3 Ph Routine Maintenance Guide En

Embed Size (px)

Citation preview

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    1/17

    Day-to-Day Routine Maintenance

    of pH Meters and Sensors

       M   a   i   n   t   e   n   a   n   c   e   G

      u   i   d   e

    Measurement Hints

    Tips for Care

    Calibration

    System Check

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    2/17

    Editorial

    Dear Reader,

    The determination of pH value, conductivity and related parameters such as ion

    concentration, resistivity, and salinity, are frequent and ordinary tasks in many labs.Samples may originate from many different areas. Sample composition, i.e. solvent and

    major components, can differ considerably and cover a wide concentration range. The most

    common solvent of pH and conductivity samples is water. However, other solvents are used

    as well. In addition, user needs in the lab reach from simple, manual determination to fully

    automated analysis systems including data gathering via software and other features.

    To meet this array of requirements, a big number of methods exist and a wide variety of

    instrument solutions have been developed by meter and sensor manufacturers. Standard

    methods are in use for numerous applications. For special needs many dedicated solutions

    exist as well. However, the vast number of possibilities can make selecting the right

    instrument and/or sensor cumbersome.

    This guide provides some insights into meter and sensor characteristics and performance,

    enabling users to make better decisions and find the right instrument and electrode. Tips and

    hints for sensor maintenance and care help to exhaust their usable life and achieve reliable

    results. This wealth of information helps finding the most suitable instrument solution but as

     well measuring successfully each time.

    METTLER TOLEDO

    Disclaimer

    This guide represents selected, possible application examples. Examples have been tested with all possible care

    in our lab with the analytical instrument mentioned in the applications. The experiments were conducted and the

    resulting data evaluated based on our current state of knowledge.

    However, this guide does not absolve you from personally testing its suitability for your intended methods,instruments and purposes. As the use and t ransfer of an application example are beyond our control, we cannot

    accept responsibility.

    When chemicals and solvents are used, the general safety rules and the directions of the producer must be observed.

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    3/17

    3METTLER TOLEDO pH and Routine Maintenance

       C   o   n   t   e   n   t Content

    1 Care and Measurement Technique 4

    2 Calibration 8

    3 Perform an Easy System Check 10

    4 pH Electrode Troubleshooting 12

    5 More Information  16

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    4/17

    4METTLER TOLEDO pH and Routine Maintenance

       C   a   r   e   a   n   d

       M   e   a   s  u

       r   e   m   e   n   t    T

       e   c   h   n   i   q  u

       e 1. Care and Measurement Technique

    This section provides an overview of how to properly care for pH and conductivity sensors and some hints

    regarding measurement techniques. In addition, the advantages of Intelligent Sensor Management (ISM®) and

    the testing of ultra-pure water are explained.

    1.1 pH Electrode Maintenance

    Regular maintenance is very important for prolonging the lifetime of any pH electrode. Electrodes with liquid

    electrolyte need the electrolyte to be topped-up when the level threatens to become lower than the level of the

    sample solution. This way a reflux of the sample into the electrode is avoided. The complete reference electrolyte

    should also be changed regularly, e.g. once a month. This ensures that the electrolyte is fresh and that no

    crystallization occurs despite evaporation from the open filling port during measurement. Be careful not to get

    any bubbles on the inside of the electrode, especially near the junction. If this happens the measurements will be

    unstable. To get rid of any bubbles, gently shake the electrode in the vertical motion like with a fever thermometer.

    1.2 pH Electrode Storage

    Electrodes should always be stored in aqueous and ion-rich solutions. This ensures that the pH-sensitive gel

    layer which forms on the pH glass membrane remains hydrated and ion rich. This is necessary for the pH

    membrane to react in a reliable way with respect to the pH value of a sample.

    Short term storage

    In between measurements or when the electrode is not being used for brief periods of time, it is best to keep

    the electrode in a holder containing the special InLab® storage solution[1], its inner electrolyte solution (e.g. 3

    mol/L KCl), or in a pH 4 or pH 7 buffer. Ensure that the level of solution in the beaker is below that of the filling

    solution in the electrode.

    Long term storage

    For long term storage, keep the electrode wetting cap filled with the InLab® storage solution[1] or, alternatively,

     with the inner electrolyte solution, pH buffer 4 or 0.1 mol/L HCl. Make sure that the filling port for reference and

    combination electrodes is closed so as to avoid loss of the electrolyte solution through evaporation, which can

    cause the formation of crystals within the electrode and junction.

    Never store the electrode dry or in distilled water as this will affect the pH-sensitive glass membrane and thus

    shorten the lifetime of the electrode.

     Although an electrode that has been incorrectly stored can be restored by regeneration procedures, following the

    above mentioned recommendations will ensure that your electrode is always ready to use.

    Temperature sensors

    Rinse the temperature sensors after use and store dry in the packing box to prevent damage.

    1.3 pH Electrode Cleaning

    To clean the electrode, rinse it with deionized water after each measurement but never wipe it clean with a tissue.

    The rough surface of the paper tissue will scratch and damage the pH-sensitive glass membrane removing the

    gel-layer and creating an electrostatic charge on the electrode. This electrostatic charge causes the measured

    signal to become very unstable. Special cleaning procedures may be necessary after contamination with certain

    samples. These are described in greater detail below.

    [1]  This InLab® storage solution can be ordered from METTLER TOLEDO (30111142)

    [2] This thiourea solution can be ordered from METTLER TOLEDO (51340070)

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    5/17

    5METTLER TOLEDO pH and Routine Maintenance

       C   a   r   e   a   n   d

       M   e   a   s  u

       r   e   m   e   n   t    T

       e   c   h   n   i   q  u

       e Blockage with silver sulfide (Ag2S)If the reference electrolyte contains silver ions and the sample being measured contains sulfides, the junction will get

    contaminated with a silver sulfide precipitate. To clear the junction of this contamination, clean it with 8% thiourea

    in 0.1 mol/L HCl solution.[1] 

    Blockage with silver chloride (AgCl)

    The silver ions from the reference electrolyte can also react with samples that contain chloride ions, resulting inan AgCl precipitate. This precipitate can be removed by soaking the electrode in a concentrated ammonia solution.

    Blockage with proteins

    Junctions contaminated with proteins can often be cleaned by immersing the electrode into a pepsin/HCI

    (5% pepsin in 0.1 mol/L HCl) solution for several hours.[2]

    Other junction blockages

    If the junction is blocked with other contaminations, try cleaning the electrode in an ultrasonic bath with water

    or a 0.1 mol/L HCl solution.

    1.4 pH Electrode Regeneration and Lifetime

    Even elect rodes that have been well maintained and properly stored may start performing poorly after some

    time. In such cases it may be possible to regenerate the pH-sensitive glass membrane and restore the

    electrode to its previous level of performance using an ammonium bifluoride regeneration solution [3]. This

    regeneration solution is based on a highly diluted solution of hydrofluoric acid which etches away a very thin

    layer of the glass membrane, exposing a fresh surface area.

    When using the regeneration mixture, do not to leave the electrode in the solution for longer than 1–2 minutes

    or the whole pH-sensitive membrane will be corroded away and the electrode rendered useless.

    The expected lifetime of a correctly used and maintained pH electrode is around one to three years. Factors

    that contribute to a reduction of the lifetime of an electrode include high temperatures and measuring at

    extreme pH values.

    1.5 Measuring pH – Temperature is a Critical Component

    pH results are only correct if the sample temperature is taken into account. With

    these simple but effective rules for avoiding negative temperature effects, it’s easy to

    obtain accurate, reproducible results.

    Automatic Temperature Compensation (ATC)

     ATC works best with normal-size samples.

    • Use a sensor with integrated temperature probe and wait for a stable signal. The meter

    automatically corrects the pH signal. ATC works best in samples larger than 10 mL. 

    • Any “Pro” type InLab® sensor – InLab® Micro Pro, Science Pro, Expert Pro – has

    integrated temperature probes, eliminating worries over wrong temperature

    settings or not capturing temperature.

    • For sensors without an integrated temperature probe, using a separate temperature probe is recommended.

    Figure 1: Temperature sensor of

    an InLab® electrode

    [1] [3] This regeneration solution can be ordered from METTLER TOLEDO (51350104)

    [2] This pepsin solution can be ordered from METTLER TOLEDO (51340068)

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    6/17

    6METTLER TOLEDO pH and Routine Maintenance

       C   a   r   e   a   n   d

       M   e   a   s  u

       r   e   m   e   n   t    T

       e   c   h   n   i   q  u

       e Manual Temperature Compensation (MTC)MTC is extremely accurate, but can be time-consuming.

    • If the temperature of your sample is known (you are working in a climate-

    controlled room or the samples just came out of the refrigerator) enter this

    known temperature in the measuring settings of your instrument to correct

    the pH (or conductivity) signal.

    • When measuring samples with different temperatures, MTC can be time

    consuming, because the setting must be changed with every temperature change.

    Measure the sample, not your sensor

    With very small samples, the sensor can take so long to reach equilibrium that the sensor temperature is

     wrongly interpreted as the sample temperature. The sample mass is negligible compared with the sensor mass,

    so take the time necessary to ensure that you actually measure the sample temperature. Best practice is to keep

    the sensor with the sample. Make sure temperatures match by storing the sensor with samples in the refrigerator

    or incubator, or at room temperature. This guarantees the highest accuracy because the pH membrane, reference

    system and sample are at the same temperature.

    1.6 Contamination Control of pH Electrodes

    When measuring samples there is always the risk of contamination, either by sample carry-over or by

    microbiological or genetic contamination. Conventional pH electrodes can also be damaged by electrolyte out-

    flow when measuring TRIS-based buffers or proteinaceous samples. This is not the case when working with

    InLab® electrodes.

    Avoid sensor contamination with TRIS buffers

     Accurate pH measurement is a key factor in buffer quality. TRIS-based buf fers – widely used in biologicalresearch ranging from molecular biology to histology – can damage standard pH equipment.

    How does TRIS do its damage?

    When measuring pH during TRIS buffer preparation, the reference junction on conventional pH electrodes can clog

     when TRIS reacts with silver ions in the fill solution. This reaction can also occur with protein in the buffer, such as

    BSA (bovine serum albumin). The eventual result is slow or fluctuating readings, or even entirely wrong results.

    InLab® electrodes by METTLER

    TOLEDO are specifically designed

    for compatibility with TRIS-basedbuffers, assuring reliable results

    and accurate buffer values. The

    electrolyte in InLab® electrodes

    is guaranteed to be free of silver

    ions, eliminating the possibility of

    contamination.

    Figure 3: SevenExcellence meter and InLab® electrode during calibration

    Figure 2: Temperature and

    MTC indication on a pH meter

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    7/17

    7METTLER TOLEDO pH and Routine Maintenance

       C   a   r   e   a   n   d

       M   e   a   s  u

       r   e   m   e   n   t    T

       e   c   h   n   i   q  u

       e Clean with RNase and DNase cleansers and autoclave to eliminate biohazardThe pH electrode models InLab® Power, Power Pro, Viscous and Viscous Pro can be sterilized by autoclaving.

    By cleaning the sensors with RNase and DNase decontamination solutions first, the potential for biological

    contamination is significantly reduced.

    Figure 4: pH electrode sterilization by autoclaving

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    8/17

    8METTLER TOLEDO pH and Routine Maintenance

       C

       a   l   i   b   r   a   t   i   o   n 2. Calibration

     

     A pH electrode needs to be calibrated regularly. It is recommended that you do this at least once a day before

    you start measuring. In a calibration the slope and offset of an electrode are determined. The theoretical slope

    and offset are given by the Nernst equation:

     E = E0 + 2.3RT / nF * log [H3O+] = E0 + 2.3RT / nF * pH Slope = 2.3RT / nF 

    Offset = Should be 0 mV at pH 7.00

    The calibration is necessary to adjust the slope and offset of an electrode to their true values for the measuring

    system in question. The calibration curve is then used to correlate the measured mV values of the electrode to

    the pH value of the solution measured.

    Figure 5: Correlation between mV value measured by pH electrode and pH value in sample. Curves shown are for the theoretical behavior,

    for offset compensated behavior and slope & offset compensated behavior.

    Since an electrode is characterized by both its zero point and its slope, it is advisable to do a minimum of a

    two point calibration for reliable measurements and better precision. When measurements are performed over a

    large range of pH values it is recommended that one takes at least 3 calibration points. Most pH meters can do

    3–5 point calibrations.

    It is important to note that one should only measure samples within the chosen region of calibration.

    When calibrating an electrode, most pH meters request that you input the type of buffers which will be used.

    There are several manufacturers of buffer solutions and the specifications of the most commonly used brands

    normally already come programmed as tables in the pH meters. These tables cover groups of buffers for a

    range of temperatures. In this way a whole group can be chosen at once allowing the temperature dependence

    of the individual buffers used for calibration, to be taken into account. If no internal or external temperature

    sensor is used, ensure that you calibrate and measure at the same temperature. In this case remember to

    manually input the temperature to allow the meter to perform the buffer temperature correction.

    The buffers which are used for the calibration are very accurate solutions with a guaranteed value and precision.

    To keep the buffer solutions suitable for calibrations for as long as possible after opening it is advisable that you

    follow these guidelines:

    • Mark the date of first use on the bottle of the buffer solution.

    • Keep the buffer solution bottles tightly sealed at all times and use the decanted buffer immediately.

    • Never return used buffer back into the original bottle or mix calibration standards from different manufacturers.

    • Ensure that no contaminants enter the buffer solution bottle and always keep the bottle sealed.

    • Store the calibration standard at ambient temperature.• Do not store the bottles of buffer solution in direct sunlight.

    • Clean the electrodes before calibration and do not calibrate directly in the original buffer solution bottle.

    mV

    pH

    7

    Theoretical behaviour(Slope –59.16 mV/pH, offset: 0 mV)

    Offset correction➀

    Slope and offset correction ➀+ ➁

           ▲

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    9/17

    9METTLER TOLEDO pH and Routine Maintenance

       C

       a   l   i   b   r   a   t   i   o   n • Never use a calibration standard with an expired use by date or that you suspect is contaminated.

    • Replace the buffer solution with a new bottle after it has reached its expiry date.

     Always repeat the calibration after cleaning your electrode, after electrode maintenance, regeneration or long

    term storage of an electrode, as all these factors have an influence on the pH electrode potential.

    Figure 6: Recommended calibration points and range

    1312111098654321 7 14

    ∆pH ≈ 6

    3 or more P

    recommended e.g. pH 4, 7, 10

    ∆pH

    ≈ 1

    1P

    e.g.pH 7

    ∆pH ≈ 3

    2P

    recommended

    e.g. pH 4 and 7

    measured value withincalibration range

    measured value outside ofcalibration range

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    10/17

    10METTLER TOLEDO pH and Routine Maintenance

       P   e   r   f   o   r   m

        a   n   E   a   s  y

       S  y

       s   t   e

       m    C

       h   e   c   k 3. Perform an Easy System Check 

    Locating the problem of a pH measuring system that has suddenly started performing badly is the first step to

    restoring it to its original level of performance.

    3.1 Where could the problem lie?

    With the meter set to read mV, dip the electrode into pH 7 buffer. The reading should be 0 mV ±30 mV with an

     Ag/AgCl reference. Next read a pH 4 or pH 10 buffer – the solution should be greater than 150 mV different f rom

    the pH 7 potential. If not then test the following…

    3.2 Application

     Are you using the right electrode for your application? There are different types of

    pH electrodes for special applications: non-aqueous, low conductivity, TRIS etc. To

    make sure that you are using the right electrode visit the METTLER TOLEDO Sensor

    ProductGuide at: www.electrodes.net

    3.3 Operator

    It is sometimes worthwhile to check the obvious:

    • Is the unit properly grounded or plugged into the wall outlet?

    • Are the electrodes plugged into proper terminals and seated firmly?

    • Is the meter properly calibrated with the correct buffers?

    Before taking a measurement, check that the wetting cap has been removed and that

    the side filling aperture is open. Remember to rinse the electrodes before measuring

    a different buffer or sample.

    3.4 pH Meter

    Test the pH meter with the shor ting clip (standard delivery) or Test Plug Set. If this

    plug does not set the potential to 0 mV, the meter may be the problem. In this case

    call METTLER TOLEDO Service.

    3.5 Buffers

    Ensure that you are using the correct buffers in the correct sequence. Always use

    fresh buffers. Check expiry date.

    3.6 Cable and Connector

    Test your detachable cable by replacing it with an identical one. If you do not have

    a spare cable or are using a hard wired electrode, then check to see whether there is

    a change in the signal on the instrument when you bend the cable.

    Inspect and clean all connectors including the meter socket. If you are using an

    electrode with a MultiPin™ or S7 connector, make sure that they are free from KCl

    crystals or other deposits. Dirty or corroded connectors lead to erroneous readings.

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    11/17

    11METTLER TOLEDO pH and Routine Maintenance

       P   e   r   f   o   r   m

        a   n   E   a   s  y

       S  y

       s   t   e

       m    C

       h   e   c   k 3.7 Electrodes

     Visual inspection of the electrode can often provide important clues about the cause

    of the problem:

    Filling solution

    • Ensure that the electrolyte level is above the internal elements.

    • Empty, rinse and refill the electrode reference chamber.• Ensure that you are using the correct electrolyte as written on the electrode shaft or in the operating

    instructions, and that the electrolyte fill port is open.

    Air bubbles

    • Check for air bubbles inside the electrode. If some are present remove them by gently shaking the electrode

    downward or in the case of electrodes with gel electrolyte placing the electrode upright in warm water.

    Blocked junction

    • see next chapter 4

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    12/17

    12METTLER TOLEDO pH and Routine Maintenance

       p   H   E   l   e   c   t   r   o   d   e   T   r   o  u

       b   l   e   s   h   o   o   t   i   n   g 4. pH Electrode Troubleshooting

    The electrode is the key to successful analysis. Since the electrode is the only part of the analytical instrument

    that is in direct contact with the sample, its selection and maintenance (and therefore sensitivity) has the

    strongest influence on precise and accurate measurements. However, an electrode that has been correctly

    selected and that has been working properly may nevertheless suddenly start performing badly. In this article

     we will help you to identify possible reasons for this and suggest a number of procedures for restoring the

    electrode to its original performance.

    4.1 Before beginning the diagnostic procedure

    Before testing an electrode, make sure that the electrode cable and the

    instrument are working properly. Then examine the sensor closely. Visual

    inspection can very often provide important clues about the cause of the

    problem, e.g. a clogged diaphragm or an air bubble in the tip of the electrode.

    In general, three procedures can be followed to restore an electrode to its

    normal working state. First of all, the glass membrane can be regenerated,

    secondly, the reference diaphragm may have to be cleaned and thirdly,

    it may be necessary to replace the electrolyte. This latter point will not

    be discussed in this article because it forms part of the normal electrode

    maintenance procedure.

    4.2 Glass membrane

    a) Diagnosis

    One symptom may very often have various causes. The following table will

    help you to find out what caused your electrode to fail with regard to the pH-

    sensitive glass membrane:Figure 7: What can be fixed?

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    13/17

    13METTLER TOLEDO pH and Routine Maintenance

       p   H   E   l   e   c   t   r   o   d   e   T   r   o  u

       b   l   e   s   h   o   o   t   i   n   g

    b) Regeneration procedure A reduced calibration slope as a result of changes in the gel-layer on the glass membrane can very of ten be

    observed with older electrodes or electrodes that have been stored dry. Similar effects are noticeable when an

    electrode is used for non-aqueous applications because the gel-layer is dehydrated. The pHsensitive glass

    membrane is reactivated using a regeneration solution. This solution is a mixture of hydrochloric (HCl) and

    hydrofluoric acids (HF).

    Since these acids are extremely aggressive, make sure that you observe all the necessary safety

    precautions, e.g. wear protective goggles, a laboratory coat and chemical-resistant gloves! 

    Cause

    Symptom

    Ageing ofmembraneglass 

    Scratches

    on

    membrane

    Broken

    membrane

    or shaft

    Gel layer

    destroyed or

    dehydrated

    Dry storage

    of

    electrode

    Calcium on

    glass

    membrane

    (whitish

    film)

    Oil, fat or

    tar residues

    (visible?)

    Deposits of

    unknown

    substances

    (visible?)

    Reducedslope(>80% -

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    14/17

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    15/17

    15METTLER TOLEDO pH and Routine Maintenance

       p   H   E   l   e   c   t   r   o   d   e   T   r   o  u

       b   l   e   s   h   o   o   t   i   n   g

    Table 2: Clogged reference diaphragm: Causes and cleaning procedures

    Despite all the corrective procedures mentioned above, it is not always possible to repair a faulty electrode.

    Damage that has occurred to the wiring or casing of an electrode during use, is irreparable. In such cases the

    electrode must be replaced.

     

    Type of

    contamination

    Cleaning agent Reaction time Remarks

    Silver sulfide Thiourea 5 –60 mins. Leave until discolorationdisappears.

     All possible types ofcontamination. Firstrecommendation for

    removing unknownsubstances.

    HCl 0.1 mol/L Approx. 12 hrs. Can also be used forinternal cleaning.

     All possible types ofcontamination.Secondrecommendation forremoving unknownsubstances.

    Chromic-sulfuric acidmixture

     Approx. 30 mins. Also cleans deposits onthe membrane very well.Sensor must beregenerated after thisprocedure.

    Proteins Pepsin / HCl cleaningsolution

    > 1 hr. Can also be used forinternal cleaning.

    Proteins NaOH 2% Approx. 20 mins.

    Lipophilicsubstances

    Ethanol, acetone Approx. 30 mins. Highly suitable for edibleoils. Possibly with support

    of a soft brush.Calcium, scale Acetic acid Approx. 30 mins. Sensor must be

    regenerated after thisprocedure.

    Soaps, tensides Hot water (80° C) Approx. 12 hrs. Rinse sensor well with hot water. Afterwards, immersein hot water and leave tocool, approximately 12hours. Only use tap water,not distilled or de-ionised water.

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    16/17

    16METTLER TOLEDO pH and Routine Maintenance

       M   o   r   e   I   n   f   o   r   m   a   t   i   o   n 5. More Information

    5.1 Guides

     A Guide to pH Measurement, Mettler-Toledo AG, 51300057, (2013)

    Guide pour les mesures de pH, Mettler-Toledo AG, 51300185, (2013)

     Anleitung zur Messung von pH, Mettler-Toledo AG, 51300058, (2013)

     A Guide to Conductivity Measurement, Mett ler-Toledo AG, 30099121, (2013)

    Guide des mesures de conductivité, Mettler-Toledo AG, 30099123, (2013)

    Ein Leit faden für Leitfähigkeitsmessungen, Met tler-Toledo AG, 30099122, (2013)

    Guía para la medición de la conductividad, Mettler-Toledo AG, 30099124, (2013)

     A Practical Guide for Life Scientists – pH and Conductivity, Mettler-Toledo AG, (2014)

    Selected Water Analysis Methods, Application Brochure 37, Mettler-Toledo AG, 51725072 (2007)

    5.2 Webinars

    We provide web-based seminars (webinars) on different topics. You can participate in on-demand webinars

    at any convenient time and place.

    Live webinars offer the added benefit of allowing you to ask questions and discuss points of interest with

    METTLER TOLEDO specialists and other participants.

     www.mt.com/webinars

  • 8/17/2019 3 Ph Routine Maintenance Guide En

    17/17

    For more information

    Mettler-Toledo International Inc

    Laboratory Division

    CH-8606 Greifensee, Switzerland

    Subject to technical changes

    © 06/2015 Mettler-Toledo AG

    Global MarCom Switzerland / MC

    Learn more about Good Electrochemistry Practices program

    www.mt.com/GEP

    www.mt.com

    Good Measuring Practices

    Five Steps to Improved Measuring Results

    The five steps of all Good Measuring Practices guidelines start with

    an evaluation of the measuring needs of your processes and their

    associated risks.

    Using this information, Good Measuring Practices provide straight

    forward recommendations for selecting, installing, calibrating and

    operating laboratory equipment and devices.

    • Preservation of the accuracy and precision of results

    • Compliance with regulations, secure audits

    • Increased productivity, reduced costs

    • Professional qualification and training

    Good Electrochemistry Practice™

    Reliable pH measurements – thanks to GEP™