Characterization of low energy ionization signals from Compton scattering in a CCD Dark Matter detector

Karthik Ramanathan University of Chicago

arXiv:1706.06053 (Accepted PRD) TeVPA 2017/08/07

1

Motivation

2

}  Solid state ionization detectors are integral component of next-generation dark matter searches due to their very low noise and the small band gap of semiconductor targets.

}  However in this low energy search regime (2-1000e-) dominant background from environmental radiation are low-energy electron recoils due to small-angle Compton Scattering of external gammas.

}  Flux is orders of magnitude higher than fast neutrons – the usual consideration for external source of signals

}  Irreducible Electron Recoil background à any potential Dark Matter can only be identified by energy spectrum.

à Need complete understanding of low-energy spectral features. à Expose UChicago Silicon CCD detector to gamma source

Motivation II

3

}  Expose to γ-­‐ray  source  }  Compton  features  +  ?

Modeling

4

}  Generically scattering cross-section given by textbook Klein-Nishina. However dealing with bound electrons. }  Expect effectively flat spectrum (with these added steps)

}  Impulse Approximation: Each atomic shell treated independently. Bound electrons are modeled as free with constrained momentum distribution derived from bound-state wave function. }  Ribberfors 1982 (https://doi.org/10.1103/PhysRevA.26.3325) }  Valid in our region of interest with low energy and momentum

transfers

}  Useful since we can obtain differential cross section expressions per atomic electron with quantum numbers n, l

Expected Spectrum

5

}  Silicon Target }  Visible Step

features }  Binding Energies

}  Provide linear parameterization (since in aggregate an unknown spectrum can be fit with straight lines...)

Testbed (UChicago)

6

cv

cv

cv

cv

Detection Principle

7

1x1 Data & MCNP Simulation Model

8

0 20 40 60 80 100 120 140

E [keV]

101

102

103

104

Numberofevents[(50eV)!1]

A Compton Edges

#

Photoelectric Absorption !Peaks

A Fluorescence

57Co 1#1MCNP

5 5.5 6 6.5 7 7.5 8 8.5E [keV]

45

50

55

60

65

70

<e[eV]

Fano Best Fit ! F = 0.128(2)

1000

2000

3000

4000

5000

6000

7000

Num.of

events

[(10

eV)!

1]

Cr K,

Mn K,

Fe K,

Fe K-

Ni K, Cu K,

57Co 4#4 DataGaussians + O,set

Fano Factor (@ 130 K)

9

Results - Cobalt

10

Results - Americium

11

L-Step

12

}  Fano model should be valid }  External modeling of all low-energy electrons emitted in Auger

cascade (RELAX atomic relaxation spectra code)

}  Calibration with Oxygen fluorescence x-rays à 21 eV resolution at Eg = 525 eV

}  Interpret decreased resolution as coming from softened L step in electron spectrum à Assumption that each atomic shell can be treated as independent does not hold? Many-body effects?

Model

13

}  From 0.5-4 keV }  Initial 3 parameter

model with fixed step heights discarded

}  6 parameter model à 2 slopes à K step height à L step location and

resolution (σL ) à Normalization

Model II

14

}  6 parameter model à 2 slopes à K step height à L step location and

resolution à Normalization

}  Able to model fit in <4 keV range to within 5% without accurate background knowledge

}  Flattens out at high γenergies

Takeaway

15

Primary }  Report, for first time, spectral Compton features associated

with the atomic structure of the target. }  Characterize the spectrum of low-energy ionization signals

from electrons Compton scattered by radiogenic γ-rays, vital for future DM searches

}  Validate applicability of simple linear model Secondary }  Demonstrate again CCD resolution down to ~60 eV }  Measure Fano Factor @ operating temperature

Remains an open question as to what happens at low energies?

Questions?

16

Exclusion Plot

17

Impulse Approximation

18

Expression valid only for E > Enl, the target electron’s binding energy. Otherwise it’s 0 as the min. energy photon can lose is that required to free the target electron

are the Compton profile functions, which encode the momentum distribution of the target electron and is taken from tabulated data. Further the integral can only be evaluated numerically.

Projection of electron momentum on scattering vector

Source Selection

19

Binning

20

}  Hardware adding of neighboring pixels at serial register }  e.g. 1x100 à 100 rows (y) transferred into serial register

before clocking in x (column) direction à Fewer pixels but same noise per pixel

Dataset

21

}  Cobalt dataset taken early 2016, Americium early 2017 }  Single 4k x 2k CCD (2.2 g mass) }  Analysis conducted using 4x4 data (1x1 used for

validation)

Image Processing

22

}  Pedestal (DC offset) subtraction à Pixel values centered at 0 with noise σpix

}  Mask “hot” pixels & lattice defects (~10% removed) }  Energy calibration done with fluorescence & P.E peaks

}  Linearity previously demonstrated using this setup

}  1x1 datasets }  Clustering done by 11x11 moving window maximizing difference

in log-likelihood between 2 hypotheses: 2D Gaussian+Noise or just Noise.

}  4x4 datasets }  Clusters identified as ionization events with contiguous pixels > 4 σpix

1x1 Diffusion Modeling

23

}  Simulated events with uniform energy distribution between 0-1 keV and uniform spatial distribution, using diffusion parameters tuned at high energies, and compared to data.

Verifies that recorded spatial distribution is consistent with the signal from Compton scattering, with negligible contamination from surface events.

Efficiency

24

}  Based on simulation, able to construct detector efficiency curves

90%+ efficient at energies > 60 eV; > 99% above 80 eV

Pixel Cuts

25

}  Energy threshold chosen to exclude readout noise.

}  Negligible single pixel readout noise > 60 eV, but present for 2+ pixels until 80 eV.

Consider only single pixel events between 60-80 eV

“Sensei”

26

}  Repeat measurement in near future

}  Non destructive “skipper” readout R&D project.

}  Perform N uncorrelated measurements for ~1/Sqrt(N) noise reduction