Left ventricular pressure tracings

Preview:

Citation preview

PRESSURE TRACING OF LEFT VENTRICLE

ByGopal C Ghosh

History of cardiac catheterisation

“The cardiac catheter was......the key in the lock”-Andre Cournand (1956)

Cardiac catheterization was first performed by Claude Bernard in 1844(horse)

Stephen Hales – as first to do cardiac catheterisation in animals(1711 on horse)

Mueller RL et al. Am Heart J 1995;129:146.

Cardiac catheterisation in human

• Werner Theodor Otto Forssmann( August 1904 – June 1979)

• First to do cardiac catheterisation in human heart

• Retrograde left heart catheterization was first reported by Zimmerman, Scott & Becker and Limon-Lason and Bouchard in 1950.

Zimmerman HA, Scott RW, Becker ND. Catheterization of the left side of the heart in man. Circulation 1950;1:357.

Dynamic pressure monitoring

• Dynamic blood pressure has been of interest to physiologists and physicians

• 1732- Stephen Hales measured the blood pressure of a horse by using a vertical glass tube

Fluid-filled standard catheter system

• Proper equipments for high quality catheterisation recordings

• Coronary angiography: smallest-bore catheters, 5F & even 4F

• Complex hemodynamic catheterization is optimally performed with larger-bore catheters that yield high-quality hemodynamic data. To obtain proper hemodynamic tracings, 6F or even 7F catheters may be required

Manifold system

Fluid-filled catheter is attached by means of a manifold to a small-volume-displacement strain gauge type pressure transducer

Wheatstone bridge

Pressure MeasurementTerminology

• Natural frequency– Frequency at which fluid oscillates in a catheter when it is tapped– Frequency of an input pressure wave at which the ratio of output/input

amplitude of an undamaged system is maximal

Grossman W. Cardiac Catheterization, Angiography, and Intervention. 5th Edition. Baltimore: Williams and Wilkins, 1996.

Shorter catheterLarger catheter lumenLighter fluid

Higher natural frequency

Damping

• Dissipation of the energy of oscillation of a pressure management system, due to friction

Greater fluid viscositySmaller catheter radiusLess dense fluid

Greater damping

Damped natural frequency• Frequency oscillations in the catheter when friction

losses are taken into account

Natural frequency = Damping System critically dampedNatural frequency < Damping OVERdampedNatural frequency > Damping UNDERdamped

Less damping greater artifactual recorded pressure overshoot above true pressure when pressure changes suddenly

More damping less responsive to rapid alterations in pressure

Sensitivity

• Ratio of amplitude of the recordedsignal to the amplitude of the inputsignal

Optimal damping can maintain frequency response flat (output/input ratio = 1) to 88% of the natural frequency of the system

Frequency response

• Ratio of output amplitude to input amplitude over a range of frequencies of the input pressure wave

• Frequency response of a catheter system is dependent on catheter’s natural frequency and amount of damping

• The higher the natural frequency of the system, the more accurate the pressure measurement at lower physiologic frequencies

Grossman W. Cardiac Catheterization, Angiography, and Intervention. 7th Edition.

What is the optimal frequency response?

• The essential physiologic information is contained within the first 10 harmonics of the pressure wave's Fourier series

• The useful frequency response range of commonly used pressure measurement systems is usually <20 Hz

• Frequency response was flat to <10 Hz with small-bore (6F) catheters

Reflected waves

• Both pressure and flow at any given location are the geometric sum of the forward and backward waves

FACTORS THAT INFLUENCE THE MAGNITUDE OF REFLECTED WAVES

Sources of Error

• Tachycardia If pulse is too fast for natural frequency of system, the fidelity of

the recording will drop. Pulse = 120 10th harmonic = 20 Hz Damped natural frequency

should be at least 60 Hz• Sudden changes in pressure• Deterioration in frequency response• Catheter whip artifact• End-pressure artifact• Catheter impact artifact• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia

• Sudden changes in pressure Peak LV systole, trough early diastole, catheter bumping against

wall of valve Artifact seen due to under damping

• Deterioration in frequency response• Catheter whip artifact• End-pressure artifact• Catheter impact artifact• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure

• Deterioration in frequency response Introduction of air or stopcocks permits damping and reduces

natural frequency by serving as added compliance When natural frequency of pressure system and low frequency

components falls, high frequency components of the pressure waveform (intraventricular pressure rise and fall) may set the system into oscillation, producing “pressure overshoots” in early systole & diastole

• Catheter whip artifact• End-pressure artifact• Catheter impact artifact• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure• Deterioration in frequency response

• Catheter whip artifact Motion of the catheter within heart or large vessels accelerates fluid in

catheter and produces superimposed waves of 10 mm Hg Common in pulmonary arteries & unavoidable• End-pressure artifact• Catheter impact artifact• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure• Deterioration in frequency response• Catheter whip artifact

• End-pressure artifact Pressure from endhole catheter pointing upstream is artifactually

elevated. When blood flow is halted at tip of catheter, kinetic energy is converted in part to pressure. Added pressure may range 2-10 mm Hg.

When endhole catheter is oriented into the stream of flow, the “suction” can lower pressure by up to 5%

• Catheter impact artifact• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure• Deterioration in frequency response• Catheter whip artifact• End-pressure artifact

• Catheter impact artifact Pressure transient produced by impact on the fluid-filled catheter by an

adjacent structure (i.e. heart valve) Any frequency component of this transient that coincides with the natural

frequency of the catheter manometer system will cause a superimposed oscillation on the recorded pressure wave

Common with pigtail catheters in the left ventricular chamber, where the terminal pigtail may be hit by the mitral valve leaflets as they open

• Systolic pressure amplification in the periphery• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure• Deterioration in frequency response• Catheter whip artifact• End-pressure artifact• Catheter impact artifact

• Systolic pressure amplification in the periphery Consequence of reflected wave Peripheral arterial systolic pressure commonly 20 mm Hg higher than central

aortic pressure (mean pressure same or slightly lower) Masks pressure gradients in LV or across aortic valve Use of a double-lumen catheter (e.g., double-lumen pigtail) or trans-

septal technique with a second catheter in the central aorta can be helpful

• Errors in zero level, balancing, calibration

Sources of Error• Tachycardia• Sudden changes in pressure• Deterioration in frequency response• Catheter whip artifact• End-pressure artifact• Catheter impact artifact• Systolic pressure amplification in the periphery

• Errors in zero level, balancing, calibration Zero level must be at mid chest level All manometers must be zeroed at same point Zero reference point must be changed if patient repositioned Transducers should be calibrated against standard mercury

reference (rather than electrical calibration signal) and linearity of response should be verified using 25, 50, and 100 mm Hg

Intracardiac micromanometers (Catheter-tip pressure manometer)

• High fidelity transducer catheter with miniaturized transducer placed at tip (Millar Instruments)• Improved frequency response characteristics

and reduced artifact• Measurement of myocardial mechanics

(dP/dt of LV)

Cardiac cycle

Wiggers Diagram

• Systolic ejection phase - QT interval on the ECG

• LV systolic pressure is measured at the peak pressure of the ejection phase

Left ventricular end diastolic pressure

• End Diastolic pressure can be measured on the R wave of the ECG, which will coincide just after the ‘a’ wave on the LV trace. This is called the post ‘a’ wave measurement of EDP

• To be measured at end expiration

Aortic Pressure

Anachrotic Notch

• During the first phase of ventricular systole (isovolumetric contraction), a presystolic rise may be seen

• Occurs before the opening of the aortic valve

Dichrotic notch

• Blood flow attempts to equalize by flowing backwards - results in closure of the aortic valve

• This event marks the end of systole and the start of diastole

Pulse Pressure

• The difference between the systolic and diastolic pressure

• Factors Factors that affect pulse pressure pressure

– changes in stroke volume – aortic regurgitation – changes in vascular compliance

Common cardiological conditionsNeeds invasive monitoring

Aortic Stenosis(severity)• Discrepancy between the physical examination

and the elements of the Doppler echocardiogram• Optimal technique: record simultaneously

obtained left ventricular and ascending aortic pressures

• Mean gradient differences not the peak to peak gradient (peak left ventricular pressure does not occur simultaneously with the peak aortic pressure)

• Pullback traces with a single catheter from the left ventricle to the aorta can be helpful, but only if the patient is in normal sinus rhythm with a regular rate

• Carabello sign:Critical aortic stenosisCatheter across the valve itself will cause

further obstruction to outflowThis sign occurs in valve areas 0.6 cm2 or less

& when 7F or 8F catheters are used to cross the valve

Carabello sign

Use side hole catheters. Aortic pressure damping can occure with end hole catheters

Dont use femoral arterial pressure:1. Peripheral amplification of arterial wave-decrease the gradient falsely2. PAD-increase gradient falsely

Evaluation of low flow low gradient aortic stenosis(LVEF<40%)

• Gradient (<30 mm Hg) and a low output, resulting in a small calculated valve area(<1cm2)

• These patients with low-flow/low-gradient AS (LF/LGAS) may truly have severe AS with resultant myocardial failure (true AS) or may have more moderate degrees of AS and unrelated myocardial dysfunction (pseudo-AS)

• True AS: increased flow across a fixed valve orifice results in increased transvalvular flow velocity and gradients, without a change in calculated valve area.

• Pseudo-AS: augmented flow results in only a mild increase in transvalvular gradient and an increase in valve area by ≥0.2 cm2

Dynamic LVOT obstruction

• Visual assessment:Aortic stenosis:• Delay(tardus) and reduction (parvus) in the

upstroke of the central aortic pressureDynamic LVOT obstruction:• Spike-and-dome pattern with an initial rapid

upstroke• Late peaking left ventricular pressure

Braunwald-Brockenborough sign

Mitral Stenosis

• Continuous-wave Doppler echocardiography is highly accurate

• Noninvasive estimations are inconsistent & pulmonary hypertension out of proportion to the apparent severity of the mitral valve disease

• Simultaneous pulmonary artery wedge pressure and left ventricular pressure

Overestimates

Evaluation of valvular regurgitation

• left ventriculography and aortic angiography are the modalities most often used to assess the severity of valve regurgitation

• When there is a discrepency between clinical assessment and dopplar echocardiographic measurement

• Sellar criteria used

Sellers criteria

Precautions

• Large-bore catheters and a large amount of contrast to completely opacify the cardiac chambers(small amount underestimate)

• Avoidance of ventricular ectopy and entrapment of the mitral valve apparatus by the catheter

• High right anterior oblique views for left ventriculograms may be necessary to avoid the retrograde contrast from being superimposed on the spine or descending aorta

Evaluation of aortic regurgitation

• Grade 1: a small amount of contrast material enters the left ventricle in diastole; it is essentially cleared with each beat and never fills the ventricular chamber

• Grade 2: More contrast material enters with each diastole and faint opacification of the entire chamber occurs

• Grade 3: the LV chamber is well opacified and equal in density with the ascending aorta

• Grade 4: complete, dense opacification of the LV chamber in one beat, and the left ventricle appears more densely opacified than the ascending aorta

Part of the angiographic assessment of aortic regurgitation involves assessment of (LAO view):

1. Aortic valve leaflets (mobility, calcification, number of leaflets)

2. Ascending aorta (extent and type of dilatation)

3. Possible associated abnormalities (e.g., coronary lesions, sinus of Valsalva aneurysm, dissecting aneurysm of the aorta, and ventricular septal defect)

Constrictive Pericarditis VersusRestrictive Cardiomyopathy

CCP RCM1. Ventricular interdependence

Present Absent

2. Pulmonary arterial systolic pressure> 55-60 mm hg

No Yes

3. Equalisation of Right & Left ventricular EDP

Yes No

4. Dip and platue pattern of left ventricular end diastolic pressure

Yes No

5. RVEDP/RVSP> 1/3 Yes No

David G Hurrell et al. Circulation 1996

Ventricular interdependence

• Highly sensitive & specific• Also seen in cardiac tamponade, acute right

ventricular infarction, subacute tricuspid regurgitation

• Inspiratory decrease in pulmonary venous and intrathoracic pressure is not transmitted into the cardiac chambers

Hatle et al.  Circulation.1989;79:357-370. 

Irina Kozarez et al. Grand Rounds Vol 11 pages 111–114

Hypertrophic Cardiomyopathy

• There is frequently a dynamic left ventricular outflow tract obstruction that is highly dependent on loading conditions and the contractile state of the ventricle

Indication for septal ablation

• Suitable anatomy• Severe symptoms unresponsive to medical

management• Documented left ventricular outflow gradient

of 50 mm Hg either at rest or during provocation

2011 ACCF/AHA Guideline for the Diagnosis and Treatment of Hypertrophic Cardiomyopathy: Executive Summary

• If there is a gradient 50 mm Hg at rest, provocative maneuvers such as the Valsalva maneuver or induction of a premature ventricular contraction should be performed

• However, if a gradient is not provoked with these maneuvers, infusion of isoproterenol is helpful because direct stimulation of the beta 1 and beta 2 receptors simulates exercise and may uncover a labile outflow tract gradient

Hemodynamics

• The left ventricular outflow tract gradient is dynamic and can change significantly during a single diagnostic catheterization

• It is recommended that catheters such as a multipurpose or Rodriquez catheter with side holes at the distal portion of the catheter should be used to determine the exact location of obstruction (pigtail with multiple side holes to be avoided & single end hole catheters to be avoided)

Nishimura et al. Circulation 2012

Brockenbrough-Braunwald sign

• The decrease in pulse pressure after a premature ventricular contraction is due to reduced stroke volume caused by increased dynamic obstruction

HCM: Typical midcavitary obstruction

Fernando Pivatto et al. Rev Bras Cardiol Invasiva. 2014

Apical HCM

C.-C. Chen et al: Clin. Cardiol. 34, 4, 233–238 (2011)

Takotsubo cardiomyopathy

• Apical ballooning with basal hyperkinesis on the left ventriculogram.

Golabchi A et al. J Res Med Sci. 2011 Mar;16(3):340-5.