10/14/2004 SPIN2004 Meeting 1

The MIT-Bates Compton Polarimeterfor the South Hall Ring

W.A. Franklin for the BLAST Collaboration

SPIN2004 Conference Trieste, Italy

October 14, 2004

1. Physics Motivation2. Experimental Apparatus3. Results

Page 2: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 2

BLAST

Collaborating InstitutionsMIT, UNH, Arizona St., Duke, Dartmouth, Vrije University, U.S.

Naval Acad., Ohio, Boston U.

12 Ph.D. students• Neutron Electric Form Factor

V. Ziskin, 10/15 Session 5• e-p Asymmetries on

DeuteriumA. Maschinot, 10/15 Session 5

• Electric-Magnetic Form Factor Ratio of HydrogenC. Crawford, 10/15 Session 5

• Research program in South Hall Ring using Bates Large Acceptance Spectrometer Toroid • Study comprehensively nucleons and light nuclei at low Q2

Experiments underway 2003, running throughout 2004

Page 3: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 3

BLAST Experiment

• South Hall Ring: Intense (175 mA) stored CW polarized electron beams in at 850 MeV

• BLAST Atomic Beam Source: (E. Tsentalovich, 10/15 Session 8)

• BLAST: Symmetric detector with wide momentum transfer bite

• Beam-target polarization product from BLAST asymmetry.• Need rapid nondestructive measurement of beam polarization. • Laser backscattering can provide.

Measure asymmetries using polarized beams and targets

Page 4: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 4

Compton Polarimetry Overview• Compton scattering in highly relativistic frame

Angular distribution compressed into narrow kinematic cone Photon frequencies shifted into gamma regime Detect backscattered photons with compact detector

• Compton scattering cross section • Well known theoretically• Term dependent on electron spin and laser helicity Can extract e- polarization by measuring asymmetries in scattering rates for circularly polarized laser light

Transverse pol. yields asymmetry in azimuthal distribution of scattered photons.Longitudinal pol. yields asymmetry in scattered photon energy spectrum

Page 5: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 5

Compton Polarimetry Below 1 GeV

• Bates seeks precise polarization measurement for each ring fill (15 minutes) for experiments with BLAST.

Bates

JLab

HERA532 nm laser light

• Compton polarimetry is well established at high energy accelerators (Apol~0.5)

• Different challenges exist in applying at energies below 1 GeV. > Analyzing power falling with energy (Apol < 0.05)> Interaction mechanism varies with gamma ray energy> Broader angular distribution for photons> Background from low energy photons> Beam lifetime less than 1 hour

Electron Energy (MeV)

Compton Analyzing Power

Ap

ol

Page 6: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 6

CsI detector

Laser hut

Remotely controlledmirrors

Electron beam

InteractionRegion

Scatteredphotons

SHR Injection Line

RingDipole

BLAST

Laser exit

Laser line

• Design Considerations• Based on NIKHEF Compton Polarimeter• Located upstream of BLAST target to reduce background • Measures longitudinal projection of electron pol.• Scattered gamma trajectory defined by electron momentum

• Polarimeter Layout• Laser in shielded hut with 18 m flight path • Interaction with electron beam in 4 m straight section• Laser mirrors moved remotely • CsI calorimeter 10 m from IR

SHR Compton Polarimeter

Page 7: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 7

Laser System• Laser

• Solid-state continuous-wave, very stable • 5W output at 532 nm

• Optical Transport•Simple, robust lens arrangement for transport to IR and focusing• Mechanical chopper wheel allowing background measurements • Circular polarization by Pockels Cell for rapid helicity reversal• Phase-compensated mirror arrangement

• Interaction Region• 4 degrees of freedom for laser scans. • Laser intercepts stored beam at < 2 mrad• EPICS Control System for slow controls

Y a

ngle

X angle

Page 8: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 8

Scattered Gamma Ray Line

• Align electron beam first to align using bremsstrahlung background. • Movable collimators used to eliminate background from beam halo• Sweep magnet, veto scintillator reject charged particles• Scintillator hodoscope provides position information for beam alignment• Pure CsI calorimeter offers resolution and speed for single photon mode • High rate PMT bases for linear response, stable gain

• Variable thickness stainless steel absorbers for high intensity operation

Page 9: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 9

High Intensity Operation• Measurements made up to 190 mA.

Stainless steel absorbers act as neutral density filter to control rate.

• Signal-to-background tractable at high currents (beam size increases)

• Energy calibration stable on short time scale for high rates in CsI

• Small systematic correction to asymmetry for absorber thickness

• Polarization reduction sometimes observed in raising beam current (tune spreading). • Rapid feedback for retuning Ring.

Page 10: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 10

Data Acquisition

• VME-based system• Rapid digitization

– 100 MHz, 12-bit buffered ADC– External triggering – Single event mode

• High readout and sorting speed

– DMA for high CW event rates (> 100 kHz)

– Pulse shape discrimination and pile-up rejection

– Rapid spin sorting capability

• Integration

– Linked with BLAST analysis

– Synch with BLAST event stream

– Include EPICS information

Fast and reliable data acquisition system is very important

Developed locally (T. Akdogan)

Page 11: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 11

• Analysis begins with raw ADC spectra

– L1 - Laser on, right-handed Pcirc

– L2 - Laser on, left-handed Pcirc

– B1 - Laser off, right-handed Pcirc

– B2 - Laser off, left-handed Pcirc

• Establish energy calibration based on Compton edge and ADC pedestal• Normalization for background subtraction from bremsstrahlung tail• Asymmetry between laser helicities formed as function of gamma energy • Fit asymmetry data with function representing polarimeter analyzing power(GEANT simulation)

Data Analysis

Page 12: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 12

Fill-by-Fill Polarization Results

• Polarization reversed in electron source on fill-by-fill basis• Polarization monitored continuously• Typical precision of 4-5% for ~15 minute fill• Gaussian profile to results

Time (hours)

Pola

riza

tion

Page 13: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 13

Cumulative Results

• Database of results for BLAST experiment in blocks of ~4 hrs• Polarization stable within few percent as a function of time. • Changes usually correlated with beam properties.• Mean polarization (2004): 0.663, (July-Sep, 2004): 0.654• Long term errors determined by systematics

Page 14: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 14

Consistency Checks

• Expected sinusoidal dependence on injection angle• Avg. Injection polarization 0.71 +/- 0.04 (20 MeV transmission pol.).• Measure zero asymmetries with unpolarized beam• Consistency of results for two helicity states

Wien Filter Voltage

Sto

red

Pola

riza

tion

Page 15: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 15

Spin Flipping in South Hall Ring

• Spin flipper

• Reverse direction of beam pol. while stored

• Separate instrumental asymmetries from polarization differences from source

• Adiabatic spin flip

• Rf magnetic field

• Ramp through resonant frequency.

• Performance in South Hall Ring

• Achieved efficiency > 98% (Michigan Spin Physics Group)

• < 1 sec to induce spin flip.

• Spin flip period of ~5 min.

Resonance Scan

frf (kHz)

Pfin

al

Page 16: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 16

Automated Spin Flip

• Spin flipper controlled through Compton DAQ for time synchronization. • Execute flip based on time or beam current • Polarization data sorted from beginning of fill to flip (dark) and post-flip (open)• Eliminates false asymmetry due to geometric effects• Verifies equality of beam polarization for two helicities at injection to < .01

Pola

riza

tion

Time (hrs)

Page 17: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 17

Systematic Error Estimation

•Total systematic error in avg. beam pol. estimated at 0.04. •Working to reduce, sufficient for BLAST experiments.

Small analyzing power makes systematic error reduction crucial

Error Contribution DP

Modeling of Apol and energy calibration

0.03

Pile-up 0.01

Beam misalignment and PITA

~0.05 single pol. state, < 0.01 average pol

Laser circular polarization 0.005

Spin precession uncertainty 0.003

Page 18: 10/14/2004SPIN2004 Meeting1 The MIT-Bates Compton Polarimeter for the South Hall Ring W.A. Franklin for the BLAST Collaboration SPIN2004 Conference Trieste,

10/14/2004 SPIN2004 Meeting 18

Summary

• MIT-Bates operates a laser backscattering Compton polarimeter during internal target experiments with BLAST at 850 MeV.

• High intensity operation has been successful with stored currents exceeding 175 mA.

• Polarization results are generally stable with time. Certain changes in beam conditions can produce depolarization.

• Systematic error control and calibration uncertainty meet level needed for BLAST experiments. Use of consistency checks provides reduction of systematic errors.