6/3/2014 1 Hareesha N G, Dept of Aero Engg, DSCE
The basics of CT
• CT imaging chain
• System components
• Acquisition methods
• Image quality
• Applications
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 2
X-ray: The beginning
• X-Rays founded in 1895 by Wilhelm Conrad Roentgen
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 3
CT: The beginning
• CT founded in 1970 by Sir Godfrey Hounsfield
– Engineer with EMI, LTD.
– first applications were in neuroradiology
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 4
CT Scanner
• X-Ray modality used to the body in cross section
• Used to determine
– extent of trauma
– location and type of tumors
– status of blood vessels
– pre surgical planning
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 5
CT System
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 6
CT System
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 7
Basic CT scanner components
• Gantry
• X-Ray Tube
• Detector
• Control Console
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 8
Gantry
• CT X-ray tube
• High voltage generator
• Detector array
• Data acquistion system
• Slip ring
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 9
The CT X-ray Tube
• Anode heat capacity
– 3.5 MHU up to 6.5 MHU
• Determines maximum mAs
• Determines volume length
• Dictates generator size
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 10
Detector Elements
• Capture energy that has not been attenuated by the patient
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 11
Control console
• Set scan parameters – kVp, mA, scan time, reconstruction filter,
etc.
• Set scan mode – Digital radiograph, axial or volume
• Houses reconstructor
• Review and archive images
• Post-processing
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 12
CT
• CT - Computed Tomography
• CAT Scan - Computerized Axial Tomography
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 13
Scanning methods
• Digital projection – AP, PA, Lat or Oblique projection
– Surview, Scanogram
• Conventional CT – Axial
• Start/stop
• Volumetric CT – Helical or spiral CT
• Continuous acquisition
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 14
Digital Projection
• X-ray tube and detector remain stationary
• Patient table moves continuously
– With X-rays “on”
• Produces an image covering a range of anatomy
– Similar to a conventional X-ray image, e.g. flat plate of the abdomen
• Image used to determine scan location
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 15
Axial CT
• X-ray tube and detector rotate 360°
• Patient table is stationary – With X-ray’s “on”
• Produces one cross-sectional image
• Once this is complete patient is moved to next position – Process starts again at the
beginning
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 16
Volume CT
• X-ray tube and detector rotate 360°
• Patient table moves continuously – With X-ray’s “on”
• Produces a helix of image information – This is reconstructed into 30 to 1000 images
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 17
Volume Scanning: Pitch
1:1 2:1
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 18
Advantages of Volume CT
• More coverage in a breath-hold – Chest, Vascular studies, trauma
• Reduced misregistration of slices – Improved MPR, 3D and MIP images
• Potentially less IV contrast required
• Gapless coverage
• Arbitrary slice positioning
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 19
Fundamentals of Multislice CT
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 20
Multislice Fundamentals
• Everything is better
• (R)esolution – Z-axis, spatial, low contrast
• (S)peed – Temporal -
bolus capture, stopped motion
• (V)olume – Thin slice -
– organ-specific coverage
• (P)ower – Enough photons -
uncompromising image quality
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 21
Single Slice = One 10mm slice per rotation
Dual Slice = Two 5mm slices per rotation
Quad Slice = Four 2.5mm slices per rotation
Multislice Effectiveness
• Everything is better
– Resolution 2x 4x-8x
– Speed same same
– Volume same same
– Power same same
Dual Quad
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 22
. .
single detector arc dual detector arc
pre-patient collimation
post-patient collimation
x-ray tube focal spot
_ _ _
Mx8000
Dual Slice
Dual Slice Detector Optimized for 2 Slice Acquisition
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 23
Approximately 10%
more efficient than
matrix detectors
Variable Wide Area Detector
Asymmetrix™
Variable detector length Fixed detector length
Quad Detector Technology
• Philips patented variable wide area detector
• Variable slice thickness
– 4 x 1mm
– 4 x 5mm
– 4 x 2.5mm
– 2 x 0.5mm
– 2 x 8mm
– 2 x 10mm
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 24
8 Element
2-D array
4 Slices
Quad Technology How it works
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 25
2x0.5mm
4x1mm
4x2.5mm
4x5mm
2x10mm
p-plane
fused fused
to FEE
Asymmetrix™ Technology Variable slice thickness
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 26
CT
• CT attenuation information
• CT image quality
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 27
Attenuation
• X-ray beam passes through patient
• Each structure attenuates X-ray beam differently
– According to individual densities
• Radiation received by detector varies according to these densities
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 28
Density information
• Transferred from detector to CT computer (A to D converter)
• Reconstructed by computer into a cross-sectional image – Displayed on screen
– Each pixel displayed on monitor has varying brightness
• The greater the attenuation, the brighter the pixel
• The less attenuation, the darker the pixel
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 29
Density information
• Density values correspond to a range of numbers
– Hounsfield scale
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 30
Window settings
• Window width – Determines range of CT numbers displayed on an image
• Values above this range = white
• Values below this range = black
– Window level
• Sets the center CT number displayed on the monitor
• Determines the location on the Hounsfield scale about which the window width will be centered
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 31
CT image quality
• Spatial resolution
– Ability to resolve small objects in an image
– Measured in lp/cm
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 32
Isotropic Imaging
• True 0.5mm Isotropic imaging
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 33
CT image quality
• Contrast resolution
– Ability to differentiate small density differences in an image
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 34
Post Processing Options
• Visualization of vasculature in relation to pathology
– Show course of vessels
– Show stent placement
– Define vascular stricture
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 35
Cervical Spine Spiral Acquisition
Rotation – 0.75 sec
Coverage – 160 mm
Pitch – 0.875
Acq. Time – 36 sec
FOV – 250 mm
ST – 1.0 mm
Rec. Incr. – 0.6 mm
Std Res. – 8 lp/cm
120 kV, 200 mAs
CTDI100w – 39 mGy
CTDIFDAw – 17 mGy
Thin-Slice Spiral Neck
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 36
Neuro-Angiography
Circle of Willis
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 37
Thin-Slice Spiral Lungs
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 38
MIP 4D-Angio
MasterCut Panoramic View
Renal Arteries
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 39
Ext. Spiral Acq. Spiral Acquisition
UltraFast – 0.5 sec
Coverage – 1400mm
Pitch – 1.75
Acq. Time – 41.5 sec
FOV – 420 mm
ST –2.5 mm
Rec. Incr. – 1.6 mm
Std Res. – 8 lp/cm
120 kV, 96 mAs
CTDI100w – 7.3 mGy
CTDIFDAw – 4.9 mGy
Extended Spiral Acquisition
6/3/2014 Hareesha N G, Dept of Aero Engg, DSCE 40
CT Scanners
• Provide a window into the body
• Customer considerations – How many patients
– Referring physicians
– Budget
– Upgrade expectations
• Philips has ALL the answers