Upload
ajoy-chanda
View
220
Download
0
Embed Size (px)
Citation preview
8/6/2019 05318277
1/6
IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 37, NO. 12, DECEMBER 2009 2347
Inactivation Characteristics of Bacteria inCapacitively Coupled Argon Plasma
Sureshkumar and Sudarsan Neogi
AbstractPlasma technology is being focused on the medical,food, and pharmaceutical fields for sterilization applications. Thesterilizing effect of the 13.56-MHz radio-frequency (RF) plasmagenerated by using argon gas was studied using Staphylococcusaureus, one of the most common pathogens liable to hospital-acquired infections. The major focus of this paper was to performa parametric study by varying the external-process parameterssuch as plasma treatment time, RF power, gas-flow rate, and pres-sure on the inactivation ofS. aureus. The results were supported byoptical emission spectroscopy and scanning electron microscopystudies.
Index TermsBiomedical engineering, plasma applications,sterilization, surface contamination, surface treatment.
I. INTRODUCTION
THE staphylococcus aureus is one of the most important
pathogens responsible for hospital-acquired infections.
Any sterilization process must ensure the complete removal of
microorganisms relevant to the targeted application and must
leave the material with no or minimal damage. Sterilization is
the complete elimination of all types of microorganisms [1].
Sterilization can be achieved by heat, chemicals like ethylene
oxide, radiation, pulsed light, and plasma. Each sterilization
method has its own limitations. For instance, heat (either dryheat or steam) sterilization cannot be used to sterilize heat-
sensitive materials since deformation of such materials may be
expected at high temperatures. Chemicals like ethylene oxide
can sterilize materials at low temperatures. However, the prod-
ucts sterilized by EtO sterilization should be aerated for a longer
time before use [2]. Since EtO is flammable and carcinogenic,
adequate safety measures are needed. Radiation sterilization is
commercially used nowadays for sterilizing single-use devices.
It needs an isolated site of operation, and proper safety mea-
sures should be taken while working with gamma radiation. It
has been reported that certain polymers like PTFE, polyacetals,
etc., are not compatible with gamma radiation [3]. Plasma
sterilization has been in focus in recent years because of itsremarkable advantage of ambient-temperature operation, high
efficiency, shorter treatment times, and nontoxicity. Contrary
to EtO, sterilization by using plasma is considered to be safe,
both for the operator and the patient [1]. Sterilization by plasma
Manuscript received August 4, 2009; revised September 15, 2009. Currentversion published December 11, 2009. This work was supported by theDepartment of Science and Technology, India (DST Grant SR/S3/CE/31/2005).
The authors are with the Department of Chemical Engineering, IndianInstitute of Technology Kharagpur, Kharagpur 721 302, India (e-mail: [email protected]).
Digital Object Identifier 10.1109/TPS.2009.2033112
can be achieved at atmospheric pressure or in vacuum. At the
present condition, the sterilization technology is still in an iso-
lated and scattered experimental stage [4]. There are also other
sterilization methods which use high electric field [5], high-
voltage pulse [6], and pulsed-light irradiation [7]. However, in
the aforementioned methods, complete sterilization is achieved
only at higher temperature. Plasma is a partially ionized, low-
pressure gas that is composed of ions, electrons, and UV
photons as well as the reactive neutral species (free radicals
and excited atoms) with sufficient energy to break covalent
bonds and initiate various chemical reactions [2]. Low-pressureplasmas are generated in the pressure range of 1100 Pa
(133 Pa = 1 torr) [8]. For any application in plasma, it isnecessary to have a basic understanding of the experimental
system used by playing with various process parameters.
In the current investigation, the sterilizing effect was studied
by varying the plasma parameters using a model organism
(S. aureus) in a low-pressure plasma reactor. The effects of
treatment time, radio-frequency (RF) power, gas-flow rate, and
gas pressure were studied using argon plasma. Similar para-
metric studies were reported using different configurations of
plasma systems such as Escherichia Coli with remote glow-
discharge air plasma [9], Bacillus subtilis with low-pressureoxygen-based plasmas in distributed electron cyclotron reso-
nance reactor [10] and Escherichia Coli with remote argon
plasma [11].
II. EXPERIMENTAL SETUP
A. Plasma Reactor
A capacitively coupled plasma reactor was used for this
study, as shown in Fig. 1. The stainless-steel reactor consists
of two parallel plates fixed inside a cylindrical chamber. The
reactor was connected to an RF generator, which generates
plasma in the RF range of 13.56 MHz, through a matchingnetwork. Gas-flow rate was controlled by means of a mass-flow
controller. The temperature inside the sterilization chamber was
measured by means of a thermocouple, and gas pressure was
monitored by means of a capacitance gauge.
The external parameters such as gas pressure, gas-flow rate,
and time of exposure to plasma for each experimental run
can be prefixed using control systems. Initially, the reactor
was pumped down to a pressure of about 0.76 mtorr using
a Roots and rotary-pump cascaded arrangement. Then, argon
gas was allowed to enter the chamber through a mass-flow
controller. Once the pressure inside the sterilization chamber
was stabilized, the RF plasma was ignited.
0093-3813/$26.00 2009 IEEE
8/6/2019 05318277
2/6
2348 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 37, NO. 12, DECEMBER 2009
Fig. 1. Reactor setup used for sterilization study.
TABLE IOPERATING CONDITIONS OF THE PLASMA REACTOR
B. Plasma Treatment
The S. aureus (NCIM 2079) chosen for the sterilization study
was procured from NCIM, Pune, India. The bacterial strain
was maintained on standard-method agar slants at 56 C. The
culture was grown on a sterile nutrient broth, containing 5 g/l of
peptic digest of animal tissue, 5 g/l of sodium chloride, 1.5 g/lof beef extract, and 1.5 g/l of yeast extract incubated in a
rotary shaker operating at 175 r/min and maintained at 37 C.
For the sterilization studies, the cells were harvested from the
exponential phase for plasma treatment. A drop volume of 50 l
of the bacterial suspension was aseptically applied over the
sterile glass slides of dimension 25 mm 25 mm 1 mm
and left in the laminar hood for drying. The samples (slides
containing dried bacteria) were placed inside the sterilization
chamber for plasma treatment. Purple, the characteristic color
of argon, was observed inside the reactor. After plasma treat-
ment, the chamber was vented to the atmosphere, and the
samples were collected. The bacterial samples were treatedfor different parameters as mentioned in Table I. One of the
samples was not exposed to plasma and kept as a control
sample. The treated samples were collected and serially diluted
using sterile 0.02 M phosphate buffer at pH 7. A drop volume
of 50 l of the diluted suspension from each test tube was
plated on the sterile nutrient agar plates using standard agar
spread-plate procedure. The plates were observed for bacterial
growth after incubation at 37 C for 24 h. After incubation,
S. aureus colonies were observed over the surface of the agar
plates, which were confirmed by gram staining.
The number of bacterial colonies formed over the agar plates
was counted visually. A semilogarithmic graph was plotted
between the number of colonies remaining after treatment andthe plasma treatment time, which is called the survivor curve, to
determine the D-value. D-value is defined as the time required
to reduce a bacterial population by 90%. D-value is obtained
from the inverse of the slope of the survivor curve [12].
To determine the effect of individual plasma parameters,
germicidal effect (GE) was calculated. GE is another way of
expressing or validating the efficiency of a sterilizing process.GE is defined as the difference between the logarithm of the
original bacterial population (No) and number of survivors(Nt) remaining after plasma treatment. It is expressed as GE =logNo logNt [13].
C. OES
Optical Emission Spectroscopy (OES) is a noninvasive
technique for the diagnosis of active species present in the
plasma. The light emitted by plasma was captured by means
of a high-resolution optical emission spectrometer (HR4000
CG-UV-NIR) from Ocean Optics Inc., U.S. The spectral res-
olution was 1 nm, and the hit interval was 0.1 nm.
D. SEM
Scanning electron microscopy (SEM) studies were car-
ried out in a scanning electron microscope (JEOL-JSM-5800,
Japan) to study the surface morphology of bacteria before and
after plasma treatment. Prior to analysis, the samples were
coated with a thin layer of gold.
III. RESULTS AND DISCUSSION
A. Effect of Plasma Treatment Time
The results for the effect of treatment time are shown in
Fig. 2. It is observed that by increasing the exposure time to
plasma, the number of colonies reduced. However, since the
only candidate for sterilization is argon plasma, the bacterial
reduction was less. This shows the effect of argon plasma on
S. aureus with respect to treatment time. The applied power
was maintained at 100 W for all treatment times. The results for
the bacteria reduction are plotted in a survivor curve as shown
in Fig. 2(a). A two-phase nature of the survivor curve was ob-
served. The first phase (D1 = 0.53 min) is due to the killing ofbacteria at the top surface of the bacterial load by UV radiation
from the argon plasma. This phase has a smaller D-value, as
already reported. It is known that UV rays in the wavelength of220280 nm are readily absorbed by DNA and, thereby, affects
8/6/2019 05318277
3/6
SURESHKUMAR AND NEOGI: INACTIVATION CHARACTERISTICS OF BACTERIA IN CAPACITIVELY COUPLED ARGON PLASMA 2349
Fig. 2. Effect of plasma treatment time (100 W, 40 sccm, 150 mtorr).(a) Survivor curve. (b) GE versus treatment time. (c) Comparison betweenexperimental and calculated data.
the replication or reproduction of the cell. On the other hand, the
second phase (D2 = 3.61 min) is due to the effect of the UVradiation and argon free radicals over stacked species, which
was revealed by a larger D-value. However, if the cells are
stacked, the cells or cell debris (formed as a result of Phase I)
lying over the top surface hinders the cells at the subsurface
from UV exposure [1]. Hence, a marginal bacterial reduction
was achieved. The GE was also plotted as a function of plasmatreatment time as shown in Fig. 2(b). Using the survivor curve,
Fig. 3. Effect of input RF power (5 min, 40 sccm, 150 mtorr).
a quantitative relationship was developed between the number
of survivors (CFUs/ml) remaining after plasma treatment and
plasma treatment time (t) which is represented by the followingpolynomial equation:
log(CFUs/ml) = 0.019 t4 0.255 t3
+ 1.261 t2 2.908 t + 8.66. (1)
The plot for (1) along with the experimental data is shown in
Fig. 2(c).
B. Effect of RF Power
The effect of input RF power on the sterilization efficiency is
shown in Fig. 3. The total energy input to the gas is increased byincreasing the input power. Hence, it is obvious that increasing
the input RF power increases the electron density by increasing
the number of active species in the plasma [14]. With pure argon
plasma, a significant bacterial reduction can be achieved by
increasing the power from 20 to 100 W. This shows that the
GE strongly depends on input RF power. At 20 W, the GE was
1.69, whereas it exceeded 3.0 while applying the input power
of 100 W for the same operating conditions. However, while
operating at higher power, the effect of temperature should
be also accounted. In this case, the temperature at the end of
treatment was observed to be 40 C.
C. Effect of Gas-Flow Rate on Sterilization
As can be observed from Fig. 4, the GE varied between
1.5 and 3.1 while varying the argon gas-flow rate from 20 to
100 sccm (standard cubic centimeters per minute). At higher
flow rates, the bacterial reduction (GE) was much weaker than
at lower flow rates. This effect has been studied by several re-
searchers [9], [10]. Increasing the flow rate reduces the average
residence time of molecules inside the sterilization chamber.
At a given constant power, only a given or possible number
of molecules gets ionized and results in plasma formation. In
the current experimental setup, a significant difference was ob-
served by varying the argon-gas-flow rate from 20 to 100 sccm.It can be concluded that the lower flow rates showed relatively
8/6/2019 05318277
4/6
2350 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 37, NO. 12, DECEMBER 2009
Fig. 4. Effect of gas-flow rate (100 W, 5 min, 150 mtorr).
better bacterial reduction than the higher flow rates [15]. Ifthe gas-flow rate is too low (as in the case of 20 sccm), the
number of reactive species generated might be less enough
for efficient sterilization as reflected from lower GE values.
Hence, an optimum flow rate should be selected for efficient
sterilization. The sterilizing effect improves until a saturation
point is reached for the applied power, i.e., the maximum
number of argon atoms is getting ionized at this point for an
applied power of 100 W and, thereby, resulted in a higher GE
value is at 40 sccm. However, the average energy of the species
in the plasma and the residence time is lower at higher flow
rates than at low flow rates [11]. This might be the reason for
greater GE at low flow rates, and after reaching a peak value
at a particular flow rate, it reduces. In the present experimentalconditions, better sterilization was achieved when the gas-flow
rate was maintained at 40 sccm.
D. Effect of Gas Pressure
It is known that an increase in pressure reduces the gas mean
free path. As the atoms get more crowded inside the chamber,
the collision rate increased, and there is a slight increase in
plasma density, i.e., the number of active species per unit
volume. In this experimental setup, the working pressure for
RF plasma generation can be varied from 100 to 200 mtorr. It
is obvious from Fig. 5 that there is no significant difference inbacterial reduction with the variation of pressure in the given
range. Since, at higher pressures, the mean free path of the
gas and electrons reduced; collisions among them increase to
generate active species in plasma. However, the average energy
of the active species becomes less due to frequent collisions
inside the sterilization chamber, which in turn, decreases the
sterilization efficiency [16]. This effect resulted in relatively
lower GE value at 200 mtorr.
E. OES
To realize the species present in argon plasma, OES studieswere performed. The optical emission spectrum generated by
Fig. 5. Effect of gas pressure (100 W, 5 min, 40 sccm).
argon plasma is shown in Fig. 6. It is understood from Fig. 6
that the neutral and atomic species of argon were present in
the plasma. The presence of UV radiation, which is consid-
ered to be the dominant species for pure argon plasma, was
also observed in the range of 308.10 nm. OES studies were
discussed with each variable plasma parameter as given next.
Neutral argon emission lines were observed at 601.75, 800.62,
912.30, 922.45, and 965.78 nm. The presence of argon ions
were indicated by peaks observed at 686.13, 771.72, 794.05,
841.57, and 922.76 nm.
1) Effect of Treatment Time: By increasing the plasma treat-
ment time, a marginal change in the plasma spectrum wasobserved. As mentioned earlier, an emission band was observed
near the 300310-nm range indicating the presence of UV
radiation. Similarly, the respective peaks for argon atoms and
ions (both single and multiple) were also observed.
2) Effect of Power: Increasing the input power increased
the number of active species present in plasma due to the
increase in energy input. This effect was well realized by OES
studies. For instance, the intensities of the argon neutral (Ar I
601.75) atoms and ion (Ar II 736.54 nm) peaks were found to
be increasing during an increase in the input RF power. This
showed the presence of large concentration of plasma species
at higher powers.3) Effect of Gas-Flow Rate: Since the flow rate of argon is
increased, there is a decrease in plasma density by flushing out
the plasma. Fig. 7(b) shows the slight increase in the intensity
of the argon neutral (601.75 nm) and ion (811.09 nm) peaks.
Since the power and the gas pressure are fixed for the constant
reactor volume, an optimum number of atoms are ionized
within the given treatment time at a given flow rate for bacterial
sterilization.
4) Effect of Pressure: Increasing the gas pressure inside the
sterilization chamber reduces the mean free path of electrons.
This, in turn, increases the number of collisions between atoms
and electrons; however, the average energy of these plasma
species gets reduced. A decrease in the intensity of the peakswas observed as the gas pressure is increased, as shown in
8/6/2019 05318277
5/6
SURESHKUMAR AND NEOGI: INACTIVATION CHARACTERISTICS OF BACTERIA IN CAPACITIVELY COUPLED ARGON PLASMA 2351
Fig. 6. Optical emission spectrograph of argon plasma.
Fig. 7. (a) Argon species intensity variation with power. (b) Argon speciesintensity variation with gas-flow rate. (c) Argon species intensity variation withgas pressure.
Fig. 7(c). For instance, by increasing the pressure from 100 to
200 mtorr, the intensity of the neutral argon peak at 601.75 nmreduced from 1962 to 1535 arbitrary units.
Fig. 8. SEM images. (a) Control S. aureus cells. (b) After 5 of min argonplasma treatment.
F. SEM
SEM studies were carried out for control and treated bacterial
cells. The images captured by scanning electron microscopeare shown in Fig. 8(a) and (b). The treated cells shown here
8/6/2019 05318277
6/6