2 4 Method Validation HPLC Case Study

Embed Size (px)

DESCRIPTION

Method Validation

Citation preview

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    1/34

    Method and Validation basics

    HPLC case study

    Hua YIN(Assessor)

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    2/34

    The prequalification programme --Assessor's training | 19-20 January 20112|

    Outline

    HPLC methodology

    - Content of HPLC test procedure

    - System Suitability Testing (SST)

    - Relative Response Factor (RRF)

    Validation of HPLC method

    case study

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    3/34

    The prequalification programme --Assessor's training | 19-20 January 20113|

    Information Sources

    FDA CDER reviewer guideline for validation ofchromatographic methods (1994)

    WHO TRS 937 Appendix 4 Analytical Method Validation

    2006

    ICH Q2(R1) 2005

    Compendial General Chapters

    Methods and Validation presentationLynda Paleshnuik

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    4/34

    The prequalification programme --Assessor's training | 19-20 January 20114|

    High Performance Liquid Chromatography (HPLC)

    HPLC is a separation technique based on a solid stationary phase and

    a liquid mobile phase. Separations are achieved by partition,

    adsorption, or ion-exchange processes, depending upon the type of

    stationary phase used.

    Chiral

    Ion--exchange

    Ion--pair/affinity

    Normal phase

    Reversed phase

    Size exclusion

    The reversed-phase HPLC with UV detectionis most commonly usedform of HPLC, is selected to illustrate the parameters of HPLC

    method and validation.

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    5/34

    The prequalification programme --Assessor's training | 19-20 January 20115|

    A f low scheme for HPLC

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    6/34

    The prequalification programme --Assessor's training | 19-20 January 20116|

    Content of HPLC test procedure

    Any analytical procedure submitted should be described in

    sufficient detail, includes:

    Preparation of mobile phase

    Chromatographic condition:

    Column: type (e.g., C18 or C8), dimension (length, innerdiameter), particle size (10m, 5 m)

    Detector: wavelength

    Injection volume

    column T

    flow rate,

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    7/34

    The prequalification programme --Assessor's training | 19-20 January 20117|

    Content of HPLC test procedure

    Elution procedure: isocratic or gradient elution

    Preparation of standards and samples

    Operation procedure: sequence of injections

    System suitability testing (SST) and criteria

    Calculations

    QOS 2.3.R.2analytical procedures and validation summaries

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    8/34

    The prequalification programme --Assessor's training | 19-20 January 20118|

    Compendial methods

    When claim a compendial method, there should be no change in:

    The type of column i.e the stationary phases

    Detector wavelength

    Components in Mobile phase

    System suitability testing and criteria

    Adjustments to ratio of components in mobile phase, flow rate, columntemp, dimension of column, particle size (reduction only), may be

    necessary to achieve the system suitability criteria. The allowablevariations for each parameter, see Int.Ph 1.14.4 or USP generalchapter .

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    9/34

    The prequalification programme --Assessor's training | 19-20 January 20119|

    System suitability testing (SST)

    Precision: Assay: RSD 1% (API) or 2% (FPP), n 5

    Impurities: in general, RSD 5% at the limit level, up to 10% or

    higher at LOQ, n 6

    Resolution (R): >2

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    10/34

    The prequalification programme --Assessor's training | 19-20 January 201110|

    System suitability testing (SST)

    Tailing factor/peak asymmetry: ( 2)

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    11/34

    The prequalification programme --Assessor's training | 19-20 January 201111|

    System suitability testing (SST)

    Number of theoretical plates (N): column efficiency 2000

    Gradient elution is one way to increase the N

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    12/34

    The prequalification programme --Assessor's training | 19-20 January 201112|

    System suitability testing (SST)

    A SST should contain:

    For Assay:

    precision + one or more other parameter

    For impurity test:

    resolution + precision+ one or more other parameter

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    13/34

    The prequalification programme --Assessor's training | 19-20 January 201113|

    Relative Response Factor (RRF)

    Quantitation of Impurities

    Against impurity RSs: when reference standard available

    Against API itself

    Relative response factor should be considered

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    14/34

    The prequalification programme --Assessor's training | 19-20 January 201114|

    Relative Response Factor (RRF)

    Response factor: the response (e.g. peak area) of drugsubstance or related substances per unit weight.

    RF= peak area / concentration (mg/ml)

    Relative response factor (RRF):

    RRF=RFimpurity/ RFAPI,OR,

    RRF=slopeimpurity/ slopeAPI

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    15/34

    The prequalification programme --Assessor's training | 19-20 January 201115|

    Relative Response Factor (RRF)

    Rifampicine:

    y =31.312x + 4.963

    Rifampicine Quinone:

    y = 26.198x + 1.154

    RRF= 26.198/ 31.312

    =0.84

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    16/34

    The prequalification programme --Assessor's training | 19-20 January 201116|

    Relative Response Factor (RRF)

    To review:

    a) RRF calculation, and

    b) if RRF is properly used in the final calculation for %

    impurity

    If RRF within 0.8-1.2, correction may not be necessay

    Correction factor= 1/RRF, the reciprocal of the RRF

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    17/34

    The prequalification programme --Assessor's training | 19-20 January 201117|

    Review points for HPLC method

    is the analytical proceduredescribed in detail including all the parameters ? is SSTwell defined to ensure the consistency of system performance?

    The preparation of solutions:

    assay: concentration of reference standard should be close to thesample solution

    impurities: concentration of the reference standards should be close tothe limit

    The way of quantitation of impurities

    In case API is used as the reference, RRF should be used or justification ofexclusion should be provided.

    To check the determination of RRF, check the correction of calculation ofimpurities

    confirm/complete the QOS 2.3.R.2

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    18/34

    The prequalification programme --Assessor's training | 19-20 January 201118|

    Validationcompendial methods

    AssayAPI

    No validation generally required. Exception: specificity for majorimpurities not in the monograph.

    AssayFPP

    Specificity, accuracy and precision (repeatability).PurityAPI and FPP

    Full validation for specified impurities that are not included in themonograph (specificity, linearity, accuracy, repeatability,intermediate precision, LOD/LOQ)

    Validation of the limit for individual unknowns, if tighter than that inthe monograph: LOQ of the API should be below the limit forindividual unknowns

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    19/34

    The prequalification programme --Assessor's training | 19-20 January 201119|

    Non-compendial methods

    Full validation is required for purity, assay and dissolution methods(HPLC, UV) :

    Specificity

    Linearity

    Accuracy

    Repeatability

    Intermediate precision

    LOD/LOQ (not required for assay, dissolution)

    Robustness (recommended)

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    20/34

    The prequalification programme --Assessor's training | 19-20 January 201120|

    Specificity

    Blank solution to show no interference

    Placebo to demonstrate the lack of interference from excipients

    Spiked samples to show that all known related substances are

    resolved from each other

    Stressed sample of about 10 to 20% degradation is used to

    demonstrate the resolution among degradation products

    Check peak purity of drug substance by photodiode array detector (PDA): eg

    purity angle is lower than the purity threshold.

    Representative chromatograms should be provided with time scale

    and attenuation indicated

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    21/34

    The prequalification programme --Assessor's training | 19-20 January 201121|

    Linearity / Range

    The working sample concentration and samples tested foraccuracy should be in the linear range (concentrations Vs.

    Peak areas)

    Minimum 5 concentrations

    Dilute of stock solution or separate weighings

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    22/34

    The prequalification programme --Assessor's training | 19-20 January 201122|

    Linearity / Range

    Assay : 80-120% of the theoretical content of active

    Content Uniformity: 70-130%

    Dissolution: 20% of limits; eg if limits cover from

    20% to 90% l.c. (controlled release), linearity shouldcover 0-110% of l.c.

    Impurities: reporting level to 120% of shelf life limit

    Assay/Purity by a single method: reporting level ofthe impurities to 120% of assay limit

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    23/34

    The prequalification programme --Assessor's training | 19-20 January 201123|

    Linearity / Range

    Correlation coefficient (r)

    API: 0.998

    Impurities: 0.99

    y-Intercept and slope should be indicated together with

    plot of the data

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    24/34

    The prequalification programme --Assessor's training | 19-20 January 201124|

    Accuracy

    Assay

    API: against an RS of known purity, or via an alternate method ofknown accuracy; analysis in triplicate.

    FPP: samples/placeboes spiked with API, across the range of 80-120% of the target concentration, 3 concentrations, in triplicateeach.

    Report per cent recovery (mean result and RSD): 1002%

    ICH Q2 states: accuracy may be inferred once precision, linearity

    and specificity have been established. (Demonstration preferred).

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    25/34

    The prequalification programme --Assessor's training | 19-20 January 201125|

    Accuracy

    Impurities: API/FPP spiked with known impurities

    Experienced in PQ:

    Across the range ofLOQ-150%of the target concentration

    (shelf life limit), 3-5 concentrations, in triplicate each.(LOQ, 50%, 100%, 150%)

    Per cent recovery: in general, within 80-120%,depends

    on the level of limit

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    26/34

    The prequalification programme --Assessor's training | 19-20 January 201126|

    Precision

    System precision: by multiple injections (n 5) of a homogeneous sample

    (standard solution).

    RSD 1% is recommendedfor assay;

    RSD 5% is recommendedfor related substances (referencestandards at the limit)

    Indicates the performance of the HPLC system

    As a system suitability test

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    27/34

    The prequalification programme --Assessor's training | 19-20 January 201127|

    Precision

    Repeatability (method precision) Multiple measurements of a sample by the same analyst

    A minimum of 6 determinations at the test concentration (6

    times of a single batch), or

    3 levels (80%, 100%, 120%) , 3 repetitions each (combined

    with accuracy)

    For Assay: RSD 2.0%

    For individual impurity above 0.05%, in general, RSD 10%

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    28/34

    The prequalification programme --Assessor's training | 19-20 January 201128|

    Precision

    Intermediate precision (part of ruggedness) Test a sample on multiple days, analysts, equipments

    Repeat the method precision by different analyst in

    different equipment using different lot of column on

    different days

    RSD should be the same requirement as method precision

    Reproducibility (inter-laboratory trial)

    Not requested in the submission

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    29/34

    The prequalification programme --Assessor's training | 19-20 January 201129|

    LOD/LOQ

    signal to noise ratio: LOD 3:1 , LOQ 10:1 May vary with lamp aging, model/manufacturer of detector, column

    standard deviation of the response and the slope of the calibrationcurve at levels approximating the LOD /LOQ

    = the standard deviation of the response, base on the standard deviation of the blank

    The calibration curve

    should be validated by analysis of samples at the limits.

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    30/34

    The prequalification programme --Assessor's training | 19-20 January 201130|

    LOD/LOQ

    LOD: below the reporting threshold

    LOQ: at or below the specified limit

    Not required for assay/dissolution methods.

    Applicant should provide

    the method of determination

    the limits,

    chromotograms

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    31/34

    The prequalification programme --Assessor's training | 19-20 January 201131|

    Robustness

    The method's capability to remain unaffected by small butdeliberate variationsin method parameters

    Influence of variations of pH in a mobile phase

    Influence of variations in mobile phase composition

    Different columns (different lots and/or suppliers)

    Temperature

    Flow rate

    Evaluate the System suitability parameters

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    32/34

    The prequalification programme --Assessor's training | 19-20 January 201132|

    Robustness

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    33/34

    The prequalification programme --Assessor's training | 19-20 January 201133|

    Conclusion

    HPLC methods play a critical role in analysis ofpharmaceutical product

    Validation of HPLC should demonstrate that the method

    is suitable for its intended use

    Review the information in dossier against QOS 2.3.R.2

    Data for acceptance, release, stability will only be

    trustworthy if the methods used are reliable

  • 5/22/2018 2 4 Method Validation HPLC Case Study

    34/34

    The prequalification programme --Assessor's training | 19-20 January 201134 |

    Case Study

    Case Study 1--HPLC Method.doc

    QOS 2.3.R.doc

    Case Study 2 -- Validation of HPLC Method.doc

    http://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%201--HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/QOS%202.3.R.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%202%20--%20Validation%20of%20HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/QOS%202.3.R.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/QOS%202.3.R.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/QOS%202.3.R.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%201--HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%201--HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%201--HPLC%20Method.dochttp://localhost/var/www/apps/conversion/tmp/scratch_5/Case%20Study%201--HPLC%20Method.doc