Upload
ig-koprinkov
View
216
Download
2
Embed Size (px)
Citation preview
Optics Communications 225 (2003) 371–376
www.elsevier.com/locate/optcom
Tunable narrowband quasi-modeless laser
I.G. Koprinkov *,1, Akira Suda, Katsumi Midorikawa
RIKEN (The Institute of Physical and Chemical Research), 2-1 Hirosawa, Wako-shi, Saitama 351-0198, Japan
Received 18 November 2002; received in revised form 12 July 2003; accepted 19 July 2003
Abstract
A simple and reliable ‘‘modeless’’ dye laser is created. Spectrally dense, tunable generation of strongly suppressed
mode structure of about 6 GHz bandwidth is obtained. The laser parameters allow to detect simultaneously the entire
inhomogeneously broadened spectral line for a number of practical cases.
� 2003 Elsevier B.V. All rights reserved.
PACS: 42.60.By; 42.60.Da; 42.55.Mv
Keywords: Modeless laser; Dye laser; Laser cavity; Cavity modes
1. Introduction
The discrete mode/frequency structure of thelaser cavity field becomes inappropriate for some
applications. In the case of inhomogeneous
broadening of the atomic transitions, the laser
emission interacts with those atomic groups only,
whose transitions coincide with the frequencies of
the laser cavity modes. The atomic groups having
off-resonant transitions do not practically absorb
laser photons and may become lost for the ex-periment. This problem becomes crucial when the
number of the particles (atoms, molecules) to be
* Corresponding author. Tel.: +359-2-9653072; fax: +359-2-
683215.
E-mail address: [email protected] (I.G. Koprinkov).1 Present address: Department of Applied Physics, Technical
University of Sofia, 8, Kliment Ochridski blvd., 1756 Sofia,
Bulgaria.
0030-4018/$ - see front matter � 2003 Elsevier B.V. All rights reserv
doi:10.1016/j.optcom.2003.07.036
detected is very low. That is why, creation of
tunable laser of continuum spectrum without (or
having strongly suppressed) frequency structureand a bandwidth comparable to the typical
Doppler bandwidth, is important. Lasers without
mode structure, called modeless lasers (MLLs),
were demonstrated in a number of works [1–4].
Nowadays, the MLLs are important tool in the
broadband single-shot CARS spectrometers [3,4]
and allow substantial improvement of the CARS
performance. The MLLs are actually travellingwave amplifiers working in amplified spontaneous
emission (ASE) mode, i.e., the ASE originating
from the end of the active medium is amplified
during each pass experiencing [1–3] or not [3] se-
lective spectral control. The MLLs without selec-
tive spectral control [3] or using low-pass number
amplification [1] generate too wide spectrum,
which, while suitable for broadband CARS spec-troscopy, becomes inappropriate for spectroscopy
ed.
372 I.G. Koprinkov et al. / Optics Communications 225 (2003) 371–376
of Doppler broadened atomic or molecular lines,
whose bandwidth is in the GHz range. At properly
designed MLLs, generation of tunable modeless
emission of single-shot bandwidth of about 6 GHz
is achieved [2]. This, however, is obtained at the
expense of complications in the optical scheme,thus imposing additional requirements in the laser
alignment and operation. Here, we propose an
alternative approach, which allows to create a
simple and reliable (quasi) MLL, generating nar-
rowband tunable emission. The laser we propose
actually has cavity and thus, cavity modes. How-
ever, due to the very low Q-factor used, the mode
bandwidth is wide and the adjacent modes aresubstantially overlapped so that the emission is, in
fact, continuum or quasi-modeless. To the best of
our knowledge, such an approach in the creation
of MLL is proposed for the first time [5] and its
design and properties are considered in the present
paper. The choice of the particular laser medium
and the other laser parameters was also dictated by
the intended application of such a MLL, namely,optical detection of inner shell ionized lithium at-
oms [6] using 1s2s(3S)–1s2p(3P) triplet transition at
548.5 nm.
2. Quasi-modeless cavity
The present approach of ‘‘modeless’’ laser isbased on a concept of mode overlapping. The mode
bandwidth DmMB, the free spectral range (FSR)
DmFSR, and the finesse F of the laser cavity are
related by the following expression [7,8]:
DmMB ¼ DmFSRF
; ð1Þ
Table 1
The finesse, mode bandwidth, and mode filling factor (assuming Lor
cavity consisting of combinations of total reflectance mirror (R ¼ 0:9
single Fresnel reflection wedge glass plate (R ¼ 0:04)
R1 0.99 0.99
R2 0.99 0.08
F 310 2.32
DmMB (GHz) 0.005 0.72
MFF 0.005 0.50
where the finesse is determined by the reflectivity
of the cavity mirrors R1 and R2
F ¼ pðR1R2Þ1=4
1� ðR1R2Þ1=2: ð2Þ
Consequently, if the finesse is less than unity, the
adjacent modes of a bare laser cavity must strongly
overlap each other.
For the quantitative estimation of the modeless
properties of given laser cavity, a quantity called
mode-filling factor (MFF) will be introduced. It
can be expressed as a fraction of the rectangular‘‘area’’ of mode peak intensity (gð0Þ) height, and 1
FSR (DmFSR) width, covered by two normalized (or
equal peak intensity) adjacent cavity modes gðmÞ
MFF ¼2R DmFSR=20
gðmÞdmgð0ÞDmFSR
: ð3Þ
The relevant parameters of a two-mirror bare
cavity for particular values of the mirror reflectivity
are shown in Table 1. As can be seen, finesse below
unity (F ¼ 0:65) can be achieved (based on thecavity mirrors only) using the Fresnel reflection of
two uncoated wedge glass plates as cavity mirrors.
The MFF of such a cavity approaches 0.90. The
corresponding Q-factor at k ¼ 550 nm is Q ¼ ð2L=kÞF � 2� 105, and the effective photon lifetime in
the cavity, s ¼ Qk=2pc, is 60 ps. Thus, in the
nanosecond time scale, the emission of such a laser
entirely follows the emission dynamic of the activemedium. Preliminary prescribed cavity parameters
can be obtained using dielectric mirrors of suitable
reflectivity. Approaching the limiting case R1 ! 0,
R2 ! 0, the laser is, of course, converging on the
true, ASE-MLL, whose spectral selection and
tuning requires an auxiliary optics [2].
entzian mode shape gðmÞ) of a L ¼ 9 cm long two-mirror bare
9), double Fresnel reflections parallel glass plate (R ¼ 0:08) and
0.99 0.04
0.04 0.04
1.75 0.65
0.95 2.57
0.60 0.89
Fig. 1. Optical scheme of the laser: DC, dye cell; G, diffraction
grating; M1 and M2, wedge glass plates.
I.G. Koprinkov et al. / Optics Communications 225 (2003) 371–376 373
In a real laser, other optical elements also in-
troduce losses. For the MLL considered here,Fig. 1, the main additional reduction of the effec-
tive cavity Q-factor comes from the grazing inci-
dence arrangement of the grating. That is why, the
actual finesse could be substantially smaller than
that one given in Table 1. Such extra losses are
inevitable for a selective cavity and, in the same
time, enforce the modeless properties. Overlapping
of the cavity modes, however, is not enough toobtain pronounced modeless operation because
the emission strongly narrows due to the optical
gain. That is why, besides of low cavity Q-factor,
other conditions that contribute to the modeless
operation, must be also assured. The most im-
portant of these are the relatively low pump in-
tensity (to reduce the single-pass optical gain) and
appropriate short pump pulse (to reduce thenumber of cavity round-trips). Both conditions
can be met, e.g., keeping the pump energy close to
the laser threshold. The modeless properties of the
generated laser emission can be also characterized
by the MFF.
For the experimental check of the above ap-
proach we have used a dye laser. Laser dyes,
having large stimulated emission cross-section(r � 10�16 cm2), are suitable for the present MLL
concept because high enough gain can be easily
achieved at low pump energy.
3. Experimental set-up
The scheme of the ‘‘modeless’’ laser is shown inFig. 1. For the sake of simplicity, the laser cavity is
formed by two uncoated wedge glass plates
(R1 ¼ R2 ¼ 0:04) as an output mirror M1 and a
tuning mirror M2. The laser active medium is an
ethanol solution of Rhodamin 575 (Exciton, Inc.).
The dye cell DC is 2 mm thick static cell, placed at
about 45� (or Brewster angle, if necessary) with
respect to the optical axis to avoid feedback from
the cell walls as well as to extend the length of theactive medium. Dye concentration of about
1� 10�3 M/l was found to be optimal for the
present purpose. The length of the laser cavity was
set at L � 9 cm. At these parameters, the FSR and
the bare cavity mode bandwidth (based on the
cavity mirrors, only) are DmFSR ¼ 1:67 GHz and
DmMB ¼ 2:57 GHz, respectively. The frequency
selection element is 2400 l/mm grazing incidencediffraction grating G, placed at about 88� angle ofincidence. The MLL has been pumped by the
second harmonic of a Nd:YAG laser having 11 ns
pulse duration. Longitudinal pumping has been
used to ensure better matching between the pump
and the dye laser modes. The pump emission was
focussed by a f ¼ 30 cm spherical lens. The dye
cell was placed at about 3 cm before the focalpoint. The spectrum of the laser was analyzed by a
Fabry–Perot etalon of 30 GHz FSR and finesse of
about 30. The energy of the dye laser was mea-
sured by a Rm-3700 Universal Radiometer – Laser
Probe Inc. and a RjP-735 energy probe. The pulse
duration was measured by a fast digital oscillo-
scope Tektronix DSA 602 and a PIN photodiode
S-5972.
4. Results and discussions
The mode properties of the laser spectrum have
primary importance for the present study and the
resolution of the available Fabry–Perot etalon will
be discussed first. The Fabry–Perot cavity can betreated in terms of dumping oscillator, whose fre-
quency transmission function has Lorentzian
shape of, in our case, 1 GHz mode bandwidth.
Thus, if not wider than the etalon transmission
bandwidth, two Lorentzian shaped adjacent laser
modes (1.67 GHz separation) will mutually over-
lap each other slightly below 0.25 level from their
peak intensity. The relation between the resultantintensity minimum and maximum of their super-
position is thus Imin=Imax 6 0:50. This is well below
374 I.G. Koprinkov et al. / Optics Communications 225 (2003) 371–376
the critical value of about 0.8 for the resolution of
two spectral lines [9], and two such modes must be
well resolved by the etalon. Usually, the experi-
mentally observed laser modes do not have so wide
frequency wings as the Lorentzian profile and, at
equal other conditions, Imin=Imax below 0.50 mustbe expected. Thus, if higher than 0.5 Imin=Imax is
observed, it must be attributed to the laser mode
bandwidth. In relation with the above discussion,
the following criterion of modeless emission will be
proposed: the laser emission will be considered
modeless if two adjacent cavity modes are unre-
solved according to the criterion of resolution of
two spectral lines, i.e., the condition Imin=Imax P 0:8must hold [10].
A single-shot dye laser spectrum at 0.5 mJ
pump energy is shown in Fig. 2(a). As can be seen
from the Fabry–Perot fringes, the laser mode
structure is well suppressed and practically – in-
distinguishable. The bandwidth of the generation
is about 6 GHz and consists of 3–5 well-over-
lapped cavity modes. Increasing the pump energyresults in increased spectral bandwidth and more
pronounced mode structure. Due to fluctuations,
this also happens in about 20–30% of the cases at
Fig. 2. Single-shot laser spectra at 0.5 mJ average pump energy.
For more details see the text.
0.5 mJ (average) pump energy when high-energy
pulse hits the active medium. The mode structure
may look better suppressed than it really is if the
modes have some smooth, e.g., bell-shape, peak
intensity distribution, Fig. 2(a). The actual mode
overlapping will look realistic if the modes havenearly equal peak intensity (‘‘top-flat’’ distribu-
tion). This, however, does not usually take place in
the laser operation because of the mode gain
competition and the selective cavity losses, and
may happen accidentally as a fluctuation. Such a
case is shown in Fig. 2(b), where the generated six
modes have nearly equal peak intensity. The rela-
tion between the adjacent intensity minima andmaxima in that case is about Imin=Imax � 0:9. Thisis well above the accepted 0.8 value for the mode
resolution and represents an indication of good
mode overlapping. The tuning range of the dye
laser at 0.5 mJ pump energy is 544–554 nm.
The dependence of the pulse energy of the MLL
(M1 output) vs pump energy and the correspond-
ing energy conversion efficiency are shown inFig. 3. Despite the very high cavity losses, the
MLL does not require high pump energy. Stable
laser operation can be achieved, e.g., at about 150
lJ pump energy. Pulse energy of about 5 lJ is
obtained at 1 mJ pump energy. This determines
the conversion efficiency of about 0.5%, based on
the M1 output only. As must be expected, the en-
ergy conversion efficiency of such an oscillator isnot high because of the very low cavity feedback
and the high losses of the grazing incidence grat-
ing. Within the energy range studied, the energy
conversion efficiency increases when the pump
0
2
4
6
0
0.2
0.4
0.6
0 200 400 600 800 1000
Dye
lase
ren
ergy
[µJ]
Dye
lase
ref
fici
ency
[%]
dye laser energy
dye laser efficiency
Fig. 3. Laser energy and energy conversion efficiency vs pump
energy relationships.
I.G. Koprinkov et al. / Optics Communications 225 (2003) 371–376 375
energy increases, Fig. 3. This, however, also in-
creases the spectral bandwidth and the pulse du-
ration. The oscillator output can be amplified by a
standard amplifier if higher energy is required for
particular application. In general, better efficiency
can be expected using Rhodamin 6G dye. The low-energy pump regime, used here, ensures long life-
time of the laser dye even using a static dye cell.
The pulse duration of the laser, measured at
0.35 mJ pump energy, was 2 ns. Thus, the laser
emission above the threshold is formed for about
three cavity round-trips. The peak power is 500 W
and the power conversion efficiency is 1.5%. The
pulse duration gradually increases when the pumpenergy increases, Fig. 4. The oscilloscope traces
show that this results mainly from the extension of
the trailing edge of the pulse. The later tends to
transform into a second pulse at pump energy
above 1 mJ. The second pulse becomes well ex-
pressed if replacing the glass plate M2 by a high-
reflectivity (aluminum) mirror. Thus, at low pump
energy/feedback, the trailing part of the pumppulse is not effectively converted in laser genera-
tion. This shows that the energy conversion effi-
ciency can be improved using shorter pump pulses.
The general advantage of the present approach
of MLL is that a tunable, narrowband, spectrally
dense emission is generated by means of simple
laser scheme, while still preserving the cavity
control on the laser generation. The alignment ofthis laser is as simple as the other similar grazing
incidence dye lasers. In the lack of experience,
lasing can be initially achieved using an auxiliary
total reflectance mirror placed behind the glass
plate M2. Then, the glass plate M2 is aligning so as
Fig. 4. Laser pulse duration vs pump energy.
to get generation by means of M1–M2 cavity, while
the auxiliary mirror is misaligning and removing
from the laser.
The MLL described here was particularly de-
signed for optical detection of inner shell lithium
ions using 1s2s(3S)–1s2p(3P) triplet transitionat 548.5 nm. The transition bandwidth at high-
density lithium vapors (about 1017 cm�3 at 900 �Cvapor temperature) and low degree of ionization
(Stark broadening due to the electric field of the
ions [6] can thus be neglected) is dominated by the
Doppler broadening and is estimated to be about 5
GHz. The lifetime of 1s2s(3S) (radiatively meta-
stable) and 1s2p(3P) states is estimated to be about20 and 15 ns, respectively, based on the collisions
with neutrals and spontaneous emission (3P state).
Thus, the MLL parameters (tuning range, band-
width, and pulse duration) well satisfy the re-
quirements for optical detection of 1s2s(3S) lithium
ions. Of course, the MLL described here is also
suitable for other spectroscopic applications when
tunable and spectrally dense emission over theDoppler bandwidth of given transition is required.
In conclusion, a simple and reliable ‘‘modeless’’
dye laser has been created. Spectrally dense, tun-
able generation of about 6 GHz bandwidth is
obtained. The low Q-factor and the low-energy
pump regime help to generate a quasi-modeless
emission of short pulse duration, and long lifetime
of the laser dye. The laser parameters allow todetect simultaneously the entire inhomogeneously
broadened spectral line for a number of practical
cases.
Acknowledgements
The authors are grateful to Dr. Tohru Kobay-ashi for the loan of technical equipment.
References
[1] S. Szatmari, F.P. Schafer, Opt. Commun. 49 (1984) 281.
[2] P. Eward, Opt. Commun. 55 (1985) 124.
[3] J.W. Hahn, C.W. Park, S.N. Park, Appl. Opt. 36 (1997)
6722.
[4] D.R. Snelling, R.A. Sawchuk, T. Parameswaran, Appl.
Opt. 33 (1994) 8295.
376 I.G. Koprinkov et al. / Optics Communications 225 (2003) 371–376
[5] I.G. Koprinkov, A. Suda, K. Midorikawa, in: Proceedings
of the 34th Conference of European Group of Atomic
Spectroscopy (34th EGAS), 2002, pp. 389–390.
[6] R.G. Caro, J.C. Wang, J.F. Young, S.E. Harris, Phys. Rev.
A 30 (1984) 1407.
[7] W.T. Silfvast, Laser Fundamentals, Cambridge University
Press, Cambridge, 1996.
[8] Eq. (1) gives approximate value of the mode (cavity
eigenfrequency is assumed) bandwidth when the reflec-
tivity of the mirrors is low. The transmission function
(mode frequencies) of a Fabry–Perot cavity can be
expressed in terms of Airy function [7], from which the
mode bandwidth can be deduced more precisely. How-
ever, reducing the mirror reflectivity down to the range
of interest results in minima of the transmission function,
whose intensity is higher than the 0.5 level, and the
mode bandwidth cannot be defined in the usual way, i.e.,
as the full width at half maximum. That is why, our
conclusions will be based on Eq. (1), assuming modes of
Lorentzian spectral shape. The mode bandwidth based
on Airy function is wider than that one predicted by Eq.
(1), so that this will only enforce the predicted modeless
properties.
[9] W. Demtr€ooder, Laser Spectroscopy, Basic Concepts and
Instrumentation, Springer, Berlin, 1982.
[10] It is, of course, understood that the transmission function
of the Fabry–Perot etalon does not limit the resolution of
the laser modes.