16
Manual 400 MHz Passive High-Voltage Probe R&S ® RT-ZH10 1409.7720.02 Printed in Germany Test and Measurment www.allice.de Allice Messtechnik GmbH

400 MHz Passive High-Voltage Probe RT-ZH10 · 2020. 11. 8. · Root mean square vs. Peak Pulse Voltage RT-ZH10 Measurement Category I 0 1000 2000 3000 4000 5000 6000 7000 8000 0200

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

  • Manual

    400 MHz Passive High-Voltage Probe R&S® RT-ZH10 1409.7720.02

    Printed in Germany

    Test and Measurment

    www.allice.de Allice Messtechnik GmbH

  • 2

    R&S RT-ZH10Manufacturer

    ROHDE & SCHWARZ

    For comprehensive information about Rohde and Schwarz, please visit our homepage on the R&S Internet(http://www.rohde-schwarz.com).

    For queries regarding technical aspects of our products, please contact our support athttp://www.customersupport.rohde-schwarz.com.

    For international services, please contact our service partners at http://www.service.rohde-schwarz.com.

    For additional questions, please contact our headquarters in Munich.

    ROHDE & SCHWARZ GmbH & Co. KGMühldorfstraße 15D-81671 MünchenPhone: +9 (89) 41 29 - 0Fax: +49 (89) 41 29 - 12 164

    www.allice.de Allice Messtechnik GmbH

  • 3

    R&S RT-ZH10Declaration of Conformity

    (EC conformity marking)

    The manufacturer declares the conformity of this product with the actual required safety standards in accordance with the Low Voltage Directive (LVD) 2006/95/EC:

    CEI/IEC 61010-031:2008 Safety requirements for electrical equipment for measurement, control and laboratory use -

    Part 031: Safety requirements for hand-held probe assemblies for electrical measurement and test

    WEEE/ RoHS Directives

    (EC conformity marking)

    This electronic product is classified within the WEEE/ RoHS* category list as monitoring and control equipment (category 9). Category 9 products are exempted from the restrictions under the scope of the RoHS directive.

    Your help and efforts are required to protect and keep clean our environment. Therefore return this electronic product at the end of its life either to the manufacturer or take care of separate WEEE collection and professional WEEE treatment yourself. Do not dispose as unsorted municipal waste!

    * EC Directives:WEEE Directive 2002/96/EC - Waste Electrical and Electronic EquipmentRoHS Directive 2002/95/EC - Restriction of the use of certain Hazardous Substances in Electrical and Electronic Equipment

    www.allice.de Allice Messtechnik GmbH

  • 4

    R&S RT-ZH10IEC Measurement Categories

    Definitions and Examples (Clause 6.5.2)

    Measurement Category I Definition: Measurement category I is for measurements performed on circuits not directly connected to a mains supply.

    Examples: Measurements in circuits not derived from a mains supply and specially protected ( internal ) circuits derived from a mains supply. In the latter case, transient stresses are variable; for that reason it is required that the transient withstand capability of the equipment is made known to the user.

    Measurement Category II Definition: Measurement category II is for measurements performedCAT II on circuits directly connected to the low voltage installation.

    Examples: Household appliances, portable tools and similar equipment.

    Measurement Category III Definition: Measurement category III is for measurements CAT III performed in the building installation.

    Examples: Measurements on distribution boards, circuit breakers, wiring including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation and equipment for industrial use like for example stationary motors with per- manent connection to the fixed installation.

    Measurement Category IV Definition: Measurement category IV is for measurements CAT IV performed at the source of the low-voltage installation.

    Examples: Electricity meters and measurements on primary over current protection devices and and ripple control units.

    www.allice.de Allice Messtechnik GmbH

  • 5

    R&S RT-ZH10IEC Pollution Degrees

    Definitions (Clause 3.5.6 )

    Pollution Degree 1 No POLLUTION or only dry, non conductive POLLUTION. NOTE: The POLLUTION has no influence.

    Pollution Degree 2 Only- non conductive POLLUTION. Occasionally, however, a temporary conductivity caused by condensation must be accepted.

    Pollution Degree 3 Conductive POLLUTION occurs or dry, non-conductive POLLUTION occurs which becomes conductive due to condensation which is to be expected.

    IEC Safety Symbols

    The following symbols may appear on the product or in this instruction manual:

    Caution, risk of danger. Refer to manual.

    Caution, risk of electric shock.

    Earth (ground) TERMINAL.

    www.allice.de Allice Messtechnik GmbH

  • 6

    R&S RT-ZH10Safety Information

    To avoid personal injury and to prevent fire or damage to this product or products connected to it, review and comply with the following safety precautions. Be aware that if you use this probe assembly in a manner not specified the protection this product provides may be impaired.Only qualified personnel should use this probe assembly.

    Use only grounded instruments.Do not connect the probe ground lead to a potential other than earth ground. Always make sure the probe and the measurement instrument are grounded properly.

    Connect and disconnect properly.Connect the probe output to the measurement instrument and connect the ground lead to earth ground before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground lead from the circuit under test before disconnecting the probe from the measurement instrument.

    Observe probe and probe accessory ratings.Do not apply any electrical potential to the probe input which exceeds the maximum ratings of the probe or the accessories connected to it. In a combination always the lower rating / measurement category applies to both probe and accessories connected to it. Make sure to comply with the voltage versus frequency derating curve on page 8.

    Keep away from live circuits.Avoid open circuitry. Do not touch connections or components when power is present.

    Do not operate with suspected failures.Refer to qualified service personnel.

    Indoor use only.Do not operate in wet/damp environment. Keep product surfaces dry and clean.

    Do not operate the product in an explosive atmosphere.

    www.allice.de Allice Messtechnik GmbH

  • 7

    R&S RT-ZH10Specifications

    Specifications that are not defined to be warranted are typical and are published as general information to the user. The instrument must have warmed-up for at least 20 minutes and the environmental conditions must not exceed the probe`s specified limits.

    Electrical Specifications

    Attenuation Ratio (1) 100:1 ± 2 % at DC (warranted)Voltage Coefficient 0.0005 %/V System Bandwidth 400 MHz (-3 dB) Probe Risetime 900 ps (10 % - 90 %) Input Resistance (System) 50 MΩ ± 1 % (warranted)Input Capacitance (System) 7.5 pF

    Maximum Rated Input Voltage (2) Measurement category I: 1000 V root mean square 4000 V transient overvoltageMeasurement category II: 1000 V root mean square CAT II

    Compensation Range 5 pF - 20 pF Input Coupling of the Measuring Instrument 1 MΩ AC / DC

    Pollution Degree (2) 2

    Voltage Derating

    Note that the max. input voltage rating of the probe decreases as the frequency of the applied signal increases.

    (1) Connected to appropriate R&S oscilloscope. See R&S oscilloscope operating manual for further details.(2) As defined in IEC 61010-031. Also see definitions explained on page 4 and 5.

    Typical voltage derating RT-ZH10Measurement Category I

    0

    200

    400

    600

    800

    1000

    1200

    100K 1M 10M 100M 1GFrequency [Hz]

    Ampl

    itude

    AC

    root

    mea

    n sq

    uare

    [V]

    www.allice.de Allice Messtechnik GmbH

  • 8

    R&S RT-ZH10Specifications

    Maximum Pulse Ratings

    For pulse measurments make sure to comply with the ratings as shown on this page.

    * Values at 10 ms also apply to all Pulse Durations smaller than 10 ms.

    Root mean square vs. Peak Pulse Voltage RT-ZH10Measurement Category I

    0

    1000

    2000

    3000

    4000

    5000

    6000

    7000

    8000

    0 200 400 600 800 1000

    AC root mean square voltage or DC [V]

    Peak

    Pul

    se V

    olta

    ge [V

    ]

    Maximum Pulse Derating RT-ZH10Measurement Category I

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    5000

    5500

    6000

    10* 100 1000

    50%

    25%

    10%

    Pulse Duration [ms]

    Max

    imum

    Pea

    k Pu

    lse

    Volta

    ge [V

    ]

    Duty Cycle

    www.allice.de Allice Messtechnik GmbH

  • 9

    R&S RT-ZH10Specifications

    Input Impedance

    Note that the input impedance of the probe decreases as the frequency of the applied signal increases.

    Mechanical Characteristics

    Weight (probe) 67 g Cable Length 2 mProbe Tip Diameter 5 mm

    Environmental specifications

    Altitude operating up to 2000 m non-operating up to 15000 mTemperature Range operating 0° C to +50° C non-operating -40° C to +71° CMaximum Relative Humidity operating 80 % relative humidity for temperatures up to +31° C, decreasing linearly to 40 % at +50° C

    Typical input impedance RT-ZH10

    10

    100

    1K

    10K

    100K

    1M

    10M

    100M

    10 100 1K 10K 100K 1M 10M 100M 1GFrequency [Hz]

    |Z| [

    Ohm

    ]

    www.allice.de Allice Messtechnik GmbH

  • 10

    R&S RT-ZH10Adjustment Procedures

    The probe can be adjusted for low frequency (LF) compensation and for high frequency (HF) compensation and DC dividing ratio.

    LF Compensation

    LF needs to be adjusted when the probe is connected to the oscilloscope input for the first time. LF compen-sation matches the probes cable capacitance to the oscilloscope input capacitance. This matching assures good amplitude accuracy from DC to upper bandwidth limit frequencies. A poorly compensated probe clearly influ-ences the overall system performance (probe + scope) and introduces measurement errors resulting in inaccurate readings and distorted waveforms.

    We recommend to use a square wave or rectangular wave generator with an amplitude of 5 Vpp minimum. When using a generator with smaller amplitude it may be useful to choose "RF Reject" in the trigger settings. Adjust the LF compensation trimmer to optimum square wave response. For clarification see below figures.

    undercompensated optimum overcompensated

    LF Compensation

    www.allice.de Allice Messtechnik GmbH

  • 11

    R&S RT-ZH10

    LF Compensation

    HF Compensation

    DC

    Adjustment Procedures

    HF Compensation

    It is typically not necessary to adjust HF compensation unless the probe is being used with an oscilloscope having large differences in input characteristics to the oscilloscope that was used for factory adjustment.We recommend to use the following equiment for proper HF compensation:

    Rectangular wave generator with a risetime faster than 700 ps, with an amplitude of 5 Vpp minimum,50 Ohm feed-through and probe BNC adapter.

    HF adjustment is performed by connecting the probe to the rectangular wave generator. Adjust trimmers for optimum square wave response.

    DC Adjustment

    In order to provide highest accuracy over the voltage range this probe´s DC dividing ratio is factory adjusted at 500 V using a measuring device with a precision input impedance of 1MΩ ± 0.01 %.

    optimum

    www.allice.de Allice Messtechnik GmbH

  • 12

    R&S RT-ZH10

    Flexible Adapter 5.0-L to 4 mm safety banana plug

    Sprung Hook 5.0-Lred

    Safety Alligator Clip red

    Safety Alligator Clip black

    Insulating Cap 5.0-Lblack

    BNC Adapter 5.0-L

    Ground Lead 22 cm to 4 mm banana plug

    Protection Cap 5.0-Lblack

    Ground Lead 22 cm

    Coding

    Rings (

    set)

    3x4 col

    ors

    Spring Tip 0.8 mm

    Solid Tip 0.8 mm

    Adj

    ustment

    Tool

    Probe Accessory

    www.allice.de Allice Messtechnik GmbH

  • 13

    R&S RT-ZH10Scope of Delivery

    The following items are included in the scope of delivery. Please check the delivery for completeness.

    RT-ZH10 400 MHz passive high-voltage probe (Order No. 1409.7720.02)

    Item Qty Probe RT-ZH10 1 Operating Manual 1 Adjustment Tool 1 BNC Adapter 5.0-L 1 Insulating Cap 5.0-L (black) 1 Protection Cap 5.0-L (black) 1(1) Coding Rings (set) 3x4 Colors 1 Ground Lead 22 cm 2 Ground Lead 22 cm to 4 mm banana plug 1 Safety Alligator Clip (black) 1 Safety Alligator Clip (red) 1 Solid Tip 0.8 mm 5 Spring Tip 0.8 mm 5(2) Sprung Hook 5.0-L (red) 2 Flexible Adapter 5.0-L to 4 mm safety banana plug 1

    (1) plugged on probe (2) one is installed in probe

    Use ground lead only for connections to earth ground.

    The BNC Adapter is rated: 100 V root mean square CAT II, Pollution Degree 1

    The accessories provided with the probe have been safety tested. Do not use any other accessories than those “originally” provided.

    www.allice.de Allice Messtechnik GmbH

  • 14

    R&S RT-ZH10Handling

    Handle with care especially when fitted with the extra thin and sharp spring contact tip to avoid any injury. Note that the probe cable is a sensitive part of the probe. Do not damage

    through excessive bending or pulling. Avoid mechanical shock to this product in general to guarantee accurate performance and protection.

    Maintenance

    Cleaning

    To clean the exterior of the probe use a soft cloth moistened with either distillated water or isopropyl alcohol. Before use allow the probe to dry completely.

    Changing the Probe Tip

    To change the probe tip use pliers to grip and pull it carefully straight out of its contact socket, along the axis of the probe. Do not grip the white plastic insulator or the housing with pliers, because the tip could be squeezed and cannot be removed and respectively the probe could be damaged.If the probe tip is removed, the new tip can be inserted with pliers into the contact socket, along the axis of the probe. In order to insert the probe tip completely into the housing, press the probe tip against a hard surface carefully.

    Use pliers to grip and pull the probe tip carefully out of its contact socket.

    Do not grip the white plastic insulator or the probe housing with pliers.

    www.allice.de Allice Messtechnik GmbH

  • Copyright®2011 ROHDE & SCHWARZ All rights reserved.Information in this publication supersedes that in all previously published material.Specifications are subject to change without notice.

    M-RT-ZH10 Revision A - Juli 2011

    www.allice.de Allice Messtechnik GmbH

  • © 2018 Allice Messtechnik GmbH – Alle Rechte vorbehalten.© 2018 Allice Messtechnik GmbH – All rights reserved

    make ALLICE your partner

    ALLICE Messtechnik GmbH

    Kelsterbacher Strasse 15-19 60528 Frankfurt am Main

    Tel.: +49(0)69-67724-583 Fax: +49(0)69-67724-582

    [email protected]

    www.allice.de

    Verwendete Warenzeichen und Schutzrechte sind Eigentum der jeweiligen Hersteller.Logos and company names listed are trademarks or trade names of their respective owners.

    AlliCEMesstechnik GmbH

    www.allice.de Allice Messtechnik GmbH