Tests of scintillators

CsI, CsI(Na), BGO, NaI(Tl)

CsI(Tl)

Page 2: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

CsI(Na)

170 mm

15 x 15 mm9 x 9 mm

Page 3: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Energy of PMT 1 (small window)

Ene

rgy

of P

MT

2 (

big

win

dow

)

CsI(Na) scintillator with 2 PMT readout

Full absorption

Measurement without collimation 137Cs, 661 keV

Page 4: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Energy of PMT 1 (small window)

Ene

rgy

of P

MT

2 (

big

win

dow

)

CsI(Na) measured with a collimated source

Sum (e

nerg

y)

Ratio (position)

Page 5: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Energy spectrum (sum of 2 PMT’s energies)

Sum of 2 PMTs energies

Energy resolution 9.6% at 661 keVConstant for the position of the interaction

Page 6: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution (ratio of the PMT’s energies)

Light division between 2 PMTs

15 cm

10 cm

5 cm

Page 7: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution

0 2 4 6 8 10 12 14 160.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

1.1 Y =0.29835-0.00312 X+0.0029 X2

light

div

isio

n be

twee

n P

MT

s

position in scintillator, cm

Page 8: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution

0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.00

2

4

6

8

10

12

14

16

18po

sitio

n in

sci

ntill

ator

, cm

light division between PMTs

Position resolution 1.5 – 3 cm

Page 9: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

0.0 0.2 0.4 0.6 0.8 1.00

1

2

3

4

5ab

sorp

tion

dist

ance

, cm

energy, MeV

CsI

Ge

Mean absorption distance

Page 10: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Readout with one PMT

1000 1500 2000 2500 3000 3500 4000 4500 50000

100

200

300

400

coun

ts

energy of one PMT

5 cm

10 cm15 cm

Page 11: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

CsI, pure

Page 12: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Energy resolution

Resolution 17 – 19 %

Sum energy

Page 13: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution

Light division between 2 PMTs

15 cm

10 cm

5 cm

Page 14: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution

2 4 6 8 10 12 14 16 180.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0 Y =0.32407+0.16521 Xlig

ht d

ivis

ion

betw

een

PM

Ts

position, cm

Page 15: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Position resolution

0.5 1.0 1.5 2.0 2.5 3.0 3.5

2

4

6

8

10

12

14

16

18

20

posi

tion,

cm

light division between PMTs

Position resolution 2 – 8 cm

Page 16: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

BGO

Page 17: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Energy of PMT 1 (small window)

Ene

rgy

of P

MT

2 (

big

win

dow

)

BGO scintillator with 2 PMT readoutScintillator was illuminated by Cs137 source (661 keV) without a collimator

Strong nonuniformity of light yieldalong the crystal distance

Page 18: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

BGO collimated measurement

Full absorption

Page 19: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

BGO collimated measurement

Full absorption

Page 20: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

1)Light yield is strongly non uniform along the crystal distance

2)Light separates always in the same proportion between left and right sides

3)Energy resolution ≈18% (big window) ÷ 24% (small window)

BGO results

Page 21: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

CsI(Tl)

7% at 661 keV

2.5 cm2.5 cm

Page 22: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

to continue with..

Page 23: Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)