104
Care Fresenius Medical 4008 HDF Technical Manual

TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Embed Size (px)

Citation preview

Page 1: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

CareFreseniusMedical

4008 HDF

Technical Manual

Page 2: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4008 HDF

Technical Manual

Page 3: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 0-1Part No. 674 398 1

4008 HDF Technical Manual

The Technical Manual contains all information necessary for performing maintenance and repair work.

The 4008 HDF option reflects the latest state of technology and complies with the requirements ofEN 60601-1

Assembly, extension, adjustment, modification or repair may only be carried out by the manufacturer orpersons authorized by him.

Any inquiries should be addressed to:

Manufacturer:Fresenius Medical CareDeutschland GmbHPlant SchweinfurtHafenstraße 9D-97424 Schweinfurt, GermanyTel.: 09721-678-0Fax: 09721-678-200

Head office:Fresenius Medical CareDeutschland GmbHBorkenberg 14D-61440 Oberursel/Ts., GermanyTel.: 06171-60-0Telex: 410805 fres dFax: 06171-251-58

Local Service:

Page 4: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

0-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 5: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 0-3

How to use the Technical Manual

Search and find What? Where?Tables Contents Pages 0-5 and at the beginning of each chapter

Purpose This manual is intended for:– first studies (to acquire basic knowledge)– reference purposes (for start-up, maintenance and repair)

Organization The manual is divided into 6 chapters:0 General Notes1 Description of machine functions2 Technical safety checks3 Calibration instructions4 Circuit descriptions and circuit diagrams5 Spare parts

Numbering system Page number 1-3 is to be interpreted as: Chapter 1, Page 3

Qualification This manual is intended for service technicians– who are familiar with the current Operating Instructions (Operating Instruc-

tionseratining to this Technical Manual are available under part no. 674 406 1)– who have the necessary background experience in mechanics, electrical and

medical engineering– who have been authorized by the manufacturer to perform maintenance and

repair work– who have access to the necessary auxiliary and measuring equipment

Restrictions The study of this manual does not represent an alternative to the training coursesoffered by the manufacturer.

Manual changes Manual changes will be released as new editions, supplement sheets or productinformation.

Note:Modifications relating to circuit diagrams and component layouts (SP/BP) do notnecessarily involve a change of the footer (edition).Refer to the index field of the respective circuit diagram or component layout forthe respective state of these diagrams. The identification on the P.C.B. permitsthe user/technician to verify if the circuit diagram / component layout matches theP.C.B. actually installed.

In general, this manual is subject to modification.

Page 6: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

0-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Representation New circuit symbols are used in the circuit diagrams. Potential data given in thecircuit diagrams and setting instructions refer to the respective earth.For example: 24 means ground for 24 V voltage.

Component marking Example:in circuit diagrams

This refers to a resistor with a position number 75 with a resistance of 1.5 Ohm.

The decimal point is replaced by a unit symbol (to reduce the possibility of errors).

Resistors: Capacitors:R1: 0.1 Ω µ 1: 0.1 µF1R5: 1.5 Ω 1µ5: 1.5 µF1K5: 1.5 kΩ 1000µ: 1000 µF

Note: When repairing or exchanging replacement parts make sure to take the applica-ble ESD precautions (e.g. EN 100 015-1).During repair/troubleshooting in the hydraulic unit, protect the components fromdialysate.

Technical data: The technical data for the 4008 HDF option is to be found in Chapter 1 of theOperating Instructions.

75

1R5

Page 7: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 0-5

Table of contents

Section Page

1 Description of machine functions ............................................................................... 1-1.1 Description ...................................................................................................................... 1-31.2 Description of extended T1 test ...................................................................................... 1-81.3 Error messages............................................................................................................... 1-10

2 Technical safety checks ............................................................................................... 2-

3 Calibration instructions ............................................................................................... 3-3.0 General information on the calibration instructions ....................................................... 3-33.1 Calibrating the UF2 pump ............................................................................................... 3-53.2 VDE inspections.............................................................................................................. 3-73.3 Calibrating the 4008 HDF scale ...................................................................................... 3-83.4 Calibrating the substituate sensor .................................................................................. 3-133.5 Calibrating the HDF blood pump .................................................................................... 3-153.6 Repair instructions .......................................................................................................... 3-17

4 Circuit descriptions and circuit diagrams .................................................................. 4-4.1 LP 625 display board ...................................................................................................... 4-34.2 LP 754 control board HDF .............................................................................................. 4-94.3 LP 760 HDF motor control .............................................................................................. 4-194.4 LP 761 HDF drive keyboard ........................................................................................... 4-274.5 LP 762 scale comm ........................................................................................................ 4-334.6 LP 763 SSE serial interface extender ............................................................................. 4-39

5 Spare parts .................................................................................................................... 5-5.0 How to use the spare parts catalog ................................................................................ 5-35.1 P.C.B.s ............................................................................................................................ 5-45.2 Scales ............................................................................................................................. 5-65.3 Substituate lift ................................................................................................................. 5-85.4 UF2 pump / blood pump (HDF) / valve V 126 ................................................................. 5-10

Page 8: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

0-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 9: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-1

Table of contents1 Description of machine functions

Section Page

1.1 Description .................................................................................................................... 1-31.1.1 Components .................................................................................................................... 1-31.1.2 Description ...................................................................................................................... 1-31.1.3 Component tests ............................................................................................................. 1-3

Fig.: Block diagram ......................................................................................................... 1-4Fig.: Flow diagram .......................................................................................................... 1-5Fig.: Flow diagram .......................................................................................................... 1-6

1.2 Description of extended T1 test / error messages in T1 test ................................... 1-81.2.1 Test UF-Function ............................................................................................................ 1-8

1.3 Error messages during treatment and the HDF test ................................................ 1-10

1.4 Error messages, substituate pump ............................................................................. 1-13

Page 10: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 11: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-3

1.1 Description

1.1.1 Components

The components of the 4008 HDF option are permanently connected to the hemodialysismachine. The 4008 HDF option comprises the following components:– Substituate lift– Scales (weighing range < 20 kg) with data interface and taring facility– Substituate pump with bidirectional data interface– UF2 pump

1.1.2 Description

The scales determine the actual weight of the substituate reserve and signals this to thehemodialysis machine. This calculates the substituate rate depending on the set treatment time.

A substituate sensor mounted on the substituate lift recognises whether:– Line set is fitted without substituate or– Line set is fitted with substituate.

The substituate pump delivers the substituate solution to the venous bubble catcher. The filtratefrom the dialysis fluid circuit is removed by the UF2 pump.

1.1.3 Component tests

The scales are checked by the operator by means of a plausibility test before treatment. Thesubstituate sensor is tested while the 4008 HDF is filling.The 4008 HDF test tests:– The substituate pump– The UF2 pump (electrically and hydraulically)(If the 4008 HDF test is not carried out then the connections are to be tested but once).The UF2 pump is tested cyclically during treatment.The electric control system of the UF2 pump is checked continuously.

Page 12: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Scales

CPU 1

UF2 pump Substituatepump

TMP

CPU 2W Serial communication

Serial communication

Ser

ial c

omm

unic

atio

nOperating system Safety system

Fig.: Block diagram

Page 13: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-5

UF

2p

um

pU

Fp

um

pF

low

pu

mp

Dia

lysa

tep

ress

ure

tran

sdu

cer

V12

6

V26

V24

V24

b

Dia

lyze

r

Air

det

ecto

r

Ven

ou

s b

ub

ble

cat

cher

Art

eria

lb

loo

d p

um

p

Ven

ou

s lin

ecl

amp

Su

bst

itu

ate

pu

mp

Su

bst

itu

ate

sen

sor

Sca

les

Su

bst

itu

ate

bag Pat

ien

t

Bal

anci

ng

ch

amb

er

LP

632

CP

U 2

LP

631

CP

U 1

LP

763

SS

ES

eria

lin

terf

ace

exte

nd

er

LP

633

Inp

ut

bo

ard

LP

634

Ou

tpu

tb

oar

d

seria

lin

terf

ace

X63

2/B

24X

632/

C27

X63

4 R

/C

11X

634

R/

A24seria

l int

erfa

ce

seria

l int

erfa

ce a

nd ta

ring

line

X63

2/A

7

X63

2/C

7

X63

2/A

29X

634

L/A

–C24

X63

4 L/

A–C

23

X63

3 L/

C14

X63

3 L/

C23

X63

3 L/ B6

Con

trol

, UF

2 pu

mp

Con

trol

, UF

pum

p

Ack

now

ledg

emen

t, di

alys

ate

pres

sure

Con

trol

, UF

pum

p

Con

trol

, UF

2 pu

mp

Ack

now

ledg

emen

t, U

F p

ump

Ack

now

ledg

emen

t, U

F2

pum

p

Ack

now

ledg

em.,

dial

y. p

ress

ure

XH

DF

2–4

X34

8v/4

–5

X63

3 R

/C

28–3

0

Fig.: Flow diagram

Page 14: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Fig.: Flow diagram

Page 15: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-7

Legend

2 Temperature sensor3 Temperature sensor5 Float switch6 Level sensor7 Conductivity measuring cell8 Blood leak detector9 Pressure transducer10 Reed contact for concentrate12 Reed contact for bicarbonate20 UF2 pump21 Flow pump22 UF pump23 Concentrate pump24 Dialyzer valve 124b Dialyzer valve 225 Bicarbonate pump26 Bypass valve29 Degassing pump30 Drain valve31 Balancing chamber valve 132 Balancing chamber valve 233 Balancing chamber valve 334 Balancing chamber valve 435 Balancing chamber valve 536 Balancing chamber valve 637 Balancing chamber valve 738 Balancing chamber valve 841 Water inlet valve43 Fill valve54 Heater rod61 Pressure reducing valve63 Filter/water inlet65 Loading pressure valve66 Heater block66a Water inflow chamber66b Heater rod chamber66c Float chamber68 Balancing chamber71 Filter/concentrate72 Filter/bicarbonate73 Filter/external dialysate74 Filter/UF75 External flow indicator76 Filter/fill valve

Hydraulic measurement points

A Reduced water inlet pressureB Balancing chamber loading pressureC Flow pump pressureD Degassing pump pressure

77 Heat exchanger78 Relief valve84 Disinfection valve85 Disinfectant conncector86 Recirculation valve87 Discharge valve88 Multifunction block88a Degassing chamber88b Secondary air separator88c Primary air separator89 Degassing orifice90a Rinse chamber concentrate90b Rinse chamber bicarbonate91 Rinse valve92 Vent valve94 Concentrate suction tube95 Bicarbonate suction tube97 Ventilation pump99 Rinse valve100 Rinse valve102 Concentrate valve in central delivery system104 Bicarbonate valve in central delivery system109 Temperature sensor111 Hydrophobic filter112 Vent valve114 Dialysate filter115 Disinfection valve sensor116 Sampleing valve117 Check valve (concentrate)118 Check valve (bicarbonate)119 Filter (concentrate)120 Filter (bicarbonate)121 Concentrate connector in central delivery system122 Bicarbonate connector in central delivery system123 Pressure switch for V 102124 Pressure switch for V 104125 Adapter plate126 4008 HDF vent valve148 Filter/rinse valve149 Filter/rinse valve150 Filter151 Orifice152 4008 HDF hydrophobic filter

Page 16: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-8 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1.2 Description of extended T1 test / error messages in the T1 test

Possible error messages displayed in the T1 test are listed in the Technical Manual for thehemodialysis machine. If the HDF option is installed and activated the following error codes canalso appear:

1.2.1 Test UF-Function

Test description:

CPU 1 starts up the UF2 pump at a defined rate.CPU 2 controls the hydraulic and electrical function of the UF2 pump.CPU 2 blocks the control line of the UF2 pump and checks for standstill.UF2 counter check.

Figure:

Page 17: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-9

Error description:

Error message Description

F11 UF-Function The interval between the strokes of the UF2 pump was less than220 ms. Correct delivery of the required volume is not safeguard-ed because the stroke return travel is too short.– A pump rate which was too high was signalled by the CPU 1.

F12 UF-Function The pulse time for the UF2 pump is less than 180 ms. Correctdelivery of the required volume is not safeguarded because thedischarge time is too short.– Monoflop on LP 634 defect (IC 42/R65/C45).

F 13 UF-Function The pulse for the UF2 pump is longer than 500 ms. A maximumrate of 5000 ml/h is not possible.– Monoflop on LP 634 defect (IC 42/R65/C45).

F14 UF-Function UF2 pump not active during the test (4 s).– Acknowledgement (UF_P2, X637/B26) → X632/C7 no LOW

pulse– Control line (UF_P2, X634L/A–C24) → X637/B26 no LOW

pulse

F15 UF-Function The UF2 pump cannot be stopped by the CPU2.– Control line (UF_P_EN, X632/C28) → X634R/A22 no 5 V.– Reset input to IC 42/pin 13 on LP 634 defect.

F16 UF-Function The UF pump acknowledgement from CPU1 is faulty.– Acknowledgement (UF_P2, X637/B26) → X633/C23 no LOW

pulse

F17 UF-Function The change in pressure after a UF2 pump stroke less than20 mmHg.– UF2 pump mechanically defect.– Control line (UF_P2_CRTL, X632/B24) → X634R/C11 no

HIGH pulse.

F20 UF-Function Difference in pressure between UF pump stroke and UF2 pumpstroke more than 20 %.– Stroke rate for UF pump or UF2 pump not set correctly.

Page 18: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1.3 Error messages during treatment and the HDF test

Message on thedisplay

Cause

F321 HDF-failure

F322 HDF-failure

F323 HDF-failure

F324 HDF-failure

F325 HDF-failure

The UF pump or UF2 pumptest will not be started as longas the level sensor (6) detectsair. The level sensor (6) willdetect air one minute after the4008 HDF test has been se-lected

The UF pump or UF2 pumphas carried out more than 50strokes and still no air hasbeen detected.

The substituate pump runs at arate of 10 ml/min. to raise thefluid level by a defined amountin the secondary air separator(the fluid will just be detected).Air will still be detected after 2minutes.

Since fluid has been recog-nized in the secondary air sep-arator the substituate pumpstarts at a rate of 10 ml/min.until a change in weight followson the scales (i.e. a jump ingrammes). There was no dif-ference in weight establishedafter one minute.

If the expected change inweight (i.e. a jump ingrammes) is higher or lowerthan –1 g, the respective UFpump test will be repeated.The UF pump test was repeat-ed more than 5 times.

Possible error elimination

Check level sensor (6) (de-fect).

Check V126 (does not open,venous bubble catcher col-lapses), check V43 (whetherleaking or open).Check UF pump or UF2 pump,respectively (mech. defect).

Check substituate tubing(clamped off, not connected tothe venous bubble catcher).

Check substituate tubing(clamped off).

Check bag (fluctuations in airdraft)Check scales (for drifting).

Page 19: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-11

The fluid level in the second-ary air separator is raisedagain by the substituate pumpafter 100 strokes of the UFpump or UF2 pump, respec-tively, until fluid is detected.Plus run-on time at a rate of 10ml/min. for the gramme jump.Air will still be detected after aspecific period of time (de-pending on the delivery rate ofthe substituate pump).

One UF pump did not pass thetest. The filling volume for thesecondary air separator is notwithin the given tolerance of100 ml; ± 5 ml. Should the testgive a reading of over 105 mlthe cause could also be airtaken in the flow from a badlyvented dialysator.

Interval between two strokesof the UF pump less than220 ms.

Pulse time of a stroke of UFpump less than 180 ml.

Pulse time of a stroke of UFpump more than 500 ms.

Starting time for the UF pumpmore than 10 seconds.

Difference between desired oractual delivery rate of the UFpump greater than + 10%.

UF pump stops longer than themaximum period time.

Despite switched off ultrafiltra-tion the change in delivery rateof the UF pump is greater than10 ml.

Interval between two strokesof the UF2 pump less than220ms.

Check substituate hose(clamped off, not connectedwith the venous bubblecatcher).

Check UF pump or UF2 pump,respectively (not calibrated,mechanical defect).

Check CPU 1 (defect).

Check LP 634 (regulatingmonoflop defect).

Check LP 634 (regulatingmonoflop defect).

Check LP 634 (regulating endstage defect).

Check CPU 1 / CPU 2(communication problems).

Check LP 634 (regulating endstage defect).Check UF pump(interruption, control).Check CPU 1 / CPU 2(communication problems)

Check CPU 1 / CPU 2(communication problems)

Check CPU 1 (defect).

F326 UF-failure

UF1 volume-ErrororUF2 volume-Error

F327 UF-failure

F328 UF-failure

F329 UF-failure

F330 UF-failure

F331 UF-failure

F332 UF-failure

F333 UF-failure

F334 UF-failure

Message on thedisplay

Cause Possible error elimination

Page 20: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Check LP 634 (regulatingmonoflop defect).

Check LP 634 (regulatingmonoflop defect).

Check LP 634 (regulating endstage defect).

Check CPU 1 / CPU 2(communication problems).

Check LP 634 (regulating endstage defect).Check UF pump (interruption,control).Check CPU 1 / CPU 2(communication problems).

Check CPU 1 / CPU 2(communication problems).

Check UF pump(spring, screen)

Check UF2 pump(spring, screen)

Check UF pump and UF2pump (delivery volume)

Check monitor / scales(communication problems).

Check monitor / scales(communication problems).

Check V24B and V43 feed-back lines.Check sense of rotation of as-piration pump.Check hydraulics unit forleaks.

Check scalesCheck substituate bag

Check substituate bag(weight fluctuations)Check HDF control system(greater fluctuations)

Pulse time of a stroke of theUF2 pump less than 180 ms.

Pulse time of a stroke of theUF2 pump more than 500 ms.

Reaction time of the UF2pump more than 10 seconds.

Difference in desired or actualdelivery rate of UF2 pumpgreater than ± 10 %.

UF2 pump stops longer thanthe maximum period time.

Despite switched off 4008HDFchange in delivery rate of theUF2 pump greater than 10 ml.

Failure of UF pump.

Failure of UF2 pump.

Volume difference between UFpump and UF2 pump.

Balance failure recognized byCPU2 greater than ± 500 ml.

Bolus exceeded by more than+ 20 ml, detected by CPU2.

CPU2 failed to perform the cy-clic UF pump test within 5 min-utes.

Weight change on the scalesduring the weighing test great-er than 2g.

The delivery rate correctionfactor could still not be deter-mined after 20 minutes treat-ment time.

Possible error eliminationMessage on thedisplay

Cause

F335 UF-failure

F336 UF-failure

F337 UF-failure

F338 UF-failure

F339 UF-failure

F340 UF-failure

F341 UF-failure

F342 UF-failure

F343 UF-failure

F344 HDF-failure

F345 HDF-failure

F346 HDF-failure

F348 HDF-failure

F349 HDF-failure

Page 21: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 1-13

Code Description of failure

E.01 Line diameter outside the permitted range

E.02 Non-defined hexaswitch position

E.03 Venous pressure transducer not balanced

E.04 Failure, running time monitoring system, SN operation

E.05 SN stroke volume outside the permitted range

E.06 The SN pressure thresholds outside the value range of the AD converter

E.07 Not defined

E.08 Failure in the AD conversion

E.09 Not defined

E.10 Not defined

E.11 Not defined

E.12 Failure, speed monitoring (hall sensor)

E.13 Failure, monitoring system, current sensing resistors

E.14 Failure, monitoring system, current sensing resistors

E.15 Failure, speed monitoring system

1.4 Error messages, substituate pump

Page 22: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

1-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 23: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 2-1

2 Technical safety checks

General Notes

This chapter includes all necessary technical safety checks (TSC).

These inspections must be carried out every 12 months.

The technical safety inspections stipulated for the hemodialysis machine must be carried out in addition tothese technical saftey checks.

The technical safety checks are to be recorded in the equipment log.

A technical safety checks report is to be found at the page 2-3.

Page 24: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

2-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 25: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 2-3

Technical safety checks report Seite 1/1

Manufacturer: ............................. Date: .............................

Machine: ............................. Technician: .............................

Operating hours: .............................

The following checks and inspections have to be carried out on the machine at least every 12months by persons who are capable of carrying out such technical safety checks efficiently asa result of their training, knowledge and experience gained in practice and are not subject toinstructions as to their checking activity.(See also BMA Bulletin dated 02.04.1987)

TSC WA No. OKDescription Desired value/function

Extent of checks and time limits for technical safety checks.

TIME LIMIT: every 12 months

1 UF2 pump

TSC 1.1 Delivery volume (1 ml/stroke) 60 strokes = 60 ml ± 1 %

2 VDE inspections (values according to EN 60601-1)

TSC 2.1 Protective ground resistance max. 0.3 Ω

TSC 2.2 Summarized leakage current Must fulfill both conditions:1. No more than 1.5 times the “summari-

zed value first measured”(“Summarized value first measured”:Refer to the machine card enclosed withthe machine.)

2. No more than 1 mA

The technical safety checks are to be recorded in the equipment log and the results of thechecks documented.

If the machines are not safe in function and/or operation they are to be repaired or the operatoris to be informed as to the dangers involved when using the machine in its given state.

The correct completion of the listed work and the correctness of statements made are herebycertified.

Signature, technician: Signature, customer:

........................... , date ............................. ................................ , date .............................

.................................................................... .........................................................................

Page 26: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

2-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 27: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-1

Table of contents3 Calibration instructions

Section Page

3.0 General information on the calibration instructions ................................................ 3-3

3.1 Calibrating the UF2 pump ............................................................................................ 3-5

3.2 VDE inspections ............................................................................................................ 3-7

3.3 Calibrating the 4008 HDF scales ................................................................................. 3-8

3.4 Calibrating the substituate sensor .............................................................................. 3-13

3.5 Calibrating the HDF blood pump ................................................................................. 3-153.5.1 Pressure transducer ....................................................................................................... 3-153.5.2 Checking SN switchover points ...................................................................................... 3-16

3.6 Repair instructions ....................................................................................................... 3-17

Page 28: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 29: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-3

I/O

Spülen

Heißreinigen

Desinfektion

SingleNeedle

Test

Vorbereiten

Dialyse Start

AlarmTon Aus

Best. Auswahl

Luftdetektor ArteriellerDruck

VenöserDruck TMP Leitfähigkeit

Blutleck

Überbrücken

Ultrafiltration

Reset

UF Menge

UF Variation

Fluß

Temperatur

Prog.I/O Prog.I/O

Bic.Konz.

SetI/O

mmHgkPa mmHgkPa mmHgkPa

ml

mS/cm(25°C)

ml/min

°C

Set

UF Rate ml/h

UF Ziel ml

UF Restzeit h:min

4008

StartStop

Rate: ml/min(Ø: mm)

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

Pven.

StartStop

Rate: ml/h(Bolus: ml)

Bolus

Rate

( :h.min)

–40

0

–100

–200

–300

100

200

280

0

10

20

30

–10

–20

–30 0

100

200

0

10

20

30

300

400

500

40

50

60

0

100

200

0

10

20

30

300

400

500

40

50

60

13

13.5

14

14.5

15

15.5300500700900

35

37

39

Fresenius Medical Care

Fresenius Medical Care

LP 763

LP 762

LP 761

LP 625

LP 754

LP 760

3.0 General information on the calibration instructions

Measuring instruments:The same measuring instruments as used for the hemodialysis machines are to be employed.Required in addition is a calibration weight PT6 – 5000 g.

Fig.: P.C.B. overview 4008 HDF

Page 30: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1 2 3 4 5 6 7 8ON

OFF

LP 632 CPU 2

SW1

Fig.: DIP switch 4008 HDF

SW 1, DIP switch 8:ON: 4008 HDF test selectable (required/not required)OFF: 4008 HDF test carried out automatically

Page 31: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-5

Select ADJ. UF-PUMP VOLUME bypressing the and keys.

Press the Confirm key.

Press the Confirm key.

Select UF-PUMP 2 by pressing the and keys.

Press the Confirm key.

3.1 Calibrating the UF2 pump

Switch off hemodialysis machine.

Set service switch to ON (top).

Switch on hemodialysis machine.

Confirm SelectCALIBRATION

Confirm SelectADJ. UF-PUMP VOLUME

Confirm SelectUF-Pump 2

Enter the number of strokes by pressingthe and keys.

Confirm Selectpulse-amount = 60 –40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display for approx 3 sec.Confirm SelectACKNOWLEDGED

Page 32: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Text display

Draw off UF 2 pump line, close T-piece,hang line in a measuring cylinder.

Confirm Selectpress uf key

Press the key UF I/O.

The UF LED emits light.

The UF2 pump delivers.

35

37

39

UF Variation

Prog.I/O Prog.I/O

BicConc

Set

Time Left h:min

The remaining UF Pulses are shown inthe display.

Confirm Selectuf pulses left = 59

Ultrafiltration

ResetVolume

UF

Prog.I/O

UF Volume ml

UF Rate ml/h

UF Goal ml

Time Left h:min

0060

The number of preselected pulses isshown in the UF Goal display.

Text display after completion.

Check delivered quantity.If necessary adjust UF 2 pump and re-peat process.

Delivery rate: 1 ml/stroke(60 strokes = 60 ml)Tolerance: ± 1 %.

Confirm SelectUF-Pump 2

Page 33: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-7

3.2 VDE inspections

(Values in accordance with EN 60601-1)

.1 Protective ground resistance:

– max. 0.3 Ω.

Measuring pointProtective

ground resistance test

.2 Summarized leakage current

Must fulfill both conditions:1. No more than 1.5 times the “summarized value first measured”

(“Summarized value first measured”: Refer to the machine card supplied with the machine.)2. No more than 1 mA

The measurements must be taken in the dialysis mode of operation in the “ON phase” of theheating control system.The scales must be travelled out to such an extent that neither of the two end switches areactuated (middle position).The two leakage currents are to be measured each time with different mains connection poling.The two leakage currents are added together and give the summarized leakage current.

Page 34: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-8 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

3.3 Calibrating the 4008 HDF scales

Note:The prompts shown in the text dis-play have to be run through correctlyby the technician since no direct re-turn signal is given by the scales.

Fresenius Medical Care

Select CALIB. HDF-SCALE by pressingthe and keys.

Press the Confirm key.

Brief text display.

Display on scales is not illuminated.

Switch off hemodialysis machine.

Set service switch to ON (top).

Switch on hemodialysis machine.

Confirm SelectCALIBRATIONPress the Confirm key.

Confirm SelectCALIB. HDF-SCALE

Confirm SelectCHECKING HDF-SCALE!

Text display.Confirm Selectcalib. Switch ON?

➀ Open housing of scales at the back.

➁ Set calibration switch for the scales inON position.

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Page 35: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-9

Text display.Confirm SelectHDF-DISPL shows C?

500 100

g

C

Fresenius Medical Care The display on the scales is active.A “C” appears at the top right.

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.No weight is to be on the scales.

Confirm SelectHDF-SCALE unloaded ?

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.Confirm SelectHDF-DISPL shows 0 ?

Scales display.

500 100

g

C

Fresenius Medical Care

Page 36: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.

Scales display: 5000 g

Hang calibration weight PT6 – 5000 g onscales.

Confirm Selectload 5 kg CAL. WEIGHT

500 100

C

+

Fresenius Medical Care

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.Confirm SelectDISPL shows UNIT “g”?

Scales display: 5000 g.

500 100

g+

Fresenius Medical Care

Page 37: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-11

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.Confirm Selectcalib. Switch OFF ?

500 100

g+

Fresenius Medical Care

➀ Set calibration switch for the scales toOFF position.

➁ Close housing.

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Text display.

Remove calibration weight.

Confirm SelectHDF-SCALE unloaded ?

–40

0

–10

–20

–30

SingleNeedle

Test

Prime

Override

Press the Override key.

Brief text display.Confirm SelectSCALE CALIBRAT. done

Text display.Confirm SelectCALIB. HDF-SCALE

Page 38: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Switch off hemodialysis machine.

Set service switch to OFF (bottom).

Switch on hemodialysis machine.

Hang calibration weight PT6 – 5000 g onscales.

Scales display: 5000 g

500 100

g+

Fresenius Medical Care

Remove calibration weight.

Calibration of 4008 HDF scalescompleted.

Page 39: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-13

3.4 Calibrating the substituate sensor

I/O

Spülen

Heißreinigen

Desinfektion

SingleNeedle

Test

Vorbereiten

Dialyse Start

AlarmTon Aus

Bes

Luftdetektor ArteriellerDruck

VenöserDruck TMP

Blutleck

Überbrücken

Ultrafiltration

Reset

UF Menge

UF

Prog.I/O

mmHgkPa mmHgkPa mmHgkPa

ml

UF Rate ml/h

UF Ziel ml

UF Restzeit h:min

4008

Rate: ml/min(Ø: mm)

Rate: ml/h(Bolus: ml)

Bolus

( :h.min)

–40

0

–100

–200

–300

100

200

280

0

10

20

30

–10

–20

–30 0

100

200

0

10

20

30

300

400

500

40

50

60

0

100

200

0

10

20

30

300

400

500

40

50

60

Fresenius Medical Care Place filled substituate line in sub-stituate sensor.

Set potentiometer P1 on the LP 754 sothat the LED D26 just illuminates.

XHDF

XW

AA

GE

X348V_SN

P1

IC14

X348V

IC17

IC2

D26

IC8 TP

0FED

CB A9 876

54

321

I/O

Spülen

Heißreinigen

Desinfektion

SingleNeedle

Test

Vorbereiten

Dialyse Start

AlarmTon Aus

Bes

Luftdetektor ArteriellerDruck

VenöserDruck TMP

Blutleck

Überbrücken

Ultrafiltration

Reset

UF Menge

UF

Prog.I/O

mmHgkPa mmHgkPa mmHgkPa

ml

UF Rate ml/h

UF Ziel ml

UF Restzeit h:min

4008

Rate: ml/min(Ø: mm)

Rate: ml/h(Bolus: ml)

Bolus

( :h.min)

–40

0

–100

–200

–300

100

200

280

0

10

20

30

–10

–20

–30 0

100

200

0

10

20

30

300

400

500

40

50

60

0

100

200

0

10

20

30

300

400

500

40

50

60

Fresenius Medical Care Place empty substituate line in sub-stituate sensor.

Page 40: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Adjust potentiometer P1 on LP 754 untilLED D26 just turns off.Count the number of revolutions thepotentiometer is turned. Keep thisnumber in your mind.0FE

DC

B A9 876

54

321

XHDF

XW

AA

GE

X348V_SN

P1

IC14

X348V

IC17

IC2

D26

IC8 TP

Turn back the potentiometer P1 on theLP 754 by half this number.The LED D26 will illuminate again.

XHDF

XW

AA

GE

X348V_SN

P1

IC14

X348V

IC17

IC2

D26

IC8 TP

0FE

DC

B A9 876

54

321

Page 41: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-15

3.5 Calibrating the HDF blood pump

3.5.1 Pressure transducer

X348V_SN XHDF

XW

AA

GE

0FE

DC

B A9 876

54

321 Switch the hemodialysis machine off.

Set HEX switch on the LP 754 toposition “F”.

Switch the hemodialysis machine on.

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

Display 000

Open SN pressure transducer to atmos-phere.

Press the Start/Stop key.

Display 250

Apply a pressure of 250 mmHg to the SNpressure transducer.

Press the Start/Stop key.

Page 42: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-16 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

Display – – –

XHDF

XW

AA

GE

0FED

CB A9 876

54

321

X348V_SN

Switch off hemodialysis machine.

Set HEX switch on the LP 754 toposition 1.

Balancing out has been completed.

3.5.2 Checking SN switchover points

Bottom switchover point: fixed at 75 mmHg (± 6 mmHg)

Top switchover points (± 6 mmHg):

Stroke volumes (ml) 10 15 20 25 30 35 40 45 50

Top switchover point 110 130 150 172 195 219 244 270 299

Page 43: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-17

3.6 Repair instructions

Important:If the substituate lift has to be dis-mounted for repair the seal is to bechecked and if necessary renewedbefore fitting the lift again.Drip water protection!

Page 44: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

3-18 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 45: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-1

Table of contents4 Circuit descriptions and circuit diagrams

Section Page

Fig.: Connection diagram 4008 HDF .............................................................................. 4-3Fig.: Wiring diagram 4008 HDF ...................................................................................... 4-5Fig.: Block circuit diagram, mains voltage 4008 HDF .................................................... 4-7Fig.: Block circuit diagram, low voltage 4008 HDF ........................................................ 4-7

4.1 LP 625 display board .................................................................................................... 4-94.1.1 Description ...................................................................................................................... 4-94.1.2 Circuit and component layout diagram ........................................................................... 4-11

4.2 LP 754 control board HDF ............................................................................................ 4-154.2.1 Description ...................................................................................................................... 4-154.2.2 Circuit and component layout diagram ........................................................................... 4-17

4.3 LP 760 HDF motor control ............................................................................................ 4-234.3.1 Description ...................................................................................................................... 4-234.3.2 Circuit and component layout diagram ........................................................................... 4-25

4.4 LP 761 HDF drive keyboard ......................................................................................... 4-294.4.1 Description ...................................................................................................................... 4-294.4.2 Circuit and component layout diagram ........................................................................... 4-31

4.5 LP 762 scales comm ..................................................................................................... 4-354.5.1 Description ...................................................................................................................... 4-354.5.2 Circuit and component layout diagram ........................................................................... 4-37

4.6 LP 763 serial interface extender .................................................................................. 4-414.6.1 Description ...................................................................................................................... 4-414.6.2 Circuit and component layout diagram ........................................................................... 4-43

Page 46: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 47: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-3

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

XHDF

X63

Z

X348v

LP 630motherboard

LP 754control board

HDF

SNXHDF

X348v

XWAAGE

Substituatepump

X63

9B

LP 639power logic(4008 / 4008 E)

LP 744power control 1(4008 B)

MMotorSubstituate lift

Auxiliary socket

X5

X4

X2

X3

X1

LP 760HDF motorcontrol

X2

X1

LP 761HDF drivekeyboard

Substituatesensor

X2X1

X3

X4

LP 762scalescomm

ModuleScales

ModuleScales

SwitchSubstituate lift

LP 763SEE serial interface

extender

X637C

X637B X637B

X637C

XA1P

XA4V0

UF2 pump

V126

LP 747 distribution board

(X74

4AX

744C

)

Fig.: Connection diagram 4008 HDF

Page 48: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 49: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-5

no assignment

24V1SE12

AGND (SE2)3SE34SE45

WRXD6WTXD7TARA8

no assignment9no assignment10no assignment11no assignment12MOTOR_CTRL13

PWR_GND14

123456789

1011121314

XW

AA

GE

X3

BS

ST

1S

N2

BP

ST

3T

XD

4R

XD

5B

PS

B_V

EN

6A

DK

S7

BP

US

_VE

N8

no a

ssig

nmen

t9

BP

R_V

EN

10S

NS

T11

no a

ssig

nmen

t12

no a

ssig

nmen

t13

VE

N_B

PR

_SE

T14

AG

ND

15+

12V

16P

WR

_GN

D17

PW

R_G

ND

1824

V19

24V

20

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

HD

F_O

N1

WT

XD

2W

RX

D3

TAR

A4

BL_

AL

5H

DF

_LO

G2

6H

DF

_LO

G1

7no

ass

ignm

ent

8+

12V

9A

GN

D10

1 2 3 4 5 6 7 8 9 10

X348v XHDF

X348v XHDF

LP 754control board HDF

+12

V1

AG

ND

2TA

RA

3R

XD

4T

XD

5E

ND

_SW

6+

12V

7–

8

1 2 3 4 5 6 7 8

X1

X1

+12V1UP2

STOP3DOWN4

SE15SE26SE37SE48

9no assignment10

123456789

10

X2

LP 760HDF motor

control

X2

X1

LP 761HDFdrive

keyboard

DG

ND

A1

DG

ND

C1

PW

R_G

ND

A28

PW

R_G

ND

C28

+24

VA

29+

24V

C29

+12

VA

31+

12V

C31

+5V

A32

+5V

C32

RX

D_E

XT

_1C

20T

XD

_EX

T_1

C21

A1

C1

A28

C28

A29

C29

A31

C31

A32

C32

C20

C21

X63Z

X63Z

LP 630motherboard

LP 763SEE serial

interface extender

no assignment1no assignment2no assignment3

GND4no assignment5

TXD6GND7

no assignment8GND9

PRINT10TARA11

no assignment12no assignment13no assignment14

123456789

1011121314

X3

W1IW21F22F13CF4

ON/OFF5no assignment6

123456

X4

LP 762scalescomm

Modulescales

+12

VE

ND

_SW

1 2

Switch substituate lift

X2

Substituate sensor

SU

B_A

1S

UB

_K2

SU

B_C

3P

WR

_GN

D4

no a

ssig

nmen

t5

no a

ssig

nmen

t6

L11

L12

N2

3P

E4

X4

PE

1P

E2

N3

LC4

X5

M

X63

9B

N1L12 3 4

LP 639power logic

(4008 / 4008 E)

LP 744power control 1

(4008 B)

X637B X637C

X637B X637C

UF

_P2

26–2

826

–28

V12

614

14

XA1P XA4V0

UF

_P2

10

V12

69

LP 747distribution

board

UF2 pump V 126

X744A/1X744C/3

Fig.: Wiring diagram 4008 HDF

Page 50: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 51: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-7

Mainsplug

Mainsfilter

Mainsswitch

(on backof machine)

511 K 1 µF T 2,5 A

T 2,5 A

Mainsrelay

T 2,5 A T 2,5 A

Maintrans-

former

Standby-trans-

former

Auxiliarysocket

Direction of turnswitchover

Motor relay

M

24 Vswitch controller

Imax ≈ 15 A(short circuit

cut-out)

12 Vswitch controller

Imax ≈ 4 A(current

limitation)

LP 630motherboard

LP 633input board

LP 634output board

LP 754control board

HDF

LP 760HDF motor control

+24 V +12 V

Linear controllercurrent restrictedinternally to approx. 1 A

LP 761HDF drivekeyboard

+12 V

Scales+12 V

Maintransformer

Fig.: Block circuit diagram, mains voltage 4008 HDF

Fig.: Block circuit diagram, low voltage 4008 HDF

Page 52: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-8 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 53: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-9

4.1 LP 625 display board

4.1.1 Description

General

The displays and the keyboard are on this P.C.B..

Plug connections to LP 625:– X189, connection to LP 760

Display

All necessary information is given to the user through 6 multiplex 7-segment displays and 2 LEDswhich are mounted on the socket.

Keyboard

The keyboard comprises 3 single non-illuminated keys and 1 key with integrated LED (HDF). Thekeys are soldered directly to the P.C.B..The user receives a acknowledgement through a spring contact when the keys are pressed.

Page 54: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 55: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-11

4.1.2 Circuit and component layout diagramLP 625 display board

LP 625Circuit diagram

ABCDEFF E D C B

AENDERUNG

NAME

NAME

AUS

TAG

ERSATZ

ERSATZ

FORMAT

BE

GE

NO

TAG

12

34

56

78

12

34

5

BLATT

01234567891011121314151617181920

MEDIZIN-

TECHN.

SYSTEME

SP

DISPLAY BOARD

1(1)

673872

1992

ST

LP625

30.06.

24.11.

D.KOE

aneue Zeichnung

19.0

2.93

Asch.

Asch.

bR8

6(R1

3)be

st.

12.0

3.93

cSH

5,IC

1 ne

u12

.07.

94

AN

AN

a gb

ec

df

DP

AN

AN

a gb

ec

df

DP

AN

AN

a gb

ec

df

DP

AN

AN

a gb

ec

df

DP

AN

AN

a gb

ec

df

DP

AN

AN

a gb

ec

df

DP

A8

A7

A6

A5

A4

A3

A2

A1

Y1

Y2

Y3

Y4

Y5

Y6

Y7

Y8

G2

G1&

EN

GND

+5V

+24V

X

+24V

+24V

+5V

+5V

+5V

+5V

+5V

+5V

+5V

+5V

+5V

1 2 3 4 5 6

10

9 8 7

DP1

HDSP7801

1 2 3 4 5 6

10

9 8 7

DP2

HDSP7801

1 2 3 4 5 6

10

9 8 7

DP3

HDSP7801

1 2 3 4 5 6

10

9 8 7

DP4

HDSP7801

1 2 3 4 5 6

10

9 8 7

DP5

HDSP7801

1 2 3 4 5 6

10

9 8 7

DP6

HDSP7801

119 2 3 4 5 6 7 8 9

18

17

16

15

14

13

12

11

IC1

74HCT541

87 6 5 43 2 1

LD1

HLMP2655

12

LD2

HLMP2400

34

LD2

HLMP2400

R3

4K99

R1

4K99

R6

7K50

R5

7K50

R4

7K50

R9

7K50

R8

7K50

R7

7K50

R2

4K99

R10

4K99

R11

121R

1 2 3 4 5 6 7 8

16

15

14

13

12

11

10

9

RN1

8X220R

R12

150R

R13

4K99

12

SH1

RF15

12

SH2

RF15

12

SH3

RF15

21

SH4

MTG

SH5

RF15

T8

BS170

T3

BC327-40

T1

BC327-40

T6

BC327-40

T4

BC327-40

T2

BC327-40

T7

BC547B

T5

BC327-40

1X189

10X189

11X189

12X189

13

X189

14

X189

15

X189

16

X189

17

X189

18

X189

19

X189

2X189

20

X189

21

X189

22X189

23

X189

24

X189

25X189

26

X189

3X189

4X189

5X189

6X189

7X189

8X189

9X189

Page 56: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 57: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-13

LP 625Component layout diagram

Page 58: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 59: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-15

4.2 LP 754 control board HDF

4.2.1 Description

General

The control, section and power pack are on the this P.C.B..

Plug connections to LP 754:– X186, connection to position sensor– X188, connection to stepper motor– X189, connection to LP 748– X190, connection to cover switch– X192, connection to pressure transducer– X348, connection to dialysis monitor

Hexswitches on the LP 754:0 No function1 Single needle blood pump / 4008 HDF / ON LINE HDF2 – A No functionB Blood pump stop alarm (15, 30 secs.)C No functionD Call-up operating time meter (display x 100 = number of hours)E Test operation (only manufacturer)F Pressure transducer calibration

Voltage generation

The +24 V and +12 V voltage supply is made available to the blood pump by the monitor. The +5V supply voltage is generated on the module by the switch controller IC 20 from the +24 V voltageto minimize loss of power.

Stepper motor control

The stepper motor is run in microstep operation to reduce noise. The resolution is 60 microstepsper step. The RISC processor transmits a 8-bit dataword alternately on the pins 3 and 5 of the DAconverter from IC 7. Two sine form voltages are available on the output of the converter, phaseoffset by 90°. They are conducted to the stepper motor controller IC 2 together with the currentdirection signals. Together with the two SM drivers IC 1/IC 22 and the current sensor resistorsR58/R59 as feedback these form a closed control circuit which determine two sine-shapedcurrents, phase offset by 90°, in the two coils.

Page 60: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-16 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Microprocessor

The cycle frequency of the microprocessor is determined by the quartz Q2 between the processorconnections 39 and 40. Quartz oscillation is made possible by the capacitors C5 and C6.The keyboard signals and the signals BSST, BPST are read-in through the port P4.The IC 9 serves as intermediate memory for the addresses AO/A7.The ALE signal on pin 50 of the microprocessor is the control line of the data address latch.The signal from the revolution and direction of turn sensor (position sensor) in the pump bed isread-in through port T1.The operation data of the pump is saved in the NOVRAM IC 21 through ports 5.1 to 5.3 when thedialysis machine is switched off or there is a voltage failure. Undervoltage detection (power down)is through the comparitor IC 23.With the WR line the data takeover for the display is controlled in the external data latch IC 14.

PLL intermediate circuit

PLL component IC 4 together with the counter IC 19 causes the frequency coming from theprocessor to control the stepper motor to be multiplied otherwise the processor would be too slowto generate this frequency.

RISC processor

The RISC processor IC 5 is supplied with a cycle signal from the processor. With each cycle an8-bit data word is read from a look up table by the processor alternately for each phase of thestepper motor, respectively. Included in this data word is both the current direction as well as thecurrent desired value.A watchdog for the CPU is also incorporated in the RISC processor. When a cycle is missing onpin 8/IC 5 the RISC processor triggers a reset by the CPU through pin 7.

Display control

The data word for controlling the display is stored in IC 14. The multiplex operation of theindividual digits is made possible through the decoder IC 18.

Speed and flow

The speed is transmitted by the processor through port P 1.1 and the flow through port P 1.2 tothe dialysis monitor.

Pressure measurement

The differential measurement amplifier for the pressure transducer is formed by the IC 6 (1/2/3)and IC 6 (5/6/7). The measuring signal is read-in by the processor through the AD converter inputAN 7 and calibrated for zero and steepness by the software. Afterwards the measuring signal ismodulated to pulse width through port P1.3 and transmitted through a subsequent DA converterIC 11 (5/6/7) to the monitor.

Page 61: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-17

4.2.2 Circuit and component layout diagramLP 754 control board HDF

LP 754Circuit diagram 1/2

Page 62: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-18 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 63: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-19

LP 754Circuit diagram 2/2

Page 64: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-20 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 65: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-21

LP 754Component layout diagram

Page 66: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-22 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 67: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-23

4.3 LP 760 HDF motor control

4.3.1 Description

General

The LP 760 serves to control the lift drive.Lift operation presupposes that the scales are swung away. This is recognized by a limit switch.The scales can be moved upwards or downwards by pressing the or keys. The drive isswitched off automatically, likewise by limit switches, when reaching the uppermost or bottom-most position. When the scales are being travelled upwards or downwards they can be stoppedat any time with the Stop key.The motor for the lift drive is safeguarded against overloading by means of an integratedthermoswitch. If this protection facility responds the lift will come to a standstill. The lift can beused again after the motor has cooled down.

Circuit structure

Supply voltageThe control circuit is supplied with +12 V. This voltage is generated from the +24 V voltage of thehemodialysis machine by the linear controller T1 in conjunction with the capacitors C4, C5 andC6.

Motor controlTwo types of circuit are implemented to control the motor:a) continuous operation (R21 with components, R22 without componentsb) inching operation for downwards movement (R21 without components, R22 withcomponents).

To Point a):One D flip-flop is available for each direction of movement (IC 3). The flip-flops are set by a highlevel on X2/2 (key ) or X2/4 (key ). The flip-flop for the respective other direction of movementis reset at the same time by pressing the appropriate key. The two flip-flops are reset through highlevel on X2/3 (Stop key). This also takes place should the head of the scales be swung over themachine (high level on X1/6).

To Point b):There is only one flip-flop available for the upwards movement (IC 3/pin 1/2). The flip-flop for thedownward movement is set through IC 5. The control logic for the downward operation is notinfluenced by this (see Point a). However, the drive is only active as long as the key is pressed.

The IC 5 works as an 8 to 1 multiplexer. The address (0 to 7) of the input to be interconnected isprovided through the pins 11 to 13. Only the address 3 leads to a low level on output IC 5/14 and,as a result, to the motor starting. R 20 serves as a pull-up resistor. A defined output level isobtained with the IC 5/14 in high ohmic state, caused by pressing the Stop key (high level on IC5/15).

Page 68: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-24 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Other circuit sections

The motor feed lines are circuited through the relay RL 1 and RL 2. The position of RL 1determines the travel direction of the lift. The motor is switched on and off through RL 2. Bothrelays are controlled by the transistors T 2 (RL 1) and T 3 (RL 2). T 2 is controlled directly from thenon-inverted output IC 3/1. The output signals of both flip-flops, IC 3/2 and IC 3/12 (inverted) areused to control T 3 and are linked with each other through IC 1/1 and IC 1/2. In the case of lowlevel on one of the two flip-flop outputs IC 3/2 or IC 3/13 (i.e. the respective flip-flop has been set)the IC 1/pin 3 links up high level and T 3 switches RL 2.

Motor current recognition

The motor current is monitored while the lift is operating. The internal thermal-protective switchresponds in the case of the motor becoming overheated. Both flip-flops (IC 3) are reset by motorcurrent recognition so that the motor does not start up again automatically after the drive hascooled down. The voltage drop is evaluated through the high load resistor R 18 for motor currentrecognition. The generated alternate voltage caused by the current flow through R 18 isconnected to the capacitor C 14 and C 13 through the diodes D 4 and D 5. The resulting directvoltage through R 13 is approx. double so high as the peak value of a half wave of the alternatevoltage on R 18 less the on state voltages of D 4 and D 5. The direct voltage controls theoptocoupler IC 4 through R 14. R 13 serves to completely discharge C 13 and C 14 when themotor is switched off.IC 4 ensures protective isolation of the control circuit from the mains voltage conductingcomponents of the current recognition system.The output of the optocoupler is circuited with the pull-up resistor R 19 against 12 V. High level onthe collector of IC 4 resets both flip-flops through IC 1/pin 12 and 13 and IC 1/8. This would leadto continuous blocking of the control circuit unless further measures were taken. This is why theoutput of IC 4 is connected up parallel to a further transistor T 4. In switched off state the capacitorC 12 is charged through high level on IC 1/4 through D 3 and R 10. This makes the transistor T 4conductive. After setting one of the flip-flops (IC 3) the IC 1/4 goes to low. C 12 discharges itselfthrough R 11, R 12 and the B-E path from T 4. T 4 blocks after approx. 500 ms. The outputtransistor must be linked by IC 4 before this time has elapsed otherwise both flip-flops will be resetagain.

Page 69: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-25

4.3.2 Circuit and component layout diagramLP 760 HDF motor control

LP 760Circuit diagram

ABCDEFG

12

34

56

78

910

BCDEFGH

12

34

5

H

BLATT

NAM

TAG

AUS

BE

GE

NO

AENDERUNG

TAG

NAME

ERSATZ

ERSATZ

FORMAT

01234567891011121314151617181920

MEDIZIN-

TECHN.

SYSTEME

A3

1994

1(1)

SP

673 720

LP760

21.09.

OTT

C1-C3 STUETZKONDENSATOREN

HDF-

MOTO

R-ST

EUER

UNG

HDF-

MOTO

R-CO

NTRO

L

(1)

(2)

(3)

T100mA

*

*

R22

NICH

T BE

STUE

CKT

/ NO

T FI

TTED

M

aX1

.1wa

r +2

4V

>1_

>1_

I

AD

O

1D

S R

C1

1D

S R

C1

X7

X6

X5

X4

X3

X2

X1

X0

CBAZINH

EN

+24V

AGND

AGND

AGND

+12V

AGND

UP

STOP

AGND

DOWN

AGND

+12V

AGND

AGND

+12V AGND

END_SW

AGND

AGND

SE1

SE2

SE3

SE4

+12V

+12V

SE1

SE2

SE3

SE4

+24V

RXD

+24V

AGND

AGND

+24V

AGND

AGND

+24V

L1

L1

N2

PE

PE

N LC

TXD

TARA

MOTOR_CTRL

AGND

AGND

TARA

RXD

TXD

+12V

+12V

AGND

+12V

+12V

AGND

+12V

+12V

AGND

AGND

AGND

AGND

PE

+12V

AGND

R5

73K2

2 3 4 5

1IC2

4072

910

11

12

13

IC2

4072

1

2

3

T1

7812

C5

100N

C4

10U

C6

1U0

6 5 3 4

1 2

IC3

4013

8 911

10

13

12

IC3

4013

R2

10K0

R3

10K0

R4

10K0

2X2

3X2

4X2

C7

10U

C8

1N0

C10

1N0

R1

10K0

6X1 5

X2

6X2

7X2

8X2

1X2

2X3

3X3

4X3

5X3

1X3

6 X3

45

31

2

RL1

VSB24

D1

1N4448

R7

0R0

R6

0R0

C9

100N

C11

100N

R8

10K0

R9

10K0

D2

1N4448

abRL2

VSB24

12 RL2

VSB24

1X4

2X4

3X4

1X5

2X5

3X5

4X5

7 X3

8 X3

13 X3

14 X3

2 X1

3 X1

4 X1

5 X1

7 X1

1 X1

C1

100N

C2

100N

C3

100N

D6

ZPD15

D7

ZPD15

3 T5 Q6025

&9 8

10

IC14011

&5 6

4IC1

4011

&1 2

3IC14011 &12

13

11

IC1

4011

R19

20K5

T4

BC337-40

1 2

4 3

IC4

SFH617G-3

D3

1N4448

R10

20K5

R11

20K5

R12

40K2

R14

499R

R13

10K0

R16

10R0

R15

10R0

D5

1N4007

D4

1N4007

R17

1K00

R18

2R4

4X4

F1

C12

10U

C14

47U

C13

47U

T2

BSS91

T3

BSS91

15

10

14

11

12

13

1 2 3 4 5 6 7 9

IC5

4512

R20

10K0

R21

0R0

R22

0R0

Page 70: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-26 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 71: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-27

LP 760Component layout diagram

Page 72: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-28 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 73: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-29

4.4 LP 761 HDF drive keyboard

4.4.1 Description

General

The keys for controlling the substituate lift are found on this P.C.B..

Page 74: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-30 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 75: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-31

4.4.2 Circuit and component layout diagramLP 761 HDF drive keyboard

LP 761Circuit diagram

B

C

D

E

F

1 2 3 4

B

C

D

E

F

AENDERUNGAUS TAG NAM

BEGENO

TAG NAME

ERSATZERSATZ FORMAT

BLATT

A401234567891011121314151617181920

MEDIZIN-TECHN.SYSTEME

LP761 673 766

HDF-ANTRIEB-TASTATUR

UP

DOWN

STOP

HDF-MOTOR-KEYBOARD 1(1)

SP

04.08. OTT1994

05.08. ASCH

a X1 geandert

+12V

UP

STOP

DOWN

SE1

SE2

SE3

SE4

+12V

SUB_A

SUB_K

SUB_C

PWR_GND

1 2

S2RF15

1 2

S1RF15

1 2

S3RF15

1X2

2X2

3X2

4X2

5X2

6X2

7X2

8X2

1 X1

2 X1

3 X1

4 X1

5 X1

6 X1

Page 76: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-32 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 77: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-33

LP 761Component layout diagram

Page 78: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-34 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 79: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-35

4.5 LP 762 scales comm

4.5.1 Description

The LP 762 serves as interface between the dialysis machine and the scales.

The serial transmission from the dialysis machine arrives on the plug X1 pin 4 (RXD) and isconverted at the interface IC 1 (max. 232) to 5 V level. IC 2 makes a serial parallel conversion andpasses the signal on to the scales. The scales are switched on or off, respectively, by thetransistor T1.

The serial carry forward (weight) comes from the scales to the plug X3 pin 6 (TXD) and isconverted by IC 1 from 5 V to RS 232 level. The signal arrives at the dialysis machine throughplug X1 pin 5 (TXD).

SP1 (ML78L05) converts the 12 V into 5 V which serves as the supply voltage for LP 762.

Page 80: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-36 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 81: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-37

4.5.2 Circuit and component layout diagramLP 762 scales comm

LP 762Circuit diagram

Page 82: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-38 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 83: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-39

LP 762Component layout diagram

Page 84: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-40 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 85: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-41

4.6 LP 763 SSE serial interface extender

4.6.1 Description

General

The LP 763 is necessary for serial communication of the monitor with the scales and the HDFpump module.

Circuit structure

The circuit of the LP 763 comprises two circuit units independent of each other.

Serial interfacesHeart of the LP 763 P.C.B. is an 8-fold UART (universal asynchronous receiver/transmitter), i.e.an interface IC (IC 1) with eight independent serial channels. Also to be found on the P.C.B. aretwo TTL/RS232 level converters (IC 2 and IC 3) and an interrupt logic IC (IC 4), as well as diverseresistors, capacitors and an oscillation quartz Q1 for generating the working cycle for IC 1.

COMMCO connectionThe slot 63Z used by the LP 763 P.C.B. is also provided for the installation of P.C.B. LP 752. TheLP 752 takes over the connection of the 4008 monitor signals to the COMMCO P.C.B. LP 729. Inorder to be able to also continue the linking to COMMCO/FINESSE in 4008 HDF machines thecircuit section of the LP 752 P.C.B. is taken over in the LP 763 P.C.B..

Function

After switching on the machine the UART is initialled with the transmission parameters.

Transmission parameter channel A:1200 Baud, 8 databits, 1 stopbit, no paritibit

Transmission parameter channel B:9600 Baud, 8 databits, 1 stopbit, no paritibit

Channel A is responsible for communication with the scales, channel B for communication withthe HDF pump module.

After going through the watchdog test the scales are switched on through the transmission linefrom channel A and the weight requirements of the 4008 monitor are communicated to the scales.The weight value of the scales is read by the 4008 monitor on the reception line.

Commands and data from the 4008 monitor are sent through the transmission line from channelB to the HDF pump modules. Data which comes from the HDF pump module is received by the4008 monitor through the receiving line.

The remaining six communication channels are not used in 4008 operation.

If UART data is ready for collecting or if the transmission buffer is ready to accept new digitsduring a transmission the component causes an interrupt at CPU 1. CPU 1 processes thisinterrupt demand (in so far as nothing more important is to be done) and reads a digit from thereception buffer or writes a digit on the transmission buffer.

Page 86: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-42 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 87: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-43

4.6.2 Circuit and component layout diagramLP 763 SSE serial interface extender

LP 763Circuit diagram 1/2

Page 88: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-44 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 89: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-45

LP 763Circuit diagram 2/2

Page 90: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-46 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 91: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 4-47

LP 763Component layout diagram

Page 92: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

4-48 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 93: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-1

Table of contents5 Spare parts

Section Page

5.0 How to use the spare parts catalog ............................................................................ 5-3

5.1 P.C.B.s ........................................................................................................................... 5-4

5.2 Scales ............................................................................................................................. 5-6

5.3 Substituate lift ............................................................................................................... 5-8

5.4 UF2 pump / blood pump (HDF) / valve V126 ............................................................. 5-10

Page 94: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Page 95: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-3

5.0 How to use the spare parts catalog

Objective:

Spare parts can be defined and ordered.

Organization:

This spare parts catalogue contains 4 assemblies.5.1 P.C.B.5.2 Scales5.3 Substituate Lift5.4 UF2 pump / blood pump (HDF) / valve V126

Each assembly has its assembly number. 5.X.0 refers to the basic version.Major changes (modifications) within an assembly can be identified by an increasing decimal(e.g. 5.X.1, 5.X.2, …)

In the spare part lists it is refered to:Built-inFrom Equipment CodeTo Equipment Code

Update service:

Updates to this spare parts catalog will be released as:– Replacement pages– Supplementary pages– Technical information

Generally we reserve the right to make changes.

Requirements for correct spare part orders:

The equipment code must be entered in the machine record.All modifications must be entered in the machine record. Update the equipment code, ifnecessary, to ensure that the correct spare parts will be ordered.

Page 96: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

5.1 P.C.B.s

I/O

Spülen

Heißreinigen

Desinfektion

SingleNeedle

Test

Vorbereiten

Dialyse Start

AlarmTon Aus

Best. Auswahl

Luftdetektor ArteriellerDruck

VenöserDruck TMP Leitfähigkeit

Blutleck

Überbrücken

Ultrafiltration

Reset

UF Menge

UF Variation

Fluß

Temperatur

Prog.I/O Prog.I/O

Bic.Konz.

SetI/O

mmHgkPa mmHgkPa mmHgkPa

ml

mS/cm(25°C)

ml/min

°C

Set

UF Rate ml/h

UF Ziel ml

UF Restzeit h:min

4008

StartStop

Rate: ml/min(Ø: mm)

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

Pven.

StartStop

Rate: ml/h(Bolus: ml)

Bolus

Rate

( :h.min)

–40

0

–100

–200

–300

100

200

280

0

10

20

30

–10

–20

–30 0

100

200

0

10

20

30

300

400

500

40

50

60

0

100

200

0

10

20

30

300

400

500

40

50

60

13

13.5

14

14.5

15

15.5300500700900

35

37

39

Fresenius Medical Care

Fresenius Medical Care

1

2

3

4

5

6

Page 97: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-5

Built-inFrom/until Equipment Code 000 – 000

Item Part number Description

1 673 790 1 P.C.B. LP 762 scales comm2 674 182 1 P.C.B. LP 763 SEE serial interface extender3 673 766 1 P.C.B. LP 761 HDF drive keyboard4 673 872 1 P.C.B. LP 625 display board5 673 720 1 P.C.B. LP 760 HDF motor control6 673 041 1 P.C.B. LP 754 control board HDF

Page 98: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

5.2 Scales

1 2 3 4 5 6 7 8

19 161718 15 14 13 12 10 9

11

20 21 2322 24

30 29 28 27 25

26

Page 99: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-7

Built-inFrom/until Equipment Code 000 – 000

Item Part number Description

11 673 983 1 Sliding bush12 552 244 1 Screw13 646 352 1 Screw14 674 409 1 Microswitch15 552 312 1 Screw16 553 330 1 Screw17 673 635 1 System unit, scales18 643 161 1 Screw19 673 988 1 Hook10 673 987 1 Latch11 642 861 1 Screw12 553 621 1 Nut13 673 986 1 Adapter14 563 151 1 Retaining washer15 674 106 1 Line holder16 552 275 1 Screw17 673 971 1 Fastening plate18 642 693 1 Screw19 673 985 1 Bearing20 672 530 1 Cover21 552 276 1 Screw22 674 551 1 Caution label: Stability limit23 673 977 1 Display plate24 673 922 1 Back shell25 673 921 1 Front shell26 553 620 1 Nut27 559 690 1 Washer28 642 648 1 Spacer sleeve29 640 143 1 Spacer30 673 923 1 Front foil display

674 068 1 Scales, complete

Page 100: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-8 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

5.3 Substituate lift

1

2

3

4

5

6

7

8

9

11

14

16

17

6

28

27

25

26

6

24

22

21

20

19

18

35

9

34

33

32

313029

23

10

37

36

38

12

39

40

41

Page 101: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-9

Built-inFrom/until Equipment Code 000 – 000

Item Part number Description

11 674 399 1 Connection cable12 644 205 1 Screw13 673 990 1 Locking ring14 674 106 1 Line holder15 653 099 1 Sensor16 552 244 1 Screw17 674 334 1 Pipe18 673 975 1 Clamping plate19 552 275 1 Screw10 553 733 1 Toothed washer11 674 395 1 Drive12 673 973 1 Dog14 673 755 1 Screw16 553 621 1 Nut17 537 203 1 Fastening clip18 646 274 1 Tolerance sleeve19 674 078 1 Stripper20 673 974 1 Collar bush21 673 920 1 Housing, lift22 673 639 1 Front foil, lift23 559 690 1 Washer24 674 400 1 Connection cable25 674 358 1 Connection cable26 673 918 1 Base plate27 673 972 1 Funnel28 674 255 1 P.C.B. cover29 646 606 1 Spacer bolt (M 3 x 8.5)30 650 439 1 Spacer bolt (M 3 x 12.5)31 559 690 1 Washer32 574 617 1 Spacer sleeve33 644 040 1 Spacer bolt (M 3 x 13)34 673 982 1 Bush35 674 487 1 Sealing tape36 675 052 1 Split pin37 553 694 1 Washer38 674 333 1 Locating pin39 552 312 1 Screw M5x1640 675 222 1 Fork41 675 051 1 Bolt with head

673 915 1 Substituate lift, complete

Page 102: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

5.4 UF2 pump / blood pump (HDF) / valve V126

StartStop

Rate: ml/min(Ø: mm)

HDFReset

Vol : l

40

Hydraulic section 4008 / 4008 E Hydraulic section 4008 B

Concentratepump(23)

UF pump(22)

UF2 pump(20) UF2 pump

(20)Concentratepump(23)

Bibcarbonatepump(25)

1

2

3

4

5

6

Page 103: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

Fresenius Medical Care 4008 HDF 3/11.97 (TM) 5-11

Built-inFrom/until Equipment Code 000 – 000

Item Part number Description

UF2 pump

The UF2 pump is identical with the ultrafiltration pump(see appropriate Spare Parts Catalogue 4008)

670 706 1 Ultrafiltration pump, complete

Blood pump (HDF)

The blood pump (HDF) is similar to the blood pump (single needle)(see appropriate Spare Parts Catalogue 4008)

40 671 726 1 Front foil

673 924 1 Blood pump module HDF, complete

Valve V126

1 645 881 1 Filter2 545 361 1 Silicon tube (0.12 m)3 674 349 1 Solenoid valve MV 126 / solenoid valve MV434 674 347 1 Valve block5 552 275 1 Screw6 579 072 1 O-ring

674 941 1 Valve block, complete

Page 104: TM 4008 HDF - interest€¦ · 1.3 Error messages ... X632/ B24 X632/ C27 X634 R/ 1 serial interface and taring line serial interface X634 R/ A24 ... 75 External flow indicator

5-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)