CONE HEALTH Cancer Center 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CONE HEALTH CANCER CENTER

3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

Caroline Yount, MSCONE HEALTH CANCER CENTER

Page 2: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

3D QA

The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose distributions and the methods for IMRT dose delivery demand that the dosimetry measurement techniques… be reviewed and adapted for the unique challenges posed by IMRT.

TG 120

Page 3: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

QA Challenges

• Modern technologies are utilizing rotation delivery to speed up treatment time and decrease tissue toxicity (RapidArc®, VMAT, TomoTherapy®, 

IMAT, SmartArc)

• Rotational beam delivery creates a challenge for patient specific QA 360 degrees of delivery

Page 4: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

3D QA 

Page 5: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

3D QA

Page 6: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

3D QA

Page 7: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

3D QA

Page 8: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

ArcCHECK Introduction

• Designed for Helical & Arc Delivery

• 1386 diodes in a helical geometry21cm diameter, 21cm length1cm spacing, 3.28cm equivalent depth0.64mm2 active detector area

• 4th Dimension = Time50ms update frequency

• Optional cavity plug insert with ion chamber holder PMMA construction

Page 9: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Detector Geometry

• Entrance and exit dose are measured Effectively doubling the detector density in the measurement field. 

• Central 10x10 contains approximately 230 detectors • Detectors are arranged on a HeliGrid™

Increases sampling rate and reduces detector overlap from Beams Eye View (BEV)

• Entrance and exit dose can be correlated to determine gantry angle

Page 10: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Question

“How can I interpret and trust 3D QA?”

Page 11: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Commissioning Measurements

1. Detector spacing2. Linearity3. Sensitivity4. Individual detector response variations5. Absolute dose accuracy

Then… Trend analysis

Page 12: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Detector spacing

kVCT MVCT

Page 13: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Linearity – Conventional Dose Rates

6MV500MU/min10x10100SAD

200200200300

400500

600700

800

1000

2000

R² = 1

050

010

0015

0020

0025

0030

00

0 500 1000 1500 2000 2500

Dos

e @

a c

entr

al d

iode

MU1.

001.

051.

101.

151.

201.

25

0 1000 2000 3000R

eadi

ng/M

UMU

Ratio

Page 14: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

50200200200

500

1000

2000

4000

R² = 1

010

0020

0030

0040

0050

00

0 1000 2000 3000 4000 5000

Dos

e @

a c

entr

al d

iode

MU

Linearity – High Dose Rates

6MV FFF1400MU/min10x10100SAD

1.00

1.05

1.10

1.15

1.20

1.25

0 2000 4000 6000R

eadi

ng/M

UMU

Ratio

Page 15: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Linearity – High Dose Rates

10MV FFF2400MU/min10x10100SAD

50200200200500

1000

2000

4000

R² = 1

010

0020

0030

0040

0050

0060

00

0 1000 2000 3000 4000 5000

Dos

e @

a c

entr

al d

iode

MU1.

001.

101.

201.

30

0 2000 4000 6000R

eadi

ng/M

UMU

Ratio

Page 16: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Spatial Sensitivity

Table = 0mm Table shifted laterally = 0.5mm

Page 17: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Spatial Sensitivity

0.5mm

1mm

2mm

3mm

86%88%90%92%94%96%98%

0 1 2 3 4Shift (mm)

Pass Rate vs Shift

Lateral Table Shifts

Page 18: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Spatial Sensitivity

0.5mm

1mm

2mm

3mm

86%88%90%92%94%96%98%

0 1 2 3

Shift (mm)

Pass Rate vs Shift

Longitudinal Table Shifts

Page 19: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Angular Sensitivity

0.5°

Gantry Angle Changes

86%88%90%92%94%96%98%

0 1 2 3Change in Angle (degrees)

Pass Rate vs Angle Change

Page 20: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Detector Response Variations

• Sensitivity differences between the ArcCHECK detectors 

• These differences can be measured• Individual correction factors applied to subsequent raw measurements

Page 21: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Detector Response Variations

Page 22: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Detector Response Variations

Page 23: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Calibration Validity Check

PAAP

Page 24: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Calibration Validity Check

In‐house array calibration:

Factory array calibration:

Page 25: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Calibration Validity Check

1.5%

New In‐house array calibration:

1%

Page 26: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

ArcCHECK Calibration

• Array calibration on standard linac• Dose calibration on standard linac SAD hand calc at 

3.28 cm depth (water equivalent depth)

Page 27: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Dose Accuracy

Measured 10x10 Planned 10x10

Page 28: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Clinical Example ‐ Chest

Page 29: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Phantom Plan

Page 30: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

QA Result

Measured Calculated

Comparison

Profiles

Ion Chamber = -2%

Page 31: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Clinical Example – Head & Neck

Page 32: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Phantom Plan

Page 33: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

QA Result

Measured Calculated

Comparison

Profiles

Ion Chamber = 0%

Page 34: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Clinical Example ‐ Prostate

Page 35: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Phantom Plan

Page 36: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

QA Result

Measured Calculated

Comparison

Profiles

Ion Chamber = -1.2%

Page 37: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Trend Analysis

“When did 3D QA indicate clinical issues?”

Page 38: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Trend Analysis

Patient‐specific Chamber Readings

Setup error, high gradients, modulation too high

‐13.5%‐12.0%‐10.5%‐9.0%‐7.5%‐6.0%‐4.5%‐3.0%‐1.5%0.0%1.5%3.0%4.5%6.0%7.5%9.0%10.5%12.0%13.5%

Machine dose rate low ~900 QAs

Page 39: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Trend Analysis

Patient‐specific Gamma Pass Rates

Gamma 3%/3mm <1

0%

10%

20%

30%

40%

50%

60%

70% Last 100 QAs

Page 40: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Future work

1. Better characterization with  FFF2. Higher resolution characterization for SRS3. Applying results to patient’s CT for DVH 

analysis

Page 41: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Conclusions

1. Commissioning2. Ongoing Quality Improvement

Page 42: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry

CONE HEALTHCancer Center

CONE HEALTHCancer Center

Thank you!