Digi-Pas™ A division of Group of Companies DIGITAL POST LEVEL DWL1000 XY ISO 14001:2000 Cet No.: TWO4/00450EM ISO 9001:2000 Cet No.: QSP00050 OHSAS 18001:2007 Cet No.: SG08/50066 Tested to Comply with FCC Standards Instruction Manual Congratulations on your choice on this Digi-PasDigital Level DWL-1000 We suggest you read this instruction manual thoroughly before using the digital post level. Save thisinstruction manual for future reference. Roof/drainage Leveling Roof/drainage Leveling 1. Kit Contents Description Quantity Digital Post Level Device “AAA” Alkaline Batteries Aluminium Jig 1 2 1 2. Technical Specification Accuracy Post Level Measuring Range Digital Level Angle Measuring Range ± 0.10° 0° ~ ± 45° 0° ~ ± 90° Resolution ± 0.05° Operating Temperature Power Supply Estimated Battery Life Dimension Weight 3. Operation Instruction Battery Installation Battery Compartment 1. Open the battery cover, located as vertcial side plastic (See Figure 1). 2. Insert 2 new “AAA” alkaline batteries with polarity (+/-) as indicated on the battery compartment. 3. Close the battery door securely in place. Fig 1 IMPORTANT NOTICE: It is the responsibility of the user to verify the calibration of the post level before each use. Display LCD screen with Cross hair Vial Power ON/OFF Power on and off the digital post level by pressing this key Mode Key To Switch the resolution reading R1: 0.05 R2: .01 R3: 0.5 % Back Light Key Turn on the back light of LCD for brighter LCD 4. Using the Product HOLD Key a) Press HOLD , the digital reading blink. The angle is in the memory. b) Press HOLD again to release the function. CAL Key Periodic checking should be conducted to this device. If it is out of alignment, recalibration will be necessary. This key is used for calibration to absolute horizontal and absolute vertical of the digital post level. See “Calibration Procedures” in section 7. 6. Special Features and Functions Buzzer Function Hold the “Mode” button for 4 sec to switch the buzzer on and off. Automated Shut-Off This digital post level will shut-off automatically after 10 minutes. Low-power indication When voltage is low, battery symbol will be shown on the LCD. This indicate batteries need to be replaced. Dual Angle Measurement Provide the measurement reading for both pitch and roll axis. It is suitable to determine exact angle between two surfaces. Show that digital post level is at zero level Show the proximity towards the zero level Figure above shown the correct usage in digital level mode Digital Level Mode 7. Calibration Procedures Step 1 (CAL 1) Step 2 (CAL 2) Step 3 (CAL 3) Step 4 (CAL 4) Step 5 (CAL 5) Step 6 (CAL 6) Caution: Digital Post Level device is calibrated in the factory for maximum accuracy. User only need to re-calibrates when necessary. 1) Press the CAL button. “CAL 1” show on digital display. Place the device as shown in the step 1. Ensure both surfaces of the aluminum bar are in full contact with the calibration bar. Press the “MODE” button, wait for the countdown from 10 to 1. ”CAL 2” show on display. 2) Place the device on the reverse side of calibration bar as shown in the step 2. Press the “MODE” button and wait for countdown from 10 to 1. ”CAL 3” will be shown on display. 3) Place the device as shown in the step 3. Press the “MODE” button and wait for the countdown from 10 to 1. ”CAL 4” will be shown on display. 4) Rotate the device 180 degree horizontally as shown on the step 4. Press the "MODE" button and wait for the countdown from 10 to 1. ”CAL 5” will be shown on the screen. 5) Place the device as shown in the step 5. Press the “MODE” button and wait for the countdown from 10 to 1. ”CAL 6” will be shown on display. 6) Rotate the device 180 degree horizontally as shown on the step 6. Press the “MODE” button and wait for the countdown from 10 to 1. 7) A long beep sound will signal, calibration is completed. 8. Relationship of Vial and Digital Reading Paired 9. Care and Handling • Avoid exposing unit to shock vibrations and extreme temperatures. • Before moving or transporting the unit, make sure that the unit is turned off. • Remove the batteries when storing the unit for an extended time(more than three months) to avoid damage to the unit should the batteries deteriorate. • Always store the unit in its case when not in use. • Keep the level dry and clean. Remove any moisture or dirt with a soft, dry cloth. • Do not use harsh chemicals, strong detergents or cleaning solvents to clean the level. 0°c ~ ± 50°c 2x “AAA” 1.5V batteries 150 hours (backlite on) 350 hours (backlite off ) 120 x 60 x 50 mm WARRANTY The Digi-Pas level is warranted to the original purchaser to be free from defects in workmanship and material. Our corporation will repair or replace any detective part which may malfunction under normal and proper use within a period of 30 days from the date of purchase. This warranty is not subject to misuse, abuse, assignment, or transfer. The exclusive remedy under any and all warrants and guarantees, expressed herein, and we shall not be liable for damages from loss or delay of equipments uses, consequential, or incidental damage. Distributed by: Manufactured by: JSB TECH PTE LTD 41 Science Park Road #04-02, The Gemini, Singapore Science Park 2 Singapore 117610 Tel: (65) 6778 0640 Fax: (65) 6777 0460 Email: [email protected] Website: www.digipas.com JIM1000XY-001-F09 1 2 3 5. LCD Screen Icon Indicate direction of inclination Indicate proximity to target Indicate zero level achieved Indicate batteries need to be replaced Indicate beep sound is activated via “Mode” button Indicate single axis function Indicate the selection of resolution: R1: 0.05 R2: 0.1 R3: 0.5 Indicate percentage mode Indicate angle of pitch and roll in percentage or degree. (See the relationship of vial and digit reading pairing details) 4 5 6 7 Switching Single & Dual Axis a) Press HOLD, then press mode when display, it is i n single axis function. b) To switch to dual axis, Press HOLD then press MODE. When disappear, it is in dual axis mode. 210g

Dwl1000xy

Embed Size (px)

Citation preview

Page 1: Dwl1000xy

Digi-Pas™A division of Group of Companies

DIGITAL POST LEVELDWL1000 XY

ISO 14001:2000Cet No.: TWO4/00450EM

ISO 9001:2000Cet No.: QSP00050

OHSAS 18001:2007Cet No.: SG08/50066

Tested to Complywith FCC Standards

Instruction ManualCongratulations on your choice on this Digi-Pas™ Digital Level DWL-1000We suggest you read this instruction manual thoroughly before using thedigital post level. Save thisinstruction manual for future reference.

Roof/drainage LevelingRoof/drainage Leveling

1. Kit Contents

Description QuantityDigital Post Level Device

“AAA” Alkaline Batteries

Aluminium Jig

1

2

1

2. Technical Speci�cation

Accuracy

Post Level Measuring Range

Digital Level Angle Measuring Range

± 0.10°

0° ~ ± 45°

0° ~ ± 90°

Resolution ± 0.05°

Operating Temperature

Power Supply

Estimated Battery Life

Dimension

Weight

3. Operation Instruction

Battery Installation

Battery Compartment

1. Open the battery cover, located as vertcial side plastic (See Figure 1).

2. Insert 2 new “AAA” alkaline batteries with polarity (+/-) as indicated on the battery compartment.

3. Close the battery door securely in place.

Fig 1

IMPORTANT NOTICE:It is the responsibility of the user to verify the calibration of the post level before each use.

Display LCD screen with Cross hair

Vial

Power ON/OFFPower on and o� the digitalpost level by pressing this key

Mode KeyTo Switch the resolution reading

R1: 0.05R2: .01R3: 0.5%

Back Light KeyTurn on the back light of LCDfor brighter LCD

4. Using the Product

HOLD Keya) Press HOLD , the digital reading blink. The angle is in the memory.b) Press HOLD again to release the function.CAL Key

Periodic checking should be conductedto this device. If it is out of alignment,recalibration will be necessary. This key isused for calibration to absolute horizontaland absolute vertical of the digital post level.See “Calibration Procedures” in section 7.

6. Special Features and Functions

Buzzer FunctionHold the “Mode” button for 4 sec to switch the buzzer on and o�.

Automated Shut-O�This digital post level will shut-o� automatically after 10 minutes.

Low-power indicationWhen voltage is low, battery symbol will be shown on the LCD. This indicate batteries need to be replaced.

Dual Angle MeasurementProvide the measurement reading for both pitch and roll axis. It is suitable todetermine exact angle between two surfaces.

Show that digitalpost level is at zerolevel

Show the proximitytowards the zerolevel

Figure above shown the correct usage in digital level mode

Digital Level Mode

7. Calibration Procedures

Step 1 (CAL 1) Step 2 (CAL 2) Step 3 (CAL 3)

Step 4 (CAL 4) Step 5 (CAL 5) Step 6 (CAL 6)

Caution: Digital Post Level device is calibrated in the factory for maximum accuracy. Useronly need to re-calibrates when necessary.

1) Press the CAL button. “CAL 1” show on digital display. Place the device as shown in the step 1. Ensure both surfaces of the aluminum bar are in full contact with the calibration bar. Press the “MODE” button, wait for the countdown from 10 to 1. ”CAL 2” show on display.

2) Place the device on the reverse side of calibration bar as shown in the step 2. Press the “MODE” button and wait for countdown from 10 to 1. ”CAL 3” will be shown on display.

3) Place the device as shown in the step 3. Press the “MODE” button and wait for the countdown from 10 to 1. ”CAL 4” will be shown on display.

4) Rotate the device 180 degree horizontally as shown on the step 4. Press the "MODE" button and wait for the countdown from 10 to 1. ”CAL 5” will be shown on the screen.

5) Place the device as shown in the step 5. Press the “MODE” button and wait for the countdown from 10 to 1. ”CAL 6” will be shown on display.

6) Rotate the device 180 degree horizontally as shown on the step 6. Press the “MODE” button and wait for the countdown from 10 to 1.

7) A long beep sound will signal, calibration is completed.

8. Relationship of Vial and Digital Reading Paired

9. Care and Handling

• Avoid exposing unit to shock vibrations and extreme temperatures.

• Before moving or transporting the unit, make sure that the unit is turned off.

• Remove the batteries when storing the unit for an extended time(more than three months) to avoid damage to the unit should the batteries deteriorate.

• Always store the unit in its case when not in use.

• Keep the level dry and clean. Remove any moisture or dirt with a soft, dry cloth.

• Do not use harsh chemicals, strong detergents or cleaning solvents to clean the level.

0°c ~ ± 50°c

2x “AAA” 1.5V batteries

150 hours (backlite on)350 hours (backlite o�)

120 x 60 x 50 mm

WARRANTY

The Digi-Pas level is warranted to the original purchaser to be free from defects in workmanship andmaterial. Our corporation will repair or replace any detective part which may malfunction under normaland proper use within a period of 30 days from the date of purchase. This warranty is not subject to misuse,abuse, assignment, or transfer. The exclusive remedy under any and all warrants and guarantees, expressedherein, and we shall not be liable for damages from loss or delay of equipments uses, consequential, orincidental damage.

Distributed by:

Manufactured by:

JSB TECH PTE LTD41 Science Park Road #04-02, The Gemini, Singapore Science Park 2 Singapore 117610

Tel: (65) 6778 0640Fax: (65) 6777 0460

Email: [email protected]: www.digipas.com

JIM

1000

XY-0

01-F

09

1 2 3

5. LCD Screen Icon

Indicate direction of inclination

Indicate proximity to target

Indicate zero level achieved

Indicate batteries need to be replaced

Indicate beep sound is activated via “Mode” button

Indicate single axis function

Indicate the selection of resolution:

R1: 0.05R2: 0.1R3: 0.5

Indicate percentage mode

Indicate angle of pitch and roll in percentage or degree.(See the relationship of vial and digit reading pairing details)

4 5 6 7

Switching Single & Dual Axisa) Press HOLD, then press mode when display, it is i n single axis function. b) To switch to dual axis, Press HOLD then press MODE. When disappear, it is in dual axis mode.

display, it is i n single axis function.

press MODE. When disappear,

210g