The CATANA Protontherapy Beam Delivery System. Cuttone G., Cirrone G.A.P., Lojacono P.A., Lo Nigro S., Raffaele L., Rovelli A.,Sabini M.G., Salamone V. Laboratori Nazionali del Sud (INFN) Via S. Sofia 44 Catania, ITALY. - PowerPoint PPT Presentation
Citation preview
The CATANA
Protontherapy Beam Delivery System
Laboratori Nazionali del Sud (INFN)Via S. Sofia 44Catania, ITALY
Cuttone G., Cirrone G.A.P., Lojacono P.A., Lo Nigro S., Raffaele L., Rovelli A.,Sabini M.G., Salamone V.
LNS Superconducting Cyclotron is the
unique machine in in Italy and South Europe used for protontherapy
Treatment of the choroidal melanoma
In Italy about 300 new cases for year
•0 ° respect the switching magnet
•80 meter after extraction
•3 m proton beam line
LAYOUT OF LNSPRESENT TREATMENT ROOM
OUR HISTORY STARTED THREE YEARS AGO......
1999
WITH A BARE BUT VERY STABLE TABLE
AND A SUPERCONDUCTOR CYCLOTRON SUITABLE FOR62 AMeV PROTON BEAM
FIRST OF ALL WE TOOK IN CONSIDERATION TWO
PARAMETER FOR THE DEVELOPMENT OF OUR PROTON
BEAM DELIVERY SYSTEM:
ENERGY In order to obtain ranges up to 30 mm E.T.
OMOGENEITY In terms of lateral dose distribution in air
In order to reach a proton beam irradiation field:
Omogeneus within 2.5 %
With lateral penumbra of about 1mm
At least 30 mm in diameter
at isocenter in air
WE DECIDED FOR A PASSIVE SCATTERER SYSTEM AND STUDIED THREE DIFFERENT EXPERIMENTAL
CONFIGUARATIONS
SIMULATION TOOLKIT : TRIM2000 UNDER NT OS
GEANT3.0 UNDER LINUX RED HAT 6.1 OS
SINGLE SCATTERER FOIL300 m Tantalum
DOUBLE SCATTERER FOIL80 m + 20 m Tantalum
DOUBLE SCATTERER FOIL WITH CENTRAL STOPPER15 m + 25 m + 7 mm thick copper beam stopper
SIMULATION RESULTS
TRANSMITTED ENERGY [AMeV]
PRACTICAL RANGE
[mm WATER]
LATERAL OMOGENEITY [mm][FROM 95% TO 95%]
SINGLE FOIL
57.248 28.43 27
DOUBLE FOIL
57.625 29.12 24
DOUBLE FOIL WITH
BEAM STOPPER
60 30.97 47
0
20
40
60
80
100
120
140
-40 -30 -20 -10 0 10 20 30 40
Distance from Isocenter [mm]
Nor
mal
ized
Dos
e D
istri
butio
n [%
]
MONTE CARLO SIMULATION
EXPERIMENTAL RESULT(Silicon Diode)
EXPERIMENTAL RESULTS
OUR BEAM PROFILE WITH A 50 mm DIAMETER FINAL COLLIMATOR
SIMULATION VS. EXPERIMENTAL RESULT
0
20
40
60
80
100
120
0 5 10 15 20 25 30 35 40
Depth in Water [mm]
No
rmal
ized
Do
se
EXPERIMENTAL RESULTS
LNS PURE BRAGG PEAK
PRACTICAL RANGE
[mm water]
PEAK/PLATEAU RATIO
FWHM [mm
water]
DISTAL FALL-OFF [mm water]
Energy [AMeV]
EXPERIMENTAL DATA 30.9 4.5 3 0.8 60
SIMULATION DATA 30.97 4.47 2.7 0.6 60
Scatterer system
Modulator & Range shifter
Monitor chambers
Ligth field
Laser
WE EVALUATED THE PROTON BEAM LATERAL DISTRIBUTION ALONG OUR BEAM LINE
1. JUST AFTER THE FIRST SCATTERER
0
20
40
60
80
100
120
140
-35 -25 -15 -5 5 15 25 35
2. AFTER THE SECOND SCATTERER
0
20
40
60
80
100
120
140
-35 -25 -15 -5 5 15 25 35
3. AT ISOCENTER WITH A 25 mm COLLIMATOR
0
20
40
60
80
100
120
140
-35 -25 -15 -5 5 15 25 35
The actual distance beetwen modulator and range shifter is 33 cm
IN COLLABORATION WITH CCO WE PRODUCED A PRELIMINARY SET OF MODULATORS
0
10
20
30
40
50
60
70
80
90
100
110
0 5 10 15 20 25 30 35
Depth in water [mm]
No
rmal
ized
Dep
th D
ose
Dis
trib
utio
n
It is interesting to note we didn’t find any difference in the experimental results, varing this distance from 15 cm up to 33 cm
1999
ACTUALLY THE BEAM LINE IS READY
2002, February
WE ACKNOWLEDGE FOR THEIR SUPPORT AND ENCOURAGEMENT: