Upload
s-m
View
35
Download
3
Embed Size (px)
Citation preview
DIFFERENT METHODS OF MEASURING BLOOD PRESSURE
Stephen Hales
• 1727- First Blood Pressure measurements by inserting glass tubes inside arteries of a Mare and recording pressure from the column of blood that rose
• Forssmann, Counard and Richard
• Nobel Prize 1956
• Procedure of cardiac catherisation
MEASURING BLOOD PRESSURE
BP MEASUREMENT
DirectIndirect
TonometricUltrasonicOscillometricAuscultatory
HybridAneroidMercury
Palpatory
Palpatory Method
• Rough Estimation of Systolic Pressure
• Carotid Femoral radial >70 mm Hg
• Carotid and Femoral > 50 mm Hg
• Carotid >40 mm Hg
Deakin CD, Low JL (September 2000). Accuracy of advanced trauma life support guidelines for predicting systolic blood pressure using carotid, femoral and radial pulses: Observational ; BMJ 321 (7262): 673–4.
Direct Method
• Direct Intra arterial Measurement of Pressure using a pressure transducer
• Systolic is recorded 5-10 mm Hg higher and Diastolic is 5-10 mm lower than non- invasive techniques
• Gold Standard• Accurate beat-to-beat monitoring
Matthew Ward, Jeremy A Langton,
Disclosures Cont Edu Anaesth Crit Care and Pain. 2007;7(4):122-126.
Mercury Sphygmomanometer
• Samuel Siegfried Karl Ritter von Basch (1881)
• Riva Rocci (1896)
• Modern Version by Harvey Cushing (1901)
GOLD STANDARD IN CLINICAL PRACTICE
Aneroid Sphygmomanometer
• No Mercury• Metal Bellow and
Lever system• Inaccurate if not
calibrated
.Mion D, Pierin AM. How accurate are sphygmomanometers? J Hum Hypertens. 1998; 12: 245–248 Yarows SA, Qian K. Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience. Blood Press Monit. 2001; 6: 101–106.
Hybrid BP monitors
• Combine features of mercury and aneroid sphygmomanometers
• Mercury replaced by Electronic Pressure Gauge
• Auscultatory method used
PHYSICS OF BP MEASUREMENT
• Reynolds number
Re = ρ D v
μ
ρ- Density of the fluidD- Diameter of the vesselV- Velocity of the fluidμ- Viscosity of the fluid
Re < 2100 – Laminar flow
Re > 4000- Turbulent flow
Korotkoff SoundsPhase 1 Phase 2 Phase 3 Phase 4 Phase 5
Silence Tapping Soft Swishing
Crisp Blowing Silence
Cuff pressure
SYSTOLIC DIASTOLIC
Oscillometric Sphygmomanometer
• Records Oscillations in BP cuff
• Blood pressure determined by an algorithm
• Calibration Must
• Ambulatory and Home BP
Auscultatory Vs Oscillatory
Other Techniques
• Ultrasonic probes under cuff• Tonometry• Doppler
HOW TO MEASURE BLOOD PRESSURE IN
CLINICAL SETTING
Subject Position
• Quitely Seated atleast for 5 min in a chair• Feet on the floor• Arms at Level of Right Atrium (mid point of
sternum)• Avoid Caffeine, Exercise and Smoking 30
min prior to recording
Body Position
• Back supported
• Legs Uncrossed
• Arm supported
• Cuff at level of Right Atrium- Midpoint of the Sternum
• Arm to be at level of right atrium even when patient is supine (P=hρg)
• Support with pillow
Cuff Size
• Length: Width ratio ideally 2:1
• Tubing at least 70 cm long
Approx Size of Cuff
Length Width
Arm Circumference
Cuff Cuff Size
Upto 10 cm Newborn 4X8 cm
11-15 cm Infant 6X12 cm
16-21 cm Child 9X18 cm
22-26 cm Small Adult 12 X22 cm
27-34 cm Adult 16X30 cm
35-44 cm Large Adult 16X36 cm
45-52 cm Adult Thigh 16X 42 cm
Cuff Placement
• Midline of bladder of cuff over arterial pulsation
• Lower end 2-3 cm above cubital fossa
• Midthigh level in lower limb
Instrument at Level of Cuff
Position of Eye of Observer
• Eye at level of upper meniscus of mercury column
Inflation and Deflation
• Initial Inflation atleast 30 mm above point where radial pulse disappears
• Deflation at 2-3 mm per second• First and Last sounds to be taken as
systolic and Diastolic Pressures• Column to be read to nearest 2mm Hg
Number of Measurements
• Minimum of two readings at least intervals of atleast one min
• Average of both readings to be taken
• If diff > 5 mm Hg, more readings to be taken and averaged out
Both Arm Measurements
• Ideally, first visit should include measurement of blood pressure in both arms
• If Consistent difference in measurements without identifiable cause, higher reading should be used as baseline
Blood Pressure in Paediatric Age Group
• BP in all four limbs to be checked on first visit
• Auscultatory method satisfactory • Doppler• Oscillometric Most accurate • Palpatory method for systolic pressure • Flush method for infants – Systolic only
Self Monitoring at Home
• Useful in White Coat Hypertension• Smokers
Continuous Ambulatory BP
• White Coat Hypertension• Drug Resistance• Hypotension with Anti hypertensive
therapy• Episodic Hypertension• Autonomic Dysfunction
TAKE HOME MESSAGE
• Comfortable Sitting position of Subject• Appropriate Cuff Size• Calibrated Instrument• Proper Inflation and Deflation Procedure• Average of Readings