Istrumentazioni Sistemi Automatici S.r.l.

VIA BERGAMO 41 - 21020 TAINO (VA) - ITALY OFFICES TEL. +39.0331.956081 - FAX +39.0331.957091 LAB: TEL. +39.0331.956483 - E-MAIL [email protected] WEB www.isatest.com

DATE: 08/06/2006 DOC.MIE12110 REV.2

T2000 INTRODUCTORY GUIDE

Page 2: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 2 of 46

REVISIONS SUMMARY VISA N PAGE DATE 1 All 25/04/2006 Issued Lodi.

2 39 to 43 8/06/2006 Added the Calibration chapter Lodi

Page 3: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 3 of 46

SHORT FOREWORD ....................................................................................................................................................... 4

INTRODUCTION ............................................................................................................................................................ 5

1 TEST SET DESCRIPTION AND USE ....................................................................................................................... 7 1.1 THE FRONT PANEL ..................................................................................................................................................... 7 1.2 THE POP-UP MENU ..................................................................................................................................................... 9

1.2.1 Transformers selection ................................................................................................................................... 10 1.3 POWER-ON............................................................................................................................................................... 13 1.4 OUTPUTS DESCRIPTION AND HAZARDOUS SITUATIONS ............................................................................................ 14

1.4.1 Main current and voltage outputs .................................................................................................................. 14 1.4.2 HV voltage output ........................................................................................................................................... 15 1.4.3 Hazardous situation summary........................................................................................................................ 16

1.5 CURRENT GENERATION............................................................................................................................................ 17 1.6 OPTIONAL THERMAL PRINTER.................................................................................................................................. 18 1.7 PROTECTIONS .......................................................................................................................................................... 18

2 TRANSFORMERS TESTING FUNDAMENTALS ................................................................................................ 21 2.1 CURRENT TRANSFORMERS....................................................................................................................................... 21 2.2 REMANENCE............................................................................................................................................................ 21 2.3 HIGH CURRENT TESTS.............................................................................................................................................. 22 2.4 WINDING RESISTANCE TEST..................................................................................................................................... 22

3 THE HELL, IT DOESN’T WORK............................................................................................................................ 23 3.1 INTRODUCTION........................................................................................................................................................ 23 3.2 ERROR MESSAGES.................................................................................................................................................... 24 3.3 TROUBLE SHOOTING ................................................................................................................................................ 25 3.4 PHYSICAL DESCRIPTION........................................................................................................................................... 27 3.5 NO OUTPUT FROM CURRENT AND VOLTAGE............................................................................................................. 28 3.6 THE AC VOLTAGE MEASUREMENT IS NOT STABLE................................................................................................... 29 3.7 ERROR ON THE EXTERNAL 10 V OR 600 V MEASUREMENT ..................................................................................... 29 3.8 THE FAULT CANNOT BE FIXED.................................................................................................................................. 39

4 CALIBRATION .......................................................................................................................................................... 41 4.1 INTRODUCTION....................................................................................................................................................... 41 4.2 CALIBRATION PROCEDURE ..................................................................................................................................... 41 4.3 T/2000 OUTPUT CALIBRATION................................................................................................................................ 42

4.3.1 OUTPUT CURRENTS CALIBRATION ....................................................................................................... 42 4.3.2 OUTPUT VOLTAGES CALIBRATION ....................................................................................................... 43

4.4 EXTERNAL INPUT CURRENTS......................................................................................................................... 44 4.5 EXTERNAL INPUT VOLTAGES ........................................................................................................................ 45

APPENDIX 1 SPARE PARTS LIST ............................................................................................................................ 46

Page 4: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 4 of 46

SHORT FOREWORD Dear T2000 user, I often wondered why the user’s manual is not very much used, even if it includes valuable information. As me too I am a user of such manuals, the answer I have given myself is that valuable information are concealed somewhere in the thick thing, and I do not have time to waste to find it. So, either the manual is actually of help, or I ignore it. This is why I decided to split the T2000 manual in three: specification, with all performance details; application manual, with instructions about how to use it one its operation is understood; introductory guide, with the device description and basic information. The idea is that you may read once the device description, while you need application examples more than once; so, why not to split the manual in three? The only exception to this organization is next page: it reminds to be cautious when using these test sets. We are on the field since more than 50 years, and no injury has ever been reported; yet, your kids want you back home after work. Have a good work with T2000! Primo Lodi Q&A Manager

Page 5: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 5 of 46

INTRODUCTION T2000 allows performing all the test to be performed on current and voltage transformers. The following table lists the tests that can be performed on CT and VT.

N. TEST OF

TEST DESCRIPTION

1 CT Ratio, Voltage mode 2 CT Ratio, polarity and burden 3 CT Burden; secondary side 4 CT Excitation curve 5 CT Winding or burden resistance 6 CT Voltage withstand 7 CT Polarity by impulses 8 VT Ratio; polarity 9 VT Burden, secondary side 10 VT Ratio, electronic transformers 11 VT Voltage withstand 12 VT Secondary over-current protection 13 PT Ratio per TAP 14 PT Resistance of Tap Changer contacts 15 PT Tap Changer dynamic resistance test16 R Ground resistance and resistivity

With external options, T2000 can test: . With the High IDC module, up to 400 A: contact resistances, in the micro-Ohm range; . With the current booster: primary tests, up to 4000 A. The basic T2000 function is to generate current and voltages, as requested by the type of test to be performed, that is selected on the LCD screen by means of the multi-function knob. Test results are kept in memory, and can be transferred to a PC at a later time, along with settings. The instrument contains one generator with many outputs: High AC current; Low AC current; Low DC current; Current impulses; High AC voltage; Low AC voltage; Low DC voltage. All outputs are adjustable and metered on the large, graphic LCD display. With the multi-purpose knob and the LCD display it is possible to enter the MENU mode, that allows to set many functions, that make T2000 a very powerful testing device, with manual and semi-automatic testing capabilities, and with the possibility to transfer test results to a PC via the RS232 interface. These results can be recorded, displayed and analysed by the powerful TDMS software, which allows creating a data base of test results for all the plant. The ease of operation has been the first goal of T2000: this is why the LCD is graphic, and so large. With it, the dialogue in MENU mode is made easy. Besides, all T2000 outputs relevant to the selected test are continuously measured, and output values are displayed, with no extra effort to the operator.

Page 6: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 6 of 46

Additional features are: . Two meters, current and voltage, with independent inputs, and with High and Low inputs each, allow measuring CT or VT outputs or any other source; . The optional thermal printer gives the immediate printout of the CT saturation curve; . An auxiliary contact, that follows START and STOP inputs, allows simulating the circuit breaker. At power-on, there is a message asking to wait for the self diagnosis is completed. After this, the following menu is displayed.

The desired operation is accessed by the multi-function wheel. The following chapters explain how to operate with the above selections. NOTE: WINDOWS is a trademark of MICROSOFT inc.

Page 7: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 7 of 46

1 TEST SET DESCRIPTION AND USE In this chapter we describe the test set and inform about the operations that are common to relay or transformers testing.

1.1 THE FRONT PANEL

The following list includes the key components inside T2000. 1) Supply socket and main supply fuse, rated T5A, incorporated in the supply socket. 2) Power-on switch, with light. 4) Emergency push-button with lock-in. 5) Main outputs transformer. 6) Main outputs adjustment. 7) INTE ON-OFF printed circuit board; code YWA11410. 8) Earth socket. 11) Back panel board ; code YWA11413. 12) High output current measurement transformer. 13) Output sockets for high current output. 15) Microprocessor board, code YWA41300. 16) Converter board code YWA11401. 17) Front panel board; code YWA11411. 22) MENU control knob, with switch.

Page 8: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 8 of 46

23) Display. 25) Serial interface connector. 26) Main current outputs measuring transformer. 27) Main current output sockets. 28) High Voltage enable key. 29) Variable transformer home position switch. 44) 800 A range selection light. 45) 40 A or 25 V range selection light. 46) 10 A or 90 V range selection light. 49) ON + TIME light: current is generated and time metered until STOP is detected. 50) OFF light: no current generation. 51) ON light: current is generated. 52) OFF + TIME light: current is removed and time metered until STOP is detected. 53) Main AC voltage selection light. 54) Main DC current or current impulses selection light. 55 and 56) START and STOP push-buttons. 57) Push-button for the selection of main output. 58) HV voltage on light. 60) Main AC voltage safety sockets. 61) Main DC current or current impulses safety sockets. 67) External current meter safety sockets. 68) External voltage meter safety sockets. 69) 25 mA metering input protection fuse. 70) 10 V measurement input connector. 80) Transformer protection thermal switches. 81) Serial interface supply transformer. 82) Electronic power switch (SCR). 83) TRANFORMERS board for the auxiliary supplies, PWA11412. 84) HV output measurement transformer. 85) Fan. 86) Primary circuit breaker. 87) HV opening relay. 88) HV output current meter. 89) Input current meter. NOTE: the connector to the 10 V input (70) can be removed only acting on the connector body. DON’T PULL THE CABLE, AS YOU COULD DAMAGE IT!

Page 9: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 9 of 46

1.2 THE POP-UP MENU The following is the list of features that are menu selected. The menu is operated by means of the control knob marked MENU, that incorporates a switch. The menu is entered pressing the knob and selecting the item moving the knob. Once the item has bee found and programmed, pressing the arrow the menu moves back of one step, so that other programming can be performed; else, selecting ESC the menu returns to the main window. The first display is the following.

Next screens depend upon the type of selection: . For transformer tests, there is a dedicated screen for each type of test: only relevant parameters are displayed and recorded. . Other selections: Resistance, Results, Preferences, Test header. Any setting can be saved to and recalled from the memory. Up to 10 settings can be stored and recalled; setting no. 0 is the default one, and pops up at power-on. Settings are permanently stored in the memory; new settings can be written to the same address after confirmation. For normal mode operation it is possible to recall the standard setting, that cannot be modified. During the test, test results can be stored in the memory (up to 500 results may be stored). At the end of test, settings and test results can be transmitted to a PC provided with TDMS. The software allows saving test results, examining them and so on. The specification of TDMS is given in a separate document. When the PC is connected, settings can also be created and transferred into T2000 using TDMS.

Page 10: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 10 of 46

1.2.1 Transformers selection With this selection, the choice is: CT; VT; PT. After entering the final selection, the operator can input the relevant parameters, still by using the multi-function knob: turning it allows changing the parameter; pressing it makes it possible to go to next parameter. Once all parameters are set, it is possible to start the test and execute it. Test time is kept to the minimum to avoid the excess of heating. The following table summarizes all tests and the corresponding performances.

Page 11: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 11 of 46

TEST OF

TEST DESCRIPTION

INPUT DATA CONN. OUT CONN. IN MEASUREMENTS

CT N. 1

Ratio Voltage mode - I primary; - I secondary (nominal values) - Voltage output - Voltage input

High/Low V AC to CT secondary

CT primary to low or high Vin

1) High / Low VAC out; 2) Low V in; 3) Polarity; 4) Actual ratio; 5) Ratio error %; and excitation curve, if selected

CT N. 2

Ratio, polarity and burden

- I primary; - I secondary (nominal values); - Clamp Y/N; - Clamp ratio; - Voltage input.

High I AC to CT primary

CT secondary to high I in; (Low Iin with Clamp); CT secondary to Vin low or high.

1) High I AC out (primary); 2) I in (secondary); 3) Nominal ratio; 4) Actual ratio; 5) Ratio % error; 6) Polarity 7) VA rating 8) Power factor;

CT N. 3

Burden, secondary side

- IN secondary (nominal value); - Voltage input. - Current output

Low I AC to CT burden

CT burden to Vin

1) I out (secondary); 2) V out (secondary); 3) Phase V-I out (secondary); 4) Power factor; 5) VA rating;

CT N. 4

Excitation curve - Voltage output - I nom secondary - VA rating - Accuracy class - Overload - Internal loss - Standard (IEC, ANSI)

High V AC to CT secondary

1) High V AC out; 2) I out of High V AC; 3) Iout-Vout curve; 4) Current at knee, IKm; 5) Voltage at knee, VKm

CT N. 5

Winding or burden resistance

- Temperature compensation Y/N - Ambient and target temperatures

Low I DC to CT burden or winding

CT burden to Vin

1) Low I DC out; 2) V of lowI DC out; 3) Resistance; 4) Compensated resistance

CT N. 6

Voltage withstand - Max High V AC ; - Max I test - Tmax

High V AC to: Primary and secondary;

1) High V AC out; 2) I out of High V AC ; 3) Elapsed Time

CT N. 7

Polarity by impulses Low IDC to CT primary

CT sec. to Iin 1) I DC out; 2) I secondary; 3) Polarity

Page 12: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 12 of 46

TEST OF

TEST DESCRIPTION

INPUT DATA CONN. OUT CONN. IN MEASUREMENTS

VT N. 8

Ratio; polarity V primary in kV; V secondary; Connection LL, LN for primary and secondary (nominal values)

High V AC to VT primary

VT secondary to V in

1) High VAC (primary) 2) V in (secondary); 3) Phase shift ; 4) Actual ratio; 5) Ratio error %; 6) Polarity

VT N. 9

Burden, secondary side

- V secondary (nominal value) - Connection LL, LN - Voltage output - Voltage input

Low V AC to VT burden

VT burden to V in (if enabled)

1) V out (secondary); 2) I out (secondary); 3) Phase V-I ; 4) Power factor; 5) VA rating

VT N. 10

Electronic Voltage Transformers

- V primary; - V secondary; - Connection LL, LN for primary and secondary (nominal values)

High V AC to VT primary

VT secondary to V in

1) High VAC (primary) 2) V in (secondary); 3) Actual ratio; 4) Ratio error %; 5) Polarity

VT N. 11

Voltage withstand - Max High V AC; - Max I test; - Test duration.

High V AC to Primary and secondary;

1) High V AC out; 2) I out of High V AC ; 3) Elapsed Time

VT N. 12

Over-current protection

- I Trip - Output current

Low I AC to VT protection

1) I out (secondary) 2) I trip

PT N. 13

Ratio per Tap - V primary in kV; - V secondary; - Connection LL, LN for primary and secondary

High V AC to VT primary

VT secondary to V in

1) High V AC out; 2) I of High V AC; 3) Phase V-I 4) V in; 5) Actual ratio; 6) Ratio error %.

PT N. 14

Resistance of Tap Changer contacts

- Temperature compensation Y/N - Ambient and target temperatures

Low I DC V in 1) I DC out; 2) V of IDC out; 3) Resistance; 4)Compensated resistance

PT N. 15

Dynamic tap changer test

- Time base - trigger level

Low I DC V in 1) I DC out; 2) V of IDC out; 3) Resistance; 4) Resistance waveform

R Grid N. 16

Resistance or resistivity of earthing grid

- Output voltage - Input voltage

Low V AC to auxiliary spike

V input from measurement spike(s)

1) Output voltage 2) Output current 3) Input voltage 4) Ground resistance or Ground resistivity

Page 13: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 13 of 46

1.3 POWER-ON At first, be sure that the main control knob (6) is turned (rotated) to the zero position (complete counter-clockwise). The reason is that the current generator is actually a high current voltage generator. If the output is connected to the load (typically low impedance), as soon as the test is started, a very high current can circulate in the circuit. Next, connect the mains supply cable to the instrument and then to the supply. The power supply must be made of three wires: phase, neutral and ground. It is also intended that the neutral will have a low voltage with respect to ground (20 V maximum). If you need an extension cable, use a cable where the copper cross section of 6 sq. mm. minimum: this serves to prevent heating and voltage drops, that would cause also a reduction of the performances. Power-on T2000: the test set checks if it is connected to the ground; if not, it alerts the operator with the following message.

All operations will be inhibited until the ground is available; power-off, connect to the ground and then power-on again. Usually the grounding connection is provided by the power supply cord; if the ground connection is not available at the power outlet, connect T2000 to ground using the yellow/green cable with crocodile provided. The connection to water pipes or to a big metal frame will do. NOTE: this protection checks the voltage between phases and ground; at least one of them (the neutral) must have a low voltage with respect to the ground. This means that T2000 cannot be powered by two phases of a 220 V (or 110 V) three-phase supply (127 V or 63.5 V phase to neutral). If you are in this situation, and you are sure that the ground is there, you can over-ride the alarm as follows: . Press the > key; . The alarm will show up again; . Press the > key four times in all: the message will disappear, and you can continue. T2000 diagnostic tests also the voltage supply value: if it exceeds the limits of ± 20%, one of the following messages will be displayed.

If the ground is connected, the following message is displayed.

Page 14: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 14 of 46

During this time, a diagnostic sequence controls: . Key microprocessor board components; . Auxiliary supply voltages. If something is wrong, the operator is alerted by one of the following messages.

At the end of it, default selections are active; T2000 is in the OFF state. If some selection keys are pressed at power-on, the following message warns the operator to release them.

After power-on, the following control lights turn on (default situation): . TEST OFF (50); . MENU SELECTION: 40 A SOCKET (45). To operate the test set it is necessary to perform the pop-up menu selections. During operation, if something goes wrong the test set protections intervene; usually, the operator is alerted by a message.

1.4 OUTPUTS DESCRIPTION AND HAZARDOUS SITUATIONS

1.4.1 Main current and voltage outputs The first thing to be understood is that all main outputs are generated by a sole transformer, and are adjusted by knob (6); as a consequence: . It is impossible to adjust separately the AC current and the AC voltage; . The output selection does not open the unselected outputs; . Only one connection should be made to main outputs. In particular, if there are two connections to output sockets (27) and (13), the displayed current value is meaningless, currents will divide as a function of the burden. The situation of zeros is the following: . There are three independent zero: 800 A; 3000 V; all other main outputs; . The zero of low current outputs is in common with the 250 V AC output, but it is not directly connected to the zero of the DC current output, as there is a rectifying bridge between them. DO NOT CONNECT THESE ZERO BETWEEN THEM: this would cause a short-circuit on one of the rectifying diodes. The only exception to the above is the HV output: it is not connected if it is not selected. However, when HV is selected, if a load is connected to the main AC current or AC voltage or DC current, as HV is adjusted these outputs will generate in a not controlled way. Coming to the characteristics of main supplies, the following applies.

Page 15: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 15 of 46

. The quality of the waveform is depending upon the quality of the voltage supply. Please do not blame T2000 for distorted waveforms! . Current outputs are generated by a voltage transformer with low voltage and high current ratings. As a consequence, the current is a function of the load: you cannot pre-set the current value by adjusting knob (6). These outputs are not hazardous, as voltage is low; however, current capability is high; so, take care not to touch both conductors with a ring, watch or metal tool: it would heat-up very quickly. . AC voltage can be dangerous, as the range is 250 V: do not touch both ends. . DC current is low voltage and current limited; so, it is not dangerous. . Do not generate high power for a long while. If you are overloading the output, the test set will cut it off after a while, according to the table in the cover; in hot weather, the thermal protection could intervene, cutting the operation for a long while. Besides, repeating tests in overload conditions reduces the life of the components. . If in a plant being build up the AC supply is missing, sometimes a stand-by generator or a DC to AC converter can be used. In this instance, the quality of the supply can be very poor, to the point that it could be impossible to perform tests. In particular, if a DC to AC converter is used, it must be generating a real sinusoidal waveform: cheaper converters generate a trapezoid shaped waveform. This can also be the problem if a filter is used on the mains.

1.4.2 HV voltage output Coming now to the HV output, IT IS VERY DANGEROUS; so, please take the maximum care when using it. . Never leave connection cables connected, even if the output is unused. T2000 HV connectors are suited for HV isolation; instead, connection cables have at their ends some accessible clip-on crocodiles. It is true that HV is disconnected when not used; however, there is a relay that performs this, and if the relay fails, you have HV on your leads while performing other tests. . Connect the cable to the test device before enabling the HV test. Here again, a mistake is always possible. DO NOT CONNECT THE TEST DEVICE WHEN OUTPUTS ARE ACTIVE! . If possible, do not connect to ground any of HV voltage outputs. The high voltage generator has both sockets not connected to ground. In this situation, the accidental contact of the operator with either output would be of no consequence; this means that the operation would be safer. The connection of one side of the output to ground reduces the safety margin. The contact to ground of the other lead would cause a short circuit to the HV output, with immediate disconnection of the test set. Much worse, the accidental contact of the operator with the other lead would submit him to the high voltage, WITH NO PROTECTION FROM THE TEST SET. This is why an isolated operation is STRONGLY RECOMMENDED WHENEVER POSSIBLE. The recommendation does not apply to the isolation test, where one of the two leads must be grounded. During this test the above recommendations apply, so, take special care.

Page 16: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 16 of 46

The HV is generated only while the START button (56) is pressed: this is an additional safety feature. The operator will have both hands occupied: one for the knob (6), the other for the button (56); however, this is much safer than starting the HV generation, and then forgetting it in the ON state. When the connection is performed, connect the crocodile first, then connect the cable to the HV connectors. The HV connector is made of a moving body, a threaded part on which is located a nut: see sketch. You cannot remove the connection pulling the cable: it is necessary to act on the body. If, for safety reasons, you do not want that the connection is removed, it is possible to lock the moving body with the nut. After this, the connector cannot be removed, until the nut is unscrewed to its former position.

BODY

THREAD

NUT

1.4.3 Hazardous situation summary The following table summarizes the most hazardous situations discussed above. Please consider this list, and check the situation in case of doubt. SITUATION CAUSE OF RISK CONTROL Long generation of outputs Possible danger of over-heating components, specially

with high ambient temperature Check burden and duration

HV cables connected while not used

Risk to the user if the isolation relay fails No HV cable

High AC voltage output connected to ground

In this situation, if the operator touches the other terminal, he would be subject to a mortal voltage: it is necessary to be extremely cautious.

Ground connection

Connection while the output is active

In this situation there is voltage on the connection cables: there is the risk of being hit touching them.

The test set must be OFF

Disconnection while the output is active

The inductive load would cause a very high voltage spike. The test set must be OFF

Connection to a live wire The connection can be dangerous to the test set. Test before connecting

Page 17: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 17 of 46

1.5 CURRENT GENERATION If the following current limits and time duration of main current outputs are trespassed, the generation is interrupted, and the operator is warned by an alarm message. HIGH AC CURRENT CURRENT OUTPUT

A

OUTPUT POWER

VA

ON TIME

s

OFF TIME min

100 600 STEADY - 150 800 15 min 30 200 1000 4 min 15 400 1600 15 5 600 2000 5 3 800 2000 1 2

With the T2000E model, the performance is the following. CURRENT

A MAX

POWER VA

MAX ON TIME

s

OFF TIMEmin

MAX TOTAL LOAD mOhm

LOAD W/O

CABLES mOhm

POWER W/O

CABLES VA

100 850 CONT. - 86 80 800 150 1200 15 min 30 55 49 1100 200 1550 4 min 15 39 33 1320 300 2050 15 5 23 17 1530 400 2400 15 5 15 9 1440 600 2600 5 3 7 1 360 800 2100 1 2 3 - -

LOW AC CURRENT RANGE

A AC CURRENT OUTPUT

A

OUTPUT POWER

VA

LOAD TIME

s

RECOVERY TIME min

12 300 STEADY - 18 15 min 30 24 4 min 15 36 800 15 5 48 5 3

40

60 1000 1 2 5 400 STEADY -

7.5 15 min 30 10 800 60 15 15 30 10

10

20 1000 15 5 This generator serves for the test of VT protections, where low current is necessary. The procedure is the following.

Page 18: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 18 of 46

. At first, be sure that the main control knob (6) is turned (rotated) to the zero position (complete counter-clockwise). . Power-on T2000. . Select by the push-button (57) the measurement on the desired output sockets (13), according to the maximum current to be generated: the LED turns on; the AC voltage value is displayed. . Connect the device to be tested to sockets (13). Consider that for tests of 40 A up it is necessary to connect the relay by a wire having at least a cross section of 10 sq. mm; for lower currents, a cross section of 2.5 sq. mm can be used. . Press ON and adjust the output current to the desired value with knob (6). . After you have started the test, if the burden is a short circuit made of a short cable, you measure at zero knob position a current that usually is less than 3% of the range. This value does not influence at all the measurement of the current you are generating: it is not an error of the measurement instrument. . There is also another possible problem: the desired current cannot be reached. If it is impossible to reach the desired value, this is because the burden is too high. Very often the problem comes from connection wires; so, to perform the test it is necessary either to shorten them, or to increase the cross section (or both). Note that the test starts and stops as the current passes the zero.

1.6 OPTIONAL THERMAL PRINTER The optional thermal printer allows to print the selected test result. The connection is performed via the RS232 interface, using the serial cable provided. To get the printout, select the desired result and reach for the printer icon on the screen by means of the multifunction selector: the screen is printed.

1.7 PROTECTIONS - If the test set is not connected to the ground, the test set does not allow for power generation, and warns the operator with the above diagnostic message. - If the power supply value is not within 20% of the nominal, the test set does not allow for power generation, and warns the operator with the above diagnostic message. - Fuse on the mains supply: if open, the power-on switch does not turn on. - At power-on, a diagnostic sequence controls: . Key microprocessor board components; . Auxiliary supply voltages. If something is wrong, the operator is alerted by the above diagnostic messages. - Emergency pushbutton: if pressed, main transformer primary circuit is open; all main outputs are removed. In this instance, the following message is displayed.

Page 19: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 19 of 46

After pressed, the pushbutton stays in the emergency position until released by rotating it; as it is, the message disappears. - The high voltage output has the following protections: . HV outputs are disconnected when not used; . It is possible to access HV generation only in some tests; otherwise, the access is inhibited; . Confirmation key: if not turned ON, the HV output is not generated. If the key is turned ON while T2000 is generating other outputs, the generation is interrupted, and the following message is displayed:

. When a suitable test is selected, the test set informs the operator when the HV key should be turned ON:

. It is impossible to start generating the HV unless the adjustment knob is at zero; in this instance, the following message is displayed.

. The HV output is generated only until the START (H.V.) button is pressed: releasing the button interrupts immediately the generation. The test has to be started over again. Note that this feature applies only to HV generation: for all other generations, a single touch on START is enough. - The transformer under test cannot be left with some remaining magnetization. On some tests (like the saturation curve test), if the test is broken during the generation, and the output has not been adjusted to zero prior to stop, the following message is displayed.

In this instance, the operator has to start again the test, slowly reach the same test value and slowly decrease to zero. - Thermal (NTC) sensor on the main and auxiliary transformers. In case of over-temperature, the following message is displayed. It is necessary to power-off and wait until T2000 cools down.

- Thermal sensor on the T2000 inside. If T2000 is too hot, the following message is displayed. It is necessary to power-off and wait until T2000 cools down.

- Thermal sensor on the T2000 inside. If T2000 is too cold, the following message is displayed. The low temperature affects the display operation.

- If maximum time duration of power transformer generators are trespassed (see the table in the Specifications chapter), the generation is interrupted, and the operator is warned by the following message.

Page 20: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 20 of 46

Pressing the knob, T2000 displays the pause time duration:

- The main AC voltage source is protected against overloads. In this instance, the following message is displayed.

- The main DC current source is protected against over-voltages, that could be caused by the load inductance. In addition, the output is automatically kept to zero as test stops, so that any residual energy on the external load is discharged: during this time, the following message is displayed.

- The 20 mA measurement input is protected by fuse (69), rated F63mA, against wrong connections.

Page 21: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 21 of 46

2 TRANSFORMERS TESTING FUNDAMENTALS Many of tests explained in this manual involve high voltage and/or current, and should be performed by experienced personnel, familiar with the peculiarity or dangers that may exist during test setup and procedures. While some dangers have been pointed out, it is impractical to list all necessary precautions.

2.1 CURRENT TRANSFORMERS Most used current transformers have no primary winding, as the primary is made of the primary conductor passing once through the center of a toroidal core to perform its function. Since the secondary winding is uniformly distributed about the core and only a single primary turn is used, essentially all magnetic flux is linked to the secondary winding, and there is essentially no leakage flux; this means also that the leakage reactance is negligible. CT’s of this type are classified C type per ANSI/IEEE C57.13-1978, indicating that ratio correction at any current can be calculated once the burden, the secondary winding resistance and the excitation characteristics are known. With this type of CT’s the same standard states that all tapped sections shall be so arranged that the ratio can be calculated from the excitation characteristics. Measuring the burden of the CT is important because the data allows verifying the accuracy of the transformer itself. The internal burden is the resistance of internal windings plus connection cables, and can be measured. Also the external burden can be measured, both as resistance and as impedance. For the correct measurement of the external burden, please consider the following notes. . All relays and other devices should be connected before the measurement, on the correct tap. . If a parallel CT exists, it should be disconnected. . The burden of electromechanical relays can be quite high, particularly with ground protection relays.

2.2 REMANENCE All transformers with iron cores have the problem of residual magnetization. The core materials are subject to hysteresis: if a current through the transformer is interrupted, the flux density does not become zero while the current is zero. If the current interrupted is high, or if it is a DC current, remanence can be substantial, and can even be higher of the saturation knee. When the current is next energized, the flux change will start from the remanent value, and if the flux change is in the direction of the remanence, a large part of the cycle may find the transformer being saturated. When this occurs, much of the primary current is required for excitation, while the secondary current is significantly reduced, and distorted on alternate half cycles. The condition can be corrected by demagnetizing the current transformer. This is performed by selecting the saturation curve test, and applying to the secondary winding a voltage that should be starting from zero and then slowly increased until the saturation knee is passed; then, slowly decreasing (10 seconds) the voltage down to zero. The amount of saturation depends upon the value of the interrupted current: the higher the value the higher the saturation degree.

Page 22: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 22 of 46

2.3 HIGH CURRENT TESTS The high current generator is actually a low voltage, high current voltage generator. As a consequence, the voltage output is safe: touching conductors during the test would not be hazardous. However, there are two hazards to be considered: . Touching the high current leads with a metal object, such as the ring, watch, chain, would cause an extremely high current to the object, that would quickly increase its temperature: this would cause severe burns to the operator. . The high current generates high heat, specially at clamps: do not touch the cable without gloves, after a long test or an high current test.

2.4 WINDING RESISTANCE TEST Usually transformers are used at 50 Hz, so their inductance is very low, as it is due to magnetic field losses. When the winding resistance is tested, we apply a DC current to a load that is heavily inductive, in the order of hundred or thousand of Henry. The T2000 generator is actually a voltage generator, with a resistor in series and one in parallel to the load. With this type of load, the time constant for the current to reach its maximum is in the order of seconds, possibly minutes. For the sake of measuring the resistance it is not important that the current is perfectly stable; however, this gives the idea of the fact that if the adjustment knob is moved quickly, the current does not follow, and continues to grow until its asymptotic value. All this is when the current is increased: more important to be considered is what happens when the current has to be decreased to zero. When the resistance is measured, and the test should be stopped, T2000 does not accept the stop command, and awaits for the knob to be adjusted to zero. The movement should be slow enough to allow for the magnetic energy stored into the transformer to be dissipated into the parallel resistor; when the knob is at zero, T2000 accepts the STOP command, but keeps on measuring the current, and shows a message until it is decreased to zero. The load can be disconnected only after the message is disappeared; however, for added safety, you can short circuit the winding prior to remove the connections. Do not remove the connection while the test is on, or until the message is not disappeared: interrupting the circuit with a non dissipated magnetic energy would generate an over-voltage on the connection lead in your hand, that can be in the order of thousand of Volts: this is hazardous for you and can damage T2000.

Page 23: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 23 of 46

3 THE HELL, IT DOESN’T WORK

3.1 INTRODUCTION Sometimes, when my ears whistle, I wonder if it is because of some of my customers being angry at us because the test set doesn’t work: according to Murphy’s law, when it was most necessary. We at ISA do our best efforts to filter the so-called infant mortality of electronic components prior to delivery of all our test sets; and this after extensive testing of prototypes and pre-production units. Yet, sometimes faults occur, because everything dies, including electronic components; so, please, before shooting at us, see if the following instructions can serve you to fix the problem. If not, e-mail us the problem, not forgetting to mention the unit’s serial number: our business is to minimize your downtime. My e-mail address is: [email protected] Last, our experience is that our test sets withstand very heavy duty cycles for long wiles, if correctly used; most problems arise because of: . Voltage or current neutral is connected to earth; . Severe spikes on the mains (spikes are not always so kind to respect standard limits); . Transit, with the associated drops and thermal cycling; . Errors, such as the connection to live wires, or generating current on a short circuit with the handle completely turned to the maximum current. Please mention in your e-mail how did the fault occur: this serves us for our continuous improvement program.

Page 24: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 24 of 46

3.2 ERROR MESSAGES The test set performs a number of tests at power-on and during the generation. The following table lists all the messages, and the corrective action. A) TESTS AT POWER-ON (and runtime) ERROR MESSAGE CONSEQUENCE CORRECTIVE ACTION Test set not connected to ground; power off and correct

The test cannot proceed Connect the test set to ground using the dedicated socket and ground cable.

Power supply too low! Power off and correct

The message is repeated twice; then the test can proceed

Correct the supply. A slightly reduced voltage causes the corresponding reduction in maximum current and power; a very low voltage can damage the test set.

Power supply too high! Power off and correct

The test cannot proceed Correct the supply, as there is the risk to damage the test set.

Diagnostic error + 15 V auxiliary supply

If the error is not confirmed it is possible to continue; else, the test cannot proceed (NOTE 1)

Try some power on – off; if persists, three steps: . Check the fuses; . Try to replace the CONV board; . Return the test set.

Diagnostic error - 15 V auxiliary supply

If the error is not confirmed it is possible to continue; else, the test cannot proceed (NOTE 1)

Try some power on – off; if persists, three steps: . Check the fuses; . Try to replace the CONV board; . Return the test set.

Diagnostic error NV memory access

If the error is not confirmed it is possible to continue; else, the test cannot proceed (NOTE 1)

Fault in the MICR board: replace it

Test results corrupted It is necessary to erase corrupted data to continue

Using the menu, erase all results

Invalid test result data It is necessary to erase invalid data to continue

Using the menu, erase all results

B) RUNTIME TESTS ERROR MESSAGE CONSEQUENCE CORRECTIVE ACTION The test lasted too long. Test set is pausing

Tests are blocked for the specified duration

Wait until the message disappears

Pause. Residual time XXX seconds

If you try to restart before the above message disappears, the test set informs you about the residual time

Wait until the message disappears

Main V AC overload: reduce voltage or burden

Generation is stopped Reduce voltage or burden of main AC voltage

Main IAC output connection error

Generation is stopped The test set is sensing two burdens on main IAC and VAC: remove one of them

Page 25: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 25 of 46

Auxiliary AC voltage overloaded: reduce voltage or burden

Generation is stopped Reduce voltage or burden of the auxiliary voltage

Transformer over temperature: cool down lasts 15 minutes

Generation is stopped (NOTE 1)

The main transformer temperature is too high because of heavy loads or very long test duration: wait until it cools down

T/3000 is too hot: power off for 15 minutes

Generation is stopped (NOTE 1)

The test set temperature is too high because of heavy loads or very long test duration: wait until it cools down

T/3000 is too cold: let it warm-up

Generation is stopped (NOTE 1)

The test set temperature is too low: let it powered-on until it heats up

10 V external input too high!

Measurement is stopped Connect the input to the 600 V sockets

600 V external input too high!

Measurement is stopped The input cannot exceed 600 V AC.

NOTE 1: Sometimes the alarm is false, i.e. the fault is in the diagnostic circuits (they also can fail), and the test set is OK. In this case, you can proceed by pressing the buttons <, > and the control wheel at power on. TAKE CARE, because this operation skips ALL faults, and the test set is no more providing alarms in case of too long test duration during high current generation: MAKE THESE TESTS AS SHORT AS POSSIBLE!

3.3 TROUBLE SHOOTING In case of malfunctioning, the following table lists the more likely types of faults, and the correction. Following paragraphs give more details about the corrections. SYMPTOM POSSIBLE CAUSE CORRECTION Problems after transport Heavy shock Open T2000 and check for

loose boards or connections At power-on does not turn on Mains supply fuse open The fuse is located in the power

supply plug, in the small drawer. Replace it.

The variable transformer gets very hot as soon as the test set is powered-on

The INTE board is faulty Replace INTE

After the general over-temperature alarm message it does not turn on

The power transformer was over-heated

Wait at least 30 minutes before trying again: it needs time to cool down.

It is impossible to measure on the 20 mA range

The input was overloaded; the PTC protection is not restored

Wait 15 minutes before trying again.

No measurement of I main Broken connection See text Measurements are completely false

The microprocessor has lost its correction factors

Go to the calibration procedure and repeat it.

I cannot continue the test with the High IDC module

The thermal sensor is intervened.

Await for the module to cool down.

Page 26: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 26 of 46

Page 27: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 27 of 46

3.4 PHYSICAL DESCRIPTION T2000 components are accessed by removing the five screws that are located below it, and lifting the central part by the handles. There is a main board mounted just below the front panel, and a number of components mounted on the bottom panel. The arrangement of the main components mounted on the bottom panel is shown in the figure below.

85 30

86 5

15 7 16 83

87

On the upper left corner are mounted the control cards; their location is shown in the drawing below. The first three cards are kept in place by the piece of aluminium sheet on the right. If a board is failing, it is necessary to remove the two screws. The last board, transformers, is kept in place by the screw on the front, as shown: it must be removed for card replacement.

Page 28: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 28 of 46

FRONT PANEL

15

14 SCREWS

7

SCREW 83

SERIAL i/f TRANSFO

On the edge of the CONV board are mounted a number of ring-shaped test points. They carry the key auxiliary supplies, which operate the test set. The arrangement of the test points is the following.

-15

+15

+50

+5

+5

+5

0 V

0 If any of these voltages problem of blown fuse, o

3.5 NO OUTPUT FROM This is the case when yovoltage outputs. In this sin the boards pack. The rof T2000 and board repla

+1 2

VV V V V V

is missing, the tesr repair the auxilia

CURRENT AND

u can start the tesituation the fault ieplacement is vercement.

V

t sery

VO

t, bs loy fa

V V

t cannot operate: it is necessary to find out if it is a voltage circuit.

LTAGE

ut no output is generated from the main current and cated in the INTE ON-OFF board (7), that is located st: just follow the instructions above for the opening

Page 29: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 29 of 46

3.6 THE AC VOLTAGE MEASUREMENT IS NOT STABLE First of all, check that there is no unplugged connector or broken wire. Next, it is necessary to understand if the problem is in the measurement or if the output is actually unstable. To this purpose, connect the output to an AC voltage meter and verify the reading. If it is unstable as well, the problem is in the amplifier; otherwise, it is in the measurement circuit. In the first instance, amplifier, we have to verify if the problem comes from the adjustment potentiometer. To this purpose, proceed as follows: . Go to the menu and select PREFAULT + FAULT; . In the PREFAULT mode, the output voltage is adjusted by the multi-function knob, and not by the potentiometer: please verify the output. If it is stable, the problem is located in the adjustment knob. If it is unstable, it can be located on the AP/50 amplifier, or on the CONV board. The problem when replacing CONV is that, as it hosts the measurement circuits, after replacing it, it is necessary to repeat the calibration. It is up to you if you want to receive the replacement boards or to deliver the test set to us. In the second instance, the fault is in the measurement, we have to replace the CONV board, with the problem explained above.

3.7 ERROR ON THE EXTERNAL 10 V OR 600 V MEASUREMENT If you doubt that the voltage measurement is wrong, proceed as follows: . Select the test of the CT ratio, voltage mode; . Select the 90 V output; . Select a ratio 1:1; . Connect the 90 V output to the voltage measurement input to be tested; . Start the test, and compare the measured output voltage to the measured input: they should be the same value, with a maximum deviation of 1%; . If measurements are different, use an external voltmeter to decide which is the faulty one. If the voltage measurement is faulty, the repair involves four steps: . Dismounting the front panel and the main board from the rest of the test set; . Testing the circuit and replacing the faulty components; . Re-assembling the test set; . Final test. 1) Dismounting the front panel and the main board from the rest of the test set. The process is made of: . Unplugging all the connectors from the upper part to the lower part; . Unsoldering the connection wires; . Unscrewing the wires to the VARIAC; . Unscrewing the two high current cables from their sockets. The following pictures have been taken on a T/3000, so, on T/2000 two connectors and two boards (auxiliary AC and DC supplies) are missing. On the following picture is shown the lower T/3000 side.

Page 30: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 30 of 46

Connectors to be removed are all those going to the main board and coming from the lower part of the instrument, and the three screw connections to the VARIAC: please mark them, in order to avoid errors when the cabling is restored. The fast-on connections to the front power supply plug are not to be removed. Mark connectors with the same number of pins: the length should avoid confusion, but it is safer to have an additional reference. The following picture shows the detail of two more wires to disconnect.

Page 31: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 31 of 46

The two black wires, connected to the HV relay with FAST-ON, are to be unfastened. These wires are connected to the HV sockets; one of them enters a CT toroidal transformer that measures the HV output current. The CT transformer is tied to the lower part; so, it is necessary to remove the wire through it. Please note the path of these wires, as it is to be repeated EXACTLY THE SAME as the test set is re-assembled: this path is critical because of inductions from the HV voltage. Next picture shows the connectors on the right side.

Page 32: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 32 of 46

It is necessary to disconnect the connectors, as shown, and to unsolder all the wires shown. About this, PATE CARE TO MARK THEM, IN ORDER TO ENSURE THE SAME SEQUENCE WHEN THEY WILL BE SOLDERED BACK. Next picture shows the upper side of the test set.

Page 33: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 33 of 46

Arrows show the two high current wires, to be unscrewed, and the connectors to remove. Next picture shows the detail of three flat cable connections to remove: take care as they are hard to remove and fit, but the body is a bit fragile.

Page 34: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 34 of 46

Last, as all connections have been removed, remove the front panel and mother board from the rest, by unscrewing the screws shown in the next picture.

Page 35: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 35 of 46

After removing these boards, the support legs will remain attached to the front panel, so that it can be moved without damaging the components. Take care when you unscrew, in case some connection was not removed. 2) Testing the circuit and replacing the faulty components First of all, refer to the attached circuit schematic, that shows the failing circuit. The first thing is to verify if there is some component burnt or physically damaged. The circuit is located on the mother board, near to the measurement sockets: please refer to the following mounting drawings. In particular, we suspect that the failing component is U68: have a close look at it. There is a possibility that the gas discharger D49 is in short circuit, or that zeener diodes TZ9 and TZ10 are in short circuit: please verify.

Page 36: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 36 of 46

Components of the voltage measurement circuit.

Page 37: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 37 of 46

Power supply of the voltage and current measurement circuits.

Page 38: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 38 of 46

If there is some damaged component, try to replace it; test also if fuses marked F8 and F9 are open; in case, replace them. After this, it is necessary to verify if the measurement circuit is operational. To this purpose, it is necessary to power-on the test set; this can be done as follows: . Connect to the power supply the connectors J10 (lower side of the test set) and J18 (see mounting schematic above), using the two extension cables provided; . Remove the three flat cables from the lower part, replace them with the three longer cables provided, and connect them to the upper part. Power-on: the display shows the starting message, as usual. Select a test that involves the external voltage measurement, and then the two voltage measurement inputs: they should display zero. . If the measurement is zero, apply the voltage to one input, and verify that the measurement is correct; then repeat on the other. If everything is OK, the repair is finished. . If the measurement is not zero, it is necessary to perform other tests. First thing, we should verify that the auxiliary DC supplies, + 5 V_MV, + 15 V_MV and – 15 V_MV, are correct. To this purpose, refer to the test point TP19, shown above: it is the zero of these voltages. Then, measure that the voltage on fuses F8 and F9 is 17.5 V AC nominal: this ensures that the transformer is OK.

. If one or both voltages are zero, temporarily remove fuses F8 and F9, as necessary, power-on and measure again the voltage.

IF ONE OR BOTH OUTPUTS on the transformer side ARE ZERO, THEN THE FAULT COMES FROM THE TRANSFORMER ITSELF. The recovery of such a fault implies removing the transformer from the main board: in this case, please inform us immediately.

If voltages are OK on the transformer side, then there is a short-circuit on some component. In this instance, replace all the IC’s of the circuit, and try to power-on again: the short-circuit should be disappeared. If not, it is necessary to look at the schematic, and try removing all components that could cause the short-circuit.

. If AC voltages are OK, please verify the DC voltages between TP19 and: U69 pin 4 (+15 V); U69 pin 11 (-15 V); U61 pin 1 (+ 5 V). If some is missing, either there is a short circuit (in this instance remove all IC’s we have delivered, and try again), or the corresponding series regulator is broken (very unlikely). Once AC and DC voltages are OK, if the measurement is still wrong, replace all IC’s we have delivered; if still the measurement is wrong, it is necessary to follow the circuit schematic to locate the fault. Key components are: . PGA204AU, U68: it is a programmable gain amplifier. The gain is set by the microprocessor, as a function of the input amplitude; . OPA2477, U69, is a dual 477 op amp; . MAX312CSE, U70, is a four way analog switch; . HCPL7800, U61, is a fixed gain isolation amplifier. 3) Re-assembling the test set The re-assembling will be performed the other way round with respect to the dismounting; so: . Remove the two extension cables;

Page 39: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 39 of 46

. Replace the long flat cables with the original ones;

. Screw in place the front panel;

. Solder and screw wires;

. Fit back the connectors;

. Re-locate the HV cable, AS IT WAS BEFORE. 4) Final test Once the test set is restored, power-on again and verify that all outputs are correctly operational.

3.8 THE FAULT CANNOT BE FIXED If the fault is too hard to be fixed, you have to deliver it back to your agent. We have encountered problems caused by a poor packing of instruments that have been delivered us for calibration or repair. In order to avoid such inconveniences, please apply the following procedure. First of all, compile the following form, and attach it to the instrument. Please do not forget to compile it. With the instrument should come the mains supply cable, and the serial interface cable. The user’s manual originally delivered with the test set is not necessary. Cover the instrument itself with a polyester film, in order to protect it against dust and foam. The instrument should be protected by anti-shock foam having a minimum thickness of 5 cm ON ALL SIDES. Use a new carton box as a container. On the box apply the UP and the FRAGILE labels. In the box the instrument will be placed horizontal or standing; not upside down. If the set is heavier than 20 kg it is better to use also a pallet: this ensures that the box will not be packed upside down. Last but not least, do not declare an high value for customs: this expedites clearance of the good and lowers fees.

Page 40: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 40 of 46

INSTRUMENT RETURN FORM

DATE ____________ AGENT _________________ COUNTRY ___________________ TYPE OF INSTRUMENT ____________________ SERIAL NO. __________________ INSTRUMENT RETURNED FOR: CALIBRATION ____ REPAIR ____ In case of repair, please specify the following. DATE OF FAULT _______________ REPORTED BY E-MAIL, PHONE ___________ COMPANY _____________________ USER’S REFERENCE ____________________ FAULT DESCRIPTION ____________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________ HOW DID IT OCCUR ______________________________________________________ ___________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________ LOCAL ANALYSIS OR ATTEMPTS TO REPAIR _______________________________ ___________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________ RECOMMANDATIONS AND NOTES _________________________________________ ____________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________

Page 41: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 41 of 46

4 CALIBRATION

4.1 INTRODUCTION T/2000 does not need to be calibrated, as metering circuits employ high stability components. It is suggested to check the unit every 3 years, by comparing T/2000 measurements to external meters. Tests should be performed with an high accuracy multi-meter, that should guarantee a maximum AC measurement error of 0.1%, both for voltage and current. Besides, as such multi-meters do not have current ranges greater than 2 A, for the test of the high current ranges it is necessary a class 0.1 measurement Current Transformer. The adoption of lower-class instruments may cause false interventions, that introduce errors into the test set. At the end of the test, if deviations are not acceptable it is possible to enter the calibration mode, as explained in the followings.

4.2 CALIBRATION PROCEDURE The calibration mode is accessed by pressing push-buttons < and > at the meantime, and powering on T/2000. The windows that opens up allows for first selections. The operating mode is the following: . Select the range to be calibrated. Calibrated ranger are checked. . Set to zero the adjustment knob. . With no output, ad just the off-set. . Press > to go ON. The external meter measures the output; T/2000 displays its measurement. . There are two adjustment parameters: COARSE and FINE. Go to the COARSE adjustment, and modify it until the T/2000 measurement is greater than the one displayed by the external meter. Reduce the COARSE adjustment until T/2000 measurement is less than the external one, and save the “coarse adjustment. . Now, go to the FINE adjustment, and modify it to the best match with the external meter (the difference should be less than 0.2% of the reading). . Press the multifunction knob to confirm. T/2000 continues to generate: it is possible to check that the measurements match. If necessary, it is possible to re-enter the same range and repeat the adjustment; else, go to the next range: T/2000 goes OFF. . Take now the output to the value to be tested, to verify for the linearity of the measurement. If the measurement has a negative error, go back to range adjustment, reduce the COARSE adjustment parameter and increase the FINE one. . Go to next range, and so on, until the calibration is completed. . Once coarse and fine adjustments are performed, it is possible to correct the off-set readings. This is performed selecting the range without generation, correcting the reading and pressing “set”. Don’t perform the offset calibration before the coarse adjustment, because the offset would be modified. . At the end of off-set calibration, it is necessary to verify the full scale reading: in case you find some (minor) deviation, adjust it, acting on “fine” only. Take care because on some ranges there are more adjustments to be performed on the same range (this is provided in order to null the linearity error:

- EXT I, 20 mA: 2 calibrations on the AC range, 4 calibrations in the DC range;

Page 42: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 42 of 46

- EXT I, 10 A: 4 calibrations in the DC range; - EXT V, 10 V: 2 calibrations on the AC range, 4 calibrations in the DC range; - EXT V, 600 V: 2 calibrations on the AC range, 4 calibrations in the DC range; - MAIN I DC: 2 calibrations in the DC range.

These calibrations are performed with “Coarse” and “Fine” on the range setting; for other settings, only the “Fine” adjustment is performed. DC adjustments are: the range, positive; 20% of the range, positive, 20% of the range, negative, 80% of the range, negative. During these adjustments, the display shows the input to be applied (for instance, 10.66 mA on the 20 mA DC, nominal range). The actual applied value should deviate from the displayed one by no more than 5%. The program alerts if a scale is not calibrated. At the end of the session, all parameters are saved in the non volatile FLASH EPROM.

4.3 T/2000 OUTPUT CALIBRATION The internal measurements calibration is performed by generating the following current and voltage outputs. For each output range there is a number of measurement scales, that are separately adjusted: during the use T/2000 automatically selects the best scale. This increases the measurement accuracy.

4.3.1 OUTPUT CURRENTS CALIBRATION NOTE: the current output from the main AC voltage is also calibrated. OUTPUT SOCKET CONTROL 300 V AC; current measurement (Mis_IAC_main)

Range Calibration value Notes 0.008 A 5 mA Output 300 Vac Main ; series resistor 1000 Ohm 0.019 A 14 mA Output 300 Vac Main ; series resistor 1000 Ohm 0.079 A 50 mA Output 300 Vac Main ; series resistor 1000 Ohm 0.199 A 140 mA Output 300 Vac Main ; series resistor 100 Ohm 0.799 A 500 mA Output 300 Vac Main ; series resistor 10 Ohm 1.999 A 1,4 A Output 300 Vac Main ; series resistor 10 Ohm

10 A ; current measurement (Mis_IAC_main)

Range Calibration value Notes 0.199 A 140 mA Copy values from the 40 A range 0.799 A 500 mA 1.999 A 1,4 A 7.999 A 5 A 19.99 A 14 A

40 A ; current measurement (Mis_IAC_main)

Range Calibration value Notes 0.199 A 140 mA Output 40A; series resistor 10 Ohm 0.799 A 500 mA Output 40A; series resistor 10 Ohm 1.999 A 1,4 A Output 40A; series resistor 10 Ohm 7.999 A 5 A Output 40A; Current Transformer 10//1 19.99 A 14 A Output 40A; Current Transformer 20//1 79.99 A 50 A Output 40A; Current Transformer 50//1

800 A;current measurement (Mis_IAC_main)

Range Calibration value Notes 0.799 A 500 mA Output 800A; series resistor 10 Ohm 1.999 A 1,4 A Output 800A; series resistor 1 Ohm 7.999 A 5 A Output 800A; series resistor 1 Ohm 19.99 A 14 A Output 800A; series resistor 0.47 Ohm 79.99 A 50 A Output 800A; Current Transformer 50//1 199.9 A 140 A Output 800A; Current Transformer 200//1

Page 43: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 43 of 46

799.9 A 400 A Output 800A; Current Transformer 400//1

4.3.2 OUTPUT VOLTAGES CALIBRATION OUTPUT SOCKET CONTROL 300 V a.c.; voltage measurement (main Vac)

Range Calibration value Notes 1.999 V 1.2 V 7.999 V 5 V 19.99 V 13 V 79.99 V 50 V 199.9 V 140 V 299.9 V 250 V

3000 V a.c. voltage measurement

Range Calibration value Notes 19.99 V 13 V Start with VARIAC at zero 79.99 V 50 V Start with VARIAC at zero 199.9 V 130 V Start with VARIAC at zero 799.9 V 500 V Start with VARIAC at zero 1999 V 1400 V Use a VT with ratio 3000//100 2999 V 2500 V Use a VT with ratio 3000//100

260 Vac auxiliary

Range Calibration value Notes 1.999 V 1.2 -1.3 V 7.999 V 5 V 19.99 V 13 V 79.99 V 50 V 199.9 V 140 V 299.9 V 250 V

130/260 V d.c.

Range Calibration value Notes 19.99 V 13 V The offset adjustment is performed adjusting the output to the

minimum, and then short-circuiting it 79.99 V 50 V The offset adjustment is performed adjusting the output to the

minimum, and then short-circuiting it 199.9 V 140 V The offset adjustment is performed adjusting the output to the

minimum, and then short-circuiting it 299.9 V 250 V The offset adjustment is performed adjusting the output to the

minimum, and then short-circuiting it

Page 44: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 44 of 46

4.4 EXTERNAL INPUT CURRENTS INPUT SOCKET

CONTROL

10 A AC

Range Calibration value Notes 0.039 A Follow display

indications Use a stable source, such as a DRTS test set. 33mA

0.099 A as above as above 72mA 0.399 A as above as above 330mA 0.999 A as above as above 720mA 3.999 A as above as above 3,3A 9.999 A as above as above 7,2A

10 A dC (Mis_25m_10Aext)

Range Calibration value Notes 0.039 A Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

0.099 A as above as above 0.399 A as above as above 0.999 A as above as above 3.999 A as above as above 9.999 A as above as above

25mA dC (Mis_25m_10Aext)

Range Calibration value Notes 0.2 mA Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

0.8 mA as above as above 2 mA as above as above 8 mA as above as above 25 mA as above as above

20 mA AC (Mis_25m_10Aext)

Range Calibration value Notes 0.2 mA Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

0.8 mA as above as above 2 mA as above as above 8 mA as above as above 20 mA as above Use a stable source, such as a DRTS test set.

On this measurement there are 2 calibrations.

Page 45: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 45 of 46

4.5 EXTERNAL INPUT VOLTAGES INPUT SOCKET

CONTROL

10 Vac (Mis_600Vext)

Range Calibration value Notes 0.010 V Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

0.039 V as above Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

0.099 V as above Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

0.399 V as above Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

0.999 V as above as above 3.999 V as above as above 9.999 V as above as above

10 Vdc (Mis_600Vext)

Range Calibration value Notes 0.010 V Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

0.039 V as above Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

0.099 V as above Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

0.399 V as above as above 0.999 V as above as above 3.999 V as above as above 14.999 V as above as above

600 VdC (Mis_600Vext)

Range Calibration value Notes 0.999 V Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 4 calibrations.

3.999 V as above as above 15.99 V as above as above 79.99 V as above as above 199.9 V as above as above 799.9 V as above as above

600 Vac (Mis_600Vext)

Range Calibration value Notes 0.999 V Follow display

indications Use a stable source, such as a DRTS test set. On this measurement there are 2 calibrations.

3.999 V as above as above 11.99 V as above as above 59.99 V as above as above 149.9 V as above as above 599.9 V as above as above

Page 46: T2000 INTRODUCTORY GUIDE - User Equipuserequip.com/files/specs/5693/Service _ Calibration.pdf · 2016. 8. 28. · T2000 INTRODUCTORY GUIDE . DOC. MIE12210 Rev. 1 Page 2 of 46 REVISIONS

DOC. MIE12210 Rev. 1 Page 46 of 46

APPENDIX 1 SPARE PARTS LIST The following is the suggested list of spare parts for a duration of five years and for up to five test sets. The reference is made to the list of components. 1) N. 10 Fuses in the supply socket (1), rated T10A, code XFU22101. 2) N. 10 Fuses for the 25 mA metering (69), rated F63mA, code XFU23115. 3) N. 1 DC to DC converter for the display (23) backlight, code XCA10348. 4) N. 1 Encoder with switch, for MENU control knob (22), code XCM10160. 5) N. 1 Printed circuit board INTE ON-OFF (7), code PWA11410.