Upload
alynie-gomes
View
222
Download
0
Embed Size (px)
Citation preview
8/17/2019 3 Ph Routine Maintenance Guide En
1/17
Day-to-Day Routine Maintenance
of pH Meters and Sensors
M a i n t e n a n c e G
u i d e
Measurement Hints
Tips for Care
Calibration
System Check
8/17/2019 3 Ph Routine Maintenance Guide En
2/17
Editorial
Dear Reader,
The determination of pH value, conductivity and related parameters such as ion
concentration, resistivity, and salinity, are frequent and ordinary tasks in many labs.Samples may originate from many different areas. Sample composition, i.e. solvent and
major components, can differ considerably and cover a wide concentration range. The most
common solvent of pH and conductivity samples is water. However, other solvents are used
as well. In addition, user needs in the lab reach from simple, manual determination to fully
automated analysis systems including data gathering via software and other features.
To meet this array of requirements, a big number of methods exist and a wide variety of
instrument solutions have been developed by meter and sensor manufacturers. Standard
methods are in use for numerous applications. For special needs many dedicated solutions
exist as well. However, the vast number of possibilities can make selecting the right
instrument and/or sensor cumbersome.
This guide provides some insights into meter and sensor characteristics and performance,
enabling users to make better decisions and find the right instrument and electrode. Tips and
hints for sensor maintenance and care help to exhaust their usable life and achieve reliable
results. This wealth of information helps finding the most suitable instrument solution but as
well measuring successfully each time.
METTLER TOLEDO
Disclaimer
This guide represents selected, possible application examples. Examples have been tested with all possible care
in our lab with the analytical instrument mentioned in the applications. The experiments were conducted and the
resulting data evaluated based on our current state of knowledge.
However, this guide does not absolve you from personally testing its suitability for your intended methods,instruments and purposes. As the use and t ransfer of an application example are beyond our control, we cannot
accept responsibility.
When chemicals and solvents are used, the general safety rules and the directions of the producer must be observed.
8/17/2019 3 Ph Routine Maintenance Guide En
3/17
3METTLER TOLEDO pH and Routine Maintenance
C o n t e n t Content
1 Care and Measurement Technique 4
2 Calibration 8
3 Perform an Easy System Check 10
4 pH Electrode Troubleshooting 12
5 More Information 16
8/17/2019 3 Ph Routine Maintenance Guide En
4/17
4METTLER TOLEDO pH and Routine Maintenance
C a r e a n d
M e a s u
r e m e n t T
e c h n i q u
e 1. Care and Measurement Technique
This section provides an overview of how to properly care for pH and conductivity sensors and some hints
regarding measurement techniques. In addition, the advantages of Intelligent Sensor Management (ISM®) and
the testing of ultra-pure water are explained.
1.1 pH Electrode Maintenance
Regular maintenance is very important for prolonging the lifetime of any pH electrode. Electrodes with liquid
electrolyte need the electrolyte to be topped-up when the level threatens to become lower than the level of the
sample solution. This way a reflux of the sample into the electrode is avoided. The complete reference electrolyte
should also be changed regularly, e.g. once a month. This ensures that the electrolyte is fresh and that no
crystallization occurs despite evaporation from the open filling port during measurement. Be careful not to get
any bubbles on the inside of the electrode, especially near the junction. If this happens the measurements will be
unstable. To get rid of any bubbles, gently shake the electrode in the vertical motion like with a fever thermometer.
1.2 pH Electrode Storage
Electrodes should always be stored in aqueous and ion-rich solutions. This ensures that the pH-sensitive gel
layer which forms on the pH glass membrane remains hydrated and ion rich. This is necessary for the pH
membrane to react in a reliable way with respect to the pH value of a sample.
Short term storage
In between measurements or when the electrode is not being used for brief periods of time, it is best to keep
the electrode in a holder containing the special InLab® storage solution[1], its inner electrolyte solution (e.g. 3
mol/L KCl), or in a pH 4 or pH 7 buffer. Ensure that the level of solution in the beaker is below that of the filling
solution in the electrode.
Long term storage
For long term storage, keep the electrode wetting cap filled with the InLab® storage solution[1] or, alternatively,
with the inner electrolyte solution, pH buffer 4 or 0.1 mol/L HCl. Make sure that the filling port for reference and
combination electrodes is closed so as to avoid loss of the electrolyte solution through evaporation, which can
cause the formation of crystals within the electrode and junction.
Never store the electrode dry or in distilled water as this will affect the pH-sensitive glass membrane and thus
shorten the lifetime of the electrode.
Although an electrode that has been incorrectly stored can be restored by regeneration procedures, following the
above mentioned recommendations will ensure that your electrode is always ready to use.
Temperature sensors
Rinse the temperature sensors after use and store dry in the packing box to prevent damage.
1.3 pH Electrode Cleaning
To clean the electrode, rinse it with deionized water after each measurement but never wipe it clean with a tissue.
The rough surface of the paper tissue will scratch and damage the pH-sensitive glass membrane removing the
gel-layer and creating an electrostatic charge on the electrode. This electrostatic charge causes the measured
signal to become very unstable. Special cleaning procedures may be necessary after contamination with certain
samples. These are described in greater detail below.
[1] This InLab® storage solution can be ordered from METTLER TOLEDO (30111142)
[2] This thiourea solution can be ordered from METTLER TOLEDO (51340070)
8/17/2019 3 Ph Routine Maintenance Guide En
5/17
5METTLER TOLEDO pH and Routine Maintenance
C a r e a n d
M e a s u
r e m e n t T
e c h n i q u
e Blockage with silver sulfide (Ag2S)If the reference electrolyte contains silver ions and the sample being measured contains sulfides, the junction will get
contaminated with a silver sulfide precipitate. To clear the junction of this contamination, clean it with 8% thiourea
in 0.1 mol/L HCl solution.[1]
Blockage with silver chloride (AgCl)
The silver ions from the reference electrolyte can also react with samples that contain chloride ions, resulting inan AgCl precipitate. This precipitate can be removed by soaking the electrode in a concentrated ammonia solution.
Blockage with proteins
Junctions contaminated with proteins can often be cleaned by immersing the electrode into a pepsin/HCI
(5% pepsin in 0.1 mol/L HCl) solution for several hours.[2]
Other junction blockages
If the junction is blocked with other contaminations, try cleaning the electrode in an ultrasonic bath with water
or a 0.1 mol/L HCl solution.
1.4 pH Electrode Regeneration and Lifetime
Even elect rodes that have been well maintained and properly stored may start performing poorly after some
time. In such cases it may be possible to regenerate the pH-sensitive glass membrane and restore the
electrode to its previous level of performance using an ammonium bifluoride regeneration solution [3]. This
regeneration solution is based on a highly diluted solution of hydrofluoric acid which etches away a very thin
layer of the glass membrane, exposing a fresh surface area.
When using the regeneration mixture, do not to leave the electrode in the solution for longer than 1–2 minutes
or the whole pH-sensitive membrane will be corroded away and the electrode rendered useless.
The expected lifetime of a correctly used and maintained pH electrode is around one to three years. Factors
that contribute to a reduction of the lifetime of an electrode include high temperatures and measuring at
extreme pH values.
1.5 Measuring pH – Temperature is a Critical Component
pH results are only correct if the sample temperature is taken into account. With
these simple but effective rules for avoiding negative temperature effects, it’s easy to
obtain accurate, reproducible results.
Automatic Temperature Compensation (ATC)
ATC works best with normal-size samples.
• Use a sensor with integrated temperature probe and wait for a stable signal. The meter
automatically corrects the pH signal. ATC works best in samples larger than 10 mL.
• Any “Pro” type InLab® sensor – InLab® Micro Pro, Science Pro, Expert Pro – has
integrated temperature probes, eliminating worries over wrong temperature
settings or not capturing temperature.
• For sensors without an integrated temperature probe, using a separate temperature probe is recommended.
Figure 1: Temperature sensor of
an InLab® electrode
[1] [3] This regeneration solution can be ordered from METTLER TOLEDO (51350104)
[2] This pepsin solution can be ordered from METTLER TOLEDO (51340068)
8/17/2019 3 Ph Routine Maintenance Guide En
6/17
6METTLER TOLEDO pH and Routine Maintenance
C a r e a n d
M e a s u
r e m e n t T
e c h n i q u
e Manual Temperature Compensation (MTC)MTC is extremely accurate, but can be time-consuming.
• If the temperature of your sample is known (you are working in a climate-
controlled room or the samples just came out of the refrigerator) enter this
known temperature in the measuring settings of your instrument to correct
the pH (or conductivity) signal.
• When measuring samples with different temperatures, MTC can be time
consuming, because the setting must be changed with every temperature change.
Measure the sample, not your sensor
With very small samples, the sensor can take so long to reach equilibrium that the sensor temperature is
wrongly interpreted as the sample temperature. The sample mass is negligible compared with the sensor mass,
so take the time necessary to ensure that you actually measure the sample temperature. Best practice is to keep
the sensor with the sample. Make sure temperatures match by storing the sensor with samples in the refrigerator
or incubator, or at room temperature. This guarantees the highest accuracy because the pH membrane, reference
system and sample are at the same temperature.
1.6 Contamination Control of pH Electrodes
When measuring samples there is always the risk of contamination, either by sample carry-over or by
microbiological or genetic contamination. Conventional pH electrodes can also be damaged by electrolyte out-
flow when measuring TRIS-based buffers or proteinaceous samples. This is not the case when working with
InLab® electrodes.
Avoid sensor contamination with TRIS buffers
Accurate pH measurement is a key factor in buffer quality. TRIS-based buf fers – widely used in biologicalresearch ranging from molecular biology to histology – can damage standard pH equipment.
How does TRIS do its damage?
When measuring pH during TRIS buffer preparation, the reference junction on conventional pH electrodes can clog
when TRIS reacts with silver ions in the fill solution. This reaction can also occur with protein in the buffer, such as
BSA (bovine serum albumin). The eventual result is slow or fluctuating readings, or even entirely wrong results.
InLab® electrodes by METTLER
TOLEDO are specifically designed
for compatibility with TRIS-basedbuffers, assuring reliable results
and accurate buffer values. The
electrolyte in InLab® electrodes
is guaranteed to be free of silver
ions, eliminating the possibility of
contamination.
Figure 3: SevenExcellence meter and InLab® electrode during calibration
Figure 2: Temperature and
MTC indication on a pH meter
8/17/2019 3 Ph Routine Maintenance Guide En
7/17
7METTLER TOLEDO pH and Routine Maintenance
C a r e a n d
M e a s u
r e m e n t T
e c h n i q u
e Clean with RNase and DNase cleansers and autoclave to eliminate biohazardThe pH electrode models InLab® Power, Power Pro, Viscous and Viscous Pro can be sterilized by autoclaving.
By cleaning the sensors with RNase and DNase decontamination solutions first, the potential for biological
contamination is significantly reduced.
Figure 4: pH electrode sterilization by autoclaving
8/17/2019 3 Ph Routine Maintenance Guide En
8/17
8METTLER TOLEDO pH and Routine Maintenance
C
a l i b r a t i o n 2. Calibration
A pH electrode needs to be calibrated regularly. It is recommended that you do this at least once a day before
you start measuring. In a calibration the slope and offset of an electrode are determined. The theoretical slope
and offset are given by the Nernst equation:
E = E0 + 2.3RT / nF * log [H3O+] = E0 + 2.3RT / nF * pH Slope = 2.3RT / nF
Offset = Should be 0 mV at pH 7.00
The calibration is necessary to adjust the slope and offset of an electrode to their true values for the measuring
system in question. The calibration curve is then used to correlate the measured mV values of the electrode to
the pH value of the solution measured.
Figure 5: Correlation between mV value measured by pH electrode and pH value in sample. Curves shown are for the theoretical behavior,
for offset compensated behavior and slope & offset compensated behavior.
Since an electrode is characterized by both its zero point and its slope, it is advisable to do a minimum of a
two point calibration for reliable measurements and better precision. When measurements are performed over a
large range of pH values it is recommended that one takes at least 3 calibration points. Most pH meters can do
3–5 point calibrations.
It is important to note that one should only measure samples within the chosen region of calibration.
When calibrating an electrode, most pH meters request that you input the type of buffers which will be used.
There are several manufacturers of buffer solutions and the specifications of the most commonly used brands
normally already come programmed as tables in the pH meters. These tables cover groups of buffers for a
range of temperatures. In this way a whole group can be chosen at once allowing the temperature dependence
of the individual buffers used for calibration, to be taken into account. If no internal or external temperature
sensor is used, ensure that you calibrate and measure at the same temperature. In this case remember to
manually input the temperature to allow the meter to perform the buffer temperature correction.
The buffers which are used for the calibration are very accurate solutions with a guaranteed value and precision.
To keep the buffer solutions suitable for calibrations for as long as possible after opening it is advisable that you
follow these guidelines:
• Mark the date of first use on the bottle of the buffer solution.
• Keep the buffer solution bottles tightly sealed at all times and use the decanted buffer immediately.
• Never return used buffer back into the original bottle or mix calibration standards from different manufacturers.
• Ensure that no contaminants enter the buffer solution bottle and always keep the bottle sealed.
• Store the calibration standard at ambient temperature.• Do not store the bottles of buffer solution in direct sunlight.
• Clean the electrodes before calibration and do not calibrate directly in the original buffer solution bottle.
mV
pH
7
Theoretical behaviour(Slope –59.16 mV/pH, offset: 0 mV)
Offset correction➀
Slope and offset correction ➀+ ➁
▲
➀
➁
▲
▲
8/17/2019 3 Ph Routine Maintenance Guide En
9/17
9METTLER TOLEDO pH and Routine Maintenance
C
a l i b r a t i o n • Never use a calibration standard with an expired use by date or that you suspect is contaminated.
• Replace the buffer solution with a new bottle after it has reached its expiry date.
Always repeat the calibration after cleaning your electrode, after electrode maintenance, regeneration or long
term storage of an electrode, as all these factors have an influence on the pH electrode potential.
Figure 6: Recommended calibration points and range
1312111098654321 7 14
∆pH ≈ 6
3 or more P
recommended e.g. pH 4, 7, 10
∆pH
≈ 1
1P
e.g.pH 7
∆pH ≈ 3
2P
recommended
e.g. pH 4 and 7
measured value withincalibration range
measured value outside ofcalibration range
8/17/2019 3 Ph Routine Maintenance Guide En
10/17
10METTLER TOLEDO pH and Routine Maintenance
P e r f o r m
a n E a s y
S y
s t e
m C
h e c k 3. Perform an Easy System Check
Locating the problem of a pH measuring system that has suddenly started performing badly is the first step to
restoring it to its original level of performance.
3.1 Where could the problem lie?
With the meter set to read mV, dip the electrode into pH 7 buffer. The reading should be 0 mV ±30 mV with an
Ag/AgCl reference. Next read a pH 4 or pH 10 buffer – the solution should be greater than 150 mV different f rom
the pH 7 potential. If not then test the following…
3.2 Application
Are you using the right electrode for your application? There are different types of
pH electrodes for special applications: non-aqueous, low conductivity, TRIS etc. To
make sure that you are using the right electrode visit the METTLER TOLEDO Sensor
ProductGuide at: www.electrodes.net
3.3 Operator
It is sometimes worthwhile to check the obvious:
• Is the unit properly grounded or plugged into the wall outlet?
• Are the electrodes plugged into proper terminals and seated firmly?
• Is the meter properly calibrated with the correct buffers?
Before taking a measurement, check that the wetting cap has been removed and that
the side filling aperture is open. Remember to rinse the electrodes before measuring
a different buffer or sample.
3.4 pH Meter
Test the pH meter with the shor ting clip (standard delivery) or Test Plug Set. If this
plug does not set the potential to 0 mV, the meter may be the problem. In this case
call METTLER TOLEDO Service.
3.5 Buffers
Ensure that you are using the correct buffers in the correct sequence. Always use
fresh buffers. Check expiry date.
3.6 Cable and Connector
Test your detachable cable by replacing it with an identical one. If you do not have
a spare cable or are using a hard wired electrode, then check to see whether there is
a change in the signal on the instrument when you bend the cable.
Inspect and clean all connectors including the meter socket. If you are using an
electrode with a MultiPin™ or S7 connector, make sure that they are free from KCl
crystals or other deposits. Dirty or corroded connectors lead to erroneous readings.
8/17/2019 3 Ph Routine Maintenance Guide En
11/17
11METTLER TOLEDO pH and Routine Maintenance
P e r f o r m
a n E a s y
S y
s t e
m C
h e c k 3.7 Electrodes
Visual inspection of the electrode can often provide important clues about the cause
of the problem:
Filling solution
• Ensure that the electrolyte level is above the internal elements.
• Empty, rinse and refill the electrode reference chamber.• Ensure that you are using the correct electrolyte as written on the electrode shaft or in the operating
instructions, and that the electrolyte fill port is open.
Air bubbles
• Check for air bubbles inside the electrode. If some are present remove them by gently shaking the electrode
downward or in the case of electrodes with gel electrolyte placing the electrode upright in warm water.
Blocked junction
• see next chapter 4
8/17/2019 3 Ph Routine Maintenance Guide En
12/17
12METTLER TOLEDO pH and Routine Maintenance
p H E l e c t r o d e T r o u
b l e s h o o t i n g 4. pH Electrode Troubleshooting
The electrode is the key to successful analysis. Since the electrode is the only part of the analytical instrument
that is in direct contact with the sample, its selection and maintenance (and therefore sensitivity) has the
strongest influence on precise and accurate measurements. However, an electrode that has been correctly
selected and that has been working properly may nevertheless suddenly start performing badly. In this article
we will help you to identify possible reasons for this and suggest a number of procedures for restoring the
electrode to its original performance.
4.1 Before beginning the diagnostic procedure
Before testing an electrode, make sure that the electrode cable and the
instrument are working properly. Then examine the sensor closely. Visual
inspection can very often provide important clues about the cause of the
problem, e.g. a clogged diaphragm or an air bubble in the tip of the electrode.
In general, three procedures can be followed to restore an electrode to its
normal working state. First of all, the glass membrane can be regenerated,
secondly, the reference diaphragm may have to be cleaned and thirdly,
it may be necessary to replace the electrolyte. This latter point will not
be discussed in this article because it forms part of the normal electrode
maintenance procedure.
4.2 Glass membrane
a) Diagnosis
One symptom may very often have various causes. The following table will
help you to find out what caused your electrode to fail with regard to the pH-
sensitive glass membrane:Figure 7: What can be fixed?
8/17/2019 3 Ph Routine Maintenance Guide En
13/17
13METTLER TOLEDO pH and Routine Maintenance
p H E l e c t r o d e T r o u
b l e s h o o t i n g
b) Regeneration procedure A reduced calibration slope as a result of changes in the gel-layer on the glass membrane can very of ten be
observed with older electrodes or electrodes that have been stored dry. Similar effects are noticeable when an
electrode is used for non-aqueous applications because the gel-layer is dehydrated. The pHsensitive glass
membrane is reactivated using a regeneration solution. This solution is a mixture of hydrochloric (HCl) and
hydrofluoric acids (HF).
Since these acids are extremely aggressive, make sure that you observe all the necessary safety
precautions, e.g. wear protective goggles, a laboratory coat and chemical-resistant gloves!
Cause
Symptom
Ageing ofmembraneglass
Scratches
on
membrane
Broken
membrane
or shaft
Gel layer
destroyed or
dehydrated
Dry storage
of
electrode
Calcium on
glass
membrane
(whitish
film)
Oil, fat or
tar residues
(visible?)
Deposits of
unknown
substances
(visible?)
Reducedslope(>80% -
8/17/2019 3 Ph Routine Maintenance Guide En
14/17
8/17/2019 3 Ph Routine Maintenance Guide En
15/17
15METTLER TOLEDO pH and Routine Maintenance
p H E l e c t r o d e T r o u
b l e s h o o t i n g
Table 2: Clogged reference diaphragm: Causes and cleaning procedures
Despite all the corrective procedures mentioned above, it is not always possible to repair a faulty electrode.
Damage that has occurred to the wiring or casing of an electrode during use, is irreparable. In such cases the
electrode must be replaced.
Type of
contamination
Cleaning agent Reaction time Remarks
Silver sulfide Thiourea 5 –60 mins. Leave until discolorationdisappears.
All possible types ofcontamination. Firstrecommendation for
removing unknownsubstances.
HCl 0.1 mol/L Approx. 12 hrs. Can also be used forinternal cleaning.
All possible types ofcontamination.Secondrecommendation forremoving unknownsubstances.
Chromic-sulfuric acidmixture
Approx. 30 mins. Also cleans deposits onthe membrane very well.Sensor must beregenerated after thisprocedure.
Proteins Pepsin / HCl cleaningsolution
> 1 hr. Can also be used forinternal cleaning.
Proteins NaOH 2% Approx. 20 mins.
Lipophilicsubstances
Ethanol, acetone Approx. 30 mins. Highly suitable for edibleoils. Possibly with support
of a soft brush.Calcium, scale Acetic acid Approx. 30 mins. Sensor must be
regenerated after thisprocedure.
Soaps, tensides Hot water (80° C) Approx. 12 hrs. Rinse sensor well with hot water. Afterwards, immersein hot water and leave tocool, approximately 12hours. Only use tap water,not distilled or de-ionised water.
8/17/2019 3 Ph Routine Maintenance Guide En
16/17
16METTLER TOLEDO pH and Routine Maintenance
M o r e I n f o r m a t i o n 5. More Information
5.1 Guides
A Guide to pH Measurement, Mettler-Toledo AG, 51300057, (2013)
Guide pour les mesures de pH, Mettler-Toledo AG, 51300185, (2013)
Anleitung zur Messung von pH, Mettler-Toledo AG, 51300058, (2013)
A Guide to Conductivity Measurement, Mett ler-Toledo AG, 30099121, (2013)
Guide des mesures de conductivité, Mettler-Toledo AG, 30099123, (2013)
Ein Leit faden für Leitfähigkeitsmessungen, Met tler-Toledo AG, 30099122, (2013)
Guía para la medición de la conductividad, Mettler-Toledo AG, 30099124, (2013)
A Practical Guide for Life Scientists – pH and Conductivity, Mettler-Toledo AG, (2014)
Selected Water Analysis Methods, Application Brochure 37, Mettler-Toledo AG, 51725072 (2007)
5.2 Webinars
We provide web-based seminars (webinars) on different topics. You can participate in on-demand webinars
at any convenient time and place.
Live webinars offer the added benefit of allowing you to ask questions and discuss points of interest with
METTLER TOLEDO specialists and other participants.
www.mt.com/webinars
8/17/2019 3 Ph Routine Maintenance Guide En
17/17
For more information
Mettler-Toledo International Inc
Laboratory Division
CH-8606 Greifensee, Switzerland
Subject to technical changes
© 06/2015 Mettler-Toledo AG
Global MarCom Switzerland / MC
Learn more about Good Electrochemistry Practices program
www.mt.com/GEP
www.mt.com
Good Measuring Practices
Five Steps to Improved Measuring Results
The five steps of all Good Measuring Practices guidelines start with
an evaluation of the measuring needs of your processes and their
associated risks.
Using this information, Good Measuring Practices provide straight
forward recommendations for selecting, installing, calibrating and
operating laboratory equipment and devices.
• Preservation of the accuracy and precision of results
• Compliance with regulations, secure audits
• Increased productivity, reduced costs
• Professional qualification and training
Good Electrochemistry Practice™
Reliable pH measurements – thanks to GEP™