Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Application Note
Using Advanced Polymer Chromatography with Quaternary Solvent Manager and Evaporative Light Scattering Detection for Gradient Polymer Elution Chromatography
Jennifer Gough, Will Martin, Neil J. Lander
Waters Corporation
Abstract
The APC with QSM is specifically designed for aggressive solvents used in the polymer
industry. The Waters ACQUITY Advanced Polymer Chromatography (APC) System is designed to
operate consistently using harsh solvents like THF, chloroform, and dimethyl sulfoxide. By adding
the flexibility of a quaternary pump to the industry-accepted APC system, polymer analysis
techniques are no longer limited to HPLC technology.
Benefits
p-QSM with aggressive solvent compatibility right out of the box.■
Reverse-phase, normal-phase, and size-based separations provide flexible analysis.■
The low dispersion APC with p-QSM uses less solvent than HPLC or GPC, which makes it more
sustainable
■
High resolution with sub-3-μm BEH column technology■
Introduction
The gradient polymer elution chromatography (GPEC) chromatogram in Figure 1 has a run time of
30 minutes using an Alliance HPLC (high-performance liquid chromatography) system. If this
method is transferred to an UltraPerformance Liquid Chromatography (UPLC) system, the run time
could be significantly shortened. However, most UPLC systems are not compatible with organic
solvents, such as tetrahydrofuran (THF), and are not robust enough to constantly use 100% organic
solvent.
The Waters ACQUITY Advanced Polymer Chromatography (APC) System is designed to operate
consistently using harsh solvents like THF, chloroform, and dimethyl sulfoxide. Originally, the APC
system was designed for gel permeation chromatography (GPC) using an isocratic pump. By adding
the polymer quaternary solvent manager (p-QSM) for gradient elution and an evaporative light
scattering detector (ELSD) to the APC system, GPEC can be accomplished in less than 10 minutes
(Figure 1).
Figure 1. GPEC comparison of HPLC and UPLC analysis times.
Experimental
System: APC with p-QSM
Column: XBridge BEH C8, XP, 2.5 μ
m, 3.0 × 100 mm (p/n
186006047)
Mobile phase A/B: Methanol
Mobile phase C/D: Tetrahydrofuran
Flow rate: 0.7 mL/min
Column temp.: 40 °C
Injection volume: 1 μL
Sample concentration: PMMA (1 mg/mL), PS
(0.12 mg/mL), PBD (0.5
mg/mL)
Gradient:
Sr.
No
Time Flow
(mL/min)
%A %B %C %D Curve
1 Initial 0.700 100.0 0.0 0.0 0.0 Initial
2 1.00 0.700 60.0 0.0 40.0 0.0 11
3 2.00 0.700 35.0 0.0 65.0 0.0 11
4 3.00 0.700 15.0 0.0 85.0 0.0 11
5 4.00 0.700 15.0 0.0 85.0 0.0 11
6 5.00 0.700 0.0 0.0 100.0 0.0 6
7 6.00 0.700 100.0 0.0 0.0 0.0 6
8 7.00 0.700 100.0 0.0 0.0 0.0 11
In order to precipitate a polymer onto the column, the polymer solubility in the gradient blend of
solvents needs to be determined. The accepted method in literature is to titrate a known
concentration of polymer solution (10–2 to 10–3 mg/mL) with a non-solvent (poor solvent). As the
titration progresses, the polymer solution will appear cloudy as the polymer aggregates and
precipitates: the polymer is repelled by the non-solvent and the polymer molecules are forced
together in aggregation. The volume fraction of the non-solvent is calculated using Equation 1,
where phi (F) equals the cloud point (CP) divided by 100, which equals the volume of the non-
solvent (Vns) divided by the sum of polymer volume, good solvent volume, and non-solvent volume
(Vp + Vs + Vns). This technique is called cloud-point determination.4,5
Eq 1. F = CP/100 = Vns / Vp + Vs + VNS
If the linear gradient cannot be optimized to achieve reproducible peak areas, a step gradient may
be needed. The GPEC of three polymers, polystyrene (PS), polybutadiene (PBD), and
polymethylmethacrylate (PMMA), are used for this experiment (Table 1). The linear gradient
produces poor peak area reproducibility for PS and PMMA, and the PBD does not elute.
Table 1. Polymers used in GPEC experiment.
Therefore, a step gradient method is used to control the amount of time the polymers spend in the
insoluble regions of the gradient. A steep curve is used for the insoluble regions (11) of the gradient
profile, while the soluble region is a less steep curve (6) as seen in the gradient table and resulting
chromatography (Figure 2).
Figure 2. Empower overlay of the GPEC analysis of PMMA, PS, and PBD.
By using the Empower 3 CDS custom calculations and report features, there is no need to export
data into a separate spreadsheet. The report in Figure 3 has the peak area ratios and the gradient
ratios calculated within Empower. The complete GPEC project and detailed GPEC guide are
available on the Waters website and in the Empower marketplace.
Figure 3. Empower report for the GPEC analysis of three separate
injections with gradient custom calculation.
Results and Discussion
GPEC is not GPC. In the example chromatograms of Figure 4, three chemically different polymers
co-elute using a GPC separation, because they have similar hydrodynamic volumes. If reversed-
phase separation is used to separate these polymers, they would remain partially soluble throughout
the separation and not baseline separate: partially overlapping. The GPEC method purposefully
precipitates the polymers temporarily onto the column or column frit until the gradient ratio of good
solvent reaches the polymer’s solubility point, releasing the polymer from the surface of the column.
Instead of using the size-based separation of GPC, the polymers are separated by their chemical
solubility using GPEC. GPEC is an LC application and is completely different from a GPC
application.1
Figure 4. GPC and GPEC of three polymers using an Alliance System.
The APC system designed for GPEC has a new configuration (Figure 5). The polymer isocratic
solvent manager (p-ISM) has been replaced with a polymer quaternary solvent manager (p-QSM) to
accommodate gradient elution with 100% organic solvent. The refractive index (RI) detector has
been replaced with an evaporative light scattering detector (ELSD), because RI is not compatible
with gradient elution. A photodiode array (PDA) is not used for this application example, because
not all polymers are UV active.2,3
Figure 5. APC System with p-QSM
and ELSD.
Conclusion
GPEC is a complementary analysis to GPC, when co-eluting peaks interfere with determining
polymer ratios in a mixture. This uncommon analysis technique has been used for decades but has
been limited to traditional HPLC instruments. UPLC instruments have solvent compatibility
limitations that can limit analysis options like GPEC. The APC with QSM is specifically designed for
aggressive solvents used in the polymer industry. By adding the flexibility of a quaternary pump to
the industry-accepted APC system, polymer analysis techniques are no longer limited to HPLC
technology.
References
Uliyanchenko, E.; van der Walac, S.; Schoenmakers, P.J., Challenges in Polymer Analysis by
Liquid Chromatography. Polym. Chem. 2012, 3, 2313–2335.
https://pubs.rsc.org/en/content/articlelanding/2012/py/c2py20274c#!divAbstrac.
1.
Waters ACQUITY UPLC ELS Detector, https://www.waters.com/waters/nav.htm?cid=514219.2.
ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide,
https://www.waters.com/webassets/cms/support/docs/71500109303rd.pdf.
3.
Klumperman, B.; Cools, P.; Philipsen, H.; Staal, W. (1996), A Qualitative Study to the Influence of
Molar Mass on Retention in Gradient Polymer Elution Chromatography (GPEC). Macromol.
Symp., 110: 1–13. doi:10.1002/masy.19961100102,
https://onlinelibrary.wiley.com/doi/abs/10.1002/masy.19961100102.
4.
Klenin, V. J.; Shmakov, S.L. Features of Phase Separation in Polymeric Systems: Cloud-Point
Curves (discussion). Universal Journal of Materials Science 2013, 1 (2), 39–45.
https://pdfs.semanticscholar.org/44cf/8e3fc4396d2d8bc6bc55 b5be0b2648e70619.pdf.
5.
Featured Products
ACQUITY Advanced Polymer Chromatography System
Empower 3 Chromatography Data Software
720006705 November 2019
©2019 Waters Corporation. All Rights Reserved.