35
1 VARIAN Medical Systems Quality Assurance for IGRT Roman Wicha, PhD Varian Medical Systems Stanford IGRT Course 2008 VARIAN Medical Systems IGRT QA Agenda VMS IGRT Solution IGRT QA Tests QA Tools Examples of QA Protocols Summary

QA For IGRT

  • Upload
    neelu

  • View
    99

  • Download
    6

Embed Size (px)

DESCRIPTION

QA

Citation preview

  • 1VARIAN Medical Systems

    Quality Assurance for IGRT

    Roman Wicha, PhDVarian Medical Systems

    Stanford IGRT Course 2008

    VARIAN Medical Systems

    IGRT QA

    Agenda VMS IGRT Solution

    IGRT QA Tests

    QA Tools

    Examples of QA Protocols

    Summary

  • 2VARIAN Medical Systems

    VMS IGRT Solution -On-Board ImagerTM (OBI)

    X-ray Source- Varian G242

    aSi Imaging Panel- Varian PaxScan 4030CB

    Robotic Arms- 3 pivot points:

    Shoulder, elbow and wrist

    - Completely retractable

    - Position feedback control

    Software- Custom software for image

    acquisition and registration

    VARIAN Medical Systems

    VMS IGRT Solution

    On Board Imager Radiographic mode

    Fluoroscopic mode

    Interaction with Respiratory Gating (RPM) System available

    Cone Beam CT (CBCT) Acuity

    Hi-Energy Linear Accelerator

  • 3VARIAN Medical Systems

    Radiographic Mode

    VARIAN Medical Systems

    1. Setup patient

    2. Extend arms

    3. Exit treatment room

    4. Remotely acquire AP and lateral kV images

    5. Compare current images to reference images (DRRs)

    Automated

    Manual

    6. Calculate couch motion required to correct setup

    7. Move couch remotely

    8. Reacquire and reanalyze to confirm setup correction

    9. Retract arms remotely

    10. Treat

    Radiographic Mode: Anatomy

  • 4VARIAN Medical Systems

    Sample images

    Images courtesy of Karolinska Institute

    VARIAN Medical Systems

    Kilovoltage Radiographs

  • 5VARIAN Medical Systems

    Cone-Beam CT

    VARIAN Medical Systems

    1. Setup patient

    2. Extend arms

    3. Exit treatment room

    4. Acquire CBCT Dataset

    5. Compare current images to reference CT images

    Manual

    Automated

    6. Calculate couch motion required to correct setup

    7. Move couch remotely

    8. Reacquire and reanalyze to confirm setup correction?

    9. Retract arms remotely

    10. Treat

    CBCT Mode

  • 6VARIAN Medical Systems

    Clinical Results Hirslanden Klinik, Aarau

    CT

    CBCT

    VARIAN Medical Systems

    CBCT

    Clinical Results Emory

    CBCT CT Simulator

  • 7VARIAN Medical Systems

    Intrafraction motion

    VARIAN Medical Systems

    Pretreatment Fluoroscopic Verification of Gating

    Setup patient

    Extend arms

    Rotate gantry to first field

    Exit treatment room

    Remotely acquire fluoro

    MLC aperture turns green/red, indicating when the gate is open/closed

    Verify target remains within aperture

    Repeat for other fields

    Retract arms remotely

    Treat

  • 8VARIAN Medical Systems

    IGRT QA

    Number of IGRT installations is increasing

    There are limited guidelines for QA available as of now Yoo et all., A quality assurance program for the on-

    board imager, Medical Physics, Vol. 33, No. 11, November 2006

    IGRT QA should contain All the tests done for 3D CRT and IMRT

    Additional test related to image guidance

    VARIAN Medical Systems

    IGRT QA

    Primarily a machine QA

    There are no patient specific tests for IGRT (on top of the ones performed for IMRT)

    Parameters to be tested Safety

    Mechanical and optical

    X-Ray beam

    Imaging

    Functionality

  • 9VARIAN Medical Systems

    IGRT QA Daily Checks

    Anti-collision system

    Warning lights and audio warning for kV beam, interlocks

    Motion enable from OBI control console and hand pendants

    kV source / kV detector / MV detector positioning accuracy

    VARIAN Medical Systems

    Daily checks Anti-collison

    Detector Front

    Side

    Arm

    Source Front

    Side

    Back

    Arm

    Detector Front

    Side

    MV detector arm kV source arm kV detector arm

  • 10

    VARIAN Medical Systems

    IGRT QA Safety

    Daily checks kV source door interlock

    Warning lights and audio warning for kV beam

    Motion enable from OBI control console

    VARIAN Medical Systems

    Motion Enable

    Detector and arm motion is only possible when motion enable bars on the hand pendant or motion enable button on the OBI console are pressed

  • 11

    VARIAN Medical Systems

    Position calibration Daily Option 1

    Vertical axis calibration Position arms to the predefined position

    (e.g. kVS 100cm, kVD 50 cm, MVD 50 cm)

    Measure distance from detector cover / kV source plate to the isocenter

    VARIAN Medical Systems

    Position calibration Daily Option 2

    Longitudinal and lateral axis calibration Position kV Blade Calibration Plate on the table

    Move table to the isocenter

    Align kV Blade Calibration Plate with Clinac crosshair

    Acquire an image

    Compare position of digital graticule with the center cross of the plate

  • 12

    VARIAN Medical Systems

    Position Calibration Daily - Option 2

    Tolerance displacement < 2 mm

    Procedure Set Gantry 90o, measure the Lat displacement of

    KVD using ceiling laser and measure the Vertical displacement using cross-hair

    Set Gantry 270o, measure the Lat displacement of KVS using ceiling laser and measure the Vertical displacement using cross-hair

    VARIAN Medical Systems

    Position calibration Daily Option 3

    The Cube Isocenter Phantom was used for this daily QA procedure Two sets of 1-mm lead shots (bbs) were used to define 2 separate

    spatial locations (points)

    In each set, one bb is on the anterior side (AP bb) and the other (Lat bb) is on the lateral side of the phantom

    The first set defines the isocenter and the second set defines a point 1.5 cm anterior, inferior, and lateral from the isocenter

    Adapted from Rabih Hammoud

  • 13

    VARIAN Medical Systems

    The OBI daily QA procedure starts with setting up the Cube phantom center (central bb set) at the mechanical isocenter of the linac by using the cross hairs projections anteriroly and laterally

    Position calibration Daily Option 3

    Adapted from Rabih Hammoud

    VARIAN Medical Systems

    Position calibration Daily Option 3

    Lat DRR Lat KV Image

    Using the plan created from Eclipse, AP MV and Lat KV images were acquired (shown only Lat image)

    Adapted from Rabih Hammoud

  • 14

    VARIAN Medical Systems

    Lat DRR AP DRRKV Lat Image MV AP Image

    The alignment of the central bbs with the digital graticule (representing the imager isocenter) will be checked from both AP and Lat images

    Position calibration Daily Option 3

    Adapted from Rabih Hammoud

    VARIAN Medical Systems

    Position calibration Daily Option 3

    Long: __ mm OK 1.5 mm Long: __ mm OK 1.5 mm

    Lat: __ mm OK 1.5 mm Vert: __ mm OK 1.5 mm

    Adapted from Rabih Hammoud

  • 15

    VARIAN Medical Systems

    Results

    MV Isocenter Check

    00.20.40.60.8

    11.21.41.6

    1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31Daily measurement (days)

    Dai

    ly m

    ism

    atch

    di

    stan

    ce (m

    m)

    S/IR/L

    Courtesy of Rabih Hammoud, presented at AAPM

    VARIAN Medical Systems

    Results

    KV Isocenter Check

    00.20.40.60.8

    11.21.4

    1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31Daily Measurement (days)

    Dai

    ly m

    ism

    atch

    dist

    ance

    (mm

    )

    S/IA/P

    Courtesy of Rabih Hammoud, presented at AAPM

  • 16

    VARIAN Medical Systems

    IGRT QA Monthly Checks

    Safety

    OBI laser crosshair

    Arms isocentric calibration

    Arm run out

    Blades calibration

    Couch motion Linear

    Rotation

    VARIAN Medical Systems

    IGRT QA Safety

    Emergency off (part of Clinac QA)

    Backup motion controller

  • 17

    VARIAN Medical Systems

    OBI Laser Crosshair

    Using Clinac lasers position Isocenter cube to the isocenter

    With gantry at 0, position kV source arm to 100 cm

    Switch on OBI laser and verify its projection on the cube

    Repeat for gantry angles 90, 270

    VARIAN Medical Systems

    Arms Isocenter Calibration

    Position kV arm to 0,0,0 and verify that the center of the detector is aligned with the isocenter of the machine

    Repeat for MV arm

  • 18

    VARIAN Medical Systems

    Arm run out - I

    Position arm to 0,0,20

    Verify that the crosshair or OBI laser projection is aligned with the center of the detector

    Move detector along the vertical axis and verify the run out throughout the whole range of the motion

    VARIAN Medical Systems

    Arm run out - II

    Position kV Blade Calibration Plate on the table

    Move table to the isocenter

    Align kV Blade Calibration Plate with Clinac crosshair

    Position arm to 3-4 different positions along the vertical axis (e.g. 20, 40, 50, 70)

    Acquire image at each position

    Compare position of digital graticule with the center cross on the Calibration Plate

  • 19

    VARIAN Medical Systems

    Blades calibration

    Done using fluoro mode

    Blade calibration tool placed at isocenter

    Set up blades to given size

    Verify blades alignment with the calibration tool

    VARIAN Medical Systems

    IGRT QA Monthly Checks

    Isocenter alignment of MV and kV beam

    MV Image Quality

    kV Image quality

    Shift calculation

    Measurement tools

  • 20

    VARIAN Medical Systems

    Isocenter Accuracy

    Tolerance Coincidence of treatment isocenter and

    imaging center < 1.5 mm

    Preparation Phantom with a center marker

    QA plan with a setup field (AP or Lat) and DRR

    Courtesy of Dr. Sua Yoo, Duke University

    VARIAN Medical Systems

    Procedure Set the phantom center at the isocenter based on calibrated

    lasers, crosshair, and SSD

    Acquire a kV image in OBI

    Measure the displacement of the center marker from the center of the graticule in OBI

    Blade Calibration PlateCube phantom

    Marker phantom

    Isocenter Accuracy

    Courtesy of Dr. Sua Yoo, Duke University

  • 21

    VARIAN Medical Systems

    Tolerance Accuracy isocenter over gantry rotation < 2 mm

    Preparation Volume phantom with a center marker QA plan with setup fields of various gantry/source

    angles 0o, 90o, 180o, and 270o

    Isocenter over Gantry Rotation

    Modified from Dr. Sua Yoo, Duke University

    VARIAN Medical Systems

    IsocenterIsocenter accuracyaccuracy(May 2005 (May 2005 Apr 2006)Apr 2006)

    : Error in phantom setup Re-QAX: Laser misalignment Laser adjusted

    : Error in KVD calibration KVD re-calibrated

    IsocenterIsocenter over gantry rotationover gantry rotation (Dec, 05 (Dec, 05 July, 06)July, 06)

    G270 S

    I

    S

    I

    S

    I

    S

    I

    L R P A R L A P

    G0 G90 G180

    -0.2

    -0.1

    0.0

    0.1

    0.2

    1 2 3

    Dis

    plac

    emen

    t (cm

    )

    Mechanical accuracyMechanical accuracy(May 2005 (May 2005 Apr 2006)Apr 2006)

    Ver Long Lat

    KVS

    KVDKVS

    KVD KVSKVD

    Courtesy of Dr. Sua Yoo, Duke University

  • 22

    VARIAN Medical Systems

    MV Image Quality

    Set up detector

    Place phantom on top of the detector

    Acquire and analyze image (use Quality setting in the sequence template) max R determines contrast

    max C determines resolution

    VARIAN Medical Systems

    IGRT QA Imaging chain

    Weekly checks Image quality with imaging phantom

    Monthly checks High contrast resolution

    Gray scale linearity

    Low contrast sensitivity

    Image geometry Distance

    Angle

  • 23

    VARIAN Medical Systems

    Low contrast resolution Tolerance: > 11 12 disksFluoro: 70 kVp, 32 mA, 6ms 11 13 disksRadio: 75 kVp, 25 mA, 6ms 13 15 disks Spatial resolution Tolerance: > 11th groupFluoro: 50-80kVp, 80mA, 32ms 9 11th groupRadio: 50-80kVp, 80mA, 120ms 10 12th group

    Fluoroscopic Radiographic

    Results Image Quality

    Courtesy of Dr. Sua Yoo, Duke University

    VARIAN Medical Systems

    IGRT QA OBI Shift calculation

    OBI Geometric Phantom Crosshair #1 reference

    Crosshair #2 test

    Crosshairs shifted by the known distance

    4 markers inside phantom

    Reference image for setup fields at reference crosshair

    kV/kV, MV/kV and Marker Match for fields setup at test crosshair

    Verify distance calculated by OBI algorithm

  • 24

    VARIAN Medical Systems

    Tolerance Accuracy of 2D2D match and couch shift <

    2 mm Preparation

    Phantom1. Phantom with off-center markers at

    known positions, or 2. phantom with a center marker placed

    at known position. QA plan with orthogonal setup fields and

    DRR

    Phantom Setup at Iso

    kV images (AP/Lat)

    2D2D match

    Apply couch shift

    Remote shift couch

    Match DRRs graticuleto off centered marker

    Verify couch in room

    Before match After match Courtesy of Dr. Sua Yoo, Duke University

    Shift Calculation 2D / 2D

    VARIAN Medical Systems

    2D2D match and couch shift2D2D match and couch shift(May 2005 (May 2005 Apr 2006)Apr 2006)

    -1

    0

    1

    2

    Dis

    agre

    emen

    t (m

    m)

    -2

    -1

    0

    11 2 3

    Diff

    eren

    ce (m

    m)

    2D2D match accuracy Couch shift accuracy

    Ver: Ant +/Post ; Long: Inf +/Sup ; Lat: Rt +/Lt

    Ver Long

    LatVer Long Lat

    Results

    Courtesy of Dr. Sua Yoo, Duke University

  • 25

    VARIAN Medical Systems

    Measurement tools

    Measurement of

    Distance

    Area

    Angle

    VARIAN Medical Systems

    IGRT QA Annual Checks

    Annual checks kV beam output parameters

    kV, mA, exposure time

    HVL

    Automatic Brightness Control (ABC) for fluoroscopy

  • 26

    VARIAN Medical Systems

    IGRT QA CBCT Tests

    Daily checks Geometry

    HU calibration spot check

    Annually (Monthly) Uniformity

    Contrast

    Resolution

    HU calibration

    VARIAN Medical Systems

    CBCT Daily Check

    Position CATPHAN on the table

    Acquire CBCT

    Verify geometry

    Verify HU

  • 27

    VARIAN Medical Systems

    CBCT Annual test

    Position Norm Phantom on the table

    Acquire CBCT

    Measure profile and verify uniformity of HU

    VARIAN Medical Systems

    Half fanFull fanROI

    Center

    Left

    Bottom

    Right

    Top

    15.6 11.318.0 9.0

    - 15.9 9.6- 8.5 7.8

    - 13.4 10.1- 7.3 8.4

    - 14.6 11.6- 3.4 8.9

    - 14.6 10.4- 4.8 8.7

    Full-fan Half-fan

    Results HU Uniformity

    Courtesy of Dr. Sua Yoo, Duke University

  • 28

    VARIAN Medical Systems

    Contrast

    Performance 0.5 %

    VARIAN Medical Systems

    Resolution

    Performance 10 lp/cm

  • 29

    VARIAN Medical Systems

    Half fanFull fanTolerance

    Spatial linearity

    Slice thickness

    Spatial resolution

    Low contrast resolution

    49.8 50.0 49.8 50.2 0.5 mm

    2.45 2.512.39 2.60 0.1 mm

    Group 6Group 8Group 6

    Disk 4Disk 5Disk 4

    Full-fan Half-fan

    Low

    con

    tras

    tre

    solu

    tion

    Spat

    ial

    reso

    lutio

    n

    Results CBCT Image Quality

    Courtesy of Dr. Sua Yoo, Duke University

    VARIAN Medical Systems

    HU Calibration

    1. Air =-1000

    2. LDPE = -100

    3. Acrylic = 120

    4. Teflon = 990

  • 30

    VARIAN Medical Systems

    Air

    PMP

    LDPE

    Poly-

    Acrylic

    Delin

    Teflon

    - 995.7 13.3- 998.3 2.4

    - 206.5 12.7- 196.6 8.6

    - 106.9 16.1- 101.2 8.8

    - 48.5 10.5- 42.8 8.8

    116.1 12.396.0 9.0

    354.0 15.1363.1 10.7

    1007.8 22.8989.0 10.5

    Half fanFull fan Full-fan Half-fan

    Results HU linearity

    Courtesy of Dr. Sua Yoo, Duke University

    VARIAN Medical Systems

    QA Tools delivered with OBI and CBCT

    On Board Imager Isocenter cube

    Blade calibration plate

    Leads TORR-18FG phantom

    OBI geometric phantom on Exact Bar

    RPM RPM Gating phantom

  • 31

    VARIAN Medical Systems

    On-Board Imager Phantoms

    RPM-Gating phantom:gated radiographs

    and pretreatment fluoroscopic verification of gating

    Leeds TORR-18FG Phantom: kV image quality

    OBI Geometric Phantom: overall accuracy

    and precision of the system

    kV Blade Calibration Plate: calibrate the kV blades

    Isocenter Cube:alignment of the kV to the

    MV beam

    VARIAN Medical Systems

    QA Tools delivered with OBI and CBCT

    Cone Beam CT NORM calibration phantoms

    CATPHAN

  • 32

    VARIAN Medical Systems

    Norm calibration

    Norm Phantom To correct for scatter

    2 Norm phantoms Head

    Body

    CATPHAN Geometry

    Resolution

    HU

    Contrast

    VARIAN Medical Systems

    Examples of QA Protocols Courtesy of Sua Yoo, PhD, Duke University

  • 33

    VARIAN Medical Systems

    Daily OBIDaily OBI Weekly OBIWeekly OBI

    VARIAN Medical Systems

    Monthly OBIMonthly OBI

    (continues)

  • 34

    VARIAN Medical Systems

    Monthly CBCTMonthly CBCT

    (continues)

    VARIAN Medical Systems

    Automated IGRT QA Works in Progress

    IGRT QA is straightforward, but may require some time to acquire and analyze data

    It maybe possible to automate some of the QA steps Weihua Mao, Louis Lee, and Lei Xing -

    Development of a QA phantom and automated analysis tool for geometric quality assurance of on-board MV and kV x-ray imaging systems, submitted to Med. Phys.

  • 35

    VARIAN Medical Systems

    Automated IGRT QA Works in Progress

    Courtesy Lei Xing, Stanford

    VARIAN Medical Systems

    Summary

    OBI and CBCT are increasingly used in the clinical routine

    Development of commissioning and QA procedures is needed to accomplish safe delivery of IGRT

    QA should cover Safety

    Mechanical and optical

    Imaging

    Functional tests