1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Embed Size (px)

Citation preview

Page 1: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

1

Page 2: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

2

Page 3: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

3

Page 4: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

4

Page 5: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356-456 Micro-Moist

Moisture Content in Coal is an important parameter:

Preparation-Regulate Moisture Content

Handling-Dust-Ice

Transport-Weight

Trading-Weight

Calorific Value

5

Page 6: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356-456 Micro-Moist

• Microwave Transmission Techniques are the most accurate method of determining moisture content of Coal

• Attenuation/Area Weight

• Phase Shift/Area Weight

• Phase Shift and or/Attenuation

6

Page 7: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356-456 Micro-Moist

Attenuation measurementcan provide accuracy up to 1wt% moisture

Temperature dependence error

Bulk Density dependence error

Particle size dependent error

Conductivity dependent error

7

Page 8: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 456 Micro-Moist

Phase Shift Measurement:or (Combined Attenuation and Phase Shift)

Independent of Temperature of Coal

Independent of Particle Size variations

INTEGRAL Area Weight Compensation

Accuracy of 0.2 wt% moisture achievable

8

Page 9: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Principle of Operation

Microwave Measuring System, LB 356-456 Micro-Moist

9

Page 10: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 354-356-456 Micro-MoistThe choice of 2.7-3.4 GHz frequency range:

• Measurement of bulk, not surface. This provides a representative measurement

• The temperature coefficient of free water crosses zero at 3 GHz

• Attenuation due to ionic conductivity (salt content,…) is reduced at higher frequencies

• Frequency analysis over the frequency range provides phase measurement greater than 360 deg with plausibility control

10

Page 11: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356 Micro-Moist

11

Page 12: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356 Micro-Moist Coal Preparation Plant:

12

Page 13: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 356 Micro-Moist

Coal Preparation Plant management reported precision of 0.4% moisture (one sigma RMS)

13

Page 14: 1. 2 3 4 Microwave Measuring System, LB 356-456 Micro-Moist Moisture Content in Coal is an important parameter: Preparation-Regulate Moisture Content

Microwave Measuring System, LB 456 Micro-Moist

Product Improvements:

• Reference cable can be replaced by short link if To constant

• PT-100 Temperature Compensation with direct connection

• Detector to Evaluation Unit uses FSK (Frequency Shift Keying)

14