3D-FBK pixel sensors with CMS readout: first tests results M. Obertino, A. Solano, A. Vilela Pereira, E. Alagoz, J. Andresen, K. Arndt, G. Bolla, D. Bortoletto, M. Boscardin, J.M. Brom, R. Brosius, M. Bubna, J. Chramowicz, J. Cumalat, G-F. Dalla Betta, F. Jensen, A. Krzywda, A. Kumar, S. Kwan, CM Lei, D. Menasce, L. Moroni, , J. Ngadiuba, I. Osipenkov, L. Perera, M. Povoli, A. Prosser, R. Rivera, P. Tan, S. Terzo, N. Tran, L. Uplegger, S. Wagner More details in “Test-beam studies of diamond sensors for SLHC” L. Uplegger 12th Pisa Meeting on Advanced Detectors 3D sensors consist of an array of columnar electrodes (radius ∼ 5µm) of both doping types which penetrate entirely in the silicon substrate, perpendicularly to the surface. Fermilab testbeam Telescope with eight pixel planes of CMS Forward Pixel Detector. Pixel size is 100 µm x 150 µm, but enhanced resolution is derived from charge sharing by tilting the telescope planes at 25°. Telescope resolution on the DUT (Detector Under Test) ~ 6 µm. Rotation and cooling of DUTs possible Alignment done using a software developed at the University of Milano Bicocca. This structure decouples the electrode distance from the sensor substrate thickness. The close electrode spacing provides several advantages compared to the planar sensor design: low full depletion voltage (about -10 V), fast charge collection time, reduced charge trapping probability and therefore high radiation hardness. Spatial resolution (20° angle) 1 P I X E L C L U S T E R 2 P I X E L C L U S T E R 3D pixel sensors compatible with the CMS PSI46 readout chip were first fabricated at SINTEF (Oslo, Norway), and more recently at FBK (Trento, Italy) and CNM (Barcelona, Spain). Several sensors with different electrode con gurations (single n-type electrode (1E), two n-type electrodes (2E), and four n-type electrodes (4E) per pixel cell), all bump-bonded with the CMS pixel PSI46 readout chip, were characterized in laboratory and tested at Fermilab with a proton beam of 120 GeV/c, before and after irradiation. New productions of 3D sensors with different characteristics are foreseen in view of the CMS tracker upgrade. Results presented in this poster concentrate on one un-irradiated FBK 3D sensor tested in April 2012. FBK WAFER 3 1E CMS sensors with: 80 x 52 pixels of size 100 µm x 150 µm passing through columns (full 3D) 200 mm thickness slim edges (200 mm on two sides) bump bonded at Selex (Rome) and tested at Fermilab in April 2012 120 GeV/c protons Noise vs V BIAS behaviour as expected from CV measurements On wafer electrical test performed at FBK I-V curve measured after bump bonding Efficiency measurement Preliminary Preliminary Total collected cluster charge is ~13.3 ke -. The plots below show that the collected charge is what expected in 200 mm thick silicon (~16 ke - ) only if the track hits the central part of the pixel. Elsewhere it is reduced by the combined effect of charge sharing and readout chip threshold (~2500-3000 e - ). 0° angle 20° angle Noise measurement Charge distribution Detector efficiency versus bias voltage at 0 o angle: it saturates at 20 V, where the sensor is completely depleted absolute value related to the geometrical acceptance of the sensor which at 0 o is limited by the vertical electrodes (inactive volumes) Efficiency increases by rotating the DUT with respect to the beam. Further improvement expected by a better treatment of charge sharing. Preliminary test beam results Laboratory measurements No B field V BIAS = -20 V 0° angle 20° angle Readout n + electrode s x = 14.4 mm s y = 12.8 mm Ohmic p + electrodes Readout n + electrode V BIAS = -20 V 0° angle - - Noise measurement from S-curve 1,E-09 1,E-08 1,E-07 1,E-06 1,E-05 1,E-04 -60 -50 -40 -30 -20 -10 0 Itotal FE-I3 (A) Vback (V) ATLAS09 -w 4 -CM S -tem porary m etal CM S 1 CM S 2 CM S 3 Breakdown @ ~ -40 V -V [V] I[A]

3D-FBK pixel sensors with CMS readout: first tests results

  • Upload
    wiley

  • View
    32

  • Download
    3

Embed Size (px)

DESCRIPTION

3D-FBK pixel sensors with CMS readout: first tests results - PowerPoint PPT Presentation

Citation preview

Page 1: 3D-FBK pixel sensors with CMS readout:    first tests results

3D-FBK pixel sensors with CMS readout: first tests results  

M. Obertino, A. Solano, A. Vilela Pereira, E. Alagoz, J. Andresen, K. Arndt, G. Bolla, D. Bortoletto, M. Boscardin, J.M. Brom, R. Brosius, M. Bubna, J. Chramowicz, J. Cumalat, G-F. Dalla Betta,

F. Jensen, A. Krzywda, A. Kumar, S. Kwan, CM Lei, D. Menasce, L. Moroni, , J. Ngadiuba, I. Osipenkov, L. Perera, M. Povoli, A. Prosser, R. Rivera, P. Tan, S. Terzo, N. Tran, L. Uplegger, S.

Wagner

More details in “Test-beam studies of diamond sensors for SLHC” L. Uplegger

12th Pisa Meeting on Advanced Detectors

3D sensors consist of an array of columnar electrodes (radius ∼ 5µm) of both doping types which penetrate entirely in the silicon substrate, perpendicularly to the surface.

Fermilab testbeam Telescope with eight pixel planes of CMS Forward Pixel Detector. Pixel size is 100 µm x 150 µm, but enhanced resolution is derived from

charge sharing by tilting the telescope planes at 25°. Telescope resolution on the DUT (Detector Under Test) ~ 6 µm. Rotation and cooling of DUTs possible Alignment done using a software developed at the University of Milano Bicocca.

This structure decouples the electrode distance from the sensor substrate thickness. The close electrode spacing provides several advantages compared to the planar sensor design: low full depletion voltage (about -10 V), fast charge collection time, reduced charge trapping probability and therefore high radiation hardness.

Spatial resolution (20° angle)

1 PIXEL CLUSTER2 PIXEL CLUSTER

3D pixel sensors compatible with the CMS PSI46 readout chip were first fabricated at SINTEF (Oslo, Norway), and more recently at FBK (Trento, Italy) and CNM (Barcelona, Spain). Several sensors with different electrode configurations (single n-type electrode (1E), two n-type electrodes (2E), and four n-type electrodes (4E) per pixel cell), all bump-bonded with the CMS pixel PSI46 readout chip, were characterized in laboratory and tested at Fermilab with a proton beam of 120 GeV/c, before and after irradiation. New productions of 3D sensors with different characteristics are foreseen in view of the CMS tracker upgrade. Results presented in this poster concentrate on one un-irradiated FBK 3D sensor tested in April 2012.

FBK WAFER

3 1E CMS sensors with: 80 x 52 pixels of size 100 µm x 150 µm passing through columns (full 3D) 200 mm thickness slim edges (200 mm on two sides)bump bonded at Selex (Rome) and tested at Fermilab in April 2012

120 GeV/c protons

Noise vs VBIAS behaviour as expected from CV measurements

On wafer electrical test performed at FBKI-V curve measured after bump bonding

Efficiency measurement

Preliminary

Preliminary

Total collected cluster charge is ~13.3 ke-.

The plots below show that the collected charge is what expected in 200 mm thick silicon (~16 ke-) only if the track hits the central part of the pixel. Elsewhere it is reduced by the combined effect of charge sharing and readout chip threshold (~2500-3000 e-).

0° angle 20° angle

Noise measurement

Charge distribution

Detector efficiency versus bias voltage at 0o angle: it saturates at 20 V, where the sensor is completely depleted absolute value related to the geometrical acceptance of the sensor which at 0o is limited by the vertical electrodes (inactive volumes)

Efficiency increases by rotating the DUT with respect to the beam. Further improvement expected by a better treatment of charge sharing.

Preliminary test beam results

Laboratory measurementsNo B field

VBIAS = -20 V

0° angle 20° angleReadout n+ electrode

sx= 14.4 mm sy= 12.8 mmOhmic p+ electrodes

Readout n+ electrode

VBIAS = -20 V

0° angle

-

-

Noise measurement from S-curve

1,E-09

1,E-08

1,E-07

1,E-06

1,E-05

1,E-04

-60-50-40-30-20-100

I tot

al F

E-I3

(A)

Vback (V)

ATLAS09 - w4 - CMS - temporary metal

CMS 1

CMS 2

CMS 3

Breakdown @ ~ -40 V

-V [V]

I[A]