CHAPTER- 4

INSTRUMENTATION

AND

METHOD CHARACTERIZATION

Page 2: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

4.1. Characterization technique of ARC glass substrate

Following instruments would be require for method characterization of ARC solution and

ARC glass substrate-

4.1.1. pH meter

4.1.2. UV-VIS Spectrometer

4.1.3. Atomic Force Microscope (AFM)

4.1.4. Climate test chamber-WK340/40

4.1.1. pH meter:

A pH meter is an electronic device, which is used for measure the pH of a liquid. A typical PH

meter consists of a special measuring probe that is a glass electrode, connected to an

electronic meter that displays & measures the pH value. A pH meter measure a value that how

liquid is acidic or alkaline. The theory of the pH meter is to determine the conc. of H+. When

dissolved the acid in water form positively charged H+ and the higher conc of H+ ions show the

acid is stronger. In the same way when dissolved alkali in water form negative charged hydroxyl

ions and the higher conc of hydroxyl ions show the base is the stronger. The amt of hydrogen

ions present in sol is dissolved in some amt of water find out the value of pH.

A pH = 7.0, show that, solution is a neutral means pure water having a pH = 7.0. When pH

values < 7 it’s indicate solution is acidic while a pH value > 7 will indicate sol is basic. pH value

of sol is 1.0 than it’s a extremely acidic sol and a sol of pH = 14.0 is than its extremely alkaline.

Page 3: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 41: pH meter

Probe:

Probe is a key part of a pH meter. It is a rod like structure generally made by glass. At the

bottom of the probe there is a bulb, the bulb is a very sensitive part of a probe that contains the

sensor. To determine the pH of a soln, the probe is proper dipped into the soln. Probe also

making with 02 nos. electrodes. These electrical signals go through a probe to a meter its

shows the pH value reading. Probe having 02 nos. electrodes; first one is a glass sensor

electrode and another than is a reference electrode. Both type of electrodes are hollow bulbs,

which contain a KCl sol, and also suspended AgCl wires in it. Glass sensing electrode making

with an extraordinary glass coated with metal salts & silica [101].

Figure 42: pH Electrode

Calibration and Operation of pH Meter:

pH Meter Instrument, calibration and operating procedure can be define shortly in following

simple steps:

Page 4: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

• First prepare buffer solution for 4.01, 9.18 and 10.01. So for dissolve one capsule of

each buffer in 100 ml distilled water or HPLC grade water.

• Dip the electrode in PH 4.01 buffer solution and adjust display to get the reading 4.01.

• Carried out same exercise for buffer solution of 9.18 and 10.01.

• Display should be adjusted to specified buffer value.

• Ensure that temperature sensor is dipped in solution while taking pH and adjust knob to

at room temperature.

• Check the instrument twice with the above buffer solution.

• If the value observed is with in range of ± 0.05 from specific value, then pH meter is

ready for use.

• Wash electrode every time with distilled water or HPLC grade water, when buffer or test

solution is to be checked.

• Keep the electrode in distilled water or HPLC grade water after used.

• Select alkaline or acidic both buffer solution for calibration.

• Fresh buffer solution prepared after two weeks.

Precautions of glass probes during handling:

• All time glass probe tip kept proper wet.

• When a glass electrode alone then stored it’s in deep in an acidic soln.

• Distilled or DI water not used for longer-term.

• Joint electrodes are better stored immersed in the bridge electrolyte frequently

potassium chloride 3.0 M.

4.1.2. UV-VIS Spectrometer:

Page 5: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

UV Spectrophotometer (Perkin-Elmer Lambda 35) work on Beers & Lambert's Law, which

states that absorbance is depends on width of the absorbing layer & concentration. Here we

keep width constant. This is used for concentration determination in organic Solution [102]. In

this project we used UV-VIS spectrophotometer for treated glass substrates transmittance

analysis. The operating software of instrument is UV Winlab. UV Spectrophotometer (Perkin-

Elmer Lambda 35) image is as follow

Figure 43: UV-VIS Spectrometer for analysis of transmittance of glass substrate

The functional part image of UV –Vis Spectrophotometer for absorbance analysis of liquid

sample are as follow

Page 6: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 44: Functional part of UV-VIS Spectrometer

Operating procedure of UV-VIS Spectrometer:

• Power switch on of UV VIS Spectrometer instrument and computer.

• After initialization of instrument, open the UV Winlab software.

• Set all instrument parameters as well as analysis parameters

• Give command for instrument auto zero

• After completion of auto zero, give command for blank with BaSO4 std.

• After completion of auto zero, give command for blank with BaSO4 std.

• After completion of blank measurement take %transmittance for substrate.

Page 7: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Calibration of UV-VIS Spectrometer: Calibration of UV-VIS Spectrometer can be illustrate by

following flow chart-

Figure 45: Flow chart for calibration of UV-VIS Spectrometer

Page 8: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is
Page 9: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

External Calibration of UV-VIS Spectrometer:

• Baseline Flatness

• Wavelength Accuracy

• Stray Light

• Wavelength repeatability

• Photometric Accuracy

• Photometric Repeatability

Instrumental trouble during analysis:

During analysis of any solution we can be facing following problem

1) Lambda shifting.

2) Base line becomes broad.

3) Start up lambda did not show 650 nm.

4) Base line noise developing during the period of scanning.

5) Week lamp changing.

1) Lambda shifting: Following reason for lambda shifting-

• In proper solution

• Contamination in sample

• Low lamp energy

• Solution should be homogeneous

• Check the contamination

• Change the lamp.

2) Base line become broad: Following reason for base line become broad-

• Filter wheel motor problem

• Sample not proper

• Any alignment disturb in monochromator

Solution:

Page 10: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

• Call to service engineer

• Sample should be homogeneous

3) Start up lambda did not show 650 nm: It may be occurs due to D2 or Tungsten lamp is

fused or low energy.

Solution: To change the lamp. After Every change of lamp, fitment should be verified through

described method. These two are the observing points.

• Noise observation at initializing stage.

• Both light beam observation on the paper at single wavelength (326 nm).

4) Base line noise developing during the period of scanning: Following factor responsible

for base line noise.

• Due to lamp

• Due to filter wheel

• Due to open mirror

Solution:

• If lamp energy is low then change the lamp.

• Do not touch the mirror or filter wheel during lamp changing.

5) Week lamp changing: If during initialization 650 nm wavelength not show, that means lamp

is fused or low energy.

Solution: Change the lamp.

Another trouble causes during analysis can be show by following flow chart-

Page 11: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 46: Flow chart for cause of trouble in UV-VIS Spectrometer

Precautions during handling UV-VIS Spectrometer:

Page 12: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 47: Do/Don’t Chart for UV-VIS Spectrometer

Analysis of %Transmittance of glass substrate by using Integrating Sphere in UV-

VIS spectrometer:

For analysis of transmittance or Reflectance by UV-VIS Spectrometer, that provides us.

Integrating Labsphere (RSA-PE-20) accessory with instrument. The Labsphere RSA-PE-20 is a

diffuse reflectance and transmittance accessory designed for a number of PerkinElmer UV-VIS

spectrometers. The list of compatible spectrometers includes the Lambda 25/35/45 series as

well as the older models Lambda 2/12/14/20/40.

Description of Integrating Sphere (RSA-PE-20):

The Integrating Sphere (RSA-PE-20) accessory is specifically designed to measure the

reflectance or transmittance of solids, liquids, powders, or other small objects that can fit in the

transmittance or reflectance ports. Basic components of the accessory include the integrating

Page 13: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

sphere, transfer optics and detector pre amplification module. Although the RSA-PE-20 is a

single beam construction, the accessory is compatible with the double beam configuration of the

Lambda Series spectrometers and your UV Winlab software. The integrating sphere assembly,

shown in below Figure, is 50 mm in diameter and constructed of Spectralon. Spectralon is the

same material used on the reflectance standard supplied with the accessory.

Figure 48. Integrating sphere (RSA-PE-20)

The RSA-PE-20 integrating sphere features two ports: a transmittance port where the beam

enters the sphere, and a sample reflectance port. The illustration in Figure 7 shows the

accessory integrating sphere along with the transfer optics and preamplification board. The

silicon detector is mounted directly to the preamp board underneath. Two Spectralon baffles

inside the sphere shield the detector sensor from direct illumination by the ports.

Page 14: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 49: Constructions of Integrating Sphere

The sample holders, mounted either at the transmittance or sample reflectance port, clamp the

Sample Reflectance standard or sample against the respective port. Three sample holders are

provided with the accessory for mounting a sample either at 0° or 8° angles of incidence. The

sample holders mount directly to the port frame at each end of the sphere. The sphere detector

is a silicon photodiode installed at the detector port at the bottom of the integrating sphere. The

photodiode operates in photovoltaic mode; the preamplifier board for compatibility with the

instrument electronics conditions the detector output.

Integrating sphere accessory alignment: Proper alignment of integrating sphere is requiring

during installation of instrument. The Optical setup of the RSA-PE-20 accessory can be show

as in following figure-

Page 15: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 50: Optical setup of the RSA-PE-20 accessory

Sample Beam Energy Scan: Before analysis the sample we require sample beam energy scan

of instrument for that setup the following instrument parameter according to below table-

Instrument Parameters Setting

Lamps UV/Vis ON

Beam Single Beam

Method Scan

Data Interval 1 nm

Abscissa Start 250

Abscissa End 1100

Slit Mode Fix

Slit 2 nm

Page 16: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Integration Time 0.08 s

Ordinate Mode E1

Sample Beam Front

Attenuator 100%

Table 2: Instrument setup for sample beam energy scan

Page 17: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

100% Baseline Scan: After sample beam energy scan we require 100% Baseline

Scanning of instrument. For that purpose, setup the following instrument

parameter according to below table-

Instrument Parameters Setting

Lamps UV/Vis ON

Beam Double Beam

Method Scan

Data Interval 1 nm

Abscissa Start 250

Abscissa End 1100

Slit Mode Fix

Slit 2 nm

Integration Time 0.08 s

Ordinate Mode %T

Sample Beam Front

Attenuator 100% /100%

Table 3: Instrument setup for 100% Baseline Scane

Loading the Sample Holders:

Both transmittance and sample reflectance ports are 5/8-inch holes at either end of the

accessory sphere. The transmittance port is closest to mirror M3. A sample holder clamp fits

Page 18: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

over a dovetail extension located underneath each port, securing the reflectance standard or

sample tightly against the port opening.

The sample holder with the 8 ーwedge should be installed at the sample reflectance port when

taking 8 ーhemispherical measurements. The 0 ーsample holder should be installed at the sample

reflectance port when taking diffuse measurements with the specular component excluded.

Figure 8 describes the proper way to load the sample holder. When mounting a sample or

reflectance standard at any port, you should make sure the

reflecting surface lies flat against the wedge and completely fills the port surface area. To

remove the sample holder at the sample reflectance port, loosen the setscrews along the side of

the respective wedge and remove the entire sample holder assembly.

Figure 51: Loading a standard or sample in to sample holder

In some instances, you may need to load a light trap at the sample reflectance port. The light

trap can be loaded using either wedge or fits into the sample holder in the same manner as a

reflectance standard.

Theory of Operation:

Page 19: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

During the traditional measurement of sample absorption by a spectrometer, the relationship

between absorptance and transmittance of the sample beam is described by the Kirchoff

equation:

A + T = 1

Where: A = absorbance & T = transmittance.

When measuring the absorbance of a sample in the sample compartment, the detector signal of

the spectrometer represents the portion of the sample beam i.e. not absorbed or scattered by

sample.

When using the RSA-PE-20 accessory, it is convenient to use the Kirchoff relationship:

A + T + R = 1

where R = reflectance of the sample beam at the sample surface.

During reflectance analysis, the reflected component of the sample beam is collected by the

integrating sphere and detected by sphere detector.

Zero line Correction:

An integrating sphere is sensitive to small-angle scatter from the sample beam coupling optics.

The scattered radiation strikes the wall of the integrating sphere near the sample reflectance

port, creating a "halo" surrounding the port. This "halo-effect" gives rise to a small error in

measurement of reflectance factor which is most significant when measuring samples of very

low reflectance.

Fortunately, this error is relatively easy to characterize and correct. The following procedure

may be used to correct reflectance factor measurements of diffuse or specular samples, in

either the 8°/Hemispherical or 0°/Diffuse geometries:

1. Perform a reflectance factor measurement of a given sample as described in the operating

procedures.

2. Replace the sample with a light trap. The reflectance of the light trap is approximately

0%.

3. Execute a sample scan to record the measurement data on the light trap

Page 20: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

4. Compute the corrected reflectance factor for the sample, Ps, as follows:

Ps = (R – Z) Pr / (100 – Z)

where:

Pr = the reflectance value of the reference standard,

R = the sample data displayed on the instrument expressed in percent, and

Z = the data from the light trap measurement, expressed in percent.

Operating Procedures for the RSA-PE-20 Accessory:

The RSA-PE-20 accessory can measure transmittance factors at either 0° or 8° angles

of incidence. The angle of incidence is controlled by the sample holder configuration at

the sample reflectance port. You should keep in mind, however, that 8° hemispherical

measurements will include the specular component reflecting off the sample surface.

Although calibrations using the 0° geometry are available on a custom basis, the

standard calibrations at Labsphere are performed at 8° If your application requires NIST

traceability, you should use a calibrated reflectance standard and the 8° reflectance

geometry.

Since any integrating sphere accessory acts as a signal attenuator, you should run scans at a

slower scan speed than in ordinary transmittance work without the accessory, or increase the

integration time if collecting data at a fixed wavelength.

Transmittance Measurements:

ASTM E 179-91a describes three categories of transmittance measurement: regular

transmittance, diffuse transmittance, and total transmittance. These three terms relate to each

other in the same fashion as regular or specular, diffuse and total reflectance. Regular

transmittance is the ratio of the undiffused transmitted flux to the incident flux. Diffuse

transmittance is the ratio of transmitted flux measured at all forward angles except the regular

Page 21: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

transmittance angle, to the incident flux. Total transmittance is the ratio of flux transmittance at

all forward angles to the incident flux. The following procedure can be used to measure total

transmittance of a sample-

• Load a diffuse reflectance standard at the sample reflectance port.

• If you are scanning a liquid sample, load an empty cuvette into the cuvette holder, and

load the entire cuvette holder assembly at the transmittance port. Otherwise, leave the

transmittance port empty.

• Choose your transmittance sample method from the UV Winlab software or manually set

all instrument parameters as desired.

• Perform a background correction scan.

• Load your sample at the transmittance port. If you are measuring a liquid, replace the

empty cuvette with your sample cuvette.

• Execute a sample scan and save the scan data.

• The value displayed by the instrument is the sample transmittance.

The procedure here for total transmittance measurement will produce accurate results for non-

scattering samples. For transmittance samples exhibiting high degree of scattering, the scan

results will be instrument-specific.

We can measure the diffuse transmittance of a test sample using the following procedure:

• Perform a background correction scan on the accessory with a diffuse reflectance

standard loaded at the sample reflectance port, and an empty cuvette, if used, fitted in

front of the transmittance port.

• Replace the reflectance standard at the sample reflectance port with a light-trap. This

light trap will capture and exclude from measurement the undiffused portion of the flux

transmitted by the sample.

• Load the test sample or sample cuvette at the transmittance port and execute a sample

scan.

• Measuring the total and diffuse transmittance, as described above, and subtracting the

diffuse component from the total transmittance can determine an approximate regular

transmittance of a sample [103].

Page 22: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

4.1.3. Atomic Force Microscope (AFM):

After chemical treatment of glass substrate we require analysis of surface roughnes because

due to that type of data or image show the effect of chemical tretment or chemical etching on

glass surface, so for AFM is a most suitable instrument for that purpose. The surface

morphology of the glass substrate was examined by AFM.

AFM is a very high-resolution type of scanning probe microscopy, with demonstrated resolution

on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction

limit. In some variations, electric potentials can also be scanned by using conducting cantilevers

[104].

Page 23: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 52: Atomic Force Microscope (AFM)

AFM provides a 3D profile of the surface on a nano scale, by measuring forces b/w a sharp

probe (<10 nm) and surface. The probe is supported on a flexible cantilever. The AFM tip

“gently” touches the surface and records the small force b/w the probe and the surface. The amt

of force b/w the probe & sample is dependent on the spring constant of the cantilever & the

distance b/w the probe and sample surface. This force could be described using Hooke’s law:

F=-K.X

F= Force

K= Spring constant

X= Cantilever deflection

Probes are typically made from Si3N4 or Si. Different cantilever lengths, materials, and shapes

allow for varied spring constants and resonant frequencies.

Probes may be coated with other

materials for addition SPM applications such as chemical force microscopy (CFM) and magnetic

force microscopy (MFM).

Page 24: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 53: Tip of AFM

Scanning probe microscopes (SPM) describe a broad group of instruments used to image &

determine properties of material, chemical, and biological surfaces. SPM images are find out by

scanning a sharp probe across a surface while monitoring & compiling tip–sample interactions

to provide an image. The 02 primary forms of SPM are scanning tunneling microscopy (STM)

and AFM. The functional schematic flow diagram of atomic force microscope (AFM) is show as

like

Page 25: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 54: Schematic diagram of AFM

Systematic operating procedure (SOP) of AFM:

• AFM Instrument operating procedure can be define in following simple steps:

• Switch on the main power supply.

• Ensure air pressure must be open.

• Switch on the AFM control P.C. or monitor.

Page 26: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

• Switch on the nano scope scanning probe controller.

• Switch on the nano scope dimension controller.

• Now open nano scope SPM icon than open real time mode icon.

• Now select profile and load tapping AFM and go for initialization.

• Change the stage chuck as per sample requirement. During changing of chuck,

switch off the vacuum.

• Now put the sample on stage chuck.

• Move the table under scanner head and then focus the surface.

• Set the tip and adjust cantilever in central during surface examine or sampling

focusing.

• Care must be taken during focusing the sample that tip does not hit the sample

surface.

• Move the start of groove area measurement point under the cross hair.

• Adjust scan parameters, adjust groove by X, Y offset in scan control.

• If image is comfortable than capture it and save data or images in 2D or 3 D mode.

Page 27: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

4.1.4. Climate test chamber-WK340/40 (Environment Chamber):

So many chemical treated methods are available for increasing transmittance on glass

substrate for application in solar module but environmental stability is crucial factor for this layer.

Solar modules are make use in all our world for exclusion power crisis so keep this mind many

environment factor which will be directly effect anti reflection coating of solar module glass. The

above observations point to the fact that environmental test is must for solar module glass which

treated by any chemical process for improving transmittance. During this test we observe the

effect of different environment conditions like temperature, humidity on anti reflective coated

glass substrate and analyze visual defects, peak wavelength shifting of transmittance and the

reduction of transmittance with respect to chemical treated anti reflected coated glass.

It is the main focus area of our research that treated substrate layer will be environment stable.

It characteristics will isolate this process to other exist developing process. Due to analysis of

process stability we would be require following environment test i.e. (a) Thermal cycle test (b)

Damp heat test (c) Humidity freeze test. For those entire environment tests we require climatic

chamber or Environment chamber. Environmental chamber is a system of sufficient thermal and

humidification capacity shall be used to change ambient conditions to meet test necessities and

to reach specified temp. conditions. The temp. change test shall be executed in accordance IEC

61646 standard.

The International Electro technical Commission (IEC) is a worldwide organization for

standardization comprising all national electro technical committees (IEC National Committees).

IEC technical committee 82 has prepared International Standard IEC 61646: Solar photovoltaic

energy systems [105].

We have carried out all environment tests in Climate test chamber-WK340/40 (Weiss Technik).

Performance for temperature tests of Climate test chamber-WK340/40 are as

Maximum temperature (°C): +180

Minimum temperature (°C): -45

Page 28: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Temperature changing rate cooling (K/min): 2.0

Temperature changing rate heating (K/min): 3.0

Temperature changing rate linear (K/min): 1.0

Humidity range (% r.h.): 10 to 98

Calibration values (°C): +23°C / 50% r.h. and +95°C/ 50

Figure 55: Climate test chamber

Transmittance would be measure before and after taken environment test. We mention

conditions for different environment test i.e.

Thermal cycle test: This environment test consist 50,100 & 150 thermal cycle and each cycle

consist - 40* to 85* C temperature for 6.0 Hrs.

Damp heat test: During this environment test substrate putting for 1000Hrs.at temperature 85*

C and 85%RH.

Humidity freeze test: 85*C to –

4.2. Characterization Technique of Solar Modules:

Temperature changing rate cooling (K/min): 2.0

Temperature changing rate heating (K/min): 3.0

Temperature changing rate linear (K/min): 1.0

Humidity range (% r.h.): 10 to 98

Calibration values (°C): +23°C / 50% r.h. and +95°C/ 50 % r.h.

Figure 55: Climate test chamber-WK340/40 (Environment Chamber)

Transmittance would be measure before and after taken environment test. We mention

conditions for different environment test i.e.

This environment test consist 50,100 & 150 thermal cycle and each cycle

40* to 85* C temperature for 6.0 Hrs.

During this environment test substrate putting for 1000Hrs.at temperature 85*

– 40*C @ 85%RH 10Cycle each cycle=24Hrs [106].

4.2. Characterization Technique of Solar Modules:

WK340/40 (Environment Chamber)

Transmittance would be measure before and after taken environment test. We mention

This environment test consist 50,100 & 150 thermal cycle and each cycle

During this environment test substrate putting for 1000Hrs.at temperature 85*

40*C @ 85%RH 10Cycle each cycle=24Hrs [106].

Page 29: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Following instruments would be require during fabrication of Solar module and their

characterization-

4.2.1.Differential Scanning Calorimeter (DSC)

4.2.2. Sun simulator machine for I – V Characterization

4.2.1. Differential Scanning Calorimeter (DSC):

In our experiment Differential Scanning Calorimeter (DSC) is used for thermal analysis of tedlar

and EVA materials, which materials used in fabrication of solar modules. Typical applications of

this instrument like reaction kinetics, purity analysis, polymer cures through heats and

temperatures of transitions and reactions. A Differential Scanning Calorimeter (DSC) image is

show as follow

Figure 56: Differential Scanning Calorimeter instrument

A typical DSC cell uses a Cu-Ni disk for transferring heat to the sample and reference. The

differential heat flow to the sample and the reference is monitoring by the constantan/chromel

thermocouples formed by the junction of the constantan plate and the chromel wafer covering

the underside of each pan. DSC provides max calorimetric accuracy from -170°C to 750°C and

sample size from 0.10 mg to 40 mg. A typical DSC Machine flow diagram can be show as follow

Page 30: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is
Page 31: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 57: A typical DSC Machine flow diagram

Analysis Procedure of DSC:

DSC instrument operating procedure can be define in following steps

• Clean sample pan and cover by means of slow pressure nitrogen gas.

• Then tare weight pan and cover, weigh accurately (1.0mg to 9.0 mg) by a presicion

balance.

• Now pan and cover crimped by crmping machine to sealed pan.

• Then this kept in left side of the furnace and reference was placed in right hand positions

during the experimentations.

• Sample and reference lid handled with the help of tweezers only to minimize

interference.

• Keep nitrogen gas flow 20 ml/min, heating r

• After analysis, furnace cooled to at ambient conditions before next experiment.

• After reaching the furnace temperature at room temperature swith off the instrument and

now close the software.

Figure 57: A typical DSC Machine flow diagram

DSC instrument operating procedure can be define in following steps-

Clean sample pan and cover by means of slow pressure nitrogen gas.

Then tare weight pan and cover, weigh accurately (1.0mg to 9.0 mg) by a presicion

Now pan and cover crimped by crmping machine to sealed pan.

Then this kept in left side of the furnace and reference was placed in right hand positions

during the experimentations.

Sample and reference lid handled with the help of tweezers only to minimize

Keep nitrogen gas flow 20 ml/min, heating rate 10°C/minute from ambient to 500°C.

After analysis, furnace cooled to at ambient conditions before next experiment.

After reaching the furnace temperature at room temperature swith off the instrument and

Then tare weight pan and cover, weigh accurately (1.0mg to 9.0 mg) by a presicion

Then this kept in left side of the furnace and reference was placed in right hand positions

Sample and reference lid handled with the help of tweezers only to minimize

ate 10°C/minute from ambient to 500°C.

After analysis, furnace cooled to at ambient conditions before next experiment.

After reaching the furnace temperature at room temperature swith off the instrument and

Page 32: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Thermal Analysis of polymer by DSC:

Thermal analysis includes a group of technique in which specific physical properties of a

material are measured as a function of temperature. Different of thermal analysis techniques are

Differential Scanning Calorimetry (DSC), Differential Thermal Analysis (DTA),

Thermogravimetric Analysis (TGA), Thermomechanical Analysis (TMA), Dynamic Mechanical

Analysis(DMA), Evolved Gas Analysis (EGA).

In this technique, the sample and reference materials are subjected to a precisely programmed

temperature changed. Thermal energy is added to either the reference or the sample chambers

in order to maintain both reference and sample at same temp since the energy transferred is

exactly equivalent in magnitude to energy evolved or absorbed in the transition, the balancing

energy yields a direct calorimetric measurement of the transition energy. A polymer DSC curves

shown below in figure [102].

Figure 58: Example of a polymer curve in DSC Machine

Also show another curve in figure 59, which shows the temperature curve and indicates

exothermic and endothermic process.

Page 33: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 59: DSC Thermo gram different temperature points

The heat flow into the sample holder can be calculated by using the following equation:

( )TTKdT

dQb −=

Where,

T: Sample temp.

Tb: Programs block temp.

K: Thermal conductivity

qtTTb +=0

---------- [1]

Where,

T0 : Initial temp.

q : Programs heating rate.

Heat capacity expressed, as the amt of heat energy required for raise the temp of a body by 1K.

For a substance with a constant heat capacity, defined by the following eq;

( )0

TTCQ P −= --------- [2]

Page 34: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

To derive an eq. from eq. [1] & [2], that express for the DSC as follows:

K

CqT P=∆ ------------ [3]

Where,

∆T: Diff. in temp b/w the reference mat. & sample.

The heat capacity show as:

PP mcC = ---------- [4]

Where,

c: The specific heat

For a given phase transition the enthalpy change may found by integrating over the area where

the transition is seen to arise on the DSC curve. This change show with following integral eq:

∫∫

∆==∆

f

i

f

i

T

T

T

T

P dTq

TKdtCH ----------- [5]

Where,

Tf and Ti are the final and initial temperatures of integration limits for graph integration. So, it

indicates enthalpy changes involved in the various transition phases. Different transition

temperatures describe in footnotes as follows:

Glass Transition Temperature-Tg:

It is a temp at which, on heating, an amorphous polymer changes from being hard, brittle and

glass like to transformed a soft rubber like substance. The factors with relating to polymers and

their effect on Tg are show in following table:

Page 35: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Table - 4: Effect of Glass Transition Temperature (Tg)

Page 36: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Crystallization Temperatures- Tc:

It is the temp at which molecules gained sufficient freedom to move into a stable phase that is in

crystalline phase and a transition occurs.

Melting Temperatures- Tm:

It is temp at which the crystalline molecules turns liquid from solid state that have gained

enough vibrational freedom. For the increased freedom of these molecules, the thermo gram of

DSC shows sudden dip at this temp. M.P. of any materials could be show by following

equations:

m

m

mS

HT

∆= ----------- [6]

Where,

∆Hm: Melting temperature

∆Sm: Entropy change.

Degradation Temperature- Td:

It is the temperature, at which the molecules decompose; atoms start to break as vibrational

energy more in this stage. This temperature may be exothermic or endothermic depending upon

the nature of samples.

Page 37: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

4.2.2. Sun Simulator Machine (Spectranova) for I – V Characterization: Sun simulator

machine (Spectranova) is used for analyze electric parameters (current-voltage

characterization) of solar module. The image of Spectranova Sun simulator machine is as follow

Figure 60: Sun Simulator (Spectranova) Machine

The inside image of Spectranova Sun simulator, where we put the solar module or solar cell for

electric parameter analysis can be show as following

Page 38: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 61: Insight image of Sun Simulator (Spectranova) Machine

Page 39: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

For different electrical parameters like current, voltage, fill factor etc measured by Sun simulator

machine which having a light source and their filtering unit. When filtered light flashed on solar

module, measuring system of sun simulator machine analyze the electrical parameters due to

the variable change and their correction in temperature. We can show the sun simulator

machine process flow diagram is as follow in figure –

Figure 62: Schematic process flow diagram of Sun Simulator machine

Page 40: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Current-Voltage Characterization:

Photovoltaic cells modeled as a current source in parallel with a diode. When there is no light to

produce any current, then PV cell worked as a diode. When intensity of incident light raises, the

PV cell, as illustrated in Figure, generates current like

Figure 63: Current generation in PV cell

The current-voltage graph of a PV cell has illustrated in Fig as the V across the measuring load

is swept from zero to VOC and different electric parameters for the cell could be resolute from

this data

Page 41: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 64: I-V curve of PV cell

1) Open circuit voltage-Voc

2) Short circuit current-Isc

3) Max power (Pmax), Current @ Pmax (Imax), Voltage @ Pmax (Vmax)

4) %Fill Factor-FF

5) %Efficiency

1) Voc:

The VOC found when there is no any current passing through the cell like

V (at I=0) = VOC

2) Isc:

The ISC corresponds to the short circuit condition when the impedance is low and is calculated

when the voltage equals 0.

Page 42: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

I (at V=0) = ISC

3) Max. Power (Pmax), Current at Pmax (Imax), Voltage at Pmax (Vmax):

Power generated by the cell in watts could be calculated along the current-voltage sweep with

the equ. P=IV. At ISC and VOC points, the power would be zero (0) & the max value for power

will occur b/w the two. The voltage (V) and current (I) at this max power point are denote

as VMP and IMP correspondingly.

Figure 65: Maximum power (Pmax), Maximum Current (Imax) and Maximum Voltage

(Vmax)

4) %FF:

It is finding out by compare the max power to the hypothetical power (PT) that would be output

at both the Voc & Isc together. Fill factor could also be interpreted graphically as in Fig.

Page 43: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 66: Fill factor calculated by I

5) %Efficiency:

Efficiency is the ratio of electrical power output

into the photovoltaic cell. Pout can be taken to be

to its max power output to obtain the max efficiency.

I-V Test system:

First a light source providing incident radiation to excite the PV cell or module. After that, an

SMU like “PXI-4130” is used to sweep the voltage (V) & determine the current (I) from the photo

voltaic cell. In addition, several sensors are used to find

atmosphere.

Figure 66: Fill factor calculated by I-V curve

ratio of electrical power output Pout, with compare to the solar power input,

can be taken to be PMAX since the solar cell could be operated up

to its max power output to obtain the max efficiency.

First a light source providing incident radiation to excite the PV cell or module. After that, an

4130” is used to sweep the voltage (V) & determine the current (I) from the photo

voltaic cell. In addition, several sensors are used to find out the ambient conditions of

, with compare to the solar power input, Pin,

since the solar cell could be operated up

First a light source providing incident radiation to excite the PV cell or module. After that, an

4130” is used to sweep the voltage (V) & determine the current (I) from the photo

out the ambient conditions of

Page 44: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is

Figure 67: Test arrangement for Current-Voltage Characterization

Software like “LabVIEW, Kiethley” is used to obtain measure & display the value of current-

voltage characterization tests and evaluate the main performance parameters for solar panel.

Generally this instrument is used for small solar module current voltage characterization

purpose. To uphold the ambient temp and the temp of the photovoltaic cell, a cooling system

also be used to counter balance the heat from the source of light [107]. A Keithley (2400- A

source meter) instrument figure can be show as following

Figure 68: Keithley (2400- A source meter)

Page 45: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is
Page 46: 12 Chapter-4 Instrumentation and method characterizationshodhganga.inflibnet.ac.in/bitstream/10603/57564/12/12_chapter-4... · meter consists of a special measuring probe that is