VENTURACPAPUSER GUIDE
vent
ura
E M E R G E N C Y U S E O N LYn o t F O R R O U T I N E C L I N I C A L U S E W I T H O U T F U R T H E R A P P R O V A L
A L L H O S P I T A L S M U S T W O R K W I T H T H E I R O X Y G E N e n g i n e e r i n g T E A M S T O A S C E R T A I N T H E I R V I E O U T F L O W , A N D D O W N S T R E A M F L O W S A N D P R E S S U R E S T O S P E C I F I C W A R D A R E A S , B E F O R E D E P L O Y I N G T H I S D E V I C E
OXYGENINLET
flowadjustment
oxygenadjustment
FLOW ON/OFF
TOpatient
high
low
high
low
ON
OFF
AIR INTAKE
v e n t u r a
vent
ura
A
B
C
v 2 . 0
ONLY TO BE USED WITH THE PATIENT CIRCUIT PROVIDED
flow generator
oxygen analyser(MODELS VARY)
safety peepvalve
Patient Interface
Ab
c
flow flagindicator
(20 cmh 0)2
PEEP VALVE(cpap: 5-10 cmh 0)2
INCREASEFLOW
CORRECTFLOW
REDUCEFLOW
POSITIONDURING
INSPIRATION
1. Turn all controls (A, B, C) OFF (clockwise) and connect the UCL-Ventura to the 4-bar oxygen supply directly or by using an extension hose.
2. Connect the flow generator to the patient circuit shown.
3. Turn the FLOW ON/OFF control (A) fully ON (anticlockwise) and open the FLOW ADJUSTMENT control (B) (3-4 complete rotations anticlockwise).
4. Increase the oxygen concentration by turning the OXYGEN ADJUSTMENT control (C) (anticlockwise) to the desired level.
5. Apply mask to the patient with appropriate straps and ensure a good fit.
6. Alter the FLOW ADJUSTMENT control (B) to ensure optimal flow to:
• Flutter the mask outlet valve (larger port) throughout the respiratory cycle (the valve will flutter less during inspiration).
• Hold the blue flow indicator flag (where fitted) horizontal or just below horizontal on inspiration (the flag will remain below horizontal on expiration).
• A virtually continuous flow at the outlet filter should be felt on the back of a gloved hand throughout the respiratory cycle.
7. Alter the OXYGEN ADJUSTMENT control (C) if the oxygen concentration has changed.
If the device becomes damaged it should be withdrawn from use.
USING THE VENTURA CPAP
Recommended