10
242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate- Range Scheme for Energy and Signal Transmission by Magnetic Coupling for Implantable Devices Sai Chun Tang, Senior Member, IEEE Abstract— The increasing sophistication and power demands of medical implantable devices have motivated a variety of research on wireless energy transmission, aiming to provide higher power capability, wider receiver position, and improved performance, including safety and efficiency. This paper demonstrates a low- operating-voltage wireless energy and signal transmission method to achieve these aims. We developed a 30-cm transmitting coil in a Helmholtz-coil configuration. Innovative coil segmentation and distributive resonance techniques are employed to substantially reduce the excitation voltage and the voltage over the coil. Magnetic field simulations show that the magnetic field generated by the scheme is uniform and applicable in a wide region bounded by the coils. The receiving coil has three turns and a diameter of 2 cm. The receiver can deliver 350 mW when the transmitting coil excitation is 1 A rms /6.68 V rms and the maximum voltage over the coil is 32 V rms , which is much lower than that required in the most up-to-date method (5 A/3.5 kV). This method can provide sufficient power to operate many kinds of medical implants, especially deep-seated and locomotive devices, such as capsule endoscopes. An operating frequency of 6.1 MHz is chosen. At such a high frequency, the receiving coil does not need ferromagnetic core so magnetic resonance imaging compatible implants can be accomplished. Index Terms— Coil segmentation, implantable medical devices, low-voltage operation, wireless intermediate-range scheme for energy and signal transmission (WISEST). I. I NTRODUCTION R ESEARCH on wireless transcutaneous energy and sig- nal transmission has been motivated by the growing energy requirements of the increasingly sophisticated medical implantable devices. Magnetic coupling is probably the most widely used method among various transcutaneous wireless energy transmission techniques. It has already been used in various applications such as implantable cardiac pacemakers, drug delivery systems [1], defibrillators [2], spinal cord stimu- lator [3], [4], and middle ear hearing devices [5]. The clinical need for advanced implants, for example, next-generation capsule endoscopy [6], [7] and artificial hearts [8], [9], has necessitated the demand for more power. Manuscript received December 13, 2013; revised February 18, 2014; accepted March 7, 2014. Date of publication March 28, 2014; date of current version January 29, 2015. Recommended for publication by Associate Editor H. S. H. Chung. The author is with the Radiology Department, Harvard Medical School, Brigham and Women’s Hospital, Boston, MA 02115 USA (e-mail: [email protected]). Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org. Digital Object Identifier 10.1109/JESTPE.2014.2314388 In transcutaneous magnetic coupling, the transmitting coil placed parallel to the skin surface is driven by a high- frequency current that generates magnetic field concentrated around the coil [10]–[13]. The receiving coil located under the skin converts the magnetic energy to electric energy to operate the implantable circuit. To maximize power transmis- sion, the transmitting coil has to be well-aligned and close to the receiving coil [14]. A transcutaneous transformer could transfer 12–48 W of power for an artificial heart when the sep- aration between the transmitter and receiver was 1–2 cm [8]. However, this method is only suitable for applications that the receiving coil is placed beneath and close to the skin with a fixed location. The separation between the transmitting and receiving coils is limited to 1–2 cm. In addition, the received voltage and power are very sensitive to the coil alignment error. Wireless energy and signal transmission for a deep-seated implantable device located 15 cm from the skin was proposed in [15]. A single-turn current-carrying coil wrapped around the body is used to generate an evenly distributed magnetic field deeply inside the body. A 2-cm diameter air-core receiving coil at or around the center of the transmitting coil is used to receive the magnetic energy to operate an implantable ultrasonic pulser-receiver circuit without embedding a battery. Although the magnetic field around the transmitting coil center is even, the receiving coil position is limited to about 3 cm from the plane of the transmitting coil. Wireless power and data transmission strategies that use a pair of 30-cm diameter Helmholtz coils to generate an evenly distributed magnetic field in a wider range deeply inside the body were proposed in [6] and [7]. These methods can provide sufficient energy to power up a capsule endoscope, which is locomotive and deeply inside the body, without the need of embedded batteries. While most commercial capsule endoscopes use two watch batteries, in which the stored energy is limited by their sizes, the operating time of such endoscopes is limited and the image resolution is not comparable with that provided by traditional endoscopes. The wireless energy transmission strategies were proven to transfer power over 300 mW to a locomotive capsule endoscope, in which three orthogonal coils were wrapped around a ferrite core to receive magnetic energy in all directions. This method can allow the endoscope to take higher resolution pictures as needed without the time limitation imposed by the embedded batteries. However, to transmit sufficient power to the capsule endoscope, the transmitting coil requires a driver that can 2168-6777 © 2014 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.

242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015

A Low-Operating-Voltage Wireless Intermediate-Range Scheme for Energy and Signal Transmission

by Magnetic Coupling for Implantable DevicesSai Chun Tang, Senior Member, IEEE

Abstract— The increasing sophistication and power demands ofmedical implantable devices have motivated a variety of researchon wireless energy transmission, aiming to provide higher powercapability, wider receiver position, and improved performance,including safety and efficiency. This paper demonstrates a low-operating-voltage wireless energy and signal transmission methodto achieve these aims. We developed a 30-cm transmitting coil ina Helmholtz-coil configuration. Innovative coil segmentation anddistributive resonance techniques are employed to substantiallyreduce the excitation voltage and the voltage over the coil.Magnetic field simulations show that the magnetic field generatedby the scheme is uniform and applicable in a wide regionbounded by the coils. The receiving coil has three turns anda diameter of 2 cm. The receiver can deliver 350 mW when thetransmitting coil excitation is 1 Arms/6.68 Vrms and the maximumvoltage over the coil is 32 Vrms, which is much lower thanthat required in the most up-to-date method (5 A/3.5 kV). Thismethod can provide sufficient power to operate many kinds ofmedical implants, especially deep-seated and locomotive devices,such as capsule endoscopes. An operating frequency of 6.1 MHz ischosen. At such a high frequency, the receiving coil does not needferromagnetic core so magnetic resonance imaging compatibleimplants can be accomplished.

Index Terms— Coil segmentation, implantable medical devices,low-voltage operation, wireless intermediate-range scheme forenergy and signal transmission (WISEST).

I. INTRODUCTION

RESEARCH on wireless transcutaneous energy and sig-nal transmission has been motivated by the growing

energy requirements of the increasingly sophisticated medicalimplantable devices. Magnetic coupling is probably the mostwidely used method among various transcutaneous wirelessenergy transmission techniques. It has already been used invarious applications such as implantable cardiac pacemakers,drug delivery systems [1], defibrillators [2], spinal cord stimu-lator [3], [4], and middle ear hearing devices [5]. The clinicalneed for advanced implants, for example, next-generationcapsule endoscopy [6], [7] and artificial hearts [8], [9], hasnecessitated the demand for more power.

Manuscript received December 13, 2013; revised February 18, 2014;accepted March 7, 2014. Date of publication March 28, 2014; date of currentversion January 29, 2015. Recommended for publication by Associate EditorH. S. H. Chung.

The author is with the Radiology Department, Harvard Medical School,Brigham and Women’s Hospital, Boston, MA 02115 USA (e-mail:[email protected]).

Color versions of one or more of the figures in this paper are availableonline at http://ieeexplore.ieee.org.

Digital Object Identifier 10.1109/JESTPE.2014.2314388

In transcutaneous magnetic coupling, the transmitting coilplaced parallel to the skin surface is driven by a high-frequency current that generates magnetic field concentratedaround the coil [10]–[13]. The receiving coil located underthe skin converts the magnetic energy to electric energy tooperate the implantable circuit. To maximize power transmis-sion, the transmitting coil has to be well-aligned and closeto the receiving coil [14]. A transcutaneous transformer couldtransfer 12–48 W of power for an artificial heart when the sep-aration between the transmitter and receiver was 1–2 cm [8].However, this method is only suitable for applications that thereceiving coil is placed beneath and close to the skin with afixed location. The separation between the transmitting andreceiving coils is limited to 1–2 cm. In addition, the receivedvoltage and power are very sensitive to the coil alignmenterror.

Wireless energy and signal transmission for a deep-seatedimplantable device located 15 cm from the skin was proposedin [15]. A single-turn current-carrying coil wrapped around thebody is used to generate an evenly distributed magnetic fielddeeply inside the body. A 2-cm diameter air-core receivingcoil at or around the center of the transmitting coil is usedto receive the magnetic energy to operate an implantableultrasonic pulser-receiver circuit without embedding a battery.Although the magnetic field around the transmitting coil centeris even, the receiving coil position is limited to about 3 cmfrom the plane of the transmitting coil.

Wireless power and data transmission strategies that usea pair of 30-cm diameter Helmholtz coils to generate anevenly distributed magnetic field in a wider range deeply insidethe body were proposed in [6] and [7]. These methods canprovide sufficient energy to power up a capsule endoscope,which is locomotive and deeply inside the body, withoutthe need of embedded batteries. While most commercialcapsule endoscopes use two watch batteries, in which thestored energy is limited by their sizes, the operating time ofsuch endoscopes is limited and the image resolution is notcomparable with that provided by traditional endoscopes. Thewireless energy transmission strategies were proven to transferpower over 300 mW to a locomotive capsule endoscope, inwhich three orthogonal coils were wrapped around a ferritecore to receive magnetic energy in all directions. This methodcan allow the endoscope to take higher resolution pictures asneeded without the time limitation imposed by the embeddedbatteries. However, to transmit sufficient power to the capsuleendoscope, the transmitting coil requires a driver that can

2168-6777 © 2014 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.

Page 2: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

TANG: LOW-OPERATING-VOLTAGE WIRELESS INTERMEDIATE-RANGE SCHEME 243

produce an output as high as 5 A and 3.5 kV at 1 MHz [6].Another study demonstrated by Xin et al. [7] also shows thatthe transmitting coil requires a high-voltage resonant drivingcircuit, in which the LC tank voltage is up to several kilovolts.Although the driving circuits require lower supply voltage,high voltage presents at both the driver output and also overthe transmitting coil. Clearly, electric field shielding and high-voltage insulation for both the driver and the transmitting coilare required for the patient’s safety and to comply with safetyregulations. The required shielding, electrical insulation, andhigh-voltage driver represent a large percentage of the over-all manufacturing cost. Moreover, the electric field shieldingwould attenuate the magnetic field coupling to the implant.

This paper presents a novel low-operating-voltage wirelessintermediate-range scheme for energy and signal transmission(WISEST), especially suitable for deep-seated or locomotiveimplantable devices, using magnetic coupling while traditionaltranscutaneous transformers can only transmit energy to thereceiver fixed at 1–2 cm beneath the skin. With the proposedscheme, in which a novel coil segmentation technique isapplied, the required transmitting coil excitation voltage issubstantially reduced from several hundred volts to less than7 Vrms at the input, and the maximum voltage over the trans-mitting coil is less than 32 Vrms. The voltage over the coil canbe further reduced by increasing the number of segments. Sucha reduced coil voltage represents a means for reducing theoverall manufacturing cost, and improving the device’s safetywhile increasing the power transfer capability. In addition, therequired transmitting coil keeps a low voltage even when theoperating frequency is high, therefore the operating frequencycan be increased until the magnetic energy absorbed by bodytissue is noticeable. The operating frequency used in this paperis 6.1 MHz, at which the magnetic energy absorption ofbody tissue is negligible. At such a high frequency, ferritecores that usually needed in the receiving coils of typicalimplantable devices can be eliminated. Since the receivingcoil used in the proposed method does not contain ferritecore, there is no power loss due to eddy current and hysteresisin the magnetic core. Moreover, magnetic resonance imaging(MRI) compatible implantable devices can be achieved as thereceiving coil does not contain ferromagnetic material.

It is important to note that because the magnetic field gen-erated by the Helmholtz-coil is uniform over a wide region inthe body, the proposed method does not require the additionalintermediate resonant coils used in wireless power transfersystem via strongly coupled magnetic resonance [16]–[20].However, in the applications that the intermediate resonantcoils are required, our proposed coil segmentation method canalso be applied to reduce the resonant coil voltage, which isextremely high as reported in [21].

II. SINGLE TRANSMITTING COIL

Energy and signal transmission from a single 30-cm diam-eter external transmitting coil to a deep-seated implantabledevice was demonstrated in our previous study [15]. Theenergy transfer was achieved by magnetic coupling betweenthe transmitting coil and a 2-cm diameter receiving coil in the

Fig. 1. Single transmitting coil carrying current I .

implant located around the center of the transmitting coil. Thetransmitting coil was driven by an ac current of 1.2 Arms at afrequency of 5.7 MHz. The magnetic field intensity around thecenter of the transmitting coil was uniform, thus the voltageat the receiving coil was not very sensitive to the alignmenterror between the coils within a few centimeters.

In the analysis of the magnetic field pattern induced by thetransmitting coil shown in Fig. 1, the following equations [22]describing the magnetic field intensity (H -field) in the radial,Hr , and axial, Hx , directions will be applied:

Hr = I

x

r√

(a + r)2+x2

[−K (k)+ a2 + r2 + x2

(a − r)2 + x2 E(k)

](1)

Hx = I

1√

(a + r)2 + x2

[K (k) + a2 − r2 − x2

(a − r)2 + x2E(k)

](2)

where K (k) and E(k) are the first and the second kinds ofcomplete elliptic integrals, k2 = 4ar /[(a + r)2 + x2], a is thecoil radius, x and r are the distances from the coil center alongthe coil’s axial and radial axes, respectively, and I is the coilcurrent.

Equations (1) and (2) show that the H -field induced by thecurrent flowing through a circular filament is frequency inde-pendent in free space. However, in the actual situation, the coilhas a finite thickness and wraps around the conductive humanbody. These factors will be considered in our magnetic fieldsimulations using finite-element-analysis (FEA) with AnsoftMaxwell 3-D to investigate the high-frequency effects of theconductivity and permittivity of human tissues on the magneticenergy absorption and energy transfer. At high frequencies,eddy current induced in the conductive human tissue canattenuate the magnetic field intensity. The operating frequencyof the energy transmission system in this paper is chosen as6.1 MHz (to be explained in Section VI). In the simulation,the field intensity is simulated at dc (without the eddy currenteffect), 6.1 and 10 MHz to investigate the eddy currenteffects on the H -field attenuation at different frequencies.The conductivity and relative permittivity of some tissuesin the human body at 6.1 and 10 MHz are listed in Table I.The parameters are computed on the basis of the 4-Cole-ColeModel described in [23] by using the RF_Tools MATLABprogram developed by the Center for NMR Research at PennState University [24].

Fig. 2 shows the calculated and simulated H -field of the30-cm diameter transmitting coil excited by a 1-A current.The transmitting coil with a thickness of 0.2 mm woundaround a cylinder (29.8 cm in diameter) filled with a materialapproximating the conductivity and permittivity of humantissue. According to Table I, because the small intestine has

Page 3: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

244 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015

TABLE I

CONDUCTIVITY AND RELATIVE PERMITTIVITY

OF SOME TISSUES [23], [24]

Fig. 2. Calculated and FEA simulated H -field along (a) radial, r , and(b) axial, x , directions of the single transmitting coil shown in Fig. 1.

the highest permittivity, the second highest conductivity, and arelatively large volume in the abdomen, we choose its electri-cal parameters to represent the worst-case in terms of magneticfield attenuation by the body tissue. Although the conductivityof the small intestine is a few percent less than that of the gall-bladder bile, its volume and permittivity are much larger, soits electrical parameters are adopted in the simulation model.

Fig. 3. Multiple transmitting coils that create a more even magnetic field ina wider applicable region or a desired magnetic field pattern.

In the Ansoft FEA program setting, the solution type wasset as magnetic eddy current, and the percentage error wasset to 1%. The simulation volume was bounded by a cylinderwith a height of 80 cm and a diameter of 100 cm, and thecoils were located at the middle of the cylindrical boundary.

From Fig. 2, the results calculated by (1) and (2) areconsistent with the FEA simulation when the eddy currenteffect is not included. However, as the operating frequencyincreases, the H -field attenuation increases because of theeddy current effect in the body tissue. Maximum H -fieldattenuation occurs at the center of the coil, which is also themiddle of the cylinder. The maximum H -field attenuation is1.7% at 6.1 MHz and increases to 8.8% at 10 MHz.

From Fig. 2(a), the field variation along the coil’s radialdirection is 5% within a region of 7.5-cm diameter aroundthe coil center. However, outside this flat region, the fieldintensity increases rapidly by 50% at 9.8 cm and 100%at 11.6 cm from the coil center. Along the coil’s axis, theH -field drops 5% at 3 cm, and 25% at 6.8 cm from the coilcenter as shown in Fig. 2(b).

III. MULTIPLE TRANSMITTING COILS

Although energy and signal transmission to an implantabledevice can be achieved using a single transmitting coil, theposition of the implant is limited to a few centimeters fromthe transmitting coil plane. For locomotive implants, such ascapsule endoscope and endoscopic capsule robots [25], thatneed to move and perform procedures in different positionsinside the body, a more even magnetic field pattern generatedfrom the transmitting coil is required to power up such devices.

To generate a desired or evenly distributed magnetic fieldintensity in a wider applicable region, an additional transmit-ting coil with a distance from the original coil can be adoptedas shown in Fig. 3. For instance, the Helmholtz coil, whichcomprised a pair of coils with the same diameter, separated bya coil radius, and placed in parallel and co-axially, is a well-known structure that can provide even magnetic field boundedby the coils. Other desired field patterns may be achievedwith different coil parameters, such as number of coils, coildiameters, number of turns in each coil, coil separation,and inclination. In this paper, we use a 15-cm Helmholtz-coil-configured transmitting coil (R1 = R2 = 15 cm,

Page 4: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

TANG: LOW-OPERATING-VOLTAGE WIRELESS INTERMEDIATE-RANGE SCHEME 245

Fig. 4. Calculated and FEA simulated H -field along (a) radial, r , and(b) axial, x , directions of the transmitting Helmholtz coil shown in Fig. 3when R = d1 + d2 = 15 cm.

d1 = d2 = 7.5 cm). When I = 1 A, the calculated H -fieldin free space is shown in Fig. 4. Similar to the single coilsimulation, the H -field of the Helmholtz coil was simulatedwhen the coil was wrapped around the conductive cylinderapproximating the electrical properties of the small intestineat dc (without the eddy current effect), 6.1 and 10 MHz. Thecalculated results are also included in Fig. 4 for comparison.The calculated H -field is consistent with the simulated onewhen the eddy current effect is not included in the conductivetissue. When the operating frequency increases, the H -fieldattenuation increases due to the eddy current effect. TheH -field attenuation at the midpoint between the centers of thecoils is 3.2% at 6.1 MHz and increases to 11.5% at 10 MHz.

The two-coil configuration generates a more even H -fieldpattern in a wider applicable region than that generated by thesingle coil configuration. From Fig. 4(a), the variation of theH -field at 6.1 MHz is about 2.7% within a region of 15 cmdiameter at the middle plane between the two coils. FromFig. 4(b), the H -field variation is 5.8% in the region boundedby the two coils along the coil axis. The simulated H -fieldon the r -x plane at 6.1 MHz is shown in Fig. 5. It showsthat the field is even and the applicable area is wide enoughfor many fixed or locomotive medical implant applications.

IV. METHODS FOR REDUCING THE

TRANSMITTING COIL VOLTAGE

In the design of a practical energy transmitter for med-ical implantable devices, the patient’s safety, overall cost,and power consumption are the prime factors to consider.

Fig. 5. Simulated H -field induced by the Helmholtz coil current on ther-x plane using FEA. The coil positions and current directions are indicatedby the circles with dot and cross signs.

The most up-to-date method for driving a transmitting coilfor providing sufficient power of over 300 mW for next-generation capsule endoscopes requires a several kilovoltspower converter [6], [7]. This high voltage not only presentsat the output of the power converter, but also presents overthe transmitting coil, which is designed to wrap around tothe patient body. This high-voltage requirement could be anobstacle for the development of such an energy transmissiontechnology being adopted in patients owing to safety and costissues. Although electric shielding and high-voltage insulationmay be used to reduce the risk caused by the high-voltagecoils, the overall cost will be significantly increased and themagnetic energy coupling will be deteriorated.

In our prototype design, the transmitting coil diameter is30 cm and each coil has three turns. The operating frequencyis 6.1 MHz. The separation between the two coils is 15 cm.To obtain a power level of 350 mW from a three-turn2-cm diameter air-core receiving coil, the required transmittingcoil current is 1 Arms, or 6 A-turn of magneto-motive force(mmf). The measured inductance of the transmitting coil is20 μH. Neglecting the wire resistance of a few ohms, thetransmitting coil impedance at 6.1 MHz is 2π f L = 767 �,thus the required driving voltage is 767 Vrms (or 2.17 kVpp).The driving voltage is proportional to the operating frequency,transmitting coil current, and inductance, which is determinedby the coil size and the number of turns. Therefore, ifthe transmitting coil has a larger dimensions or number ofturns, operates at a higher frequency, or carries a highercurrent, the required driving voltage would be increasedproportionally. In this paper, we propose to significantlyreduce both the required transmitting coil excitation voltageand the voltage over the whole coil by using a distribu-tive resonance technique and dividing the coil into multiplesegments.

Page 5: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

246 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015

Fig. 6. Coil divided into four smaller subcoils.

A. Dividing a Coil into Multiple Smaller Subcoils

The required driving voltage can be reduced by dividing thetransmitting coil into multiple smaller subcoils. Fig. 6 showsan example of a coil divided into four identical subcoils. Sincethe currents flowing through each pair of wires that connectingthe subcoil circumferences to the junctions are in oppo-site direction, the magnetic field generated by each current-carrying wire pair can be canceled each other theoretically.Thus, the voltage across these connecting wire pairs, which ismainly caused by the wire resistance rather than its inductance,is negligible compared with the coil segment voltage, Vseg. Therequired voltage for driving each subcoil approximately equalsto the required voltage for driving the whole coil divided bynumber of subcoils. The driving voltage for the transmittingcoil can be further reduced by connecting a capacitor inseries to each subcoil and operating the coil-capacitor circuitat its resonant frequency, at which the input impedance andthus the required driving voltage are minimized. The appliedvoltage is proportional to the resonant circuit’s equivalentseries resistance (ESR), which is very small compared with thecoil impedance without using the resonance and segmentationtechniques, and the coil current.

Although dividing the coil into multiple subcoils can sig-nificantly reduce the driving voltage and is feasible for energyand signal transmission, it increases the conduction loss causedby the additional wires for connections. In addition, thetolerances in the dimensions of the subcoils will result indifferent subcoil inductance values, so each of the subcoil hasto be fine tuned with different series capacitors individually.However, although all the subcoils can be fine tuned to thedesired resonant frequency, the tolerance of the capacitor ESR,which is hard to control and usually not specified in thedatasheet, will cause different subcoil currents, and thereforeunbalanced field pattern will occur. A representative exampleof this unbalanced situation is demonstrated by the simulatedmagnetic field pattern on the coil plane shown in Fig. 7, inwhich the subcoil currents are 0.9, 1, and 1.1 A, respectively.As shown in the figure, the magnetic field intensity at a point

Fig. 7. Simulated H -field induced by four smaller unbalanced subcoils.

Fig. 8. Coil divided into four segments by four series capacitors.

5 cm from the coil center in the 1.1-A subcoil is higher thanthat at the corresponding point in the 0.9-A subcoil by 69%.Apart from the unbalanced field pattern, the simulated fieldpattern also shows that this configuration generates undesiredmagnetic field around the connection wire pairs since themagnetic field generated by the unequal opposite currents inthe wire pairs cannot be entirely canceled each other. Thisundesired field pattern may create stability and safety issuesin the wireless energy transfer system.

B. Dividing a Coil by Multiple Series Capacitors

To avoid the disadvantages caused by the method mentionedin Section I, we propose a novel method to divide the trans-mitting coil into multiple segments by connecting low-losshigh-frequency capacitors in series to the coil segments. Fig. 8shows an example of this configuration with four segmentsin a single-turn coil. Compared with the multiple subcoil

Page 6: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

TANG: LOW-OPERATING-VOLTAGE WIRELESS INTERMEDIATE-RANGE SCHEME 247

configuration, the benefits of the series capacitor segmentationmethod are twofold. First, it does not need the additionalbundles of wires for the connections, so its conduction lossis lower and it is easier to design and use, especially when thecoil is designed to wrap around the patient’s body. Second,because all the coil segments and capacitors are connectedin series, even there are tolerances in the arc shape, the seriescapacitance value, and the capacitor ESR, the currents flowingthrough all the coil segments are the same, so unbalancedmagnetic field can be avoid.

The proposed method is different from the traditionaldesigns [6], [7], in which the transmitting coil is driven bya resonant converter or a series capacitor is connected to thecoil to form an LC resonant tank. In these traditional designs,although the system input voltage, that is, the dc supplyvoltage of the resonant converter or the driving voltage of theLC tank, can be very low, the excitation voltage and the voltageover the transmitting coil is still very high (several kilovolts).In the proposed capacitor-segmented coil, the segment voltage,which represents the maximum voltage over the whole coil,equals to the required voltage for the unsegmented coil dividedby the number of segments. The coil segment voltage canbe reduced proportionally by increasing the number of seriescapacitors and segments. When operating the circuit at theresonant frequency, if the series resistance is ignored, thevoltage across each capacitor equals to the coil segmentvoltage, Vseg, but in opposite polarity, as shown in Fig. 8. Theinput impedance equals to the ESR of the resonant circuit,which is much smaller than the impedance of the coil withoutthe series capacitors. Therefore, for a given transmitting coilcurrent, the proposed segmentation and resonance techniquescan significantly reduce both the required driving voltage andthe voltage over the whole coil.

The transmitting coil in our design consists of two 3-turncoils connected in series. The coil diameter is 30 cm andthe coil separation is 15 cm. The inductance of the transmittingcoil is measured with an HP4194A Impedance Gain PhaseAnalyzer (Hewlett Packard, Santa Clara, CA) at 6 MHz. In ourpower transmission experiments, we divided the coil into24 segments, that is, each turn is divided into four segmentsas shown in Fig. 9. With this configuration, the segmentvoltage (32 Vrms) is 24 times less than the voltage (767 Vrms)of the whole coil without using the segmentation technique.The required driving voltage is also significantly reduced byoperating the coil at the resonant frequency. The operatingfrequency for the energy transmission is selected at around6 MHz in our experiments [15]. The resonant frequency ofthe transmission network is given by

f = 1

2π√

LTXCTX(3)

where LTX is the transmitting coil inductance and CTX isthe required resultant capacitance of the series capacitorsconnecting to the coil segments.

The closest capacitance value available that achieves aresonant frequency at around 6 MHz is 820 pF. The resultantseries capacitance is 820 pF/24 = 34.17 pF and the calculatedresonant frequency is 6.088 MHz. The impedance of the

Fig. 9. Transmitting coil divided into 24 segments by 24 series capacitors.

Fig. 10. Measured and calculated impedance of the transmitting coil withdistributed series capacitors.

transmitting coil network versus frequency was measured,calculated and is shown in Fig. 10. The calculated impedanceis consistent with the measurement. The impedance is about5 � at the resonant frequency. It is noted that the resonantfrequency can be fine tuned to a desired value by connectingcapacitors in parallel or series to any of the series capacitorsthough it may be more practical to put a parallel capacitor toone close to the coil input. Since the input impedance aroundthe resonant frequency is also low, low-driving-voltage can beachieved by operating the transmitting coil around the resonantfrequency.

V. ENERGY AND SIGNAL-RECEIVING CIRCUITS

A. Energy-Receiving Circuit

An air-core receiving coil with a diameter of 2 cm is used toconvert the magnetic energy generated by the transmitting coilcurrent to electrical energy to operate the potential implantabledevices. The advantages of using an air-core receiving coilincludes low cost, very high-power density, no frequencylimitation due to magnetic cores, no magnetic loss, and easeof manufacturing [26], [27]. The receiving coil has threeturns and is made of 26 AWG single-strand enameled wire.A resonance technique is used in the receiving circuit to

Page 7: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

248 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015

Fig. 11. Schematic of the power- and signal-receiving circuit.

increase the energy receiving capability. Here, the three-turn receiving coil is used to demonstrate the proposed coilsegmentation method and it has not been optimized. Itsdimensions and number of turns should be tailored to fitthe actual situation. In addition, more advanced tuning andcontrol can be used to optimize the stability and maximizethe power-receiving capability. A schematic of the power-receiving circuit is shown in Fig. 11. The functions of thecapacitors C1 and C2 are twofold. First, they determine theresonant frequency of the receiving circuit which is given by

fr = 1

2π√

Lr Cr(4)

where Lr is the inductance of the receiving coil andCr = (C1 ·C2)/(C1 +C2). Second, because they are connectedin a potential divider configuration, the desired output voltagelevel can be chosen by an appropriate ratio of C1 and C2. Thevoltage across C2 is given by

Vo = C1

C1 + C2Vr (5)

where Vr is voltage across the receiving coil. High-frequencylow-loss capacitors are used in the resonant network. In thispaper, we select C1 = 2.18 nF and C2 = 7.5 nF to achieve aresonant frequency of 6.1 MHz and a voltage of about 6.6 Vdc at the output of the rectifier connected in the Greinachervoltage doubler configuration. Schottky diodes are used in therectifier to reduce the power dissipation due to the forward-voltage drop. It is noted that a half-wave rectifier, which ismade of only one diode, cannot be used in this configurationsince it loads the resonant capacitors with unbalanced current,that is, conducts in the positive half-cycle only. The net currentflow of the resonant capacitors must be zero in each cycle tomake the circuit operates properly. Thus, the presented rectifieror a full-wave rectifier, which conducts in both the positive andnegative cycles, should be used.

B. Power Measurement

The output power delivered from the receiving circuitwas measured when the transmitting coil was driven bya current of 1 Arms using a radio frequency (RF) poweramplifier (240L, Electronics & Innovation, Rochester,NY). The circuit and mechanical parameters in the testingsetup are summarized in Table II. The operating frequency

TABLE II

PARAMETERS OF THE TESTING SETUP

Fig. 12. Measured transmitting coil voltage and current.

Fig. 13. Measured receiving circuit output power versus load resistancewhen the transmitting coil current is 1 A and the receiving coil is located atthe midpoint between the two transmitting coils’ centers and 1 cm from thecenter of one of the transmitting coils.

was 6.1 MHz and the measured transmitting coil voltagewas 6.68 Vrms. The transmitting coil voltage and currentwaveforms were measured and are shown in Fig. 12. Thewaveforms were captured with an oscilloscope (Tektronix,DPO 3034, Beaverton, OR) and the current was sensedby a current probe (Tektronix, TCP 312) with an amplifier(Tektronix, TCPA 300). The dc output power was measuredwith different load resistance and receiving coil locations.

Page 8: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

TANG: LOW-OPERATING-VOLTAGE WIRELESS INTERMEDIATE-RANGE SCHEME 249

Fig. 14. Measured receiving circuit output power versus transmitting coilcurrent when the receiving coil is located at the midpoint between the centersof the two transmitting coils.

Fig. 15. Measured modulated transmitting coil current and demodulatedsignal at the receiver output.

Fig. 13 shows the measured output power when the receivingcoil is placed at the midpoint between the centers of thetwo transmitting coils and 1 cm axially from the center of atransmitting coil, respectively. From the measured results, theenergy level at the midpoint is higher than that at 1 cm axiallyfrom the center of a transmitting coil. This variation can bepredicted by the theoretical magnetic field plot in Fig. 4(b).The maximum output power delivered from the receivingcircuit is about 350 mW. The output power level can beincreased by increasing the transmitting coil current. Fig. 14shows the maximum output power from the receiving circuitversus transmitting coil current. The output power is about3 W when the transmitting coil current is increased to 3.1 A.

C. Signal Transmission

Signal transmission using the proposed transmitting coiland air-core receiving coil is demonstrated with on-off keying(OOK). The signal to be transmitted modulates the 6.1 MHzcarrier frequency as shown in Fig. 15. The frequency ofthe modulation signal in this demonstration is 10 kHz. Thedemodulation circuit shown in Fig. 11 is composed of adiode-resistor-capacitor envelope detector and a buffer. Therecovered waveform from the output of the demodulator isshown in Fig. 15. In this paper, we demonstrate the signal

transmission using OOK, but other modulation techniques,such as frequency modulation, can be adopted depending uponthe needs of the application.

VI. DISCUSSION

A. Transmitting Coil Configurations and Positions

This paper demonstrates a novel low-operating-voltagetechnique for transmitting energy and signal to medicalimplantable devices deep-seated in the body by magneticcoupling. The primary advantages of the proposed methodare to significantly reduce the transmitting coil excitationvoltage to less than 7 Vrms, and the voltage over the coilto less than 32 Vrms while the voltage over the coil usingexisting methods require several kilovolts [6], [7], and there-fore improve safety, reduce overall manufacturing costs, andpower consumption. Since the voltage over the whole coiland the potential difference between windings are substantiallyreduced, the dielectric loss caused by the displacement currentflowing between windings through the coil former materials isnaturally eliminated. Besides, since the magnetic field patterngenerated by the transmitting coils is uniform and applicablein a wide region, this method can deliver stable power to deep-seated implantable or even locomotive devices.

In this paper, we used a Helmholtz-coil configured trans-mitting coil to demonstrate the low-operating-voltage coilsegmentation method but the coil configuration has not beenoptimized for power transfer. First, in the testing setup, thetransmitting coil was tightly wound, the separation betweenwindings can be increased to reduce the ac resistance causedby proximity effect. Second, the coils’ geometry and numberof turns can be varied to obtain the required magnetic fieldpattern and fit the patient’s body. For example, a more evenfield pattern can be obtained by adding one more coil with alarger diameter at the middle between the two coils, which isknown as the Maxwell coil [28]. Third, to transmit maximumpower to the implant in any orientation, the transmitting coilscan be in the following configurations: 1) both coils wrappingaround the patient’s body; 2) one in front and one on the backof the patient; 3) one on the left and one on the right of thepatient; or 4) any combination of 1) to 3).

The wireless energy transmission scheme can be used topower up locomotive devices, such as capsule endoscopes.In these applications, a navigation system can be used to tracethe implant position so that the position of the transmittingcoil can be adjusted accordingly and the implantable devicecan be always around the midpoint between the centers of thetransmitting coils to achieve maximum power transmission.

B. Transmitting Coil Excitation

The desired resonant frequency of the transmission net-work can be chosen flexibly by using appropriate capacitorsin series to the transmitting coil segments. It was reportedthat [26] the energy transmission capability between air-corecoils increases as frequency increases. On the other hand, theH -field attenuation by body tissues increases as the operatingfrequency increases. The FEA simulated results in Fig. 4 showthat the attenuation increases from 3.2% to 11.5% when the

Page 9: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

250 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015

magnetic field frequency increases from 6.1 MHz to 10 MHz.In our demonstration, we select an operating frequency of6.1 MHz, but it can be increased and optimized to maximizethe energy transmission capability without causing excessivetissue energy absorption.

In our power transmission measurement, an mmf of6 A-turns is required to deliver 350 mW of power at thereceiver output with a three-turn and 2-cm receiving coil.The output power can be increased by increasing the mmf,that is, by increasing the current or number of turns ofthe transmitting coil, or by optimizing the dimensions andnumber of turns of the receiving coil. In this paper, both thetransmitting and receiving coils are made of 26 AWG single-strand enameled wires. The energy efficiency can be improvedusing thicker wires or using bundles of wires to reduce thewinding resistance. The efficiency can also be increased usinga larger receiving coil when space is allowed.

The required maximum transmitting coil voltage in theproposed method is noticeably lower than that in the mostrecent studies [6], [7]. This means that the proposed methodhas a significantly lower specific absorption rate (SAR) inbody tissue due to the electric field (E-field) emitted fromthe transmitting coils. If necessary, the E-field can be furtherreduced by increasing the number of segments and capacitorsbecause the voltage division ratio is proportional to numberof segments. Although using identical series capacitors evenlydistributing around the transmitting coil are not necessarily,this configuration can minimize the voltage, and thus theE-field, at any point over the coil.

C. Energy-Receiving Circuit

Sophisticated and multifunctional implantable devices oftenrequire power supplies with multiple output voltage levels.Multiple linear voltage regulators may be used to deliverdifferent voltage levels, but this configuration is not energyefficient, especially when the voltage difference between theregulator’s input and output is large. Energy-saving passivepotential dividers constructed with low-loss high-frequencycapacitors, C1 and C2, as shown in Fig. 11, can be used to stepdown the receiving coil voltage to voltage levels just higherthan, for example, 0.5 V, the desired voltage. The desiredvoltage levels can be obtained by connecting low-drop-out(LDO) voltage regulators at the outputs of the rectifiers. If theimplant requires a voltage level higher than the receiving coilvoltage, a multistage low-loss diode-capacitor-based voltagemultiplier proposed in [15] can be used to step up the coilvoltage to the desired voltage level.

It is expected that the proposed method can provide suf-ficient power to many kinds of implants directly withoutembedded batteries. However, if the implanted device requiresa higher current in a certain time interval, a temporary energystorage device, such as a supercapacitor or a miniaturizedbattery, can be used to store the energy and then discharge tothe load at the desired current level. Because the H -field gen-erated by the transmitting coil is even and applicable in a wideregion, this method allows multiple implantable devices atdifferent locations operating at the same time. The dimensions

and number of turns of the 2-cm diameter air-core receivingcoil used in our demonstration could be further optimized.The receiver output power depends on the coil dimensionsand number of turns. Multistrand enameled wires (litz wires)can be used to reduce the ac resistance, and thus powerloss, due to the skin effect at high frequency. The receivingcoil can also be in other shapes to fit into the implanteddevice. Apart from capsule endoscopies, it is envisioned thatthe proposed method can deliver sufficient energy to operateother implantable devices deep in the body without embeddedbattery, such as implantable blood flow monitoring devices[29], [30], drug delivery systems [1], spinal cord stimulators[3], [4], middle ear hearing devices [5], and neuromuscularstimulator [31]. In these applications, the proposed methodcan transfer energy deep in the body and precise alignment isnot required. Besides, it can further reduce the manufacturingcost by eliminating the need of the implantable medical gradebattery and the cost of the replacement surgeries every severalyears due to the limited life of the battery.

VII. CONCLUSION

A low-operating-voltage WISEST for medical implantabledevices is proposed. The operating principle of the coil seg-mentation and distributive resonance techniques for substan-tially reducing both the required transmitting coil excitationvoltage and the voltage over the coil were demonstrated withexperiments. With the proposed coil segmentation method, theexcitation voltage for the transmitting coil is less than 7 Vrmsand the voltage over the coil is less than 32 Vrms, while themost up-to-date method requires a several kilovolts powerconverter to drive an unsegmented transmitting. We demon-strated that the proposed configuration can deliver an outputpower of 350 mW from a receiving circuit composed ofa 2-cm diameter receiving coil, a simple resonant network,and a rectifier. The proposed method can significantly reducethe overall manufacturing cost, relieve safety concerns, andincrease the power transmission capability. Since the receivingcoil does not contain ferromagnetic material, MRI-compatibleimplants can be achieved. The proposed scheme generatesevenly distributed magnetic field in a wide region boundedby the transmitting coils; it is especially suitable for deep-seated and locomotive implantable devices, such as capsuleendoscopes and endoscopic capsule robots, without using theadditional intermediate resonant coils that required in wirelesspower transfer system via strongly coupled magnetic reso-nance. The proposed low-operating-voltage coil segmentationmethod can also be used to reduce the extremely high voltagein the resonant coils in the applications that the coils arerequired.

REFERENCES

[1] M. T. Salam, M. Mirzaei, M. S. Ly, D. K. Nguyen, and M. Sawan,“An implantable closedloop asynchronous drug delivery system for thetreatment of refractory epilepsy,” IEEE Trans. Neural Syst. Rehabil.Eng., vol. 20, no. 4, pp. 432–442, Jul. 2012.

[2] W. Greatbatch and C. F. Holmes, “History of implantable devices,” IEEEEng. Med. Biol. Mag., vol. 10, no. 3, pp. 38–41, Sep. 1991.

Page 10: 242 IEEE JOURNAL OF EMERGING AND SELECTED ......242 IEEE JOURNAL OF EMERGING AND SELECTED TOPICS IN POWER ELECTRONICS, VOL. 3, NO. 1, MARCH 2015 A Low-Operating-Voltage Wireless Intermediate-Range

TANG: LOW-OPERATING-VOLTAGE WIRELESS INTERMEDIATE-RANGE SCHEME 251

[3] C. H. Hsu, S. B. Tseng, Y. J. Hsieh, and C. C. Wang, “One-time-implantable spinal cord stimulation system prototype,” IEEE Trans.Biomed. Circuits Syst., vol. 5, no. 5, pp. 490–498, Oct. 2011.

[4] S. Kunnumpurath, R. Srinivasagopalan, and N. Vadivelu, “Spinal cordstimulation: Principles of past, present and future practice: A review,”J. Clin. Monitor. Comput., vol. 23, no. 5, pp. 333–339, 2009.

[5] I. Y. Park, H. G. Lim, Y. H. Yoon, M. K. Kim, B. S. Song, andJ. H. Cho, “A transcutaneous recharging system with the function of bi-directional signal transmission for fully-implantable middle ear hearingdevices,” IEICE Trans. Fundam. Electron. Comput. Sci., vol. 89, no. 6,pp. 1962–1694, Jun. 2006.

[6] R. Puers, R. Carta, and J. Thoné, “Wireless power and data transmis-sion strategies for next-generation capsule endoscopes,” J. Micromech.Microeng., vol. 21, no. 5, p. 054008, 2011.

[7] W. Xin, G. Yan, and W. Wang, “Study of a wireless power transmissionsystem for an active capsule endoscope,” Int. J. Med. Robot. Comput.Assist. Surgery, vol. 6, no. 1, pp. 113–122, Mar. 2010.

[8] G. B. Joung and B. H. Cho, “An energy transmission system foran artificial heart using leakage inductance compensation of transcu-taneous transformer,” IEEE Trans. Power Electron., vol. 13, no. 6,pp. 1013–1022, Nov. 1998.

[9] Q. H. Chen, S. C. Wong, C. K. Tse, and X. B. Ruan, “Analysis, design,and control of a transcutaneous power regulator for artificial hearts,”IEEE Trans. Biomed. Circuits Syst., vol. 3, no. 1, pp. 23–31, Feb. 2009.

[10] C. M. Zierhofer and E. S. Hochmair, “Geometric approach for couplingenhancement of magnetically coupled coils,” IEEE Trans. Biomed. Eng.,vol. 43, no. 7, pp. 708–714, Jul. 1996.

[11] H. Miura, S. Arai, Y. Kakubari, F. Sato, H. Matsuki, and T. Sato,“Improvement of the transcutaneous energy transmission system utiliz-ing ferrite cored coils for artificial hearts,” IEEE Trans. Magn., vol. 42,no. 10, pp. 3578–3580, Oct. 2006.

[12] A. M. Pernia, I. C. Orille, J. A. Martinez, J. Martin-Ramos, J. A. Canal,and F. Zacouto, “Transcutaneous microvalve activation system usinga coreless transformer,” Sens. Actuators A, Phys., vol. 136, no. 1,pp. 313–320, May 2007.

[13] E. S. Hochmair, “System optimization for improved accuracy in tran-scutaneous signal and power transmission,” IEEE Trans. Biomed. Eng.,vol. 31, no. 2, pp. 177–186, Feb. 1984.

[14] M. Soma, D. C. Galbraith, and R. L. White, “Radio-frequency coilsin implantable devices: Misalignment analysis and design procedure,”IEEE Trans. Biomed. Eng., vol. 34, no. 4, pp. 276–282, Apr. 1987.

[15] S. C. Tang, F. A. Jolesz, and G. T. Clement, “A wireless batterylessdeep-seated implantable ultrasonic pulser-receiver powered by magneticcoupling,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 58,no. 6, pp. 1211–1221, Jun. 2011.

[16] A. Kurs, A. Karalis, R. Moffatt, J. D. Joannopoulos, P. Fisher, andM. Soljacic, “Wireless power transfer via strongly coupled magneticresonances,” Sci. Exp., vol. 317, no. 5834, pp. 83–86, Jun. 2007.

[17] C. K. Lee, W. X. Zhong, and S. Y. R. Hui, “Effects of magnetic couplingof nonadjacent resonators on wireless power domino-resonator systems,”IEEE Trans. Power Electron., vol. 27, no. 4, pp. 1905–1916, Apr. 2012.

[18] B. L. Cannon, J. Hoburg, D. D. Stancil, and S. C. Goldstein, “Magneticresonant coupling as a potential means for wireless power transfer tomultiple small receivers,” IEEE Trans. Power Electron., vol. 24, no. 7,pp. 1819–1825, Jul. 2009.

[19] B. H. Waters, A. P. Sample, P. Bonde, and J. R. Smith, “Poweringa ventricular assist device (VAD) with the free-range resonant electri-cal energy delivery (FREE-D) system,” Proc. IEEE, vol. 100, no. 1,pp. 138–149, Jan. 2012.

[20] M. Kiani and M. Ghovanloo, “The circuit theory behind coupled-modemagnetic resonance-based wireless power transmission,” IEEE Trans.Circuits Syst. I, Reg. Papers, vol. 59, no. 9, pp. 2065–2074, Sep. 2012.

[21] K. Chen and Z. Zhao, “Analysis of the double-layer printed spiral coilfor wireless power transfer,” IEEE J. Emerg. Sel. Topics Power Electron.,vol. 1, no. 2, pp. 114–121, Jun. 2013.

[22] W. R. Smythe, Static and Dynamic Electricity, 2nd ed. New York, NY,USA: McGraw-Hill, 1950, pp. 270–271.

[23] C. Gabriel, “Compilation of the dielectric properties of body tissues atRF and microwave frequencies,” Air Force Mater. Command, BrooksAir Force Base, Antonio, TX, USA, Tech. Rep. AL/OE-TR-1996-0037,1996.

[24] (2014, Apr. 8). RF Tools [Online]. Availble: http://www.pennstatehershey.org/web/nmrlab/resources/software/rftools

[25] M. Quirini, R. J. Webster, A. Menciassi, and P. Dario, “Design of apill-sized 12-legged endoscopic capsule robot,” in Proc. IEEE Int. Conf.Robot., Apr. 2007, pp. 1856–1862.

[26] S. C. Tang, S. Y. R. Hui, and H. Chung, “Coreless planar printed circuitboard (PCB) transformer—A fundamental concept for signal and energytransfer,” IEEE Trans. Power Electron., vol. 15, no. 5, pp. 931–941,Sep. 2000.

[27] S. C. Tang, S. Y. R. Hui, and H. Chung, “A low-profile wide-bandthree-port isolation amplifier with coreless printed-circuit-board (PCB)transformers,” IEEE Trans. Ind. Electron., vol. 48, no. 6, pp. 1180–1187,Dec. 2001.

[28] J. C. Maxwell, Treatise on Electricity and Magnetism, vol. 2. Oxford,U.K.: Clarendon Press, 1873, sec. 715.

[29] D. Vilkomerson and T. Chilipka, “Implantable Doppler system for self-monitoring vascular grafts,” in Proc. IEEE Ultrason. Symp., vol. 1.Aug. 2004, pp. 461–465.

[30] D. Vilkomerson et al., “Implantable ultrasound devices,” Proc. SPIE,Med. Imag., Ultrason. Imag. Signal Process., vol. 6920, p. 69200C,Mar. 2008, doi: 10.1117/12.772845.

[31] S. Y. Lee and S. C. Lee, “An implantable wireless bidirectional com-munication microstimulator for neuromuscular stimulation,” IEEE Trans.Circuits Syst. I, Reg. Papers, vol. 52, no. 12, pp. 2526–2538, Dec. 2005.

Sai Chun Tang (S’97–M’01–SM’11) was born inHong Kong, in 1972. He received the B.Eng. degree(with first class Hons.) and the Ph.D. degree inelectronic engineering from the City University ofHong Kong, in 1997 and 2000, respectively, wherehe was a Research Fellow after he graduated.

He joined the National University of Ireland,Galway, as a Visiting Academic in 2001, and thenthe Laboratory for Electromagnetic and ElectronicSystems at the Massachusetts Institute of Technol-ogy, Cambridge, in 2002. Since 2004, he has been

with the Radiology Department, Brigham and Women’s Hospital, HarvardMedical School, Boston, MA, for the developments of ultrasound diagno-sis devices and noninvasive treatment systems using high-intensity focusedultrasound. In 2008, he became a Faculty in Radiology at Harvard MedicalSchool. His current research interests include electronic medical devices, high-frequency electromagnetism, low-profile power converter design, and analogelectronics.