

sharif_geraldizo
Tequipment
Page 2: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained
Page 3: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

Contents

Introduction 1Inspection 1Preliminary Checks 1Safety Notes 2Notes on Use 5

Chapter 1 Overview 91.1 Product Overview 91.2 Names of Parts 10

Chapter 2 Measurement Procedure 112.1 Measurement Procedure 112.2 Demagnetizing (DMAG) 14

Chapter 3 Specification 15

Chapter 4 Maintenance and Service 194.1 Maintenance 194.2 Service 19

Page 4: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained
Page 5: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

1__________________________________________________

_____________________________________________

Introduction

Inspection

Preliminary Checks

Thank you for purchasing the HIOKI "9279UNIVERSAL CLAMP ON CT". To obtainmaximum performance from the product, pleaseread this manual first, and keep it handy forfuture reference.

When you receive the product, inspect itcarefully to ensure that no damage occurredduring shipping. In particular, check theaccessories. If damage is evident, or if it fails tooperate according to the specifications, contactyour dealer or Hioki representative.

Before using the product the first time, verifythat it operates normally to ensure that the nodamage occurred during storage or shipping.If you find any damage, contact your dealeror Hioki representative.Before using the product, make sure that theinsulation on the cables is undamaged andthat no bare conductors are improperlyexposed. Using the product in suchconditions could cause an electric shock, socontact your dealer or Hioki representative forrepair.

Page 6: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

2__________________________________________________

_____________________________________________

DANGERThis product is designed to conform to IEC 348Safety Standards, and has been thoroughlytested for safety prior to shipment. However,mishandling during use could result in injury ordeath, as well as damage to the product. Becertain that you understand the instructions andprecautions in the manual before use. Wedisclaim any responsibility for accidents orinjuries not resulting directly from productdefects.

The symbol printed on the product indicatesthat the user should refer to a correspondingtopic in the manual (marked with the symbol)before using the relevant function.In the manual, the symbol indicatesparticularly important information that the usershould read before using the product.

Safety Notes

Safety symbolsThis manual contains information and warningsessential for safe operation of the product andfor maintaining it in safe operating condition.Before using the product, be sure to carefullyread the following safety notes.

Page 7: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

3__________________________________________________

_____________________________________________

DANGERIndicates that incorrect operation presents anextreme hazard that could result in seriousinjury or death to the user.

WARNINGIndicates that incorrect operation presents asignificant hazard that could result in seriousinjury or death to the user.

CAUTIONIndicates that incorrect operation presents apossibility of injury to the user or damage tothe product.

NOTEAdvisory items related to performance orcorrect operation of the product.

The following symbols in this manual indicatethe relative importance of cautions andwarnings.

Page 8: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

4__________________________________________________

_____________________________________________

f.s. (maximum display or scale value, or length ofscale)Signifies the maximum display (scale) value or thelength of the scale (in cases where the scaleconsists of unequal increments or where themaximum value cannot be defined).In general, this is the range value (the valuewritten on the range selector or equivalent)currently in use.

rdg. (displayed or indicated value)This signifies the value actually being measured,i.e., the value that is currently indicated ordisplayed by the measuring instrument.

AccuracyThe specifications in this manual include figuresfor "measurement accuracy" when referring todigital measuring instruments, and for"measurement tolerance" when referring toanalog instruments.

Page 9: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

5__________________________________________________

_____________________________________________

DANGERTo avoid short circuits and potentially life-threatening hazards, never attach theUNIVERSAL CLAMP ON CT to a circuit thatoperates at more than the 600 Vrms, or overbare conductors.UNIVERSAL CLAMP ON CT should only beconnected to the secondary side of a breaker,so the breaker can prevent an accident if ashort circuit occurs. Connections should neverbe made to the primary side of a breaker,because unrestricted current flow could cause aserious accident if a short circuit occurs.

WARNINGTo avoid electric shock, do not allow theproduct to get wet, and do not use it when yourhands are wet.To avoid electric shock when measuring livelines, wear appropriate protective gear, such asinsulated rubber gloves, boots and a safetyhelmet.

Notes on UseFollow these precautions to ensure safeoperation and to obtain the full benefits of thevarious functions.

Page 10: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

6__________________________________________________

_____________________________________________

CAUTION

Note that the product may be damaged if currentexceeding the selected measurement range isapplied for a long timeDo not store or use the product where it could beexposed to direct sunlight, high temperature orhumidity, or condensation. Under such conditions,the product may be damaged and insulation maydeteriorate so that it no longer meets specifications.To avoid damage to the product, protect it fromvibration or shock during transport and handling,and be especially careful to avoid dropping.Measurements are degraded by dirt on the matingsurfaces of the clamp-on sensor, so keep thesurfaces clean by gently wiping with a soft cloth.Avoid stepping on or pinching the cable, whichcould damage the cable insulation.Keep the cables well away from heat sources, asbare conductors could be exposed if the insulationmelts.To avoid damaging the cables, do not bend or pullthe cables.This product should be installed and operatedindoors only, between 0 and 40 . However, it canbe safely operated down to -10 .When measuring the high-frequency large current,over-heating of the core may occur. Note that theinput never exceed the maximum permissible inputrange. See the figures in Chapter 3,"Specifications".

Page 11: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

7__________________________________________________

_____________________________________________

NOTE Accurate measurement may be impossible inthe presence of strong magnetic fields, suchas near transformers and high-currentconductors, or in the presence of strongelectromagnetic fields such as near radiotransmitters.

Page 12: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

8__________________________________________________

_____________________________________________

Page 13: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

9__________________________________________________

Chapter 1 Overview_____________________________________________

Chapter 1Overview

1.1 Product OverviewThe 9279 was developed for to provide a 500 Aclamp sensor corresponding to AC/DC current.The 9279 makes it possible to measure AC/DCcurrent in live power lines without cutting intothe lines. The sensor features good frequencyresponse (amplitude and phase), goodtemperature response (sensitivity and offset),and dielectric strength and is easy to connectand use. Its versatility will find application in awide variety of fields dealing with current andpower measurement.

Page 14: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

10__________________________________________________

Chapter 1 Overview_____________________________________________

Current direction indicationLever

Clamp core

Cable

Output connector

DMAG switch

1.2 Names of Parts

Page 15: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

11__________________________________________________

Chapter 2 Measurement Procedure_____________________________________________

CAUTION

To avoid damaging the output cable, grasp theconnector, not the cable, when unplugging the cable.To prevent damage to the product and sensor, neverconnect or disconnect a sensor while the power is on.The maximum continuous input range is based on heatthat is internally generated during measurement. Neverinput current in excess of this level. Exceeding therated level may result in damage to the cable.The maximum continuous input range varies accordingto the frequency of the current being measured. Seethe figures in Chapter 3, "Specifications"

Chapter 2Measurement Procedure

2.1 Measurement Procedure

Page 16: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

12__________________________________________________

Chapter 2 Measurement Procedure_____________________________________________

NOTE To measure small current levels, multiplewindings may be used to increase the relativesensitivity. 10 windings multiplies themeasured current by about a factor of 10.However, in this case the diameter of thewinding should be 20 cm or more, and radial.

This product has a resistance in series with thesignal output circuits. Therefore, if measuringan output signal directly, ensure that the inputresistance of the measuring instrument issufficiently high.

Page 17: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

13__________________________________________________

Chapter 2 Measurement Procedure_____________________________________________

Current directionindication

NOTE Grip the clamp lightly so that the lever snapssecurely shut. At that time, avoid to subjectingthe facing surface of the core to shock.Make sure that only one conductor is in thecore. Single-phase (2-wire) and three-phase(3-wire) lines clamped together will not producereading.

1. Close the clamp jaws whenthe clamp jaws are opened.

2. Make sure that the unit'spower switch is OFF.

3. Connect the clamp-on CTconnector to the unit.

4. Turning the power on the unit.5. Press the DMAG switch to

magnetize.(See 2.2 Demagnetizing(DMAG))

6. Lightly press the lever andspread the tips of the clampwith both hands.

7. Align the sensor so that thecurrent direction indicationcorresponds to the directionof current flow through theconductor to be measured,and clamp so that theconductor is in the center ofthe sensor aperture.

Page 18: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

14__________________________________________________

Chapter 2 Measurement Procedure_____________________________________________

NOTE After powering on, press DMAG.It is possible to magnetize even if pressingDMAG during input.Always make sure that the clamp jaws arecompletely closed during demagnetising.Demagnetization occurs for about 3 seconds.If an excessive amount of current was input sothat one attempt at demagnetization does notcompletely rid of the remanence, open theclamp jaws and press DMAG, then close theclamps, and press DMAG again.In case of closing the clamp and if the facingsurface of the core is subjected to shock,offset maybe output. In this case, press theDMAG switch with no input on the clamp.When measuring DC current and applyingexceed continuous the maximum permissibleinput, the malfunction may occur. Press theDMAG switch with no input on the clamp, toreturn to normal.If it does not return to normal condition, oreven pressing DMAG, the magnetic isremained in the core, contact your dealer orHioki representative.

2.2 Demagnetizing (DMAG)After measuring rush current or similar of AC orDC current measurement, micro voltage isoutput even if there is no input on the clamp.This is caused by the core characteristic usingthe clamp and magnetic is remained in thecore. This remanent cause measuring error, butit can be removed by pressing DMAG with noinput onto the clamp.

Page 19: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

15__________________________________________________

Chapter 3 Specification_____________________________________________

Rated current(AC/DC)

500 A f.s.

Output voltage(AC/DC)

2 V/500 A

The maximumpermissible inputrange (DC to 3 kHz)

650 Arms (920 Apeak)

Input resistance (DC) 0.001 mΩ or lessOutput resistance 50 ΩBasic accuracy

23 3(73 5 )

Period of guaranteedaccuracy

DC and 45 Hz to 66 Hz, 30 min ormore warming-up after degaussingAmplitude: 0.5% rdg. 0.05% f.s.Phase: within 0.2 (DC has no provision)1 year

Amplitude-frequencycharacteristic(deviation fromaccuracy)

DC to 1 kHz within 1.0%1 k to 10 kHz within 2.5%10 k to 20 kHz within 5.0%

Phase-frequencycharacteristic

DC to 1 kHz within 0.51 k to 10 kHz within 2.510 k to 20 kHz within 5.0

Temperaturecoefficient

Sensitivity: within 0.05% rdg. /Offset: within 0.005% f.s. /

Operating temperatureand humidity range

0 to 40 (32 to 104 ),Max. 80%RH (no condensation)

Storage temperatureand humidity range

-10 to 50 (14 to 122 ),Max. 80%RH (no condensation)

Effect of conductorposition

Within 1.5% (DC, 55 Hz)

Effect of externalmagnetic field (400A/m, 55 Hz and DC)

Max. 2 A

Chapter 3Specification

Page 20: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

16__________________________________________________

Chapter 3 Specification_____________________________________________

Dielectric strength 2200 Vrms for 1 min.(between electric circuit and case,between electric circuit and core)

Insulation resistance 500 V DC 100 MΩ or more (betweenelectric circuit and case, betweenelectric circuit and core)

Maximum ratedvoltage to earth

600 Vrms (850 Vpeak) insulatedconductor

Diameter ofmeasurableconductors

40 mm (1.57") or less

Supply voltage 12 V to 15 V (with accuracyguaranty but tracking)

Power supplycapacity

90 mA (with no load)250 mA (with rated input)

Supply consumption Max. 7.2 W (with rated input)Dimensions andmass

Approx. 220W 103H 43.5D mm(8.66"W 4.06"H 1.71"D)Approx. 470 g (16.6 oz.)

Cord length Approx. 3 m (9.84 feet)Accessories 9375 CARRYING CASE 1

Instruction manual 1Markband 6 (3 set)

Page 21: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

17__________________________________________________

Chapter 3 Specification_____________________________________________

Power supply GND 1Power supply (+) 2Power supply (-) 3

Output (-) 4Output (+) 5Output (-) 6

Demagnetizingsignal

ID3 7Identifying signal ID1 8Identifying signal ID2 9

Shield (cable) 10

GND+12 V (to +15 V)-12 V (to -15 V)

Vout (2 Vf.s./rated input)GND

Output connector pin array

9279

ID1 N.CID2 Connect to GND

CAUTION

Be careful to avoid connecting voltage improperly,as the internal circuitry may be destroyed.The capacity of the power supply is at least 0.5 A.Demagnetization occurs after pin 7 is shorted toground and then opened.

Output connector pin array

Mating receptacleRM515ERB-10SD (HIROSE)

Page 22: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

18__________________________________________________

Chapter 3 Specification_____________________________________________

Frequency

Inpu

t cur

rent

9279 The maximum permissible input range (continuous)

Page 23: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

19__________________________________________________

Chapter 4 Maintenance and Service_____________________________________________

Chapter 4Maintenance and

Service

4.1 Maintenance

4.2 Service

To clean the product, wipe it gently with a softcloth moistened with water or mild detergent.Never use solvents such as benzene, alcohol,acetone, ether, ketones, thinners or gasoline, asthey can deform and discolor the case.

If the product seems to be malfunctioning,contact your dealer or Hioki representative.Pack the product carefully so that it will notbe damaged during shipment, and include adetailed written description of the problem.Hioki cannot be responsible for damage thatoccurs during shipment.

Page 24: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

20__________________________________________________

Chapter 4 Maintenance and Service_____________________________________________

Page 25: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

HIOKI 9279 UNIVERSAL CLAMP ON CTInstruction Manual

Publication date: March 2004 Revised edition 3

Edited and published by HIOKI E.E. CORPORATIONTechnical Sales Support Section

All inquiries to International Sales and MarketingDepartment

81 Koizumi, Ueda, Nagano, 386-1192, JapanTEL: +81-268-28-0562 / FAX: +81-268-28-0568E-mail: [email protected] http://www.hioki.co.jp/

Printed in Japan 9279A981-03

All reasonable care has been taken in the production of thismanual, but if you find any points which are unclear or inerror, please contact your supplier or the International Salesand Marketing Department at HIOKI headquarters.

In the interests of product development, the contents of thismanual are subject to revision without prior notice.

Unauthorized reproduction or copying of this manual isprohibited.

Page 26: 9279 UNIVERSAL CLAMP ON CTg...DMAG switch with no input on the clamp, to return to normal. If it does not return to normal condition, or even pressing DMAG, the magnetic is remained

Printed on recycled paper

9279A981-03 04-03H