Selecting the OptimumSample Introduction System for ICP‐OES
AAPFCO 2017 ANNUAL MEETINGFebruary 21st, New Orleans , LA
Dr. Sergei Leikin, Texas Scientific Products LLC
ICP-AE Spectrometer$70,000-100,000
ICP Nebulizer $400-1500
Investing into ICP Technology
Analytical Nebulizers
Analytical Nebulizers:Are these THE SAME TYPE?
CONCENTRIC NONCONCENTRIC
SL1
Slide 4
SL1 Sergei Leikin, 6/22/2016
What is the “Industry Standard”?
Concentric Nebulizer – still the most commonly used
Why ‐ it “comes” with an instrument
Is it the best for your application?
What are the Fertilizer Samples ?
0.11 M Ammonium Citrate and 0.033 M Disodium EDTA pH neutral Solution
High Particulate Content !
Concentric Nebulizer Configuration:Meinhard, Conikal, Seaspray, Aerosalt
Good nebulization efficiency and precision
Low tolerance to particles of only < 75 m
Small Sample Channel ID (a): 0.15-0.25 mm
Concentric Nebulizer Configuration:Typical Blockages
Blocked Gas Channel
Sample material deposits at the orifice (José‐Luis Todoli, Jean‐Michel Mermet, Liquid Sample Introduction in ICP Spectrometry: A Practical Guide, 2011).
What is an Alternative to an “Industry Standard?”
A NonConcentric Design
Babington V‐Groove Type
Parallel Path
V‐Groove Type Nebulizers
Glass or Quartz
Inert
Vee Spray
What are the Performance Challengesfor NonConcentric Nebulizers?
Sensitivity and
Precision!
Opti‐Series Nebulizers:Design Details
Gas
Liquid Sample
Mist
Sample Orifice
Gas Orifice
OptiSolids XL™ Nebulizer Nonconcentric Configuration
Sample ChannelID = 1.0-2.0 mml
Gas Channel
Handles VERY large particulates over 1000 μm and high amount of TDS up to 30%
Good precision and sensitivity comparable to Concentric
High Salts/High TDS
OptiSolids XL™ Nebulizer: Sample Introduction
1.0 mm ID 0.5–0.75 mm
Sample Introduction System Setup for the ICP-OES Fertilizer Extracts Analysis
OptiSolids XL™ Nebulizer Cyclonic Spray Chamber
High Salt Sample Analysis Data
Dilution Factor – 1:10
0.0720.163RSD, %
Data is a courtesy of an Oilfield Labs of America
Data is a courtesy ofUniversity of KY, Division of Regulatory Services Fertilizer Program
Represented by analyses of 210 fertilizer extracts over two months period
Concentration ranges, mg/L :K 20 -1000; P 2 – 500
Spectro BLUE SOP (Radial) ICPOptiSolids XL and Concentric Nebulizers
Calibration Curve for K 769.89 nm Analytical Line
Calibration Curve for K 769.89 nm Analytical Line
0.999968
Calibration Data for K 769.89 nm Analytical Line
Precision Comparison
0
0.5
1
1.5
2
2.5
K 766.4 K 769.9 P 177.5 P 178.2 P 213.6 P 214.9
RSD, %
OptiSolids XLConcentric
Precision Comparison forSc Internal Standard
0
0.2
0.4
0.6
0.8
1
1.2
Sc 361 nm
RSD, %OptiSolids XL Concentric
Cost Comparison: OptiSolids XL™ ‐ Concentric
Lifetime:Concentric ‐ 2 per 1 year OptiSolids XL™ ‐ 1 over 2 years
Cost estimation over 2 years:
Concentric $750x2x2= $3000OptiSolids XL™ $565
Analytical Benefits of the OptiSolids XL™Nebulizer for Fertilizer Extracts Analysis