Upload
phungtu
View
258
Download
2
Embed Size (px)
Citation preview
Scanning Electron MicroscopyBasics and Applications
Dr. Julia DeuschleStuttgart Center for Electron Microscopy
MPI for Solid State Research
Room: 1E15, phone: 0711/ 689-1193
email: [email protected]
Outline
J. Deuschle Scanning Electron Microscopy SS 2017
1.Setup and Instrumentation
2.Electron-Matter-Interactions
3. X-Ray Analysis
4. Literature:Goldstein et al.: Scanning Electron Microscopy And X-Ray Analysis,Springer Verlag 2003Reimer und Pfefferkorn: Rasterelektronenmirkoskopie,Springer Verlag 1973
1.1. The electron column
Additional components:
• vacuum system (pre-pump and turbo molecular pump for
chamber, ion getter pumps for column)
• electronic controls (high voltage, lens current etc.)
• software controls (software interface for the operator to
control the microscope)
• stage (for sample mounting and positioning)
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.2. The electron gun
• generation of „free“ electrons
• acceleration of the electrons
• collimating or focusing of the beam
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.2. Electron guns
J. Deuschle Scanning Electron Microscopy SS 2017
1.2.1 Tungsten Hairpin Cathode
• Thermoionic electron source, i.e. the emission is due to heating of the cathode• The emitted electrons are focussed into a bundle by the grid cap (= beam formation) and „crossover“. • The electrons are then accelerated towards the anode and enter the column.
1.2. Electron guns
J. Deuschle Scanning Electron Microscopy SS 2017
Emission of Tungsten Hairpin Cathode
For a stable emission and hence a stable electron beam with constant beam current, the current through the filament must be sufficiently high, i.e. operated in saturation.
1.2. Electron guns
J. Deuschle Scanning Electron Microscopy SS 2017
1.2.2 Lanthanum-Hexaboride-Cathode
• Thermoionic electron source
• lower work function compared to W, thus higher emission at the same temperature
• longer lifetimes but more expensive
•Degradation due to oxid formation and evaporation
LaB6 crystal mounted on C or Re, 500 x100 µm
Failed tip
1.2. Electron guns
J. Deuschle Scanning Electron Microscopy SS 2017
1. Cold field emission gunEmission is due to strong electrostatic fields, which concentrate at the tip. The W-tip apex is typically several nm wide. UHV conditions are needed to allow tunneling of electrons.
2. Schottky field emission gunA combination of heating and electrostatic fields are used to set electrons free from a small region of 10-30 nm. A ZrO2-coating on the W-tip further lowers the work function. An additional suppresor grid is used to bundle the electrons to a beam.
1.2.3 Field-Emission Guns
1 2
1.2. Electron guns: comparision
J. Deuschle Scanning Electron Microscopy SS 2017
Emitter Type Thermionic Thermionic Schottky TFE Cold FE
Tip image comparison
Cathode Material W LaB6 ZrO/W (100) W (310)
Operating Temp (K) 2700 1800 1800 300
Cathode Radius (nm) 60,000 10,000 <1000 <100
Effective Source Radius (µm) 25 10 0.015 0.0025
Emission Current Density (A/cm2) 3 30 5300 17,000
Total Emission Current (µA) 200 80 200 5
Brightness (A/cm2.sr.kV) 1x104 1x105 1x107 2x107
Maximum Probe Current (nA) 1000 1000 10 0.2
Energy Spread @ Cathode (ev) 0.59 0.4 0.31 0.26
Energy Spread @ Source Exit (eV) 1.5 – 2.5 1.3 – 2.5 0.35 – 0.7 0.3 – 0.7
Beam Noise (%) 1 1 1 5 – 10
Emission Current Drift (%/h) 0.1 0.2 <0.5 5
Operating Vacuum (hPa/mbar) <10-5 <10-6 <10-8 <10-10
Typical Cathode Life (h) 100 >1000 >5000 >2000
Cathode Regeneration (h) None None None 6-12
Sensitivity to External Influences Minimal Low Low High
Stability Standard High Very high Low
X-ray analysis EDS / WDS EDS / WDS EDS / WDS EDS Source: micro-to-nano
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.3. Electron lenses
1. Condensor lens(es)to demagnify the beam and adjust the spot size to the experimental needs
2. Objective lensto focus the beam on the sample surface
1.3 Electron lenses
J. Deuschle Scanning Electron Microscopy SS 2017
Built up of a ferromagnetic mantle, which contains copper windings
When entering the inhomogeneous magnetic field inside the lens the electrons are bend towards the optical axis and the beam is focused. The electron path is a spiral.
1.3.1 Electromagentic lenses
1.3 Electron lenses
J. Deuschle Scanning Electron Microscopy SS 2017
trajectory
1.3.2 Electro-static lenses
Principle of a “Einzel”-lens:
The combination of 3 tubes at different potentials acts as focusing device. The electrons travel towards positive potentials and are repelled by negative potentials.
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.3. Electron lenses
1. Condensor lens(es)to demagnify the beam and adjust the spot size to the experimental needs
2. Objective lensTo focus the beam on the sample surfacedesign must also containscanning coils, stigmator and the final aperture.
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
Scanning coils
for adjusting the position of the beam on the sample surface in x- and y-direction
Image formation
the beam is moved step by step. First along on „line“ in x-direction, then
one step down in y-direction etc.
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.3. Electron lenses
1. Condensor lens(es)to demagnify the beam and adjust the spot size to the experimental needs
2. Objective lensTo focus the beam on the sample surfacedesign must also contain scanning coils, stigmator and the final aperture.
1.4 Apertures
J. Deuschle Scanning Electron Microscopy SS 2017
Aperture size is important for controlling the beam and imaging parameters as follows:
• determines the probe current
• controls the image depth of focus
• minimizies the effects of aberrations
The beam is characterized by
1. acceleration voltage2. probe current
3. convergence angle4. Spot size
1.4 Apertures
J. Deuschle Scanning Electron Microscopy SS 2017
Aperture size
determines the probe currentin limiting the amount of electrons passing through
1.4 Apertures
J. Deuschle Scanning Electron Microscopy SS 2017
Aperture size
controls the image depth of focus
Source: ammrf
1.4 Apertures
J. Deuschle Scanning Electron Microscopy SS 2017
Aperture size
minimizes the effects of aberrations
Spherical aberration ds and aperture diffraction dd cause the point spot to blur to an enlarged spot of size d.
ds= 0.5Cs3
dd= 0.61/
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.3. Electron lenses
1. Condensor lens(es)to demagnify the beam and adjust the spot size to the experimental needs
2. Objective lensTo focus the beam on the sample surfacedesign must also contain scanning coils, stigmator and the final aperture.
1.5 Stigmator
J. Deuschle Scanning Electron Microscopy SS 2017
Astigmatismof lenses is due to machining inaccuracies,
material or winding imperfections
When the probe cross-section that hits the surface is not circular, steaking of the image
features appears and needs to be corrected
Spot shape Octupole stigmatorcorrects asymmetric probe shape into a circular probe
J. Deuschle Scanning Electron Microscopy SS 2017
1.5 Stigmator
Examples for astigmatic images and corrections
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
1.6. Electron detectors
1. Setup and Instrumentation
J. Deuschle Scanning Electron Microscopy SS 2017
Source: wikipedia
1.6.1 Everhart-Thornley detector
usable for SE and BSE detection depending on the bias of the grid
1.6.2 BSE (semiconductor) detector
BSE hitting the detector create electron-hole pairs, which are seperated. The
current is proportinal to the energy of the incident electron.
J. Deuschle Scanning Electron Microscopy SS 2017
1. Setup and Instrumentation
View inside the SEM chamber
BSE-detector:angular disc inserted betweenpolepiece and samlpe
SE-detector:positioned at the side of the chamber/sampleCollector grid is biased for effiecient collection of electrons
2. Electron-Matter-Interactions
J. Deuschle Scanning Electron Microscopy SS 2017
Signal generation upon beam-solid-interaction
Signal used for imaging
Signal used for spectroscopy, i.e. chemical analysis of the material
Source: ammrf
J. Deuschle Scanning Electron Microscopy SS 2017
2.1 The Interaction Volume
Vizualizing the interaction volume
PMMA sample irradiated for a certain time with a 20 kV beam, which has been chemically etched for successively longer time increments.
Depending on the energy deposition in the material, the molecular bonds weaken, resulting in faster etching rates than the surrounding material.
C, Z=6
J. Deuschle Scanning Electron Microscopy SS 2017
The interaction volume is much lager than the beam diameter.It depends on
acceleration voltageThe higher the acceleration voltage the bigger the interaction volume will be
sample materialThe higher the amount of high atomic number elements inside the sample the smaller the interaction volume will be.
10 kV
20 kV
30 kVFe, Z=26
U, Z=91
Simulation of the interaction volume
2.1 The Interaction Volume
J. Deuschle Scanning Electron Microscopy SS 2017
Vizualizing the interaction volume as a function of atomic number
The possibility for backscattering of PE increases with increasing atomic number. Thus, the yield of BSE can be used for z-contrast imaging, i.e. material contrast.
N2
Z=7Ar Z=18
Gases „glow“ upon entering of the electron beam
2.1 The Interaction Volume
J. Deuschle Scanning Electron Microscopy SS 2017
2.2 Signals used for imaging
Primary electrons (PE): Incident (= beam) electrons hitting the sample
Secondary electrons (SE):Electrons that are created from inelastic scattering processes with beam electrons.They possess low energies and can only escape from shallow depths
Backscattered electrons (BSE): Incident electrons, that undergo elastic scattering and escape from the surface again.
J. Deuschle Scanning Electron Microscopy SS 2017
2.2 Signals used for imaging
surface
BSE
BSE
to detector
envelope
range
escape depth
chamber
SE 3
SE 1
SE 2
J. Deuschle Übungen Materialcharakterisierungs- und Testmethoden Sommersemester 2010
acquired using SE
3D appearance due to „edge effect“
Useful imaging mode for topography
2.3 Examples
How does the signal type affect an image?
acquired using BSE
inclusions with different composition become more visible
useful imaging mode for z-contrast
J. Deuschle Übungen Materialcharakterisierungs- und Testmethoden Sommersemester 2010
How does the signal type affect an image?
2.3 Examples
acquired using SE
Surface topography is clearly visible
acquired using BSE
different compositions can be seen
J. Deuschle Scanning Electron Microscopy SS 2017
2.3 Examples
beam incident angle = 0° Tilted: beam incident angle ≈ 80°
How does geometry affect an image?
SE are emitted only from very shallow depth
increasing the beam angle will raise the SE yield
Surface topography leads to changes in beam incident angle, thus the different SE yields will give „contrast“
J. Deuschle Scanning Electron Microscopy SS 2017
2.3 Examples
Topological contrast and edge effect
Cleavage surface
J. Deuschle Scanning Electron Microscopy SS 2017
3. X-Ray Analysis
Combining SE- and BSE- signal provides complementary information a more complete picture can be achieved
acquired using BSEacquired using SE
For a “full” characterization of the material, spectroscopy methods, using x-ray signals, can be used to add chemical information.
J. Deuschle Scanning Electron Microscopy SS 2017
3. X-Ray Analysis
3.1. Creation of x-rays under electron beam irradiation
Characteristic x-rays are generated upon transitions between subshells. The energy difference is characteristic for each element, so detection of the energy of the x-ray photon gives information about the elements present in the sample.
J. Deuschle Scanning Electron Microscopy SS 2017
3. X-Ray Analysis
3.2. Energy Dispersive Spectroscopy (EDS)
Capturing of a photon in the detection system creates a current pulse, which is analyzed. The height of the pulse is proportional to the energy of the incoming photon.
EDX: a complete spectrum is obtained from one measurement.
fast, but poor energy resolutionmostly used for qualitative analysisin scanning mode elemental distribution maps can be acquired
J. Deuschle Scanning Electron Microscopy SS 2017
3. X-Ray Analysis
3.3. Wavelength Dispersive Spectroscopy (WDS)
The x-ray photons emitted from the sample undergo a diffraction process, which separates one specific wavelength, for which the Bragg condition is met. The “proper” photons are focused on the detector and captured for analysis.
WDX:Only one element can be measured at a time
Slow and complicated, but with 10 x better resolution than EDXmostly used for quantitative analysis or trace element identification
J. Deuschle Scanning Electron Microscopy SS 2017
3. X-Ray Analysis
Quantification of a WDX-spectrum
The intensities (counts per second) measured on the sample of unknown composition can be compared to intensities from a standard with known composition.Thus, the weight fractions of elements in the sample can be calculated.