RDS-CAA Qualification and Programming Primer for Calypso version 5.2

Preview:

DESCRIPTION

RDS-CAA Qualification and Programming Primer for Calypso version 5.2. The MasterProbe will have to be of type RDS-CAA in order to proceed with RDS-CAA qualification. The “Fitting Position” button should be visible. - PowerPoint PPT Presentation

Citation preview

RDS-CAA Qualification and Programming Primerfor Calypso version 5.2

The MasterProbe will have to be of type RDS-CAA in order to proceed with RDS-CAA qualification. The “Fitting Position” button should be visible.

RDS-CAA Primer (5.2) 2 of 36

If MasterProbe is not already of type RDS-CAA, drop down the “Mode” menu, choose “CFS qualification,” and proceed with locating Reference Sphere Position

RDS-CAA Primer (5.2) 3 of 36

If MasterProbe is already of type RDS-CAA, select “Ref. Sphere position” and follow prompts. NOTE: If not already in A=08, B=08, be sure to drive the RDS clear of all

obstacles.

RDS-CAA Primer (5.2) 4 of 36

When qualifying RDS-CAA, it is recommended that the Reference Sphere is oriented vertically.

Unscrew the reference sphere from rotating plate, loosen rotating plate and rotate 180 and reattach the reference sphere to the top threaded hole.

RDS-CAA Primer (5.2) 5 of 36

Be sure that the “Tilt” is set correctly. If the Reference Sphere is oriented vertically, “Tilt” should be set to 1808. It matters not what Rotation angle is set. Take the first point in the

direction of the stylus, as always.

RDS-CAA Primer (5.2) 6 of 36

After setting the Reference Sphere position, it is necessary to set the Fitting Position of the MasterProbe. In this process, the MasterProbe will automatically find the Reference

Sphere position in all 12 orientations.

RDS-CAA Primer (5.2) 7 of 36

The progress of the Fitting Position is posted in the window. Status will indicate an OK status unless there has been a collision. Check that the Standard Deviation for

MasterProbe is acceptable.

RDS-CAA Primer (5.2) 8 of 36

To create a new stylus system, attach the stylus system, select “Pick Up Stylus System.”

Choose “New”

RDS-CAA Primer (5.2) 10 of 36

Check the Probe type (RST, XDT, XXT TL3)NOTE: XXT types TL1 and TL2 are not compatible with RDS-CAA qualification.

RDS-CAA Primer (5.2) 11 of 36

Enter the NAME and STYLUS NAME “1.” Check box for RDS-CAAFor qualification of individual positions, the box must remain unchecked.This is the only place where qualification type (normal or RDS-CAA) may be set

RDS-CAA Primer (5.2) 12 of 36

Select “Qualify stylus.” NOTE: If not already in A=08, B=08, be sure to drive the RDS clear of all obstacles. Take point in the direction of stylus shaft. Qualification of all 12

positions will be automatic.

RDS-CAA Primer (5.2) 13 of 36

Progress of qualification will be posted in the window.

RDS-CAA Primer (5.2) 14 of 36

To add a new stylus, select the “Add New Stylus” buttonEnter Name “2.”

RDS-CAA Primer (5.2) 15 of 36

Select “Qualify Stylus,” Choose “Full Qualification.”

RDS-CAA Primer (5.2) 16 of 36

In the window, select each position, choosing “Show” for each. Be sure head is free for rotation.

RDS-CAA Primer (5.2) 17 of 36

If any position is not able to be qualified, choose “Alternative.” Click “Show” for the new angle.

RDS-CAA Primer (5.2) 18 of 36

Once all positions have been proofed for collisions, Click OK.

RDS-CAA Primer (5.2) 19 of 36

Progress will be posted in the window.

RDS-CAA Primer (5.2) 20 of 36

Once the Stylus System has been qualified with RDS-CAA, it may be used in measurement programs.

RDS-CAA Primer (5.2) 21 of 36

Create the measurement program as usual.

RDS-CAA Primer (5.2) 22 of 36

In The Features Settings Editor, navigate to Stylus System.

RDS-CAA Primer (5.2) 23 of 36

If not already assigned, select features and “Set to” the desired Stylus system.

RDS-CAA Primer (5.2) 24 of 36

If not already assigned, select features and “Set To” desired Stylus (“1” or “2”).

RDS-CAA Primer (5.2) 25 of 36

In the new menu, “RDS Angles” it is possible to set the angle for each feature, however this method is rather tedious and fraught with errors.

RDS-CAA Primer (5.2) 26 of 36

Instead, go to the Stylus Rotation window.

RDS-CAA Primer (5.2) 27 of 36

A CAD representation of the head, probe, and stylus appears. The current position is listed under “New Angle Position”

A=0.0, B=0.0 indicates a “down” position.

RDS-CAA Primer (5.2) 28 of 36

Angle Cheat Sheet

A=-90, B=90indicates a “back” orientation

A=0, B=90indicates a “right” orientation

A=-90, B=-90indicates a “front” orientation

A=0, B=-90indicates a “left” orientationRDS-CAA Primer (5.2) 29 of 36

Alternatively, the angle can be set by the orientation of a feature. Select the feature and select “To Feature.”

RDS-CAA Primer (5.2) 30 of 36

If the preview of the angle of the stylus is incorrect, select “Invert Position.”

RDS-CAA Primer (5.2) 31 of 36

Once satisfied with the angle, accept the angle.

RDS-CAA Primer (5.2) 32 of 36

To apply current angle to features, select the appropriate features and select “Apply Angle.”

RDS-CAA Primer (5.2) 33 of 36

Changes should be reflected in the Features Settings Editor.

RDS-CAA Primer (5.2) 34 of 36

Select orientation and apply for remaining features.

RDS-CAA Primer (5.2) 35 of 36

Changes should be verified in Features Settings Editor.

RDS-CAA Primer (5.2) 36 of 36

Recommended