    Model 2350/2350A GC Controller Modbus registers that may be ofinterest to the developer are of two varieties, SIM_2251 and

    User_Modbus. Differences are summarized in the following table.

    Section G.2 lists only variables included in the User_Modbus BooleanModbus registers. These registers are not user-defined and primarilycontain alarm flags, possibly useful for debugging purposes. To use the

    Modbus Test software to view the contents of these registers, you willneed to set the  Function parameter to “1 (Read Coil)”. See Section 9.2 through Section 9.4 for details on using the Modbus Test software.

    Table G-1 Comparison of SIM_2251 and User_Modbus

    SIM_2251 User_Modbus

    Serial slave port. Serial slave port.

    Modified protocol that allows floating pointnumbers to be transmitted over Modbus via

    2251 emulation slave type.

    The standard Gould Modbus protocol thataccommodates PLC Emulation LO-HI.

    Nearly all register contents are predefined;

    a few 9000-series registers can be user-

    defined (i.e., read-write).

    Boolean (coils) are predefined. Numeric

    (registers) are user-defined.

    Variables assigned to registers can be listed

    in contents of a PC-Config Report (for

    instructions, see Section 2.9.4; for an

    example report, see Appendix A,). For more

    detail about register contents, see

    Section G.2.1.

    Variables assigned to registers can be listed

    in contents of a PC-Config Report (for

    instructions, see Section 2.9.4; for an

    example report, see Appendix A,). Boolean

    (coils) contents also listed in Section G.2.

    When using the Modbus Test software, set

     Register  Mode to “DANIEL” to view register

    contents (see Section 9.2.3).

    When using the Modbus Test software, set

     Register  Mode to “PLC LH” to view register

    contents (see Section 9.2.3).

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    2/22

     All other User_Modbus registers can be defined by the user. To defineUser_Modbus register contents (through assignment of variables), see

    Section 5.14.3.

    To obtain a complete list of register assignments, both SIM_2251 and

    User_Modbus, use MON to produce a PC-Config Report.

    To use the Modbus Test software and view the contents of SIM_2251registers, you will need to set Register Mode to “Daniel”, as noted in Table

    G-1.

    To obtain a complete list of register assignments, both SIM_2251 and

    User_Modbus, use MON to produce a PC-Config Report.

     

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    3/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    4/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    5/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    6/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    7/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    8/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    9/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    10/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    11/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    12/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    13/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    14/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    15/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    16/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    17/22

    To make a warranty claim, you, the Purchaser, must:

    1. Provide Daniel with proof of the Date of Purchase and proof of the

    Date of Shipment of the product in question.

    2. Return the product to Daniel within twelve (12) months of the date of original shipment of the product, or within eighteen (18) months of thedate of original shipment of the product to destinations outside of the

    United States. The Purchaser must prepay any shipping charges. Inaddition, the Purchaser is responsible for insuring any productshipped for return, and assumes the risk of loss of the product during

    shipment.

    3. To obtain Warranty service or to locate the nearest Daniel office,sales, or service center call (713) 827-6314, Fax (713) 827-6312, orcontact:

    4. When contacting Daniel for product service, the purchaser is asked toprovide information as indicated on the following page entitled

    "Customer Problem Report".

    5. For product returns from locations outside the United States, it will benecessary for you to obtain the import consignment address so thatDaniel's customs broker can handle the importation with the U.S.

    Customs Service.

    6. Daniel Measurement Services offers both on call and contractmaintenance service designed to afford single source responsibility for

    all its products.

    7. Daniel reserves the right to make changes at any time to any productto improve its design and to insure the best available product.

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    18/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    19/22

     

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    20/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    21/22

  • 8/17/2019 00522-G Mon 2000 Software Manual Sec_G 1MB.pdf

    22/22