102
Theoretical Evaluation of Moderately Focused Spherical Transd-ucers and Mult i-focus Acoust ic Lens/Transducer Combinations for High Intensity Focused Ultrasound Thermal Therapy Xia Wu A thesis submitted in conformity with the requirements for the degree of Master of Science Graduate department of Medical Biophysics The University of Toronto @ Copyright Xia Wu 2001

Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Theoretical Evaluation of Moderately Focused Spherical Transd-ucers and Mult i-focus Acoust ic

Lens/Transducer Combinations for High Intensity Focused Ultrasound Thermal Therapy

Xia Wu

A thesis submitted in conformity with the requirements

for the degree of Master of Science

Graduate department of Medical Biophysics

The University of Toronto

@ Copyright Xia Wu 2001

Page 2: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

A uisitions and Acquisitions et ~ 8 b g m p h i c Seivices senrices Wbîtographiquro

The author has granted a non- exclusive licence allowing the National Library of Canada to reproduce, loan, disiribute or seU copies of this thesis in microform, paper or electronic formats.

The author retains ownership of the copyright in this thesis. Neither the thesis nor substantid extracts fkom it may be printed or othewise reproduced without the author's permission.

L'auteur a accordé une licence non exclusive permettant à la Biblioth6que nationale du Canada de reproduire, prêter, distribuer ou vendre des copies de cette thèse sous la forme de microfiche/fdm, de reproduction sur papier ou sur format 6lectronique.

L'auteur conserve la propriété du droit d'auteur qui protège cette thèse. Ni la thèse ni des extraits substantiels de celle-ci ne doivent être imprimés ou autrement reproduits sans son autorisation.

Page 3: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,
Page 4: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Theoretical Evaluation of Moderately Focused

Spherical Transducers and Multi-focus Acoustic

Lens/Transducer Combinations for High Intensity

Focused Ultrasound Thermal Therapy

Xia Wu, B.A.Sc. Department of Medical Biophysics

The University of Toronto, 2001

Impractically long treatment tinies are required for highly focased spherical trans-

ducers [HF] to destroy tumours because thermal lesions generated by tliese trans-

ducers are srnall and a large number of such lesions are required. Moderately focused

spherical transducers [MF] and multi-focus acoustic lens/transducer combinations

[LTC] can generate larger lesions compared to those produced by highly focused

spherical t ransducers, and t herefore shorter treatment times are expected. The

degree of improvement in total treatment time by the use of MFs and LTCs was

quantified in this study. A 3-D ultrasound thermal mode1 and a target mode1 were

developed to calculate treatment times required for various ultrasound transducers,

under identical treatment conditions. A LTC design method was developed to de-

termine the thickness of lens elements for production of specified multi-focus fields.

The simulation results show for the treatment of a 2 x 2 x 2cm3 tumour, a HF, MF and LTC require 150, 42 and 30 min respectively.

Page 5: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Acknowledgment s

1 would like to thank the following people My supervisor, Dr. Michael Sherar, for his scientific inspiration, encourage-

ment and continuous support during this project, My cornmittee members, Dr. John Hunt and Dr. Brian Wilson, for their

advice and helpful discussions, Dr. Michael Kolios, for his scientific insight and experience, Dr. Mark Gertner for his invaluable guidance on writing and his great pa-

t ience, Car1 Kumaradas for his advice and help in science, computers and many other

things Mihaela Pop, Lee Chin, Peter Bevan, Arthur Worthington and Sean Davidson

for sharing their knowleclge and for their assistance. Mum and Dad for their unending support and al1 their sacrifices, Jack, for his loving encouragement.

iii

Page 6: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Contents

Abstract

Acknowledgment s

List of Figures

Nomenclature

iii

viii

Chapter 1 Introduction 1

. . . . . . . . . . . . . . . . . 1.1 Thermal Therapy in Cancer Tkeatment 1

1.1.1 Biological Rationale and Clinical Experience . . . . . . . . . . 1

. . . . . . . . . . . . . . . . . . . . . . . . . 1.1.2 Thermal Delivery 4

1.2 High Intensity Focused Ultrasound Thermal Therapj . . . . . . . . . 7 . . . . . . . . . . . . . . . . . . . . . . . . 1.2.1 Clinical Experience 7

1.2.2 Focused UI trasound Transducer Designs . . . . . . . . . . . . 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3 Objective 17

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 ThesisOutiine 17

Chapter 2 Theoretical Evaluation of Moderately Focused Spherical lkansducers for High Intensity Focused Ultrasound Thermal Ther-

aPY 19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 Abstract 19

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3 Introduction 20

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3 Methods 23 . . . . . . . . . . . . . . . . . . . 2.3.1 Ultrasound-Thermal Mode1 23

. . . . . . . . . . . . . . . . . . . . . . . . . . 2.3.2 Tumour Mode1 25

Page 7: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4 Results 28

. . . . . . 2.4.1 Verification of Ultrasound-Thermal Mode1 Accuracy 28

2.4.2 Comparison of Highly and Moderately Focused Spherical Trans- . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ducers 32

2.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6 Conclusions 41

Chapter 3 Theoretical Evaluation of Multi-focus Acoustic Lens/Tkansducer Combinations for High Intensity Focused Ultrasound Thermal Ther-

aPY 43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1 Abstract 43

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 Introduction 44 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Methods 47

. . . . . . 3.3.1 LTC Ultrasound Intensity Distribution Calcuiations 47 . . . . . . . . . . . . . . . . . . . . . . . 3.3.2 LTC Design Method 48

. . . . . . . . . . . 3.3.3 Methods for Determining aeatment Times 52 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4 Results 52

. . . . . . . . . . . . . 3.4.1 Comparison of LTCs and phased arrays 53 . . . . . . . . . . . . . . . . . . . . . . 3.4.2 Effect of Focus Spacing 58

. . . . . . . . . . . . . . . . . . . . . 3.4.3 Effect of Number of Foci 61

. . . . . . . . . 3.4.4 Multiple Exposure Tkeatments of the Tumour 64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5 Discussion 65

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6 Conclusions 69

Chapter 4 Summary and F'uture Work 71 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 Summary 71

. . . . . . . . . . . 4.1.1 Moderately Focused Spherical lkansducers 71

. . . . . 4.1.2 Multi-focus Acoustic Lens/Transducer Combinations 72 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2 Future Work 73

. . . . . . . . . . . . . . . . . . 4.2.1 Ultrasound-Thermal Modeling 73 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.2 LTCs 77

References 81

Page 8: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

List of Figures

Surviving fraction as a function of heating time . . . . . . . . . . . . 2

Schemat ic representation of using high intensity focused ultrasound

to treat tissue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Schematic representation of the therapeutic ultrasound applicator . . . 9

. . . . . . . . . . . . . . . . . . Focused splierical transducer designs 10

. . . . . . . . . . . . . . Schernatic of the reflector focusing applicator 13

. . . . . . . . . . . . . . . . . . . . Schematic of the 2-D acoustic lens 16

During the transducer movement from left to right to form lesions side

by side. some normal tissues in the nearfield remain in the pathway

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . of the beam 21

Block diagram of numerical model to simulate ultrasound-thermal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . effects in tissue 2 3

. . . . . . . . . . . . . . . . . nimour geometry employed in this work 26

Thermal computation domain and mesh spacing dx. dy and dz . . . . 26

. . . . . . . . . . . . . Thermal Dose threshotds of the tumour mode1 28

Comparison of measiired lesion diameters with the results predicted

. . . . . . . . . . . . . . . . . . . . . . . . . by our theoretical mode1 31

Thermal dose distributions calculated using our theoretical model . . 31

Lesion lengt h and diameter and transducer 6dB beamlengt h and

beamwidth versus radius of curvature of focused spherical transducers . 33

Ultrasound intensity distributions and thermal dose distributions pro-

. . . . . . . . . . . . . . . . . . . . . . . . . . duced by SPI and SP2 34

. . . . . . . . . . . . . . . . . . . . . . . . . 2.10 Transducer step patterns 35

. . . . . . . 2.1 1 Temperatures reached in the tissue as a function of time 37

Page 9: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.12 Thermal dose profiles for treatments of the tumour using SPI and SP2 38

2.13 Thermal dose distributions for treatments of the tumour using SPI

Coordinates and parameten used in the calculation of LTC field dis-

tributions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Coordinates and parameters used in the pseudoinverse method . . . . 49

Geometry of the turnour mode1 in Cartesian coordinates . . . . . . . 53

9 focus pattern with 2 mm focus spacing . . . . . . . . . . . . . . . . 54

The intensity distributions generated by LTCl and PA . . . . . . . . 55 The lens surface profile of LTCl . . . . . . . . . . . . . . . . . . . . . 56

Thermal dose distributions generated by LTCl and PA . . . . . . . . 57 The iiltrasound intensity distributions generated by LTCl and LTC2 59

Thermal dose distributions generated by LTCl and LTC2 . . . . . . . 60

2 mm spaced. 16-focus pattern . . . . . . . . . . . . . . . . . . . . . . 61

Ultrasound intensity distributions generated by LTCl and LTC3 . . . 62 3.12 Thermal dose distributions generated by LTCl and LTC3 . . . . . . . 63

3.13 Lateral step pattern for LTCl . . . . . . . . . . . . . . . . . . . . . . . 64

3.14 Thermal dose distributions generated by LTC1. SP2 and SPI at the

end of the treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

3.15 Thermal dose isosurface distributions generated by LTC1. SP2 and

SPI at the end of the treatrnent . . . . . . . . . . . . . . . . . . . . . 67

. . . . . . . . . 4.1 Diagam of a beam traversing t hrough layered tissues 75

4.2 Schematic of the rotation of a 9 focus LTC with 2 mm focus spacing . 79

4.3 Thermal dose distributions generated by LTCl with and lwithout the

pseudo-rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

Page 10: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Nomenclature

S ymbols

a, Distance between the surface of the transducer and the lens

element n [cm]

6, Thickness of the lens element n [cm]

c Speed of sound [cm S-'1

cb Heat capaci ty of blood [J g-' 0" C-'1

ci,. Speed of sound in the lens [cm s-'1

q Heat capacity cf tissue [J g-' - O C-'1

c,.t,, Speed of sound in water [cm S-'1 p p p p p p p p - - - - - - - -

d Distance [cm]

d, Distance between the lens element n and the field point (x,y,z) [cm]

dS Surface area of the source element [cm2]

D 'Ikansducer diameter [cm]

f F'requency [MHz]

G Intensity gain

H Matrix describing the propagation of the ultrasound wave

Page 11: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

I Intensity [W cm-*]

jd=-l

6 Thermal conductivity of tissue [W **cm-' 9 " C-'1

k, Wave number [cm-']

M Number of field points intended to be foci

N Total number of lens elements

P Pressure [Pa]

P Complex field pressure [Pa]

R Radius of curvature [cm]

s Axial distance between the skin surface and the field [cm]

S Surface area of a lens element [cm2]

SAR Specific Absorption Rate [W

t Time [s]

Total lleatment Time [s]

T Temperature [OC]

Ta Temperature of arterial blood ['Cl

TD Thermal Dose [EMd3 ]

TDd3,, Thermal dose where attenuation coefficient starts to change [E& ]

TD43i Thermal dose where attenuation coefficient reaches a plateau [EMd3 ]

un Amplitude of the complex particle velocity of the lens element n [cm a * s - ' 1

U Particle velocity [cm *-s-']

Û Complex particle velocity [cm *es-1]

Page 12: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

wb Volumetric perfusion rate [g s-l cm-3]

wm Initial blood perfusion [g s-' cm-3]

a Amplitude absorption coefficient [Np cm-' MHZ-'1

p Amplitude attenuation coefficient [Np cm-' MHZ-'1

po Initial amplitude attenuation coefficient [Np cm-' MHz-']

pl Plateau amplitude a t tenuation coefficient [ J g-l OC-']

p Density of the medium [g cm-3]

pt Density of tissue [g cm-3]

0, Phase of the complex particle velocity of the Iens eleinent n

Abbreviat ions

BHTE Bioheat Dansfer Equation

EMd3 equivalent minutes a t 43 OC

HIFU High Intensity Focused Ultrasound

LTC Lens/Transducer Combination

LTCl LTC designed to produce a 2mm spaced, 9 focus field

LTC2 LTC designed to produce a 2.5rnm spaced, 9 focus field

LTC3 LTC designed to produce a 2mm spaced, 16 focus field

MNS Minimum Norm Solution

MW Microwave

PA Phased array designed to produce a 2mm spaced, 9 focus field

RF Radiofrequency

Page 13: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

SPI Highly focused spherical transducer

SP2 Moderately focused spherical transducer

Page 14: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Chapter 1

Introduction

1.1 Thermal Therapy in Cancer Treatment

Thermal therapy refers to the clinical use of heat. Hippocrates recorded the use of

red-hot irons to treat small, non-ulcerating cancers in 400 B.C. Modern interest in

thermal therapy began with Coley [9], who used bacterial pyrogens to deliberately

induce fevers in cancer patients. In 1906, Clowes [8] demonstrated that murine carci-

noma cells heated to 45OC failed to produce tumours on subsequent inoculation into

rnice. In 1921, it was shown that diathermy (hyperthermia) and ionizing radiation

had a synergistic anti-cancer effect [58]. Although pursued with great enthusiasm,

much of the early clinical work provided only anecdotal evidence that elevated tem-

peratures have any clinical benefits. Over the past 30 years, a better understanding

of the behaviour of tissues a t elevated temperatures has developed. Effort has also

been focused on development of heating technology, to improve heat localisation

and energy penetration in tissues.

1.1.1 Biological Rationale and Clinical Experience

Heat damages cells directly, causing membrane disruption and cellular protein de-

naturation. Survival curves of chinese hamster ovary cells, as shown in figure 1.1,

demonstrate that the percentage of cells killed is an exponential function of heat-

ing time for temperatures above 43°C. This relationship is valid for most cells

in vitroand tissues in vivo (3, 16,281. Based on this relationship, Sapareto and

Page 15: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

THERMAL

Figure 1.1: Surviving fraction as a function of heating time for Chinese hamster ovary cells at various temperatures. The figure was modified from [29].

Dewey [60] developed the concept of thermal dose, a formalism to quantify ther-

mal treatments. The thermal dose, descri bed by equation 1.1, converts different

heating protocols incorporating various heating times at different temperatures into

equivalent heating durations a t a reference temperature of 43°C.

where

T = temperature [OC]

TD = thermal dose [EMe (equivalent minutes a t 43"C)I

ttotai = total treatment time [s]

The thermal dose equation indicates that for each I0C increase above 43°C the

heating time can be halved to achieve the same biological effect.

Based on the thermal dose equation, two general treatment strategies can

be adopted to achieve the same biological endpoint. The first strategy, known as

hyperthermia, is to expose tumour tissues to mild temperatures (43OC-45OC) for long

Page 16: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.1. THERMAL THERAPYINCANCER TREATMENT

periods of time (from 30 minutes to several hours). This is usually combined with

radiotherapy or chemotherapy as an adjuvant. In the second strategy, known as high

temperature thermal therapy, the minimum target temperature is approximately

50°C such that tissue coagulation is achieved in seconds to a few minutes. This can

be given as a stand-alone treatment.

Hypert her mia

No ciinical studies have been able to validate the efficacy of hyperthermia as an

independent therapeutic agent. Difficulties in raising the temperature of the entire

tumour volume to the target range using current heating applicator technology have

been reported 1721. Studies on combining hyperthermia with other treatment modalities including

radiotherapy and chemotherapy have produced more favourable results. In vivo and

in vitro studies have demonstrated that hyperthermia is especially effective against

hypoxic cells, which are radioresistant, in the centre of tumours, while radiation

eliminates the cells in the peripheral, well-vascularized regions of the tumours which

are difficult to heat but sensitive to radiation (541. A study of 70 patients with

recurrent melanomas, performed by the European Society of Hyper theniu Oncology,

reported complete response of 62% for patients treated with thermoradiotherapy,

compared to 35% of patients treated by radiotherapy alone 1551. A collaborative

study between Dutch Hyperthennia Group, the Medical Research Council (England),

the Evropean Society ojHyperthennHa Oncology and the Princess Margaret Hospital

(Canada) of patients with advanced primary or recurrent breast cancer reported

complete response in 59% of patients treated with thermoradiot herapy, compared

to 41% for patients treated by radiotherapy alone 1761. In addition, the cytotoxicity

of many chemotherapy drugs are enhanced by moderately elevated temperatures

[27,51].

High Temperature Thermal Therapy

Poor temperature distributions achieved in tumours dunng hypert hermia treatments

are caused by inadequate heat ing technologies, tissue inhomogenei ties, and heat

transfer due to conduction and blood flow [5,37]. Consequently, several groups

Page 17: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1 . l , THERMAL THERAPY IN CANCER TREATMENT

suggested the use of high temperature short exposure thermal treatmeuts. In such

treatments, the high temperatures (> 50°C) attained would rapidly coagulate tu- mour tissues, allowing the tumour response rate to be less dependent on the local

tissue physiology such as blood flow and tissue thermal conduction [2,30]. In the

past 10 years, interest in high temperature thermal therapy has been propelled by

advances in medical imaging that have allowed the near real-t ime monitoring of ther-

mal treatments. Temperature distribut ions and coagulated regions of heated tissues

have been measured non-invasively using Magnetic Resonance Imaging [33,50,64,78]

and ultrasound imaging [l, 621.

Clinical trials to evaluate high temperature thermal therapy are ongoing.

Phase 1/11 trials have been aimed at establishing treatment feasibility and safety

[59,77]. Because of relatively mal1 tissue volumes that can be heated to high

temperatures effectively by currently heating techniques, target sites have been focal

cancers such as liver metastases 167,771, prostate cancer [74,77] and renal tumours

[77]. A considerable portion of work in this field is now focusing on the development

of heating applicators that can treat larger, deep-seated tumour volumes.

1.1.2 Thermal Delivery

There are many ways to heat a tumoiir and the choice of a particular delivery

modality depends on several factors, including the location, shape and size of the

tumour and the proximity of critical normal tissues. Temperature rise in tissue is

induced by the deposition of energy, which can be delivered to the region of the

tumour by using a laser, an ultrasound transducer, a radiofrequency or microwave

applicat or.

Thermal therapy applicators are generally classified as i) external, (applied

from outside the body), ii) intracavitary, (applied from within a body cavity such as the rectum) and iii) interstitial, (embedded directly into the tumour tissue). The

goal of thermal therapy treatments is to coagulate the entire tumour volume while

sparing overlying and surrounding normal tissues. Therefore, the ability of an appli-

cator to deliver energy that will be localised in the target volume and, if necessary,

penetrate deeply into the body is of primary importance. Intracavitary and inter-

stitial applicators are effective in achieving these goals. However, treatment sites

Page 18: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.1. THERMAL THERAPY IN CANCER TREATMENT 5

accessible to intracavitary applicators are limited. Intentitial applicaton are in-

vasive and are not well suited to heating large irregular tissue volumes. External

heating devices are non-invasive, and have the flexibility to access many more treat-

ment sites than is possible with intracavitary applicators. The three most common

energy sources used in externd heating devices are radiofrequency, microwave and

ultrasound.

Radiofkequency Heating

Radiofrequency [RF] heating involves the use of electromagnetic waves in the range

of 0.1-27MHz. In capacitive RF heating, the tissue volume to be heated is sand-

wiched between two electrodes, forming a circuit element similar to a parallel plate

capacitor. The energy dissipated in the tissue volume is determined by the current

between the two electrodes and the tissue resistivity. Capacitive RF heating is the

only form of thermal therapy in which the energy absorption does not decreiise expo-

nentially with increasing deptb in tissue. However, maximum energy deposition for

capacitive RF heating occurs in tissues with high electrical resistivity, such as fat,

where excess heating can limit power deposition. Inductive RF heating is produced

by using a coil which is either placed near the body of the patient or completely

surrounds the body. The electric field induced in the body by the magnetic flux

from the coil produces an electric current within the tissues, leading to tissue heat-

ing. Maximum energy deposition for inductive heating occurs in tissues with low

resistivity, such as muscle.

The drawback of RF heating is that, because of the long electromagnetic

wavelength in tissue, it is very difficult to localise heating at depth to a specific

target volume. Selective heating of tumour tissue can only rely on the fact that

tumours often have reduced blood flow compared to normal tissues [65].

Microwave Heating

Microwave [MW] applicators produce electromagaetic waves which carry energy

into tissue. Due to international agreements, only the frequencies of 433, 915 and

2450 MHz can be used in thermal therapy treatments in electrically un-shielded facil-

ities. Energy delivered by microwaves is attenuated exponentially in tissue. Hence,

Page 19: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.1. THERMAL THERAPY IN CANCER TREATMENT 6

increasing heating at depth has to be achieved through surface cooling. Because

of the large values of MW attenuation coefficient at high frequencies, the energy

penetration in tissue is poor. For example, the penetration depthl of a plane MW field is 1.2 cm in muscle tissue a t 915 MHz. Considerable irnprovement in energy

penetration occurs for sources with frequencies lower than 500 MHz. However, be-

cause of the longer wavelengths for frequencies of less than 500 MHz, constraint of

the waves into an applicator of practical size is difficult.

Single MW waveguide radiators have been developed for use in surface heat-

ing. Their power deposition patterns are often uneven and cannot be corrected

for variations in tissue cooling. Recently, special boluses have been developed [63]

employing different concentrations of saline solution, which are capable of both cre-

ating a more uniform power deposition during the treatment and increasing the

penetration depth.

Ultrasound Heating

Ultrasound devices produce mechanical waves in tissue which cause heating. Ultra-

sound applicators for therapeutic use operate in a frequency range from 0.5 MHz

to 5MHz. The energy transported by an ultrasound wave is attenuated exponen-

tially as it propagates through tissue. The greatest advantage of ultrasound heat-

ing over MW and RF heating is that ultrasound beams can be focused, because

the ultrasound wavelengths (0.3-3 mm) a t frequencies in the therapeutic range are

much shorter than the diameter of the applicator. Focused ultrasound waves can be

generated by either a geornetrically focused applicator or an electronically phased

multi-element applicator. Focused ultrasound in particular is well suited to the lo-

calised heating of deep-seated tumours. However, typical single-focus ultrasound

applicators produce a beam only a few millimetres wide at the focus, and therefore

can only heat small volumes.

Because of large changes in acoustic impedance, ultrasound is strongly re-

flected a t tissue-bone and tissue-gas interfaces. These reflections can cause intense

heating at tissues immediately ahead of these interfaces and prevent heating be-

'Penetration depth refers to the depth at which the intensity reduces to 50% of the surface value.

Page 20: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERIMAL THERAPY 7

yond the interfaces. Hence, external ultrasound heating is only useful for regions

with "acoustic windows", i. e. regions in the body where ultrasound is not blocked

by bone or gas.

1.2 High Intensity Focused Ultrasound Thermal

Therapy

The schernatic in figure 1.2 illustrates the principle of external High Intensity Fo-

cused Ultrasound [HIFU] thermal therapy. Water is used between the transducer

1 Waier Tank 1

Figure 1.2: Schematic representation of using high intensity focused ultrasound to treat tissue.

and the body for efficient coupling of ultrasound into the body. Ultrasound pen-

etrates intervening tissues focusing on the target. Thus, sufficient heat can be

generated in the focal zone in a short period of time, while the intervening tissues

are spared.

1.2.1 Clinical Experience

HIFU heating for cancer treatment was first proposed by Burvo [6], who had in-

vestigated the effect of HIFU on Brown-Pierce tumours in rabbits and reported a

cure rate of 60%-80%. Since the 19606, many studies have been devoted to HIFU thermal therapy. Complete tumour destruction was noted in the range in 744%

of animal tumours treated with HIFU [82]. The lack of tumour response in some

animals may be attributed to the HIFU beam missing a portion of the tumour be-

Page 21: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1 .P. HIGH INTENSITY FOCUSED ULTRASOUAD THERMAL THERAPY 8

cause of animal movement, under-assessrnent of tumour size, insufficient coverage

of surroundiiig tissue or blood flow cooling.

Clinical trials of HIFU thermal therapy have produced encouraging results.

Katsumi [36] and Tsuchidate [69] reported a clinical study of 35 patients. Different

types of advanced tumours were treated. Malignant tissue was destroyed, tumours

decreased in size and there was no increase in metastasis. A phase 1 study conducted

by Vallancien et al. [73] reported that 11 patients with superficial bladder tumours of

4-20 mm diameter were treated. No specific side-effects were encountered, although

al1 patients developed transient haematuria Iasting 1-2 days. In an update on this

series, 50% of (87) patients treated by HIFU remained tumour-free, similar to that of

patients treated with conventional surgery. A more recent phase 1 trial, conducted

by ter Haar and CO-worken (771, demonstrated that HIFU treatment of tumours

of the liver, kidriey and prostate could be performed in fully conscious patients.

Regions of tumours situated up to 12cm below the skin surface were heated, while

al1 normal tissue lying in the beam path was spared.

No phase II or phase III trials have been conducted, probably because of

inadequate HIFU applicator technology for treatment of the entire tumour volume.

In the phase 1 study conducted by Vallancian et al. [73], it was found that a mean of

395 heating exposures was required to treat a target volume with a mean diarneter

of 1.4 cm, resulting in relatively long treatment times of 45 min to 68 min. The

phase 1 trial conducted by Visioli et al. [77] also reported similar problems. The

treatment of a 1.3cm3 turnour volume lasted for approximately 35min. Because

of the impractically long treatrnent time, complete tumour coagulation was not

performed.

HiFU, as a non-invasive technique, is suited to many oncological and uro-

logical applications. Prostatic carcinoma, bladder carcinoma, renal carcinoma, Iiver

metastasis and benign prostatic hyperplasia are suitable targets due to good trans-

abdominal access. Several of the above clinical studies, however, reported long

treatment times for the entire tumour volume using current HIFU applicator tech-

nology. This establishes the need for investigation into HIFU applicator designs to

reduce treatment times.

Page 22: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUMI THERMAL THERAPY 9

1.2.2 Focused Ultrasound Transducer Designs

Ultrasound transducers employ piezoelectric materials, whose dimensions change

in the presence of an electric field. When an oscillating electric voltage is applied

across a piezoelectric transducer, the material expands and contracts. Electric en-

ergy is converted into mechanical energy which rnanifests itself as a traveling ultra-

sound wave. To ensure al1 mechanical energy is emitted via the front surface of the

transducer, the acoustical impedance mismatching between the transducer and the

backing should be maximized. Air backing has proven to be almost ideal for the

high intensity therapy applications. Figure 1.3 illustrates a schematic of a thera-

peutic ultrasound applicator. Early therapy transducers employed cross-cut quartz

Figure 1.3: Schematic representation of the therapeutic ultrasound applicator.

crystals. Most therapy transducers a t present adopt piezoelectric ceramics such as PZT4 made of lead zirconate titanate. The advantages of ceramics over quartz are

the low cost and the possibility to manufacture transducers in many shapes. How- ever, mechanical and electrical losses are greater in ceramics. Also, the properties of

ceramics change with temperature and age. In contrast, quartz crystals are stable

at high power output and during prolonged use.

Focused ultrasound applicators have been constructed using many different

designs, the most common which is the focused spherical transducer. More complex

applicator designs include phased arrays and acoustic lens/transducer combinations.

Page 23: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERMAL TIIERAPY 10

Highly Focused Spherical lkansducer Design

The cornmonest way to produce a focused ultrasound field is to form a transducer

into a spherically-shaped radiator or to couple a spherically-shaped acoustic lens

with a planar transducer, as shown in figure 1.4. The major advantages of these

Figure 1.4: (a) Focused spherical transducer. (b) Planoconcave acoustic lens combined with a planar transducer. Because the speed of sound in most acoustic lens materials is greater than the speed of sound in water, focusing acoustic lenses are concave in shape.

two configurations, both referred to here as spherical transducer designs, are t heir

simplicity and low cost. However, spherical transducers can only produce single

focus ultrasound fields. The focal zone dimension of a spherical transducer is deter-

mined by physical parameters of the transducer [31], including the diameter, radius

of curvature and operat ing frequency (equation 1.2).

R c GdB beamwidth î! 1.41-9

of where

6dB beamwidth = full width at 25% of intensity maximum in the focal plane

of the transducer [cm]

R = radius of curvature [cm]

D = diameter or aperture [cm]

c = speed of sound [cm d] f = operating frequency [MHz]

Page 24: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERMAL THERAPY 11

The focused spherical ultrasound transducers employed in thermal therapy

usually have an aperture being larger than 5cm so that sufficient acoustic power

deposition as well as sufficient focusing can be achieved a t depths in the body. If

such a focused spherical transducer has a 6dB beamwidth of less than 2mm or

a 6dB beamlength2 of less than 10 mm, it is categorized here as a highly focused

transducer. Because of their small focal zones, highly focused spherical transducers

have been widely used in neurosurgical and opthalmological HIFU treatments, where

the treatment goal is to destroy small, precisely located tissue volumes. A study of

the use of HIFU in neurosurgery conducted by Lele 1461 demonstrated that a highly

focused spherical t ransducer successfull y generated "t rackless" thermal lesions at

preselected regions in cat brain. The diameter and length of these thermal lesions

were only 0.1 mm and 1 mm respectively.

When highly focused spherical transducers were applied to thermal treat-

ments of large target volumes, these transducers proved to be problematic. A study

of HIFU thermal treatment of rat liver tissue reported that 1000 exposures were re-

quired for a highly focused transducer to destroy a tissue volume of 1 x 1 x l cm3 [81].

Each of these exposures lasted for 4s followed by a 10s off-time to cool the inter-

vening tissue, resulting in a total treatment time of 4 hours. A theoretical study

conducted by Fan and Hynynen 1241 reported the same problem (10 houn for a

treatment of a 3 x 3 x 3 cm3 tissue volume).

The size of thermal lesions can be controlled by varying the exposure duration

or intensity in a certain range. Increasing exposure duration allows lesion broadening

through thermal conduction. However, these exposures should still be sufficiently

brief to avoid blood flow cooling effects, which can become significant after tens

of seconds. Since perfusion in tissue is difficult to measure accurately and can

be altered by temperature increase during heating, blood flow cooling introduces

uncertainties into the dimension of thermal lesions [39].

Increasing the ultrasound exposure intensity can also lead to enlargement of

thermal lesions. However, exposure intensity should be lower than a certain thresh-

old to avoid non-linear ultrasound beam propagation, which can lead to uncontrolled

mechanical tissue destruction (321. Moreover, in order to precisely control the for-

*6dB beamlength refers to the fuil length (dong the applicator axis) at 25% of intensity maxi- mum of the transducer focal zone.

Page 25: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. H M INTENSITY FOCUSED ULTRASOUND THERMAL THERAPY 12

mation of thermal lesions, tissue temperatures close to lOO0C should be avoided

because cavitation and the formation of vapor in tissue distort the deposition pat-

tern of ultrasound energy and modify tissue acoustic properties, making the control

of Iesion format ion difficul t (321.

Moderately Focused Spherical Transducer Design

Compared to highly focused spherical transducers, moderately focused spherical

transducers have a larger f-number or a lower operating frequency. Due to the weak

focusing, energy is less localised in the field of moderately focused transducers. If a

spherical transducer has a beamwidth geater than 2 mm and a beamlength geater

than 10 mm, it is categorized here as a moderately focused transducer.

Among al1 the reported HIFU heating systems, only the one developed by ter

Haar and CO-workers employs a moderately focused spherical transducer [77]. In an

investigation into the feasibility of HIFU treatments of kidney in pig, this moder-

ately focused spherical transducer was used to generate thermal Iesions of 2.5 mm

in diameter and 17mm in length. Applying this heating system to a phase 1 clin-

ical study, Visioli et al. [77] reported encouraging results in that the treatment of

a 1.3 cm3 volume of tumour tissue in kidney required approximately 30-35 thermal

exposures, each of which lasted for 1 s followed by a 1 min cooling period. The total

treatment time was therefore approximately 30-35 min. This result appears to be a

significant improvement on the treatment time required with highly focused spher-

ical transducers. However, the conclusion should be drawn with caution as many

treatment parameters, including the shape and location of the target volume and

the criterion for determining cooling periods are different between these treatments

and those using highly focused spherical transducers. To allow a useful cornparison

and to quantitatively evaluate the improvement of the use of moderately focused

spherical transducers, a study is needed to estimate treatment times required for

highly and moderately focused spherical transducers to treat an identical target un-

der identical treatment conditions. This question forms the fint part of this thesis.

Page 26: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSiTY FOCUSED ULTRASOUND THERMAL THERAPY 13

Other Single-focus lkansducer Design

In addition to the focused spberical transducer designs, other transducer designs

can produce single focus ultrasound fields. Fry [25] described the use of four plane

transducers, each with a planoconcave spherical lens. The four focused beams were

brought to a coincident focus and were individually phased to maximize the intensity

output at the focal spot. This niulti-transducer single focus design was also adopted

in the HIFU heating system build by Vallancian et al. [73]. Because of the small focal

zones, these two heating systems can be categorized as highly focused transducer

designs. Fky [25] applied their HIFU heating system to the treatment of Parkinson's

disease. Following craniotomy, which was necessary due to the problem of ultrasound

transmission through the skull, the target tissues were exposed to HIFU. Syrnptorns

of Parkinsonism were claimed to be eliminated. Vallancian et al. [73] applied their

heating system to the treatment of benign prostatic hypertrophy and superficial

bladder tumours, the results of which were discussed in section 1.2.1.

Fry [25] also reported the use of a reflector transducer design, as shown in

figure 1.5, in HIFU neurosurgical treatments. This applicator design is equivalent

to a conical transducer design. Because the focal zone is small, this design is also

categorized as the highly focused transducer design. The major drawback of the

reflector design is its large aperture, which limits its use in clinical applications due

the size of acoustic "windows" available in the body surface.

Figure 1.5: Schematic of the reflector focusing applicator used by Fky [25].

Page 27: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERMAL THERAPY 14

Phased Array Design

Since the size of individual thermal lesions generated by single-focus transducers is

small compared to the size of tumours, several groups have suggested the use of

multi-focus transducer designs that would produce large thermal lesions [19,23,45].

The simplest way to produce a multi-focus ultrasound field is to place several single-

focus transducers adjacent to each other. Since the size of the applicator aperture is

limited by the size of available acoustic "windows" at the body surface, individual

transducen must have a small aperture. However, the small aperture reduces the

transducer focusing, which may lead to damage to the surrounding tissue during the

treat ment.

A phased array applicator uses many small transducer elements to create indi-

vidual foci. These small transducer elements are individually electronically phased

such that waves generated by the elenients interfere constructively at the focus.

Since the transducer elements that contribute to the constructive interference are

distributed over the entire aperture, sufficient focusing can be achieved.

Phased array applicators were introduced to the field of ultrasound thermal

therapy in the early 1980s. Several different designs have been proposed and inves-

tigated including annular or concentric-ring arrays [li', 521, sec tor-vortex arrays (711,

spherically sectioned phased 1-D arrays [47,48], cylindrically and spherically sec-

tioned phased 2-D arrays [14,19,23]. Annular phased arrays, consisting of CO-planar

concentric ring elements, generate annular foci. Annular foci can be used to heat

the periphery of the tumour, where the cooling is expected to be highest, with the

interior of the field being heated over time by conduction. A major drawback of

annular arrays is that unwanted, intense secondary foci can be produced on axis be-

yond the annular focus [35]. These secondary foci can damage normal tissues during

treatment. The second drawback of annular array is that if strong localised cooling

occurs inside the annulus, for example near a large blood vessel, the interior region

may not reach a therapeutic temperature [35]. A modified annular array design, i.e.

sector-vortex array design [71], and other array designs have been proposed to elim-

inate these drawbacks. The sector-vortex phased array is geometrically focused and

consists of concentric rings, each of which further consists of sector elements. Like

annular arrays, the sector-vortex phased array also produces annular foci. However,

Page 28: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERMAL T K E W Y 15

secondary foci on the axis are eliminated by controlling the phases of individual

sector elements,

Phased 2-D arrays, consisting of a 2-D grid of square shaped elements, can

produce multiple focus fields of various shapes and sizes, allowing greater flexibility

to control thermal lesion formation than sector-vortex arrays. Another advantage of

2-D arrays is that the foci of these arrays can be scanned electronically, eliminating

the need for mechanically moving the applicator. The general drawback of phased

2-D array designs is they require a large aperture and small element spacing to

produce good focusing without the production of gating lobes [70]. This results in

a large number of array elements, and consequently great cost and complexity. The

use of geometrically focused 2-D arrays, such as cylindrically or spherically sectioned

arrays. reduces the requirement for small element spacing. However, larger array

elements then limit the range over which foci may be created to the vicinity of the

geometric focus of the array. Due to the complexity and cost, the most sophisticated

phased 2-D array built for thermal therapy applications is a 256 element spherically

sectioned array [14]. Although this array offered much better focusing than its

predecessor, it still suffered from significant secondary foci when phased to generate

cornplex focus patterns (the number of foci was more than 8). When synthesizing 9

focus or 16 focus fields to heat large tissue volumes, this 256 element phased array

had to switch ternporally betwcen a series of multiple focus patterns each with

<8 foci, appearing a t various positions. Thermal lesion volumes of approximately - 5.4 cm3 were successfully produced in pig muscle tissue by using this 256 element

phased array in a single exposure.

Multi-focus Acoustic Lens/'liansducer Combination Design

Due to the great cost and complexity of phased 2-D array applicators, multi-focus

acoustic lens/transducer combinations [LTC] have been developed to mimic their

focusing ability for mild heating ultrasound applications 1451. Figure 1.6 shows a

schematic of an LTC. Unlike the smooth surfaces of single focus converging lenses,

surfaces of the multiple focus lenses are machined into a 2-D grid of small elements

of different thicknesses. The differences in thickness between lens elements results in

phase shifts, similar to those introduced electronically in the case of phased arrays.

Page 29: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.2. HIGH INTENSITY FOCUSED ULTRASOUND THERMAL THERAPY 16

rA S a t h A-A

Figure 1.6: Schematic of the 2-D acoustic lens.

The primary advantage of LTCs over phased arrays is that LTCs consisting of

a large number of small elements are simple to build and inexpensive. The greater

number of elements rnay allow more comylex multiple focus patterns. For example,

LTCs of 2000 elements have been constructed, which were ahln to produce cornplex

focus patterns such as a 2 plane 18 focus field (451. Although LTCs are less flexible

than phased arrays in that an LTC can only produce a single multi-focus pattern,

many different lenses may be coupled to a single transducer, and lenses may be

designed for individual patients.

Whereas elements of a phased 2-D array can generate waves of both different

amplitudes and phases, elements of an LTC produce waves of identical amplitude,

because the LTC elements are activated by a single ultrasound source. The lack of amplitude modulation in the case of LTCs may result in some degradation of

the focus pattern relative to that created by a phased array using both modulated

amplitude and phase. However, a theoretical study, by Lalonde et al. [45], com-

paring focal plane intensity distributions of the 8 focus fields generated by a 2000

element phased array using only phase modulation or using both phase and ampli-

tude modulation reported only srnaIl differences. There was up to 10% reduction on

intensity gain3 at the foci in the case of "phase only modulation" compared to phase

and amplitude modulation. This difference was found to have only a small effect

on the simulated as well as measured temperature distributions in hypertherrnia

treatments [45].

31ntensity gain at foci is defined as the ratio of the sum of intensities at al1 focus points to the sum of intensities at the applicator surface.

Page 30: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.3. OBJECTIVE 17

To date, LTCs have only been investigated for mild heating ultrasound treat-

ments [44,45]. The ability of LTCs to mimic the focusing of phased arrays and

generate large thermal lesions for high temperature thermal t herapy applications

has not yet been explored. The effect of "phase only modulation" of LTCs on the

shape and size of high temperature thermal lesions remains to be investigated. These

questions form the second part of this thesis.

1.3 Objective

HIFU thermal therapy is a promising treatment modality for oncology applications.

However , thermal lesions produced by highly focused transducers are too small to

treat the entire tumour volume in a clinically practical t h e period. The use of

moderately focused spherical transducers may be able to rediice the treatment time.

The first objective of this study is to theoretically evaluate the difference between

the total times required by a highly and a moderately focused spherical transducer

to treat the same target under identical treatment conditions.

LTCs can mimic phased 2-D arrays to produce multi-focus fields, but with

substantially lower cost and complexity. Multi-focus fields allow the production of

large thermal lesions. It has been demonstrated that the use of ptiased 2-D arrays

can reduce the treatment times for large tissue vdumes compared to the use of

highly focused spherical transducers. However, LTCs have only been investigated

for mild heating applications. The second objective of this study is to demonstrate

theoretically that an LTC requires significantly shorter time for the treatment of a

large tissue volume than either a highly or a moderately focused spherical transducer.

1.4 Thesis Outline

In Chapter 2, a 3-D ultrasound-thermal model is developed for the calculation of

thermal dose distributions in tissue caused by ultrasound heating. The accuracy of

the model is verified by comparing the results with published experimental thermal

lesion data. A theoretical tumour model is developed which provides a standard

target to compare different transducer designs under identical treatment conditions.

Page 31: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

1.4. THESIS OUTLINE 18

Thermal dose profiles generated by a highly and a moderately focused spherical

transducer and total times required for them to treat the turnour are compared.

The moderately focused transducer is constrained to have the same aperture and

frequency as the highly focused transducer, while the focal length of this transducer

is selected such that the number of lesions required to ''filln the tumour is a minimum.

In Chapter 3, a design method of LTCs to produce multi-focus fields is de-

veloped. The method then is used to design LTCs in order to examine the effects

of LTC focus spacing and number of foci on the shape and size of thermal lesion.

The effect of "phase only modulation" of LTCs on the shape and size of thermal

lesions is also exarnined. Thermal dose profiles generated by an LTC design and the

time required for it to treat a 2 x 2 x 2 cm3 tumour are determined and compared

to those for the highly and moderately focused spherical transducers investigated

in Chapter 2. The ultrasound-thermal model and the tumour rnodei developed in

Chapter 2 are also adopted in this study.

In the final chapter, the results of the thesis are summarized and future work

is outlined. Future direct ions in ultrasound-t hermal modeling include extending

the ultrasound-thermal model to layered tissue geometries, taking into account the

change in tissue properties with temperature and time and incorporating the effect

of large blood vessels. Future directions in development of 2-D LTC designs include

validating beam profiles and thermal lesion predictions and investigating the effect

of layered tissue geornetries on complex focus patterns of LTCs. Furthermore, pre-

liminary work on rotating LTCs during heating to irnprove the shape of thermal

lesions for tumour treatmeots is presented.

Page 32: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Chapter 2

Theoret ical Evaluat ion of

Moderately Focused Spherical

Transducers for High Intensity

Focused Ultrasound Thermal Therapy

Abstract

High intensity focused ultrasound thermal therapy can be used to destroy deep

seated tumours non-invasively. Due to the strong focusing and simplicity, highly

focused spherical transducers have been widely adopted for HIFU applications.

However, highly focused spherical transducers are not optimal because the ther-

mal lesions they generate are very small, leading to impractically long treatment

times. Moderately focused spherical transducen can produce larger thermal lesions

compared to highly focused transducers. However, due to the weak focusing, more

heat may be deposited in surrounding normal tissues. Although some encouraging

results have been reported, it is difficult to quantitatively evaluate the ability of mod-

erately focused spherical transducer to reduce the treatment times for large tissue

volumes. A theoretical study of comparing treatment times required for highly and

Page 33: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.2. INTRODUCTION

moderately focused spherical transducers to treat the same target under identical

treatmeut conditions is presented here. A 3-D ultrasound thermal model was de-

veloped to calculate thermal dose profiles generated by ultrasound applicators. The

accuracy of the model was verified by comparing the results to published in vivo

thermal lesion data. A tumour model was constructed with a 2 x 2 x 2cm3 tumour

volume located 5cm below the body surface. The treatment goal was to deliver

to the entire tumour volume a thermal dose higher than 240 EM& while sparing

the surrounding tissue by limiting the thermal dose level to less than 60EM43 in

regions more than 5 mm beyond the tumour edge. The highly and moderately fo-

cused spherical transducers studied here had an identical aperture and operating

frequency. The focal length of the moderately focused spherical transducer was cho-

sen so that the number of thermal lesions required for this transducer to treat the

tumour was minimized. It was demonstrated thet the moderately focused spherical

transducer required a total time of 40 min to treat the 2 x 2 x 2 cm3 tumour, while

the highly focused spherical transducer required 150 min. However, it was also found

that the moderately focused transducer produced more sub-lethal thermal dose in

the intervening tissue regions. For example, the 30EM13 thermal dose contour ex-

tended to approximately 1.5cm in front of the target, compared to lcm in the case

of the highly focused spherical transducer.

2.2 Introduction

High intensity focused ultrasound [HIFU] thermal therapy involves raising the tem-

perature of a pre-selected tissue volume to between 50°C and 90°C. The high tem-

perature causes rapid coagulation of the target tissue. Moreover, large thermal

gradients in the target region are desired so that the surrounding normal tissue is

spared. l k a t ment pro tocols adop t ing highly focused transducers' were successful

in achieving this goal. However, such protocols were problematic for the treatment

of large tissue volumes. Due to the small thermal lesion volume produced for each

individual exposure, a large number of ultrasound exposures was required for the

complete destruction of the target volume [81]. Moreover, when the t ransducer was

'See the definition in section 1.2.2.

Page 34: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

moved to deliver the exposures, some normal tissue in the nearfield remaineci in the

overlapping beam pathways, as illustrated in figure 2.1. Although the temperature

rise in the nearfield after a single exposure was low compared to that in the focal

region, continua1 exposures of the same tissue region in the nearfield could result

in a large temperature rise and significant thermal damage [34]. Damianou and

Hynynen [IO] have demonstrated that a cooling period must be employed between

exposures to allow tissues in the nearfield to cool. Consequently, the treatment time

for large tissue volumes were impractically long 1231. For example, a study of HIFU

Figure 2.1: During the transducer movement from left to right to form lesions side by side, some normal tissues in the nearfield remain in the pathway of the beam.

thermal treatment by Yang et al. [81] found that 4 hours were required for a highly

focused transducer to destroy an 1 cm3 tissue volume in rabbit liver.

One approach to reducing the treatment time is to enlarge individual thermal

lesions so that their number can be reduced. This can be achieved by increasing

exposure duration and/or exposure intensity. However, blood Bow cooling effects as-

sociated with long exposures and non-linear ultrasound beam propagation associated

with high intensities may make the control of lesion formation difficult. Enlarging

thermal lesions can also be achieved by employing a transducer with a larger focal

zone such as the moderately focused spherical transducer*. However, the large fo-

cal zone associated with the weak focusing results in more heat being delivered to

tissues surrounding the target. Due to this safety issue, only one system arnong al1

the reported HIFU heating systems has employed a moderately focused spherical

transducer [77]. Applying this heating system to a phase 1 clinical study, Visioli

2See the definition in section 1.2.2.

Page 35: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.2. INTRODUCTION 22

et al. [77] reported encouraging results that the treatment of a 1.3cm3 volume of

tumour tissue in kidney required approximately 30-35 min. This result appean to

be a significant improvement on the treatment times required with highly focused

spherical transducers. However, the conclusion should be drawn with caution as

many t reat ment parameten, including the shape and location of the target volume

and the criterion for determining cooling periods, are different between the exper-

imental studies using the highly focused transducer designs and those using the

moderately focused spherical transducer designs.

A second approach to increasing the volume of individual thermal lesions is

through the use of a phased array. Fan and Hynynen [23] have developed phased

2-D array ultrasound applicators to generate large thermal lesions, where the focal

zone volume was significantly enlarged compared to those of single-focus ultrasound

applicators. Daum et al. [14] reported that an approximately 3 cm3 volume of tissue

could be thermally destroyed in a single 20s exposure by a 256 element phased 2-D array using a 16 focus field. Compared to phased array designs, moderately fo-

cused spherical transducen are simple to construct, inexpensive and commercially

available. In addition, unlike the 256 element phased array which required long

exposures (20s) to ensure that tissue volumes were completely destroyed by multi-

ple foci, moderately focused spherical transducers, which generate single foci, can

adopt short exposures, making lesion formation less dependent on blood flow cooling

effects.

To quantitatively evaluate the use of moderately focused spherical transduc-

ers, a theoretical study is presented here to compare treatment times required for

a highly and a moderately focused spherical transducer to treat a 2 x 2 x 2cm3

turnour volume under identical treatment conditions. An ultrasound-thermal mode1

was developed and used to predict treatment times. Thermal dose profiles generated

by these two transducers were also compared to examine the effect of transducer

focusing on thermal damage produced in the surrounding tissues.

Page 36: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.3. METHODS

2.3 Methods

2.3.1 Ultrasound-Thermal Mode1

A three dimensional mathematical model was developed to predict ultrasound in-

duced thermal damage in tissues. The model consists of three parts (see figure 2.2).

Figure 2.2: Block diagram of numerical model to simulate ultrasound-thermal effects in tissue

Ultrasound Intensity Distributions

The first modeling step was to calculate the ultrasound intensity distribution gen-

erated by a transducer in a non-attenuating medium. The transducer surface was assumed to be composed of an array of small element sources of ultrasound energy.

The contributions €rom these sources to the intensity a t each point in the field were

superimposed, according to the Rayleigh-Sommerfeld integral [53].

= intensity at the field point (x,y,z) [W cm-*]

= particle velocity amplitude a t the transducer surface [cm s-'1

= wave number [cm-']

= speed of sound in the medium [cm s-'1

= density of the medium [g - cmw3]

= distance [cm] between the source element (xo,yo,xo) and the field point

= surface area of the source element [cm2]

Page 37: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.3. METHODS

Heat transfer in tissue

The temperature rise in tissue due to ultrasound exposures was predicted by the

Bioheat 'Pransfer Equation [BHTE] [57]

Wh Y, 2, t ) ~ t c t + wbca(T(x, y,z, t ) - Ta) = k V* T(x, Y, 2, t ) + SAR(x, y, z ) (2.2) dt

where

T

Pt

ct

Wb

Cb

T a k

SAR

= temperature [OC] at the field point (x,y,z) at time t [s]

= density of tissue [g -cm-3]

= heat capacity of tissue [J g-' C-'1

= volumetric perfusion rate [g -s-' - cm-3]

= heat capacity of arterial blood [J m g - ' e o C-'1 = temperature of arterial blood [OC]

= thermal conductivity of tissue [W -cni-' 0" Cs'] = Specific Absorption Rate [W - cm-3], which is the amount of

power absorbed by tissue per unit volume

In this equation, it was assumed that heat transfer between blood vessels and tis-

sues occurred rnainly across the rnicrovasculature. The blood in the capillary bed

would instantly thermally equilibrate to the temperature of the surrounding tissues.

Heat transfer by large vessels which create significant temperature gradients in their

vicinity was not taken into account in this model.

The SAR was celculated using equation 2.3. To sirnplify the calculations,

it was assumed that the attenuated energy can be calculated by integating the

attenuation coefficient over the axial distance traversed.

where

P = amplitude attenuation coefficient [Np **cm-' -.MHz-', 1NP x 8.686 = ldB]

s = axial distance [cm] between the skin surface and the field

Page 38: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

point a = amplitude absorption coefficient [Np - cm-' - MHZ-'1

A finite difference algorithm was used to solve the BHTE in Cartesian coordi-

nates [41]. The temperature change was calculated at every node in a computation

domain, the size of which will be specified in section 2.3.2. Temperatures at the

boundaries of this computation domain were set to be constant at 37°C. The initial

temperature at each node was assurned to be 37°C. To simplify the calculations, al1

tissues wit hin the computation domain were assumed to possess identical acoustic

and thermal properties which remained constant during heating.

Thermal damage in tissue

Thermal dose to the tissue was quantified using the following equation [61].

where

TD

ttatd

t

= thermal dose [EMIS ] = total treatment time [s]

= time [SI

This formula has been verified for temperatures up to 57°C [3] and bas been suc-

cessfully used in predicting thermal lesion size in vivo [12].

In order to evaluate different transducer geometries in terms of treatment time under

identical treatment conditions, a tumour mode1 was developed which included the

tumour geometry, tissue properties, thermal dose targets for the tumour and thermal

dose limits for the protection of the normal tissue.

Page 39: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

liimour Geometry

The tumour geometry employed in this work is illustrated in figure 2.3. The tumour

Figure 2.3: Tumour geometry employed in this work.

selected was in the shape of a cube (2 x 2 x 2cm3), with its centre located 5cm

below the skin surface. A cube shaped target was useful as a first approach in

this theoretical investigation because the volume can be easily filled by any array

of individual, identical lesions, as illustrated in figure 2.3. Given this tumour size,

the size of the thermal computation domain was chosen to be 4 x 4 x 9cm3, to

achieve reasonable computation times (figure 2.4). The mesh spacings dx, dy and

Figure 2.4: Thermal computation domain and mesh spacing dx, dy and dz.

dz were chosen to be 0.5, 0.5 and 1 mm (figure 2.4). The time step was chosen to be

O. 1 S. These were the smallest achievable values given a pract ical computation time.

Page 40: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Al1 the computations were performed on Sun Enterprise 450 Server (400 MHz Ultra

SPARC-IICPU), and they were usually completed in from 36 hours to 1 week.

Tissue Properties

The values of acoustic and thermal properties of human muscle tissue were used in

this study (table 2.1). Muscle tissue has a lower perfusion rate than most other tis-

Ampl. atten. coeff. [Np cm-' * MHz-'] Ampl. absor. coeff. [Np cm-' MHZ-'1 Speed of sound [cm s-'1 Tissue density [g Blood density [g cm-=] Tissue heat capacity [J - g-' -O C-'1 Blood heat capacity [J g-' C-'1 Tissue conductivity [W cm-' a' C-'1 Perfusion rate [g - sdl cm-3]

Table 2.1: The acoustic and thermal parameters used in this study. These values are taken from Duck [Ml. ' The value of the absorption coefficient is chosen to be the same as that of the attenuation coefficient, in order to simplib the model.

sues, and therefore represents a "worst case" scenario because the nearfield heating

would be large.

Thermal Dose Thresholds

A thermal lesion was defined as the volume bounded by the 240 EM43 thermal dose

isosurface. The thermal dose of 240 EMd3 was selected because it lead to complete

necrosis in a variety of tissues [15]. The cooling time was chosen such that the

60 Endd3 thermal dose contour just extended t o 5 mm beyond the tumour boundary

(figure 2.5). Therefore, the normal tissue 5 mm beyond the tumour boundary would

receive a thermal dose of less than 60 EM4. This was to spare the normal tissue. The

value of 60 EM43 was selected because a thermal dose of less than 60 EM43 would not

result in severe damage in most critical tissues and organs (151. The 5 mm normal

tissue zone (figure 2.5) allowed the 240 EM43 thermal dose isosurface to completely

Page 41: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

encompass the tumour volume by sacrificing some normal tissue directly adjacent

to the tumour boundary.

Figure 2.5: Thermal doses to d l of the tumour tissue were intended ta be higher than 240EW3 and thermal doses to ail of the n o r d tissue 5 mm beyond the turnour boundary were intended to be lower than 60 EIbbJ.

Cooling periods were determined iteratively by simulating entire multi-exposure

treatments using the ultrasound-thermal model. After each simulation, the calcu-

lated 60 EMd3 thermal dose contour, based on an initially selected cooling time, was

compared with the specified 5mm limit. The cooling time was then adjusted and

the simulation repeated as required until the thermal dose limits shown in figure 2.5

were achieved. The cooling time in the first iteration was chosen such that tissues

everywhere could cool to 43OC before the next exposure was delivered.

2.4 Results

2.4.1 Verification of Ultrasound-Thermal Mode1 Accuracy

The accuracy of the mathematical model was determined by comparing predicted

transducer 6dB beamlengths3, beamwidthsJ and thermal lesion dimensions with

publis hed experimental data.

3See the definition in section 1.2.2. 4See the definition in section 1.2.2.

Page 42: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Applicator used by

Focal length (cm) Diameter (cm) Operating frequency (MHz) 6dB beamlength

Table 2.2: Measured and predicted 6dB beamlength and beamwidth of the applicators used by Sanghvi et al. [59], Chen et al. [7) and Watkin et al. [79]. The %difference between the measured and predicted values was caiculated as P " d i ~ ~ ~ ~ ~ ~ U r e ~ .

Sanghvi et al. i591 7.5 5.5

measured (mm) :predicted (mm) (% difference) 6dB beamwidth measured (mm) : predicted (mm) (% difference)

Verifying Intensity Distribution Predictions

4 7:6.6

Intensity distribution calculations for transducers in a non-attenuating medium were

Watkin et al. [791 15 10

(-5%)

0.65:0.75 (15%)

verified by comparing predicted beamlengths and beamwidths to the published ex-

Chen et al. 171 14 10

1.68 19:18.5

perimental data of three applicators (table 2.2). Two of these applicators, used

1.7 18:16.2

(-2%)

1.7:l.g (12%)

by Sanghvi et al. [59] and Watkin et al. [79], consisted of single spherical shaped

(-10%)

1.6:1.8 (12.5%)

transducers. The third applicator, used by Chen et al. [7], was a combination of

a planar transducer and a converging lens. The results indicate that the theoreti-

cal model overestimated the 6dB beamwidth for all three transducers. The largest

difference (15%) occurred for the highly focused transducer of Sanghvi et al. [59].

The difference reduced to 12% for the moderately focused transducer of Watkin

et al. (791. The model underestimated the 6dB beamlength by 5% for the highly fo-

cused transducer of Sanghvi et al. [59] and 2% for the moderately focused transducer

of Watkin et al. [79]. The large difference of 10% between the measured and pre-

dicted -6dB beamlengths for the lens/transducer combination was probably caused

by the spherical aberration of the lens.

Page 43: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.4. RESULTS 30

Verifying Thermal Dose Predictions

Thermal dose calculations were tested by comparing the predicted lesion size with

experimentally measured data [7]. Cases with and without blood flow were tested.

In Chen et al. [7], thermal lesions were consistently formed 2 mm below the surface

of rat liver using the focused spherical applicator described above. During these

experiments, the rat liver was exposed so that the liver could be accessed directly

with the ultrasound beam without passing through intervening tissue layers. In the

theoretical predictions, tissue acoustic and thermal properties were chosen to match

the experimental conditions (table 2.3). The volume bounded by the thermal dose

isosurface of 30 EMd3 (as opposed to the 240 EMd3 used in subsequent investigations)

was defined here as the thermal lesion volume. The thermal dose of 30EMd3 was

found to induce heptocyte loss and fibrosis in dog liver t i s s ~ e [15].

Ampl. atten. coeff. [Np cm-' +

Ampl. absor. coeff. [Np cm-' MHz-'] Speed of sound [cm s-'1 Tissue density [g cm-3] Blood density [g cm-3] Tissue heat capacity[J g-1 -O C-'1 Blood heat capacity [J g-' e 0 C-'1 Tissue conductivity [W cm-' 6" C-'1 Perfusion rate [g s-' cm-3]

Table 2.3: Tissue acoustic and thermal properties used in prediction of lesion size in rat liver [18].

Figure 2.6 compares the measured and predicted lesion diameters. The the-

oretical values of lesion diameter were measured from the calculated thermal dose

profiles, as illustrated in figure 2.7.

In most cases, the predicted thermal lesion dimensions fell within the ex-

perimentai error range, and agreement was better for the non-perfused cases. For

the perfused cases, the mode1 overestimated Iesion diameters by approximately 12%

for short exposures of 3s and 6s and underestimated lesion diameters by 10% for

the long exposure of 20 S. A possible explanation of the overestimation for 3 s and

6 s exposure cases was that the perfusion rate used in the theoretical predictions

Page 44: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

5 -

I 2 4 6 8 10 12 14 16 10 20 22

pulse durath: second

Figure 2.6: Cornparison of meavured lesion diameters [Cl with the results predicted by oiir theo- retical model.

Figure 2.7: Thermal dose distributions calculated using our t heoretical model. Treatment param- eters, tissue acoustic and thermal properties were chosen to match the experimentd conditions (table 2.3). The non-perfused, 12 s heating exposure case was shown here.

Page 45: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

was lower than that in the experimental situation. For the case of 20s exposure,

micro-vasculature collapse is expected to occur during heating [go], which would

have reduced the perfusion cooling effect and allowed for a larger lesion volume. In

our model, however, the perfusion rate was assurned to be constant regardless of

heating.

2.4.2 Cornparison of Highly and Moderately Focused Spher-

ical Transducers

Physical parameters of the highly focused (denoted as SPI) and the moderately

focused (denoted as SP2) spherical transducen investigated in this study are given

in table 2.4. Exposures of 10 s were adopted for both designs, which are sufficiently

brief to limit the blood flow cooling effects in the case of low perfused tissues [Il], but

also long enough to allow relatively large thermal lesions to be produceci. SP2 had

Radius of curvature (cm) Aperture (cm)

Beamlength (mm) 1 10 1 23

Operating frequency(MH2) Beamwidth (mm)

Table 2.4: Physical parameters of two focused spherical transducers studied in this work.

SP1 8.5 10

the same aperture and operating frequency as SPI, but a larger radius of curvature.

The radius of curvature of SP2 was selected such that the thermal lesion generated by

it in a 10s exposure would be slightly larger in the axial direction than the tumour.

The determination of the radius of curvature of SP2 was based on figure 2.8, which

shows lesion length and diameter5 after a 10 s exposure venus radius of curvature of

focused spherical transducers. The thermal lesions were placed at the centre of the

tumour. Spatial peak intensities [I,,] were chosen such that the predicted maximuni

temperature in the tissue after the exposure [TpeaL] was approximately 85OC to avoid

tissue vaporization. The values of I,,used are aven in table 2.5.

SP2 13 10

1 1.9

51n this work, the length or the diameter of a thermd lesion refen to the maximum dimension of the lesion measured axially or laterdy.

1 3

Page 46: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

- diameter of lesion .O 1 .-* -WB beamlength -.- -6dB beamwidth

l

# B

# 4

80 90 100 110 120 130 140 150 Radius of curvature: mm

Figure 2.8: Lesion length and diameter and transducer 6dB beadength and beamwidth versus radius of curvature of focused spherical transducers. The transducers were assumed to give a 10 s exposure to the tumour.

Table 2.5: Spatial peak intensities used in figure 2.8.

Focal length (cm) 8.5

Tpeak (OC) 85.1

ISP (W cm-2)

1003

Acoustic power output (W) 14.4

Page 47: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.4. RESULTS 34

The relative intensity distributions of SPI and SP2 and thermal doses gen-

erated by these two transducers For a single 10 s exposure are plotted in figure 2.9.

Both these transducen produced single, ellipsoid-sbaped focal zones and similarly

Figure 2.9: Figures (a) and (b) show relative ultrasaund intensity distributions produced by SPI and SP2 respectively. The distributions are shown as contour plots relative to the peaii value in 10% intervais starting at 10%. The axial planes at y=Omrn are displayed. The focal plane is at z=50mm. Figures (c) and (d) show the thermal dose profiles generated by SPI and SP2 respectively. The thernial dose profiles are displayed as contour plots, showing contours representing 240, 60, 30 and 5 EMd3, progressing outwards from the lesion centre. Dashed lines represent the tumour boundary. The lesions were produced by placing the centre of the focal zone at the centre of the tumour and deiivering a 10 s e-xposure.

shaped thermal lesioos (the volume bounded by the 240 thermal dose con-

tour). Figure 2.9 (a) and (b) show that intensity gradients generated by SP2 are

smaller than those generated by SPI. Figure 2.9 (c) and (d) show that both SPI and SP2 produced sharp thermal dose gradients over a single exposure. The dis-

tance betwecn the 240 EMd3 and 5 EMs3 contours in the axial direction for SPI was

Page 48: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.4. RES ULTS 35

slightly shorter than that for SP2.

Since the length of the thermal lesion generated by SPI was only 50% of the

tumour length (z direction), two lesions arrays at two different depths, separated by

10 mm, were required for the treatment. Only one lesion array was required for SP2

because the size of thermal lesions was equal to the size of the tumour along the z

direction. Moreover, more thermal lesions were required to cover al1 of the tumour

in the lateral direction for SPI because individual lesions were smaller in diameter

compared to those generated by SP2. Figure 2.10 illustrates transducer step patterns employed for SPI and SP2.

Axial movemcnt of SPI focus was chosen to step from the distal region of the tumour

towards the proximal region (figure 2.10(a)). This pattern was chosen to avoid

coagulated tissue being present in the pre-focal regions of subsequent exposures.

Coagulated tissue, which attenuates ultrasound more than uncoagulated tissue [13,

261, would prevent penetration of ultrasound fields to distal regions.

The lateral step patterns (figure 2.10 (b) and (c)) were suggested by Fan and

Hynynen (1996). The advantage of these patterns was that excessive heating was

avoided by delivering successive exposures to regions which were not adjacent to

each other. After determination of the transducer step patterns, treatments were

Figure 2.10: Tcansducer step patterns for SPI ((a) rutid and (b) lateral) and SP2 ( ( c ) lateral).

simulated to determine the cooling times required to spare the surrounding normal

tissue. Table 2.6 compares the number of exposures, exposure times, cooling times

and total treatment times required for SP1 and SP2 to treat the tumour. The

maximum tissue temperatures predicted are also given in table 2.6. They are higher

Page 49: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

than the Tpeakin single exposures, because of the heat accumulation, but lower than

the vaporization temperature of 100°C. Temperatures reached in the tissue during

Number of exposures Exposure time [sj spatial peak intensity [W cm-*] Acoustic power output [W] Maximum temperature [ O C ] Cooling time [s] Total time [hour]

Table 2.6: Total treatment times, number of exposures, exposure times and intensities, and cool- ing times for SP1 and SP2 to treat the tumour. The acoustic power outputs over each exposure and the maximum temperatures occurred in the tissue during multiple exposures are aiso listed.

the entire treatment using SP2 were shown in figure 2.11. Figure 2.1 1 shows that

immediately after the first heating exposure, which was given to the center of the

tumour, the tissue temperature at (O, 0, 4 cm) reached the maximum value. The

temperature then dropped during the cooling period. As the transducer moved to

heat the peripheral region of the tumour, the peak temperatures at (O, 0, 4cm)

after individual exposures gradually decreased, because little amount of energy was

deposited at this location. The gradua1 increase in temperature at (O, 0, 3.5cm)

and (0, O, 1 cm) during the initial 15 exposures indicates that heat accumulates at

these locations. The highest tissue temperature ever reached during the treatment

at Icm below the skin surface was approximately 4 l 0 c .

Thermal dose profiles produced at the end of the treatments using SPI and

SP2 are displayed in figure 2.12. Figure 2.12 shows that in both treatments, some

small volumes of tumour tissue received a sub-lethal thermal dose of less than

240 EM13. This could have been avoided if more lesions had been formed. Compar-

ison of figure 2.12(a) and figure 2.12(c) indicates that, a t the end of the treatments,

the thermal dose delivered to the nearfield by SP2 was greater than for SPI. For cornparison, thermal dose profiles generated by SPI and SP2 with longer

cooling times are displayed in figure 2.13, where the cooling times were chosen such

that the 30 EM43 thermal dose contour extended approximately 5 mm beyond the

Page 50: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.5. DISCUSSION

351 I l 1 I O 500 Io00 1500 2000 2500

Treatment time: second

Figure 2.11: Temperatures reached in the tissue as a function of tirne during the tumour treatment using SP2. Temperatures of the tissue that was located on the central axis (x=O mm, y=O mm) were plotted.

tumour boundary a t the end of the treatnient (as opposed to the 60 EMd3 threshold

used in figure 2.12). Given this criterion, the cooling periods chosen were 70s and

120 s for SPI and SP2 respectively. The resulting total treatment times were 2.8

houn and 0.9 houn for SPI and SP2 respectively. Cornparison of figure 2.12(c) and

figure 2. U(c) shows t hat the thermal dose delivered to the nearfield was significantly

reduced due to the longer cooling period.

2.5 Discussion

The results showed that for the treatment of a 2 x 2 x 2 cm3 tumour 5 cm deep in the

body, the moderately focused spherical transducer required 42 min, 28% of the time

required for the highly focused spherical t ransducer . This demonstrates a significant

improvement in total treatment time by the use of the moderately focused spherical

t ransducer.

Alt hough the cornparison was performed under identical treatment condi-

tions, this conclusion should still be drawn with caution. Firstly, if the criterion for

choosing cooling periods is changed, the difference between treatment times required

Page 51: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.5. DISCUSSION

Figure 2.12: Therrnd dose profiles for tretttmcnts of the tumour using SPI (a, b) and SP? ( c d ) . The cooling time in the treatments was chosen such that the 60 EMdt3 thermai dose contour just extended to 5mni beyond the turnour boundary at the end of the treatments. Figures (a), (c) show the axial plane y=Omm, where the 5, 30 a d 60E& thermal dose cantours are ma.uimally c-xtended beyond the tumour boundary. Figures (b), (d) show the laterd planc z=50 mm. Thermal dose profiles are displayed as contour plots, representing 240, 60, 30 =and 5 Dashed lines represent the tumour boundary.

Page 52: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.5. DISCUSSION

Figure 3.13: Tïcemd dose distributions for treatments of the tuniour using SPI (a, b) and SP2 (c. d). The cooling time in the treatments was chosen such that the 30 EMt3 thermai dose contour just extended to 5mm bcyond the tumour boundary at the end of the trcatments. Figures (a), ( c ) show the axial plane y=Omm, where the 5, 30 and 60E& thermal dose contours are maximally extended beyond the tumour boundary. Figures (b), (d) show the lrrtcrd plane z=50mni. Thermal dose profiles are displayed as contour plots, representing 240, 60, 30 and 5 E&. Dashed lines represent the turnour boundary.

Page 53: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.5. DISCUSSION

by the highly and moderately focused spherical transducers may also change. For

example, when the 30 EM43 thermal dose contour extending no further than 5 mm beyond the tumour boundary was employed as the normal tissue thermal dose limit,

the moderately focused spherical transducer required 54 min to treat the same tu-

mour volume. This indicates that as the thermal sensitivity of the normal tissues

surrounding the target increases, the ability of moderately focused spherical trans-

ducen to reduce the treatnient time rnay be reduced, due to the increase in cooling

period required. Secondly, the ability of moderately focused spherical transducers

to reduce the treatment time also depends on the shape of the target. Figure 2.8 in-

dicates that thermal lesions generated by moderately focused spherical transducers

over single exposures are elongated. Thus, the use of moderately focused spherical

transducers enlarges thermal lesinns mainly in the axial direction compared to the

use of highly focused spherical transducers. Hence, for the treatment of tumour

volumes which are small in the axial direction, moderately focused spherical trans-

ducers will not offer any improvement in treatment time. On the other hand, the

use of moderately focused spherical transducers will be more effective in reducing

treatment time if the axial dimension of the tumour volume is larger than that

investigated here.

To simplify the calculations and reduce computation time, constant tissue ul-

trasound attenuation and constant perfusion were assumed in the ultrasound ther-

mal model. Tissue ultrasound at tenuation, however, has been demonstrated to

increase with temperature and time during heating [13,26]. In addition, perfusion

has beeii found to decrease with increasing temperature and time (41. The assump-

tion of constant perfusion rnay lead to an underestimation of the thermal lesion

volume (section 2.4.1). Since a relatively low perfusion rate was assumed in tbis

study, the assumption of constant perfusion was expected to have a small effect on

treatment time cornparison result. A preliminary study conducted by Kolios [38]

has demonstrated that the assumption of constant ultrasound at tenuation resulted

in an overestimation of the axid dimension of the thermal lesion volume and an

underestimation of the lateral dimension of the thermal lesion volume. However,

the efFect of the assumption of constant ultrasound attenuation on treatment time

calculation result is not known.

Page 54: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.6. CONCL USIONS 41

This study also examined the effects of the weak focusing on treatrnent out-

cornes including thermal dose profiles and the selection of cooling periods. To limit

the thermal dose to the surrounding tissue below the threshold, the moderately

focused spherical transducer required a cooling period of 90 s between exposures,

as compared to 60s for the highly focused spherical transducer. Although the dif-

ference between thermal dose gradients generated by the highly and moderately

focused spherical transducers is small over single exposures, it increases over multi-

ple exposures in the nearfield region, due to heat accumulation. This indicates that

the use of a moderately focused transducer will lead to more (sub-lethal) thermal

dose deposited in the intervening tissue regions.

In order to furtlier reduce the number of thermal lesions to make treatments of

large tissue volumes more efficient, transducer designs capable of producing thermal

lesions larger in the lateral dimensior must be employed. Phased 2-D arrays [14]

and rnulti-focus acoustic lens/transducer combinations [45], which are capable of

generating multiple focus fields, are potential solutions. These transducer designs

will be investigated in Chapter 3.

2.6 Conclusions

A 3-D mathematical model was developed to predict ultrasound-induced thermal

damage in tissue. The accuracy of the model was verified by comparing the predicted

results with published in vivo lesion data. The predicted thermal lesion size agreed

with measured results [7] for the non-perfused cases. For the case of high perfusion

rates, the difference between measured and predicted lesion sizes was attributed

to a difference between the perfusion rate in the experimental conditions and the

value used in simulations and to changes in perfusion during heating that were not

modelled. By using the ultrasound-thermal model, treatments of a 2 x 2 x 2cm3

tumour volume using a highly and a moderately focused spherical transducer were

simulated, and the total treatment times were determined. It was demonstrated

that the moderately focused spherical transducer required a total time of 42 min to

treat the tumour, 28% of the time required by the highly focused spherical tram-

ducer. However, the use of the moderately focused spherical transducer resulted

Page 55: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

2.6. CONCLUSIONS 42

in more (sub-let hal) thermal dose being delivered to the intervening tissue regions. The study suggests that transducer designs capable of generating larger thermal le-

sions should be employed to further reduce the total treatment time for large tissue

volumes.

Page 56: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Chapter 3

Theoret ical Evaluat ion of

Multi-focus Acoustic

Lens/Transducer Combinat ions

for High Intensity Focused

Ultrasound Thermal Therapy

3.1 Abstract

The diameters of thermal lesions generated by single-focus transducers are on the

order of a few millimetres, which are usuaiiy small compared to the diameter of

the tumour volume. The combination of a single transducer and an acoustic lens

incorporating a 2-D grid of "elements" can produce multiple foci. Thermal lesion

volumes generated by these multi-focus 1ensJtransducer combinations [LTC] may be greater than those generated by single-focus spherical transducers. Compared to

phased array designs, which can also generate multi-focus fields, LTCs are simple,

inexpensive, and may offer greater flexibility for producing complex focus patterns,

due to the greater number of elements. However, unlike phased arrays, whose ele-

ments can produce amplitude as well as phase modulation, LTCs can only use phase

modulation because al1 LTC elements are activated by a single source. A theoret-

Page 57: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.2. INTRODUCTION

ical evaluation of the ability of LTC designs to generate large thermal lesions for

high temperature thermal treatments is presented. A design rnethod, based on the

pseudoinverse method, was developed to determine uniform amplitude activation

signals required for LTCs to generate specified multi-focus fields. We demonstrate

that LTC "phase only modulation" bas very small effects on thermal lesion shape

and size. To use LTCs for thermal lesion formation, longer exposure durations of

50s are suggested even though variable blood flow may produce large thermal does

changes in the field. The effect of focus spacing and number of foci produced by the

LTC on the shape and size of thermal lesions was investigated. Specifically, thermal

lesions volumes generated by 9 focus LTCs with 2 m m or 2.5 mm focus spacing and

by a 16 focus LTC with a 2 mm focus spacing were compared. The results indi-

cate that increasing focus spacing or number of foci lead to enlargement of thermal

lesion volumes in both the lateral and axial directions. I t was demonstrated that

a 9 focus LTC with 2mm focus spacing required a total tiine of 30 min to treat a

2 x 2 x 2 cm3 tumour, compared to approximately 40 min and 150 min required by

moderately and highly focused spherical transducers respectively.

3.2 Introduction

It was found in Chapter 2 that moderately focused spherical transducers offered

significantly shorter tirnes for the treatment of a 2 x 2 x 2 cm3 tissue volume than

highly focused spherical transducers. However, there is a limitation on reducing the

lesion number because thermal lesions generated by focused spherical transducers

are longer in the axial direction than in the lateral direction. Thus, even if the

lesion length is identical to the tumour length, the lesion diameter is far from being

sufficiently large to cover the tumour in the lateral direction over a few exposures.

One approach to enlarge the lesion diameter is to use transducer designs which can

generate multiple laterally adjacent foci. When the focus spacing is sufficiently small

or the exposure duration is sufficiently long, individual thermal lesions will coalesce

into one single large lesion, due to thermal conduction.

Three ways to create multi-focus fields are being considered in this chapter.

The first way is to use several focused spherical transducers placed adjacent to each

Page 58: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.2. INTRODUCTION 45

other. Each transducer must have a small aperture because the size of available

acoustic "windows" a t the body surface is limited by bones or air cavities. However,

the small aperture reduces the transducer focusing which may lead to damage to

tissues surrounding the target. A second way is to use a phased 2-D array [14,23].

A phased 2-D array applicator consists of a 2-D grid of srnall piezoelectric elements,

each of which is an individually activated ultrasound source. By activating these

elements in a phase delay sequence, interference fields can be created which contain

multiple foci. Phased 2-D arrays can also be used to scan and switch between various

multiple focus fields electronically. To increase the focusing and to avoid grating

lobes [70], a large aperture and small element spacing are required, resulting in a

large number of array elements. However, the complexity and cost to build phased

array applicators increases dramatically with the number of array elements.

A third way to generate multi-focus fields is to use a combination of a single

transducer and a conjugate acoustic lens [45]. Such a lens is formed by cutting the

surface of a flat plastics into a 2-D grid of small elements of different thicknesses.

The different thicknesses create the phase shifts that are introdiiced electronically

in the case of phased arrays. Whereas elements of a phased array can generate

waves with modulated amplitude and phase, elements of a conjugate lens/transducer

combination [LTC] produce waves of identical amplitude because the LTC elements

are activated by a single ultrasound source. The lack of amplitude modulation may

result in some degradation of the focus patterns generated by LTCs. However, the

primary advantage of LTCs over phased arrays is that LTCs consisting of a large

number of small elernents are simple to build and inexpensive. A greater nuniber

of elements may allow more complex multi-focus patterns. Although LTCs are less

flexible than phased arrays in that an LTC can only produce a single multi-focus

pattern, many different lenses rnay be coupled to a single transducer, and Ienses

rnay be designed for individual patients.

A theoretical study conducted by Fan and Hynynen [24] has demonstrated

t hat phased 2-D arrays required significantl y shorter treatnient times for large tissue

volumes than highly focused spherical transducers. The phased 2-D arrays that Fan

et al. [23] investigated were spherically focused, to reduce the requirement for srna11

element size. However, the large element size of these arrays, which was 2 x 2 cm2,

Page 59: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

limited the range over which foci may be created to the vicinity of the geometric

focus of the arrays. The smallest element size being reported for the phased 2-D

array design is 0.65 x 0.65 cm2 [14]. This axray, which consisted of 256 elements and

was geometrically focused to achieve strong focusing, could coagulate tissue volumes

of 0.7cm(x axis) x 0.7cm(y a i s ) x 1.2 cm(z axis) using a 9 focus field for a single

exposure of 20 S.

To date, LTCs have been investigated for mild heating ultrasound treatments

by Lalonde and Hunt [44,45]. The element size of 2 x 2 mm2 was chosen for the

LTC designs as a compromise between increase in grating lobe levels and surface

roughness [43]. Ultrasound scattering from the rough surfaces reduced the efficiency

of transmission of ultrasound through the lenses. LTCs of 1000 elements have been

constructed, which were able to uniformly heat a 1 x 1 x l cm3 tissue volume to the

hyperthermie temperature by using a 12 focus field [43]. Conjugate lenses can be

coupled to either focused or unfocused transd~icers. I t was found that by combining

conjugate lenses wit h focused t ransducers, the roughness of the lenses was reduced.

The intent of this chapter is to evaluate theoretically the ability of LTCs to

mimic the focusing of phased arrays and to generate large thermal lesions for high

temperature thermal treatments. A design method, based on the pseudoinverse

method, was developed to determine the uniform amplitude activation signals for

LTCs to produce specified multi-focus fields. The effect of LTCs being restricted to

"phase only modulation" on the shape of the t herrnal lesion volume were examined.

The effect of the LTC focus spacing and number of foci on the thermal lesion shape

and size was also examined. The time required for a 2 mm spaced, 9 focus LTC to treat a 2 x 2 x 2cm3 tumour volume was deterrnined, and compared with those

required for the highly and moderately focused spherical transducers investigated in

Chapter 2. The ultrasound thermal model and turnour model developed in Chapter

2 were also employed in this study.

Page 60: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.3. METHODS

3.3 Methods

3.3.1 LTC Ult rasound Intensity Distribution Calculations

The ultrasound intensity distribution of an LTC was determined using equation 3.1.

This equation was modified from equation 2.1 to take account of the phase delays

created by different thicknesses of LTC elements. Note that al1 LTC elements must

share a single particle velocity amplitude value because they are activated by a single

source.

where

I = intensity at the field point (x,y,z) [W -cm-2]

P = complex pressure a t the field point (x,y,z) [Pa]

Un = amplitude of the complex particle velocity of the lens element n

(u, = [cm s-l]

0, = phase of the complex particle velocity of the lens element n

N = total number of lens elements

b a v e = wave number [cm-']

d, = distance [cm] between the lens element n and the field point (x,y,z)

(see figure 3.1)

S = surface area of a lens element [cm2]

j =J-i

The phase delay (-r < On < T ) was determined from the thickness of the LTC element n by equation 3.2.

where

f = operat ing frequency [MHz] an = distance [cm] between the surface of the transducer and the lens

Page 61: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4:

Figure 3.1: Coordinates and parameters used in the calculation of LTC field distributions.

element n (figure 3.1)

th = thickness [cm] of the lens element n

cm&, = speed of sound [cm s-*] in water

C~ens = speed of souiid [cm s-'1 in the lens

3.3.2 LTC Design Method

An LTC design method, based on the pseudoinverse method [19], was developed

here to determine the phase delay (O,, see equation 3.1) required for elements of an

LTC to produce a specified multi-focus field.

3.3.2.1 Pseudoinverse Method

The pseudoinverse method was developed by Ebbini et al. f19) to determine the

amplitude and phase (un and O,, see equation 3.1) required for phased array elements

to produce specified intensity values a t intended foci and minimum acoustic power

elsewhere (figure 3.2). The amplitude values (un, n = 1 N) determined by the

pseudoinverse method usually are not identicai, whereas in the case of the LTC the

amplitude values of al1 elements must be identical because the lens is coupled to

a single transducer . Hence, the pseudoinverse met hod canno t be directly applied

to LTC design. Because only small modifications to the pseudoinverse method was

made for the design of LTCs, the Following discussion det ails the pseudoinverse

method as applied to the phased array, and the LTC case is presented after.

Page 62: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

cle

field point M

Figure 3.2: Coordinat es and parameters used in the pseudoinverse method.

To apply the pseudoinverse method, equation 3.1 is rewritten in the matrix

form:

P = HU

= vector representing complex field pressure, where p, and 4, represent the amplitude and phase of the complex pressure at the field point m

respectively

M = number of field points interested

H = M x N matrix describing the propagation of the ultrasound

wave from each element of the array to each field point

Û = vector representing complex particle velocity, where un and 0,

represent the amplitude and phase of the complex particle velocity of the phased array

element n respectively

N = number of phased array elements

d n ~ = distance between the field point rn and the phased array element n

(see figure 3.2)

Page 63: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

S = surface area of the phased array element

Equation 3.3 is solved for u (If equation 3.3 had been solved for the design

of an LTC, there would be a constraint of un = un+l). To ensure that solutions

to equation 3.3 exist, the number of field points (M) in P must not be larger than

the number of elements (N) contained in a phased array. When M is less than N ,

and the positions of these M field points are such that al1 of the row vectors of H are linearly independent, equation 3.3 possesses an infinite number of solutions [66].

One of these solutions is called the minimum norm solution (equation 3.4 [49]),

because the norm of this solution (zSi l n2 ) is the minimum of al1 solutions.

where

H* = conjugate transpose of H (661

Because un2 is proportional to the intensity of the wave generated by the nth element,

the minimum norm solution [MNS] results in a total energy present a t the phased

array field that is minimized for a given P. Thus, with the MNS, the likelihood

is that the greatest constructive interference, and therefore the greatest pressure

amplitude, will occur at the field points specified in P, the intended foci. However,

any solution to equation 3.3. including the MNS, cannot guarantee that the intended

foci will be produced. It is possible that field points not specified in can possess

greater constructive interference than the field points specified in P. Because the

MNS is the most likely solution to produce an intended multi-focus pattern, it was

adopted in the design of phased arrays [19,21,23]. Thus, M can represent the

number of foci in the specified field, dm, is the distance between the focus rn and

the array element n, where p, and #m are the relative amplitude and phase of the

complex pressure at the focus m. Because H*(HH*)-l is called the pseudoinverse

matrix [66], the use of the MNS (equation 3.4) is also called the pseudoinverse

method.

Since the pseudoinverse method could not be directly applied to the design of

LTCs, a method based on the pseudoinverse method was developed for LTC design.

Page 64: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

The first step in this method was to use the pseudoinverse method to determine the

amplitude and phase required for applicator elements to produce specified foci. The

second step was to calculate the thicknesses of LTC elements from the phase values

determined by the pseudoinverse rnethod using equation 3.2, where the amplitude

values determined by the pseudoinverse method was ignored.

3.3.2.2 Phase Optirniration

The values specified in f represent the complex pressures at the intended foci. Since

the intensity of the wave is a phase-independent quantity, changing the pressure

phase value of the acocistic waves a t the intended foci will not affect the intensity

values at these foci. By varying the relative pressure phase values at the intended

foci, it is possible to increase the constructive interference at these foci, and therefore

increase localization of energy in the field. This can be explained rnathernatically.

For eacli choice of P, there is an MNS. Given the nurnber, position and pressure am-

plitudes of foci, there exists one particular phase pattern of foci ([$1 * - * 4, * * - $bM])

which will produce an MNS of the lowest norm. Thus, this phase pattern of foci,

compared to other phase patterns of foci, results in the greatest likelihood that the

pressure amplitude values at each focus in P will be greater than that a t any other

field points. To obtain this particular phase pattern, the following equation was

derived [21]

where

G = intensity gain, a measure of energy concentration at the foci relative

to the entire field

The phase pattern of P which minimizes the norm of the MNS (CL, un2) maximises

G. Substit uting H*(HH*) P (equation 3.4) for Ûdnnorm, equation 3.5 is expressed

Page 65: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Based on this equation, the phase pattern of P which maximizes G can be found

iteratively (201. A direct solution, according to Ebbini and Cain [21j, is the phase

pattern of the eigenvector that corresponds to the smallest eigenvalue of matrix

( H H ) Although this solution results in a suboptimal value of G, it was employed

because of its simplicity.

3.3.2.3 Summary of LTC Design Procedure

The procedure for the design of an LTC to produce a multi-focus field is summarized

as follows

Step 1: A multi-focus pattern was specified, including the number and position of

foci and the pressure amplitudes at these foci. The specified pressure amplitude

values were normalized to the maximum of al1 foci. Parameters of the transducer

source were also specified, including the aperture, radius of curvature and operating

frequency.

Step 2: Optimization of the phase pattern of the intended foci was performed ac-

cording to the method discussed in section 3.3.2.2.

Step 3 The MNS was obtained using equation 3.4.

Step 4: The relative intensity distribution of this LTC was calculated using equation

3.1 where 0, were equal to the phases of the MNS.

3.3.3 Methods for Determining 'Ikeatment Times

The total time required for an LTC to treat a 2 x 2 x 2 cm3 tumour, illustrated in

both figure 2.3 and figure 3.3, was determined using the ukrasound-thermaI mode!

developed in Chapter 2.

3.4 Results

The LTCs investigated here employed identical spherically focused transducers as

the energy source, which had a 10 cm aperture, an 8.5 cm radius of curvature and was

operated at a frequency of I MHz. This transducer was highly focused so that when

multi-focus lenses, which were diverging lenses, were placed in front, sufficiently

Page 66: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.3: Geometry of the tumour model in Cartesian coordinates. Dashed lines represent the tumour boundary. The body surface is located at the z=0 mm plane.

strong focusing could still be maintained in the field. Al1 the LTCs consisted of

2000 elements with each element being 2 x 2 mm2.

3.4.1 Comparison of LTCs and phased arrays

To evaluate if the shape and size of thermal lesions generated by LTCs are similar

to those produced by phased array, ultrasound intensity distributions and thermal

dose distributions produced by a 2-D phased array and an LTC were compared

where these two applicators possessed identical physical parameters, including the

diameter, radius of curvature, number of elements and operating frequency, and

were designed to produce identical multi-focus patterns. The pseudoinverse method

was used in the design of the phased array. The LTC was designed using the method

developed in this study. The thermal dose distributions were calculated using the

ultrasound-thermal model and tumour model developed in Chapter 2.

The ultrasound intensity distributions produced by an LTC (denoted as

LTC1) and a phased array (denoted as PA) that possessed identical physical pa-

rameters and were designed to produce an identical 9-focus pattern (figure 3.4) are

shown in figure 3.5. In order to compare the intensity distributions of multi-focus

applicators in a useful way, both 2-D contour plots of the applicator focal plane in-

tensity distribution and 6dB and lOdB isosurfacel plots of the entire applicator field

are shown. The 2-D contour plots show that, whereas the intensity values at the - - - -

'The 6dB and lOdB isosurfaces refer to the isosurfaces of25% and 10% of the intensity maximum respectively.

Page 67: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.4: 9 focus pattern with 2 mm focus spacing. All foci were assigned to have the identical pressure amplitude values.

foci of PA are identical, intensity values at the foci of LTCl Vary considerably, but

these values were still in a symmetrical pattern. The 6dB isosurface plots indicate

considerable difference between the shapes of the focal zone volumes of LTCl and

PA. Whereas the individual focal zones of LTCl varied in length and diverged at

angles at the proximal and distal ends, the individual focal zones of PA were nearly

identical in length and al1 parallel to the applicator axis. The lOdB isosurface plots

show the similarity between the focal zone shapes of LTCl and PA. In both cases,

individual focal zones varied in length and diverged at angles at both the proximal

and distal ends. Figure 3.6 plots the lens surface profile of LTC1. The maximum

thickness of the elements is 5 mm. The minimum thickness is approximately 0.9 mm.

The thermal dose profiles generated by LTCl and PA for exposure durations

of 20 and 50 s are shown in figure 3.7. The exposures were delivered such that the

focal plane of the applicators was a t a depth of 5cm below the skin surface. The

values for IsPys (table3.1) were chosen such that the 60 E M 4 thermal dose contour

extended approximately 5 mm beyond the tumour boundary after the exposure. A sufficiently long cooling period (200 and 400 s for exposure durations of 20 and 50 s

respectively) was included to allow the tissue to cool to approximately 40°C. This

was necessary for calculating the thermal dose to surrounding tissues because a

considerable thermal dose was delivered to these tissues after the power was turned

off. The thermal lesion volumes (volumes enclosed by the 240EMd3 isosurface)

generated by both LTCl and PA were irregular in shape, with "wings" outside

the main lesion volumes. To avoid t hese "wings" damaging surrounding tissues,

Page 68: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.5: The intensity distributions generated by LTCl (a, c, e), PA (b, d, f ) . Figures (a) and (b) display the contour plots at the focal plane (z=50 mm), wit h contours in 10% intervals of the peak value, beginning with the 10% contour. Figures (c) and (d) display the -6dB isosurfaces. Figures (e) and (f) display the -10dB isosurfaces.

Page 69: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

50

Element number 50 Element number

Figure 3.6: The lens surface profile of LTC1. The element size is 2 x 2 mm2.

LTCl (figure 3.7 (c))

Table 3.1: Values of I,,used in the study.

LTCl (figure 3.7 (e)) LTCZ (figure 3.9 (b)) LTC2 (figure 3.9 (c)) LTC3 (figure 3.12 (b)) LTC3 (figure 3.12 (c))

PA (figure 3.7 (d)) PA (figure 3.7 (f))

(SI 20 50 50 50 50 50 20 50

(OC) 66.5

(W cm-*) 510

71.9 72.4 65.8 70.6 65.6 80.3 81

270 270 225 270 230 530 323

Page 70: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.7: Thermal dose distributions generated by LTCl (a, c, e) and PA (b, d, f ) for 20s (a, b, c, d) and 50s (e, f) exposure duratious. Figures (a) and (b) display the thermal dose distributions at y=0 mm and z=50 mm. Figures (b), (c), (d) and (f) display 240 (opaque surface) and 60 Eh& (transparent surface) thermal dose isosurfaces. Dashed lines represent the t umour boundary.

Page 71: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

sufficiently small 1,'s were required. However, this also resulted in thermal lesion

volumes that were not sufficiently large to cover the tumour volume along the axial

direction. Hence, the use of LTCl and PA to treat a target volume of 0.8 cm(x axis) x

0.8 cm(y axis) x 2 cm(z axis) for a 20s exposure would lead t o 19% and 18% of the

target volume being delivered with a thermal dose of less than 240 EMd3. In the case of 50s exposure durations, the thermal lesion shapes were less

irregular due to the thermal conduction effect. The lateral dimension of the lesion

volumes also increased. To treat a larger target volume of 1 x 1 x 2cm3 for a

50s exposure duration, the use of LTCl would lead to a thermal dose of less than

240EM43 delivered to 10% of the target volume, compared t o 19% for the use of

PA.

3.4.2 Effect of Focus Spacing

The effect of LTC focus spacing on the shape and size of thermal lesions was exam-

ined by comparing the thermal dose profiles generated by two LTCs that produced

an identical number of foci, but with different focus spacings.

The ultrasound intensity distri butions produced by LTCl (investigated in

section 3.4.1) and an LTC (denoted as LTCZ) which was designed to produce the

same 9 focus pattern (figure 3.4) but with a focus spacing of 2.5mm are shown in

figure 3.8. The 2-D contour plots show that the difference between the intensity

values a t foci was larger in the case of LTC2. The isosurface plots show that the

difference between the lengths of individual focal zones was larger in the case of

LTCZ. The focal zones of LTCZ also diverged at larger angles tlian those of LTC1. The thermal dose distributions generated by LTCl and LTC2 for an exposure

durations of 50 s are displayed in figure 3.9. The exposures were delivered such that

the focal plane of the applicators was at a depth of 5 cm below the skin surface. In

the cases of figure 3.9 (a) and (c), the values for I,(table 3.1) were chosen such that

the 60 EM43 thermal dose contour extended approximately 5 mm beyond the tumour

boundary after the exposure and a cooling period of 400s, which allowed the tissue

to cool to approximately 40°C. In the case of figure 3.9 (b), the 1,value (table 3.1)

was chosen to be the same as that used in the case of figures 3.9 (a). Figure 3.9

shows that, given identical exposure intensities and durations, LTC2 generated a

Page 72: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.8: The ultrasound intensity distributions generated by LTCl (a, b) and LTC2 (c, d). Figures (a) and (c) display the contour plots at the focal plane (z=50 mm), with contours in 10% intervals of the peak value, beginning with the 10% contour. Figures (b) and (d) display the 6dB isosurfaces.

Page 73: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Figure 3.9: 240 Ebh3 (opaque surface) and 60Ebh3 (transparent surface) isosurface plots of thermal dose distributions generated by LTCl (a) and LTC2 (b and c) for a 50s exposure duration. Dashed lines represent the tumour boundary. In the case of figure (b), the Ispvalue was chosen to be the same as that used in the case of figure (a). In the case of figure (c), the values for i,,was chosen such that the 60 E1N3 thermal dose contour extended approximately 5 mm beyond the tumour boundary after the exposure and a cooling period of 400 S.

Page 74: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.4. RESULTS 61

larger thermal lesion volume, in both the axial and lateral directions, than LTC1. If

the 60 EMd3 thermal dose contour a t 5 mm beyond the tumour boundary was used

as the criterion to choose the exposure intensity, the sizes of thermal lesion volumes

generated by LTCl and LTC2 were nearly identical. To treat a 1 x 1 x 2 cm3 target

volume for an exposure duration of 50s, the use of LTCZ would lead to a thermal

dose of less than 240 EMd3 in 7% of the target volume, compared to 10% for the use

of LTC 1.

3.4.3 Effect of Number of Foci

The effect of number of foci on the shape and size of thermal lesions was examined

by cornparing the thermal dose profiles generated by two LTCs with identical focus

spacings, but different numbers of foci.

The ultrasound intensity distributions produced by LTCl (investigated in

section 3.4.1) and an LTC (denoted as LTC3) which was designed to produce the

16 focus pattern (figure 3.10) are illustrated in figure 3.11. The 2-D contour plots

Figure 3.10: 2mm spaced, 16-focus pattern.Al1 foci were assigned to have the same pressure amplitudes. Foci are numbered as shown.

show that the difference between the intensity values at foci was larger in the case

of LTC3 than for LTC1. The isosurface plots indicate that the difference between

the lengths of individual focal zones was larger in the case of LTC3. The focal zones

of LTC3 also diverged a t larger angles than those of LTC1. The thermal dose distributions generated by LTCl and LTC3 after a 50 s

exposure duration are displayed in figure 3.12. The exposures were delivered such

Page 75: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.4. RESULTS

5 - 4 -

3. 2 -

O v ii O 0 -1 .

-2. -3 - -4.

-5- -6 20 t m m -6- ' ' y: mm -6 -5 -4 -3 -2 -1 O 1 2 3 4 5 6

Figure 3.11: Ultrasound intensity distributions generated by LTCl (a, b) and LTC3 (c, d). Fig- ures (a) and (c) display the contour plots at the focal plane (z=50mm), with contours in 10% i n t e d s of the peak value, beginning with the 10% contour. Figures (b) and (d) display the 6dB isosurfaces.

Page 76: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.4. RESULTS 63

that the focal plane of the applicators was at a depth of 5 cm below the skin surface.

In the cases of figure 3.12 (a) and (c), the values for I,,(table 3.1) were chosen such

that the 60 EMd3 thermal dose contour extended approximately 5 mm beyond the

tumour boundary after the exposure and a cooling period of 400 s, which allowed the

tissue to cool to approximately 40°C. In the case of figure 3.12 (b), the &,value (table

3.1) was chosen to be the same as that used in the case of figures 3.12 (a). Figure

Figure 3.12: 240 E h 3 (opaque surface) and 60EM43 (transparent surface) isosurface plots of thermal dose distributions generated by LTCl (a) and LTC3 (b and c) for a 50s exposure duration. Dashed Lines represent the tumour boundary. In the case of figure (b), the I,,value was chosen to be the same as that used in the case of figure (a). in the case of figure (c), the values for I,, was chosen such that the 60 EWs thermal dose contour extended approximately 5 mm beyond the tumour boundary after the exposure and a cooling period of 400 S.

3.12 shows t hat , given identical exposure intensities and durations, LTC3 generated

Page 77: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

a larger thermal lesion volume, in both the axial and lateral directions, than LTC1.

If the 60 EMd3 thermal dose contour at 5 mm beyond the tumour boundary was used

as the criterion to choose the exposure intensity, the axial dimensions of thermal

lesion volumes generated by LTCl and LTC3 were nearly identical. The lateral

dimension of the thermal lesion volume generated by LTC3 was slightly larger than

that generated by LTC1. To treat a 1 x 1 x 2cm3 target volunie, the use of LTC3 would lead to a thermal dose less than 240EM43 in 14% of the target volume,

compared to 10% for the use of LTCl.

3.4.4 Multiple Exposure neatments of the nimour

Since the thermal lesion volumes generated over single exposures by the LTC designs

investigated here were not sufficiently large to cover a 2 x 2 x 2 cm3 tumour volume

in the lateral direction, multiple exposures treatments were required. Among al1

LTC designs investigated here, LTCl and LTC2 appeared to be preferable because

both LTCs can be used to achieve the destruction of 90% of the target volume over a

single exposure with the restriction on the thermal dose to the surrounding normal

tissue. Thermal lesion volumes generate by LTC2 had more "wings" than those

generated by LTC1. T h e "aings" might overlap when multiple adjacent exposures

were delivered to treat large tissue volumes, resulting in damage in surrounding

normal tissue regions. Therefore, LTCL was chosen to be the "optimal" design and

the total time required for LTCl to treat the 2 x 2 x 2cm3 tumour volume was

determined. Four exposures were required to treat this tissue volume with LTCl (figure 3.13). The cooling periods were chosen such that the tissue couid cool to

Figure 3.13: Lateral step pattern for LTC1.

approxiinately 40°C after each exposure. To avoid damage to surrounding tissues

due to overlaps between adjacent exposures, the exposure intensities used in the

Page 78: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.5. DISCUSSrON 65

multi-exposure treatment were reduced from 270 W cm-* to 250 W cm-*. The

number of exposures, exposure intensity and duration, cooling period and total

time required for LTCl to treat the tumour volume are given in table 3.2. For

comparison, the corresponding values for the highly (SPI) and moderately (SP2)

focused spherical transducers investigated in Chapter 2 are also given. Figures 3.14

Table 3.2: Total treatment times, number of exposures, e.xposure durations and intensities, cool- ing times and tumour volume under 240E& for LTCl, SP2 and SPI to treat the tumour volume.

Number of exposures Exposure time (s) spatial peak intensity (W - cm-2) Cooling period (s) Total time (hour) Tumour volume under 240 EMr3

and 3.15 displays the thermal dose distributions generated by LTC1, SP2 and SPI

at the end of the treatment. Due to the gaps between thermal lesion volumes

and the irregularity of the thermal lesion shapes, approximately 21% of the tumour

volume received a thermal dose of less than 240 EMA3 in the case of LTC1, compared

to 26% and 6% for SP2 and SPI respectively. The axial plane thermal dose contour

plots indicate that thermal dose gradients generated by LTCl in the nearfield are

higher than those produced by SP2 and SPI.

' LTCl 4 50 250 400 0.5 21%

3.5 Discussion

The results demonstrated that the shape and size of thermal lesion volumes gen-

erated by LTCs were similar to those generated by the equivalent phased arrays,

which had the same physical parameten as the LTC including the number of ele-

ments, aperture, radius of curvature and operating frequency. Hence, the effect of

LTCs being restricted to "phase only modulation" had a very small efFect on the

shape and size of thermal lesion volumes. The LTC design method was evaluated

for designing LTCs to produce symmetrical focus patterns with uniform intensity

Page 79: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.5. DISCUSSION

i -20 / O 10 M 30 4û 50 60 70 86 -% -16 0 .O 20

t m m 'I' mn

Figure 3.14: Thermd dose distributions generated by LTCl (a, b), SP2 (c, d) and SPI (e, f ) at the end of the treatment. Dashed lines represent the tuniour boundary. Figures (a). ( c ) and (e ) display the central axial plane y=Omm. Figures (b), (d) and (e) display the lateral plane z=50 mm.

Page 80: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.5. DISCUSSION

Figure 3.15: 240 EW3 (opaque surface) and 60 EM43 (transparent surface) isosurface plots of thermal dose distributions generated by LTCl (a), SP2 (b) and SPI (c) at the end of the treat- ment. Dashed lines represent the tumour boundary.

Page 81: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

a t foci. It was found that the resulting focus patterns were symmetrical, but with

reduced intensity values at the central foci. The effect of this "non-uniform focus

intensity" on the thermal lesion shape was small because the tissue located at the

central region of the focal zone could be heated through thermal conduction by the

foci a t the periphery. It can be predicted that the method may not be suitable for

designing LTCs to produce the focus fields which contain a large number of foci or

large focus spacings. This is because the degradation of intensity values at central

foci was already much greater for the 2.5 mm spaced, 9 focus LTC and the 2 mm

spaced, 16 focus LTC compared to that for the 2 mm spaced, 9 focus LTC. The results also demonstrated that both the lateral and axial dimension of

the thermal lesion volumes increased with LTC focus spacing or number of foci.

This finding was expected because an increase in focus spacing or number of foci

results in a decrease in applicator focusing, given that the aperture of the applicator

remains the same. The decrease in the applicator focusing leads to an enlargement

of the focal zone, and therefore the thermal lesion volumes, in both the lateral

and axial directions. Increasing the lateral dimension of the thermal lesion volume

while keeping the axial dimension unchanged is desirable because it can further

decrease the number of lesions required to treat the target volume. A prerequisite

for achieving this, by increasing the LTC focus spacing or number of foci, is a

corresponding increase in the focusing of the transducer source. Increasing the

focusing of an applicator can be achieved by increasing operating frequency, or

decreasing the f-number . For treatments of deep-seated tumour volumes, a sufficiently large radius

of curvature of the applicator is necessary. Tissue ultrasound attenuation increases

linearly with the applicator operating frequency. Therefore, increasing the operating

frequency essentially reduces the ability of the applicator to deliver sufficient energy

to deep seated tissue volumes. A larger aperture of the applicator is desirable for

increasing the focusing. However, in practice, the aperture size is limited by the

size of available acoustic "windows" at the body surface. Hence, given a target

volume, the minimal number of thermal lesions required for a multi-focus applicator

to treat this volume is limited by physical parameters including the depth of the

target, the ultrasound attenuation of the intervening tissues and the size of available

Page 82: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.6, CONCL USIONS

acoustic "windows" . This minimum lesion number can be achieved by optimizing

the treatment parameters such as the physical parameters of the applicator, and

exposure duration.

The physical parameters of those LTC designs investigated here, including

the focus pattern, aperture, radius of curvature and operating frequency, and the

exposure durations investigated here were not optimized for the treatment of the

target volume. The intent of this study is to demonstrate the potential of LTCs for

the treatrnent of large tissue volumes. The results showed that a 2 mm spaced, 9

focus LTC required a total time of 30 min to treat a 2 x 2 x 2cm3 tissue volume,

70% and 20% of the time required by the moderately and highly focused spherical

transducers investigated in Chapter 2 respectively. Furthermore, compared to the

use of the highly and moderately focused spherical transducers the use of this LTC design resulted in the sharpest thermal dose gradients in the nearfieid region at the

end of the treatment, due to the significantly long cooling periods which minimized

the heat accumulation in the normal tissue regions. One drawback of the use of

LTCs is that long exposure durations of 50s must be adopted to allow the thermal

conduction to improve the shape of thermal lesions. When these longer exposure

durations, compared to those (10s) used for simple spherically focused transducers,

were adopted, the temperature distributions as well as the thermal dose distributions

achieved in the tissue would be more dependent on the perfusion. Since perfusion

is one of the most uncertain factors in actual treatments, the use of longer exposure

durations may lead to difficulties in the control and prediction of thermal lesion

formation.

3.6 Conclusions

A theoretical evaluation of the ability of multi-focus acoustic lensltransducer corn-

binations [LTC] to produce large thermal lesion volumes for high temperature focus

ultrasound thermal treatments was presented. LTCs were demonstrated to be able

to produce thermal lesions with shape and size similar to those generated by phased

arrays. Since thermal lesion volumes generated by LTCs were irregular in shape,

long exposure durations of 50 s were required for the LTCs to produce thermal lesion

Page 83: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

3.6. CONCL USIONS 70

volumes of a useful shape. It was also found that, in order to enlarge thermal lesion

volume solely along the lateral direction, the LTC focusing needs to be increased.

The results dernonstrated that LTCs required significantly shorter treatment times

for large tissue volumes t han highly and moderatel y focused spherical tramducers.

Page 84: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

Chapter 4

Summary and Future Work

4.1 Summary

The goal of this thesis was to theoretically evaluate novel HIFU applicator designs,

including the moderately focused spherical transducer design and the LTC design,

for enabling practical treatment times for large tissue volumes. A 3-D mathemati-

cal model was developed to simulate ultrasound-induced thermal damage in tissue.

The accuracy of the model was verified by comparing predicted transducer focal

zone sizes and thermal lesion sizes with published experimental data. A theoretical

tumour model was constructed to allow treatment times of different transducer de-

signs to be compared under identical treatment conditions. The following sections

summarize the conclusions of each chapter and their contribution to the literature.

4.1.1 Moderately Focused Spherical Tkansducers

In Chapter 2, treatments of a 2 x 2 x 2 cm3 tumour volume using a highly and a mod-

erately focused spherical transducer were simulated using the ultrasound-thermal

model. Physical parameten of the moderately focused spherical transducer were

chosen such that the number of thermal lesions required for it to treat the tumour

was minimized. The simulation results demonstrated that the total time required

for the moderately focused sphericai transducer to treat the tumour was approxi-

mately 40 min, compared to 2.5 hours for the highly focused spherical transducer.

However, due to the weak focusing, more sub-lethal thermal dose was delivered

Page 85: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.1. SUMMARY 72

to the nearfield. Since the diameters of thermal lesion volumes generated by the

moderately focused spherical transducer were still small compared to the tumour

diameter, it was suggested that multiple focus transducer designs, which are able to

produce thermal lesions with a greater diameter, sbould be investigated to further

reduce the treatment time for tumoun of typical size.

This work was the first study to quantitatively examine the difference between

treatment times offered by highly and moderately focused spherical transducers for

coagulation of large tissue volumes. Since moderately focused spherical transducers

are simple, inexpensive and commercially available, treatment protocols adopting

these transducers may be preferable to those adopting more complex and expensive

transducer designs such as phased arrays. Examination of the thermal dose in

the nearfield generated by the highly and moderately focused spherical transducers

suggested that the moderately focused spherical transducer design should be adopted

with caution, because the use of this design resulted in more heat deposited in

the intervening tissue regions. Careful treatment planning including determining

sufficient cooling durations is therefore necessary when using the moderately focused

spherical transducer design to produce a safe and effective treatment .

4.1.2 Multi-focus Acoustic Lens/Transducer Combinations

In Chapter 3, the potential of LTCs to produce large thermal lesions for high temper-

ature thermal therapy applications was evaluated. A design method was developed

to determine the activation signals required for LTCs to produce speciîied multi-

focus fields. The results indicate that the shape and size of thermal lesion volumes

generated by LTCs and phased arrays are similar, although LTC elements only pro-

duce phase modulation. Heating exposures of 50s were found to be necessary for

LTCs to produce thermal lesions of sufficiently regular shape in order to cover most

of the tumour tissue and spare the surrounding normal tissue. The effects of LTC

focus spacing and number of foci on the shape and size of thermal lesions were also

investigated. As the LTC focus spacing or number of foci was increased, both the

lateral and axial dimensions of the thermal lesion volumes were increased. An in-

crease in LTC focusing was found to be necessary to increase the lateral dimension

of thermal lesion volume while keeping the axial dimension unchanged. The total

Page 86: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE WORK 73

time required for a 2mm spaced, 9-focus LTC to treat the 2 x 2 x 2cm3 tumour

was 30 min, compared to 40 min and 150 min required for the highly and moderately

focused spherical transducers investigated in Chapter 2 . Furthermore, the resulting

thermal dose gradients in the nearfield were largest in the case of the LTC. For ex-

ample, the thermal dose level of 5 EMd3 extended to approximately 1.7 cm in front

of the target volume, compared to 2cm and 2.9cm in the cases of the highly and

moderately focused spherical transducers.

The work presented in Chapter 3 is the fint investigation of the potential

of the use of LTCs in high temperature thermal therapy applications. The effect

of LTCs being restricted to "phase only modulation" was found to be srna11 on

the shape and size of thermal lesion volumes. Since long exposure durations of

50s are necessary for LTCs to produce thermal lesion volumes of a regular shape,

thermal treatments using LTCs may be liable to significant blood flow cooling effect.

Hence, treatment planing for LTCs in particular should take account of the blood

flow cooling effect. In summary, this study demonstrates that LTCs are prornising

applicator designs for high temperature thermal treatments of large tissue volumes,

and suggests that they should be further investigated for these applications.

4.2 Future Work

4.2.1 Ultrasound-Thermal Modeling

Ultrasound beams must penetrate intervening tissue layers, including skin, fat and

muscle tissue, to access the target in HIFU thermal treatments. These tissue layers

are known to possess varied acoustic and thermal property values, as indicated in ta-

ble 4.1. However, to simplify the calculations and reduce the computation tirnes, the

ultrasound-thermal model developed in this work assumed al1 these tissue layers to

possess identical acoustic and thermal properties. The model also assumed that val-

ues of al1 the tissue properties remained constant during heating. Changes in tissue

ultrasound attenuation and absorption coefficients as well as changes in perfusion

with temperature and time have been observed in experinental studies [4,13,26].

Comparing the results predicted by this rnodel with thermal Iesion sizes measured in

perfused liver tissues indicates t hat the assumption of constant perfusion resulted in

Page 87: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE WORK 74

tissue type ( Speed of Sound 1 Ampl. Atten 1 Density ( Perfusion

fat muscle kidney liver

skin

Table 4.1: Acoustic and thermal properties for normal human tissues [18]. - data are not d l - able.

an underestimation of lesion size for the case of long exposures. Kolios [40] developed

a non-linear thermal model to take into account the tissue ultrasound attenuation

changes during heating. The preliminary data indicates that thermal lesion volumes

predicted by the non-linear model were greater in diameter and smaller in length

compared to those predicted by the simplified model. Moreover, lesions "rnoved"

towards the transducer in the non-linear model. In addition, Fan and Hynynen [22]

demonstrated that a model which did not take into account the differences in speed

of sound between tissue layers underestimated the focal depth of a spherical trans-

ducer, and overestirnated the peak intensity value in tissues.

For the purpose of comparing thermal lesion volumes and thermal dose gradi-

ents generated by different transducer designs under identical treatment conditions,

the simplified rnodel developed in t his work is a reasonable approach, because al1 the

assumptions affect the results of different transducers in a similar manner. However,

to predict actuel t reatment resufts or to select treatrnent parameters (the exposure

intensity, cooling period and number of lesions) for a particular transducer design

to destroy a particular tumour volume and spare surrounding normal tissues, t his

model is not sufficient, and should be extended to simulate more realistic treatment

conditions.

Three improvements to this model are proposed. The first is to extend the

calculations of ultrasound intensity distributions to layered tissues, including skin,

fat, muscle, organ and tumour. The ultrasound intensity distributions in these

tissue layen can be calculated using a method developed by Fan and Hynynen (221.

This method takes account of beam reflection and refraction at tissue interfaces.

(m S-l) 1498

(NP cm-' MHZ-l)

0.4 (g

1.2 (g S-l cm-3)

0.0024

Page 88: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE: WORK 75

By determining the exact path of an ultrasound beam traversing through tissue

layers, this method can be used to accurately calculate the phase delay acquired

by the beam as it propagates through the tissue layers (figure 4.1). Thus, the

intensity distribution generated by the applicator in layered tissues can be accurately

predic t ed.

Figure 4.1: Diagram of a beam traversing through layered tissues. The phase delay this beam acquires from point source A to field point B ig cdculated as &($ A + $ +. :+ C. COS B. 1 7

where c, is speed of sound in tissue layer i.

The mode1 should also take account of changes in tissue ultrasound attenu-

ation and absorption coefficients and perfusion rate during heating. The method

described in Patankar [56] can be used to solve the nonlinear form of the bioheat

transfer equation. The decrease in perfusion rate due to heating can be modelled

by equation 4.1 based on the experimental data of Brown et al. 141.

where

Page 89: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE WORK

ww = initial blood perfusion [Np cm-' MHZ-'1

"Y = constant [s-'1

The data of Damianou et al. [13] c m be used to model the thermal dose dependency

of the ultrasound attenuation coefficient (equation 4.2).

where

Po = initial amplitude attenuation coefficient [Np cm-' MHZ-'1 PL = plateau amplitude attenuation coefficient [Np - cm-l MHZ-'1 TD430 = thermal dose where attenuation coefficient starts to increase [EM43 ] TD4% = thermal dose where attenuation coefficient reaches a plateau [EM43]

C = constant of fitting procedure [Np + cmm1]

The mode1 should also take into account large blood vesse1 effects. Kolios

et al. 1391 demonstrated that temperature gradients generated by the large vessels

of 0.3 mm or greater diameter dominated temperture distributions during high tem-

perature thermal treat ments. A 3-D thermal dose calculat ion model incorporating

the vasculature data (geometry and flow) of the target volume will be developed.

The thermal modelling method developed by van Leeuwen [75] will be adopted here.

In this method, heat transfer to large vessls is modelled by calculating the entire

volumetric temperature distribution using heat transfer coefficients derived from

analytical expressions [42], while heat transfer due to smaller vessels is modelled

by attaching local heatsinks at the terminal ends of the feeding vessels or by using

the bioheat transfer equation. 3-D vasculature data of the target volume will be

obtained by either Magnetic Resonance Angiography, Doppler Ultrasound or CT Angiography.

Page 90: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE: WORK

LTCs

4.2.2.1 Validation of Theoretical Predictions

A theoretical study of LTC designs for thermal therapy applications is presented in

Chapter 3. The next step is to experimentally validate these t heoretical predictions

by constructing a prototype LTC. The 3-D beam profiles of this LTC will be mea-

sured in a water tank. Thermal lesions will be formed in excised pig muscle tissues,

where the sizes and shapes of these lesions are measured. Of particular concern here

will be the ability of the LTC to deliver sufficient energy into the target zone to

form lesions. This concerns arises because the energy absorbed by the acoustic iens

increases in proportion to the energy delivered into the target zone. Since the lens

material possesses a low thermal conductivity, the energy absorbed by the lens may

result in melting.

4.2.2.2 Focusing in Inhomogeneous Tissues

Complex multi-focus patterns of LTCs were designed assuming that these patterns

were produced in homogeneous media. Tissue inhomogeneities are expected to in-

troduce phase shifts into the beams, and therefore might degrade the focus patterns.

Pliase errors in tissues can be divided into two parts: large, non-random erron due

to layered tissues discussed earlier and small, random errors [68]. Lalonde et al. [45]

added randorn phase errors from O to 0.2 n radians to the LTC element phases for an

8 focus LTC. No substantial changes in the positions of the foci were found. How-

ever, the intensities a t individual foci were reduced by up to 50%, due to increasing

phase incoherence. Random phase errors from O to 0.2 s radians were added to the

LTC element phases for the 2 mm spaced, 9-focus LTC (LTC1 in Chapter 3). No

changes in the sizes or shapes of the focal zone were found (data not shown). These

results suggest that multi-focus patterns of LTCs in real tissues may be insensitive

to random phase errors. Non-random phase errors, which could cause greater focus

degradation, ha?a not yet been investigated. Therefore, the next step is to deter-

mine the effect of layered tissues on the focal zone shape of LTCs and the shape of

thermal lesion volumes generated by these LTCs. The method described in section

4.2.1 will be employed for this study. Non-random phase errors can be corrected by

Page 91: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE WORK 78

incorporating tissue geometries into the calculation of the phase delays needed for

LTC elements. However, an LTC built with the phase-error correction is applicable

to a single tissue geometry. Moving the LTC to treat a large tissue volume may not

be successful when the variation in tissue geometry is large in the target region.

Rotation of LTCs

The results in Chapter 3 demonstrated that individual focal zones of LTCs varied

in length and diverged at angles at both the proximal and distal ends. As a result,

thermal lesion volumes generated by individual focal zones did not coalesce a t the

proximal and distal regions. To make these LTCs useful for tumour treatments,

long exposure durations were necessary to allow thermal lesion volumes generated by

individual focal zones to completely coalesce through thermal conduction. However,

due to the blood flow cooling effects, the use of long exposure durations may result

in difficulties in control and prediction of thermal lesion formation.

An alternative way to improve the shape of thermal lesion volumes generated

by LTCs is presented here. An examination of the focal zones of the LTCs investi-

gated in Chapter 3 revealed that the 4 individual focal zones located at the corners

of the focus pattern always had a greater length than other focal zones. If the LTC is

rotated upon its axis during heating (figure 4.2), the locations of the "corner" focal

zones in tissue will keep changing so that the energy deposited by these focal zones

in a certain tissue volume will be reduced. Some preliminary results are presented

here to demonstrate the efficacy of this method.

The 9 focus LTC with 2 mm focus spacing investigated in Chapter 3 (LTCl) was employed in this preliminary study. To simplify the computation, the case that

LTCl switched continuously between the orientations of 0,45, 90, 135, 180,225, 270

and 315" (figure 4.2) during heating was simulated. The LTC was assumed to stay

at each of the orientations for 0.5 S. The time spent between the orientations was

assumed to be zero. This continuous switching was used to mimic the rotation, and

is referred to here as pseudo-rotation. The thermal dose distributions generated by

this LTC with and without pseudo-rotation for a single exposure duration of 20s are

displayed in figure 4.3. The exposures were delivered such that the focal plane of

LTCl was at a depth of 5 cm below the skin surface. Spatial peak intensities were

Page 92: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE: WORK

Figure 4.2: Schematic of the rotation of the 9 focus LTC with 2mm focus spacing (LTC1 in Chapter 3). The orientations of 0, 45, 90, 135, 180, 225, 270 and 315' are as iiiustrated.

chosen such that the 60 EMd3 thermal dose contour extended approximately 5 mm

beyond the tumour boundary after the exposure and a cooling period which allowed

the tissue temperature to reduce to approximately 40°C. Comparison of the thermal

lesion shapes demonstrates the potential of the rotation method in that "wings"

nearly disappeared. To treat a target volume of 0.8 cm(x axis) x 0.8 cm(y axis) x

2 cm(z axis), the use of the LTC with the pseudo-rotation would lead to the thermal

dose of less than 240 EMd3 to be delivered to only 6% of the target volume, compared

to 19% for the use of the LTC wzthout the pseudo-rotation. The preliminary results

demonstrate that the rotation of LTCs during heating can significantly improve the

shape of the thermal lesion volume. As a result, short exposure durations (20s) as

compared to the 50s exposure durations can be adopted in treatments using LTCs,

and the treatment results will be less dependent on the blood flow cooling effect.

However, the rotation method may not be applicable to an LTC built with the

phase-error correction because the tissue geometry for which the LTC is corrected

is dependent on the LTC orientation.

Page 93: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

4.2. FUTURE WORK

Figure 4.3: 240 EW3 (opaque surface) and 60EW3 (transparent surface) isosurface plots of thermal dose distributions generated by LTCI with (a) and without (b) the pseudo-rotation for a 20s exposure duration. Dashed lines represent the turnour boundary.

Page 94: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

References

[l] P. D. Bevan and M. D. Sherar. B-scan ultrasound imaging of thermal coagu-

lation in bovine liver: log envelope dope attenuation mapping. Ultrasound in

Medicine and Bzology, 27(3) :379-387, 2001.

[2] B. E. Billard, K. Hynynen, and R. B Roerner. Effects of physical parameters

on high temperature ultrasound hyperthermia. Ultmsound in Medicine and

Biology, 16:409-420, 1990.

(31 J . Borrelli, L. L. Thompson, C. C. Cain, and W. C. Dewey. Tirne-temperature

analysis of ce11 killing of BHK cells heated at temperatures in the range of 43.5

C to 57.0 C. International Journal of Radiation Oncology, Biology and Physics,

19389-399, 1990.

[4] S. L. Brown, J. W. Hunt, and R. P. Hill. Differential thermal sensitivity of

tumour and normal tissue rnicrovascular response during hyperthermia. Inter-

national Journal of Hyperthennia, 8(4):501-514, 1993.

[5] S. L. Brown, X. L. Li, H. H. Pai, R. P. Hill, and J. W. Hunt. Observations of

thermal gradients in perfused tissues using water bath heating. International

Journal of Hyperthermia, 8(2):275-287, 1992.

[6] A. K. Burvo and G D . Andreevskaya. Action of high intensity ultrasonic vibra-

tions on malignant tumours in anirnals and man. Dokludy Akad Nuuk, 106:445-

448,1956.

[7] L. Chen, G . ter Haar, C. R. Hill, M. Dworkin, P. Carnochan, H. Young, and

J. P.. Bensted. Effect of blood perfusion on the ablation of liver parenchyma

Page 95: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

with high-intensity focused ultrasound. Physics in Medicine and Biology, 38(11):1661-1673, 1993.

[BI G. H. A. Clowes. A study of the influence exerted by a variety of physical and

chernical forces on the virulence of carinoma in mice. British Medicine Jozrrnal [clinàcal resource], 2: 1548-1554, 1906.

(91 W. B. Coley. The treatment of malignant tumours by repeated inoculations of

erysipelas with a report of original cases. Arnerican Journal of Medical Science,

105:497-511, 1893.

(10) C. Damianou and K. Hynynen. Focal spacing and near-field heating during

pulsed high temperature ultrasound therapy. Ultrasound in Medicine and Bi- ology, 19(9) : 777-787, 1993.

[Il] C. Damianou and K. Hynynen. The effect -.f various physical parameters on the

size and shape of necrosed tissue volume during ultrasound surgery. Journal of

the Acoustzcal Society of Arnerica, 95(3): 1641-1649, 1994.

[12] C. A. Damianou, K. Hynynen, and X. Fan. Evaluation of accuracy of a theoret-

ical mode1 for predicting the necrosed tissue volume during focused ultrasound

surgery. IEEE Transactions on Ultmsonics, Ferroelectrics and Frequency Con- trol, 42(2):182-187, 1995.

1131 C. A. Damianou, N. T. Sanghvi, F. J. Fry, and R. Maass-Moreno. Dependence

of ultrasonic attenuation and absorption in dog soft tissues on temperature

and thermal dose. Journal of the Acowticol Society of America, 102(1):628-

634, 1997.

[14] D. R. Daum and K. Hynynen. A 256-element ultrasonic phased array system

for the treatment of large volumes of deep seated tissue. IEEE Tkansactions

on Ultmsonics, Ferroelectrics, and Frequenc y Control, 46(5) : 1254-1268, 1999.

[15] W. C. Dewey. Arrhenius relationships from the molecule and ce11 to the clinic.

International Journal of Hyperthennio, 10(4):457-483, 1994.

Page 96: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

REFERENCES 83

[16] W. C. Dewey, L. E. Hopwood, S. A. Sapareto, and L. E. Gerweck. Cellular

responses to combinûtions of hyperthermia and radiation. Radiology, 123:463-

474, 1977.

[l?] J. P. Do Huu and P. Hartemann. Annular array transducer for deep acoustic

hyperthermia. IEEE Ultrasonics Symposium Proceedings, CH1698-9:705-710,

1981.

[18] F. A. Duck. Physical Properties of Tissues: A Comprehensiue Reference Book.

Academic Press, San Diego, 1990.

[19] E. S. Ebbini and C. A. Cain. Multiple-focus ultrasound phased array pattern

synthesis: optimal driving signal distribut ions for hyperthermia. IEEE ?hm-

actions on Ultmsonics, Femuelectrics and Freqvency Control, 36(5):540-548,

1989.

(201 E.S. Ebbini. Deep localised hyperthermia with ultrasound phased arrays using

the pseudoinverse pattern synthesis method. PhD dissertation, University of

Illinois, fibuna, IL, 1990.

[21] E.S. Ebbini and C.A. Cain. Optimization of the intensity gain of multiple-

focus phased-array heating patterns. International Journal of Hyperthennza,

7:953-9'73, 1991.

[22] X. Fan and K. Hynynen. The effect of wave reflection and refraction at soft

tissue interfaces during ultrasound hyperthermia treatments. Journal of the

Acoustical Society of Arnerica, 91:1727-1736, 1992.

[23] X. Fan and K. Hynynen. A study of various parameters of spherically curved

phased arrays for noninvasive ultrasound surgery. Physics in Medicine and

Biology, 41 (4):591-608, April 1996.

[24] X. Fan and K. Hynynen. Ultrasound surgery using multiple sonications-

treatment time considerations. Ultrasound in Medicine and Biology, 22(4) :Hl- 82, 1996.

Page 97: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

[25] W. J. Fry and F. J. Fry. hndamental neurological research and human neu-

rosurgery using intense ultrasound. IRE Iirrnsactions in Medical Electronics,

ME-7:166-181, 1960.

[26] M. R. Gertner, B. C. Wilson, and M. D. Sherar. Ultrasound properties of liver

tissue during heating. Ultmsound in Medicine and Biology, 23(9):1395-1403,

1997.

1271 G. M. Hahn. Hyperthermia and cancer. Plenum Press, New York, 1982.

[28] E. J. Hall. Radiobiology for the Radiologzst. J. B. Lippincott Company, Philadel-

phia, 1994.

[29] R. P. Hill and J. W. Hunt. The Basic Science of Oncology, chapter 21, pages

337-373. Pergamon Books Inc., 1987.

1301 J. W. Hunt, R. J. Lalonde, H. Ginsberg, S. Urchuck, and A. Worthington.

Rapid heating: critical t heoretical assessments of thermal gradients found in

hyperthermia treatments. International Journal of Hyperthermia, 7(5):703-718,

1991.

[31] J.W. Hunt, M. Arditi, and F.S. Foster. Ultrasound transducers for pulse-echo

medical imaging. IEEE transactions on Biomedical Engineering, 30:453-48 1,

1983.

[32] K. Hynynen. The threshold for thermally significant cavitation in dog's thigh

muscle in vivo. Ultmsound in Medicine and Biology, 17(2):157-169, 1991.

[33] K. Hynynen, A. H. Chung, V. Colucci, and F. A. Jolesz. Potential adverse

effects of high-intensity focused ultrasound exposure on blood vessels in vivo.

Ultmsound in Medicine and Biology, 22 (2) : 193-201, 1996.

[34] K. Hynynen, A. Darkazanli, E. Unger, and J. F. Schenck. Mri-guided noninva-

sive ultrasound surgery. Medical Physics, 20: 107-1 15, 1993.

[35] M. S. Ibbini and C. A. Cain. Problems associated with directly synthesized

annular heating patterns. IEEE îYansactions on Ultrasonics, Ferroelectrics

and Fkquency Control, 36:574-577, 1989.

Page 98: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

(361 S. Katsumi, M. Tsuchidate, M. Takagaki, and T. Wagai. The destructive action

of intense focused ultrasound to the malignant tumour tissues (the 4th and 5th

reports)-clinical study. Japan Med. Ultrason., 3:6-41, 1965.

[37] M. C. Kolios. Models of bioheat transfer dunng hyperthermia. Master's thesis,

University of Toronto, M. Sc. Thesis, 1994.

[38] M. C. Kolios. Applications and tmlidation of models of heat tmnsfer in perfused

tissues. PhD thesis, University of Toronto, Ph. D. Thesis, 1998.

[39] M. C. Kolios, M. D. Sherar, and J. W. Hunt. Blood flow cooling and ultrasonic

lesion formation. Medical Physics, 23(7): 1287-1298, 1996.

[40] M. C. Kolios, M. D. Sherar, and J. W. Hunt. The effects of temperature

dependent tissue attenuation,absorption and blood flow on ultrasonic lesion

formation. Ph ysics in Medicine and Biolog y, in preparation, 1998.

(411 M. C. Kolios, M. D. Sherar, A. E. Worthington, and J. W. Hunt. Modeling

temperature gradients near large vessels in perfused tissues. In M. A. Eba-

dian and P. H. Oosthuizen, editors, Fundamentals of Biomedical Heat Tans fer,

pages 23-30, HTD-Vo1.295, 1994. American Society of Mechanical Engineers.

[42] A. Kotte, G. Van Leeuwen, J. Debree, J. Vanderkoijk, H. Crezee, and J. La-

gendijk. A description of discrete vesse1 segments in thermal modeling of tissues.

Physics in Medicine and Biology, 4 1 (5):865-884, 1996.

[43] R. J. Ldonde. Field conjugate acoustic lenses for ultrasound hyperthermia.

PhD thesis, University of Toronto, Ph. D. Thesis, 1994.

[44] R. J. Lalonde and J. W. Hunt. Optimizing ultrasound focus distributions for

hypert hermia. IEEE Transactions on Biomedical Engineering, 42(10) :98l-90,

1995.

[45] R. J. Lalonde, A. E. Worthington, and J. W. Hunt. Field conjugate acoustic

lenses for ultrasound hyperthermia. IEEE Transactions on Ultmsonics, Ferro-

electrics and F~quency Control, 29:495-507, 1993.

Page 99: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

REFERENCES 86

[46] P. P. Lele. A simple method for production of trackless focal tissue lesions with

focused ultrasouad: physical factors. Journal of Ph ysiology, 160:494-512, 1962.

[47] F. L. Lizzi, M. A. Astor, C. Deng, and A. Rosado. Asymmetric focused ar-

rays for ultrasonic tumour therapy. IEEE Ultrusonics Symposium Proceedings,

1996:1281-1284, 1996.

[48] F.L. Lizzi, C.X. Deng, P. Lee, A. Rosado, R.H. Silverman, and D.J. Coleman.

A cornparison of ultrasonic beams for thermal treatment of ocular tumors.

European Journal of Ultrasound, 9:71-78, 1999.

[49] D. G. Luenberger. Optirnization by vector space rnethods. John Wiley and Sons,

Inc., 1969.

[50] J. R. Macfall, D. M. Prescott, H. C. Charles, and T. V. Samulski. H-1 MRI phase thermometry in-vivo in canine brain, muscle, and tiimor-tissue. Medical Physics, 23(lO): l?i'5-1782, 1996.

[51] J. B. Marmor. Interactions of hyperthermia and chemotherapy in animals.

Cancer Research, 39:2269-2276, 1979.

[52] K. B. Ocheltree, P.J. Benkeser, L. A. Frizzell, and C. A. Cain. An ultrasonic

phased array applicator for hypertherniia. IEEE Tmnsactions on Ultmsonics,

Ferroelectrics and A-equency Control, SU31:526-531, 1984.

[53] H.T. O'Neil. Theory of focusing radiators. Journal of the Acovstical Society of

America, 21516-526, 1949.

[54] J. Overgaard. The current and potential role of hyperthermia in radiotberapy.

International Journal of Radiation, Oncology, Biology and Physics, 16(3):535-

549, 1989.

[55] J. Overgaard, D. Gonzalez, M. C. C. M. Hulshof, G. Arcangeli, O. Dahl, and

S. M. Bentzen. Randomized trial of hyperthermia as adjuvant to radiotherapy

for recurrent or met astatic melanoma. The Lancet, 345540-543, 1995.

[56] S. V. Patankar. Numerical heat tmnsfer and fluid jiow. Hemisphere, Washing-

ton, D. C., 1980.

Page 100: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

[57] H. H. Pennes. Analysis of tissue and arterial blood temperatures in the resting

human forearm. Journal of Applied Physiology, 1:93-122, 1948.

[58] G. L. Rohdenburg and F. Prime. The effect of combined radiation and heat on

neoplasms. Archive Surgery, 2:1116-129, 1921.

[59] N. T. Sanghvi, F. J. Fry, R. Bihrle, R. S. Foster, M. H. Pliillips, J Syrus, A. V. Zaitsev, and C. W. Hennige. Noninvasive surgery of prostate tissue by high-

intensity focused ultrasound. IEEE Tmnsactions on Ultrasonics, Ferroelectta'cs

and Frequency Control, 43(6): 1099-1 1 10, 1996.

(601 S. A. Sapareto and W. C. Dewey. Thermal dose determination in cancer

therapy. International Journal of Radiation, Oncology, Biology and Physics,

10(6):787-800, 1984.

[61] S. A. Sapareto and W. C. Dewey. Thermal dose determination in cancer ther-

apy. International Journal of Radiation Oncology, Biology and Physics, 10:787-

800, 1984.

1621 R. Seip and E. S. Ebbini. Noninvasive estimation of tissue temperature response

to heating fields using diagnostic ultrasound. IEEE %nsactions on Biomedical

Engineerhg, 42 (8):828-839, 1995.

[63] M. D. Sherar, H. Clark, B. Cooper, J. Kurnaradas, and F. F. Liu. A dynamic

array filter for use with single element waveguide applicators. International

Journal of Hypertherrnia, 1993.

1641 M. D. Sherar, J. A. Moriarty, M. C. Kolios, J. C. Chen, R. D. Peters, L. C. Ang,

R. S. Hinks, R. M. Henkelman, M. J. Bronskill, and W. Kucharcyk. Compari-

son of thermal damage calculated using magnetic resonance thermornetry, with

magnetic resonance imaging post-treatment and histology, after interstitial mi-

crowave thermal therapy of rabbit brain. Physics in Medicine and Biology.

45:3563-3576, 2000.

[65] F. K. Storm, W. H. Harrison, and R. S. Elliott. Normal tissue and solid tumour

effects of hyperthermia in animal models and clinical trials. Cancer Research,

39:2245-225 1, 1979.

Page 101: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

[66] G. Strang. Linear olgebm and its applications. Harcourt Brace Jovanovich

College Publishers, 1986.

[67] Gai1 ter Haar. Ultrasound focal beam surgery. Ultrasound in Medicine and

Biology, 21 (9) : 1089-1 100, 1995.

[68] GE. Trahey, P.D. Freiburger, L.F. Nock, and D.C. Sullivan. in vivo rnea-

surements of ultrasonic beam distortion in the breast. liitmsound Imaging, 13:71-99, 1991.

[69] M. Tsuchidate, S. Katsumi, and T. Wagai. Influence of focused ultrasound on

tumours (the 6th report). Japun. Med. Ultmson., 4:44, 1966.

[70] D. H. Turnbull and F. S. Foster. Beam steering with pulsed two dimensional

transducer arrays. IEEE Tmnsactions on Ultmsonics, Ferroelectrics and Fre-

qvency Control, 38:320-333, 1991.

[71] S. Umemura and C. A. Cain. The sector-vortex phased array: acoustic field

synthesis for hyperthermia. IEEE IkBnsoctions on Ultmsonics, Ferroelectrics

and F'quency Control, 36:249-257, 1989.

1721 M. Urano and E. Douple, editors. The mtional for clinical trials in hyperther-

mia. Hyperthermia and Oncology. VSP BV, Utrecht, 1988.

[73] G. Vallancien, E. Chartier-Kastler, D. Harouni, D. Chopin, and J. Bougaran.

Focused extracorporeal pyrotherapy: experimental study and feasibility in man.

Seminars in h~logy, 11:7-9, 1993.

[74] G. Vallancien, M. Harouni, B. Guillonneau, B. Veillon, and Bougaran J. Ab-

lation of superficial bladder tumors with focused extracorporeal pyrotherapy.

UroIogy, 47(2) :2O4-207, 1996.

[75] Gerard van Leeuwen. Numerical nodelling of heat tmnsfer in hyperthermia.

PhD thesis, University of Utrecht, 1998.

[76] C. C. Vernon, J. W. Hand, S. B. Field, D. Machin, J. B. Whaley, J. van

der Zee, W. L. J. van Putten, G. C. van Roon, J. D. P. van Dijk, D. Gonzhles

Page 102: Theoretical Evaluation of Moderately Focused Spherical ... · Theoretical Evaluation of Moderately Focused Spherical Transducers and Multi-focus Acoustic ... Mihaela Pop, Lee Chin,

REFERENCES 89

Gonzales, F-F. Liu, P. Goodman, and M. Sherar. Radiotherapy with or without

hyperthermia in the treatment of superficial localized breast cancer - results

from five randomized controlled trials. International Jovrnal of Radiation,

Oncology, Ba'ology and Ph ysics, 35 (4) :731-44, 1996.

[77] A. Visioli, 1. H. Rivens, Gai1 ter Haar, A. Horwich, R. Huddart, E. Moskovic,

A. Padhani, and J. Glees. Preliminary results of a phase 1 dose escalation

clinical trial using focused ultrasound in the treatment of localised tumours.

European Journal of Ultmsound, 9: 11-18, 1999.

[78] 1. A. Vitkin, J. A. Moriarty, R. D. Peters, M. C. Kolios, A. S. Gladman, J. C.

Chen, R. S. Hinks, J. W. Hunt, B. C. Wilson, A. T. Easty, M. J. Bronskill,

W. Kucharczyk, M. D. Sherar, and R. M. Henkelman. Magnetic resonance imaging of temperature changes during interstitial microwave heating: a phan-

tom st ud y. Medical Ph ysics, 24:269-277, 1997.

[79] N. A. Watkin, G. A. ter Haar, and 1. Rivens. High-intensity focused ultra-

sound ablation of the kidney in a large animal model. Journal of Endourologg,

11 (3): 191-196, 1997.

[80] R. Yang, C. R. Reilly, F. J. Rescorla, P. R. Faught, N. T. Sanghvi, F. J.

fiy, Tg D. Franklin, L. Lumeng, and J. L. Grosfeld. High intensity focused ultrasound in the treatment of experimental liver cancer. Archives oj Surgery,

126:1002-1010,1991.

[81] R. Yang, N. T. Sanghvi, 3. R. Rederick, C. A. Galliani, F. J . Fry, S. L. Griffith,

and J . L. Grosfeld. Extracorporeal liver ablation using sonography-guided high-

intensity focused ultrasound. Investiqative Rad2ology, 27:796-803, 1992.

(821 R. Yang, N. T. Sanghvi, F. J. Resorla, K. K. Kopecky, and J. L. Grosfeld. Live

cancer ablation with extracorporeal high-intensity focused ultrasound. Euro- pean Umlogy, 23(suppl 1): 17-22, 1993.