Cerebral Oximetry Wesam Farid Mousa Assist prof Anesthesia & ICU Dammam University

Cerebral oximetry

Embed Size (px)

Citation preview

Page 1: Cerebral oximetry

Cerebral Oximetry

Wesam Farid MousaAssist prof Anesthesia &

ICUDammam University

Page 2: Cerebral oximetry

Experts have created helmets to protect the brain of motorcyclists

Page 3: Cerebral oximetry

Under general Anesthesia, the human brain can be in danger and the anesthesiologists are

sometimes faced with neuronal dysfunction/death during anesthesia

Page 4: Cerebral oximetry

NEURODETECTION & NEUROPROTECTION

• No “real time” bioassays to detect the onset of brain injury

•CT, ECHO, EEG & MRI are late indicators of brain injury

Page 5: Cerebral oximetry

• The need is critical: Cerebral ischemia is the

leading cause of compromised neurocognitive outcomes

The duration of reduced oxygenation has a direct impact on brain function

Page 6: Cerebral oximetry

• THE EVIDENCES Approximately 465,000

cardiopulmonary bypass surgeries performed in the U.S. each year

Cerebral oxygen desaturation occurs in 17%-23% of cardiopulmonary bypass surgeries (or about 79,000-107,000 cases per year) Most desaturation events cannot be detected by common monitoring devices

Page 7: Cerebral oximetry

• THE EVIDENCES• Adverse neurologic outcomes have been

reported in 6% to 53% of patients undergoing cardiopulmonary bypass surgeries

• Neurological complications vary in

severity; some are transient and some are permanent

• Adverse cerebral outcomes include focal injury, stupor, coma, seizures, memory deficit or deterioration in intellectual function

Page 8: Cerebral oximetry

Incidence of Cerebral Desaturation Events

Page 9: Cerebral oximetry

WHY CEREBRAL OXIMETRY?

• NIRS cerebral oximetry is non-invasive and continuous “real-time” detection that guides clinicians in their interventions

Page 10: Cerebral oximetry

WHY CEREBRAL OXIMETRY?

• The brain: Complex and fragile system The brain weighs about 1200 to 1500

grams, accounts for only 2% of body weight, but receives 15% of total cardiac output and utilizes 20% of the oxygen

Elapsed time is critical in desaturation events

Page 11: Cerebral oximetry

Why is Cerebral Oximetry different from other parameters?

° Cerebral Oximetry (SctO2) measures regionalmetabolism and the balance of local

tissue oxygen supply and demand.

° SctO2 is a combined saturation measurementof arterial blood and venous blood.

° other monitoring modalities are invasive

Page 12: Cerebral oximetry

Cerebral Oxygen Saturation Consensus Level

Page 13: Cerebral oximetry

From Pulse Oximetry to Cerebral Oximetry•Both are based on Beer-Lambert’s Law that describes light absorption

• Pulse Oximetry: Only Two wavelengths! Using light loss (Systole – Diastole) to calculate only signals due to pulsation of arterial blood to derive arterial O2 saturation.

• Cerebral Oximetry: Looking into entire nonpulsating field to derive tissue O2 saturation. It requires precise light sources and more wavelengths to account for light lost from other elements -as melatonin- in addition to Oxy- and Deoxy-Hemoglobin.

Light lost due to pulsation of arterial blood

Light lost due to non-pulsating arterial blood

Light lost due to venous blood

PulseOximetry(2 λ)

CerebralOximetry(multiple λ)

Light lost from the Background:• Light lost due to other molecules• Light scattering lost

time

Inci

dent

Lig

ht In

tens

ity

Ligh

t Los

t Sys

tole

tim

eLi

ght L

ost D

iast

ole

Page 14: Cerebral oximetry

Arterial Pulse Oximetry

• Tells how much oxygen blood is carrying in arterial system (oxygenated Hb)

• Unable to determine how much oxygen is used (venous system)

• Unable to determine how different organs use oxygen (global vs regional) Accuracy about 2% (not accurate below 85%)

Page 15: Cerebral oximetry

SpO2 is100% arterial blood

i.e., all pulsing

SctO2 is mix of:• 30% arterial • 70% venous

• no pulse needed

Page 16: Cerebral oximetry

VeinsArteries

Tissue (Arterioles, capillaries, venules)

Pulse Oximetry SpO2 : Arterial

Oxygen Saturation Normal range:

90-100%

Cerebral OximetrySctO2: Regional TissueOxygen SaturationNormal Range: 60-80%

SjvO2: Jugular VenousOxygen SaturationNormal Range: 50-70%

Page 17: Cerebral oximetry

sats95%

VBAG65%

NIRS

Page 18: Cerebral oximetry

The Scientific Basis For CerebralOximetry NIRS

The Spectrum of Light The NIRS “Window” From 650nm to 900nm The Modified Beer-Lambert Law The Elliptical Photon Pathway In The Brain Chromophores In Cerebral Tissue Random Background Scattering Of Light More Wavelengths = Greater Accuracy Depth Of Penetration

Page 19: Cerebral oximetry

The Spectrum of Light

• The term “spectrum” for the colors produced by a prism,

• colors in sunlight can be recombined to make white light.

Sir Isaac Newton1642-1727

Page 20: Cerebral oximetry

Microwave Infrared Near-infrared Visible Ultraviolet X-ray Gamma ray

Window for NIRS Between 650 and 950 Nanometers

Page 21: Cerebral oximetry
Page 22: Cerebral oximetry

Fundamental Principle of Oximetry: Beer – Lambert Law

Page 23: Cerebral oximetry

 Emitter-detector pairs showing the banana-shaped paths of light.

Page 24: Cerebral oximetry

How Does Light Penetrate Bone?

Near infrared light and some wavelengths of visible red light easily penetrate human tissue,including bone Light is used to noninvasively interrogate gray matter in the frontal portion of the cerebral cortexThe depth of the light path is based on the separation of light source to detector and the strength of the light source

Page 25: Cerebral oximetry

Chromophores In Cerebral Tissue

NIRS is based on specific chromophores absorption of near infra red light for various wavelenghths. In case of cerebral tissues, these chromophores are: oxygenated hemoglobin, deoxygenated hemoglobin and oxidized cytochrome (Caa3)

Page 26: Cerebral oximetry

Other Chromophores

• There are a number of chromophores which absorb light in the same region as oxy-Hb and deoxy-Hb

• Interference caused by these chromophores must be detected and accounted for to get accurate SctO2 measurements

wavelengths for CO-, Met-, total-Hb, Skin pigmentation

Page 27: Cerebral oximetry

CADAVERS: NIRS Oxygenated Hb Showed a 90% DecreaseWithin 7h of Death, Continued Decrease in from 7 to 96 h.

Page 28: Cerebral oximetry

Absorption spectra for oxygenated haemoglobin (HbO2), deoxygenated haemoglobin (Hb), Caa3, melanin, and water (H2O) over wavelengths in NIR range. Note the relatively low peak for Caa3.

Page 29: Cerebral oximetry

Commercial cerebral NIRS devices currently utilize wavelengths in the 700–850 nm range to maximize separation between Hb and HbO2. The presence of melanin as found in human hair can significantly attenuate Hb, HbO2, and Caa3 signals.

Abs

orpt

ion

Page 30: Cerebral oximetry

Light Absorbance of Hemoglobin

Page 31: Cerebral oximetry
Page 32: Cerebral oximetry

Arterial Pulse Oximetry

Page 33: Cerebral oximetry

• Coronary artery bypass grafting (CABG) on/off pump

• Isolated valve replacement (AVR, MVR)• Combined valve replacement and CABG• Aortic arch replacement with DHCA• Ventricular assist device (LVAD, RVAD,

BiVAD)• Cardiac transplantation

Clinical Applications Of Cerebral Oximetry

Page 34: Cerebral oximetry

• Liver transplantation• ECMO• Carotid endarterectomy• “Beach Chair” positioning surgery (e.g.,

shoulder surgery)• Neonatal cyanotic heart defects• High-frequency ventilation• Bowel Ischemia (Neonates)

Page 35: Cerebral oximetry

Sizes are: Neonate, pediatric, adult

Page 36: Cerebral oximetry

Somatic Oximetry

Page 37: Cerebral oximetry
Page 38: Cerebral oximetry

Brain Areas Accessible by NIRSLIMITATIONS: What Part of the Brain is Involved

Page 39: Cerebral oximetry

LIMITATIONS: What Part of the Brain is Involved

Page 40: Cerebral oximetry

LIMITATIONS: What is the cause

Know how Different Organs Behave Under Different Conditons

Ischemia (Poor Flow)

Vs

Hypoxia (No Oxygen)

Vs

Venous Congestion

Page 41: Cerebral oximetry

Old Technology uses two wave lengths, with no absolute normal values and theirAlgorithms are only designated for adults

Page 42: Cerebral oximetry

AccurateImprecise AccuratePrecise InaccuratePrecise InaccurateImprecise

Precision and Accuracy

Standard Finger Oxygen Sat =2%Old technology NIRS: Two channels 9.72%Recent technology NIRS : Five channels 3%

Page 43: Cerebral oximetry