Version: Final Version 1.0 Date: 09Oct2012 Site Imaging Manual ACRIN 6701 Repeatability Assessment of Quantitative DCEMRI and DWI: A Multicenter Study of Functional Imaging Standardization in the Prostate

Site Imaging Manual ACRIN 6701

  • Upload
    others

  • View
    15

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Site Imaging Manual ACRIN 6701

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

Version: Final Version 1.0

Date:  09‐Oct‐2012

     

Site Imaging Manual 

ACRIN 6701   

Repeatability Assessment of Quantitative DCE‐MRI and DWI:A Multicenter Study of Functional Imaging Standardization  

in the Prostate   

Page 2: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.1of16

 

Table of Contents Letter of Introduction ................................................................................................................................. 3 

ACRIN 6701 Study Schema .......................................................................................................................... 4 

1.0  OVERVIEW OF IMAGING REQUIREMENTS .......................................................................................... 5 

2.0  STUDY OBJECTIVES/SPECIFIC AIMS ..................................................................................................... 6 

3.0  SCANNER QUALIFICATION ................................................................................................................... 6 

3.1  DWI Phantom Scanning ........................................................................................................ 7 

3.2  DCE‐MRI Phantom Scanning ................................................................................................. 7 

3.3  Qualification Review ............................................................................................................. 7 

3.4  Phantom Re‐scan .................................................................................................................. 8 

3.5  Qualifying Image Submission ................................................................................................ 8 

4.0  Participant Eligibility ............................................................................................................................ 9 

4.1  Inclusion Criteria ................................................................................................................... 9 

4.2  Exclusion Criteria................................................................................................................... 9 

5.0  Participant Scheduling ....................................................................................................................... 10 

6.0  Participant Preparation ..................................................................................................................... 10 

7.0  Contrast Agent Administration ......................................................................................................... 10 

8.0  Standardized Image Acquisition Protocols ........................................................................................ 11 

9.0  Image Submission .............................................................................................................................. 11 

  9.1  TRIAD .......................................................................................................................................... 11 

  9.2  Image Transmittal Worksheet (ITW) .......................................................................................... 12 

10.0  Image Quality Control (QC) .............................................................................................................. 12 

10.1  Local Site Review of Imaging .................................................................................................... 12 

10.1.1  Site Assessment Specific to MRI Scan 1 .................................................................. 12 

             10.1.2  Site Assessment Specific to MRI Scan 2 .................................................................. 13 

     10.1.3  Locally‐Determined Deficient Imaging Study ......................................................... 14 

  10.2  ACRIN Core Laboratory Quality Control Technical Review ...................................................... 15 

             10.2.1 Image Data Queries .................................................................................................. 15 

  10.3  Central Diagnostic Review of MRI SCAN 1 ............................................................................... 15 

11.0  Imaging Forms……..………………………………………………………………………………………………………..……………16 

Page 3: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.2of16

REFERENCES………………………………………………………………………………………………………………………………………17 

APPENDIX I: DWI PHANTOM TEST INSTRUCTIONS 

APPENDIX II: DCE PHANTOM TEST INSTRUCTIONS 

APPENDIX III: MRI SCAN 1 IMAGING PROCEDURES 

APPENDIX IV: MRI SCAN 2 IMAGING PROCEDURES 

APPENDIX V: MRI IMAGE ACQUISITION PARAMETERS 

 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

History of Revisions: 

 

   

   

   

   

Page 4: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.3of16

 

 

Letter of Introduction  

 

Dear Imaging Staff,  

 

This  Site  Imaging  Manual  contains  the  image  acquisition  instructions  for  the  ACRIN  6701  trial: 

“Repeatability  Assessment  of  Quantitative  DCE‐MRI  and  DWI:  A Multicenter  Study  of  Functional 

Imaging Standardization in the Prostate.” 

 

To successfully meet the study objectives,  it  is critical that the DCE‐MRI and DWI  image datasets are 

acquired according to the imaging protocol detailed in this manual.  

 

Quality Control (QC) review of the images will be performed by the ACR Imaging Core Laboratory. This 

review will be performed  in a timely  fashion, as part of ACRIN standard operating procedures.  If any 

protocol  deviations  or  technical  issues  are  identified  during  the  review,  an  ACR  Core  Lab  Imaging 

Technologist will contact your site to provide feedback expeditiously. This will allow your site to make 

any necessary adjustments early in the conduct of the study.  

 

The ACRIN 6701  Imaging Team wishes  to  thank you  in advance  for your diligence  in adhering  to  the 

procedures described in this manual to ensure the integrity of the image data collected for the study. 

Please  do  not  hesitate  to  contact  the  ACRIN  6701  Imaging  Technologist  (see  contact  information 

below) if you have any questions.  

  

Sincerely, 

 

Dena Flamini RT(R)(M)(MR) 

ACRIN 6701 Imaging Analyst 

 

American College of Radiology 

1818 Market Street‐ Suite 1600 

Philadelphia, PA 19103 

Phone: (215) 940‐8880 

Fax: (215) 923‐1737 

www.acrin.org 

[email protected] 

   

Page 5: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.4of16

ACRIN 6701 Study Schema 

42 Day Maximum Timeline from Enrollment to Completion of Both MRI Scans 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SOCAnatomicMRIplusStudy‐RelatedDWIScan

andGad‐EnhancedDCE‐MRI#1:

MRIBetween0and28Daysafter Enrollment

Study‐RelatedDWIScanandGad‐EnhancedDCE‐MRI

#2:

MRIBetween2and14CalendarDaysafterMRIScan#1

        TREATMENT

Between 0 and 28 days later

Between 2 to 14 days later

Day#0

Eligibility/Enrollment:

Calculate GFR for Renal Function ACRIN Web Registration

Page 6: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.5of16

 

1.0 OVERVIEW OF IMAGING REQUIREMENTS 

 

 

DCE‐MRI and DWI 

Qualification 

 

Submit prior to site activation (for each scanner): 

Protocol Specific Application (PSA) 

Submission of phantom‐based DCE‐MRI and DWI qualification exams  

     (acquired per the ACRIN 6701 standardized imaging protocol). 

 

TRIAD Installation Should  be  installed  prior  to  study  participant  enrollment  for  secure, 

electronic submission of imaging to ACRIN.  

Time Points  

for Trial Examinations 

DCE‐MRI  and  DWI  per  standardized  ACRIN  6701  acquisition  protocol 

provided in this imaging manual: 

MRI/DWI/DCE‐MRI‐1: between 0 and 28 days of enrollment  

(at time of standard‐of‐care anatomic MRI per institutional practice) 

MRI/DWI/DCE‐MRI‐2: between 2 and 14 calendar days after MRI Scan #1

Image Submission 

ACRIN 6701  imaging should be submitted electronically to the ACR  Imaging 

Core Lab via TRIAD. All  imaging should be submitted within 48 hours after 

acquisition and should include an Image Transmittal Worksheet (ITW). 

Data Queries  ACRIN will issue queries, as needed, based on QC review of imaging.  

Rescanning With 

Phantom for 

Maintenance 

The  qualification  phantoms  must  be  rescanned  prior  to  any  subsequent 

participant scan if there are any substantive changes in scanner hardware or 

software during the trial. 

 

General Trial Requirements: 

3T whole body MRI scanner 

Institutional expertise in prostate MRI (proven record of 50 prostate MRIs per year) 

Site must be able to utilize a torso‐array coil 

Site must be able to utilize a power injector 

The site must perform both study‐related DCE‐MRI and DWI examinations on the same MRI 

unit  for  the  same  participant  using  the  same magnet  configuration  (manufacturer;  field 

strength; coil) 

Site must submit all trial exams to ACRIN within 48 hours after acquisition 

 

NOTE: It  may  be  helpful  to  keep  a  copy  of  this  Manual  in  the  imaging  department,  so  all 

technologists  involved  in  the  image  acquisition  of ACRIN  6701  are  privy  to  the  protocol‐required 

imaging specifications. 

Page 7: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.6of16

2.0 STUDY OBJECTIVES/SPECIFIC AIMS  

The Quantitative Imaging Biomarkers Alliance (QIBA) dynamic contrast‐enhanced‐magnetic resonance 

imaging  (DCE‐MRI)  committee  is  working  to  identify  the  limits  of  reproducibility  of  DCE‐MRI 

quantitative metrics critical for useful biomarker studies. The ACRIN 6701 trial will examine the QIBA 

DCE‐MRI  profile  claims  regarding  the  repeatability  of  tumor  DCE‐MRI,  as  well  as  repeatability  of 

diffusion weighted  imaging  (DWI) metrics, and evaluate  the performance of  these metrics  in human 

subjects  with  prostate  tumors.  In  addition  to  implementing  the  imaging  protocol,  the  trial  will 

introduce  site  imaging  personnel  to  proper  procedures  for  magnet  qualification,  the  selection  of 

tumor‐bearing human subjects, and the  implementation of the proper DCE‐MRI and DWI parameters 

in general oncologic MRI. 

Primary Hypothesis 

Within the context of a multi‐site clinical trial, the upper limits of the repeatability coefficient (RC) for 

DCE‐MRI metrics Ktrans and blood‐normalized initial area under the gadolinium curve (IAUCG90bn), and 

DWI metric D(t), using the whole prostate as a target  lesion, can be demonstrated to be  less than or 

equal to the pre‐specified values that signify the minimal detectable significant biologic change of each 

quantitative MRI metric. 

Hypothesis  #1:  The  repeatability  coefficient  (RC)  of  the  DCE‐MRI metric  Ktrans,  as measured  by median pixel values of the whole prostate, is equal to or less than 22%. 

Hypothesis #2: The RC of the DCE‐MRI metric  IAUGC90bn (blood‐normalized  initial area under the gadolinium curve between 0 and 90 seconds), as measured by median pixel values of  the whole prostate, is equal to or less than 0.05 (in normalized units). 

Hypothesis  #3:  The  RC  of  DWI metric  D(t),  as measured  by median  pixel  values  of  the whole prostate, is equal to or less than 0.3 x 10‐3 mm2/sec. 

 

3.0 SCANNER QUALIFICATION 

Participation  in  the  ACRIN  6701  requires  that  all  sites  adhere  to  standardized DCE‐MRI  and DWI 

protocol parameters.   The use of standard  imaging guidelines  is an essential component of clinical 

trials in which imaging plays a central role in the research endpoints. This is of particular importance 

in  multicenter  trials  where  equipment,  personnel,  and  imaging  acquisition  protocols  can  vary 

significantly.  Thus,  the  use  of  standardized  imaging  guidelines  helps  control  the  inter‐  and  intra‐

variability inherent in multicenter imaging trials.   

The purpose of any imaging qualification process is to help ensure the trial imaging is of high quality 

and performed per the trial‐standardized acquisition protocol. To participate in the ACRIN 6701 trial, 

each  site must  qualify  by  first  scanning  the  DWI  and  DCE‐MRI  phantoms  based  on  ACRIN  6701 

phantom  scanning  protocols.  Phantom  scanning  provides  an  opportunity  to  evaluate  compliance 

Page 8: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.7of16

with  sample  imaging acquisition protocols prior  to participant  recruitment and actual  trial‐specific 

protocols.   

These qualification examinations will be  reviewed by  the ACR  Imaging Core  Lab  for both protocol 

compliance and  image quality; approval of the qualification exams  is required prior to ACRIN 6701 

site activation. Suboptimal  image quality and/or  imaging not performed per  the  trial‐standardized 

protocol  can  result  in  exclusion  of  the  imaging  exam(s)  and/or  the  entire  case  from  analysis. 

Therefore,  routine QC  and  adherence  to  the ACRIN  6701  image  acquisition  protocol  are  of  great 

importance, and sites will be asked to re‐scan using the phantoms should there be any substantive 

changes in hardware or software to the scanner during the conduct of the trial.  

3.1 DWI Phantom Scanning  

Prior to site qualification, a DWI phantom test scan must be run. Sites will be provided with a 

DWI  ice‐water  phantom;  this  phantom  consists  of  a  1.5  liter white  plastic  container with  a 

sealed  75ml  tube  pre‐filled with  distilled water.  The  design  of  this  phantom  allows  for  the 

measurement of apparent diffusion coefficient (ADC) values of a known temperature‐controlled 

fluid  to  confirm  proper MRI  system  performance  in  acquisition  of  diffusion weighted  (DW) 

imaging. Upon  completion  and  approval  of  the DWI  phantom  qualification  testing,  sites  are 

asked to keep the ice‐water phantom in the event that there is a need for DWI phantom testing 

re‐qualification. 

Please see Appendix I for detailed DWI phantom test scan parameters and instructions.  

3.2 DCE‐MRI Phantom Scanning  

Prior  to  site  qualification,  a  DCE  phantom  test  scan  must  be  run  to  assess  MRI  system 

performance.  Sites will  be  provided  a  standardized multi‐compartment  R1  phantom  having 

inserts with T1 relaxation times spanning the ranges typically obtained in blood and in tissue.  

Sites  are  to  retain  the  DCE  phantom,  until  the  ACR  Imaging  Core  Lab  has  approved  the 

submitted imaging; a Core Lab representative will then provide detailed shipping instructions.  

 

Please see Appendix II for detailed DCE phantom test scan parameters and instructions.  

3.3 Qualification Review 

The  qualification  exam  will  be  evaluated  for  image  quality  (e.g.  artifacts,  distortion,  and 

signal‐to‐noise) and compliance with the standardized DCE‐MRI and DWI acquisition protocol.  

An  ACR  Imaging  Core  Lab  Imaging  Technologist  will  notify  the  site  of  the  results  of  the 

qualification  review  via  e‐mail.    E‐mail  documentation  will  include  your  site’s  Study 

Coordinator (SC), site PI, and lead technologist to inform the site team whether the qualifying 

exams  have  been  approved  or  not.  If  the  qualifying  exams  are  not  approved,  required 

corrections for rescanning will be included in the e‐mail.   

NOTE:  Approval  of  these  qualifying  exams  is mandatory  prior  to  your  site  registering  a 

participant onto the ACRIN 6701 trial.   

Page 9: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.8of16

3.4 Phantom Re‐scan 

If DWI or DCE‐MRI initial phantom scans are not approved after initial scanning and submission, phantom re‐scans (DWI or DCE‐MRI) will be required until parameters for approval are met for site qualification.   Upon approval of both the DWI and DCE‐MRI phantom scans, no additional requalification is required unless one of the following conditions occurs: 

a. New  scanner  being  introduced.  A  new  scanner  system,  not  ACRIN  qualified  via phantom assessment, is to be used for subsequent patient studies.  

b. Major  changes  to  qualified  scanner.  A  major  change  in  scanner  hardware  or software occurs during the course of the study.  This would include major software upgrades  (i.e.,  a  software  upgrade  that  result  in  a  new  software  primary  version number) and changes to the coil system, gradient systems, or RF amplifier. Periodic minor updates to existing software versions are not considered major changes. 

c. Trial ongoing after one year. If active patient accrual and scanning are underway at one‐year after the date of the original approval of the phantoms for the trial, then requalification will be required, with replicate scanning of the DWI and/or DCE‐MRI phantoms. 

NOTE: We  strongly  encourage  sites  to  use  a  single  3T MRI  scanner  for  the  entire  study.  However, a second 3T MRI scanner for use in this study may be necessary for some sites.   In such  cases,  qualification—via  DWI  and  DCE‐MRI  phantom  scanning  and  approval—is necessary for each scanner to be used in the study.   ACRIN strongly encourages that all MRI scans  for a given patient  take place on  the same scanner unless such  is not  feasible due  to temporary scanner failure.  

3.5 Qualifying Image Submission 

Phantom  images  in DICOM  format are to be sent to the ACR  Imaging Core Lab on Compact 

Disc (CD) or DVD.   

Please label the disk, in permanent marker, with the following information: 

Site Name 

Trial Name (ACRIN 6701) 

Date of Phantom Imaging (DD‐MMM‐YYY) 

Type of Phantom Imaging (e.g. DWI or DCE imaging) 

 

For  questions  related  to  DCE‐MRI/DWI  phantom  qualification,  please  send  an  email  to: 

[email protected].  Enter “ACRIN 6701 DCE‐MRI/DWI Qualification” in the subject line. 

Ship All Qualification Imaging to: 

American College of Radiology 

1818 Market St., Suite 1600 

Philadelphia, PA 19103 

Attn: ACRIN 6701 DCE‐MRI/DWI Qualification 

Page 10: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.9of16

4.0 Participant Eligibility 

4.1 Inclusion Criteria 

Men 18 years of age or older 

Diagnosis of prostate cancer by trans‐rectal ultrasound (TRUS)‐guided biopsy 28 to 90 days 

prior to enrollment 

Minimal tumor burden as defined by at least one of the following criteria: 

o One single core with ≥ 50% cancer burden and ≥ 5 mm tumor length 

o Two or more cores in the same prostate region, each with ≥ 30% cancer burden 

o Three  or  more  cores  positive  for  prostate  cancer  (of  any  magnitude  of  cancer 

burden) in the same prostate region 

o Gleason grade 7 or higher cancer burden 

o PSA ≥ 10 mg/mL 

4.2 Exclusion Criteria 

Severe claustrophobia not relieved by oral anxiolytics per institutional standard practice 

Presence  of  MRI‐incompatible  metallic  objects  or  implanted  medical  devices  in  body 

(including  but  not  limited  to:  non‐MRI  compatible  metal  objects,  cardiac  pacemaker, 

aneurysm clips, artificial heart valves with steel parts, metal fragments in the eye or central 

nervous system) 

Renal failure, as determined by glomerular filtration rate (GFR) < 30 mL/min/1.73 m2 based 

on a serum creatinine level obtained within 48 hours prior to enrollment 

Weight greater than that allowable by the MRI table, per local institutional practice 

Anti‐androgenic therapy within 30 days prior to enrollment 

Prior external beam, proton, or brachytherapy to the prostate 

Prior hip replacement or other major pelvic surgery 

 

Page 11: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.10of16

5.0 Participant Scheduling 

Participants will undergo two 3T MRI studies for the trial. These quantitative MRI studies (DCE‐MRI and 

DWI) will be performed without  the endo‐rectal  coil.  Sites  are  to  follow  local procedures  for endo‐

rectal coil use during the first MRI, which should be the standard practice anatomical MRI.  

MRI SCAN 1 and MRI SCAN 2 must be completed no less than 2 calendar days (to ensure 24 hours for 

clearance of  gadolinium)  and no  greater  than 14 days  apart,  and both must be  completed prior  to 

treatment initiation.  

NOTE: The same MRI unit and configuration must be used for MRI SCAN 1 and MRI SCAN 2. The same 

contrast agent used for MRI SCAN 1 must be utilized for MRI SCAN 2. 

 

6.0 Participant Preparation 

Prepare participant  according  to  local  standard practice,  including  any pre‐treatment  for  severe 

claustrophobia or anxiety. 

Confirm renal function sufficiency prior to imaging (eGFR must be assessed within 48 hours prior to 

participant  enrollment  but  does  not  have  to  be  repeated  for  the  trial  as  long  as  scheduling 

timelines do not deviate). 

Same size and placement of  the  intravenous catheter  is  recommended  for both MRI scans. Each 

participant should have an  intravenous catheter with a gauge no smaller  than 20 gauge.  Ideally, 

the  injection  site  that was  used  for  the  initial  exam  (MRI  SCAN  1)  should  be  the  same  for  all 

subsequent exams  for  that  given patient  (MRI  SCAN 2) and  any additional  imaging  that may be 

required to complete the trial protocol  (see protocol Sections 8.3 and 8.4.1  for details).  Injection 

through a port‐a‐catheter or other permanent indwelling central catheter is not permitted. 

The participant will be placed supine  in the magnet; arms at side are acceptable. The torso array 

coil should be placed about the pelvis per institutional standard practice. Landmark the participant 

at the level of the symphysis pubis 

Initiate imaging … 

NOTE:  If institutional standards require the use of endo‐rectal coil for anatomic imaging during MRI 

SCAN 1, this should be placed prior to  imaging‐ as per  intuitional practice.   See note regarding the 

use of endo‐rectal coil for anatomic imaging in Appendix III‐ Part II. 

 

7.0 Contrast Agent Administration 

Each participant should have an  intravenous catheter (no smaller than 20 gauge), placed prior to 

imaging.  

Contrast agent should be administered in a dynamic fashion with a power injector at both test and 

re‐test time points.  

Page 12: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.11of16

Contrast injection should begin 30‐40 seconds after the start of data acquisition.   

For the DCE (3D T1) sequence, gadolinium contrast agent should be administered intravenously at 

a dose of 0.1 mmol/kg body weight and rate of 2 mL/second,  followed by a 20 mL saline  flush, 

also at 2 mL/second.  

The type, amount, rate of  injection, and site/gauge of  IV should comply with the study guidelines 

and should be documented by the technologist on the MRI Scan Administration form. 

NOTE: The gadolinium agents MultiHance, Eovist, and Vasovist are NOT permitted to be used as a 

contrast agent  for  the  subjects on  this  trial. Also, keep  in mind  that  the  same brand and dose of 

contrast,  same  IV  site,  and  same  rate  of  contrast  and  saline  flush  administration  should  be 

maintained for both the test and re‐test examinations. 

 

8.0 Standardized Image Acquisition Protocols 

At each time point, the ACRIN 6701 imaging protocols will include a: 

Pre‐contrast T2‐weighted sequence  

Pre‐contrast diffusion‐weighted imaging (DWI) SE‐EPI sequence  

Pre‐contrast  T1 mapping sequence set  

Pre‐contrast coil ratio map sequence set 

Pre‐/post‐contrast T1‐weighted dynamic contrast‐enhanced (DCE) sequence   

Please  review  the  participant  preparation  description  in  Section  6.0  above  and  the  following 

participant preparation instructions and acquisition parameters carefully as this imaging protocol may 

or may not align with your institution’s standard prostate MRI protocol.  

NOTE: Additional imaging for clinical staging as per institutional routine may occur during MRI SCAN 

1.  See Appendices III‐ V for time‐point specific exam sequences. 

 

9.0 Image Submission 

9.1 TRIAD 

All  trial  exams  will  be  submitted  to  ACR  Core  Laboratory  via  TRIAD.  TRIAD  is  a  software 

application  that ACRIN  provides  for  installation  on  a  site’s  PC. One  or  several  computers  of 

choice within the institutional “firewall” and on the institutional network may be equipped with 

TRIAD software; internet access is also required. The TRIAD application can then be configured 

as a DICOM destination on either scanner(s) and/or PACS system for direct network transfer of 

study related  images  into the TRIAD directory. When properly configured, the TRIAD software 

anonymizes,  encrypts,  and  performs  a  lossless  compression  of  the  images  before  they  are 

transferred to the ACRIN image archive in Philadelphia.  

Your site will need to communicate with their IT department for assistance with the initial steps 

outlined  in  the  TRIAD  Pre‐Installation  Checklist  (see Appendix VI)  prior  to  scheduling  TRIAD 

Page 13: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.12of16

installation.  

Once  equipment‐readiness  has  been  determined,  imaging  personnel  from  ACRIN  will 

coordinate installation and training for the software. 

The  TRIAD  Installation  Guide  and  TRIAD  Users  Guide  can  be  found  online  at 

https://triad.acr.org/Learning.htm  

To contact TRIAD Support call: 215‐940‐8820 or email TRIAD‐[email protected] 

 

9.2 Image Transmittal Worksheet (ITW) 

All imaging should be submitted within 48 hours after acquisition and should include an Image 

Transmittal Worksheet (ITW). An  Image Transmittal Worksheet (ITW)  is used during the exam 

QC review to verify a complete transfer of images has been submitted to the ACR Imaging Core 

Lab.  The  ITW  is  completed  in  the Medidata/Rave  data management  system  and  upon  the 

completion  of  this  form,  an  email  will  be  automatically  generated  to  the  ACRIN  Imaging 

Technologist to notify an image submission has taken place.  

 

10.0 Image Quality Control (QC) 

10.1 Local Site Review of Imaging 

10.1.1 Site Assessment Specific to MRI Scan 1 

After completion of  imaging  for MRI SCAN 1,  the  study  should be  reviewed based on 

standard institutional practices for clinical image review and reporting. This assessment 

will include: 

Adequacy of endo‐rectal coil placement for anatomic imaging (if applicable) 

Proper  placement  and  functional  of  torso  array  coils  throughout  all  imaging 

sequences 

Removal of endo‐rectal coil  (if applicable) and proper participant  repositioning 

and  landmarking  prior  to  initiation  of  non  endo‐rectal  coil  study  imaging 

sequences 

Adequacy of anatomic imaging, especially the axial T2‐weighted series, with:  

o proper  anatomic  prescription  to  include  entire  prostate  and  seminal 

vesicles 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

Adequacy of DWI imaging, with:  

o proper  acquisition  parameters,  including  FOV, matrix,  phase  direction, 

slice thickness, and use of fat saturation 

o proper  anatomic  prescription  to  include  entire  prostate  and  seminal 

vesicles 

Page 14: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.13of16

o proper number and choice of B values 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

Adequacy of DCE‐MRI imaging, with:  

o proper  acquisition  parameters,  including  FOV,  matrix,  phase  direction 

slice thickness, and slice number 

o proper anatomic prescription  including  full pelvic  imaging with prostate 

located eccentrically in the inferior aspect of the slab to minimize arterial 

inflow 

o completion of all  three aspects of  the DCE‐MRI series: T1 mapping, coil 

ratio map, and dynamic gadolinium  imaging, with proper choices of  flip 

angle, signal averaging, and number of phases for each 

o correct choice of contrast agent type, dose, rate of  injection, and timing 

of injection 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

10.1.2 Site Assessment Specific to MRI Scan 2  

After completion of  imaging  for MRI SCAN 2,  the  study  should be  reviewed based on 

standard institutional practices for clinical image review and reporting. This assessment 

will include: 

Proper  placement  and  functional  of  torso  array  coils  throughout  all  imaging 

sequences 

Adequacy of axial T2‐weighted series for prostate localization 

Adequacy of “pre‐coffee break” DWI imaging, with:  

o proper  acquisition  parameters,  including  FOV, matrix,  phase  direction, 

slice thickness, and use of fat saturation 

o proper  anatomic  prescription  to  include  entire  prostate  and  seminal 

vesicles 

o proper number and choice of B values 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

Correct  implementation  of  the  “coffee  break”  as  evidenced  by  correct 

replacement  of  torso  array  coils,  repeat  landmarking  and  axial  T2W  imaging, 

repeat  localizer  imaging,  and  adequate  temporal  gap  between  the  two  DWI 

image sets (minimum of three minutes) 

Adequacy of “post‐coffee break” DWI imaging, with:  

Page 15: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.14of16

o proper  acquisition  parameters,  including  FOV, matrix,  phase  direction, 

slice thickness, and use of fat saturation 

o proper  anatomic  prescription  to  include  entire  prostate  and  seminal 

vesicles 

o proper number and choice of B values 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

Adequacy of DCE‐MRI imaging, with:  

o proper  acquisition  parameters,  including  FOV,  matrix,  phase  direction 

slice thickness, and slice number 

o proper anatomic prescription  including  full pelvic  imaging with prostate 

located eccentrically in the inferior aspect of the slab to minimize arterial 

inflow 

o completion of all  three aspects of  the DCE‐MRI series: T1 mapping, coil 

ratio map, and dynamic gadolinium  imaging, with proper choices of  flip 

angle, signal averaging, and number of phases for each 

o correct choice of contrast agent type, dose, rate of  injection, and timing 

of injection 

o acceptable signal‐to‐noise ratio 

o absence  of  severe  artifacts  that would  render  the  series  clinically  non‐

diagnostic 

10.1.3 Locally‐Determined Deficient Imaging Study 

If the standard practice anatomic sequences of MRI SCAN 1 are deemed suboptimal for 

clinical  evaluation  by  the  local  site  imaging  team,  the  site  PI/radiologist  has  the 

discretion to recommend repeat of one or more clinical imaging series during MRI SCAN 

2. The endo‐rectal coil is not expected to be used for any imaging sequences during MRI 

SCAN 2. However, if the local site radiology investigator deems that re‐use of the endo‐

rectal coil per  standard  institutional practice  for  repeated clinical anatomic  imaging  is 

desirable, repeat endo‐rectal coil placement  is allowed for this portion of MRI SCAN 2.  

The recommendation for re‐use of the endo‐rectal this should be adequately explained 

to the participant prior to MRI SCAN 2. 

Similarly,  if  the  study‐specific DWI and/or  the DCE‐MRI  sequences of MRI SCANS 1 or 

MRI  SCAN  2  are  deemed  technically  inadequate  or  incomplete/not  per  protocol 

specifications,  the  site  PI/radiologist  and/or  ACRIN  has  the  option  to  recommend 

repetition of the deficient study sequences as a separate visit. Only willing participants 

will undergo additional imaging visits.  

Any participants unwilling to complete repeat imaging studies will be off‐study and will be replaced to meet trial accrual. 

Page 16: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.15of16

Repeat imaging must adhere to the protocol‐specific imaging procedures and performed within 14 days of the non‐deficient imaging study.  

The  site must  appropriately  document  all  cases  in  which  deficient  imaging  led  to repeat  of  imaging.  Contact  the  trial’s  lead  ACRIN  data  manager  for  assistance  as needed. 

 

10.2 ACRIN Core Laboratory Quality Control Technical Review  

Upon receipt of the images at the ACR Imaging Core Lab, an initial QC review will be conducted 

by  a  qualified  ACRIN  Imaging  Technologist.  The  ACRIN  Imaging  Technologist  will  check  for 

missing  images/sequences,  appropriate  image anonymization,  complete  anatomical  coverage 

of the prostate, adherence of all sequences to imaging protocol, and absence of image artifact.  

In cases where  image sets are  judged to be suboptimal (“technically  inadequate”), the trial PI 

will be informed, and a replacement participant will be accrued from participating institution. 

10.2.1 Image Data Queries 

If it is found during the QC review that the submitted exam has missing data or does not 

follow  the  protocol  guidelines,  detailed  in  this  manual,  the  Medidata/Rave  data 

management  system will  generate  an  auto‐query.  Sites  are  expected  to  resolve  data 

queries expeditiously. Queries not  resolved within 7 business days will be  sent  to  the 

ACRIN 6701 trial team for additional follow‐up. 

10.3 Central Diagnostic Review of MRI SCAN 1 

Prior to quantitative DWI and DCE‐MRI analysis, two central radiologists will each assess for the presence  of  visible  tumor  nodules.  If  central  diagnostic  evaluation  of MRI  SCAN  1  by  either reader  fails  to  reveal  a  definite  or  likely  tumor  nodule  of  5 mm  or  greater,  a  replacement participant will be enrolled to the trial 

Replacement participants will be enrolled for any of the following reasons 

An enrolled participant does not complete all required imaging for MRI SCAN 1 and MRI SCAN 2 

Either MRI SCAN 1 and/or MRI SCAN 2  is determined to be technically  inadequate, and the inadequate imaging series cannot be obtained during a repeat visit. 

MRI SCAN 1 does not reveal at least one definite or probable tumor nodule 5mm or greater as determined by each of two independent central readers. 

Replacement participants will be accrued at the same participating site with the same vendor equipment,  if the participant whose deficient  imaging needs to be replaced (in order to maintain accrual of 10 participants to each vendor manufacturer). 

Page 17: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual

09Oct2012 pg.16of16

Participants with complete MRI SCAN 1 and MRI SCAN 2  imaging data sets, but whose 

MRI SCAN 1 does not define a dominant  tumor nodule, will  still be evaluated  for  the 

non‐tumor endpoints (e.g. whole prostate evaluation for DWI and DCE‐MRI). 

 

11.0 Imaging Forms 

There are 5 imaging forms that must be completed and submitted via the Medidata/Rave eCRF system 

for each DCE‐MRI and DWI exam.  

Image Transmittal Worksheet (ITW) 

Imaging Procedure form 

MRI Scan Administration 

MRI Imaging Assessment‐ Anatomic 

MRI Imaging Assessment‐ DWI/DCE MRI 

It  is  recommended  that each of  the above  forms be made available,  in hard  copy  format,  for  the 

Imaging Technologists to be completed during image acquisition when the requested information on 

these forms is most readily available. Worksheets of the data asked throughout the various imaging 

forms have been created to aid the technologist during imaging and are included in the appendices 

of this document.  

These Worksheets  are meant  to  approximate  the  actual ACRIN  6701  screens  in  the Medidata  EDC system  (e.g.  provide  users  of  the  EDC  system  with  the  questions  asked  on  the  forms  within  the system). Please note that these are worksheets and are only meant as a HELP TOOL for users who wish to know/prepare for the information content of the e‐CRFs prior to actually logging into the Medidata EDC system.  

Note:  In  the event  these worksheets  are used,  they  should be  considered  source documents  and stored  in  the  participant’s medical  record.  Please  do  NOT  submit  the worksheets  to  ACRIN‐  all participant and image‐related data must be provided electronically in the Medidata database and no hard copy forms/worksheets will be accepted. 

Page 18: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page1of809Oct2012

 

   

Appendix I 

 

DWI Phantom Test Instructions 

Page 19: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page2of809Oct2012

DWI PHANTOM TEST INSTRUCTIONS 

 

Objective: 

Measure apparent diffusion coefficient (ADC) values of a known temperature‐controlled fluid to 

confirm proper MRI system performance  in acquisition of diffusion weighted  (DW)  images of 

the pelvis using a pelvis/torso array coil. 

DWI Ice Water Phantom Preparation Procedure: 

Once  prepared,  an  ice  water  phantom  requires  approximately  1  hour  to  reach  thermal 

equilibrium.   Please plan on filling the phantom well  in advance of scanning.   Once at thermal 

equilibrium,  the phantom will be usable  for several hours.   Do not dispose of  foam  insulation 

sleeves and zip‐lock plastic bag shipped with each phantom. 

 

1. The phantom consists of a 1.5 liter white plastic container with a sealed 75 ml tube pre‐

filled with distilled water.  Leave the 75 ml tube sealed and affixed within the container.  

For phantom preparation,  temporarily remove  the container  from  the  foam  insulation 

sleeve and zip‐lock bag. 

 

2. You will need a source of ice cubes or ice chips and a sink basin for phantom filling.  The 

volume  of  ice  cubes/chips  required  is  approximately  twice  the  volume  of  the white 

container.   Fill  the white  container  to  the  top with  ice  cubes/chips  then add cold  tap 

water and fill to the top – the colder the water, the better.  Loosely cap the phantom for 

an initial “cool down period” of approximately 10 minutes. 

 

3. Depending on how cold your tap water  is, much of the  ice will melt relatively quickly.  

Therefore after 10 minutes, reopen the container and pack‐in more  ice to replace the 

melted  ice.   Fill  ice to the top allowing displaced water to overflow  into the sink.   The 

objective is to have ice cubes to the full depth of the phantom and the interstitial space 

filled with water and minimal air (Figure 1).   Screw cap on tightly, dry off the phantom 

and inspect for leaks. 

 

4. Put the phantom into a foam insulation sleeve with the container top toward the open 

end of the sleeve and seal the phantom in a zip‐lock bag.  Set the phantom aside for an 

additional 50 minutes to allow the central tube to come to thermal equilibrium with 

the  surrounding  ice  water.    As  long  as  there  is  adequate  ice  in  the  phantom, 

temperature will be controlled to near 0°C, thereby holding water within the tube at a 

Page 20: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page3of809Oct2012

known diffusion coefficient.  The phantom should be usable for several hours within the 

insulation sleeve.  If desired, you can store the phantom in a refrigerator (not a freezer) 

to extend the usable time. 

 

5. After scanning, simply empty  the  ice water down  the drain but  leave the central tube 

filled and sealed for use at a later date by following the same procedure. 

 

DWI Ice Water Phantom Scan Procedure: 

 

1. Set up  the DWI phantom  in  the center of  the magnet  table, placed upright as shown  in 

Figure 2a.  If appropriate for torso array coil imaging at your center, place posterior torso 

coil under the phantom first.   

 

2. Place  foam pads or blocks on sides of  the phantom  to support  the anterior  torso/pelvic 

coil (e.g. see Figure 2). If your system requires additional coil loading, place an appropriate 

loading coil adjacent to the phantom.   Alternately, a torso type  loading coil can be used. 

Finally place the anterior torso array coil as shown in Figure 2b.   

 

3. Start a “New Patient/Exam” using  the same patient entry convention normally used  for 

pelvis (non‐eCoil) MRI exams (e.g. “head first, supine”). 

 

4. Keep phantom in the zip‐lock bag and insulation sleeve during the scan procedure to keep 

condensate off of MRI system components.   

 

5. If required, wedge or tape the phantom using patient support pads to hold the phantom 

firmly  in  place  with  the  water  tube  approximately  vertical  (e.g.  parallel  to  anterior‐

posterior  axis)  as  shown  in  Figure  1.    Precise  alignment  is  not  required,  although  the 

phantom should be positioned roughly in the center of the magnet bore. 

 

Overview of Series Required for DWI QC Phantom Scanning: 

1. Three‐plane scout/survey 

2. Axial 3D T1‐weighted spoiled gradient echo without fat‐suppression 

3. Axial DWI using single‐shot echo‐planar‐imaging (EPI) at b‐values=0, 100, 600, 800s/mm2.  

Note: A set of DWI containing all 4 b‐values is called one “pass”. 

Page 21: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page4of809Oct2012

4. “Copy‐&‐Paste” the DWI series and repeat to acquire a total of four passes in immediate 

succession in four consecutive series.  Avoid changing scanner pre‐scan/hardware settings 

between these consecutive series.  

Details of Required Series: 

1. Three‐plane  scout/survey:  Use  site‐preferred  body  sequence  as  used  for  graphic 

prescription of subsequent series. 

2. Axial 3D T1‐weighted spoiled gradient echo without fat‐suppression:  

            FOV = 320mm (A/P phase) x 320mm (R/L frequency) at 256 x 256 matrix.   

3. Acquire  4mm  slice  thickness  interpolated  to  2mm.    Acquire  sufficient  slices  to  cover 

120mm  in  the  superior/inferior  direction.    TR=4‐10ms,  TE=min,  flip  angle  10o,  single 

average, acquisition time <3min. 

4. Axial  single‐shot  isotropic  diffusion  weighted  single  SE,  echo  planar  imaging  (EPI)  by 

parameter settings listed in Table 1 below.  One DWI “pass” should contain all 4 b‐values.  

Copy and paste  the DWI  series  so  that 4  consecutive passes are acquired  in  separate 

series.  If the MRI system cannot acquire all b‐values in one series, see note in section (e) 

under Table 1.  

         Note: Total scan time for DWI is approximately 4 passes x 3min/pass = 12min. 

 

Page 22: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page5of809Oct2012

Table 1.  DWI Ice Water Phantom SE‐EPI Scan Parameters 

DWI Single‐Shot SE EPI Single Echo

Field Strength  3TDWI Sequence  Single‐Shot SE EPI Single Echo Receiver Coil  Torso/Pelvis ArrayFOV (mm)  320 RL x 260 APAcquisition Matrix  160 RL x 128 AP *(a)Reconstruction Matrix  256 x 256 *(b)Orientation  Axial (Foldover Phase AP; Freq RL)Number of Slices  27Slice Thickness (mm)  5 mmGap (mm)  0TR (ms)  >8000TE (ms)  75‐100Half‐scan, Partial‐Fourier, Frac‐NEX No *(a)Number of Gradient Directions 3 orthogonal axes *(c) Freq Enc Bandwidth (Hz/px)   1500 to 2500 *(d)Parallel Imaging Factor (e.g. SENSE) 2 b‐values (s/mm2)   0, 100, 600, 800 *(e) Number Signal Averages  2Fat Suppression   On *(f)Image Filtering (e.g. SCIC or CLEAR)  Off

 

(*) Notes 

a. Some systems may require “half‐scan”, “partial‐Fourier”, or “fractional‐NEX” in DWI 

scans.   It  is preferred to acquire the full phase‐encode matrix = 120 by setting half‐

scan/ partial‐fourier/fractional‐NEX to “No” or at least as close to 1 as possible. 

b. Interpolate image matrix to 256 x 256.  Some systems may do this automatically. 

c. Acquire  DWI  along  three  orthogonal  axes  so  that  “isotropic”  or  “trace”  diffusion 

weighted  images  are  generated  for  each  slice  and  b‐value.  “Tetrahedral”  or 

“Gradient Overplus” diffusion encoding schemes are allowed wherein combinations 

of X, Y, and Z gradients are applied  simultaneously  for each orthogonal DWI axis.  

Individual diffusion axes DW  images are not required.   Only the b=0 and DWI trace 

images at each b‐value are required. 

d. Frequency encoding bandwidth may not be under full operator control.    If possible 

use  “maximum  bandwidth”,  or  equivalently  “minimum  fat  shift  per  pixel”,  or  set 

within 1500 to 2500 Hz/pixel range.  

Page 23: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixI

MRIDWIPhantomTestInstructions Page6of809Oct2012

e. It  is preferred to acquire all four b‐values  in a single series.   The collection of all b‐

values is called one “pass.”  However, if on a particular MRI system it is not possible 

to acquire all  four b‐values  in one  series,  it  is allowable  to acquire  these  in  three 

separate sequential series: b=0, 100  followed by b=0, 600 then b=0, 800.   On such 

scanners, these three series constitute “one pass.” DO NOT allow TE to change (e.g. 

set  “OPTIMIZE  TE”  to  “NO”),  and  do  not  allow  the  system  to  adjust  hardware 

settings (e.g. transmit/receive gain) between DWI series.  

NOTE:  The  PHANTOM  PROTOCOL  CALLS  FOR  FOUR  ACQUISITIONS  AT  EACH  b‐

VALUE.   

If  the scanner acquires all b‐values  in a single series, simply  repeat  the DWI  in 

four sequential series  (e.g. Scan #1 b=0, 100, 600, 800; Scan #2 b=0, 100, 600, 

800)… 

If  the  scanner  can  only  acquire  one  b‐value  pair  per  series,  then  perform  12 

sequential scans in order: Scan #1 b=0, 100; Scan #2 b=0, 600; Scan #3 b=0, 800; 

Scan #4 b=0, 100; Scan #5 b=0, 600; Scan #6 b=0, 800; Scan #7 b=0, 100; Scan #8 

b=0, 600; Scan #9 b=0, 800; Scan #10 b=0, 100; Scan #11 b=0, 600; Scan #12 b=0, 

800. 

 

f. Use  site‐preferred  spectral  fat  suppression  technique,  such  as  “SPAIR,” but unlike 

protocol  used  on  patients  DO  NOT  USE  the  Inversion‐Recovery  fat  suppression 

technique “STIR” for DWI Phantom scans 

 

 ADC Map Generation: 

Generate ADC maps for all slices using software tools available on the scanner.   The ADC map 

should be derived from the full set of b‐values = 0, 100, 600, 800.  If the system only acquired 

one pair of b‐values in each series, generate the ADC maps using the b‐value = 0, 800 pair.   If 

possible,  take  a  “screenshot”  of  a  region‐of‐interest  (ROI)  drawn  in  a  central  tube  region 

showing ROI size/location and resultant statistics (e.g. See Figure 3).  The screenshot should be 

simply  saved within  the phantom examination  for  transfer  to ACRIN as DICOM  images.    Site 

certification and QC  calculations will be performed on ADC maps  regenerated at ACRIN  core 

labs; however screenshots are helpful to confirm proper image scaling and equivalence in ADC 

calculation routines. 

 

 

Page 24: Site Imaging Manual ACRIN 6701

ACRIN67

MRIDWI09Oct20

 

 

Figure 2:

support a

701SiteImag

PhantomTe012

: (a) Position

anterior com

gingManual

estInstructio

Figure 1: 

n DWI phan

mponent of t

ons P

MRI of DWI

tom vertica

torso/pelvic 

age7of8

 phantom ic

lly on cente

phased arra

ce water con

er of table w

ay coil, as sh

 

ntents 

with pads an

own in (b).  

Appen

nd/or coil  lo

 

ndixI

 

ad to 

Page 25: Site Imaging Manual ACRIN 6701

ACRIN67

MRIDWI09Oct20

 

Qua

Do not d

shots, to 

Please la

S

T

D

Ty

For  que

imagearc

 

701SiteImag

PhantomTe012

lifying Ima

de‐identify D

the ACRIN I

bel the disk,

ite Name 

rial Name (A

Date of Phant

ype of Phan

 

estions  rela

[email protected]

gingManual

estInstructio

Figure

age Submis

DWI phantom

maging Core

, in permane

ACRIN 6701) 

tom Imaging

tom Imaging

ated  to  D

org.  Enter “A

Attn: A

ons P

 3: Screen ca

ssion 

m images.  S

e Laboratory

ent marker, 

g (DD‐MMM

g (DWI) 

WI  phanto

ACRIN 6701 

Ship All Qu

America

1818 M

Phil

CRIN 6701 D

age8of8

apture of RO

Submit all DW

y on Compac

with the foll

M‐YYY) 

om  qualific

DCE‐MRI/DW

ualificatio

an College o

Market St., S

adelphia, PA

DCE‐MRI/DW

OI on ADC m

WI phantom

ct Disc (CD) 

lowing infor

cation,  plea

WI Qualifica

n Imaging 

of Radiology

Suite 1600 

A 19103 

WI Qualificat

map 

m images, inc

or DVD.   

rmation: 

ase  send 

ation” in the

to: 

tion 

Appen

cluding scre

an  email 

e subject line

ndixI

en‐

to: 

e. 

Page 26: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixII

MRIDCEPhantomTestInstructions1of5AppendixII09Oct2012

 

Appendix II 

 

DCE Phantom Test Instructions 

Page 27: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixII

MRIDCEPhantomTestInstructions2of5AppendixII09Oct2012

   

MRI‐DCE QIBA PHANTOM TEST INSTRUCTIONS 

 

Objective: 

This document provides information regarding the MRI‐DCE QIBA phantom tests. The QIBA 

MRI‐DCE phantom is not currently a commercially available phantom. This phantom will be 

provided by ACRIN to all sites for qualification testing purposes only. 

 

Preparing the QIBA DCE Imaging Sequence Templates: 

1. Open a protocol or folder labeled “ACRIN6701_DCE_templates.” 

2. Create an imaging series using the proper 3D spoiled gradient echo technique based on 

MRI manufacturer  (see right column of Table 1 on of Appendix V). Enter the technical 

parameters  for  the  DCE‐MRI  imaging  protocol  into  the  imaging  series.  Refer  to  the 

parameters specified  in the DCE‐MRI column  in Table 1 of Appendix V of  this manual. 

Temporarily set the number of signal averages (NEX/NSA) to one and the flip angle to 

30.   Note  specific  vendor  required  settings  in  Table  4  of Appendix V.  (GE  users: use 

Turbo=2 setting).  Adjust parameters such as BW, phase matrix, FOV, and % phase FOV 

within the allowed ranges  from Table 1 of Appendix V to ensure that the time  for the 

sequence  is between 5 and 8  seconds.   You will need  to  reference  this  time  later on 

when creating the full protocol. 

3. Save this sequence as “ACRIN6701_DCE_template_human” in the DCE_template folder.  

This template will also be used to construct the human imaging sequence for the trial. 

4. Make  a  copy  of  the  “ACRIN6701_DCE_template_human”  series  and  rename  it  as 

“ACRIN6701_DCE_template_phantom”  

5. Open  “ACRIN6701_DCE_template_phantom”  and make  the  following  changes  to  this 

sequence: 

a. Change imaging plane to coronal 

b. Set frequency direction to S/I and phase direction to R/L. 

c. Set frequency and phase FOV to 360 each.  Percent phase FOV should be 100% 

d. Keep acquisition matrix unchanged, but set reconstruction matrix  to 512 x 512 

(Siemens users: check interpolate). 

e. Remove parallel imaging. 

f. Set number of slices to 26. 

g. Change slice thickness to 4.25mm. 

 

 

Page 28: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixII

MRIDCEPhantomTestInstructions3of5AppendixII09Oct2012

Preparing the QIBA DCE Phantom Imaging Protocol: 

1. Open a new folder/protocol for “ACRIN6701_DCE_phantom.” 

2. Add localizer sequences for body imaging as per institutional standard. 

3. Copy  the  series  labeled  “ACRIN6701_DCE_template_phantom”  and  paste  into  the 

phantom scanning protocol. 

4. Open this series and make the following changes: 

a. Set number of acquisitions/signal averages to 4 

b. GE users: set Turbo=0 

5. Rename this series as “ACRIN6701_Ph_VFA_flip_30.” 

6. Copy this series and change flip angle to 25°.  Siemens users: set acquisition properties 

to “scan without further adjustment”.   

7. Rename this series as “ACRIN6701_Ph_VFA_flip_25”. 

8.  Make  five additional  copies of  this  series.   Successively open each one,  changing  flip 

angle  to  20°,  15°,  10°,  5°,  2°  sequentially.  Save  each  subsequent  series  as 

“ACRIN6701_Ph_VFA_flipXXs” where “XX”  is “20”, “15”, “10”, “05”, and “02”,  for each 

series. 

9. Make a copy of the “ACRIN6701_Ph_VFA15_degree” series.   

10. GE users: open this series and change to Turbo=2 

11. Save/rename this series as “ACRIN6701_Ph_Ratio_map_Array_Coil” 

12. Make a copy of the “ACRIN6701_Ratio_map_Array_Coil” series 

13. Open  this  and  change  to  Body  coil  acquisition.    Siemens  users:  add  comment  to 

technologist re coil change. 

14. Save this series as “ACRIN6701_Ph_Ratio_map_Body_Coil” 

15. Make a copy of the “ACRIN6701_Ph_VFA_30_degrees” series 

16. Open this and make the following changes 

a. Change number of signal averages to 1. 

b. Change number of dynamic acquisitions based on previously noted time for the 

“ACRIN6701_DCE_template_human” acquisition, as indicated below: 

i. 5.0 ‐ 5.5 seconds   65 acquisitions 

ii. 5.6 – 6.5 seconds  60 acquisitions 

iii. 6.6 – 7.5 seconds  50 acquisitions 

iv. 7.6 – 8.0 seconds  45 acquisitions  

c. GE users: set Turbo=2 and MPH option 

d. Siemens  users:  remove  “scan without  further  adjustment”  from  acquisition 

properties. 

17. Save series as “ACRIN6701_Ph_Dynamic”.   

   

Page 29: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixII

MRIDCEPhantomTestInstructions4of5AppendixII09Oct2012

This  completes  the  phantom  protocol  set  up.  Your MRI‐DCE QIBA  phantom  protocol 

should have the following series prepared, in the order shown: 

 

localizer (recommend 3 plane 

T1 Map 30° (tuned, with pre‐scan) 

T1 Map 25° (not tuned, no pre‐scan) 

T1 Map 20° (not tuned, no pre‐scan) 

T1 Map 15° (not tuned, no pre‐scan)   

T1 Map 10° (not tuned, no pre‐scan) 

T1 Map 5° (not tuned, no pre‐scan) 

T1 Map 2° (not tuned, no pre‐scan) 

Ratio map array coil 

Ratio map body coil 

Dynamic gadolinium  (tuned, pre‐scan) 

 

QIBA DCE Phantom Preparation Procedure: 

1. Use  the  surface  array  coil per  your  institutional  routine  for pelvic  imaging  (i.e.  torso‐

array) 

2. Place  the QIBA MRI‐DCE phantom  in  the  center of  the  table, on  top of  the posterior 

elements of the torso‐array coil   

3. Add the anterior torso array coil elements to the top 

4. Landmark to the center of the phantom. 

Initiate  localizer scanning.   Once a satisfactory  localizer  imaging  in  the sagittal or axial 

plane  is obtained, prescribe  the coronal slab  from a central slice.   Continue remaining 

sequences, noting switch to body coil for the “Ratio_map_Body_Coil” series. 

 

QIBA DCE Phantom Scan Procedure 

T1 Mapping Series for DCE Body 

1. Run the localizer scan to identify the phantom (may use either body coil or array coil 

for image reception). 

2. Open the T1 map flip 30°series.  Switch to array coil for image reception.   

3. Prescribe  the  3D  slab  in  the  coronal  plane  through  the  center  of  the  phantom, 

centered in all 3 planes. Pre‐scan, as usual, for this series, then run. 

4. Run the next T1 Map at 25°, copying the position of the 3D slab from the previous 

series. DO NOT pre‐scan. 

5. Run the next T1 Map at 20°, copying the position of the 3D slab from the previous 

series. DO NOT pre‐scan. 

Page 30: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixII

MRIDCEPhantomTestInstructions5of5AppendixII09Oct2012

6. Run the 15°, 10°, 5°, 2° T1 maps in the same manner as above without a pre‐scan.  

 

 

Body Coil Ratio Map Series 

1. Run the “Ratio_map_Array_Coil scan. 

2. Run the “Ratio_map_Body_Coil scan.   Make sure to switch to body coil as receive 

coil.  DO NOT run a pre‐scan. 

Dynamic Series 

1. Switch back to array coil as receive coil.  Run the DYNAMIC SCAN, copying the 

position of the 3D slab from the previous series. Pre‐ scan is allowed for this series.  

 

  

Page 31: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page1of11

 

 

   

Appendix III 

 

MRI Scan 1 

Human Imaging Procedures 

Page 32: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page2of11

 

MRI SCAN 1: Human Imaging Procedures Introduction 

 

The following  is a guideline document depicting the patient preparation and order of  imaging for ACRIN 6701 MRI Scan 1.    Note that this document is divided in to three sections: 

1. Part  I: Anatomic  imaging procedures for sites not using an endo‐rectal coil (eCoil) (see page 3) 

2. Part II: Anatomic imaging procedures for sites using an endo‐rectal coil (eCoil)  (see page 5) 

3. Part III: Non‐eCoil DWI and DCE‐MRI imaging (all sites see page 8)  Sites  should  determine  whether  they  will  follow  Part  I  or  Part  II  of  the  anatomic  imaging procedures, then all sites should proceed to Part III after completion of either Part I or Part II. 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 33: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page3of11

Part I: 

Pre‐ DWI and DCE‐MRI Imaging Procedures 

(For Institutions not Utilizing Endo‐rectal Coil (eCoil) for Anatomic Prostate Imaging) 

 

Note: Institutions utilizing eCoil for anatomic imaging during MRI 1 should skip to Part II 

Preparation  

1. Document participant body weight on the MRI Scan Administration Form 

2. Pre‐draw appropriate amount  (0.1 mmol/kg) of protocol approved gadolinium and 20 

mL normal saline into dual‐chamber power‐injector  

3. Place  intravenous  line  (20G  or  higher),  preferably  in  right  antecubital  fossa,  for 

gadolinium contrast administration,  then document  injection site and catheter size on 

the MRI Scan Administration Form 

4. Place  participant  supine  in MRI  scanner;  connect  IV  to  power  injector;  and  deploy 

scanner‐specific local torso array coils about pelvic region 

5. Landmark participant at the level of the symphysis pubis 

 

Anatomic Image Acquisition 

1. Perform pre‐contrast anatomic imaging as described below:  

Note: Minimal required imaging is listed below, but any pre‐contrast imaging series per 

institutional norms are permitted 

a. Localizer:  per  institutional  routine.  It  is  strongly  recommended  that  a  large  FOV 

sagittal localizer is acquired for the planning of the DCE‐MRI acquisition. See Figures 

1 & 2 on page 10 & 11 of this appendix for proper slab prescription. 

b. High resolution T1‐weighted axial (or oblique axial) imaging: 

Scanning should be tailored to the prostate, with the appropriate use of FOV 

and image matrix values for high resolution prostate imaging. 

Use  institutional  norms  for  type  of  T1  scanning  and  other  acquisition 

parameters.   

Note:  Oblique  axial  imaging  (i.e  perpendicular  to  the  posterior  edge  of  the 

prostate) is acceptable if per institutional norms (see Figure 2, page 11).  Use of 

standard or oblique scanning planes should be uniform throughout examination. 

c. High resolution T2‐weighted axial (or oblique axial) imaging: 

Page 34: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page4of11

Utilization of identical scan prescription (FOV, slice thickness, slice locations) 

with that of T1 imaging is strongly encouraged. 

Note acceptable parameter ranges in Table 1 of Appendix V  

d. High resolution T2‐weighted coronal (or oblique coronal) imaging: 

Utilization  of  oblique  coronal  imaging  plane  (i.e.  parallel  to  the  posterior 

surface of the prostate is acceptable).  

Utilize institutional norms for scanning parameters 

2. Perform any additional standard of care pre‐gadolinium imaging per institutional norms.  

Possible imaging series may include: 

High resolution sagittal T2 imaging of the prostate 

Large  FOV  T1  and/or  T2  imaging  to  document  nodal  or  osseous 

metastatic disease 

Supplemental DWI with eCoil (per institutional norms) 

 

3. Please proceed to Appendix III‐Page 8 for the DWI and DCE‐MRI imaging instructions, at 

the completion of the pre‐gadolinium imaging. 

 

 

   

Page 35: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page5of11

MRI SCAN 1: Imaging Procedures 

Part II: 

Pre‐DWI and DCE‐MRI Imaging Procedures 

(For Institutions Utilizing Endo‐rectal Coil (eCoil) for Anatomic Prostate Imaging) 

 

This portion of the imaging manual is for sites that choose to use eCoil for anatomic imaging of 

the prostate.   Sites  should use  this  section  for  imaging  if  their  institutional norms  for clinical 

exams of the prostate require the use of an eCoil at 3T imaging and the use of non eCoil exams 

would  result  in  imaging  not  deemed  to  be  acceptable  for  clinical  staging  per  institutional 

standards.    Note  that  even  if  an  eCoil  is  normally  utilized  for  anatomic  and/or  functional 

prostate imaging at 3T, sites should consider whether non‐eCoil anatomic imaging may produce 

clinically acceptable  imaging, as  this will  improve patient  throughput during  this study.   After 

completion of MRI Scan 1, sites may review the imaging and determine whether eCoil imaging 

for anatomic prostate evaluation may benefit the patient during MRI Scan 2. 

Sites  that  choose  to use eCoil  imaging  for anatomic prostate evaluation  should  first perform 

eCoil‐based imaging, and then perform non‐eCoil DWI and DCE‐MRI study imaging. 

 

Preparation  

1. Document participant body weight on the MRI Scan Administration Form 

2. Pre‐draw appropriate amount  (0.1 mmol/kg) of protocol approved gadolinium and 20 

mL normal saline into dual‐chamber power‐injector  

3. Place  intravenous  line  (20G  or  higher),  preferably  in  right  antecubital  fossa,  for 

gadolinium contrast administration,  then document  injection site and catheter size on 

the MRI Scan Administration Form 

4. Deploy endo‐rectal coil  

5. Place  participant  supine  in MRI  scanner;  connect  IV  to  power  injector;  and  deploy 

scanner‐specific local torso array coils about pelvic region 

6. Landmark participant at the level of the symphysis pubis 

 

Anatomic Image Acquisition with eCoil 

1. Perform pre‐contrast anatomic imaging as described below:  

Note: Minimal required imaging is listed below, but any pre‐contrast imaging series per 

institutional norms are permitted. 

a. Localizer: per institutional routine 

b. High resolution T1‐weighted axial (or oblique axial) imaging: 

Page 36: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page6of11

Scanning should be tailored to the prostate, with the appropriate use of FOV 

and image matrix values for high resolution prostate imaging. 

Use  institutional  norms  for  type  of  T1  scanning  and  other  acquisition 

parameters.   

Note:  Oblique  axial  imaging  (i.e  perpendicular  to  the  posterior  edge  of  the 

prostate) is acceptable if per institutional norms (see Figure 2, page 11).  Use of 

standard or oblique scanning planes should be uniform throughout examination.  

c. High resolution T2‐weighted axial (or oblique axial) imaging: 

Utilization of identical scan prescription (FOV, slice thickness, slice locations) 

with that of T1 imaging is strongly encouraged. 

Note acceptable parameter ranges in Table 1 of Appendix V 

d. High resolution T2‐weighted coronal (or oblique coronal) imaging: 

Utilization  of  oblique  coronal  imaging  plane  (i.e.  parallel  to  the  posterior 

surface of the prostate is acceptable.   

Utilize institutional norms for scanning parameters 

 

2. Perform  any  additional  standard  of  care  pre‐gadolinium  imaging  per  institutional 

norms.  

Possible imaging series may include: 

High resolution sagittal T2 imaging of the prostate 

Large  FOV  T1  and/or  T2  imaging  to  document  nodal  or  osseous 

metastatic disease 

Supplemental DWI with eCoil (per institutional norms) 

 

Note: DWI performed with eCoil is at the discretion of the imaging site PI. Scanning parameters 

for eCoil DWI series should be based on  institutional norms and need not utilize study‐specific 

DWI parameters.   

Note that DWI imaging without eCoil, per study guidelines, must be performed even if eCoil DWI 

imaging is utilized. 

 

3. Prepare for Non‐eCoil Anatomic Imaging  

Remove endo‐rectal coil from patient. 

Re‐place participant supine in MRI scanner; connect IV to power injector; and re‐

deploy scanner‐specific local torso array coils about pelvic region 

Page 37: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page7of11

Landmark participant at the level of the symphysis pubis 

 

Non‐eCoil Anatomic Image Acquisition  

 

1. Localizer: per institutional routine. It is strongly recommended that a large FOV sagittal 

localizer is acquired for planning the DCE MRI acquisition. See Figures 1 & 2 on pages 10 

& 11 of this appendix. 

2. T2‐weighted axial (or oblique axial) imaging  

The utilization of oblique imaging planes is permitted (see Figure 2, page 11), but 

should be used uniformly throughout all portions of this section. 

Imaging  should be  sufficient  to  localize prostate  for  comparison with DWI and 

DCE‐MRI imaging.  High resolution imaging is not required at this stage.  Rapid T2 

FSE imaging, such as SS‐FSE, HASTE, etc; is acceptable. 

3. Please proceed to Appendix III‐Page 8 for the DWI and DCE‐MRI imaging instructions, 

at the completion of the pre‐gadolinium imaging. 

 

 

   

Page 38: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page8of11

MRI SCAN 1: Imaging Procedures 

Part III: 

DWI and DCE‐MRI Imaging Procedures 

(Non‐eCoil‐ All sites) 

 

The  following  guidelines  are  for  study‐required non‐eCoil DWI  and DCE‐MRI  imaging.   These 

instructions should be followed by each imaging center after completion of either Part I or Part 

II of these instructions. 

 

DWI/DCE Image Acquisition Procedures  

1. Perform Multi‐B value axial (or oblique axial) SE‐EPI DWI (with B values of 0, 100, 600, 

and 800 s/mm2)) 

The utilization of oblique  imaging planes are permitted (see Figure 2, page 11), 

but should be used uniformly throughout all portions of this section. 

See Table 1 in Appendix V for detailed imaging instructions 

Certain scanner systems may not allow the use to run more than one non‐zero B‐

value during DWI.  If Multi‐B  value DWI  imaging  is not permitted on  your MRI 

system, this section should be run in three distinct series.  

Label series accordingly: 

o Series 1: B values 0 & 100 

o Series 2: B values 0 & 600 

o Series 3: B values 0 & 800 

 

2. Prescribe axial (or oblique axial) 3D volume for DCE‐MRI study: 

The utilization of oblique imaging planes is permitted (see Figure 2, page 11), but 

should be used uniformly throughout all portions of this section. 

Axial volume of 32 x 5mm slices must be prescribed; do not center imaging over 

the prostate. Rather, the prostate should be at the lower portion of the imaging 

volume to minimize vascular in flow affects. See Figures 1 & 2 on Pages 10 & 11 

of  this appendix  for details  regarding  slab placement.   Perform all  series using 

same exam prescription.  

Do  not  perform  pre‐scan/preparation  between  series;  unless  indicated 

otherwise  

See Table 1 of Appendix V for detailed imaging information. 

 

3. DCE‐MRI Imaging Series:  

Page 39: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII 09Oct2012Page9of11

Multi‐flip‐angle T1 mapping protocol (30° , 25°, 20°, 15°, 10°, 5°, 2°) 

Array Body coil only imaging for receive coil bias field calculation 

Dynamic Enhanced Imaging  

 

Note on SAR: 

In  the event of a  SAR error  the  technologist  should only make adjustments  to  the  flip angle; 

please do not make any additional parameter adjustments.  

If, on a subject’s initial scan, a SAR error requires dynamic enhanced imaging to be run at a flip 

angle  less  than 30°  the  flip angle adjustment  should be noted and  this parameter  should be 

replicated, if possible, at all subsequent visits.  

 

4. Contrast Administration for DCE Imaging  

The gadolinium agents MultiHance, Eovist, and Vasovist are NOT permitted for 

this study 

The same brand and dose of contrast, same  IV site, and same  rate of contrast 

and  saline  flush administration  should be maintained  for both  the  test and  re‐

test examinations. 

Contrast  agent  should  be  administered  in  a  dynamic  fashion  beginning  30‐40 

seconds after the start of the dynamic enhanced imaging.   

Contrast agent should be administered at a dose of 0.1 mmol/kg body weight 

and rate of 2 mL/second, followed by a 20 mL saline flush, also at 2 mL/second. 

Document the amount of contrast administered on the MRI Scan Administration 

Form  

5. Post gadolinium imaging, per institutional norms, may be performed after completion 

of the dynamic enhanced imaging series.

Page 40: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII09Oct2012 Page10of11

Figure 1: DCE MRI Slab Prescription for Axial Imaging 

 

  Scout View            Correct Slab Placement (Circle=Prostate C=eCoil) 

 

 

          Incorrect (Slab too thin)               Incorrect (Slab set too low) 

Page 41: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIII

MRIScan1:HumanImagingProceduresAppendixIII09Oct2012 Page11of11

Figure 2: DCE MRI Slab Prescription for Oblique Axial Imaging 

 

ScoutView CorrectSlabPlacement(Circle=ProstateC=eCoil)

 

          Incorrect (Slab too thin)               Incorrect (Slab set too low) 

Page 42: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV 09Oct2012Page1of7

 

   

Appendix IV 

 

MRI Scan 2 

Human Imaging Procedures 

Page 43: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV 09Oct2012Page2of7

MRI SCAN 2: Imaging Procedures 

 

The following guidelines are for study‐required non‐eCoil DWI and DCE‐MRI imaging for ACRIN 

6701 MRI Scan 2.  Note that MRI Scan 2 is not for clinical use unless deemed necessary by the 

local  site  radiology  PI.    In  such  cases,  the  local  site  radiology  PI  should  indicate  to  the MRI 

technologist  prior  to  the  examination  which  additional  imaging  should  be  performed.  

Specifically,  endo‐rectal  coil  (eCoil)  imaging  should  not  be  performed  at  this  time  unless 

explicitly indicated by the site PI to repeat non diagnostic anatomic imaging series obtained at 

the time of MRI Scan 1. 

 

Note:  this MRI  scan visit  includes a “coffee break” where  the  technologist will be  required  to 

briefly remove the participant from the scan table.   

Preparation  

1. Document participant body weight on the MRI Scan Administration Form 

2. Pre‐draw appropriate amount  (0.1 mmol/kg) of protocol approved gadolinium and 20 

mL normal saline into dual‐chamber power‐injector  

3. Place  intravenous  line  (20G  or  higher),  preferably  in  right  antecubital  fossa,  for 

gadolinium contrast administration,  then document  injection site and catheter size on 

the MRI Scan Administration Form 

4. Place participant  supine  in MRI  scanner; and deploy  scanner‐specific  local  torso array 

coils about pelvic region. Do not connect the participant to the power  injector at this 

time. 

5. Landmark participant at the level of the symphysis pubis 

 

 Non‐eCoil Anatomic Image Acquisition 

1. Localizer: per institutional routine. 

2.  T2‐weighted axial (or oblique axial) imaging  

The utilization of oblique imaging planes is permitted (see Figure 2, page 7), but 

should be used uniformly throughout all portions of this section. 

Imaging  should be  sufficient  to  localize prostate  for  comparison with DWI and 

DCE‐MRI imaging.  High resolution imaging is not required at this stage.  Rapid T2 

FSE imaging, such as SS‐FSE, HASTE, etc; is acceptable. 

3. Additional pre‐gadolinium anatomic imaging series for clinical care (if required) may be 

Page 44: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV 09Oct2012Page3of7

performed at this time. 

 

DWI/DCE MRI Image Acquisition Procedures 

1. Perform “Pre‐Coffee Break” Multi‐B value axial  (or oblique axial) SE‐EPI DWI  (with B 

values of 0, 100, 600, and 800 s/mm2)) 

The  utilization  of  oblique  imaging  planes  is  permitted,  but  should  be  used 

uniformly throughout all portions of this section. 

See Table 1of Appendix V for detailed imaging instructions 

Certain scanner systems may not allow the user to run more than one non‐zero 

B‐value during DWI.  If Multi‐B value DWI  imaging  is not permitted on your MRI 

system, this section should be run in three distinct series.  

Label series accordingly: 

o Series 1: B values 0 & 100 

o Series 2: B values 0 & 600 

o Series 3: B values 0 & 800 

 

2. Implement a “Coffee Break” 

Remove the participant from the MRI table for approximately 1‐2 minutes  

Place participant supine on the MRI table 

Connect IV to power injector; and deploy scanner‐specific local torso array coils about pelvic region 

Landmarked at the level of the symphysis pubis. 

 3. Imaging Prior to Post “Coffee Break” DWI Series 

Localizer: per institutional routine. It is strongly recommended that a large FOV 

sagittal localizer is acquired for planning the DCE MRI acquisition (see Figure 2, 

page 7). 

T2‐weighted axial (or oblique axial) imaging  

o The utilization of oblique imaging planes is permitted (see Figure 2, page 7), 

but should be used uniformly throughout all portions of this section. 

o Imaging should be sufficient to localize prostate for comparison with DWI 

and DCE‐MRI imaging.  High resolution imaging is not required at this stage.  

Rapid T2 FSE imaging, such as SS‐FSE, HASTE, etc; is acceptable. 

 

4. Perform “Post‐Coffee Break” Multi‐B value axial (or oblique axial) SE‐EPI DWI (with B 

Page 45: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV 09Oct2012Page4of7

values of 0, 100, 600, and 800 s/mm2)) 

The  utilization  of  oblique  imaging  planes  is  permitted,  but  should  be  used 

uniformly throughout all portions of this section. 

See Table 1 in Appendix V for detailed imaging instructions 

Certain scanner systems may not allow the user to run more than one non‐zero 

B‐value during DWI.  If Multi‐B value DWI  imaging  is not permitted on your MRI 

system, this section should be run in three distinct series.  

Label series accordingly: 

o Series 1: B values 0 & 100 

o Series 2: B values 0 & 600 

o Series 3: B values 0 & 800 

5. Prescribe axial (or oblique axial) 3D volume for DCE‐MRI study: 

The utilization of oblique imaging planes is permitted (see Figure 2, page 7), but 

should be used uniformly throughout all portions of this section. 

Axial volume of 32 x 5mm slices must be prescribed; do not center imaging over 

the prostate. Rather, the prostate should be at the lower portion of the imaging 

volume to minimize vascular in flow affects. See Figures 1 & 2 on Pages 6 & 7 of 

this appendix for details regarding slab placement.  Perform all series using same 

exam prescription.  

Do  not  perform  pre‐scan/preparation  between  series;  unless  otherwise 

indicated  

See tables in Appendix V for detailed imaging information. 

 

6. DCE‐MRI Imaging Series:  

Multi‐flip‐angle T1 mapping protocol (30° , 25°, 20°, 15°, 10°, 5°, 2°) 

Array Body coil only imaging for receive coil bias field calculation 

Dynamic Enhanced Imaging  

 

Note on SAR: 

In  the event of a  SAR error  the  technologist  should only make adjustments  to  the  flip angle; 

please do not make any additional parameter adjustments.  

 

If, on a subject’s initial scan, a SAR error requires dynamic enhanced imaging to be run at a flip 

angle  less  than 30°  the  flip angle adjustment  should be noted, and  this parameter  should be 

replicated, if possible, at all subsequent visits.  

 

Page 46: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV 09Oct2012Page5of7

7. Contrast Administration for DCE Imaging  

The gadolinium agents MultiHance, Eovist, and Vasovist are NOT permitted for 

this study 

The same brand and dose of contrast, same  IV site, and same  rate of contrast 

and  saline  flush administration  should be maintained  for both  the  test and  re‐

test examinations. 

Contrast  agent  should  be  administered  in  a  dynamic  fashion  beginning  30‐40 

seconds after the start of the dynamic enhanced imaging.   

Contrast agent should be administered at a dose of 0.1 mmol/kg body weight 

and rate of 2 mL/second, followed by a 20 mL saline flush, also at 2 mL/second.  

Document the amount of contrast administered on the MRI Scan Administration 

Form  

8. Post gadolinium imaging, per institutional norms, may be performed after completion 

of the dynamic enhanced imaging series.

Page 47: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV09Oct2012 Page6of7

Figure 1: DCE MRI Slab Prescription for Axial Imaging 

 

  Scout View        Correct Slab Placement           (Circle=Prostate C=eCoil)  

                                         

Incorrect (Slab too thin)                  Incorrect (Slab set too low)                             

Page 48: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixIV

MRIScan2:HumanImagingProceduresAppendixIV09Oct2012 Page7of7

Figure 2: DCE MRI Slab Prescription for Oblique Axial Imaging 

 

ScoutView CorrectSlabPlacement(Circle=ProstateC=eCoil)

 

        Incorrect (Slab too thin)                  Incorrect (Slab set too low) 

Page 49: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page1of7

 

   

Appendix V 

 

Image Acquisition 

Parameters 

Page 50: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page2of7

ACRIN 6701 MRI Image Acquisition Parameters 

 

This  appendix  provides  detailed  instructions  regarding  key  imaging  parameters  for  the 

experimental  imaging series  in the protocol.   The  first table represents the parameter setting 

required for T2‐weighted imaging, diffusion weighted imaging, and DCE‐MRI. Where applicable, 

vendor‐specific nomenclature  is provided  in the table.   For DCE‐MRI, the  first table highlights 

the general acquisition parameters for setting the volumetric  imaging slab.   Details specific to 

each of the three sub‐portions of the DCE‐MRI protocol (variable flip angle [VFA] mapping, body 

coil ratio map acquisition, and dynamic enhanced  imaging) are provided  in the second section 

of this appendix. 

 

Page 51: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page3of7

Table 1: Key Acquisition Parameters for T2‐weighted, DWI, and DCE‐MRI  

Parameter T2‐weighted 

(non eCoil) *(1) T2‐weighted 

(with eCoil) *(1) DWI  DCE‐MRI *(2) 

Sequence type  FSE/TSE  FSE/TSE  DW SE‐EPI  Spoiled GRE S *(3): Fl3d_vibe 

P *(3): FFE G *(3): 3D FSPGR 

2D or 3D sequence  2D  2D  2D  3D 

Slice orientation  Axial *(4)  Axial *(4)  Axial *(4)  Axial *(4) 

Frequency direction  A/P  A/P  R/L  R/L 

Phase direction  L/R  L/R  A/P  A/P 

FOV – frequency  180‐260 mm  120‐180 mm  320‐400 mm *(5) 320‐400 mm *(5) 

FOV ‐ phase  180‐260 mm  120‐180 mm  260‐320 mm*(5) 260‐320 mm *(5) 

Partial FOV  100%  100%  75‐85% *(5)  75‐85% *(5) 

Phase oversampling  As needed  As needed  No  No 

Matrix – frequency (acquired)  256‐512  256‐512  128‐192  256 

Matrix – phase (acquired)  160‐384  160‐384  128‐192  128‐192 

Reconstruction Matrix  512 x 512  512 x 512  256 x 256  256 x 256 

Parallel Imaging Factor  Per routine  Per routine  2x  2x 

Fat‐suppression  None  None  Yes *(6)  None 

Spatial Saturation  Per routine  Per routine  None  None 

TR  3000‐10000 ms  3000‐10000 ms  > 4,000 ms  3‐10 ms *(7) 

TE  70‐140 ms  70‐140 ms  <=90 ms  < 3 ms *(7) 

Echo Train Length  Per routine  Per routine  N/A  N/A 

Flip Angle  90‐180 degrees  90‐180 degrees  90 degrees  see next section 

Page 52: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page4of7

(*) Parameter Notes: 

1. Axial  (or  oblique  axial)  T1  and  coronal  (or  oblique  coronal  T2)  parameters  are  not  specified.    These may  be  performed  per 

institutional routines.  It is recommended that the slice prescription for T1 (oblique) axial series follow that of T2 (oblique) axial 

imaging.  (Oblique) coronal T2 imaging should be run with a parameter set similar to that of the (oblique) axial T2 data set. 

2. Includes geometric and general acquisition parameters for each of the three components of the DCE‐MRI portion of the study.  

Use these settings for all aspects of the DCE‐MRI acquisition.  See second section on page 7 for details specific to each portion 

(VFA mapping, coil ratio mapping, and dynamic enhanced imaging.) 

3. Vendor‐specific notation:  S = Siemens, P = Philips, G= GE. 

4. May use oblique axial  (parallel  to posterior edge of prostate). Use of oblique axial plane  should be consistent  throughout all 

imaging series. 

5. Ensure  that entire AP width of pelvis  is  included  to avoid parallel  imaging artifacts.   Recommend  increased  frequency FOV as 

needed such that at least 85% rectangular phase FOV reduction is achievable without phase wrap. 

6. Use either Short T1 Inversion Recovery (STIR) or Spectral Adiabatic Inversion Recovery (SPAIR) fat suppression. 

7. Use minimum TR and TE achievable based on other settings, but TR should be no less than 3 ms. 

8. It  is preferred to acquire all four b‐values  in a single series.    If your system cannot collect multiple b‐values  in a single “pass”, 

then  three  separate  series  should  be  acquired  and  named  as  follows:  “DWI_b100”:  b=0  &  100;  “DWI_b600”:  b=0  &  600; 

“DWI_b800”: b=0 & 800.  Do NOT allow TE to change (e.g., set “OPTIMIZE TE” to “NO”), and do not allow the system to adjust 

hardware  settings  (e.g.,  transmit/receive  gain)  between  DWI  series.    See  instructions  in  Appendix  1  for  more  details.

Parameter T2‐weighted1 (non eCoil) 

T2‐weighted1 (with eCoil) 

DWI  DCE‐MRI2 

Readout BW [S/P: (Hz/Px) G: (kHz)]  Per routine  Per routine  S/P: 1500‐2500 G:  ±128 kHz or “maximum” 

S/P: 400‐600 G: ±50 ‐ ±100 kHz 

b‐values (s/mm2)  N/A  N/A  0, 100, 600, 800 *(8)  N/A 

Slice thickness (acquired)  2.5‐4 mm  3‐5 mm  5 mm  5 mm 

Number of slices  Variable  Variable  Variable  32 

Slice gap   mm   mm  0  N/A 

No. of excitations or averages  Per routine  Per routine  3‐6  see next section 

Sequence acquisition time  Per routine  Per routine   7minutes  see next section 

Page 53: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page5of7

Table 2:  Vendor‐Neutral Acquisition Parameters for DCE‐MRI     

 

Vendor‐Neutral Settings  Variable Flip Angle Mapping  Body Coil Ratio Map  Dynamic Enhanced Imaging 

Coil  Torso array  Part 1: Body coil Part 2: Torso array

Torso array 

Number of separate series  7  2  1 

Number of dynamic acquisitions (“phases”) 

N/A  N/A  40‐60 *(1) 

Pre‐scan prior to  acquisition(s)  For first flip angle only  Yes  Yes 

Save scan(s) as  Separate series  Separate series  Single series *(2) 

Name series as  “VFA_flip_XX”  Part1: “Ratio map body” Part 2: “Ratio map torso”

“Dynamic_Gad” 

Flip angle(s)  2, 5, 10, 15, 20, 25, 30  15  30 

Acquisitions/NEX  4  4  1 

Time per acquisition  20‐40 seconds  20‐40 seconds  5‐8 seconds *(1) 

Total time  2‐5 minutes  1.0‐1.5 minutes  4‐7 minutes *(1) 

Contrast (saline) amount  N/A  N/A  0.1 mmol/kg (max. 20 mL) 

Injection rate  N/A  N/A  2 mL/second 

Start injection time  N/A  N/A  15‐20s  after scan initiation 

 

Page 54: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page6of7

DCE‐MRI  Notes: 

1. Dynamic scan time must fall between 5 and 8 seconds.   Minimum number of phases of dynamic scan should be 45, which 

corresponds to a maximum total scan time of approximately 6minutes for a per phase temporal resolution of 8 seconds.  For 

shorter temporal resolutions, scan time should be between 5‐6 minutes.  Recommended settings are as follows:   

 

Dynamic Temporal Resolution  Total Number of Phases  Total Scan Time 

7.6‐8 seconds  45  5:25‐6:00 

6.6‐7.5 seconds  50  5:30‐6:15 

5.6‐6.5 seconds  60  5:36‐6:30 

5.0‐5.5 seconds  65  5:25‐5:58 

 

   

Table 3: 

Page 55: Site Imaging Manual ACRIN 6701

ACRIN6701SiteImagingManual AppendixV

ImageAcquisitionParametersAppendixV 09Oct2012Page7of7

Table 4:  Vendor‐Specific Acquisition Parameters for DCE‐MRI   

 

Please ensure that the correct vendor‐specific settings are employed as described below: 

Vendor‐specific settings  Variable Flip Angle Mapping  Body Coil Ratio Map  Dynamic Enhanced Imaging 

GE       

User CV/Advanced  Turbo=0  Turbo=2  Turbo=2 

User CV/Advanced   Slice Resolution = 100%  Slice Resolution = 100%  Slice Resolution = 100% 

Imaging options  EDR  EDR  EDR, MPH 

SCIC / PURE  Off  Off  Off 

       

Philips       

CLEAR  Off  Off  Off 

Partial Echo  Off  Off  Off 

Shifted Echo  Off  Off  Off 

Half Scan  Yes  Yes  Yes 

   Factor Y  0.65  0.65  0.65 

   Factor X  0.85  0.85  0.85 

       

Siemens       

Partial Fourier – phase  Off  Off  Off 

Partial Fourier ‐ slice  Off  Off  Off 

Phase Resolution  60%‐70%  60%‐70%  60%‐70% 

Slice resolution  60%‐70%  60%‐70%  60%‐70% 

Normalize  Off  Off  Off 

Elliptic/Centric Scanning  Off  Off  Off 

RF Pulse Type  Normal  Normal  Normal 

Gradient Mode  Fast  Fast  Fast