Test Report
EN 149 : 2001 + A1 : 2009
Particle filtering half masks Report no: 1.18.06.57
Client: INSPEC Certification Services 56 Leslie Hough Way Salford Greater Manchester M6 6AJ United Kingdom
Manufacturer: Reckitt Benckiser Healthcare (UK) Limited Client order: TT18-0019
Order(s) received: 2 to 17 April 2018
Model(s): Mask: RB030VS and RB030VM Ventilator: RB030
Date(s) of tests: 26 April to 25 June 2018
Signed:
Issued: 29 June 2018
Heather Webb, Laboratory Supervisor Page 1 of 18
INSPEC Test Report No: 1.18.06.57 Page 2 of 18
Conditions This report may be reproduced and distributed to your clients, provided that it is reproduced and distributed in full. Unless stated otherwise, the testing is accredited under the laboratory’s ISO/IEC 17025 accreditation issued by ANSI-ASQ National Accreditation Board. Refer to certificate and scope of accreditation AT-1933. Tests marked � are not included in our ISO/IEC 17025 accreditation. Opinions, comments and interpretations expressed in this report are shown in italics. Copies of INSPEC interpretations referenced in this report are available upon request. Specimens will be disposed of four weeks from the date of this report, unless otherwise instructed. This report has been provided in accordance with our standard Terms of Business, which can be viewed at, and printed from: http://inspec-international.com/ToB.pdf If you have difficulty accessing the Terms of Business, you may contact us for a copy.
INSPEC Test Report No: 1.18.06.57 Page 3 of 18
Summary of assessment*
Clause Assessment (see Key)
Model: RB030VS + RB030C RB030VM + RB030C
7.4 Packaging Ltd
7.5 Material Ltd
7.6 Cleaning and disinfecting
7.7 Practical performance Ltd
7.8 Finish of parts Pass
7.9.1 Total inward leakage Pass
7.9.2 Penetration of filter material: Sodium chloride Pass
7.9.2 Penetration of filter material: Paraffin oil Pass
7.10 Compatibility with skin
7.11 Flammability Pass
7.12 Carbon dioxide content of the inhalation air Pass
7.13 Head harness Ltd
7.14 Field of vision Ltd
7.15 Exhalation valve(s) Pass
7.16 Breathing resistance Pass
7.17 Clogging
7.18 Demountable parts Pass
9 Marking
10 Information to be supplied by the manufacturer
*RB030C - valve cap
INSPEC Test Report No: 1.18.06.57 Page 4 of 18
Clause Assessment (see Key)
Model: RB030VS + RB030 RB030VM + RB030
7.4 Packaging Ltd
7.5 Material Ltd
7.6 Cleaning and disinfecting
7.7 Practical performance Pass
7.8 Finish of parts Pass
7.9.1 Total inward leakage Pass
7.9.2 Penetration of filter material: Sodium chloride
7.9.2 Penetration of filter material: Paraffin oil
7.10 Compatibility with skin Ltd
7.11 Flammability Ltd
7.12 Carbon dioxide content of the inhalation air Pass
7.13 Head harness Pass
7.14 Field of vision Pass
7.15 Exhalation valve(s) Pass
7.16 Breathing resistance Pass
7.17 Clogging
7.18 Demountable parts Pass
9 Marking
10 Information to be supplied by the manufacturer
Key
Shading shows the clauses requested. Any other clauses were not requested.
Pass Requirement satisfied.
Ltd Testing requested was insufficient completely to verify compliance with the clause. Refer to the “Result details” section for more information.
Fail Requirement not satisfied. Refer to the “Result details” section for more information.
NAs Assessment not carried out.
NAp Requirement not applicable.
NT Requested but not tested due to early termination following failure.
* Assessment relates only to those specimens which were tested and are the subject of this report.
INSPEC Test Report No: 1.18.06.57 Page 5 of 18
Product characteristics
Property Characteristic
Model RB030VS mask
with RB030C valve cap
RB030VM mask
with RB030C valve cap
Classification claimed FFP1 NR FFP1 NR
Exhalation valve(s) One One
Property Characteristic
Model RB030VS mask
with RB030 ventilator
RB030VM mask
with RB030 ventilator
Classification claimed FFP1 NR FFP1 NR
Exhalation valve(s) One + ventilator One + ventilator
Submission details
Product Quantity Date
received INSPEC specimen no.
(1F0170 +)
RB030VS Dettol Protect+ filtering half mask, size Small
120
19 March 2018
101 to 163 and 301 to 363
RB030VM Dettol Protect+ filtering half mask, size Medium
120 201 to 260 and 401 to 460
RB030 Dettol Protect+ ventilator 48 501 to 520
RB030 Dettol Siti shield ventilator 48 521 to 540
RB030 Siti Shield ventilator 38 544 to 563
INSPEC Test Report No: 1.18.06.57 Page 6 of 18
Procedures
The specimens detailed within the submissions above were used for the tests covered by this report.
Testing was performed in accordance with BS EN 149 : 2001 incorporating corrigendum No. 1 (July 2002) and amendment A1 (March 2009), unless otherwise specified below. Reference should be made to the standard when reading this report.
Unless stated otherwise, specimens were tested in the condition as received by INSPEC.
7.7 The client instructed that practical performance testing be carried out on one specimen of model RB030VS or RB030VM fitted with valve cap RB030C and a total of four specimens of model RB030VS or RB030VM fitted with ventilator RB030; two with the ventilator switched on and two with the ventilator switched off.
Practical performance tests were conducted in simulation of the practical use of the apparatus under the conditions prevailing in the gallery area of the laboratory. The exercises undertaken and the equipment used were as specified in the standard.
7.9.1 The client instructed that total inward leakage testing be carried out on 10 specimens of model RB030VS or RB030VM fitted with valve cap RB030C and 10 specimens of model RB030VS or RB030VM fitted with the ventilator; five with the ventilator switched on and five with the ventilator switched off.
7.9.2 Filter penetration testing by the paraffin oil method was carried out using a modified Phoenix SG-20 aerosol generator and a Phoenix model JM-6000 photometer or a TEC Services’ model PH-3 photometer. These give similar performance to the instruments specified.
For the 120 mg exposure test, the peak penetration during exposure is reported and in addition the penetration after three minutes for comparison purposes.
During the 120 mg exposure test, the sodium chloride penetration showed continued decline and the test was terminated as the product was marked NR.
The client instructed that filter penetration testing be carried out only on model RB030VS fitted with valve cap RB030C.
7.16 Exhalation resistance was tested at a continuous flow of 160 l/min.
The client instructed that breathing resistance be carried out only on model RB030VS fitted with valve cap RB030C and on model RB030VS fitted with ventilator RB030; inhalation with the ventilator switched on and exhalation with the ventilator switched off.
INSPEC Test Report No: 1.18.06.57 Page 7 of 18
Result details
7.4 Packaging
Model: RB030VS + RB030C
The masks were not packaged as offered for sale. Manufacturer to certify regarding the final packaging to be used. NAs
The masks were packaged in sealed plastic bags, in larger plastic bags inside a large cardboard box that gave some protection against mechanical damage or contamination before use. Pass
Model: RB030VM + RB030
The masks were not packaged as offered for sale. Manufacturer to certify regarding the final packaging to be used. NAs
The masks were packaged in sealed plastic bags, in larger plastic bags inside a large cardboard box that gave some protection against mechanical damage or contamination before use. Pass
7.5 Material
Model: RB030VS + RB030C
The materials used were able to withstand handling and wear during the limited laboratory testing carried out. Ltd
The effect on materials from “in-use” environmental factors could not be evaluated during laboratory tests. Manufacturer to certify regarding such factors. NAs
Specimens 120 to 125 were conditioned in accordance with 8.3.1. None of the specimens conditioned suffered mechanical failure or collapse. Pass
Specimens 106 to 110, 117 to 119, 126 to 131, 139, 140, 156 to 158 and 161 to 163 were conditioned in accordance with 8.3.2. None of the specimens conditioned suffered collapse. Pass
The effects of filter media release were not assessed. Manufacturer to certify. NAs
Model: RB030VM + RB030
The materials used were able to withstand handling and wear during the limited laboratory testing carried out. Ltd
The effect on materials from “in-use” environmental factors could not be evaluated during laboratory tests. Manufacturer to certify regarding such factors. NAs
Specimens 420 to 422 with 520 to 522 were conditioned in accordance with 8.3.1. None of the specimens conditioned suffered mechanical failure or collapse. Pass
Specimens 406 to 410, 439, 440, 446, 447, 456 to 458 with 506 to 510, 539, 540, 546, 547 and 556 to 558 respectively were conditioned in accordance with 8.3.2. None of the specimens conditioned suffered collapse. Pass
The effects of filter media release were not assessed. Manufacturer to certify. NAs
INSPEC Test Report No: 1.18.06.57 Page 8 of 18
7.7 Practical performance
Model: RB030VS or RB030VM + RB030C
Specimen and subject details:
Specimen Subject
248 AH Pass
249 - NAs
No adverse comments were made following testing.
Model: RB030VS or RB030VM + RB030
Specimen and subject details:
Specimen (with ventilator on) Subject
448 + 548 ED Pass
449 + 549 VE Pass
Specimen (with ventilator off) Subject
450 + 550 RW Pass
451 + 551 ED Pass
No adverse comments were made following testing.
7.8 Finish of parts
Model: RB030VM + RB030C
None of the specimens used in laboratory testing showed evidence of sharp edges or burrs. Pass
Model: RB030VM + RB030
None of the specimens used in laboratory testing showed evidence of sharp edges or burrs. Pass
INSPEC Test Report No: 1.18.06.57 Page 9 of 18
7.9.1 Total inward leakage (%)
Model: RB030VS or RB030VM + RB030C
Subject Specimen Cond. Walk
Head side/ side
Head up/down
Talk Walk Mean
KDS 201 A.R. 4.12 3.53 3.33 2.95 3.94 3.58
VE 202 A.R. 1.05 0.94 1.18 1.36 0.96 1.10
KRB 203 A.R. 5.12 8.05 8.22 8.60 7.10 7.42
PBU 251 A.R. 2.12 1.67 1.58 1.81 1.62 1.76
GW 252 A.R. 0.56 0.71 0.59 2.58 0.66 1.02
ED 206 T.C. 1.29 2.02 2.41 1.94 1.94 1.92
CKN 207 T.C. 1.00 1.13 1.20 2.06 1.05 1.29
AH 208 T.C. 1.52 2.06 1.95 2.55 1.83 1.98
JLS 209 T.C. 2.76 4.46 2.94 1.96 2.60 2.94
SAH 210 T.C. 4.63 3.73 4.27 1.25 2.40 3.25
Maximum permitted 25 22
All 50 individual exercise results were not greater than 25%. Pass
All 10 individual wearer arithmetic means were not greater than 22%. Pass
Subjects EM and INH were excluded from the test panel as they were unable to achieve a satisfactory fit.
Subjects KRB, PBU, GW, CKN, AH, JLS, SAH, EM and INH commented that the mask was very small.
All test subjects chose the RB030VM (size medium) mask.
INSPEC Test Report No: 1.18.06.57 Page 10 of 18
Model: RB030VS or RB030VM + RB030
Subj.
Specimen
Cond. Walk Head side/ side
Head up/
down Talk Walk Mean
Mask
Venti-
lator
GW 401 501 on A.R. 0.45 0.76 0.52 9.86 0.81 2.48
VE 402 502 off A.R. 2.21 1.75 1.85 1.81 2.01 1.93
KRB 403 503 on A.R. 3.87 4.20 4.76 6.50 6.02 5.07
PBU 404 504 off A.R. 0.70 0.45 0.48 0.43 0.41 0.49
KDS 405 505 on A.R. 4.58 5.06 4.60 4.19 4.42 4.57
ED 406 506 off T.C. 3.36 3.59 3.54 2.87 2.99 3.27
CKN 407 507 on T.C. 0.63 0.61 0.66 0.92 0.57 0.68
AH 408 508 off T.C. 10.18 10.82 8.57 4.93 6.67 8.24
SAH 409 509 on T.C. 3.41 4.07 6.89 3.39 3.22 4.20
JLS 410 510 off T.C. 1.19 2.63 1.48 3.55 1.86 2.14
Maximum permitted 25 22
All 50 individual exercise results were not greater than 25%. Pass
All 10 individual wearer arithmetic means were not greater than 22%. Pass
All test subjects chose the RB030VM (size medium) mask.
Subject facial dimensions:
Subject
Face Length (mm)
Face Width (mm)
Face Depth (mm)
Mouth Width (mm)
KDS 102 128 98 49
VE 116 132 115 45
KRB 108 130 108 49
PBU 116 141 90 52
GW 117 133 120 53
ED 114 138 100 47
CKN 112 145 114 54
AH 119 113 115 50
JLS 122 151 120 51
SAH 112 141 123 46
EM 122 142 127 51
INH 125 153 95 58
INSPEC Test Report No: 1.18.06.57 Page 11 of 18
7.9.2 Penetration of filter material
Model: RB030VS + RB030C
Sodium chloride: Pass
Specimen Condition
Penetration (%)
After 3 minutes
Max. during exposure
111
A.R.
3.41
112 1.34
113 2.06
120
S.W.
1.53
121 3.29
122 2.77
126
M.S. + T.C.
3.43 3.43
127 3.89 3.89
128 2.59 2.59
Maximum permitted 20
Paraffin oil: Pass
Specimen Condition
Penetration (%)
After 3 minutes
Max. during exposure
114
A.R.
14.66
115 15.79
116 6.77
123
S.W.
10.15
124 11.05
125 7.89
129
M.S. + T.C.
7.89 14.81
130 8.35 13.67
131 12.41 19.37
Maximum permitted 20
7.10 Compatibility with skin
Model: RB030VM + RB030
No problems were encountered during practical performance testing. Pass
No problems were encountered during total inward leakage testing. Pass
The likelihood of materials in contact with the skin causing irritation or other adverse effect on health was not assessed. Manufacturer to certify. NAs
INSPEC Test Report No: 1.18.06.57 Page 12 of 18
7.11 Flammability
Model: RB030VM + RB030C
Specimens 244 and 245 (A.R.) and 246 and 247 (T.C.) were tested. None of the specimens ignited. Pass
Model: RB030
Specimens 544 and 545 (A.R.) plus 546 and 547 (T.C.) were tested. None of the specimens ignited. Ltd
7.12 Carbon dioxide content of the inhalation air
Model: RB030VM + RB030C Pass
Specimen CO2 (%)
235 0.65
236 0.54
237 0.48
Maximum permitted 1.0
Model: RB030VM + RB030 Pass
Specimen (with ventilator off) CO2 (%)
435 + 535 0.51
436 + 536 0.59
437 + 537 0.52
Maximum permitted 1.0
7.13 Head harness
Model: RB030VS or RB030VM + RB030C
The head harness was designed to allow the particle filtering half-mask to be donned and removed easily during limited practical performance and total inward leakage testing. Ltd
The head harness was adjustable and there were no adverse comments regarding security following limited practical performance and total inward leakage testing. Ltd
The product satisfied the total inward leakage requirements. See 7.9.1 for results. Pass
Model: RB030VS or RB030VM + RB030
The head harness was designed to allow the particle filtering half-mask to be donned and removed easily during practical performance and total inward leakage testing. Pass
The head harness was adjustable and there were no adverse comments regarding security following practical performance and total inward leakage testing. Pass
The product satisfied the total inward leakage requirements. See 7.9.1 for results. Pass
INSPEC Test Report No: 1.18.06.57 Page 13 of 18
7.14 Field of vision
Model: RB030VS or RB030VM + RB030C
There were no adverse comments following limited practical performance tests. Ltd
Model: Model: RB030VS or RB030VM + RB030
There were no adverse comments following practical performance tests. Pass
7.15 Exhalation valve
Model: RB030VM + RB030C
There were no observed problems during testing of function in all orientations. See 7.16 for results. Pass
The valve was protected against dirt and mechanical damage by a cap. Pass
The product satisfied leakage requirements. See 7.9 for results. Pass
There were no observed problems when assessing operation after high exhalation flow. See 7.16 for results. Pass
The valve housing withstood 10 N applied for 10 s. Specimens 232 (A.R.), 233 (T.C.) and 234 (M.S.) were tested. Pass
Model: RB030VM + RB030
There were no observed problems during testing of function in all orientations. See 7.16 for results. Pass
The valve was protected against dirt and mechanical damage by a ventilator. Pass
The product satisfied leakage requirements. See 7.9 for results. Pass
There were no observed problems when assessing operation after high exhalation flow. See 7.16 for results. Pass
The valve housing withstood 10 N applied for 10 s. Specimens 432 (A.R.), 456 (T.C.) and 434 (M.S.) were tested. Pass
INSPEC Test Report No: 1.18.06.57 Page 14 of 18
7.16 Breathing resistance
Model: RB030VS + RB030C Pass
Specimen Condition Inhalation resistance (mbar)
Exhalation resistance (mbar)
At 30 l/min At 95 l/min At 160 l/min
111
A.R.
0.46 1.62 1.96
112 0.57 2.03 2.24
113 0.57 1.80 2.19
117
T.C.
0.53 1.85 2.02
118 0.50 1.68 2.03
119 0.59 2.07 2.20
120
S.W.
0.57 1.88 2.14
121 0.57 1.95 2.20
122 0.50 1.63 1.87
138 A.R. + F.C. 0.56 1.89 2.27
139 T.C. + F.C.
0.51 1.77 2.20
140 0.53 1.78 2.23
Maximum permitted 0.6 2.1 3.0
Model: RB030VM + RB030C
Specimen Condition Inhalation resistance (mbar)
Exhalation resistance (mbar)
At 30 l/min At 95 l/min At 160 l/min
211
A.R.
Not requested
NAs
212 NAs
213 NAs
217
T.C.
NAs
218 NAs
219 NAs
220
S.W.
NAs
221 NAs
222 NAs
238 A.R. + F.C. 0.40 1.30 1.65 Pass
239 T.C. + F.C.
0.36 1.20 1.61 Pass
240 0.36 1.26 1.65 Pass
Maximum permitted 0.6 2.1 3.0
INSPEC Test Report No: 1.18.06.57 Page 15 of 18
Model: RB030VS + RB030 Pass
Specimen
Condition
Inhalation resistance (mbar) – ventilator on
Exhalation resistance (mbar) –
ventilator off
Mask
Venti-lator
At 30 l/min At 95 l/min At 160 l/min
311 511
A.R.
0.59 1.94 2.11
312 512 0.51 1.77 2.06
313 513 0.57 1.87 2.12
317 517
T.C.
0.56 1.82 2.11
318 518 0.59 1.97 2.20
319 519 0.57 1.85 2.07
320 520
S.W.
0.59 1.88 2.24
321 521 0.54 1.75 2.22
322 522 0.57 1.71 2.26
338 538 A.R. + F.C. 0.53 1.84 2.10
339 539 T.C. + F.C.
0.51 1.65 1.97
340 540 0.51 1.69 2.10
Maximum permitted 0.6 2.1 3.0
Model: RB030VM + RB030
Specimen
Condition
Inhalation resistance (mbar)
Exhalation resistance (mbar)
Mask
Venti-lator
At 30 l/min At 95 l/min At 160 l/min
411 511
A.R.
Not requested
412 512
413 513
417 517
T.C.
418 518
419 519
420 520
S.W.
421 521
422 522
438 538 A.R. + F.C. 0.41 1.25 1.58
439 539 T.C. + F.C.
0.43 1.39 1.82
440 540 0.41 1.24 1.63
Maximum permitted 0.6 2.1 3.0
INSPEC Test Report No: 1.18.06.57 Page 16 of 18
7.18 Demountable parts
Model: RB030VM + RB030C
The following parts were demountable: the exhalation valve cover was readily connected by hand. Pass
Model: RB030VM + RB030
The following parts were demountable: the micro ventilator was readily connected by hand. Pass
INSPEC Test Report No: 1.18.06.57 Page 17 of 18
Estimates of the uncertainty of measurement
Clause Test Uncertainty
7.4 Packaging Not applicable
7.5 Material See Note 1
7.6 Cleaning and disinfecting Not applicable
7.7 Practical performance See Note 1
7.8 Finish of parts Not applicable
7.9.1 Total inward leakage ± 4.8%
7.9.2 Penetration of filter material - Sodium chloride ± 4.8%
7.9.2 Penetration of filter material - Paraffin oil ± 4.3%
7.10 Compatibility with skin Not applicable
7.11 Flammability See Note 1
7.12 CO2 content of the inhalation air ± 8.0%
7.13 Head harness Not applicable
7.14 Field of vision See Note 1
7.15 Exhalation valve(s) See Note 1
7.16 Breathing resistance ± 4.9%
7.17.2 Breathing resistance after clogging ± 7.7%
7.17.3 Filter penetration after clogging - Sodium chloride ± 4.8%
7.17.3 Filter penetration after clogging - Paraffin oil ± 4.3%
7.18 Demountable parts Not applicable
Note 1 The acceptance criterion for this test is a straightforward “Pass/Fail”, rather than a
numerical value. Consequently, as there is no value to be reported, uncertainty has not been reported either.
Note 2 The uncertainty value is based on a standard uncertainty multiplied by a coverage factor
k = 2, which provides for a confidence level of approximately 95%. Values expressed as a percentage (%) are relative.
Note 3 It should be noted that the above values have not been taken into account when making
assessment to the pass/fail criteria.
INSPEC Test Report No: 1.18.06.57 Page 18 of 18
ANNEX
This Annex comprises one section.
1. Photographs of the products tested. (5 pages)
END OF REPORT
Reckitt Benckiser Healthcare (UK) Limited Filtering half mask, model: RB030VS (size S) fitted with valve cap RB030C
INSPEC Testing Services’ specimen number 1F0170 160 29 June 2018
Reckitt Benckiser Healthcare (UK) Limited Filtering half mask, model RB030VM (size M) fitted with valve cap RB030C
INSPEC Testing Services’ specimen number 1F0170 260 29 June 2018
Reckitt Benckiser Healthcare (UK) Limited Filtering half mask, model RB030VS (size S) fitted with ventilator RB030
INSPEC Testing Services’ specimen number 1F0170 360 and 560 28 June 2018
Reckitt Benckiser Healthcare (UK) Limited filtering half mask, model: RB030VM (size M) fitted with ventilator RB030
INSPEC Testing Services’ specimen number 1F0170 460 and 560 28 June 2018
Reckitt Benckiser Healthcare (UK) Limited Ventilator: RB030 Siti Shield
INSPEC Testing Services’ specimen number 1F0170 560 27 April 2018