WHO Method Validation

  • Upload
    milonhg

  • View
    253

  • Download
    0

Embed Size (px)

Citation preview

  • 8/12/2019 WHO Method Validation

    1/34

  • 8/12/2019 WHO Method Validation

    2/34

    The prequalification programme --Assessor's training | 19-20 January 20112 |

    Outline

    HPLC methodology

    - Content of HPLC test procedure

    - System Suitability Testing (SST)

    - Relative Response actor (RR)

    !alidation of HPLC method

    case study

  • 8/12/2019 WHO Method Validation

    3/34

    The prequalification programme --Assessor's training | 19-20 January 20113 |

    Information Sources

    "# C"$R revie%er guideline for validation ofchromatographic methods (&'')

    H* TRS '+, #ppendi .#nalytical /ethod !alidation

    0112

    3CH 40(R&) 0115

    Compendial 6eneral Chapters

    /ethods and !alidation presentation7Lynda Paleshnui8

  • 8/12/2019 WHO Method Validation

    4/34

    The prequalification programme --Assessor's training | 19-20 January 20114 |

    High Performance Li9uid Chromatography (HPLC)

    HPLC is a separation techni9ue based on a solid stationary phase and

    a li9uid mobile phase: Separations are achieved by partition;

    adsorption; or ion-echange processes; depending upon the type of

    stationary phase used:

    Chiral

    3on--echange

    3on--paire eclusion

    The reversed-phase HPLC %ith ?! detectionis most commonly usedform of HPLC; is selected to illustrate the parameters of HPLC

    method and validation:

  • 8/12/2019 WHO Method Validation

    5/34

    The prequalification programme --Assessor's training | 19-20 January 2011 |

    A flow scheme for HPLC

  • 8/12/2019 WHO Method Validation

    6/34

    The prequalification programme --Assessor's training | 19-20 January 2011! |

    Content of HPLC test procedure

    #ny analytical procedure submitted should be described in

    sufficient detail; includes@

    Preparation of mobile phase

    Chromatographic condition@

    7 Column@ type (e:g:; C&A or CA); dimension (length; innerdiameter); particle si>e (&1Bm; 5 Bm)

    7 "etector@ %avelength

    7 3nection volume

    7column T

    7 flo% rate;

  • 8/12/2019 WHO Method Validation

    7/34

    The prequalification programme --Assessor's training | 19-20 January 2011" |

    Content of HPLC test procedure

    $lution procedure@ isocratic or gradient elution

    Preparation of standards and samples

    *peration procedure@ se9uence of inections

    System suitability testing (SS) and criteria

    Calculations

    4*S 0:+:R:0analytical procedures and validation summaries

  • 8/12/2019 WHO Method Validation

    8/34

    The prequalification programme --Assessor's training | 19-20 January 2011# |

    Compendial methods

    hen claim a compendial method; there should be no change in@

    The type of column i:e the stationary phases

    "etector %avelength

    Components in /obile phase

    System suitability testing and criteria

    #dustments to ratio of components in mobile phase; flo% rate; columntemp; dimension of column; particle si>e (reduction only); may be

    necessary to achieve the system suitability criteria: The allo%ablevariations for each parameter; see 3nt:Ph &:&: or ?SP generalchapter D20&E:

  • 8/12/2019 WHO Method Validation

    9/34

    The prequalification programme --Assessor's training | 19-20 January 20119 |

    System suitability testin! "SST#

    Precision@7 #ssay@ RS" F&G (#P3) or F 0G (PP); n 5

    7 3mpurities@ in general; RS" F 5G at the limit level; up to &1G or

    higher at L*4; n 2

    Resolution (R)@ E0

  • 8/12/2019 WHO Method Validation

    10/34

    The prequalification programme --Assessor's training | 19-20 January 201110 |

    System suitability testin! "SST#

    Tailing factor

  • 8/12/2019 WHO Method Validation

    11/34

    The prequalification programme --Assessor's training | 19-20 January 201111 |

    System suitability testin! "SST#

    =umber of theoretical plates (=)@ column efficiency 0111

    6radient elution is one %ay to increase the =

  • 8/12/2019 WHO Method Validation

    12/34

    The prequalification programme --Assessor's training | 19-20 January 201112 |

    System suitability testin! "SST#

    # SST should contain@

    or #ssay@

    precision I one or more other parameter

    or impurity test@

    resolution I precisionI one or more other parameter

  • 8/12/2019 WHO Method Validation

    13/34

    The prequalification programme --Assessor's training | 19-20 January 201113 |

    $elati%e $esponse &actor "$$

    'uantitation of Impurities#gainst impurity RSJs@ %hen reference standard available

    #gainst #P3 itself

    Relative response factor should be considered

  • 8/12/2019 WHO Method Validation

    14/34

  • 8/12/2019 WHO Method Validation

    15/34

    The prequalification programme --Assessor's training | 19-20 January 20111 |

    $elati%e $esponse &actor "$$

    $ifampicine)

    y K*+,*+- I :'2+

    $ifampicine 'uinone)

    y K -.,+/0 I &:&5

    RRK -.,+/0< *+,*+-

    K1:A

  • 8/12/2019 WHO Method Validation

    16/34

    The prequalification programme --Assessor's training | 19-20 January 20111! |

    $elati%e $esponse &actor "$$

    To revie%@

    a) RR calculation; and

    b) if RR is properly used in the final calculation for G

    impurity

    3f RR %ithin 1:A-&:0; correction may not be necessay

    Correction factorK &

  • 8/12/2019 WHO Method Validation

    17/34

    The prequalification programme --Assessor's training | 19-20 January 20111" |

    $e%ie1 points for HPLC method

    is the analytical proceduredescribed in detail includin! all the parameters 2 is SS1ell defined to ensure the consistency of system performance2

    he preparation of solutions)

    7 assay) concentration of reference standard should be close to the samplesolution

    7 impurities) concentration of the reference standards should be close tothe limit

    he 1ay of 3uantitation of impurities

    In case API is used as the reference( $$& should be used or 4ustification ofe5clusion should be pro%ided,

    o chec6 the determination of $$&( chec6 the correction of calculation ofimpurities

    confirm7complete the 'OS -,*,$,-

  • 8/12/2019 WHO Method Validation

    18/34

    The prequalification programme --Assessor's training | 19-20 January 20111# |

    Validation 8 compendial methods

    Assay 8 API

    =o validation generally re9uired: $ception@ specificity for maor impuritiesnot in the monograph:

    Assay 8 &PP

    Specificity; accuracy and precision (repeatability),

    Purity 8 API and &PP

    ull validation for specified impurities that are not included in themonograph (specificity; linearity; accuracy; repeatability; intermediateprecision; L*"

  • 8/12/2019 WHO Method Validation

    19/34

  • 8/12/2019 WHO Method Validation

    20/34

    The prequalification programme --Assessor's training | 19-20 January 201120 |

    Specificity

    lan8 solution to sho% no interference

    Placebo to demonstrate the lac8 of interference from ecipients

    Spi8ed samples to sho% that all 8no%n related substances are

    resolved from each other

    Stressed sample of about &1 to 01G degradation is used to

    demonstrate the resolution among degradation products

    7 Chec8 pea8 purity of drug substance by photodiode array detector (P"#)@ eg

    purity angle is lo%er than the purity threshold:

    Representative chromatograms should be provided %ith time scale

    and attenuation indicated

  • 8/12/2019 WHO Method Validation

    21/34

    The prequalification programme --Assessor's training | 19-20 January 201121 |

    Linearity 7 $an!e

    The %or8ing sample concentration and samples tested foraccuracy should be in the linear range (concentrations

    !s: Pea8 areas)

    Minimum : concentrations

    "ilute of stoc8 solution or separate %eighings

  • 8/12/2019 WHO Method Validation

    22/34

  • 8/12/2019 WHO Method Validation

    23/34

    The prequalification programme --Assessor's training | 19-20 January 201123 |

    Linearity 7 $an!e

    Correlation coefficient "r#

    API) B ;,//0

    Impurities) B ;,//

    y9Intercept and slope should be indicated to!ether 1ith

    plot of the data

  • 8/12/2019 WHO Method Validation

    24/34

  • 8/12/2019 WHO Method Validation

    25/34

    The prequalification programme --Assessor's training | 19-20 January 20112 |

    Accuracy

    3mpurities@ #P3

  • 8/12/2019 WHO Method Validation

    26/34

    The prequalification programme --Assessor's training | 19-20 January 20112! |

    Precision

    System precision@7 by multiple inections (n 5) of a homogeneous sample

    (standard solution):

    7 RS" F &G is recommended for assayM

    7 RS" F 5G is recommended for related substances (referencestandards at the limit)

    7 3ndicates the performance of the HPLC system

    7 #s a system suitability test

  • 8/12/2019 WHO Method Validation

    27/34

    The prequalification programme --Assessor's training | 19-20 January 20112" |

    Precision

    $epeatability "method precision#7 Multiple measurements of a sample by the same analyst

    7 A minimum of . determinations at the test concentration ".

    times of a sin!le batch#( or

    7 * le%els "0;

  • 8/12/2019 WHO Method Validation

    28/34

    The prequalification programme --Assessor's training | 19-20 January 20112# |

    Precision

    Intermediate precision "part of ru!!edness#7 est a sample on multiple days( analysts( e3uipments

    7 $epeat the method precision by different analyst in

    different e3uipment usin! different lot of column on

    different days

    7 $S? should be the same re3uirement as method precision

    $eproducibility "inter9laboratory trial#

    7 Not re3uested in the submission

  • 8/12/2019 WHO Method Validation

    29/34

    The prequalification programme --Assessor's training | 19-20 January 201129 |

    LO?7LO'

    signal to noise ratio@ L*" +@& ; L*4 &1@&7 /ay vary %ith lamp aging; model

  • 8/12/2019 WHO Method Validation

    30/34

    The prequalification programme --Assessor's training | 19-20 January 201130 |

    LO?7LO'

    LO?) belo1 the reportin! threshold

    LO') at or belo1 the specified limit

    =ot re9uired for assay

  • 8/12/2019 WHO Method Validation

    31/34

    The prequalification programme --Assessor's training | 19-20 January 201131 |

    $obustness

    The methodOs capability to remain unaffected by small butdeliberate variationsin method parameters

    7 3nfluence of variations of pH in a mobile phase

    7 3nfluence of variations in mobile phase composition

    7 "ifferent columns (different lots and

  • 8/12/2019 WHO Method Validation

    32/34

  • 8/12/2019 WHO Method Validation

    33/34

    The prequalification programme --Assessor's training | 19-20 January 201133 |

    Conclusion

    HPLC methods play a critical role in analysis ofpharmaceutical product

    !alidation of HPLC should demonstrate that the method

    is suitable for its intended use

    Revie% the information in dossier against 4*S 0:+:R:0

    "ata for acceptance; release; stability %ill only be

    trust%orthy if the methods used are reliable

  • 8/12/2019 WHO Method Validation

    34/34

    The prequalification programme --Assessor's training | 19-20 January 201134 |

    Case Study

    Case Study &--HPLC/ethod:doc

    4*S 0:+:R:doc

    Case Study 0 -- !alidation of HPLC/ethod:doc

    http://case%20study%201--hplc%20method.doc/http://case%20study%201--hplc%20method.doc/http://qos%202.3.r.doc/http://case%20study%202%20--%20validation%20of%20hplc%20method.doc/http://case%20study%202%20--%20validation%20of%20hplc%20method.doc/http://case%20study%202%20--%20validation%20of%20hplc%20method.doc/http://case%20study%202%20--%20validation%20of%20hplc%20method.doc/http://qos%202.3.r.doc/http://case%20study%201--hplc%20method.doc/http://case%20study%201--hplc%20method.doc/