G.A.S. Proposal#: 2013-326
Page 3 of 5
1.0 Technical 5100HD
Ametek Model 5100HD Oxygen Analyzer for the measurement of O2 in a VRU.
The 5100HD is a Non-Contact Oxygen Analyzer that uses Tunable Diode Laser Absorption
Spectrocopy (TDLAS) Technology.
The analyzer features:
• All digital signal processing – 32 bit microcontroller capable of sophisticated signal processing
• Web-Based interface - Modbus, Ethernet and analog connectivity
• Real-Time performance Monitoring – Laser line-lock verification using internal reference cell
• NEMA 4 Enclosure houses the electronic components
• Horizontal oriented cell
• Fully integrated Sample Conditioning System
• Fast response to moisture changes.
• Resistant to Contamination – No interference from gas phase amines, glycol, methanol, H2S and
mercaptans
• Proven long term reliability.
• 4 – dry relay contacts
• One – Isolated 4-20 mA loop-powered outputs
• RS-485 serial port, isolated – supports Modicon Modbus RTU
• Display – 4 line x 20 character alphanumeric VF display
• Class I, Division 2
SPECIFICATIONS FOR MODEL 5100HD
Range: 0-5% O2
Accuracy: Typically 2% of range
Repeatability: Typically 2% of range
Sample Cell Pressure: 5 BAR A max
Ambient Temperature: -20 to 50 C (-4 F to 122 F)
Gas Flow: 1 – 10 SLPM recommended
Outputs: Analog: Isolated 4-20 mA analog output.
Digital Outputs: RS-485 serial port, Ethernet
4 – Dry relay contacts. 30 VAC, 60VDC, 100VA resistive
Laser Specification: Class 1m
Power Requirements: 120 VAC, 60 Hz, < 25W
Hazardous Area Classification: Class I, Div.2, Groups A,B,C & D
Advantages to using TDL (Tunable Diode Laser) Technology – Ametek 5100HD
• Real-Time performance Monitoring – Laser line-lock verification using internal reference cell
• Analyzer has heated bench, helps in maintaining sample above dewpoint of stream
• Low maintenance, does not require weekly/monthly/yearly calibration
• Resistant to Contamination – No interference from gas phase amines, glycol, methanol, and mercaptans
• Low maintenance – in the event of process upset, simply remove the cap on the cell and clean the mirror and cell.
G.A.S. Proposal#: 2013-326
Page 4 of 5
• 5.0 Technical O’Brien Tracepak
• Preinsulated sample tube bundle with TPU Jacket
• tba - length
• (1) ¼” x 0.035" wall, 316/316L SS seamless tube
• 8 watt/ft BTV low temperature self regulating electric tracer
FM approved, CSA certified Class 1, Div.1 @ 120/240 volts.
• T-rating T6 185F/85C
• Maximum tracer continuous exposure temperature is 150F/65C with power on
• Maximum tracer intermittent exposure is 185F/85C for 1000 hrs. with power on or off.
• Confirm circuit lengths prior to purchase
• The bundle can be cut to length in the field
Notes:
• TPU is a chloride free thermoplastic polyether urethane elastomer
• Bundle is designed for permanent installation