Wiring DiagramFoxboro (Invensys) 870ITPH
“IMPORTANT WARNINGS”
1
2
3
THE WIRING SHOWS A SENSOR WITH A SOLUTION GROUND. THIS OPTION IS DESIGNATED WITH WITH “GT”, “UT” OR “LT” IN THE MODEL NUMBER.e.g. (B-V-546-X-LT-X-X-X-X-X-X)
SENSORS WITHOUT A SOLUTION GROUND HAVE “FT”, “ST” OR “DT” IN THE MODEL NUMBER. e.g. (B-V-546-X-DT-X-X-X-X-X-X)
SENSORS WITHOUT A SOLUTION GROUND REQUIRE THAT SENSOR DIAGNOSTICS BE TURNED OFF IN THE TRANSMIT-TER AND A JUMPER IS INSTALLED AS SHOWN BELOW.
FOR SENSORS WITHOUT TEMPERATURE COMPENSATION DISREGARD TC WIRES. PROGRAM ANALYZER FOR MANUAL TEMPERATURE COMPENSATION.
FOR ORP/REDOX SENSORS COAX CORE WILL BE LABELED AS “ORP IN” AND WIRING IS THE SAME.
ANALYZER CONFIGURATION & SETTINGS
• WIRING
Sensor Lead Analyzer Terminal“TC IN” - Red (1) RTD Measurement“TC RET” - Green (2) RTD Return“TC SNSE” - Black (2A) RTD 3-Wire“pH IN” - Coax Core Shield (3) Measurement Electrode“pH SHLD” - Coax Shield (3A) Meas. Screen (Shield) “SOL GRND” - Clear (Drain) (4) Solution Earth Ground“REF” - White (5) Reference Electrode
• ALL ANALYZERS ACCEPT PT100, PT1000, OR 3.01kΩ RTD TEMPERATURE INPUTS.
• SELECT TEMPERATURE COMPENSATION MODE IN ANALYZER BASED ON THE TYPE OF RTD
BARBEN SENSOR: “E” TERMINATIONe.g. (B-V-546-X-XX-X-X-X-E-XX)
BAR
BEN
AN
ALYT
ICAL
pH /
OR
P SE
NSO
R54
6/54
7/56
7/55
1/55
4
SENSOR CABLE ORB39 EXTENSION CABLE
2
RTD
ME
AS
UR
EM
ENT
RTD
RE
TUR
N
RTD
3-W
IRE
ME
AS
UR
EM
ENT
ELE
CTR
OD
E
ME
AS
. SC
RE
EN
(SH
IELD
)
SO
LUTI
ON
EA
RTH
GR
OU
ND
RE
FER
EN
CE
ELE
CTR
OD
E
RE
F. E
LEC
. SC
RE
EN
(SH
LD)
SEN
SO
R P
OW
ER
+
SEN
SO
R P
OW
ER
-
DIA
GN
OS
TIC
NO
T U
SE
D
“TC
IN” -
Red
“TC
RET”
- Gr
een
“TC
SNSE
” - B
lack
"pH
IN" C
oax C
ore
"pH
SHLD
" Gra
y
"SOL
GRN
D" C
lear
"REF
" Whit
e
1
3
SOLUTIONEARTH GROUND
REFERENCEELECTRODE
"REF
" Whit
e
Jum
per
Wiring DiagramFoxboro (Invensys) 870ITPH
Barben Analytical reserves the right to make technical changes or modify the contents of this document without prior notice. We reserve all rights in this document and in the subject matter and illustrations contained within.
© 2015, by AMETEK, Inc. All rights reserved • FOXY_870ITPH_WD_RevA • August, 2015USA • BELGIUM • CHINA • SINGAPORE
Toll Free +1(800)993-9309 • Phone +1(775)883-2500 • Fax +1(775)297-4740 [email protected] • www.BarbenAnalytical.com
Cert. No. 43271ISO 9001:2008
“IMPORTANT WARNINGS”
1
2
3
THE WIRING SHOWS A SENSOR WITH A SOLUTION GROUND. THIS OPTION IS DESIGNATED WITH WITH “GT”, “UT” OR “LT” IN THE MODEL NUMBER.e.g. (B-V-546-X-LT-X-X-X-X-X-X)
SENSORS WITHOUT A SOLUTION GROUND HAVE “FT”, “ST” OR “DT” IN THE MODEL NUMBER. e.g. (B-V-546-X-DT-X-X-X-X-X-X)
SENSORS WITHOUT A SOLUTION GROUND REQUIRE THAT SENSOR DIAGNOSTICS BE TURNED OFF IN THE TRANSMIT-TER AND A JUMPER IS INSTALLED AS SHOWN BELOW.
ANALYZER CONFIGURATION & SETTINGS
• WIRING
Sensor Lead Analyzer Terminal“TC IN” - Red (1) RTD Measurement“TC RET” - Green (2) RTD Return“TC SNSE” - Black (2A) RTD 3-Wire“pH IN” - Coax Core Shield (3) Measurement Electrode“SOL GRND” - Clear (Drain) (4) Solution Earth Ground“REF” - White (5) Reference Electrode
• ALL ANALYZERS ACCEPT PT100, PT1000, OR 3.01kΩ RTDTEMPERATURE INPUTS.
• SELECT TEMPERATURE COMPENSATION MODE IN ANALYZERBASED ON THE TYPE OF RTD
BARBEN SENSOR: “C” TERMINATIONe.g. (B-V-546-X-XX-X-X-X-C-XX)
BARB
EN A
NALY
TICA
LpH
/ O
RP S
ENSO
R54
6/54
7/56
7/55
1/55
4
SENSOR CABLE ORB39 EXTENSION CABLE
RTD
MEA
SUR
EMEN
T
RTD
RET
UR
N
RTD
3-W
IRE
MEA
SUR
EMEN
T EL
ECTR
OD
E
MEA
S. S
CR
EEN
(SH
IELD
)
SOLU
TIO
N E
ARTH
GR
OU
ND
REF
EREN
CE
ELEC
TRO
DE
REF
. ELE
C. S
CR
EEN
(SH
LD)
SEN
SOR
PO
WER
+
SEN
SOR
PO
WER
-
DIA
GN
OST
IC
NO
T U
SED
“TC
IN” -
Red
“TC
RET”
- Gr
een
“TC
SNSE
” - B
lack
"pH
IN" C
oax C
ore
"SOL
GRN
D" C
lear
"REF
" Whit
e
1
2 3
SOLUTIONEARTH GROUND
REFERENCEELECTRODE
"REF
" Whit
e
Jum
per
FOR SENSORS WITHOUT TEMPERATURE COMPENSATION DISREGARD TC WIRES. PROGRAM ANALYZER FOR MANUAL TEMPERATURE COMPENSATION.