4

Click here to load reader

Welch Allyn Propaq CS Vital Signs Monitor€¦ · • ®NIBP with Welch Allyn FlexiPort technology ... Welch Allyn Propaq® CS Vital Signs Monitor. ... without expansion module

Embed Size (px)

Citation preview

Page 1: Welch Allyn Propaq CS Vital Signs Monitor€¦ · • ®NIBP with Welch Allyn FlexiPort technology ... Welch Allyn Propaq® CS Vital Signs Monitor. ... without expansion module

•3-or5-leadECG,heartrate•MasimoSET®orNellcor®OxiMax®pulseoximetry•NIBPwithWelchAllynFlexiPort®technology•Pulserate•Respiration•Temperature•Capnography(sidestreamand/ormainstream)*•Invasivebloodpressure(oneortwochannels)*•Fullpatientalarms•Apneaalarming•Trendingonallparameters•Programmabledefaultsondisplayedparametersandalarmlimits

•Printer•RS423connectivityport

*Optimal parameters

Continuousvitalsignsmonitoringforproceduralsedationandrecovery.

WelchAllyn

Propaq®CSVitalSignsMonitor

Page 2: Welch Allyn Propaq CS Vital Signs Monitor€¦ · • ®NIBP with Welch Allyn FlexiPort technology ... Welch Allyn Propaq® CS Vital Signs Monitor. ... without expansion module

Propaq® CS | Specifications

General•Adult/Pediatric/Neonatalfactory/custompatientmodes

•Electrosurgeryinterferencesuppression(ESIS):allchannelsexceptImpedanceRespiration

•Allchannelselectricallyisolated•Communicationsoption

ECG•Automatic3-and5-leaddetection•Selectable leads:I,II,III,aVR,aVLaVF,V•Leadfaultdetectionwithauto-reconfigure• Bandwidth:0.5to40Hzmonitormode,0.05to40Hzextendedmode(Adult);0.5to120Hzmonitormode,0.05to120Hzextendedmode(Pediatric/Neonate)

•AAMI6-pinconnector•Heart rate measurement range:25to300bpm(display)

• Heart rate accuracy:±3bpmor3%(whicheverisgreater)

•Pacerdetectionanddisplay•Defibsyncconnector•Real-timeECGoutput,delay<3msec•ESUanddefibrillatorprotectedwhenusedwithspecificcables

•Sweep speed:12.5,25,and50mm/sec•User-selectable size (mV/cm):4,2,1,0.5,0.2• QRS tone:high/med/low/off•QRS detector range:

Adult or pediatric mode:0.22to5.0mV(RTI)Neonatal mode:0.1to5.0mV(RTI)Neonatal width:40to120msec(Duration)Pediatric width: 40to120msec(Duration)Adult width:70to120msec(Duration)

Noninvasive Blood Pressure (NIBP)•Oscillometricmethod•Automatic(intervalsbetween1and60min)andmanualmodes

•QuickactionNIBPStart/Stopbutton•TurboCuf:5minofrepeatedNIBPreadings•Systolic,diastolic,meandisplay•On-screenmanometer•Storeandviewalldata(upto128readings)•LargeNIBPnumericdisplayformat• Standard cuff sizes:Neonate#1-#5,Infant,Child,SmallAdult,Adult,LargeAdult,Thigh

•Cuffoverpressureprotection• Typical measurement time: 30to45secs•Smartcuf®NIBPtechnology

NIBPPatient mode-specific operation:Neonate mode• Initial cuff inflation pressure:90mmHg• Maximum allowable cuff pressure: 132mmHg• Systolic range:25to120mmHg• Diastolic range:10to105mmHg• Mean range:10to110mmHg

Pediatric mode• Initial cuff inflation pressure:120mmHg• Maximum allowable cuff pressure:170mmHg•Systolic range: 30to160mmHg•Diastolic range:15to130mmHg•Mean range:15to140mmHg

Adult mode• Initial cuff inflation pressure:160mmHg•Maximum allowable cuff pressure:270mmHg• Systolic range:30to260mmHg• Diastolic range:20to235mmHg•Mean range:20to255mmHg

Temperature•TwoYSI400/700compatiblechannels• Range:0to50°C• Display: T1,T2andDeltatemp•Units: °Cor°F• Accuracy:±0.1°C(10to50°C),±0.2°C(0to10°C)

IBP•Oneortwochannels•Pressure range:-30to300mmHg• Pulse rate measurement range:25to250bpm•Display:5scalesorsemiautomaticrescaling• Format:Sys/Dia/Mean,userselectable• User-selectable labels: P1,P2,ART,PA,CVP,ICP;UA,UVNeonatemodeonly

• Transducer requirements:5µV/V/mmHg• Zero adjustment:±200mmHgincludingtransduceroffset

•Numeric pressure accuracy:±2mmHgor2%ofreading(whicheverisgreater),plustransducererror

Impedance Pneumography Respiration• Two user-selectable leads:RA-LA,RA-LL•Automaticcardiovascularartifact(CVA)rejection•Respiration rate range:

Adult/Pediatric modes: 0(apnea),2to150bpmNeonate mode: 0(apnea),3to150bpm

•Respiration detection threshold:140milliohmsor2XCVA,(whicheverisgreater)

•Respiration rate accuracy: ±2breaths/minor±2%(whicheverisgreater)

• Apnea alarm delay settings: Adult/Pediatric modes:6,10,15,20,25,30secNeonate mode:6,10,15,20sec

• Sweep speed:3.13,6.25,12.5mm/sec

Pulse Oximetry (SpO2)Masimo® •MasimoSET®sensors•Waveformandpulseamplitudemeter•Pulsetonepitchindicator• Saturation range:1to100%

Pulse Oximetry (SpO2)Masimo® (con’t)• Probe accuracy: 25to41°C

No Motion:Adult/Pediatric:70to100%±2digits,0to69%unspecifiedNeonate: 70to100%±3digits,0to69%unspecifiedDuring Motion: Adult/Pediatric/Neonate: 70to100%±3digits,0to69%unspecified

• Pulse rate measurement range (Adult Pediatric/Neonate):No Motion:26to239bpmDuring Motion:26to239bpm

• Pulse rate accuracy: No Motion:±3digitsDuring Motion:±5digits

Nellcor®

•NellcorOxiMax®sensors•Waveformandpulseamplitudemeter•Pulsetonepitchindicator• Saturation range:0to100%• Probe accuracy: 70to100%,28to42°C,DS-100A:±3

• Pulse rate measurement range:25to250bpm• Pulse rate accuracy:±3bpm

CO2 Option Display• Screen display:CO2waveformandEtCO2andINCO2(wheninalarm)numerics

• Waveform scale:(maximum)0to100mmHg,0to14%,0to14kPa

•Numeric range display: EtCO2:0to99mmHg,0to13.2kPa,0to23.1%;INCO2:8

ato25mmHg,1.1ato5kPa,1.1ato5%

•Units:mmHg,kPa,%;user-selectable•Sweep speed: 3.13,6.25,12.5mm/sec;user-selectable

• Response modes: Fast: 15secsamplingtimeperiod;Normal: 30secsamplingtimeperiod;Slow:45secsamplingtimeperiod

• Gas compensation:Off:CO2value=calculatedCO2value;O2>50%,No N2O:CO2value=calculatedCO2valuex1.03;N2O > 50%:CO2value=calculatedCO2valuex0.952

• Alarm limit ranges: EtCO2: 0to99mmHg,0to13.2kPa,0to13.2%;INCO2:2to25mmHg,0.2to5kPa,%(nolowerlimit)

• Resolution:1mmHg• Accuracy:

Mainstreamb: 0to30mmHg,±3mmHg31to99mmHg,±10%ofvalueSidestreamc:0to30mmHg,±3mmHg31to99mmHg,±10%ofvalue

• Altitude Error: ±0.4%/1,000ft(304.8m)

Page 3: Welch Allyn Propaq CS Vital Signs Monitor€¦ · • ®NIBP with Welch Allyn FlexiPort technology ... Welch Allyn Propaq® CS Vital Signs Monitor. ... without expansion module

Breath Rate Display•Screen display:Numeric•Breath rate (BR) source: WhenCO2isactive,CO2istheBRsource.Otherwise,RESPfromECGistheRRsource

• Units:Breaths/Minute•Range:

Adult/Ped:0(apnea),2to150breaths/minNeonate:0(apnea),3to150breaths/min

•Resolution:±1breaths/min• Accuracy:±1breaths/minor±5%,whicheverisgreaterd

•Alarm limits range: Adult/Ped:2to150breaths/minNeonate: 3to150breaths/min

CO2 Performance•PerISO9918:1993(E)/EN864:1996

Apnea Alarms and Tickets• Apnea ticket:Settoautoprintafterapneaeventandafter1minutecontinuedapnea

•Apnea alarm accuracy:±2sec•Apnea delay setting:

Adult/Ped:6,10,15,20,25,30secondsNeonate:6,10,15,20seconds

Barometric Pressure• Pressure compensation:automatic•Operating range:-2,000to15,000ft,-610to4,572m(817to429mmHg)

• Screen display:Numeric(CO2StatusWindow)•Units:mmHgorkPaor%•Accuracy:±3mmHgor2.5%ofdifferencefromcalibrationpressure,whicheverisgreater

Sidestream CO2 Option (SSCO2)• Sensor type:Sidestream,internal• Principle of operation: Nondispersive,infrared,single-beam,single-path/wavelength,ratiometric

•Operating ambient temperature:5to40°C• Startup time:30secondstypical,3minutesmaximum

• Rise time:240ms(10%to90%)at175ml/min•Delay time: 1.12secondsmaximume•Total system response time:1.36secondsmax.(RiseTimeandDelayTime)e

•Calibration:Verifysemiannually,calibrateonlyasrequired

•Sampling chamber: Internal(replaceablebyservicetechnician)

•Pneumatic and exhaust system: Integral•Barometric pressure compensation: Automatic•BTPS, ATPS, STPD:CO2value=calculatedCO2valuex0.977

•Sampling line:7-footsamplingline,ID0.055in(1.4mm),forusewithdisposablesingle-usecannula(CO2onlyorCO2sampling/O2delivery)

Sidestream CO2 Option (SSCO2) (con’t)• Flow rate:90or175ml/min,user-selectableaLowerifinalarmbBasedontheseairwayconditions:sensortemperature=42°C,

airwayadaptertemperature=33°C,watervaporpressure=

38mmHg;standardgasmixture=CO2inbalanceair,fullyhydratedat

33°C;barometricpressure=760mmHgandflow=60ml/mincBasedonthefollowingadditionalairwayconditions:sampleline=7

ft,0.055inID(2.13m,1.4mmID);sampleflowrate=175ml/min;

respiratoryrate≥50breaths/min,stableto±3breaths/min;inspired/expiredtimeratio=1:2;

barometricpressure=760mmHgdForSidestreamCO2,thisappliesonlyforBR ≥50eBasedonthefollowingadditionalairwayconditions:

sampleline=7ft,0.055inID(2.13m,1.4mmID);

sampleflowrate=175ml/min

Mainstream CO2 Option (MSCO2)• CO2waveform,EtCO2,INCO2(INCO2wheninalarm),ApneaandBreathRatedisplay

• Waveformrisetime:<120msec(to90%afterstepchange)

Mainstream CO2 Sensor•MainstreamNDIRsingle-beam,singlepath/wavelength,ratiometric

•Warm-up time (CO2 sensor and monitor): 45sectypical,3minutesmaximum

•Automaticzeroing•Noroutinecalibration•Operating altitude:-2,000to15,000ft(817to429mmHg)

•Dimensions: 1.003in(H)x1.036in(W)x0.78in(L)(2.5x2.6x2.0cm)

• Weight:.53oz(15.0g)•Cable length:10ft(3.05m)nominal

CO2 Airway Adapters

•Disposable•Size:15mmID(meetsISOspecifications)•Single adult/pediatric airway adapter for

patients who weigh > 5 kg:Addeddeadspace:<6cc(0.37cubicin)

• Single-use low-deadspace airway adapter for patients who weigh < 5 kg:Addeddeadspace:<0.6cc(0.037cubicin)

Printer Option• Thermalsensitivedot• Numeric annotation:date,time,allactiveparameters,patientnameifconnectedtoAcuity

• Number of waveforms printed simultaneously: upto3

• Print speeds:6.25,12.5,25.0mm/sec

Printer Option (con’t)• Format: 53mmwideprintareaon60mmwidepaper• Operating modes:-Continuous:start/stopreal-timeprintoutswith allactivenumericsandupto3waveforms-Snapshot:8sec(32secforCO2/Resp)of immediatehistoryforallactivenumericsandup to3waveforms-Alarm print:20secondstotal/12secondsprior toalarmingparameter- Auto print: Snapshotprintoutevery15or30min,or1,2or4hours

- Trend print:OndemandorAutoTrendevery 4hours.Upto8hoursintabularformat. -OxyCRG trend print:Print-onlyfunction(CO2or Respmustbeinstalled):2minprintoutoftrended HR/PRandSpO2numericsandcompressedCO2or Respwaveforms(CO2isprioritysource),with annotationandMIN/MAXtabularreport -OxyCRG on alarm: Settoprintout60secafter HR/PRorSpO2alarmor75secafterBR/RRor apneaalarm -Apnea ticket: AutoprintHR/PR,SpO2and elapsedtimeafterapneaeventandafter1 minuteofcontinuedapnea -NIBP ticket: AutoprintNIBP,HR/PR,SpO2, CO2andRR/BRaftereachNIBPreading

Propaq® CS | Specifications

Page 4: Welch Allyn Propaq CS Vital Signs Monitor€¦ · • ®NIBP with Welch Allyn FlexiPort technology ... Welch Allyn Propaq® CS Vital Signs Monitor. ... without expansion module

8500S.W.CreeksidePlace,Beaverton,Oregon97008USA(p)503.530.7500tollfree800.289.2500(f)503.526.4200technicalsupport800.289.2501www.welchallyn.com

©2011WelchAllynSM4045RevA

Allrightsreserved.Acuity,PropaqandSmartcufareregisteredtrademarksofWelchAllyn.Othertrademarksmentionedareforidentificationpurposesonlyandareownedbytheirrespectivecompanies.

OnlyMasimoSETSpO2sensors,includingLNOPdurableadhesiveandNRvaluedisposablesensors,shouldbeusedwithMasimopulseoxi-metry.Masimo,MasimoSET,LNOP,andNRareregisteredtrademarksofMasimoCorporation.

OnlyNellcor®OxiMax®pulseoximetrysensorsshouldbeusedwiththeNellcorpulseoximetryoption.NellcorandOxiMaxareregisteredtrademarksofNellcorPuritanBennett,Inc.

AllPropaqCSvitalsignsmonitorscomestandardwithSmartcuftechnology,anewstandardinNIBPaccuracy.Thispatentedsoftwareoffersindustry-leadingperformanceandaccuracyinthepresenceofmotionartifacts.

Alarms• All parameters:upper/lowerlimits• All parameters: Adult/Pediatric/Neonatepatientmode-specificlimits

•Factorydefaultorprogrammablesettingsforallpatientmodes

•Alarm indicator:red•Alarm(s) off indicator: yellow•Audible alarm tone:high/med/low• Alarm suspend: 90seconds•One-buttonStatSetforallalarmlimits(exceptApneaDelay)

• EQ Alert:yellow

Trends• Tabularnumericformat• Allparameterstrended/viewable• Resolution: non-NIBPtrendsenteredevery2min• NIBPtrendsenteredaftereachreading• Duration:5hoursnon-NIBPtrends(upto150readings); 8hoursNIBPtrends(upto128readings)

• Pageup/downtrendview

Inservice Mode• Includestwosetsofsimulatedpatientdataincludingwaveforms,fortrainingandeducation

Color Active Matrix Display• Type:TFT(ThinFilmTransistor)LCDmodule• Resolution:640x480pixels,1pixel=R+G+Bdots•Active viewing area:6.73x5.10in(170.9x129.6mm)• Pixel pitch:0.0105in(0.267mm)•Viewing angle:U/D40°,R/L60°(typical),≥10:1contrastratio

• Contrast ratio: 150:1(typical);measuredindarkroomatcenterofscreen

•Display color: 18-bit(6bitsperprimarycolor)Waveforms and matching numerics:Predefined

• Luminance:200cd/m2(typical);measuredatsaturationpoint

•Response time:40ms(maximum);“whitetoblack”

Environmental•Operating temperature: 0to40°C(monitor),5to40°C(printer)

•Operating relative humidity:15%to95%,noncondensing(MILSTD810E),printer35%to85%,noncondensing(MILSTD810E)

• Operating altitude: -2,000to15,000ft(-610to4,572m)

• Shipping/storage temperature:-20to60°C• Shock:50g(monitor);30g(expansionmodule/printer)

• Random vibration: DesignedtomeetRTCADO-160D,categoryC(monitor)

• Electromagnetic compatibility (EMC):perEN60601-1-2,1993

• Drip-proofperIEC60529,levelIPX1(monitor)only, withoutexpansionmodule

Physical• Monitor (including handle):

Size:8.2in(H)x9.6in(W)x5.6in(D)(20.8x24.4x14.1cm)Weight: 7.6lb(3.4kg)

• Monitor with SpO2 Module: Size: 8.2in(H)x9.6in(W)x7.7in(D)(20.8x24.4x19.7cm)Weight: 10.8lb(4.9kg)

• Monitor with Expansion Module: (withPrinter/SpO2/CO2)Size: 11.4in(H)x9.6in(W)x7.7in(D)(28.8x24.4x19.7cm)Weight: 14.4lb(6.5kg)

Power•Rechargeablesealedleadacidinternalbatterypack

•Internalrecharger• Input voltage: 12to28VDC,25W•AC power adapter:100to120VAC,50to60Hz(NorthAmerica/Japan);220to240VAC,50to60Hz(International)

•Operating time on battery, typical*:MonitorwithSpO2option:5hours;MonitorwithExpansionModule(Printer,SpO2/CO2):3hours;Monitoronly: 2hours

• Battery recharge time: 6to8hours(whenmonitorOFF)

• Battery recharge time: 8to12hours(whenmonitorON)

Certifications/Standards• ComplieswithrelevantAAMI,IEC,EN,CSAandULstandards.CEMarkedaccordingtotheEuropeanMedicalDeviceDirective.CSACertificateofComplianceforuseintheU.S.andCanada.VAContractV797P-3486kDSCPContractSP0200-97-D-8021

* Fully charged, new battery at 25 °C, NIBP and Snapshot (if applicable) every 15 minutes, all accessories used.