Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Copyrights: Restek Corporation
Considerations for Choosing a different
Carrier gas in Gas Chromatography
Considerations for Choosing a different
Carrier gas in Gas Chromatography
Jaap de Zeeuw
Restek Corporation,
Middelburg, The Netherlands
Copyrights: Restek Corporation
Considerations for using a certain carrier gasConsiderations for using a certain carrier gas
• Availability / delivery
• Price
• Purity
• Speed of analysis
• Type of detection system (PDD, MS..)
• Sensitivity (TCD)
Helium supply is under
pressure
Copyrights: Restek Corporation
What are the alternatives?What are the alternatives?
Nitrogen or Hydrogen“Slow” carrier gas
Cheap
Can be generated
“ Fast” carrier gas
Cheap
Can be generated
Safety
If there is “enough”
resolution
If there is “Just enough”
resolution
Copyrights: Restek Corporation
Both gases can be generated using generators
• Unlimited supply
• Low cost
• Relative small investment
For use as carrier gas need to make sure the purity is very high;
Consider Gas filtration
Both carrier gases are relatively CheapBoth carrier gases are relatively Cheap
Copyrights: Restek Corporation
1.0
0.6
0.2
HE
TP
(m
m)
Van Deemter Plot
Average Linear Velocity (cm/sec)
N2
He
H2
10 20 30 40 50 60 70
Van Deemter curve for different gasesoptimum gas velocitiesVan Deemter curve for different gasesoptimum gas velocities
Slow
FAST
Copyrights: Restek Corporation
Impact of optimum velocity on analysis timeImpact of optimum velocity on analysis time
6
Copyrights: Restek Corporation
Changing carrier gasChanging carrier gas
If a different carrier gas is used in the optimum velocity, the
temperature program has to be changed to get the same peak
elution order;
For N2: slower temperature programs = longer analysis time
For H2: faster temperature program = shorter analysis time
We can also use carrier gases deliberately “outside” their
optimum velocity;
If we use the SAME linear velocity as with helium, we do not have to change program and injection parameters;
Copyrights: Restek Corporation
Using a different carrier gases OUTSIDE the optimumUsing a different carrier gases OUTSIDE the optimum
If there is sufficient resolution, we can afford to loose some “plates”..
Using N2 at a HIGHER average velocity
Using H2 at a LOWER average velocity
Copyrights: Restek Corporation
Option 1:
Use Nitrogen at the SAME velocity as Helium..
..work at approximate 2x HIGHER then optimal velocity for N2..
Copyrights: Restek Corporation
Operating Nitrogen at higher velocitiesOperating Nitrogen at higher velocities
Using Nitrogen at the same linear velocity as Helium
will result in an increase of HETP by approx. a factor 2
Average linear velocity [cm/s]
Copyrights: Restek Corporation
Impact on ResolutionImpact on Resolution
A factor 2 lower efficiency means
only a factor 1.4 lower
resolution..
Loss of efficiency by a factor 2 only results in a loss of 1.4 on resolution
R = resolution
Nth = theoretical Plate numberα = selectivity
K = retention factorNth is linear with Column length
Resolution is not..
Copyrights: Restek Corporation
Impact on ResolutionImpact on Resolution
A factor 2 lower efficiency means
only a factor 1.4 lower
resolution..
Loss of efficiency by a factor 2 only results in a loss of 1.4 on resolution
Nth = 200.000 Nth = 50.000Nth = 100.000
Copyrights: Restek Corporation
Example using Nitrogen under “Helium” conditionsExample using Nitrogen under “Helium” conditions
Copyrights: Restek Corporation
Important practical implications using N2 at higher velocitiesImportant practical implications using N2 at higher velocities
• Peaks are BROADER so they are LOWER
• Response (height) will be 20-25% Lower
• Because peaks are broader, they will run faster in each other upon AGING
• More frequent maintenance• Column life time will be shorter
Copyrights: Restek Corporation
Option 2:
Use Hydrogen at the SAME velocity as Helium..
..work at approximate 30% SLOWER then optimal linear velocity for H2..
15
Copyrights: Restek Corporation
Option 2: Use Hydrogen at the SAME velocity as Helium..Option 2: Use Hydrogen at the SAME velocity as Helium..
Average linear velocity [cm/s]
Using Hydrogen at the same linear velocity as Helium will
result in an increase of HETP by approx. a factor 1.2
Copyrights: Restek Corporation
Setting same linear velocity for Hydrogen / Nitrogen/ HeliumSetting same linear velocity for Hydrogen / Nitrogen/ Helium
Carrier gases can be compressed, meaning at SAME
linear velocity, the Flows will be different..
Example below for a 30m x 0.25mm capillary, set at 30 cm/s at 40ºC
Gas Average Inlet pressure Flow
linear velocity
[cm/s] [kpa] [mL/min]
Helium 30.0 98.17 1.28
Nitrogen 30.0 87.93 1.23
Hydrogen 30.0 43.05 1.02
Copyrights: Restek Corporation
Using similar linear velocitiesUsing similar linear velocities
• Allows you to use the same temperature programming and injection conditions;
• Result is the same chromatogram with the same retention times
Copyrights: Restek Corporation
Comparison Comparison
25 cm/s, 0.85 mL/min He, 25 cm/s, 0.85 mL/min
2, 25 cm/s, 0.69 mL/min H2, 25 cm/s, 0.69 mL/min
R = 2.80
R = 2.68
Copyrights: Restek Corporation
He carrier, 1.4 mL/min constant flow, 34 cm/sec at 80°C
80°C (0.1 min), 16.7°C/min to 320°C (0.5 min)
FWHH
1.1 sec
FWHH
1.1 sec
H2 carrier, 1.06 mL/min constant flow, 34 cm/sec at 80°C
80°C (0.1 min), 16.7°C/min to 320°C (0.5 min)
13.64 min
13.50 min
8.10 min
8.03 min
Copyrights: Restek Corporation
He carrier, 1.4 mL/min constant flow, 34 cm/sec at 80°C
80°C (0.1 min), 16.7°C/min to 320°C (0.5 min)
Heptachlor
epoxideDDE
DieldrinEndrin
DDD
Endosulfan II
DDT
trans-
Chlordane
cis-
Chlordane Endosulfan I
H2 carrier, 1.06 mL/min constant flow, 34 cm/sec at 80°C
80°C (0.1 min), 16.7°C/min to 320°C (0.5 min)
Copyrights: Restek Corporation
What about using MS? Running H2 at SAME speed as HeWhat about using MS? Running H2 at SAME speed as He
Despite of running hydrogen at a LOWER velocity, and using the SAME
programming conditions, there is almost no difference in separation.
22
Copyrights: Restek Corporation
Optimal conditionsOptimal conditions
If analysis is optimized, analysis time is reduced by 40% using H2
Note the higher oven temperature rate
Copyrights: Restek Corporation
Using Nitrogen and Hydrogen under non optimal conditionsUsing Nitrogen and Hydrogen under non optimal conditions
Result in a loss of efficiency..
What if there is JUST ENOUGH separation
between the peaks?
Can only be used if:
“there is “ENOUGH” separation between the peaks
Copyrights: Restek Corporation
Complex chromatogramsComplex chromatograms
T im e - - >
Need to maintain efficiency (keep plates)
Copyrights: Restek Corporation
Maintain the separation efficiency using N2Maintain the separation efficiency using N2
Using same column and nitrogen under OPTIMAL conditions (12 cm/s)
Impact
• 2 x slower then Helium, analysis time will be 2x longer
• Peaks will be lower (sensitivity)
• Need to change temperature program for same peak elution order
Copyrights: Restek Corporation
Important : linear gas velocity vs oven temperature program Important : linear gas velocity vs oven temperature program
• If a different carrier gas is used and the SAME linear gas
velocity, the temperature program can stay the same;
• If the linear gas velocity is CHANGED, and the oven temperature program is NOT CHANGED, the elution
temperatures of components will change.
• This can result in a different peak elution order.
Copyrights: Restek Corporation
If elution temperatures are changed, also the peak elution can change..
Copyrights: Restek Corporation
Changing the flow keeping SAME temperature program.Changing the flow keeping SAME temperature program.
Result: elution temperatures will change:
Higher flow = lower elution temperature
At 8.9 psi, u = 22 cm/s
At 26 psi, u = 61 cm/s
Tel. = 250ºC
Tel. = 210ºC
Ref.: Jason Thomas, restek corporation
4,4’-DDE
4,4’-DDE
Endosulfan I
Endosulfan I
Copyrights: Restek Corporation
29
Copyrights: Restek Corporation
Using a smaller column diameter
Replacing the 30m x 0.25mm for a 20m x 0.15mm and operate the 0.15mm at a little higher velocity
Analysis times are approximate 2x shorter
Maintain the separation efficiency using N2Maintain the separation efficiency using N2
Copyrights: Restek Corporation
Using a 0.15mm ID instead of 0.25mmUsing a 0.15mm ID instead of 0.25mm
0 10 20
0 2 4 6 8 10 12 14
Time (min)
20
40
28 min
15 min
30m x 0.25mm Rxi-5Sil MS,
df = 0.25 μm
20m x 0.15mm Rxi-5Sil MS,
df = 0.15 μm
Copyrights: Restek Corporation
Calculating the impactUsing method translatorCalculating the impactUsing method translator
Can use the SAME oven program
Same analysis time
Helium Nitrogen
30/0.25 20/0.15
32 21.5 cm/s
Copyrights: Restek Corporation
Using a smaller column diameter
Using 0.15/0.18mm ID and operate at higher velocity
Impact
• Comparable analysis time;
• Can use same programming conditions
• Need to purchase a different column
• Lower loadability
Maintain the separation efficiency using N2Maintain the separation efficiency using N2
Copyrights: Restek Corporation
The Practical test
Copyrights: Restek Corporation
ChallengeChallenge
• Take POLAR column : Stabilwax
• Use COMPLEX mixture : Fragrances
• Compare 30m x 0.25mm with 20m x 0.15mm
Operate with same temperature programming using pre-selected flow / linear velocities
Efficiency optimized flow setting
Copyrights: Restek Corporation
Copyrights: Restek Corporation
Optimized Flows for He and N2Optimized Flows for He and N2
Efficiency Speed
optimized Optimized
0.25mm He 1.4 mL/min 2.0 mL/min
0.15mm N2 0.27 mL/min 0.38 mL/min
0.15mm N2 0.36 ml/min 0.52 mL/min
Actual Flows used to use the same program and obtain same analysis time
Copyrights: Restek Corporation
Efficiency optimized flow
Copyrights: Restek Corporation
Copyrights: Restek Corporation
30m x 0.25mm x 0.25µm Stabilwax
He constant 1.4 mL/min
40°C (0.1 min), 6.4°C/min to 250°C
6.81 min 31.23 min
Efficiency-optimized flow
Optimal heating rateMen’s After Shave
20.2m x 0.15mm x 0.15µm Stabilwax
N2 constant 0.36 mL/min
40°C (0.1 min), 6.4°C/min to 250°C
6.75 min 31.24 min
Men’s After Shave
Copyrights: Restek Corporation
Detail of separation
Copyrights: Restek Corporation
Efficiency optimized flow
Copyrights: Restek Corporation
Copyrights: Restek Corporation
Note the Pressures and flows usedNote the Pressures and flows used
For Helium 15.27 psi 1.40 mL/min
For Nitrogen 17.42 psi 0.36 mL/min
Very small difference
4 times lower flow
Copyrights: Restek Corporation
Features/Benefits using N2 and 0.15mm columnsFeatures/Benefits using N2 and 0.15mm columns
• Separations are exact identical
• Analysis time is the same
• No change in oven temperature program
• Pressure required for N2 is just a little higher
• Nitrogen is always available
• Price advantage for nitrogen and volume used
• 0.15mm columns have the same OD as 0.25mm: can use
the same ferrules
• 0.15mm columns have proven to work for many years and can be manufactured with different phase technologies
Copyrights: Restek Corporation
LimitationsLimitations
• Loadability/capacity is about 5 times less: for real trace analysis this will be a challenge
• Nitrogen is not a good MS gas, maybe in future..
Copyrights: Restek Corporation
Or..Or..
Move to a Hydrogen as the carrier gas
Copyrights: Restek Corporation
Why is hydrogen of interest ?Why is hydrogen of interest ?
• Readily available and can be produced in the lab
• Economical
• Can be used with many applications
• Offers an option to have also a faster analysis
• Peaks are 2x higher (sensitivity increase)
• For same data, need 50% of injected amount;
32
Copyrights: Restek Corporation
What are potential issues with using hydrogen?What are potential issues with using hydrogen?
• Safety aspect
• Potential leaks in the oven
• Leaks in the lab
• Split vent flows
• High pressures
• Reactivity
• Hydrogen can react with unsaturated compounds
• Impact on detector response
• FID signal
• MS fragmentation and sensitivity
33
Copyrights: Restek Corporation
Use of H2: Readily available and can be produced in the labUse of H2: Readily available and can be produced in the lab
H2-generators
• Unlimited supply
• Low cost
For use as carrier gas need to make sure the purity is very high;
Consider Gas filtration
NOTE: Most labs already have H2-line for operating FID.. This gas can be
used also, but need to verify cleanliness..
Copyrights: Restek Corporation
Optimal linear velocities for different carrier gasesOptimal linear velocities for different carrier gases
Hydrogen Uopt. = 40 - 45 cm/s
Helium Uopt. = 25 - 33 cm/s
Nitrogen Uopt. = 8 - 14 cm/s
Hydrogen is the
fastest carrier gas..
The same separation at
approx 1.7 shorter run time
Copyrights: Restek Corporation
20m x 0.15mm x 0.15µm Stabilwax4.22 min
6.74 min
20m x 0.15mm x 0.15µm Stabilwax6.68 min
10.67 min
134.1°C
133.6°C
Helium
5°C/min
Hydrogen
8.5°C/min
Fragrance analysis: Hydrogen instead of HeliumFragrance analysis: Hydrogen instead of Helium
Copyrights: Restek Corporation
Oven temperatureOven temperature
If the oven temperature is NOT adjusted:
• Gain in shorter analysis time is not as big as expected
37
Copyrights: Restek Corporation
Impact of using hydrogen, Using the SAME oven temperature programImpact of using hydrogen, Using the SAME oven temperature program
2 4 6 8 10 12 14 16
Time (min)
0 2 4 6 8 10 12 14 16
Time (min)
30 cm/s; 1.0 mL/min
Elution temperatures are lower
Only 1.5 minute faster..
Peaks start to shift
Helium: 1.0 mL/min
Hydrogen: 2.0 mL/min
38
Copyrights: Restek Corporation
Oven temperatureOven temperature
If the oven temperature is NOT adjusted:
• Gain in shorter analysis time is not as big as expected
• Peaks may start to move:
• Some separations will improve
• Some separations will go worse
• Peak elution can change
Example:
Chlorinated
Pesticides
39
Copyrights: Restek Corporation
a -ChlordaneEndosulfan I
4,4'-DDE
Dieldrin
Program 9 ºC/minProgram 9 ºC/min
Impact of elution temperature on Separation
T el= 204ºC
T el= 223ºC
T el= 214ºC
T el= 248ºC
Copyrights: Restek Corporation
New Calculator for oven programmingNew Calculator for oven programming
Free On-Line & windows based
Method translator & Flow calculator
Copyrights: Restek Corporation
New Calculator for oven programmingNew Calculator for oven programming
On-Line & windows based Method
translator & Flow calculator
46
Copyrights: Restek Corporation
New Calculator for oven programmingNew Calculator for oven programming
On-Line & windows based Method translator & Flow calculator
44
Calculator is available for FREE available on USB or can
be downloaded or accessed from website
Copyrights: Restek Corporation
Temperature programs needed to get the SAME elution temperatureTemperature programs needed to get the SAME elution temperature
2 4 6 8 10 12 14 16
60ºC, 2 min → 250 ºC @ 10ºC/min
60ºC, 1 min → 250 ºC @ 20ºC/min
Elution temp E12= 185.9 ºC
He, 1 mL/min
H2, 2 mL/min
2x faster analysis
43
Copyrights: Restek Corporation
Industrial application: Solvent impurity analysisIndustrial application: Solvent impurity analysis
15 min
Ref: J vercammen, Interscience Belgium
Copyrights: Restek Corporation
Hydrogen : safety issuesHydrogen : safety issues
Need to make sure there is no accumulation of H2
possible in the oven..
Hydrogen is combustible over a concentration range of 4% to 74% by volume;
What can we do to reduce this risk?
Copyrights: Restek Corporation
RisksRisks
When the column breaks
Amount of H2 released depends on where the column breaks
• End
• Middle
• Front • Limit the flow of H2
• Reduce risk of column breakage
50
Copyrights: Restek Corporation
Flow restriction to set with GC (electronic flow setting)Flow restriction to set with GC (electronic flow setting)
• Using Flow-controlled operation, the amount of H2
flowing in the GC is controlled.
• When column breaks at inlet, only the volume in injector can enter the GC + column volume
If this happens the GC will jump immediate in “standby”
mode as it detects an error because the pressure is not correct..
Copyrights: Restek Corporation
Da Vincihttp://www.davinci-ls.com/
SIMhttp://sim-gmbh.de/en/
Copyrights: Restek Corporation
Use metal capillary columnsUse metal capillary columns
• Metal columns are virtually unbreakable
• Already widely used for Simdist and Biodiesel, as well as in process and portable systems
• Many phase technologies available
Today’s series metal columns are VERY inert and
compete very well with fused silica
Copyrights: Restek Corporation
Inertness Test for 10m x 0.32mm metal column @ 110 ºCInertness Test for 10m x 0.32mm metal column @ 110 ºC
This is just a 0.1 μm film!
54
Copyrights: Restek Corporation
Use of H2: how big is the risk?Use of H2: how big is the risk?
At Restek, Chrompack, Varian and J&W, millions of GC columns have been tested using hydrogen.
55
Copyrights: Restek Corporation
Split outlet: What are the chances that hydrogen built
up in the lab? Considerations
Split outlet: What are the chances that hydrogen built
up in the lab? Considerations
• Volume of the Lab
• Total split flow of Hydrogen
• Hydrogen is lighter then air
• Fast diffusion (20m/s)
Copyrights: Restek Corporation
What to do with the hydrogen that
is splitted in the lab?
What to do with the hydrogen that
is splitted in the lab?
This is how we did it 30 years ago
Copyrights: Restek Corporation
What to do with the hydrogen that
is splitted in the lab?
What to do with the hydrogen that
is splitted in the lab?
This is how we did it 30 years ago
58
Copyrights: Restek Corporation
Reality: Lab ventilationReality: Lab ventilation
• Normal lab ventilation
• Air conditioning
• Fume hoods
• Local hoods
One can also install “vent”
line at the split outlet..
Copyrights: Restek Corporation
Use of HydrogenUse of Hydrogen
Reactivity?
Copyrights: Restek Corporation
Hydrogen is a reducing gasHydrogen is a reducing gas
Styrene → ethylbenzene
Higher injector temperatures
more ethylbenzene is formed
Copyrights: Restek Corporation
Alumina columns and hydrogenAlumina columns and hydrogen
• There has been no proof that Hydrogen can cause reactions with unsaturated hydrocarbons in the alumina PLOT
• Studies were done 25 years ago and showed up to 200ºC no reactivity (using KCl deactivated columns)
Problem may be present as today there are:
• More polar alumina columns
• Using different deactivations and different types of alumina
• Higher Operation temperatures
62
Copyrights: Restek Corporation
Solvents and HydrogenSolvents and Hydrogen
Some solvents can form acids when evaporated in H2
Result: Faster activation of Injection port liners
• Methylene chloride
• Chloroform
• CS2
For Split : minimal impact
For Splitless : bigger impact
May consider to change solvent: FE use ethylacetate
63
Copyrights: Restek Corporation
Important additional advantage besides 2x fasterImportant additional advantage besides 2x faster
Peaks are 2x narrower, they are 2x higher
For same sensitivity only 50% has to be injected..
This means:
• 2x less liner contamination
• 2x less column contamination
• Reduced costs per analysis
64
Copyrights: Restek Corporation
Hydrogen and MSHydrogen and MS
• Higher flow rate (use 0.18mm columns)
• Mass spectra may show bias (polar compounds)
• Initial activity of ion source (need some conditioning)
• 3 times lower of sensitivity (compensated by lenses)
• Hydrocarbon background (contamination of EPC)
• H2 – air mixtures in MS = deadly for MS (need service)
65
Copyrights: Restek Corporation
Changing carrier gas from heliumChanging carrier gas from helium
If peaks are well resolved:
- Nitrogen at higher velocity, keep conditions the same, loose some
separation, deal with shorter column life;
- Hydrogen at slower velocity, keep condition the same, small impact on
separation, deal with safety issues;
If peaks are just resolved:
- Nitrogen at optimum speed: accept longer analysis time;
- Nitrogen and smaller bore capillary; need some method development;
- Hydrogen at optimum velocity: need method development, has benefits
on analysis time and sensitivity; deal with safety issues;
Copyrights: Restek Corporation
Thank you for your kind attention