Upload
d-rallan
View
220
Download
4
Embed Size (px)
Citation preview
Ultrasound in dermatology – basic principles and applications
D. Rallan and C. C. Harland
Department of Dermatology, St Helier Hospital, Carshalton, Surrey
Summary Ultrasound is widely used in general clinical medicine for non-invasive internal
imaging. Over the last twenty years, technological advances have enabled the
application of high-resolution ultrasonic imaging to the skin. Equipment and hardware
is now available to produce cross-section images and three-dimensional
reconstructions of selected skin segments. Resolution in vivo is not comparable to
light microscopy but continues to improve with superior transducer designs. Skin
ultrasonography has been reliably employed as an imaging modality in experimental
designs, its quantifiable parameters being a distinct advantage. In particular, increased
water content of the upper dermis, as occurs in inflammatory conditions or as a result
of photodamage, can be demonstrated clearly as an echo-poor zone. Thus, the future of
high-resolution ultrasound (HRU) may reside in its experimental role in monitoring
inflammatory or photodamage processes in response to novel treatments. With regard
to skin tumours, HRU reliably measures tumour thickness and also holds promise as a
differentiator between seborrhoeic keratoses vs. melanoma and benign naevi vs.
melanoma. While largely an experimental tool, the potential as an accurate,
quantitative and reliable diagnostic and monitoring aid, merits further attention
with an emphasis on clinical outcome measures.
Introduction
The term ‘ultrasound’ refers to sound waves of a
frequency that is above the human hearing range. In
1794, two Italian zoologists, who showed that bats use
an acoustic orientation system, discovered the phenom-
enon of ultrasound. Almost 85 years later, the Curie
brothers demonstrated the piezo-electric effect of sound
generated from crystal.1
Ultrasound became established in general medical
imaging in 1950 but it was not until 1979 that its
usefulness in dermatology was recognized.2 Today, skin
ultrasound remains an experimental tool which has
received increasing attention alongside other noninva-
sive investigative methods. Herein, the principles of skin
ultrasonography are described and potential future uses
of high resolution ultrasound are discussed.
Doppler ultrasound is alluded to only briefly.
Basic physics
All sound requires a medium in which to propagate.
When traversing media with differing density, there is a
change in velocity. This change may cause refraction,
reflection or more usually a combination of both
(scatter). Sound is also absorbed by various structures
to varying degrees. Excess sound energy absorption can
cause cavitation or heat damage. At the appropriate
frequency, ultrasound has the potential to highlight
structural heterogeneity. Increasing the frequency
improves the resolution, thereby enhancing the ability
to discriminate between structures. A greater frequency,
however, reduces the depth penetration of the sound
wave.3
In general medical imaging, a frequency of 3.5–
7.5 MHz is used producing a resolution of 2–3 mm. For
Correspondence: D. Rallan, Department of Dermatology, St Helier Hospital,
Wrythe Lane, Carshalton, Surrey, SM5 1AA, UK.
Tel. ⁄ Fax: +44 20 8296 3479.
E-mail: [email protected]
Accepted for publication 8 July 2003
Experimental dermatology • Review article
632 � 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638
skin imaging, a high frequency of 15 MHz and above is
used delivering a resolution of 60–120lm. This is
termed high resolution ultrasound (HRU).
Sound production and image generation
Figure 1 is a schematic example of a tranducer within a
scanner head used to generate A-, B-and C-scans which
will be discussed below.
A (amplitude) mode scanning
The most basic imaging relies on conversion of signals
into an amplitude value (Fig. 2). The radiofrequency
data is ‘enveloped’ to display amplitude data or a ‘peak’
on an oscilloscope. Each peak corresponds to an interface
between two structures of differing densities. The time
between the peaks on the oscilloscope enables calculation
of the distance between the corresponding interfaces and
therefore allows the determination of skin and tumour
thickness. The velocity of sound in the skin is taken as an
average of 1580 m ⁄ s. A-scanning is unidimensional.
B (brightness) mode scanning
If the transducer is moved along the skin surface,
several A-scans can be acquired. However, in this
instance, the returned signals are given a brightness
value or echogenicity on a grey scale instead of an
amplitude value. By means of dedicated software, this
Figure 1 A schematic representation of high-resolution ultra-
sound in operation; a transducer produces high-frequency sound.
A brief pulse is fired and the reflected sound is picked up by the
same transducer before the next pulse is fired (pulsed ultrasound).
The received signal is fed into a computer and can be processed in
several ways. An A-scan is unidimensional, and can only give
information at one line through the lesion; a B-scan is produced by
multiple A-scans in a vertical two dimensional plane. The
sequential back and forth movement of the transducer enables
acquisition of multiple B-scans and permits the construction of a
three-dimensional image.
Figure 2 Image generation for A- and
B- mode scanning. The raw data is elec-
tronically converted to either an amplitude
value or a brightness value to produce an
A- or a B- scan respectively.
Ultrasound in dermatology • D. Rallan and C. C. Harland
� 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638 633
radiofrequency data is used to generate a two-dimen-
sional cross-sectional image (Fig. 2).
C (computed) mode scanning
B-scans are cross-sectional images in a plane perpen-
dicular to the skin surface as shown in Fig. 3. By
combining multiple B-scans, an image can be generated
which is in a plane parallel to the skin. An image in this
plane can only be computed (and not acquired directly).
Evolution and applications
Conventional ultrasound can be helpful in certain
specific clinical circumstances in dermatology. High
resolution A- and B- mode scanning, however, has
produced a host of experimental applications as listed in
Table 1. A detailed discussion of each is outside the
scope of this article. Some studies are used to demon-
strate basic concepts.
Skin thickness and A-mode scanning
In 1979, Alexander and Miller were the first to use
A-mode ultrasound for the determination of skin
thickness in 10 healthy volunteers.2 A comparison
with the gold standard of xeroradiography (a high
resolution method involving x-rays applied parallel to
the skin surface) showed a high correlation (r ¼ 0.99).
They concluded that ultrasound was safer, cheaper, and
more versatile as xerography was confined to the limbs.
It therefore rapidly became the preferred method for
determination of skin thickness, paving the way for
further experimental uses.
Amongst the first applications was the objective
assessment of skin atrophy caused by topical steroids.4
This study also confirmed the high correlation between
the two methods for both normal and corticosteroid
treated skin. Furthermore, Capewell et al. were the first
to use A-mode scanning to quantify dermal thinning in
asthmatic subjects on inhaled corticosteroids and oral
steroids.5 Skin thickness has been used as an index of
severity, and as a means of monitoring treatment
response, of plaque psoriasis.6,7 However the surface
scale interferes with the propagation of ultrasound and
must be moisturized prior to transducer application.
Allergic patch test reactions can be quantified accu-
rately when compared with traditional calliper meas-
urements.8,9 Skin thickness has also been found to vary
during the course of the day and this is thought to be
due to gravitational translocation of dermal fluid.10,11
Until 1983, skin ultrasound was used almost exclu-
sively for skin thickness determination. Coleman and
Lizzi were the first to investigate the potential of A-mode
scanning in tissue characterization.12 They used a
spectrum of frequencies and observed that ocular
malignant melanoma and metastatic carcinoma could
be reliably classified based on their acoustic reflectance
signatures. This technique of analysing acoustic spectra
Table 1 Examples of the potential applications of ultrasound
imaging in dermatology.
Applications
Common mode(s)
of scanning
Skin thickness determination2,4,5, A
Examination of lymph nodes in
skin disease21,23,26
B, C and duplex
sonography
Differentiation of soft tissue
tumours12,13,20
B and C
Staging and follow up of malignant
melanoma17,23,24
A, B and duplex
sonography
Examination of benign and malignant
epithelial tumours20,22
A, B and C
Examination and classification of
scleroderma morphoea31
B
Examination and assessment of
plaque psoriasis6,7
A and B
Assessment of dermal oedema10,11,19 A and B
Assessment of skin ageing and
photodamage18
A and B
Screening for necrotizing fasciitis32 Conventional sonography
Post-operative and inflammatory
abdominal wall abscess detection33
Conventional sonography
Figure 3 The difference between A-, B- and C-mode imaging. A
scans are unidimensional. Multiple A- scans combined produce a
two dimensional cross-section image (B-scan) as shown. C-scans
are also two dimensional but in a plane parallel to the skin surface
and are derived from multiple B-scans. Both B- and C-images can
be used for three dimensional reconstructions.
Ultrasound in dermatology • D. Rallan and C. C. Harland
634 � 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638
has not been evaluated fully for skin tumours. Some
features of conventional A-mode scanning had been
investigated with regard to small tumour characteri-
zation13 but B-mode technology now shows more
potential.14,15
Echogenicity and B-mode scanning
B-mode software technology became sufficiently devel-
oped by 1986 making it possible to generate cross-
sectional images. Consequently, it has become apparent
that tissue characteristics, in particular collagen fibres,
keratin and dermal water content, are the major
determinants of HRU images.15,16 Furthermore, dermal
collagen content, collagen type, collagen orientation and
collagen bundle size all have an effect on dermal
reflectivity ⁄ echogenicity.15 These factors determine the
number of collagen–matrix interfaces and hence the
extent of sound scatter. Thick keratin is highly reflective
and has the effect of producing an echolucent retro-
lesional shadow (Fig. 4). In contrast, dermal water is
poorly reflective and reduces dermal echogenicity.
Figure 4 B-scan of a seborrhoeic keratosis. The difference in
density between the gel and skin surface causes significant sound
reflection at the point of entry. This entry echo line (EEL), is
prominent in heavily keratinized tumours (e.g. seborrhoeic kera-
toses) which reflect almost all sound at the point of entry and
thereby cast a shadow in the dermis (posterior attenuation).
bbl – bubbles in gel.
300250
200150
100
50
00.5 1 1.5 2.5 3.52 3
0.5
1
1.5
2
2.520 40 60 80 100 120 140 160 180 200 220
(b)
300250
200150
100
50
00.5 1 1.5 2
1
1.5
2.0
2.520 40 60 80 100 120 140 160 180 200 220
(a)
A line (88) - through normal skin
Figure 5 (a) B-scan (top) from skin of a middle aged woman who is a longterm sunbed user. A malignant melanoma (MM) is shown. EEL –
entry echo line, SC – subcutis, B – bubble in gel. Note the distinct echopoor ’sub-epidermal low echogenic band (SLEB) indicative of
photodamage. An A-scan (bottom), is taken through the dashed line. (b) B-scan (top) and one A-scan (bottom) through a compound
naevus from a young male patient. EEL: Entry echo line, CN: compound naevus, B: bubbles in gel. Compared to Figure 5a, a wide echopoor
band beneath the EEL is not seen.
Ultrasound in dermatology • D. Rallan and C. C. Harland
� 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638 635
Descriptive studies on tissue characteristics of normal
skin and pathology have been well documented. De Rigal
et al. were one of the first groups to use quantitative
B-mode scanning in their study on photodamage.18
They observed and quantified a distinct subepidermal
low echogenic band (SLEB) in photodamaged skin
(Fig. 5). This was initially attributed to altered elastotic
collagen. Alternatively, such echopoor zones could be
equated to increased water content.16 Seidinari and
co-workers capitalized on this phenomenon by quanti-
fying dermal oedema in patch test reactions (Fig. 6).
The distribution of dermal oedema has now been
elucidated in several conditions: cardiac failure, post-
thrombotic venous disease and lymphoedema.19
B-mode scanning and skin tumours
During the last decade, the emphasis of research has
shifted towards the field of noninvasive assessment of
skin tumours.15,17,20–26 Tumour depth can be deter-
mined in vivo with good accuracy. In a study involving
144 melanomas, Gassenmaier et al. showed a strong
correlation (r ¼ 0.95) between sonometric and histo-
logic measurements of tumour thickness.25 Ultrasound
tends to overestimate tumour thickness. Tissue shrink-
age following excision and associated lymphocytic
infiltrates are thought to be the main causes of this
disparity.
Tumours are generally echo-poor. Most benign skin
tumours have more internal echoes compared to
malignant melanoma (Fig. 7). This difference may be
due to a relative paucity of collagen in invasive
melanoma. Quantification of features such as internal
shadows, has not yet been shown to have greater
accuracy than a subjective assessment of such images
by a consultant dermatologist.22 Furthermore, studies
which are currently in progress have yet to show
superior diagnostic accuracy for HRU vs. clinical
judgement at any level. Duplex sonography (B-scan
combined with Doppler ultrasound), shows good poten-
tial in this regard particularly in relation to lymph node
assessment.21,23,26
Figure 7 Benign naevus (left) compared to
malignant melanoma (right). Benign
tumours generally have more internal
echoes.
Figure 6 A typical ultrasound appearance of a positive allergy test
is shown (top) compared to normal skin (bottom). The oedema in
allergic and irritant reactions was found to occur mainly in the
papillary dermis. The degree of reduction in echogenicity could be
quantified, thus giving another parameter (apart from skin thick-
ness changes) for measuring patch test reactions. (Figure adapted
from Contact Dermatitis,1991; 24: 216-222).
Ultrasound in dermatology • D. Rallan and C. C. Harland
636 � 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638
Advantages and disadvantages of HRU imaging
Compared with xeroradiography, ultrasound is cheaper
and safer. Depending on the type of scan to be acquired,
the procedure can be quick. The noninvasive nature
enables accurate measurements to be made in vivo.
Prototype equipment is cumbersome (Fig. 8) and
certain sites (e.g. inner canthus) remain inaccessible.
The resolution in vivo is not comparable to that with
conventional light microscopy. There is at present, an
upper limit to the sound energy that can be transmitted
into the skin without causing damage.
Future potential
The ability to quantify photodamage and skin ageing is
perhaps to be the most promising application of skin
ultrasound in the foreseeable future. This is likely to be
driven by the booming ‘wellness industry’.27 With
improved tissue characterization, both qualitative and
quantitative, HRU may find a role as a diagnostic aid,
especially in conjunction with other noninvasive tech-
niques such as reflectance spectrophotometry and skin
patterning.28,29
The destructive properties of sound are already used
in other medical fields but have yet to be assessed in
skin tumour therapy. In order to achieve resolution
comparable to light microscopy, a frequency of
40–60 MHz is required.30 This is achievable ex vivo.
The development of epidermal cooling devices as used
for modern laser therapies will be necessary. However,
the ability to examine living tissue at a microscopic level
stands to introduce a whole new era of clinical
investigation and experimental designs for the future.
Acknowledgements
The department’s research is funded by Epsom and St
Helier NHS Trust Fellowship Scheme. We acknowledge
the collaboration and support provided by the Depart-
ment of Physics, Institute of Cancer Research, Royal
Marsden Hospital and especially from Dr Jeff Bamber
and Nigel Bush.
References
1 Curie J, Curie P. Development par pression d’electricite
polaire dans lees cristaux hemiedres a faces incliees.
CR Acad Sci 1881; 91: 294–5.
2 Alexander H, Miller DL. Determining skin thickness with
pulsed ultrasound. J Invest Dermatol 1979; 72: 17–9.
3 Bamber JC. Acoustic characteristics of biological media.
In: Cocker MJ, ed. Encyclopaedia of Acoustics. New York:
John Wiley and Sons, Inc., 1997: 1703–26.
4 Tan CY, Marks R, Payne P. Comparison of xeroradio-
graphic and ultrasound detection of corticosteroid induced
dermal thinning. J Invest Dermatol 1981; 76: 126–8.
5 Capewell S, Reynolds S, Shuttleworth D et al. Purpura and
dermal thinning associated with high dose inhaled corti-
costeroids. Br Med J 1990; 16: 1548–51.
6 Serup J. Non-invasive quantification of psoriasis plaques –
measurement of skin thickness with 15MHz pulsed
ultrasound. Clin Exp Dermatol 1984; 9: 502–8.
7 Gupta AK, Turnbull DH, Harsiewicz KA et al. The use of
high frequency ultrasound as a method of assessing the
severity of a plaque of psoriasis. Arch Dermatol 1996; 132:
658–62.
8 Serup J, Staburg B. Ultrasound for assessment of allergic
and irritant patch test reactions. Contact Dermatitis 1987;
17: 80–4.
9 Seidinari S, Di Nardo A, Pepe P et al. Ultrasound B-scan-
ning with image analysis for assessment of allergic patch
test reactions. Contact Dermatitis 1991; 24: 216–22.
10 Gniadecka M. Localisation of dermal oedema in lipo-
dermatoslcerosis: distribution, effects of posture and
compressive therapy evaluated by high frequency ultra-
sonography. Acta Derm Venereol 1995; 75: 120–4.
11 Tsukahara K, Takema Y, Moriwaki S et al. Dermal fluid
translocation is an important determinant of diurnal
variation in human skin thickness. Br J Dermatol 2001;
145: 590–6.
12 Coleman DJ, Lizzi FL. Computerised ultrasonic tissue
characterisation of ocular tumours. Am J Ophthalmol 1983;
96: 165–75.
Figure 8 The prototype equipment used by our group. Data
acquisition software and the scanner head (resting on patients
arm) could be condensed into a laptop with a toothbrush sized
probe in future models.
Ultrasound in dermatology • D. Rallan and C. C. Harland
� 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638 637
13 Edwards C, Al-Boosi MM, Marks R. The use of A-scan
ultrasound in the assessment of small skin tumours.
Br J Dermatol 1989; 121: 297–304.
14 Hughes BR, Black D, Srivastava A et al. Comparison of
techniques for the non-invasive assessment of skin
tumours. Clin Exp Dermatol 1987; 12: 108–11.
15 Harland CC, Bamber JC, Guesterson BA et al. High
frequency, high resolution ultrasound in the assessment
of skin tumours. Br J Dermatol 1993; 128: 525–32.
16 Gniadecka M, Quistorff B. Assessment of dermal water by
high frequency ultrasound: comparative studies with nuc-
lear magnetic resonance. Br J Dermatol 1996; 135: 218–24.
17 Mende U, Petzoldt W et al. Sonography: the ideal imaging
method for staging and follow-up of malignant melanoma.
In: Ultrasound in Dermatology (Altmeyer, P, El-Gammal, S,
Hoffman, K, eds), 1st edn. Berlin: Springer-Verlag 1992,
119–29.
18 De Rigal J, Escoffier C, Querleux B et al. Assessment of
ageing of the human skin by in vivo ultrasonic imaging.
J Invest Dermatol 1989; 93: 621–5.
19 Gniadecka M. Localisation of dermal oedema in lipoder-
matosclerosis, lymphoedema and cardiac insufficiency.
J Am Acad Dermatol 1996; 35: 37–41.
20 Hoffman K, El-Gammal S, Winkler K et al. Skin tumours in
high frequency ultrasound. In: Ultrasound in Dermatology
(Altmeyer P, El-Gammal S, Hoffman K, eds) 1st edn.
Berlin: Springer-Verlag, 1992, 181–201.
21 Schmid-Wendtner M, Partscht K, Korting HC et al.
Improved differentiation of benign and malignant
lympadenopathy in patients with cutaneous melanoma
by contrast-enhanced colour doppler sonography. Arch
Dermatol 2002; 138: 491–7.
22 Harland CC, Kale SG, Jackson P et al. Differentiation of
common benign pigmented lesions from melanoma by high
resolution ultrasound. Br J Dermatol 2000; 143: 281–9.
23 Prayer L, Winkelbauer H, Gitzmann N et al. Sonography
versus palpation in detection of regional lymph node
metastases in patients with malignant melanoma. Eur J
Cancer 1990; 26: 827–30.
24 Rossi CR, Seno A, Vecchiato A et al. Ultrasound guided
fine needle aspiration cytology in the early detection of
melanoma metastases. Cancer 2000; 25: 186–93.
25 Gassenmeier G, Kiesewetter F, Schell H. Value of high
frequency sonography in the determination of maximum
vertical tumour thickness in primary malignant melanoma
of the skin. In: Altmeyer, P, El-Gammal, S, Hoffman, K, eds,
Ultrasound in Dermatology, 1st edn. Berlin: Springer-Verlag,
1992, 221–7.
26 Tschammler A, Ott G, Seelbach-Goebel B et al. Lymhpa-
denopathy: differentiation of benign from malignant dis-
ease – colour doppler ultrasound assessment of intranodal
angioarchitecture. Radiology 1998; 208: 117–23.
27 Pilzer PZ. The next trillion. In: The Next Trillion, 1st edn.
(abridged version), Dallas: Video plus Inc 2001, 27–34.
28 Round AJ, Duller AWG, Fish PJ. Lesion classification using
skin patterning. Skin Res Technol 2000; 6: 183–92.
29 Wallace VP, Crawford DC, Mortimer PS et al. Spectropho-
tometric assessment of pigmented skin lesions: methods
and feature selection for evaluation of diagnostic
performance. Phys Med Biol 2000; 45: 735–51.
30 Foster FS, Pavlin CJ, Harasiewicz KA et al. Advances in
ultrasound biomicroscopy. Ultrasound Med Biol 2000; 26:
1–27.
31 Hoffman K, El-Gammal S et al. Examination of circum-
scribed scleroderma using 20-MHz B-scan ultrasound.
In: Altmeyer, P, El-Gammal, S, Hoffman, K, eds. Ultrasound
in Dermatology, 1st edn. Berlin: Springer-Verlag, 1992,
221–7. 231–43.
32 Yen ZS, Wang HP, Ma HM et al. Ultrasonographic
screening of clinically-suspected necrotizing fasciitis. Acad
Emerg Med 2002; 9: 1448–551.
33 Yeh HC, Rabinowitz JG. Ultrasonography and computer
tomography of inflammatory abdominal wall lesions.
Radiology 1982; 144: 859–63.
Ultrasound in dermatology • D. Rallan and C. C. Harland
638 � 2003 Blackwell Publishing Ltd • Clinical and Experimental Dermatology, 28, 632–638