Frequently Asked Questions
www.reflexnow.com
FAQs
XRF Fundamentals 3
Introduction 3
Safety 3
When to use Soil Mode or Geochem Mode 3
Geochem Mode 3
Soil Mode 3
Sample Preparation 3
Instrument 4
Configuration 4
Ambient Pressure 4
S/W version 4
F/W version 4
Filter wheel failure 5
Code 923, Code 1000 & MCA Error 5
Incorrect elements in XRF profile 5
Will not connect using USB 6
Some USB ports will not connect 7
Instrument will not Cal Check (see ambient pressure) 7
XRF Kit/PC 8
Connection to Internet 8
Insufficient privilege for REFLEX User 8
Incorrect CRMs 8
Broken/damaged CRMs 9
Incorrect power adapter for XRF test bench 9
Docking station damage (screws) 9
PC configuration 10
1
Contents
FAQs
Shortcut configured to operate in "Demo" mode 10
Logmein functionality not operational 10
DHL not authorised to transport Lithium batteries 11
USB connection through Test stand will not connect 11
Default connection to REFLEX HUB is the wrong URL 12
REFLEX XRF Software 13
Elements Reported 13
CSV Export not correct 13
Data being written to incorrect sample 13
Data will not upload 13
Metadata configuration 13
Instrument trigger lock prevents reading 14
Software fails to start or is very slow in operation 15
Language/regional settings format 16
REFLEX HUB 17
Calibration 17
Data Search 17
Reporting 17
HUB Login 18
QA Reporting 18
2
Contents
FAQs
XRF Fundamentals
Introduction
Safety
When to use Soil Mode or Geochem Mode
Geochem Mode
• Mg, Al or Si are required
• Best performance for light elements (Mg, Al, Si, P, S, K)
• Elements of interest are above a few percent concentration
• Interest in both major elements and traces
• Contains mostly heavy elements
Soil Mode
• Contains mostly light elements
• Elements of interest are below a few percent (traces)
• Elements of interest are heavier than K
• Low detection limits required for Ca, Cr, Mn
Sample Preparation
3
FAQs
Instrument
Configuration
The instrument should be configured to the settings outlined in the
document 2.7 REFLEX XRF REFLEX Settings - Instrument
SOP_1.2.docx supplied with the instrument.
Under normal circumstances, the instrument should be preconfigured
to these settings, however they may have been changed on site.
The output from the instrument is unpredictable if the settings have
been modified.
Ambient Pressure
The instrument is configured to operate within a specific range of
barometric pressure. This can be adjusted on the instrument by
adjusting the barometer settings in the Safety/Hardware screen in the
instrument software.
S/W version
The minimum Delta Software version required for operation of
REFLEX XRF is 2.5.22.205. The version number is displayed in the
REFLEX XRF Help About screen
F/W version
4
FAQs
Instrument (cont)
Filter wheel failure
This is a hardware error caused by the failure of the filter wheel to
move or the sensor to detect the correct position.
There has been one issue where the filter wheel failed to move
smoothly because dust was allowed to enter through a perforation of
the instrument measurement window.
Code 923, Code 1000 & MCA Error
This indicates an empty spectrum. It basically means the system is
not recognising any X-RAYs. Most likely cause for this is a faulty
detector but totally dud X-ray tube could cause this as well (if the
issue is tube related you are likely to get tube voltage and/or current
errors as well).
Incorrect elements in XRF profile
The REFLEX XRF instruments have been configured for a specific
suite of elements. A subsequent change to the instrument profiles
introduced Ta to the instrument suite, and this initially caused a
problem with the REFLEX XRF software.
The software has been reconfigured to accept a dynamic list of
elements from the instrument and this should no longer present a
problem for the software to report.
5
FAQs
Instrument (cont)
Incorrect elements in XRF profile (cont) However, care must be taken
with additional elements in the analysis suite due to the spectral
deconvolution into element concentrations. Some elements will interfere
with the spectra of other elements, and this needs to be considered
carefully.
Will not connect using USB
• Windows Mobile Device Centre – the PC being connected to the
instrument must have the appropriate version (32-bit or 64-bit) of
WMDC installed.
• USB Cable – check connection with another cable. Suggest using the
gold plated cables supplied with the REFLEX XRF instrument
• USP Port on PC – use another USB port on the PC. Use USB2.0
ports in preference to USB3.0 ports.
• USB Port on Instrument/Test Stand – change from the test stand to
the side of the instrument or vice versa. The connection on the side
of the instrument has less connections than the back of the test stand
and in some cases this connection is more robust.
• USB Drivers on PC – ensure that the PC has the latest drivers for the
USB hardware on the PC. The hardware manufacturer generally has
more robust drivers than the generic Microsoft drivers installed
• USB Selective Suspend Settings – Windows 7 and 8 have a USB
selective suspend setting that may disable USB communication after
a period of time. This should be disabled for use with REFLEX XRF.
6
FAQs
Instrument (cont)
Some USB ports will not connect
The reason for this is unclear, but there are some PC USB ports that will
connect to WMDC (and the REFLEX XRF instrument) better and in a
more robust manner than others. It may be the USB drivers or the power
supplied to the port.
Instrument will not Cal Check (see ambient pressure)
In the situation where the ambient barometric pressure is outside the
range allowed for in the instrument configuration, the instrument will not
provide an error until it is required to run a Cal Check (or the first
reading).
Refer to the section on Ambient Pressure above to provide advice on
setting the barometric pressure to local conditions.
7
FAQs
XRF Kit/PC
Connection to Internet
The REFLEX XRF PC is configured to accept either a wired internet
connection or a wireless connection.
Should the wireless connection be used, the user must provide a valid
network and credentials to connect, however the connection must be
enabled using the REFLEX administrator account.
Refer to REFLEX support for assistance.
Insufficient privilege for REFLEX User
REFLEX User is a limited account, and can be used for data acquisition,
and transfer of data to REFLEX HUB once an internet connection is
established.
Should anything be required that needs administrator privilege, contact
REFLEX Support for assistance.
Incorrect CRMs
Incorrect CRMs used during analysis can be changed using the REFLEX
XRF Sample Inspector. If they are changed prior to upload to REFLEX
HUB, the updated information will be transferred to the REFLEX HUB,
however if it is changed after upload, it will be correct locally only.
If required, contact REFLEX Support for advice on data in the REFLEX
HUB.
8
FAQs
XRF Kit/PC (cont)
Broken/damaged CRMs
Broken or damaged CRMs are unsuitable for use. If individual CRMs are
damaged, the remainder can be used. The damaged CRMs should be
discarded and replacement sought from REFLEX.
Incorrect power adapter for XRF test bench
The REFLEX XRF Test Bench can be supplied with either a 9VDC or
18VDC power requirement. Using the incorrect power supply can cause
damage to the equipment.
Each Test Bench and Power Supply is labelled with the require/supplied
voltage. Ensure that the correct supply is used.
It has been suggested that the plugs should be colour coded, but this is
yet to occur.
Note also that the power supply for the REFLEX XRF Docking Station is
18VDC but is not to be used to power the Test Stand.
Docking station damage (screws)
A number of the REFLEX XRF Docking Stations as supplied have weak
plastic screw sockets holding the base to the cover. Many of these have
been replaced (under warranty), and should any damage/breakage occur
to the Docking Station, contact REFLEX for a replacement.
9
FAQs
XRF Kit/PC (cont)
PC configuration
Shortcut configured to operate in "Demo" mode
The early versions of REFLEX XRF software required a switch to be
appended to the executable file to operate in data capture mode. This
situation manifests itself by immediate connection to an instrument, and
rapid acquisition of Fe and Cu data only.
This has been corrected for V2.0 (and future releases), and does not
require this switch.
Should this occur, contact REFLEX for instructions.
Logmein functionality not operational
Logmein is the communication software used by REFLEX Support to
assist with troubleshooting our rental fleet. This should be installed and
activated during the PC imaging process.
If this is not enabled, contact REFLEX Support for advice on enabling or
for replacement of the PC.
10
FAQs
XRF Kit/PC (cont)
DHL not authorised to transport Lithium batteries
Some transport companies are not legally allowed to transport Lithium
batteries, the batteries used by the REFLEX XRF instruments.
Ensure prior to transportation that the transport operator is licensed to
carry the batteries, and that the packaging requirements have been met.
In all cases, the REFLEX XRF DELTA instrument MUST be transported in
the supplied Pelican Case to minimise the possibility of damage.
Also note that during air transportation; only one of the supplied batteries
can accompany the REFLEX XRF DELTA instrument in the Pelican Case.
The other must be hand carried.
USB connection through Test stand will not connect
The USB connection of the PC to the Test Stand and ultimately to the
REFLEX XRF DELTA instrument has more connections than a direct
connection to the instrument itself. Because of this, there are some
situations where the USB connection via the test stand fails.
In this situation, re-seat all connections, and if the connection failure
continues, try connecting directly to the side of the REFLEX XRF DELTA
instrument.
Refer to the section on “Will not connect using USB” should this continue.
11
FAQs
XRF Kit/PC (cont)
Default connection to REFLEX HUB is the wrong URL
The URL configured to connect to REFLEX HUB is:
https://go.reflexhub.com
Some PCs have been configured with the URL:
https://go.reflexhub.com.au which will not connect. This can be
modified in the settings of Google Chrome.
In the event that this has occurred, and immediate connection is required,
edit the address bar by removing the “.au” and press the <Enter> key to
refresh the URL in the browser. Contact REFLEX Support for information
of a permanent resolution.
12
FAQs
REFLEX XRF Software
Elements Reported
In previous versions of the REFLEX XRF software, the list of elements to
be reported was fixed in the software. From version 1.1.5 this list is
dynamic and the software will report all elements (apart from Ta) that the
instrument analyses.
CSV Export not correct
The issue reported that the elements reported does not match the headings
of the CSV export should not occur from version 1.1.5. If it does, it should
be reported immediately to Reflex Support for advice.
Data being written to incorrect sample
Data will not upload
Metadata configuration
13
FAQs
REFLEX XRF Software (cont)
Instrument trigger lock prevents reading
There is a known bug in the REFLEX XRF system that cannot control the
instrument trigger lock.
This is indicated by connection to the instrument, running a Cal Check, and
then getting an error during the running of the next sample that the trigger
is locked.
This is most easily worked around by closing the REFLEX XRF software
and restarting the REFLEX XRF DELTA instrument and the REFLEX XRF
software. When the instrument is restarted, connect to the instrument as
quickly as possible, and commence analysis.
Once analysis has started, the trigger will not lock while the REFLEX XRF
software remains connected to the instrument.
14
FAQs
REFLEX XRF Software (cont)
REFLEX XRF Software fails to start or is very slow in operation
When the database of the REFLEX XRF becomes very large (i.e. more
than 2GB), the software may become sluggish in operation or in extreme
cases, fails to start.
In this case, it is necessary to create a new database for the software using
the flowing procedure:
• Upload to Hub or export to CSV any required samples
• Close REFLEX XRF
• Navigate to the C:\ReflexXRF\V2.2.1 directory
• Export any custom Calibration or QAQC CRM files
• Rename the file ReflexXrf.db to ReflexXrf-YYYYMMDD.db (where
YYYYMMDD is the current Year-Month-Day) or something similar
• Restart REFLEX XRF. This will create a new database.
• Add the required projects and import the custom Calibration or QAQC
CRM files
• REFLEX XRF is now ready for use
15
FAQs
REFLEX XRF Software (cont)
REFLEX XRF Language/Regional Settings Format
REFLEX XRF will not operate correctly if using regional settings format and
language other than English.
The known issues are:
• Software crash with “Runtime Error” when the Format is set to French.
The runtime error is in the Microsoft Visual C++ Runtime Library.
• Incorrect numbers displayed when Format is set to Spanish or
Portuguese (this uses the comma as the decimal separator). An
example is that the analysis is displayed as 3414421689913580032
ppm instead of 341442 ppm.
16
FAQs
REFLEX HUB
Calibration
This has not yet been implemented in REFLEX HUB. Please contact
REFLEX Support for assistance should this be required.
Data Search
When logged into REFLEX HUB, click the Search option. Select the XRF
Assays view. A form will be presented to add selection criteria. Add the
required criteria and click the Search button.
Reporting
After the Search has completed, the data will be present on the screen.
Click the icon above the data to export the data as Word, Excel or PDF
document.
17
FAQs
REFLEX HUB (cont)
HUB Login
The REFLEX HUB login credentials will have been supplied by the HUB
Support team. Please note that the login to the REFLEX XRF for data
upload is case sensitive for both the username and password.
Also note that for the REFLEX XRF data upload, the URL should be:
https://api.reflexhub.com/XrfSample. If the URL is https://stg-
api.reflexhub.com/XrfSample, contact REFLEX Support for assistance.
QA Reporting
This is not yet implemented.
18
REFLEX Asia Pacific
T +61 (0) 8 9445 4020
REFLEX North America
T +1 (705) 235 2169
REFLEX South America
T +56 (2) 2589 9300
REFLEX South Africa
T +27 (0) 11 905 5595
REFLEX Europe
T +44 (0) 1273 405 975
Please contact your local
representative for further information:
www.reflexnow.com
April 2015