    1/22

    POP-8300 Free

    Chlorine/ClO2 /pH/TEMP Online

    Analysis Dosing & Control

    Integration System

    Instruction Manual

    V1.2

  • 8/15/2019 285_pop-8300a__manual

    2/22

    Introduction

    Thanks for choosing POP-8300 Free Chlorine/ClO2/ pH/TEMP Online

    Analysis Dosing &Control Integration System manufactured by Hebei Create

    Instrumentation Technologies Co., Ltd. Proper sensor installation and parameter setting would give maxium performance and advantage of this

    instrument for your good usage. So please carefully read this manual before

    installation.

    This instrument is a precise electrochemical analysis Dosing &Control

    Integration System, which should be operated by technicians with relevant

     professional knowledge.

    Please contact technical backup of CREATE when you meet any

     problems during installation and usage.

    Check the actual product with complete set after you receive the package,

    and contact us if any missing or damage.

    Our serious promise:

    1. The indicator’s quality guarantee is one year from the date of 

     purchasing. During this period, if the meter has quality problems,

    manufacturer is responsible for maintenance for free or replacement.

    2. We provide lifelong maintenance service for the product whatever you

     purchase from us or distributors.

    3. If the damage of the meter is caused by the following reasons, it is out

    of the maintenance service:

    A).The meter is burned caused by misconnection with high voltage power 

    supply or soggy.

    B).The meter is refitted or misused without permission.

    C).The meter is damaged under the condition out of use environment.

    D).The relevant damage caused by choosing the wrong type.

    E).The physical damage caused by ultimate load

    F).The meter is out of operation caused by improper storage and

    transportation (refer to SJ/T10463-93 standard)

    G).Consumable material is out of maintenance service.

  • 8/15/2019 285_pop-8300a__manual

    3/22

    Content

    I GENERAL INFORMATION..........................................................................  1

    I GENERAL..........................................................................................................................1II SYSTEM CHARACTERISTICS.............................................................................................1

    III APPLICATIONS................................................................................................................1

    IV TECHNICAL INDEX.........................................................................................................2

    Ⅱ.INTRODUCTION TO SYSTEM................................................................... 3

    I OUTLINE...........................................................................................................................3

    II CABINET I NSTALLATION.................................................................................................4

    III FLOW CELL PARTS......................................................................................................... 4

    IV CHECK BEFORE POWER SUPPLY.....................................................................................5

    Ⅲ. OPERATION................................................................................................ 8

    I OUTLINE DIMENSION OF CABINET....................................................................................8

    II K EY INTRODUCTION........................................................................................................8

    III SETTING AND OPERATION..............................................................................................9

    IV MAINTENANCE............................................................................................................ 13

    V. TROUBLE SHOOTING.............................................................................. 14

    VI. COMPLETE SET....................................................................................... 15

    VII. ORDER DIRECTORY.............................................................................  15

    APPENDIX I.......................................................................................................................16

    APPENDIX II..................................................................................................................... 16

  • 8/15/2019 285_pop-8300a__manual

    4/22

    1

    I General information

    i General

    POP-8300 Free Chlorine/ClO2/pH/TEMP Online analysis dosing &control

    integration system comes together intelligent control meter, Free Chlorine/ClO2

    sensor, pH sensor, Temp probe and matching constant flow device and other 

    relative accessories, with tight structure, easy transportation and installation and

    free of maintenance features.

    ii System characteristics

    1. The indicator, sample treatment and flow devices are integrated to build a good

    environment condition for parameter measurement. Complex on-site problems to be collected and eliminated much uncertain factors during application procedure.

    2. Wall mounted protection type , sealed cabinet, transparent panel design, system

    integration.

    3. Build-in filter, constant level and constant flow rate (patent technology) to

    ensure the water sample flow constant and measurement data stable.(free

    chlorine)

    4. Automatic exhausting and anti-siphon device ensure the equal circulation and

    always infiltration for all the sensors. (free chlorine)

    5. Measurement circuit embedded in the rotor flow meter indicator. Flow rate is

    visible. (free chlorine)

    iii Applications

    1)It is widely used for free chlorine, ClO2, pH, temperature measurement and

    time management of municipal water, potable water, food.

    2 ) It is widely used for swimming pool, SPA, water-fun facilities and

    circulating water systems that have to be monitored on-line.

    3) It is widely used for disinfection of impounding reservoir and pollution

    water, or other disinfection process concerned with free chlorine and

    ClO2.

  • 8/15/2019 285_pop-8300a__manual

    5/22

    2

    iv Technical index

    System Model POP-8300A POP-8300B

    Meter ModelPOP-8803A POP-8803B

    Free Chlorine/pH/Temp.   ClO2/Temp.

    Measurement

    range

    Free Chlorine (0.01~2.00)mg/L( ppm)(0.01~20.00)mg/L( ppm)

    ClO2   (0.01~2.00)mg/L( ppm)(0.01~20.00)mg/L( ppm)

     pH 2.00 ~ 12.00(Free Chlorine)

    Temp (0.0~99.9)℃(Temp. compensation component Pt1000)

    Resolution

    Free Chlorine 0.01mg/L

    ClO2   0.01mg/L

     pH 0.01

    Temp. 0.1℃

    Accuracy

    Free Chlorine 90% confidence interval±10%FS

    ClO2   90% confidence interval±10%FS

     pH 0.1

    Temp. ±0.5℃

    Medium pH 6.0~8.0(only for Free Chlorine application)

    Temp.   (0.0~40.0)℃

    (4~20)mA

    Output

    Channels Doubles channels

    characteristics  Reversible, fully adjustable, instrument/transmitter mode for 

    selectionChannels

    confi uration  Programmable for free chlorine, ClO2, Temp, pH

    Loop

    resistance  400Ω(max),DC24V

    Accuracy ±0.1mA

    Control port

    channels Triple channels

    Contact mode Photoelectric semiconductor relayLoad capacity Load current 50mA(max),AC/DC 30V

    Functions

    selection

    Programmable( free chlorine, ClO2, Temp, pH, timing)output;

    Communicati

    on ort  RS485 MODBUS RTU RS485

    Working

     power 

    Power supply

    voltage  System AC 220V /50Hz

    Power 

    consum tion  ≤5.5W

    Workingenvironment

      temperature:(0~50 )℃;humidity:≤85% RH(none condensation)

  • 8/15/2019 285_pop-8300a__manual

    6/22

    3

    Storage

    environment  Temperature:(0~60)℃; humidity:≤85%RH(none condensation)

    Protection

    level  IP65( plastic protection cabinet, integration system)

    Cabinet

    material  PC

    Cabinet size   460mm×350mm×165mm(H*W*D)

    installation Wall mounted

    Cabinetwei ht

      ≤10kg

    Pipeline flow   (250-350)   ml/min

     

    .Introduction to System

    i Outline

    POP-8300A /POP-8300B

    POP-8300A/POP-8300B

  • 8/15/2019 285_pop-8300a__manual

    7/22

    4

    ii Cabinet Installation

    Picture A.

    Procedures of installation

    A. Please open 4 pcs holes on the wall, reference picture A, recommended hole size

    Φ10mm

    B. Tighten the screws of rear cover (of cabinet)

    C. Fix the cabinet on the wall by setscrew

    D. Pull out plug of drain pipe of cabinet.

    Make sure the installation horizontally and choose the firm facade plate

    where is away from light.

    iii Flow cell parts

    A. There should be drainage pipe to

    lead water from main pipeline. Ball

    valve is needed to connect the

    drainage pipe with main pipeline.

    B. Drainage pipe should use 1/4  〞

     plastic pipe, inlet flow should be less

    than 300mL/min.

    C. Connect drainage pipe with cabinet. Firstly, take drain plug from drainage pipe

    outlet , insert drainage pipe with hose clamp into Outlet of cabinet. , then screwon.

  • 8/15/2019 285_pop-8300a__manual

    8/22

    5

    D. Drain pipe should be led to low level place, and

    avoid any knot on pipe.

    E. Inlet 1 is for inlet water measurement under normal status.

    Note Inlet 2 is for inlet water measurement under testing status.

    Inlet 1 is for inlet water measurement under normal status. The system will

    measure water from Inlet 1 during normal operation. Press Enter for 10 seconds,system will switch water sample from Inlet 1 to Inlet 2. Meanwhile, system will

    shut down all control output, only show measuring value. Inlet 2 is usually used

    for testing pre-ACF water (activated carbon filter process)

    Please make sure the working environment is in accordance with the

    technique requirement.

    Flow-chart of Flow Part

    1)   If there are so much impurities in water, PP filter is needed (as above

     picture) .

    2) Do not use the filter which can absorb Free chlorine; Seriously choose

    filter capacity since too big filter may cause lag of measurement data.

    3)   Floater should be at scale(250~350) , if isn’t , please increase pressure

    to promote flow speed.

    4)  Discharged water should be put into where lower than cabinet; Do not

    connect drainage pipe into pipe with positive or negative pressure.

    iv Check before power supply

    1) There is no need to active sensor, because the sensor keep in wet all the

    time.

    2) Connect with power supply AC 220V/50Hz.

  • 8/15/2019 285_pop-8300a__manual

    9/22

    6

    3) Adjust inlet water ball valve, so the floater can reach target standard.

    4)   This equipment can be used directly after ex-factory.(   free chlorine

     buffer solution pH   ( 6.2~6.4)when ex-factory. Please calibrate the

    sensor when testing free chlorine, please reference   MEASURMENT CALIBRATION.

    5) Please follow the below diagram to connect   (4~20)mA transmitting

    /control /RS485 communication (Remove the four screws from the

    upper plate, then carefully remove the back cover from the meter )

    Wire connection 

    24VA   Power input port, DC24V, no polarity connection

    24VB

    Power input port, DC24V, no polarity

    connection

    EARTHElectromagnetic compatibility ground

     protection terminal(connect the earth)

    SWITCH1   Semiconductor photoelectric Bilateral switch

    I outputSWITCH1’

    SWITCH2   Semiconductor photoelectric Bilateral switch

    II outputSWITCH2’

    SWITCH3Semiconductor photoelectric Bilateral switch

    SWITCH3’

  • 8/15/2019 285_pop-8300a__manual

    10/22

    7

    III output pH-IN   p pH sensor center line

     pH-ref    pH sensor shielding line

    CELL-PFree Chlorine/ ClO2   sensor shielding

    line(Green)

    CELL-V   Free Chlorine/ ClO2 control line (Red)

    CELL-G   Free Chlorine/ ClO2 measuring line (White

    TEMP-I   Red wire of Temperature sensor 

    TEMP-B   Yellow wire of temperature sensor 

    TEMP-G   Green wire of temperature sensor 

    RS485A   RS485 port.

    RS485B   RS485 port

    T1+/T1-First channel mA transmitting mode, powered by

    conditioning module

    I1+/I1-First channel mA instrument mode, powered by

    instrument

    T2+/T2-Second channel mA transmitting mode, powered by

    conditioning module

    I2+/I2-Second channel mA instrument mode, powered by

    instrument[note]:System with power supply AC 220V, then AC/DC electric transducer will

    convert it to low voltage.

  • 8/15/2019 285_pop-8300a__manual

    11/22

    8

     . Operation

    i Outline dimension of cabinet

    POP-8803A main interface POP-8803B main interface

    ii Key introduction

    Key Name Functions

    Back 

    key

    1.back to previous menu after setting and saving

     parameter 

    2.cancel the current setting

    3.switch between main monitor and check interface

    Select

    key

    1.cursor movement key

    2.left or right option key in parameter setting mode

    ADD

    key

    1.select number 0~9 circularly

    2.up and down option key in parameter setting mode

    Enter 

    key

    1.enter into password setting menu

    2.confirm parameter saved and enter into the next

    menu

  • 8/15/2019 285_pop-8300a__manual

    12/22

    9

    iii Setting and Operation

    Press in main display interface and enter into engineer mode, select

    engineering authority and input password 1000, then enter into the setting menu.

    【 NOTE】:1 Normal user ex-factory password 0000, which can be revised. Please remember 

    the new one.

    2. ESC can back to checking transmitting status and control status during running.

    3. Configuration: this can configure any value.

    4. SWITCH3 output is for switching inlet water sample (factory defaults).

    iv Measurement calibration

    Free Chlorine /ClO2/pH/TEMP is electrochemical sensor, which its sensitivity is

     persistent attenuated with time increased and measured media. It’s determined by

    the natural law of the sensor. Calibrating the probe slope to make sure the accuracy

    measurement and the calibration period is determined by media.

    Soft keyboard

    ( prompt)  Function

    Cancel the current setting and back to the previous menu

    1.scan the upward menu

    2. select the number 0~9 circularly

    1.scan the downward menu

    2.check the parameter setting when enter into the

     parameter checking interface,

    1.right-move operation

    2.to enter into the parameter reading interface after enter into the parameter checking interface.

    Save parameter and enter into the next menuOK

    Esc

  • 8/15/2019 285_pop-8300a__manual

    13/22

    10

    Item Menu name Menu functions

    1  System

    calibration

    Free chlorine/ClO2, zero calibration and knownconcentration calibration; pH calibration; calibrated

    according to menu prompt (the meter has been calibrated

     before leaving factory). More detail information, pleasecheck the part five.

    2  Measurement

     parameter 

    Sensor 

     parameter 

    The default value is Free chlorine/ClO2

    zero /slope parameter ,or set zero/slope

    value manually

    Unit setting

    Free chlorine/ClO2 unit could be selected

     ppm or mg/L,Celsius or Fahrenheit unit

    for temperature unit selection

    Measurement

    rangeselection

    0.01-2.00mg/L( ppm)or 0.01-20.00mg/L

    ( ppm)

    Filter depth  Adjust the filter depth for optimizing the

    measurement parameter.

     pH

    temperature

    compensation

    Manual temperature compensation or 

    auto temperature compensation for 

    selection(POP8803B no this function)

    Running

    compensation

    setting

    Running the compensation adjustment

    for parameter measurement error caused

     by uncertain factors of on-siteRestore the

    factory default

    Restore the missing data caused by

    disoperation

    3  Control

     parameter 

    Three channels photoelectric switch parameter setting:control output configuration for measured parameter /high

    /low limit alarm control

    4  Transmitting

     parameter 

    Two channels transmitting parameter setting, controloutput configuration for measured parameter and improve

    the measurement accuracy

    5   Communication parameter  setting of RS485 baud rate and address code. Refer toappendix1 for communication protocol.

    6  password

    setting

    Engineer user and common user for secondary permission

    setting, sensor permission downward compatibility

    7 Time setting Set and modify the clock 

    8  Backlight

    setting  Remain ON, delay off, brightness adjustment for selection

    9  Language

    selection

    Chinese and English for selection under engineer 

     permission

    10  Manufacturer 

    informationCheck manufacturer information and website, getcustomer service information

  • 8/15/2019 285_pop-8300a__manual

    14/22

    11

    【 NOTE】:Calibration buffer solution is pH (6.2~6.4) before leaving factory.

    Free

    Chlorine/ClO2calibrationinterface Calibration process steps

    SartSkip

    Zero calibration

    Put the sensor in water which

    without CHL.

    0%

    1. Please calibrate this sensor regularly due

    to keep measurement accuracy.

    2. Prepare water which without any

    chlorine: Boiled tap water, then after 

    cool, please put free chlorine sensor in

    the water.

    3. Enter normal user interface, choose

    SYSTEM CALIBRATION , then enter 

    the Calibration term, then calibrate the

    sensor by following pictures.

    Low concentration medium of free chlorine

    calibration: select the suitable buffer solution

    according to medium range to calibrate the

    sensor; and choose “Free Chlorine pH

    Compensation” and adjust pH value of buffer 

    solution.

    If high ClO2   concentration is known, just

    calibrate the sensor within the range which

    already indicated in the item

    After zero calibration of Free Chlorine and ClO2 probes, put the probes into Free Chlorine or 

    ClO2   solution of known concentration. Press

    Enter key for Free Chlorine and ClO2 calibration

    according to left diagram.

  • 8/15/2019 285_pop-8300a__manual

    15/22

    12

    This interface means the calibration is failed. We

    recommend to replace a new probe and

    re-calibrate.

    If the calibration is succeed, press ESC key and

    return to main menu.

     pH calibration interface  Calibration process steps(suitable for 

    POP8803A)

    1.Do pH slope calibration for the probe when the

     probe is used for a long time or long time no use.

    2. Select normal User and enter into SystemCalibration, press Enter key and enter into

    calibration options. Operate pH slope calibration

    according to left diagram.

    3.There are on-line and off-line modes for pH

    calibration.

    4.On-line calibration, means putting the probe

    into the prepared solution for  

    calibration(10.00/7.00/4.00 or select acid/neutral

    or alkali/ neutral two points).

    1. Select on-line mode and press Enter key to

    enter into the current temperature setting

    interface.

    2. Press digit select key and add key to enter into

    the current temperature of buffer solution. Then

     press Enter key to save.

    1. Please choose buffer solution from 10.00,7.00, 4.00 by order, or 9.18, 6.86, 4.01 by

    order.

    2. Confirm to continue if pH value is according

    with buffer solution

    3. Skip to continue if pH value is not

    according with buffer solution

    4. Press ENTER to confirm whether past or not

    after setting.

    5. If failed, please re-calibrate.

  • 8/15/2019 285_pop-8300a__manual

    16/22

    13

    According to left diagram, put the cleaned probe

    into the prepared buffer solution. Press enter to

    save and enter into the next menu.

    Check the displayed mA value stable or not.

    When the value is stable, press Enter key and

    start calibration.

    1.This interface shows the calibration is succeedand the program switch to the next buffer 

    solution calibration interface automatically.

    The calibration process is the same as above.

    2. If calibration is failed, the interface shows to

    Replace pH sensor.

    3.If succeed, there will be a small face.

    4.Enter next calibration item.

    IV Maintenance

    i Cabinet maintenance

    a. The installed pipeline should be unblocked. Check connection part, filter,

    constant flow device, inlet and outlet often to obverse the sedimentation

    generated or not. Please clean if there is sedimentation.

     b. Open blow down valve regularly to clean sedimentation.

    c. Please clean the cabinet by Neutral Cleaner, acidic and corrosive strong

    solution or abrasives are not allowed.

    ii pH probe maintenance

    a. pH probe will keep in wet environment, it can work directly after powered.

     b. There is not need to calibrate after a long time powered off, sensor will keep

    in wet environment.

    c. If probe is slicked by suspended matter ,can use 0.01mol/L HCl or NaOH

    solution to clean,then use clear water washing up.

  • 8/15/2019 285_pop-8300a__manual

    17/22

    14

    d. If the instrument still can not be calibrated in above methods,it means the

     probe is invalid, and should be replaced;

    e. When calibrate the new sensor please keep buffer solution temperature.

    f. Except free chlorine/chlorine dioxide, oxidation and other strong reducing

    agents of sensor is sensitive to overlapping;

    g. The sensor could not be used for measurement of electrolysis brine sodium

    hypochlorite. Since the generated hydrogen ion would damage the platinum

    layer and form inverse chlorines reverse current. This will harm the sensor and

    reduce sensor working life.

    iii. Replacement of probes

    a. Please use the original sensor, because different manufacturer different sensor.

     b. Please indicate the system which the sensor used in , manufacture will provided

    the replacement.

    c. Please do not lengthen or cut short the cable length!

    V. Trouble shooting

    Common fault inspection and trouble shooting:symptom Possible causes Trouble shooting

     No instrument display

    Incorrect connection of 

     power supply

    Check whether there is

    24v between 24V A/B.

    Whether press power 

    supply switch.

    Serious reading error of Free Chlorine and ClO2.

      Check the flow speed

    Adjust inlet ball valve for 

    stable flow speed or replace the sensor.

    Different Transmitting

    data

    A. Incorrect

    corresponding setting

    B. Incorrect receiving

    transfer.

    A. Reset 4-20mA setting

    B. Reset receiving data

    transfer 

    Incorrect pH value  A. Calibrate the sensor 

    B. pH sensor failed

    A. Calibrate pH slope

    B. Replace the sensor 

  • 8/15/2019 285_pop-8300a__manual

    18/22

    15

    VI. Complete Set

    Model

    name

    POP-8300A POP-8300B

    Main parts

    POP-8803A POP-8803BFree chlorine sensor    ClO2 sensor 

     pH sensor Temperature sensor 

    Temperature sensor 

    P33 flow device *3 P33 flow derive *3

    Filter device Filter device

    Stable flow speed device Stable flow speed device

    Cabinet Cabinet

    Manual Manual

    Certificate Certificate

    Other parts 1/4" plastic pipe(5 meters)Ball adjustment valve(1 pc)

    Drain pipe(1 pc)

    Hose clamp(1 pc)

    M6 setscrew(4 pcs)

    VII. Order Directory

    Please indicate the following items before ordering:

    1)  Free chlorine or ClO2 measurement should be indicated.

    2)The measurement range of Free Chlorine and ClO2

    3) Which industry the meter will be used for.

  • 8/15/2019 285_pop-8300a__manual

    19/22

    16

    Appendix I

    RS485 communication protocol.

    Please set baud rate and ADDRESS of meter to connect with host computer.

     NOTE: Please log on our website for this protocol

    Website: http://www.createc.cn

    Appendix II

    i mA transmitting wiring diagram

    Diagram:Internal and external

    connection in instrumentation

    mode

    Diagram: internal and

    external connection in

    transmitting mode

    [Note]  :  The second mA

    transmitting connection is

    same as the above diagram.

  • 8/15/2019 285_pop-8300a__manual

    20/22

    17

    ii Electrical wiring connection diagram

    With intermediate relay to expand drive high pressure or dynamic load

  • 8/15/2019 285_pop-8300a__manual

    21/22

    18

     photoelectric switch can control metering pump

    iii Control diagram

  • 8/15/2019 285_pop-8300a__manual

    22/22

    High and low limit can control pH